You are on page 1of 3

seca 727/ 728

QuickMedical is an authorized distributor for seca.

See the complete line of seca products at :


www.medicalscales.com/seca

Call toll free: 1-888-345-4858


Fax: 425-831-6032

1
(9) of the baseplate. pressed again and awaited.
Preparation is important... Press lock (8) in the cover of the battery With accumulator operation, appears on
Transport locking device compartment towards the text “BATTERY” and the display if the accumulators have to be
Loosen transport locking device. First turn over open the cover. Then insert the plug (7) of the recharged. This happens automatically by
the scales so that the baseplate is at the top. set of accumulators (10) into socket (11). Insert connecting the plug-in power supply unit
Loosen the two outer clamping screws (1). Pull the set of accumulators in the battery provided.
out the spacers (3) between the frame (5) and compartment, close the cover again and 16
the cradle support (2) sideways and remove the engage the lock.
clamping bar (6) with the screws and spacers. Note: For weighing independently of the mains,
Note: Before transporting over long distances, the set of accumulators must be charged up
the scales must be locked again by re-inserting before initial use by plugging the power supply
the spacers and tightening the clamping screws unit into the scales for 24 hours. 4
hand tight. Therefore keep the transport locking Connect the plug-in power supply unit to the 14
device and packaging. mains. 13
8 12
6
Model 727 Model 728
5
1
7
2
If scale is switched on
weights up to 0,4 kg will
4 be tared, more than 0,4
-Tare kg will be stored by pushing
-Hold the ’Net/HoldButton‘.
Model 727 10 (blinking) -lbs:oz
4
9
17 16 15
3
6
5 11 Changeover kg/lbs
1
2 4 If you require the weight to be displayed in lbs,
press the kg/lbs changeover switch (14) after
Model 728 switching on or while weighing. The lbs indicator
(15) lights up for the lbs setting.
How to weigh correctly…
3
On scales with the Autohold function
Power supply Press the green start button (13). and then The Autohold function is intended to save the
The scales can be either mains or accumulator appears in sequence on the display (16). user pressing the Hold key.
operated. A power supply unit and a set of The scales are ready for operation. Zeroing It is active as soon as the scale is switched on.
accumulators are supplied; both parts are takes place automatically when the scales are If a load of over 400 g is placed on the scale, the
connected by the following procedure. You can empty. However, if a value other than weight is automatically held in the display.
then choose between mains and accumulator appears on the display, the start button (13) is to Pressing the Hold key releases the weight from
operation. be pressed for zeroing prior to weighing. The Hold.
empty scales switch off after 10 minutes when The Autohold function only becomes active
Mains and accumulator operation mains operated, and after one minute when again when a load of over 400 g is placed on the
Insert the low voltage plug of the plug-in power accumulator operated. Before using the scales scale. This procedure can be repeated as often
supply unit provided into the low voltage socket for further weighing, the start button must be as required.
5
Taring device The switch (A) is located at the end of the Tare range: 0 - 4 kg 0 - 14 oz
When the scales are switched on, there is cable in the battery compartment. The 2g Permanent display: 0.4 - 20 kg 14 oz - 44 lbs
provision for taring items such as cushions resolution works up to a weight of 10 kg. Then Height of mod. 727 151 mm 6"
and towels, which have been placed on the 5g resolution takes over. Height of mod. 728 163 mm 6 1/2"
scales before weighing, up to a weight of 400 Width 551 mm 21 5/8"
grammes (14 oz). Depth 320 mm 12 5/8"
1. Place cushion or towel on the scales – the What do I do if… Empty weight: 5 kg 11 lbs
weight is displayed. Press Tare/Hold
button (12) – appears and the … The display remains dark?
indicator for Tare/Hold (17) lights up. The ● Check the mains connection Power supply
actual weighing process can now take ● Check the connection of the low voltage
place. Power supply unit depending on model
plug (9). 230 V~/12 V=/130 mA
2. When the scales are completely empty, … No weight is displayed under load?
appears automatically and the (seca no. 68-32-10-252)
● Check whether the transport locking
indicator for Tare/Hold (17) goes out. device has been removed 115 V~/230 V-/9 V=/350 mA
… appears on the display? (seca no. 68-32-10-243)
Prolonging the weight display ● Press start button.
The scales offer the facility for retaining 115 V~/9 V=/350 mA
● If taring, press the tare button.
weights in excess of 400 grammes on the (seca no. 68-32-10-259)
… appears on the display?
display after the weighing process. ● The accumulators must be recharged. Accumulator: 6 unit accumulator set
1. Switch on scales with the start button (13). … appears on the display? (seca no. 68-22-12-721)
2. Place the baby on the scales – the weight is ● The scales are overloaded, reduce
displayed. IEC 601-1: class of protection II, type B
load.
3. Press the Tare/Hold button (12) and the
… appears on the display? We only provide a guarantee for the safety,
weight is stored. The indicator for Tare/ ● Check whether scales are overloaded.
Hold (17) flashes. reliability and performance of the equipment
● The automatic self-checking system has if:
4. If the baby is taken off the scales, the
weight continues to be displayed. found a fault. Inform seca service – the equipment is used in conformance with
5. The permanent display is cancelled by department. the operating instructions, and
pressing the Tare/Hold button (12) or the – the power supply unit provided (12V/130
start button (13) again. The indicator for Technical data mA, seca no. 68-32-10-252) or the 6 unit
Tare/Hold goes out. Weighing range: 0 - 20 kg 0 - 44 lbs accumulator set supplied is used, and
Fine graduation: 5 g 0.2 oz – repairs are undertaken by authorised
2g/5g resolution persons and
changeover switch In 2g graduation mode – the transport locking device and
To change over the 0 kg to 10 kg: 2g 0.1 oz packaging supplied are used for transport
resolution from 5g to 10 kg to 20 kg: 5g 0.2 oz and
2g, please open the
A – the scales are only cleaned with normal
battery compartment. disinfectants.
6