Professional Documents
Culture Documents
The detection accuracy of the diagnostic radiologist is combinations (Alpha 4/RP, Hi Plus/RP, Alpha 4/XD, and
important in everyday medical decision making. However, Alpha 4/XM) and three focal spots (0.3, 1 .0, and 2.0 mm
little work has been done relating the detection accuracy of nominal sizes). Though rare earth screen-film systems
the radiologist to the quality of the image. This study, using a are not conventionally used for chest radiography, the
thorax and lung phantom, simulated tissue-equivalent 6.4 mm
potential for patient dose reduction is significant. How-
lesions, and a 183 cm source-to-image distance, shows that
ever, it must be assured that the detection accuracy of
the detection accuracy is not dependent on the focal spot size
(over a range of 0.3-2.0 mm). However, the false positive rate the radiologist is not impaired by the noisier, lower dose
increases when using small focal spots. In addition, the image’s.
detection accuracy decreases with increasing root-mean- The major variables studied were image quality (mod-
square (RMS) noise (a measure of the amount of quantum ulation transfer function, MTF) and noise (mottle) with
mottle in the image), while the false positive rate and intraob- differences in responses between radiologists and non-
server disagreement increase with increasing RMS noise. It is radiologists also being considered. Images of a thorax
also shown that the nonradiologist responds to changes in phantom with lungs and simulated spherical lesions
noise in exactly the same way as the radiologist. were produced. So that the radiologists would not be-
Introduction come familiar with the distributions, 14 lesion patterns
were used.
It is often assumed that any degradation in image quality,
in terms of noise or image fidelity (resolution or modula- Materials and Methods
tion transfer function [MTF]), will significantly decrease
Experimental Films
the detection accuracy of the radiologist. This is an
important assumption because of the added cost re- Films were produced using three focal spots (0.3, 1 .0, and
quired to produce imaging equipment with improved 2.0 mm) and four screen-film combinations (table 1). In all
instances the modulation transfer function (fig. 1) was deter-
resolution and because of the additional dose required
mined by means of edge-gradient analysis [8]. Only one MTF is
to produce low noise radiographs. Feddema and Botden
shown for Alpha 4 (fig. 1A) since the MTF for all three combina-
[1] concluded that the radiologist can detect the perti-
tions using that screen were identical. The effect of phase was
nent pathology on images of extremely low resolution disregarded since it was found to be negligible for the focal
but, given the choice, would normally select the most spots used [9].
aesthetically pleasing high resolution image. The Alpha 4/RP system is a mismatched system because the
Most studies associating image quality with diagnostic film is not spectrally sensitive to the green emission of the rare
accuracy have had serious limitations in the types of earth phosphors. It was used in the study so that three systems
simulated radiographs used . Several authors [2-5] lim- with different noise levels and the same MTF could be studied.
ited their studies to relatively simple stimuli, such as The radiographs for this study (fig. 2) were produced on 28 x
uniformly exposed radiographs containing circular Ic- 35.5 cm film using a thorax phantom (3M Co.) consisting of
human bone encased in Plexiglas. The phantom lungs were dog
sions. Kundel and Revesz [6] studied this problem by
lungs inflated with formalin fumes while the vascular system
superimposing simulated lesions on conventional radio- was perfused with latex. The dried, preserved lungs simulated
graphs using a video system. This approach is limited by the vasculature of the human lung. No attempt was made to
the restricted response of video systems. More recently, simulate the mediastinal, diaphragmatic, and abdominal re-
Brogdon et al. [7] used radiographs produced by super- gions.
imposing normal radiographs and radiographs of simu- The lesions consisted of Lucite spheres 6.4 mm in diameter
lated pathology. However, the image quality of the dupli- (fig. 3). Several radiologists were shown chest radiographs
cate of the superimposed films was not established. containing 6.4 and 4.8 mm lesions in identical locations. They
Studies using clinical radiographs are usually not con- detected only one or two of the nine smaller lesions, whereas
clusive due to variations in the patient and the position they detected six to eight of the larger lesions. Consequently,
the lesions of larger size were used.
of the lesion. In such studies it is difficult to determine if
All radiographs were produced with the geometry shown (fig.
changes in detectability are due to differences in image
4) using an anteroposterior projection. The phantom was posi-
quality or to lesion location. tioned identically for each radiograph, assuring that the lesions
In an attempt to overcome these problems and to and superimposed structures were in the same positions on the
determine the effect of image quality on detection accu- radiographs. The radiographs were made at 90 kVp using
racy, we carried out a study using four screen-film stationary grids (10:1 ratio, 40 lines/cm with a 183 cm source-
TABLE 1
Characteristics of Screen-Film Combinations
Root-
Rela-
Mean-
Screen-Film Screen tive
Film Type square
Combination Type Expo Noise
sure (x 10’)
Downloaded from www.ajronline.org by 103.119.54.48 on 10/27/19 from IP address 103.119.54.48. Copyright ARRS. For personal use only; all rights reserved
DuPont Hi PIus/
Kodak RP. ‘CaWO4 Blue-sensitive 0.90 0.56
3M Alpha 4:
With Kodak
RP Rare earth Blue-sensitive 1 .65 0.57
With 3M XD. Rare earth Green-sensitive 1 .00 0.61
With 3M XM. Rare earth Green-sensitive 0.30 0.73
A B
1.0
0.8
LI. 0.6
I-
0.4
0.2 High
Plus
-i I I I I I
F.,j. 2.-Radiograph of phantom with nine lesions. Obvious lesions,
0 2 4 6 8 10 0 2 4 6 8 10
deleted in analysis, are indicated by circles; remaining seven lesions are
Frequency Ic/mm) indicated by arrows.
TABLE 2
Summary of Subjects
5ubject 5ubject
ldentifica- Background ldentifica- Background
lion lion
1 Radiologist 8 . . Radiologist
2 Radiologist 9 Resident (fourth year)
3 Resident (third year) A Secretary
4 Radiologist B Graduate student
5 Radiologist C Medical physicist
6 Resident (second year) D Radiologic
7 Resident (first year) technologist
for deletion from the focal spot series were located correctly in
eight or nine of a possible nine responses and were clearly
visible on the films made with the 2.0 mm focal spot. Of the 126
lesions, 50 were deleted; of these 50, seven were lesions on
bone, 25 were lesions oft bone, and 18 were lesions on bone
edges.
To study the inconsistency, or intraobserver disagreement, of
a given observer, two identical radiographs were made with
each screen-film-focal spot combination. (All lesions, including
those deleted for the detection accuracy analysis, were included
in the intraobserver disagreement analysis). For each lesion that
the participant saw twice in the same session on films made with
the same focal spot and screen-film combination, there were
three possible responses: (1) the subject did not locate the
lesion correctly either time; (2) the subject located the lesion
Fig. 4.-Geometry for x-ray imaging system. correctly both times; and (3) the subject located the lesion
correctly only once. The ratio of the number of the responses
locating the lesion correctly only once to the total number of
lesion pairs is the intraobserver disagreement. The relative
the overall detection accuracy. The detection accuracy was intraobserver disagreement is then found by dividing the mdi-
then defined as: vidual’s intraobserver disagreement for a specific imaging con-
dition by his average disagreement, thereby reducing the varia-
A_C -O (1) tion due to differences between sessions and observers.
-
Results
0.8 3 3
3
Radiologists as Group
0.7
An attempt was made to determine the effect of 8
Fig. 5.-Detection accuracy
changes in image quality on the detection accuracy of
as function of nominal focal spot <
the radiologists using the responses from one session size for radiologists 1-9 (table 2). 2 99
for each of nine radiologists (fig. 5). Though trends may Averages for nine radiologists
Downloaded from www.ajronline.org by 103.119.54.48 on 10/27/19 from IP address 103.119.54.48. Copyright ARRS. For personal use only; all rights reserved
areindicatedbydots. 0.4 6 7
be apparent, it is not possible to determine from figure 5
if the changes in detection accuracy were significant, 4
0.3
due to the wide spread in the overall accuracies of the
0.3 1.0 2.0
individual radiologists. The wide range of individual Focal Spot Sin 1mm)
accuracies has not been considered in many previous
studies. It is clear that changes in detection accuracy
must be considered for each radiologist in order to
when reading films with a large amount of noise (fig.
determine whether some improve and some get worse
7C).
with changes in image quality.
Nonradiologists
Radiologists as Individuals
Four nonradiologists viewed the radiographs three
Five radiologists were asked to read the films several
times each and followed the same procedures as the
times. Only the change in the criterion level of the
radiologists (fig. 8). The nonradiologists responded to
individual radiologist and a possible learning effect
changes in image quality in a manner similar to radiolo-
could produce changes in the detection accuracy from
gists. The findings imply that in the future it should be
session to session. The results of the multiple sessions
possible to use nonradiologists in such studies if the
for the five radiologists evaluated are shown in figure 6.
task is well defined and if it involves only the detection of
Although their mean accuracies vary from 0.45 to 0.85,
a limited, specified pathologic condition.
all radiologists respond in a similar fashion to changes
in image quality. The detection accuracy of the individual Changes in Detection Accuracy
radiologist is not affected significantly by the change in
There were large differences between the detection
focal spot size for the imaging geometry used. The false
accuracies of individual radiologists (figs. 5-7). Since
positive rate tends to decrease with increasing focal spot
these variations are known to be present in all types of
size (fig. 7A).
responses from human observers, the foregoing data
Since the radiographs made with the smallest focal
can be analyzed by eliminating the individual differ-
spot tend to reproduce the fine detail in the phantom
ences. This can be done by looking at the changes in
clearly (i .e. , the amount of structured noise is increased),
detection accuracy of the radiologists as a group.
it would be expected that the radiologist might falsely
If A1 is the detection accuracy for a particular film
indicate the presence of a lesion in the noisy back-
quality and A2 is the accuracy for films of a different
ground. In addition, the radiologist would be expected
quality, the change in detection accuracy is given by the
to have a more difficult task in reading the radiographs
following formula:
made with the small focal spot because he is not accus-
tomed to viewing chest radiographs with considerable
fine detail. .A- ‘ 2 2
- (A1 + A2)/2
Since it has been shown that the root-mean-square
(RMS) noise will interfere with detection [10], we used This result is merely the change in the detection accu-
the RMS noise as a relatively simple way to characterize racy normalized to the average accuracy. The results are
fluctuations in the density of a radiograph. The RMS shown in figure 9, where the lowest noise and the
noise was determined by uniformly exposing a radio- smallest focal spot image are assigned a relative detec-
graph and measuring the density at 750 independent tion accuracy of unity. It is apparent that the variation
locations on a film (with fixed aperture of 1 mm). The due to differences between individual radiologists has
average and the square of the differences (deviations) been reduced considerably (cf. fig. 5). Increasing the
between each measurement were determined. The RMS focal spot size from 0.3 to 2.0 mm had no effect on the
noise is then the square root of the mean of the squared detection accuracy (figs. 9A and The detection
9B).
deviations or the standard deviation of the density mea- accuracy decreased for increasing RMS noise (figs. 9C
surements. The RMS noise has been discussed in detail and 9D).
in the literature [11] and recently applied in the radio- A fourth data point, X (figs. 9C and 9D), was plotted
graphic field [12]. for the Hi PIus/RP screen-film combination. Although its
It is evident (figs. 6C and 6D) that the detection value for RMS noise is not significantly different from
accuracy of the individual radiologist decreases with that of the Alpha 4/RP combination, the relative detec-
increasing RMS noise. Most radiologists tend to report tion accuracy is significantly lower, due to the lower
more false positives when the noise is increased (fig. modulation transfer function (MTF) for the Hi Plus screen
7B). Also, the radiologist tends to disagree with himself compared to the Alpha 4 screen.
DETECTION ACCURACY IN CHEST RADIOGRAPHY 251
C A
A 0.91.
B
0.91- 3 14
34 ll
Radiologist #1 0.8 0.7L 12
0.71 1 0.6L 10
Downloaded from www.ajronline.org by 103.119.54.48 on 10/27/19 from IP address 103.119.54.48. Copyright ARRS. For personal use only; all rights reserved
IL. 8
Radiologist #3
6
3 I 4
0.61- o.4. 2
Radiol #4 J o-
O.4
#{188}%;:
1 0.41- * ‘, 0.3
Focal
1.0
Spot Size (mm)
2.0
C
I 03
0.7’- 3
0.7
1.4
1 1.2’
Radiologist #5 0.6 - I’’ fli.o.
0.51-
o.7L
1
2
0.41-
V 0.8
0.6-
Radiologist *6
0.6L 1
0.60 0.70 0.80
RMS Noise (x102)
0411 1 Fig. 7.-A, Number of false positive responses per reading session as
f &
0.3
,
1.0
I
C
C
A Radiologists 1.0
A 0.8
1.0 - 0.9
0.8 0.7
, 0.7 0.6
0.9 - SEM 0.8 Radiologists M
0.8 - 0.7
0.6 . 0.5 0 T I I I
Downloaded from www.ajronline.org by 103.119.54.48 on 10/27/19 from IP address 103.119.54.48. Copyright ARRS. For personal use only; all rights reserved
D
8 0.5,: I I I 0.4
0 B All Particpants 1.0 6
All Participants
6 6 0.8 . 0.7
T I I ‘L
::4 4 0.3 1.0 2.0 0.60 0.70 0.80
0
Focal Spot Size (mm) RMS Noise )xlO”2)
12 2
E Fig. 9.-Relative detection accuracy for radiologists and nonradiolo-
z3 i I C . .
gists as function of nominal focal spot size and root-mean-square (RMS)
0.3 1.0 2.0 E noise. One standard error of mean (SEM) is indicated.
Focal Spot Size mm)
t 1.4
H 3.5
I’1 0.60
I
0.70
I
0.80
3.0
2.5
by changes in the focal spot size, but the false positive sheet. And remember, the 72 radiographs must be read in less
rate decreases with increasing focal spot size; (4) the than 1 hour and 15 minutes.
detection accuracy decreases for increasing root-mean- Are there any questions?
square noise; and (5) the false positive rate and intraob-
REFERENCES
server disagreement increase with increasing root-mean-
square noise. 1 . Feddema J, Botden PJM: Adequate diagnostic information,
Downloaded from www.ajronline.org by 103.119.54.48 on 10/27/19 from IP address 103.119.54.48. Copyright ARRS. For personal use only; all rights reserved