You are on page 1of 6

Preparation for Electron

Backscatter Diffraction with LaboPol-4 /


LaboForce-1
Danka Katrakova1, Morten J. representative specimen area ways that would depend on the
Damgaard2 and Frank Mücklich3 [92Ran], [93Schw], [93Wri], type of material as summarised
[98Kat]. The preparation is far in Fig. 1. Very often vibratory pol-
1
Continental AG, Hannover, simpler than that of thin foils for ishing with colloidal silica is the
Germany; previously at 3; TEM, but it is also much more best solution. The polishing time
katrakova@hotmail.com demanding than that for normal can be of about some hours.
2
Struers A/S, Kopenhagen, imaging in the SEM. This is Therefore polishing cloths should
Denmark mainly due to the very low infor- be chosen very carefully in order
3
Saarland University, Functional mation depth of the EBSD signal to avoid relief on the specimen
Materials, Saarbrücken, based on channelling effects. surface. If such equipment is not
Germany available normal rotation polish-
After standard preparation proce- ing with silica could also provide
Abstract dure - mechanical grinding and acceptable results. Some
The LaboPol-4/LaboForce-1 polishing - a thin residual surface etchant may be used for speed.
combination that allows low load distortion layer is left [82Sam], Chemical polishing and slight
and low rotation speed was used [99Pet], [99Van]. This layer is not etching are very fast and they
to prepare a large variety of sam- really disturbing when light or surely remove any residual distor-
ples for Electron Backscatter Dif- scanning electron microscopy is tion. But care should be taken to
fraction (EBSD). used, especially if it is removed avoid surface layers as well as
As the signal comes from inter- by etching. In the case of EBSD pitting and/or severe roughening.
action in a very thin surface layer, though this is often not sufficient Another very recommended
this method requires a very gen- for a good pattern quality. Some method is electropolishing.
tle preparation without any re- additional effort may be neces- Again, no surface layer should be
sidual distortion on the specimen sary. In general, classical etching left. Unfortunately electro-
surface. is not needed for EBSD because polishing is restricted to electri-
In this paper a preparation strat- the contrast is defined through cally conductive materials.
egy for EBSD is discussed and the orientation differences. It is Some physical methods are very
successful preparation proce- even avoided as the induced sur- useful for 'difficult' materials like
dures for 16 different materials face roughening is disturbing due ceramics, minerals or metals with
(metals, intermetallics and ce- to the high specimen tilt (70°). high oxygen affinity. Especially
ramics) are given. The results are oxidation layers should be
compared with alternative prepa- Preparation Strategy
ration routes like modified stand- for EBSD
ard polishing and vibratory pol- In order to obtain a flat Mechanical grinding and polishing
final step
ishing. and even distortion free
specimen surface the
Introduction standard grinding and Mechanical Chemical Physical
Electron Backscatter Diffraction polishing procedures treatment treatment treatment
(EBSD) has become very popular have to be adjusted.
among materials and earth sci- These modifications
entists since the first commercial mainly concern the final
electropolishing
colloidal silica
etch-polishing

automatic systems were avail- preparation stage.


annealing
polishing

ion beam
polishing

polishing
chemical
vibratory

cleaning

able in 1994. The main advan- As a general guideline


etching

plasma
slight

tage of this method is the possi- reduced loads and ro-


bility to link morphology (grain tation speed should be
size and shape) with used for mechanical
crystallographic features (phase, grinding and polishing.
orientation, disorientation) on the The final step can be Fig.1: Preparation strategies
microscopic scale but still in a performed in different for EBSD

19
avoided - therefore annealing of This is a disadvantage when Materials
metals for example should take compared to standard mechani- Different materials were prepared
place in inert gas atmosphere or cal polishing. Another drawback so that EBSP's of good quality
in vacuum. Beam energies if us- is the relatively poor polishing dy- were obtained. The test materials
ing ion beam or reactive plasma namics obtained when using pol- differ in hardness (which strongly
should be rather low to avoid ad- ishing speeds above 10 rpm. influences the preparation behav-
ditional damage. iour) and also in their EBSD infor-
Due to the large difference be- mation depth.
Experimental tween sample mover and disc The escape depth of the back-
rotation speed the sample is scattered electrons depends on
Equipment preferential polished in one direc- the atomic number and weight
Conventional mechanical rota- tion. When using soft polishing as well as on the density of the
tional grinding and polishing al- cloths this will lead to polishing material and on the energy of the
low 5 N force and 150 rpm rota- artefacts like comet tails [00Gee] electrons [93Joy]. Only little is
tion speed as a minimum. This It can be also quite inconvenient known about the exact informa-
equates to a relative velocity of if really a long time is needed for tion depth of EBSD. Since only
approximately 1 m/s in a 250 silica polishing, as it is not eco- backscattered electrons with low
mm system. With the LaboForce-1 nomical regarding abrasive sus- energy loss contribute to the pat-
the load can be varied between 0 pension and manpower. Com- tern formation (whereas the other
and 20 N in 2,5 N steps. The pared to vibratory polishing this build the diffuse pattern back-
LaboPol-4 allows gradual change is a disadvantage. But regarding ground) it is much lower. It de-
of the rotation speed between 0 time this gentle rotational polish- pends on the atomic number and
and 150 rpm; the specimen ing is still much faster than the weight, the density, also on the
holder rotates at 8 rpm, that al- vibratory method so that the acceleration voltage in the SEM.
lows relative velocities between 0 whole preparation and not only Some attempts have been made
and 0.88 m/s. Obviously by us- the final step can be completed to calculate the spatial resolution
ing this equipment a very gentle in reasonable time. Also wavi- (including depth) for some mate-
preparation is possible. But of ness, that is typical when vibra- rials via Monte Carlo simulation
course preparation time signifi- tory ground, is almost not ob- [98Ren]. Considering those re-
cantly increases. served. sults the information depth for

Fig. 2:
Test materials
differing in hardness
and EBSD
information depth;
R(BS) corresponds
to the escape depth
of the backscattered
electrons at 20 kV;
R(EBSD) is about
25 % of R(BS).
EBSP of some
materials are
included with their
PQI

20
Sample Grinding Final
polishing polishing

1 standard standard

2 standard new
38
Preparation for Electron
Backscatter Diffraction
with LaboPol-4 / LaboForce-1
3 standard vibratory

4 new new
the EBSD signal is ap- stage to the new machine and
proximately a fourth of 5 vibratory vibratory almost always on the vibratory
the escape depth of the one.
backscattered elec- When the new device was used
trons in general for a cut-off en- for the whole preparation all met-
ergy of 90 % that of the primary Results and Discussion als could be prepared so that
electrons. The quality of the preparation EBSP of excellent quality were
was judged by using the EBSP observed. Often an etchant was
For mechanical preparation and quality (pattern quality index) as a added to increase preparation
the residual strain induced into measure [00Kat]. speed. Hard materials (HV > 250)
the materials surface it is impor- were not easy to prepare espe-
tant to consider the hardness of A reliable automatic EBSD scan cially without chemical support.
the material. Main parameters for is only possible if the EBSP's are Ceramics could not be prepared
hardness are the chemical com- of good quality (contrast) so that satisfactorily in reasonable time.
position and processing history. they can be recognized by the Some more trials especially by
The investigated materials cover computer. The average height of varying polishing cloths and
the range from 5.5 to 1756 HV, the detected peaks in the Hough abrasives may lead to another
the roughly estimated EBSD in- transformation that is used for result.
formation depth is between 20 pattern recognition [92Kri] can be
and 370 nm (Fig. 2) [00Kat]. used as a quality parameter Vibratory polishing delivers
All materials were chosen to be [94War], [98TSL]. It should be mostly the best quality EBSP's.
coarse grained and relaxed to noticed that the so defined pat- The main drawback is the enor-
ensure that any pattern distortion tern quality index (PQI) is also mous time expense. But normally
comes from the preparation. dependent on other factors like it can be accepted as no addi-
crystal system, orientation, etc. tional abrasive or manpower are
Procedure Nevertheless it is a very good needed. If the entire preparation
From each material five samples measure, even not absolute, for (including grinding) is done vibra-
were taken. Three of them were the strain in the diffraction vol- tory a high surface waviness is
ground and polished on a stand- ume of material. For automatic observed because the scratch
ard rotational polishing machine scans a PQI above 80 is desir- depth cannot be reduced as effi-
(RotoPol-22 / RotoForce-4). The able. But depending on the crys- ciently as the deformation depth.
final step was done once on the tal symmetry and orientation pat- Due to the long polishing time
same machine, once on the new tern with much lower PQI can be strong relief may appear, espe-
one (LaboPol-4 / LaboForce-1) indexed correctly as well. cially in ceramics. Therefore
and on the vibratory polisher re- cloths must be chosen very care-
spectively. The fourth sample Using standard mechanical fully.
was prepared entirely on the new preparation procedure some- Fig. 4 and Fig. 5 compare the
machine and the last was vibra- times good results were ob- standard preparation route as
tory ground and polished (Fig. 3). tained. This was especially the well as vibratory polishing with
Preparation was stopped when case for very hard materials like mechanical polishing using re-
an optically (light microscopy) ceramics but also for a couple of duced pressure and rotation
good specimen surface was metals. It must be mentioned speed. In both cases the used
achieved. EBSP quality was though that any kind of chemical material is Titanium. In compari-
checked. If needed the final step or physical support was needed son to the standard preparation
was repeated, modified or con- as well. The most important ad- route it can be seen that there is
sidered inappropriate and conse- vantage remains the short prepa- some deformation left, especially
quently another approach was ration time. More "difficult" mate- in the edge region - i.e. deforma-
chosen. rials could be managed very of- tion twins. Also the contrast in
ten when transferred for final the light micrograph as well as in

21
the EBSP is lower that again ity. Sometimes it may be suffi- the final stage it should not be
confirms surface damage. The cient to use the standard ma- too much as it is applied just to
result with this softer rotation chine especially if supported speed up the process and solve
polishing is comparable to that chemically. But sometimes even the deformed layer, not really to
achieved using vibratory polisher. the new machine may have diffi- achieve the etch effect.
But it is much faster. Compared culties or imply a very long proc-
to a specimen prepared only by ess. In that case the vibratory Conclusions
vibratory polishing this machine polisher would be preferred. Soft LaboPol-4/ LaboForce-1 is a
has an advantage regarding materials (HV below 80) should very useful polishing device for
evenness and flatness (Fig. 5). be ground and polished on the EBSD specimen preparation. Es-
In a laboratory with all these new machine and (if needed) pecially for soft metals the whole
equipment the preparation strat- taken to the vibratory polisher for preparation can be performed
egy for materials harder than 80 the final step (Fig. 5). Often etch- very gently, time efficiently and
HV would be standard mechani- ing between the different polish- with a minimum of surface relief.
cal grinding and polishing fol- ing steps is very helpful to re- It can be also recommended for
lowed by modified final stage on move the deformation. This also the final polishing step when
the polishing device with reduced significantly speeds up the pro- silica suspension is used. This is
pressure and rotation speed abil- cedure. If etchant is added on applicable to all materials and
even to ceramics.
This machine might also be use-
ful for the preparation of other
soft materials like polymers or for
a) other surface sensitive tech-
niques like AFM / STM.

Acknowledgement
The authors would like to thank
all Struers co-workers who sup-
ported them with advice regard-
ing the preparation procedures
for the different materials.

Appendix

Preparation Procedures for the


Investigated Materials

Only preparation procedures


b) where the LaboPol-4/
LaboForce-1 is involved are in-
cluded. Alternative preparation

Fig. 5:
a) correctly prepared sample with
high PQI (s. Fig. 4a)
- low roughness: 8 nm;

b) whole preparation done by


vibratory polishing - roughness
significantly higher: 229 nm.
Measured area: 1.26 mm x 0.95 mm.

22
38
Preparation for Electron
Backscatter Diffraction
with LaboPol-4 / LaboForce-1

Material References

[00Gee] Geels, K.; Gillesberg,


Surface Grit, Lubricant Rotation Force [N] Time B.: Polishing Dynam-
Grain size speed (min) ics, Prakt Metallogr 37
[rpm]
(2000) 150
Sn SiC-paper # 4000 Ethylenglycol 150 5 until plane
MD-Pan 6 µm Blue 150 5 5
Etching: * = 5 g Ammonium Persulfate + 50 ml dist. Water, 10-15 s
[00Kat] Katrakova, D.;
Mücklich, F.;
MD-Mol 3 µm Red 100 5 30
MD-Chem OP-S* Water, Soap 75 2.5 10 Damgaard, M.J.:
Specimen Preparation
Al SiC-paper # 1200 Wax; Water 150 5 until plane
and Electron
MD-Mol 3 µm Green 150 5 15 Backscatter Diffraction,
MD-Dur OP-S Water, Soap 100 5 15 Proc. EuroMet 2000,
MD-Nap OP-S Water, Soap 100 2.5 5 Saarbrücken, Germany
(2000) to appear
Zn SiC-paper # 1200 Ethylenglycol 150 5 until plane
MD-Mol 3 µm DiaPlus 150 5 15
MD-Nap OP-U* Water, Soap 100 5 5
[99Pet] Petzow, G.:
Metallographic Etching,
* = Nital, 3 %
Ohio, ASM International
(1999)
Mg SiC-paper # 2400 Ethylenglycol 150 5 until plane
MD-Mol 3 µm Yellow 150 5 20
[99Van] Vander Voort, G.F.:
Etching: Nital, 3 %( Isoamylalkohol), 10-20 s
Metallography Princi-
MD-Nap 1 µm Yellow 100 2.5 15
MD-Chem OP-S Ethanol (1:1) 75 2.5 15 ples and Practice,
Avoid water contact, Only clean with Ethanol, Use DP-A
Ohio, ASM Interna-
tional, Materials Park
(1999)
Cu SiC-paper # 1200 Water 150 5 until plane
MD-Pan 9 µm DiaPlus 150 5 15
Etching: * = 3.3 g FeCl 3 + 10 ml HCl + 40 ml dist. H 2O, some seconds [98Kat] Katrakova, D.;
MD-Mol 3 µm DiaPlus 150 5 45
Mücklich, F.: Orienta-
MD-Chem OP-S* Water, Soap 100 5 30 tion Imaging
Microscopy - a Power-
Al (Li) SiC-paper # 1200 Water 150 5 until plane ful Tool for Materials
SiC-paper # 4000 Water 150 5 5 Characterization,
MD-Dur OP-S Water, Soap 100 2.5 25 www.uni-sb.de/
MD-Nap OP-S Water, Soap 75 2.5 60
techfak/fb15/fuwe/oim
(1998)
Fe SiC-paper # 800 Water 150 7.5 until plane
MD-Allegro 6 µm Water 150 7.5 10
MD-Dac 3 µm Blue 150 Blue 15 [98Ren] Ren, S.X.; Kenik, E.A.;
MD-Chem OP-S Water, Soap 100 5 10 Alexander, K.B.; Goyal,
A.: Exploring Spatial
Resolution in Electron
routes are given only if they lead LaboForce-1 is not the same as Back-Scattered Diffrac-
to better results by means of the 150 rpm on a standard pol- tion Experiments via
EBSP quality. Generally, Struers ishing machine. This is because Monte Carlo Simula-
tion, Microsc Microanal
consumables were used. Other- the samples rotate at 8 rpm and
4 (1998) 15
wise it is notified. Which are the not as usual at 150 rpm. If an
corresponding products from etchant was added to the silica [98TSL] TSL: OIM Data Collec-
competitors can be found in suspension, this is noted with a tion User Manual,
[99Pet]. Please keep in mind that star (*) and its chemical composi- Drapper, Utah, USA,
150 rpm on the LaboPol-4 / tion is given in the respective TexSEM Laboratories
part of the table. (1998)

23
Material

Surface Grit, Lubricant Rotation Force [N] Time


Grain size speed (min)
[rpm]
Ti SiC-paper # 500 Water 150 10 until plane
SiC-paper # 1200 Water 150 10 20
MD-Chem OP-S Water, Soap 100 7.5 90

Be SiC-paper # 1200 Water 150 10 until plane


MD-Largo 9 µm DiaPlus 150 10 10
MD-Nap 1 µm Red 150 10 15
Slight etching: H 2SO 4, 10 %, some seconds
[94War] Wardle, S.T.; Lin, L.S.;
Cetel, A.; Adams, B.L.:
Ni SiC-paper # 1200 Water 150 15 until plane
Orientation Imaging
MD-Dac 3 µm Green 150 10 20
LeCloth Alumina, 1 µm Water 100 7.5 20 Microscopy: Monitoring
LeCloth Alumina, 0.3 µm Water 100 7.5 25 Residual Stress Profiles
MD-Chem OP-S Water, Soap 100 7.5 15 in Single Crystals Using
an Image Quality Para-
TiAl SiC-paper # 500 Water 150 10 until plane meter, IQ, Proc.
MD-Dur 6 µm Blue+H 2O2,1:1 150 15 35 MSA'94, New Orleans
MD-Dur 3 µm Blue+H 2O2,1:1 150 12.5 15
MD-Chem OP-S* Water, Soap 75 7.5 15 (1994) 680
OPS* = OPS + H2O2 + NH 4OH = 8:1:1
[93Joy] Joy, D.C.: Scanning
Electron Microscopy in
RuAl MD-Pan 15 µm Red 150 15 until plane
MD-Dur 6 µm Blue 150 15 10 Materials Science and
MD-Dur 3 µm Blue 150 15 10 Technology 2A, Cahn,
LeCloth Alumina, 1 µm Water 100 10 20 R. W. ;Haasen, P., VCH
LeCloth Alumina, 0.3 µm Water 100 10 25
MD-Chem OP-S Water, Soap 100 10 15
(1993) 221

[93Schw] Schwarzer, R.A.: The


RhAl SiC-paper # 800 Water 150 15 until plane
MD-Allegro 6 µm Green 150 15 15 Determination of Local
MD-Dac 3 µm Green 150 15 20 Texture by Electron Dif-
MD-Dur 1 µm Green 150 15 30 fraction - A Tutorual
MD-Chem OP-S Water, Soap 100 10 30 Review, Tex Microstr 20
Etching: 10 ml HNO3 + 100 ml HCl + 50 ml H 2O, hot, 1-5 min (1993) 7

[93Wri] Wright, S.I.: A Review


Material of Automated Orienta-
tion Imaging
Microscopy (OIM), J
Surface Grit, Lubricant Rotation Force [N] Time Comp Assisted
Grain size speed (min) Microscopy 5 (1993)
[rpm] 207
BaTiO 3 RotoPol 22 / RotoForce 4
MD-Plan 15 µm Blue 150 25 until plane [92Kri] Krieger Lassen, N.C.;
MD-Allegro 6 µm Blue 150 25 5 Juul Jensen, D.;
MD-Pan 6 µm Blue 150 30 7
MD-Dac 3 µm Blue 150 25 5 Conradsen, K.: Image
LeCloth Alumina, 1 µm Water 150 20 5 Processing Procedures
LeCloth Alumina, 0.3 µm Water 150 20 5 for Analysis of Electron
Etching: 5 ml HCl + 5 ml HNO3 + 1 ml HF + 89 ml H 2O, 5-10 seconds Back Scattering Pat-
terns, Scanning
ZrO2 RotoPol 22 / RotoForce 4 Microscopy 6 (1992)
MD-Plan 15 µm DiaPlus 150 25 until plane 115
MD-Plan 9 µm DiaPlus 150 25 5
MD-Plan 6 µm DiaPlus 150 25 10 [92Ran] Randle, V.: Microtexture
MD-Dur 3 µm DiaPlus 150 25 10
MD-Dur 1 µm DiaPlus 150 25 10 Determination and its
MD-Plus 0.25 µm DiaPlus 150 20 10 Applications, London,
MD-Chem OP-AA Water 150 20 15 The Institute of Materi-
Annealing: 1200 °C, 10 min als (1992)

Al 2O3 RotoPol 22 / RotoForce 4 [82Sam] Samuels, L.E.:


MD-Plan 15 µm DiaPlus 150 30 until plane Metallographic Polish-
MD-Plan 9 µm DiaPlus 150 30 5 ing by Mechanical
MD-Plan 6 µm DiaPlus 150 30 10
Methods, Ohio, ASM
MD-Dur 3 µm DiaPlus 150 30 10
MD-Dur 1 µm DiaPlus 150 25 10 International, Metals
MD-Plus 0.25 µm DiaPlus 150 25 10 Park (1982)
MD-Chem OP-AA Water 150 20 15

24 Annealing: 1200 °C, 15 min

You might also like