You are on page 1of 233

Page 1 of 232

Page 2 of 232
Page 3 of 232
Page 4 of 232
ecam BOOM
Gantry ARM
HAND
CONTROL CAUDAL TILT SWITCH

E STOP
PPM

esoft
DET 1 Workstation

E STOP
C-FRAME

TOWER

PALLET

PALLET CLUTCH
RELEASE
SWITCH_LEFT HORIZ
VERTICAL ACME MOTOR ASM.
SCREW COVERS
REAR DOCKERS
BED

LEFT D
COVER FRONT
BED
LEFT CABLES
FRONT
ROLLER DET 2

FRONT BED

Page 5 of 232
E.CAM SYSTEM

YOKE MOTOR
SHAFT BRAKE
90VDC MED
ENCODER 24VDC
E.SOFT
MDB YOKE 1 HOME
10K POT CAUDAUL TILT
CAUDAL TILT CLUTCH 24VDC
YOKE 1
WITH 10K POT

TUBE GEAR
STRIP

RAIL
DETECTOR
ACQUISITION DETECTOR 1 CART 10K POT
1 SHOWN
E.SOFT WORKSTATION DETECT
SNAC VI DEO C-FRAME
MONITOR COLLIMATOR WITH EEPROM MOTOR
ETHERNET(EXT)
SEATED 24VDC
ETHERNET(SNAC) ETHERNET
BLANK
MED 2 ( NOT SHOWN )

DPC CLEAR C-FRAME


MED 1

SCSI BLANK PALLET LEFT ARM REST SWITCH


SCSI RELEASE MAGNET
RS485 LOOK AHEAD
SOUND
TAXI COMMUNICATION 140MBits PALLET

TUBE GEAR
TICKS B
BRAIN MAGNET
SCL 38. 4K BAUD RATE

RS-485 COM. A

RAIL
POSITION RIGHT ARM REST SWITCH
P/S5V&12V

J13 10K POT LOOK AHEAD


J16 PMAC SHAFT YOKE 2
OUT IN
MOTORP.S.

J7 ENCODER 10K POT


AMP 4
AMP 3
AMP 2
AMP 1
MECP

J4 BELT DRIVE
ESB/SOLENOID J12
J14 MDB YOKE 2 HOME
P/S 24V

AC POWER H2 180
J15 ON HORIZONTAL CLUTCH 24VDC RE-CONFIGURE SOLENOID 165VDC
PPM 12VDC J11 SWITCH
FAN1 MEB FAN2 MDB OUT HOME DETECTOR 2 CONNECTS TO J12
HAND CONTROLLOR

J10

HORIZONTAL MOTOR
J1
J2

J3
J5
J6
J9
J8

HOME
90VDC
OUT ER R I NG
DETECTOR 2
I NNER RI NG
RS- 485 COMM.

CLEAN BOX
ROTATE BRAKE 24VDC
ROTATE MOTOR 109VDC
TOWER E
200 - 240 VAC 30 AMP PPM
1 C ROTATE SHAFT
A ENCODER
HD3 DET ECT OR M
TO DETECTOR 1

T AXI MDB
ROTATE / DETECTOR
DETECTORP/S 1 HOME
DTPS BOARD
FLOOR
PCALCBOARD MAGNET
TAXI DOCKING
AIMD BOARD DOCKING FRONT 2
OUTSIDE TUB 00 REAR
INSIDE TUB PI59 BD
DD59 BOARD
PRE-AMP PALLET
DA C CARRIAGE
SHAF T EN CODER
PMT ’ s UP
10K POT BASE
DETECTOR 2 SHOWN MDB 10K POT
SHAF T CARRIAGE DOWN 1 0 K POT
E NCODE R HOME
FRONT UP / DOWN HOME SWITCH REAR HOME
UP / DOWN MOTOR 90VDC MDB
NOTE: AMP BOARD FUNCTION SWITCH
FLOOR
AMP 1 = ROTATE MOTOR LEFT / RIGHT MOTOR DRIVE 90VDC
AMP 2 = RADIAL 1 / PATIENT BED REAR MDB REAR UP / DOWN
AMP 3 = RADIAL 2 / PATIENT BED UP / DN MOTOR 90VDC
AMP 4 = GANTRY L / R - PATIENT BED IN / OUT

Page 6 of 232
Page 7 of 232
PPM

E STOP

TOWER
HAND
CONTROL

Page 8 of 232
1. STARTS OR STOPS IMAGE ACQUISTIONS.
1 2 2. CLEARS THE PATIENT POSITIONING MONITOR
(PPM) SCREEN SHOWN ON THE DISPLAY PANEL.
3. MOVES THE PATIENT BED TO A PREDETERMINED
SETTING FOR LOADING/UNLOADING WHEELCHAIR
3 4 PATIENTS.
5 4. MOVES THE PATIENT BED TO A PREDETERMINED
SETTING FOR LOADING/UNLOADING GURNEY
7 6 PATIENTS.
5. MOVES THE PATIENT BED IN AN UPWARD
8 DIRECTION.
NOTE: THIS BUTTON IS ALSO USED FOR AUTO
9 11 HEIGHT POSITIONING.
6. MOVES THE PATIENT BED AWAY FROM THE
10 12 GANTRY.
7. MOVES THE PATIENT BED IN TOWARDS THE
GANTRY.
8. MOVES THE PATIENT BED IN A DOWNWARD
14 13 DIRECTION.
9. MOVES DETECTOR 1 OUTWARD (RADIALLY).
10. MOVES DETECTOR 1 INWARD (RADIALLY).
15 16 11. MOVES DETECTOR 2 OUTWARD (RADIALLY).
12. MOVES DETECTOR 2 INWARD (RADIALLY).
NOTE: NUMBERS 11 AND 12 ARE NOT APPLICABLE
TO A SINGLE-DETECTOR SYSTEM.
13. ROTATES THE GANTRY IN A CLOCKWISE
DIRECTION.
14. ROTATES THE GANTRY IN A COUNTERCLOCKWISE
DIRECTION.
15. MOVES GANTRY TO LEFT DIRECTION.
16. MOVES GANTRY TO RIGHT DIRECTION.
NOTE: NUMBERS 15 AND 16 ARE NOT APPLICABLE
TO A SINGLE-DETECTOR SYSTEM.

Page 9 of 232
Page 10 of 232
CAUDAL TILT CLUTCH
GEAR TEETH

Page 11 of 232
Page 12 of 232
DET. 1 RADIUS
DRIVE ASM.
E STOP

HAND
CONTROL

RADIUS
HOME FLAG

RADIUS
POT GEAR

RADIUS
POT

Page 13 of 232
RECONFIGURE
FLAG

DETECTOR
RADIUS HOME
FLAG

FLAG SENSORS

GEAR STRIP

DETECTOR
RADIUS
DETECTOR
CALIBRATE POT
RADIUS
ADJUST
CALIBRATE POT

DETECTOR
RADIUS
CALIBRATE POT
GEAR

Page 14 of 232
Page 15 of 232
DETECTOR 1
RADIUS DRIVE
ASSEMBLY

Page 16 of 232
Page 17 of 232
GANTRY HEIGHT
ADJUSTMENT
BOLT

LEFT SIDE
GANTRY FRONT
WHEEL

STOP BOLT

LOCKING BOLT
FRONT WHEEL
WEAR PLATE

Page 18 of 232
RADIAL DRIVE
SPUR GEAR
(WORM GEAR SIDE)

RADIAL DRIVE
SPUR GEAR
(MOTOR SIDE)

Page 19 of 232
Page 20 of 232
Page 21 of 232
OUTER RING

RECONFIGURE
SOLENOID INNER RING
E STOP

A FRAME

CABLE TAKE-UP
ROLLERS HAND
CONTROL

REAR BED

ROTATE
HOME FLAG

ROTATE
MOTOR

Page 22 of 232
ROTATE
ROTATE MOTOR
HOME FLAG MOUNTING
BRACKET
ROTATE
HOME
SENSOR

ROTATE
MOTOR

Page 23 of 232
ROTATE
HOME FLAG

ROTATE
HOME
SENSOR

Page 24 of 232
ROTATE MOTOR
DRIVE GEAR
ROTATE HOME
ADJUSTMENT POT

Page 25 of 232
OUTER RING
INNER RING

ROTATE HOME FLAG

Page 26 of 232
RECONFIGURE
SOLENOID ASSEMBLY

OUTER RING

INNER RING

A FRAME

FRONT OF REAR BED


(IN UP POSITION)

Page 27 of 232
RECONFIGURE
SOLENIOD STATUS
SWITCH/SENSOR

RECONFIGURE
LATCH

RECONFIGURE
SOLENIOD

RECONFIGURE
SENSOR

A FRAME

Page 28 of 232
RECONFIGURE
LATCH

RECONFIGURE
FLAG

RECONFIGURE
SOLENOID

RECONFIGURE
A FRAME
SENSOR

Page 29 of 232
DETECTOR CABLE
TAKE-UP ROLLERS

Page 30 of 232
A FRAME

REEL SPRING

MDB

REEL SPRING
STEEL CABLE

DETECTOR
CABLES

Page 31 of 232
STEEL REEL CABLE

STEEL REEL CABLE “V”


ROLLERS - (AT REAR
BOTTOM RIGHT OF GANTRY)

Page 32 of 232
E-STOP-(GANTRY
LEFT SIDE-OUTSIDE)

E-STOP-(GANTRY
LEFT SIDE-INSIDE)

Page 33 of 232
E-STOP
(VIEWED FROM INSIDE
GANTRY REAR)

Page 34 of 232
RING LOCK
LEVER

Page 35 of 232
RING LOCK LEVER
SPRING

RING LOCK LEVER

RING LOCK LEVER


PIVOT PIN

Page 36 of 232
OUTER RING

RING LOCK
BLOCK

INNER RING

Page 37 of 232
Page 38 of 232
FRONT
BED

Page 39 of 232
ROLLER REAR ROLLER
ASM. ASM.
RIGHT LEFT

PALLET IN/OUT DRIVE BELT


T

LEF
RES

TA
ARM

RM
HT

RES
RIG

T
GROUND
ROLLER

FRONT
Page 40 of 232
CUT-OUT IN GUIDE
ROD FOR BRAIN
POSITION SWITCH

BRAIN
POSITION
SWITCH
TRAY FOR
PALLET
IN/OUT DRIVE
BELT

OUTSIDE
GUIDE ROD

FRONT BED-UNDERNEATH-COVER REMOVED


Page 41 of 232
Page 42 of 232
ACME SCREW

BRASS NUT

ACTUATOR
FOR SAFETY
SWITCH

WEAR
INDICATOR
PLATE

Page 43 of 232
BOLT FOR FRONT
COLUMN COVER

VERTICAL
MECHANICAL
STOP BOLT

VERTICAL GUIDE
RAIL

Page 44 of 232
VERTICAL

Page 45 of 232
VERTICAL
DRIVE
PULLEYS

VERTICAL
DRIVE BELT

Page 46 of 232
FRONT BED IN/OUT (HORIZONTAL) DRIVE ASSEMBLY

Page 47 of 232
FRONT BED IN/OUT (HORIZONTAL) DRIVE ASSEMBLY - LEFT SIDE

FRONT BED IN/OUT (HORIZONTAL) DRIVE ASSEMBLY - RIGHT SIDE


Page 48 of 232
PALLET
INSIDE
GUIDE RODS
PALLET REMOVED -
PULLED BACK INTO
BRAIN POSITION

Page 49 of 232
FRONT BED DOCKING ASSEMBLY

Page 50 of 232
NEW STYLE DOCKING ASSEMBLY
Page 51 of 232
SENSOR

MAGNET

Page 52 of 232
Page 53 of 232
Page 54 of 232
LOWER LIMIT
VERTICAL SENSOR
MOTOR

MDB

Page 55 of 232
REAR BED
IN
DOWN
POSITION

Page 56 of 232
Page 57 of 232
ECAM TOWER POWER SUPPLIES/AMPLIFIERS

MECP – (Motion Electronics Control Processor) – Control processor for all motions.

AMP 4 – (Analog Amplifier) – PHS IN/OUT – Gantry Left/Right.


AMP 3 – (Analog Amplifier) – Radial 2 – Front Bed UP/DN.
AMP 2 – (Analog Amplifier) – Radial 1 – Rear Bed UP/DN.
AMP 1 – (Analog Rotate Amplifier) – Rotate Axis.

Motor Power Supply – Supply voltage to motors.

+5/+12 VDC Power Supply

• +5 VDC – All logic and remote sensors (opto home sensors, magnetic reed sensors,
microswitches).
• +12VDC – Circuits related to analog and differential signals.

+24 VDC Power Supply


• Supply current for the VBUS AC relays RL4 and RL5.
• Supply the MED/MEDU boards for the Caudal Tilt clutches and C-Frame motion.
• Supply power for the Rotate, Radial 1 and Radial 2 axes brakes.
• Supply for the patient bed IN/OUT clutch.

Page 58 of 232
MOTOR
POWER +5V
SUPPLY +/-12V
POWER
SUPPLY

+24V
POWER
SUPPLY

Page 59 of 232
J7-DETECTOR 2

PMAC BOARD

J4-DETECTOR 1

MEB

Page 60 of 232
J7 J4

PMAC

MEB

Page 61 of 232
Page 62 of 232
DETECTOR PROCESSOR
COVER

Page 63 of 232
DETECTOR
PROCESSOR
ASSEMBLY

Page 64 of 232
Page 65 of 232
DEAD
TIME

AIMD
TAXI

PCALC

Page 66 of 232
DTPS

Page 67 of 232
Page 68 of 232
SIEMENS ecam ESOFT NORMAL
POWER-UP / POWER-DOWN SEQUENCE

Power Up Sequence

01. Make sure all breakers are off.


02. Turn on the MAIN breaker.
03. Turn on the COMPUTER breaker.
04. Turn on the CAMERA breaker.
05. Turn on the MOTION breaker.
06. Turn on the ECG breaker.
07. When the PPM screen comes up, turn on the esoft
workstation.
08. When the esoft workstation finishes powering up, you
may login into the system by pressing “CTRL-ALT-DEL”
and providing the proper password.

Power Down Sequence

01. At the esoft workstation, select End Session from the


Options menu.
02. Click the Shut Down button.
03. Wait until the system saves the system data.
04. Press the power button to turn off the power to the
essoft workstation computer.
05. After the esoft workstation is powered down, perform
shutdown of the SNAC computer from the PPM.
06. After the SNAC computer has shutdown, turn off the
MOTION breaker.
07. Turn off the CAMERA breaker.
08. Turn off the ECG breaker.
09. Turn off the MAIN breaker.

Page 69 of 232
ECAM NORMAL POWER UP

COMPUTER BREAKER

COMPONENT OBSERVE
SNAC computer begins SNAC green LED turns on. Hard disk access amber
to boot. LED turns on.

CAMERA BREAKER

COMPONENT OBSERVE
MEB fan #2 turns on. Fan on left side of MEB Base turns on.
MEB power on. MEB Backplane green power LED's light; (+12V, -
12V, +5v, +5V MECP, +5V SENSOR). Axes
connector LED's light; (D20 thru D26)
MECP boots. Red LED on MECP flashes once.
Red & Green LED's on MECP ON for 2 seconds.
Red & Green LED's on MECP OFF.
Green LED's on MECP ON steady.
PMAC boots. Red LED very dim while MECP is booting.
Yellow and Green LED's on after successful MECP
boot.
Green Amplifier LED's light.

MOTION BREAKER

COMPONENT OBSERVE
+24V Switched turns on. Patient bed pallet clutch engages.

Motor Power Supply LED D1 lights & AC Fan #1 turns on.


(VBUS) turns on.

Page 70 of 232
ECAM TOWER / CLEAN BOX CIRCUIT BREAKERS

Tower / Clean Box - Breaker Side

MECP AMP AMP AMP AMP MOTOR P/S

FAN #2 FAN #1

MAIN
INPUT MOTION COMPUTER

AC INPUT

ECG CAMERA

ALL SWITCHES SHOWN IN


RED INDICATOR
THE ON POSITION
GREEN INDICATORS

Clean Box - Breaker Side

MOTION COMPUTER
MAIN INPUT

AC
INPUT

ECG CAMERA

Clean Box - A/C Outlet Side

VOLTAGE SELECTION

AUX MODEM AUX AUX

200v
208v
AUX AUX AUX AUX 220v
230v
240v

ECG INPUT
AC INPUT 0.5A 120V Page 71 of 232
ECAM esoft SYSTEM POWER UP ACQUISITION esoft

FRONT REAR

FRONT - SNAC - REAR

MAIN CIRCUIT
POWER e.soft
BREAKER - 30 AMP
FD
( 200 - 240 VAC )
VOLTAGE
HD MONITOR 19"
TAP HERE PRESS SOFT START
HARD DRIVE TO POWER UP ICON

CD
PPM
CLEAN BOX

ETHERNET PLUG

COMPUTER BREAKER HOME SYSTEM


POWER TO SNAC &
ICON COMPUTER

HOME
12VDC RESET
GREEN RED
LEDs
ETHERNET PLUG POWER
CAMERA BREAKER
SWITCH
POWER TO DETECTOR S1
ASSEMBLIES & MOTION
MEB BOOTING MEB CARD CAGE POWER STEP 1
ELECTRONICS CARD

GREEN

PS 5 / 12VDC
CAGE - LOGIC PMAC DETECTOR POWER 120 VAC

RED
ON

MOTOR PS
MOTION BREAKER OEM
+5 MEB POWER SUPPLY
POWER TO MOTION

AMP 4

AMP 3

AMP 2

AMP 1
+5 MECP +/- 5 & 20 VDC
ELECTRONICS CARD

MECP

PS 24VDC
CAGE - MOTOR POWER +5 SEN
15 VDC +12 VDC
OFF
-12 VDC
ECG BREAKER
DTPS -20VDC
-5VDC
POWER TO ECG
PLUG PATIENT BED +5VDC
-15VDC +20VDC
FAN #1 FAN #2
-12VDC
ON OFF +12VDC HV
+15VDC
MEB CARD CAGE POWER STEP 2

PS 5 / 12VDC
EMERGENCY PCALC AIMD
STOP BUTTONS
ON

MOTOR PS
CPU
AMP 4

AMP 3

AMP 2

AMP 1
MECP

PS 24VDC
ECG
ON
-15VDC
+15VDC
-5.2VDC
+5VDC
+5VDC

FAN #1 FAN #2 DETECTORS 1 & 2

ON ON
Page 72 of 232
Page 73 of 232
ecam Tower AC Voltage Power Checks

MOTION COMPUTER
MAIN INPUT

AC
INPUT

ECG CAMERA

Make sure all circuit breakers are set to “ON”. Use the following table to verify
that the Tower AC voltages are within specifications.

WHAT TO CHECK VOLTAGE REFERENCE COMMENTS


200V
Tower clean box line Between load
to If line voltage is outside this range, contact an electrician.
cord voltage and neutral
240V

If the line voltage is within specifications, but the output from


Tower clean box 120V Between hot
the outlet strips is not within tolerance, change the
outlets +10% and neutral
transformer taps to bring the voltage within tolerance.

VOLTAGE SELECTION

AUX MODEM AUX AUX

200v
208v
AUX AUX AUX AUX 220v
230v
240v

ECG INPUT
AC INPUT 0.5A 120V

Page 74 of 232
Page 75 of 232
TROUBLESHOOTING ecam +5V/+12V POWER SUPPLY

Checking input to the power supply:

01. Apply power to the ecam.


02. With DVM, should measure approximately 110Vac present at the input of the
+5V/+12V power supply, black and white wires on the upper power supply. If
present, validate clean box. If clean box valid, continue.
03. With DVM, measure approximately 110Vac between E25 and E26. If no voltage,
replace MEB, if present repair cable to input of +5V/+12V power supply.

Checking power supply for voltage loss during boot process:

01. Power down the ecam.


02. Connect DVM to the +5V supply between TP31 and TP19. (On MEB)
03. Properly apply power to ecam system.
04. Monitor the +5V as the system powers up, should observe loss of only
approximately a few milli-volts as the PMAC boots. If the voltage drop is more,
replace the +5V/+12V power supply.

Checking output of the power supply:

01. Properly power down the ecam system.


02. Remove the Motor Power Supply (VBUS) power supply board from the MEB
backplane.
03. Properly power up the ecam system.
04. Measure with DVM the outputs of the power supply:
a. +12V +0.6V at both E5 and E6 (ORG), using TP19 as reference. (If not
correct, see below to adjust.)
b. +12V +0.6V at E11 (Blue), using TP19 as reference.
c. -12V +0.6V at E12 (Yellow), using TP19 as reference. (If not correct,
replace the +5V/+12V Power Supply; these voltages are not adjustable.)
d. +5V +0.25V / -0.00V at both E7 and E8 (Red), using TP19 as reference.
05. If the voltages are not present, replace the +5V/+12V Power Supply.

Adjusting the outputs of the power supply:

NOTE: If adjustment is required, it will be necessary to remove the Motor Power Supply
board and maybe the Rotate Amp board to access adjustment pot. Before removing this
or any board(s), properly shutdown ecam system, remove board(s), then properly apply
power to ecam system.

01. With power applied to the ecam system, measure with DVM the outputs of the
power supply. Adjust if Clean Box tapping matches input voltage.

Page 76 of 232
a. +5V +0.25V / -0.00V at TP31 using TP19 for reference. Adjust with
V1ADJ pot on power supply; (CW increases voltage).
b. +12V +0.6V at E5 using TP19 for reference. Adjust with V2ADJ pot on
power supply; (CW increases voltage).

Turn potentiometer CW to increase voltage.

V1ADJ (+5V)
+5/+12V Power Supply
V2ADJ (12V)

MEB Backplane
V1ADJ (+24V)
+24V Power Supply

Page 77 of 232
MOTOR
POWER +5V
SUPPLY +/-12V
POWER
SUPPLY

+24V
POWER
SUPPLY

Page 78 of 232
Page 79 of 232
TROUBLESHOOTING ecam +24V POWER SUPPLY

Checking input to the power supply:

01. Apply power to the ecam.


02. With DVM, should measure approximately 110Vac present at the input of the
+24V power supply, black and white wires on the lower power supply. If present,
validate clean box. If clean box valid, continue.
03. With DVM, measure approximately 110Vac between E27 and E28. If no voltage,
replace MEB, if present repair cable to input of +24V power supply.

Checking output of the power supply:

01. Properly power down the ecam system.


02. Remove the Motor Power Supply (VBUS) power supply board from the MEB
backplane.
03. Properly power up the ecam system.
04. Measure with DVM +24V +1.2V between E15 or E16 (GRY) to E19 or E20
(BRN).
05. If the voltages are not present, replace the +24V Power Supply.

Adjusting the outputs of the power supply:

NOTE: If adjustment is required, it will be necessary to remove the Motor Power Supply
board and maybe the Rotate Amp board to access adjustment pot. Before removing this
or any board(s), properly shutdown ecam system, remove board(s), then properly apply
power to ecam system.

01. With power applied to the ecam system, measure with DVM the output of the
power supply. Adjust if Clean Box tapping matches input voltage.
a. +24V +1.2V between E15 or E16 (GRY) to E19 or E20 (BRN). Adjust
V1ADJ pot on power supply; (CW increases voltage).

Turn potentiometer CW to increase voltage.

V1ADJ (+5V)
+5/+12V Power Supply
V2ADJ (12V)

MEB Backplane
V1ADJ (+24V)
+24V Power Supply

Page 80 of 232
MOTOR
POWER +5V
SUPPLY +/-12V
POWER
SUPPLY

+24V
POWER
SUPPLY

Page 81 of 232
Page 82 of 232
ECAM HD3 DETECTOR OEM POWER SUPPLY VOLTAGE
VERIFICATION
Use this procedure to verify the detector OEM power supply voltages.

01. Proper apply power to gantry and detector.


02. Verify the following voltages.

03. Set the oscilloscope time base at 5msec/div and check the ripple voltage on the
+5.15V and –5.2V supplies. The ripple voltage for these supplies must not
exceed 50mV.
04. There is an in-line fuse which comes off the +5.15V terminal on the OEM Power
Supply. Check this fuse if there is no +5.15V to the detector PC boards.

OEM R59

R7
R405

-20V
TP22
-5V

+5V
TP2
+20V

HV

DTPS S1

-12V

+15V

+12V

Page 83 of 232
OEM
POWER
SUPPLY

Page 84 of 232
Page 85 of 232
ECAM HD3 DETECTOR DTPS (DETECTOR POWER SUPPLY)
VOLTAGES VERIFICATION

DTPS Low Voltage Verification

Using a DVM and oscilloscope, verify the following DTPS output voltages and adjust if
necessary.

ADJUST MEASURE AT FOR RIPPLE GROUND


TP4 +10.00V +10MV
Check only, no
adjustment.
TP5 -10.00V +10MV
<20MV TP2
R102 TP20 +12.00V +10MV
R116 TP21 -12.00V +10MV
R75 TP18 +15.00V +10MV
R132 TP19 -15.00V +10MV

DTPS High Voltage Verification

01. Make sure the HV is on and allowed 15 minutes for warm-up.


02. Using a DVM and oscilloscope, verify the following DTPS high voltages and
adjust if necessary.

ADJUST MEASURE AT FOR RIPPLE GROUND


R11 TP7 -100V +100MV
<100MV TP2
R10 TP6 -200V +200MV

Page 86 of 232
DTPS Oscillator Verification

Check and adjust if necessary.

ADJUST MEASURE AT FOR GROUND


Sawtooth pulse width of 50usec +2usec
R25 TP10 TP2
(refer to picture of DTPS(E) signal)

DTPS Service Map

Page 87 of 232
Page 88 of 232
Page 89 of 232
ECAM HD3R/HD4 DETECTOR VOLTAGE CHECKS

The new detector assembly for the Siemens ecam system is called the HD3R or HD4.
Currently, we have little information concerning this new detector, however this does not
mean there is nothing we can do. We are still limited due to not having access to
“Standard Plus” diagnostics. Operational speaking, both the HD3 and HD3R detectors
function the same.

Primary Changes on the HD3R Detector

• The PI59 and the PCALC boards have been combined into the PI59R board.
(This PI59R board is located inside the detector tub.)
• The AIMD and the DTPS boards have been combined into the AIMDR board.
• The OEM power supply has been changed. (Still referred to as the OEM.)
• Both the AIMDR and the PI59R boards are “digitally” calibrated using
diagnostics which are currently not available to us.

HD3R Voltage Checks

-100V
• Can be measured on the AIMDR board, TP116 (J15) using TP63 (Z15) as
reference. (+100mV)
-200V
• Can be measured on the AIMDR board, TP115 (J16) using TP63 (Z15) as
reference. (+100mV)
HV S6
• The detector HV ON/OFF Switch S6 is located on the AIMDR board (C13).
HV Reference
• The HV Reference voltage can be measured on the AIMDR at TP109 (D12) using
TP10 (H12) as reference. The HV reference voltage adjustment is R918 (F13).

Page 90 of 232
ECAM HD3R DETECTOR OEM POWER SUPPLY VOLTAGE CHECKS

OEM Power Supply


(Top View)

R41

R16 R31 +5V


R19
-12V +12V
-5V

ECAM HD3R OEM POWER SUPPLY VOLTAGES


R41 on OEM TP58 (AIMDR, X8) +5V +5.18 to +5.22 TP63 (AIMDR, Z15)
Power Supply
R19 on OEM TP64 (AIMDR, X6) -5.20V -5.18 to -5.22 TP63 (AIMDR, Z15)
Power Supply
R31 on OEM TP121 (AIMDR, X17) +12.0V +11.98 to +12.02 TP63 (AIMDR, Z15)
Power Supply
R16 on OEM TP122 (AIMDR, X17) -12.0V -11.98 to -12.02 TP63 (AIMDR, Z15)
Power Supply
TP94 (AIMDR, P7) 3.3V +3.1 to +3.5 TP63 (AIMDR, Z15)
NOTE: Ripple voltage for these supplies must not exceed 50mV.

Page 91 of 232
Page 92 of 232
Page 93 of 232
ECAM HD3 DETECTOR GAMMA EVENT FLOW

1. Gamma ray strikes the crystal. The energy from the gamma ray is transferred to the
sodium iodide molecules in the crystal causing the excited molecules to emit light
photons which in turn strike the PMTs.
2. Energy from the photons is transferred to the molecules in the light-sensitive coating that
covers the inside base of the PMT. These exited molecules emit electrons. A focusing
electrode attracts the electrons to the first of a series of dynode plates inside the PMT.
Each plate is set to a decreasing voltage potential, the first is at a potential of
approximately –800V, the second approximately –700V and continues until the last plate
at -100V. Each successive plate more and more electrons are released multiplying the
few that strike the first plate to millions at the last plate. The result is a small but
measurable current at the output of the PMT.
3. The PMT current passes to the preamp resulting in the signal PAOUT. The signal
PAOUT represents the energy of the gamma ray event as seen by each PMT. At the
preamp, PAOUT is amplified 10 times then sent to the DD59(C) or (CF) board.
4. On the DD59, each PAOUT signal is summed through a 2K resistor network resulting in
the signal ES which is then sent to the PI59 board.
5. On the PI59, ES is now called SUM_E_MTX and applied to negative input of Energy
Sum Amp. Output of the Energy Sum Amp is not called SUM_E_AMP and used in two
places:
a. Applied to an amplifier to energy FLASH ADC on PI59 where it is digitized
representing the raw energy value of the gamma event;
b. Passes through a 49.93/4 resistor on the PI59 and renamed SUM_E_OUT and goes
to an amplifier on the AIMD. Output of the amplifier signal is now called
SUM_E.
6. Each of the 59 PAOUT signals are passed through the DD59 to the PI59 board where
they are fed into on of the 59 Flash ADCs which convert the analog signal from each of
the PMTs to a digital signal. These signals are then integrated. Integrations is
accomplished by the ASIC. These integrals represent the total energy of the detected
gamma ray and the portion seen by each PMT. (Each ASIC is connected to 12
Amplifier/FLASH ADC circuits, therefore five ASIC devices are required on the PI59
board.)

Page 94 of 232
7. While the event is being integrated on the PI59 board, the AIMD board is examining the
SUM_E signal for that same event. The analog SUM_E signal on the AIMD board is
used to drive several circuits.
a. INDET – The signal SUM_E is fed through an amplifier with a threshold detector
resulting in the output signal INDET; (Incident Detection). The signal INDET
starts the Flash ADC clock which in turn starts the integration process.
b. PILE-UP DETECT – If INDET stays high after the rising edge of PILE-UP
DETECT, then a pile-up condition exists. (In a PILE-UP, a new event has started
integrating before the previous event has completed integration.)
c. PILE-UP AMPLITUDE – This amplifier measures the amplitude of the pile-up
event. If the level is below a certain threshold, then the pile-up event is rejected.
d. EARLY DUMP – The signal SUM_E is compared to a coarse upper and lower
energy window. If a pile-up event does not fall in this range, then the EARLY
DUMP signal is sent to the PI59 board to stop integration.
e. COSMIC OVERLOAD – If SUM_E goes above +3.6V, the circuit generates a
signal which dumps the event. An OVERLOAD DELAY signal is also generated
to prevent a new event from starting right away.
f. AFTERGLOW – The afterglow circuit compensates for the baseline shift
encountered at high count rates.
8. AIMD –
a. Determines if the event is accepted or dumped.
b. Uses the signal INDET to feed the dead time circuit via a resistor/capacitor
network which is charged and discharged. This voltage is fed to an ADC in the
microprocessor controller where 5V is equal to 64% dead time and 0.3V is equal
to <4% dead time.
c. There is a 120MHz clock and phase circuitry which controls event integration on
the PI59 board.
d. There is also a microprocessor which acts as a board controller controlling board
functions such as High Count Rate, Temperature Compensation and
communicating with the DPC board in the SNAC to load the PMT gain DAC data
for tuning.
9. After the PI59 has completed the PMT integration process, this data is sent to a FIFO on
the PI59 which is connected to a FIFO on the PCALC board. From this FIFO, the data if
fed through buffers to the DSP, (Digital Signal Processing), nodes of which there are
eight. (One DSP acts as master node, the other seven are general processing nodes.)
Each node processes a full block before outputting the data to the TAXI interface and
accepting a new input block. The following events take place in the DSP nodes:
a. Correct for short integration or pile-up
b. Correct Zi magnitudes for offset errors
c. Process PMT data for LRF (Light Response Function) corrections (optical
distortion)
d. Calculate the event centroid and correct for energy
e. Scale and position the event.

Page 95 of 232
10. Event information is now sent from the PCALC board to the DPC board in the SNAC
computer via the TAXI, (Transparent Asynchronous Xmitter-Receiver Interface), which
is a high speed serial link having a transfer rate of 140Mbps. The TAXI can be thought
of as a pseudo-parallel register. It load data into one side and outputs it on the other, but
in this case the “other” side is separated by a long serial link utilizing 503/4 ohm coaxial
cable.
11. The TAXI link, Formatter and Input FIFO make up the data input function of the DPC.
a. TAXI Link – Receives the high speed serial data from the PCALC board and is
converted to a parallel format in the TAXI circuitry. The CTS, (Clear To Send)
signal is generated by the TAXI circuit to control data transfer speed between the
PCALC board and the DPC.
b. Formatter – The event data is received by the formatter in three 16-bit words
(SPECT mode) or four 16-bit words (COIN mode). The formatter rearranges the
event data into three 18-bit words (SPECT mode) or four 18-bit words (COIN
mode). This is done so the data is in a form that is easily used by the Correction
Processor.
c. Input FIFO – The purpose of the FIFO is to decouple the TAXI serial input from
the Correction Processor. Writing to the FIFO is done on the TAXI clock while
reading from the FIFO is done on the Correction Processor clock.
12. Correction Processor (COP) has two major functions:
a. Correction Processor Control (COPC) controls functions such as synchronous
state machines which keep the event processing synchronized.
b. Correction Processor Data Path (COPD) contains adders, multipliers and LUTs
that perform actual data calculations.
13. The event passes through four stages on the Correction Processor (COP)
a. Stage 1 – Zmap Coefficients are applied to the event data. The Zmap coefficients
are stored on the flash memory on the DPC board.
b. Stage 2 – Data is compared to upper and lower energy boundaries to determine if
the event falls within a certain energy window. If it is, the image event is passed
on to Stage 3. If it is tuning or peaking event, it is read out by the SNAC
computer at this point and no further processing is performed on the data.
c. Stage 3 – Performs the Linearity Correction algorithm on the event data. The
coefficients used in this process are unique to each detector and stored on flash
memory on the DPC board.

Page 96 of 232
d. Stage 4 – The event data is prepared for display.
i. The X and Y data is compared against upper and lower X/Y limits and if
within these limits, the event is considered to be within the Fine FOV.
Only events within the Fine FOV are processed further.
ii. The event is oriented for display. The X and Y values can be swapped
and/or flipped if desired.
iii. Following orientation, image panning is performed which is a function
that adds a pan offset to the detector coordinate system data.
iv. The magnify function is performed next. The algorithm used allow only
certain magnification factors.
v. Matrix sizing is now implemented. Matrix sizing allows viewing of the
image with various resolutions.
vi. The Scaler ID is determined. The Scaler ID is a function of the detector
and the associated analyzer.
vii. Counters are activated:
1. Scaler Counters – Four 32-bit counters that count events in the
Fine FOV for an associated head and analyzer.
2. Count Rate Counters – There 24-bit counters that count gamma
events.
3. Time Tick Counters – A 26-bit down counter counts COP clock
cycles and when zero is reached, a “Time Tick” is generated and
the counter is reset to a value determined by software.
14. Data from the COP is outputted to the SNAC computer by the Output FIFO, PCI
Interface and DMA Controller.
15. There types of communication between SNAC and ICON/esoft workstation:
a. Command Control – Sets up and controls motion and electronic parameters for
camera operation.
b. Growing Image Server – Sends an image in progress to the workstation for
display during acquisition.
c. Image File Server – Sends the final image to the workstation. (During acquisition,
image data is stored on the SNAC computer until acquisition is complete, then
entire acquisition is sent to the workstation.)

NOTE – All boards referenced such as the AIMD also include the AIMD(C)(CF), just as
the DD59(C)(CF) board, however does not include any boards contained in the HD3R
detectors.

• AIMD – Analog Interface Microprocessor-Digital


• ASIC – Application Specific Integrated Circuit
• DD59 – Detector Master Board
• DPC – Detector Processor Card
• DSP – Digital Signal Processor
• LUT – Look Up Table
• PCALC – Position Calculator
• PI59 – PMT Integrator
• SNAC – Siemens Nuclear Acquisition Computer

Page 97 of 232
Processing Computer
DAC Control

J17
DD59

FROM DETECTOR ONE


BUFFERS/ LINE
DRIVERS (If there are two detectors.) esoft
1 OF 16 DECODERS
PI59 J11
PCALC Input FIFO
MODIFIED 3 WORD FORMAT
DIAGNOSTIC SIGNAL PATH
. Pile-up Formatter 512x 18 Bits
. Short Int. DSP
uP . Zi Offset

DIAGNOSTIC SIGNAL PATH


ASIC
CONTROL . LRF Correction
R112 R127 . Centroid
SLAVE DATA BUS
Processor
First Word Flag
FLASH DSP Violation flag

HV
ADC . Scale (COP)
. Position

PAOUT
DSP
2K X, Y, E
Det. ID
TAXI Z MAP
Ethernet

TAXI
FLASH Stage 1

J16
FIFO

FIFO
DSP

J5
ASIC ASIC (External Network)

COP Data Path (Performs Calculations)


LUT - Z Map

COP Control (Synchronizes Operation)


ADC
R112 R127 (Flash
Memory)
DSP

TAXI IMAGE DATA


HV DPC Stage 2
- Window

CTS
TUNING PEAKING - PMT Data
FLASH
PAOUT ASIC - (Tuning Peaking)
2K ADC DSP

J9
-
Stage 3

J8
ASIC + + Ethernet
- Linearity Correction
DUMP/ MASTER
LINEARITY
CORRECTION (External Network)
R112 R127 FLASH INTEG
SUM_E DSP
ADC CONTROL NODE - LUT
HV DIAG. Stage 4
BASELINE
LOOPBACK (Flash - Fine FOV
SUM E
Memory)
OFFSET OFFSET
SCL SCL +
- Orientation ICON /
- Panning
PAOUT
2K INTERFACE INTERFACE - Magnify esoft Workstation
J3 J14 J3 J1 J2 J3 J13 J12 - Matrix Sizing

Controller
- Scaler ID
- Output Format Computer
To Detector 1 or (Acquisition)

DMA
Terminator . CTS
(If single head.)
(CLEAR TO
SEND)

J3 J5 J3 J1 J2 J3
EVENT CONTROL
SCL . Pile-Up
. Phase
INTERFACE . Indet
. Dump PCI Output FIFO
OVERLOAD
. Clock
DELAY
+ 3.6V
OVERLOAD
Interface 512x 18 Bits
4us
Ethernet
E OFFSET
BASELINE

One
Shot DUMP

PA9 EARLY
EARLY DUMP LOWER LIMIT
DUMP DISABLE
120nS 0.0V
DEADTIME
One
EARLY DUMP UPPER LIMIT Shot PA8 EARLY
INDET DUMP ENABLE DTAB
OFF ADJ.
Serial Communications Link

SUM E OUT
INDET
SERIAL I/O
INDET INDET
DC OFFSET
SUM Z 180-280nS
END PCI BUS
DACs

INDET One-Shot EARLY DUMP TIMING


ON ADJ.
PILEUP DETECT

PILEUP
ON ADJ.

uP PILEUP AMPLITUDE SNAC


Control MOTHEROBARD
PILEUP
. High Count Rate AMP. GAIN
. Temp Compensation
. Tuning ADC

AIMD
ISA BUS

Ethernet

SCL INTERFACE

SLR + / -
SLT + / -

HD3 Detector/SNAC Data Flow

Page 98 of 232
SNAC COMPUTER
Page 99 of 232
ECAM – MOTION AXES – (DUAL HEAD)

There are seven (7) axes of motion:

• Gantry Rotate CW/CCW


• Detector 1 Radial In/Out
• Detector 2 Radial In/Out
• Gantry Left/Right
• Front Patient Bed In/Out
• Front Patient Bed Up/Down
• Rear Patient Bed Up/Down

Each axis, the electrical interface is almost identical. Each axis consists of a motor, an encoder
feedback, (except Up/Down rear bed (PHS) which has no encoder), a position potentiometer,
(used at power up and with the home sensor for position verification), and a home flag sensor.

Three axes have a brake. The rotate brake is integrated into the rotate motor. The two other
brakes are on the output shaft of the radial 1 In/Out drive and the radial 2 In/Out drive. All three
brakes are controlled by the PMAC.

Other sensors such as Rear PHS carriage tilt sensors, Front Patient Bed docker sensors, etc. are
carried to the MEB for the MECP/MECP2 via the axis cable but do not report to the PMAC.
Each axis has one cable from the MEB to the axis. At each axis is a Motor Distribution Board
(MDB) which serves as a harness to break out the various signals to their individual components.

Page 100 of 232


ECAM – “HOME ALL” – (DUAL HEAD)

PREREQUISITES

• Detectors must not be tilted.


• System must be in “idle” state and not “diagnostic” mode.
• Front bed must be docked in front of gantry or in its storage position, have arm rest fully
retracted, rear bed must be fully up or down.

SEQUENCE – 180o

• Detector 1 & 2 move full out.


• Detector 1 homes.
• Patient bed In/Out homes.
• Gantry Left/Right homes.
• Front and rear patient beds Up/Down home.
• Rotate homes Detector 1 to 0o, system reads both ring home flags.
• Detector 2 homes.

SEQUENCE – 90o

• Detector 1 & 2 move full out.


• Detector 1 homes.
• Patient bed In/Out homes.
• Gantry Left/Right homes.
• Front and rear patient beds Up/Down home.
• Rotate homes (Detector 1 to 0o, system reads outer ring home flag. Gantry rotates CW to
inner ring flag, then rotates CCW until Detector 1 at 0o.
• Detector 2 homes.

Page 101 of 232


ECAM TOWER POWER SUPPLIES/AMPLIFIERS

MECP – (Motion Electronics Control Processor) – Control processor for all motions.

AMP 4 – (Analog Amplifier) – PHS IN/OUT – Gantry Left/Right.


AMP 3 – (Analog Amplifier) – Radial 2 – Front Bed UP/DN.
AMP 2 – (Analog Amplifier) – Radial 1 – Rear Bed UP/DN.
AMP 1 – (Analog Rotate Amplifier) – Rotate Axis.

Motor Power Supply – Supply voltage to motors.

+5/+12 VDC Power Supply

• +5 VDC – All logic and remote sensors (opto home sensors, magnetic reed sensors,
microswitches).
• +12VDC – Circuits related to analog and differential signals.

+24 VDC Power Supply


• Supply current for the VBUS AC relays RL4 and RL5.
• Supply the MED/MEDU boards for the Caudal Tilt clutches and C-Frame motion.
• Supply power for the Rotate, Radial 1 and Radial 2 axes brakes.
• Supply for the patient bed IN/OUT clutch.

Page 102 of 232


ECAM MOTION REQUEST

TOWER

Acquisition Computer

Programmed Motion MECP


- Motion Command PMAC
Programmed Motion - Machine Status Motion Control A MOTOR
Yes / No - Amplifiers Control Digital Velocity Value Motor Voltage
SNAC Computer Diagnostics - Logic Control M POWER
- Collision Avoidance - Brakes, Relays Control
- Encoders, Pots, HomeSwitches
P SUPPLY
- Reconfiguration
- Collimator Change
Motor Current
Hand Controller

MEB

MDB

AXIS MOTOR
NOTES:

1. The Acquisition Computer communicates to the SNAC computer via Ethernet.


2. The SNAC computer communicates to the MECP via RS485.
3. The Hand Control is plugged into the MEB.
4. The SNAC computer is housed within the Tower.

Page 103 of 232


ROTATE CABLE HANDLER POSITIONS

DETECTOR 1 at 0o REAR DETECTOR 1 at 90 o

VIEW

FLAG

HOME SENSOR SAFE TO PRESS


CW ROTATE. CAN
SAFE TO PRESS PRESS CCW
CW ROTATE ROTATE FOR 90o
ONLY.

DETECTOR 1 at 180o DETECTOR 1 at 270o

CAN PRESS EITHER


SAFE TO PRESS
ROTATE BUTTON
CW ROTATE FOR
FOR 180o OF
90o ONLY. CAN
ROTATION
RUN CCW ROTATE
FOR 270o

DETECTOR 1 at 360o

REAR VIEW
SAFE TO PRESS
CCW ROTATE
BUTTON ONLY

CAUTION:
PHYSICAL STOP

Page 104 of 232


Page 105 of 232
DIAGNOSTIC INFORMATION FOR esoft WORKSTATIONS

ECAM diagnostic tests for esoft workstations are “CD_ROM” based. The
diagnostics are presented on two levels.

STANDARD and STANDARD PLUS

The STANDARD level contains all the diagnostic test necessary to calibrate the
system. You can always run the STANDARD diagnostics by leaving the
password fields blank on the “Service Software Authentication” page.

The STANDARD PLUS level contains all the STANDARD level test, plus many
additional tests, which can be helpful in system troubleshooting.

NOTE:

• For troubleshooting a detector problem, make sure you are using the
STANDARD PLUS level of diagnostics.

The following pages discuss how to enter into a STANDARD diagnostics


session and the diagnostic programs.

SNAC application software must be running to boot diagnostics as


described in the following pages.

Page 106 of 232


STANDARD DIAGNOSTICS FOR esoft WORKSTATIONS

DETECTOR - CALIBRATE DIAGNOSTICS


DETECTOR - CALIBRATE - COIN: TIMING SKEW **
DETECTOR - CALIBRATE - COIN: WINDOW WIDTH **
DETECTOR - CALIBRATE - PI: ADC RAMP ZERO (DETECTOR 1)
DETECTOR - CALIBRATE - PI: ADC RAMP ZERO (DETECTOR 2)

DETECTOR - DISPLAY DIAGNOSTICS


DETECTOR - DISPLAY - COIN: CONFIGURATION
DETECTOR - DISPLAY - CONFIGURATION REPORT (DETECTOR 1)
DETECTOR - DISPLAY - CONFIGURATION REPORT (DETECTOR 2)

DETECTOR - LOAD DIAGNOSTICS


DETECTOR - LOAD - TIMING SKEW
DETECTOR - LOAD - WINDOW WIDTHS
DETECTOR - LOAD - INITALIZATION

DETECTOR DIAGNOSTICS
DETECTOR - TEST - PCALC: SELF (DETECTOR 1)
DETECTOR - TEST - PCALC: SELF (DETECTOR 2)
DETECTOR - TEST - AIMD: SELF TEST (DETECTOR 1)
DETECTOR - TEST - AIMD: SELF TEST (DETECTOR 2)
DETECTOR - TEST - PI: SELF TEST (DETECTOR 1)
DETECTOR - TEST - PI: SELF TEST (DETECTOR 2)
DETECTOR - TEST - PI: ADC RAMP (DETECTOR 1)
DETECTOR - TEST - PI: ADC RAMP (DETECTOR 2)
DETECTOR - TEST - COIN: SELF TEST **
DETECTOR - TEST - EVENT CONTROLLER (DETECTOR 1)
DETECTOR - TEST - EVENT CONTROLLER (DETECTOR 2)

GANTRY - CALIBRATE DIAGNOSTICS


GANTRY - CALIBRATE - C-FRAME (DETECTOR 1)
GANTRY - CALIBRATE - C-FRAME (DETECTOR 2)
GANTRY - CALIBRATE - CAUDAL TILT (DETECTOR 1)
GANTRY - CALIBRATE - CAUDAL TILT (DETECTOR 2)
GANTRY - CALIBRATE - COLLIMATOR CART OFFSETS
GANTRY - CALIBRATE - LEFT/RIGHT
GANTRY - CALIBRATE - PHS AUTO HEIGHTS
GANTRY - CALIBRATE - PHS IN/OUT
GANTRY - CALIBRATE - PHS UP/DOWN BACK
GANTRY - CALIBRATE - PHS UP/DOWN FRONT
GANTRY - CALIBRATE - RADIAL 1
GANTRY - CALIBRATE - RADIAL 2
GANRTY - CALIBRATE - ROTATE

GANTRY - DISPLAY DIAGNOSTICS


GANTRY - DISPLAY - CALIBRATED AXES STATUS
GANTRY - DISPLAY - CAUDAL TILT TYPE
GANTRY - DISPLAY - COLLIMATOR CART TYPE
GANTRY - DISPLAY - COLLIMATOR INFORMATION (DETECTOR 1)
GANTRY - DISPLAY - COLLIMATOR INFORMATION (DETECTOR 2)
GANTRY - DISPLAY - SYSTEM STATUS
GANTRY - DISPLAY - VERSION REPORT Page 107 of 232
STANDARD DIAGNOSTICS FOR esoft WORKSTATIONS
(continued)

GANTRY - EXERCISE DIAGNOSTICS


GANTRY - EXERCISE - C-FRAME (DETECTOR 1)
GANTRY - EXERCISE - C-FRAME (DETECTOR 2)

GANTRY - HOME DIAGNOSTICS


GANTRY - HOME - ALL AXES
GANTRY - HOME - GANTRY LEFT/RIGHT
GANTRY - HOME - PHS IN/OUT
GANTRY - HOME - PHS UP/DOWN
GANTRY - HOME - PHS UP/DOWN BACK
GANTRY - HOME - PHS UP/DOWN FRONT
GANTRY - HOME - RADIAL 1
GANTRY - HOME - RADIAL 2
GANTRY - HOME ROTATE

GANTRY - LOAD DIAGNOSTICS


GANTRY - LOAD - CAUDAL TILT TYPE
GANTRY - LOAD - COLLIMATOR CART TYPE
GANTRY - LOAD - COLLIMATOR INFORMATION
GANTRY - LOAD - LEFT/RIGHT AXIS FLAG
GANTRY - LOAD - NUMBER OF HEADS
GANTRY - LOAD - PROFILE OPTION
GANTRY - LOAD - RECONFIGURATION ANGLE
GANTRY - LOAD - RECONFIGURATION LATCH TYPE
GANTRY - LOAD - RECONFIGURATION MASK
GANTRY - LOAD - ROOM SIZE *

GANTRY DIAGNOSTICS
GANTRY - TEST - MECP: SELF TEST
GANTRY - TEST - MED: SELF TEST (DETECTOR 1)
GANTRY - TEST - MED: SELF TEST (DETECTOR 2)

SNAC - DISPLAY DIAGNOSTICS


SNAC - DISPLAY - CONFIGURATION REPORT

SNAC - LOAD DIAGNOSTICS


SNAC - LOAD - DPC: RELOAD LUT's

* ONLY DISPLAYED WHEN THE PROFILE OPTION IS ENABLED IN THE SNAC


CONFIGURATION FILE

** ONLY DISPLAYED WHEN THE COINCIDENCE OPTION IS ENABLED IN THE SMAC


CONFIGURATION FILE

Page 108 of 232


ECAM – STARTING DIAGNOSTIC SESSION

ON AN “ICON” (MACINTOSH) SYSTEM

Follow the steps below to start a “Standard Diagnostic” session on an ecam Macintosh based
system.

1. Insert the ICON Software CD into the ICON (Macintosh) CD drive.

2. Open the ICON 8.5 CD and then open the following folders in the order listed:

• ecam
• ICONF
• Service
• Tools

3. Double click on the ecam Diagnostics icon labeled :

4. Select New Diagnostic Session from beneath the “File” menu item.

This will bring up a list of “Standard” diagnostics.

Page 109 of 232


STARTING A “STANDARD” DIAGNOSTIC SESSION ON
esoft WORKSTATION

01. Insert the ecam STANDARD PLUS CD-ROM into the SNAC computer
CD-ROM drive. (Make sure it is the correct revision) CD needed if
running “Standard Plus” diagnostics.
02. Verify that the esoft application is running by:

• The first way is to double click on the Internet Explorer icon on your
desktop. This will take you to the “Service Software Authentication”
window if it is still setup from the factory. Now you can enter in the
passwords you received from the Uptime Center. If someone has changed
the configuration of the Internet Explorer then you will have to use the
second way of launching the diagnostics.

• The second way of launching the diagnostics is thru the esoft application
software. Follow the steps listed below to get to the “Service Software
Authentication” window:

o You must start the esoft application in order to gain access to the
diagnostic package. Launch the “Start E.Soft” application using the
shortcut located on the esoft Workstation desktop.

o The E.Soft application should now be running. (See image below)

Page 110 of 232


03. From the esoft workstation computer under the “Options” menu select
“Local Service”.

Page 111 of 232


04. The following window will appear.

05. This is the step that requires you to enter a password to continue.

• You can always run the “STANDARD” diagnostics by leaving the


password fields blank on the “Service Software Authentication” page.
• If password is known, select the check box called “Set as Default”. This
will cause the first fourteen digits of your password to be saved for future
diagnostic sessions. You will only have to key in the last six digits for your
next diagnostic session. Press “OK” button to continue. If you keyed in
the correct password you will see the next screen. If not, check to see if
you keyed in all the digits as required.

Page 112 of 232


06. From the above window select “E.CAM Diagnostics” to gain access to the
diagnostic package.
07. When the following window appears press the “Yes” button to continue.

Page 113 of 232


• When the following window appears press the “OK” button to continue.
This indicates that you have established a connection with the “Diagnostic
Server” and it is running.

08. From the Window below, select “All”.

09. From the window below, select “Open”.

Page 114 of 232


• You will then get a status bar “Connecting to Diagnostics”.

10. Once “Connecting to Diagnostics” is completed, you will see a Diagnostics


window similar to the one below.

Page 115 of 232


11. Make sure you “click & drag” at the bottom of the window pictured above, to
open the diagnostics status windows. Your Diagnostic window should now
appear like the window pictured below.

11. You can now scroll down through the diagnostic test listing until you find
the one you wish to run. Click once on the diagnostic you wish to run to
highlight it, the click on “Execute”. Make sure you follow the
instructions and read the warnings for the diagnostic you are running.

FEATURES

• ALL = All diagnostic test will display in the Selections window.


• SNAC = SNAC specific test will display in the Selections window.
• GANTRY = Gantry specific tests will display in Selections window.
• DETECTOR = Detector specific test will display in Selections window.

Page 116 of 232


HOW TO RUN A DIAGNOSTIC TEST : EXAMPLE

CONFIGURATION REPORT (DETECTOR 1) (DETECTOR 2)

This test commands the AIMD(C)(CF), PCALC(CF) and PI59(C)(CF) boards to


report their switch settings and firmware versions.

01. Enter the diagnostic package.


02. Select the desired Configuration Report and select Execute.
03. Check reports against entries in the Service Log.

• This diagnostic generates error message F1

o All switch settings and firmware versions must be as stated in the


Service Log. If not, contact Technical Support to verify the proper
switch settings and firmware versions. Be sure to set all board switch
settings correctly and to update the Service log if necessary.

Page 117 of 232


NOTES:

• Deselect the test(s) before selecting a new test.


o (Use the “Deselect All” button)
• Make sure when you have completed your diagnostic session, that you
“QUIT” the session and verify it is no longer running.

Page 118 of 232


CALIBRATE C-FRAME

DETECTOR #1

This procedure allows the calibration of the Collimator C-Frame on detector #1.

01. Power up the ecam system.


02. Home the ecam when prompted.
03. Unload the collimators from the ecam.
04. Rotate the system so that detector #1 is at 90o.
05. Start diagnostic session.
06. Select “New Diagnostic Session” from the “FILE” menu.

CAUTION:
• When a new position pot is installed, or in any case when the pot position
is unknown, leave the gear set screw loose on the pot, so the pot will not
be damaged. This is a ten turn pot and 9.5 turns are used in the range of
the C-Frame motion from open to closed. If the gear set screw is tight on
the pot, the pot could easily be damaged by the running past its end point.
The Calibrate C-Frame diagnostic does not depend on the pot value for
opening and closing the C-Frame.

07. Select the Gantry – Calibrate C-Frame (Detector 1) diagnostic.


08. Press the Execute Selections button.
• Status:
i. Gantry – Calibrate C-Frame (Detector 1).
ii. Press any PPM button when your ready to run test.

09. Press a PPM button.


• Prompt
i. The test is starting. Detector 1 C-Frame will move to its
open then its closed position. Press return to start motion.

NOTE:
• If Calibrate C-Frame initially fails with the error message “Collimator
partially installed” it may be because the Collimator Cleared Switch is
activated by the tub. It will be necessary in this case to open the C-Frame
enough to deactivate the Collimator Cleared Switch. Open the C-Frame
on one of two ways:
o Run Exercise C-Frame. The C-Frame opens and the diagnostics
pause. At this point, click the Close box in the upper left corner of
the diagnostic window. The diagnostics will exit with the C-Frame
open. Then run Calibrate C-Frame.
o Use the MEDP switches (see MEDP service map) to open the
C-Frame enough to deactivate the Collimator Cleared Switch.
Then run Calibrate C-Frame.

Page 119 of 232


10. Press return
• Status
i. MED1 Actual Pot Value: xx
ii. Acceptable Range: 12 through 16

• Prompt
i. Calibrate and lock down the potentiometer.
ii. Press return to continue.

11. Adjust the potentiometer with a 1/16” allen wrench to the expected value,
lock it down and press return.
12. C-Frame opens.
• Prompt
i. Verify the C-Frame is in the open position.
ii. Enter Y or N

13. Enter “Y” and press return.


• Status
i. Pass
ii. Test Sequence Completed

14. C-Frame closes. END.

NOTE:
• When you have completed with diagnostics, you must EXIT and CLOSE
the diagnostic session.

WARNING!
When you enter into and while you are running diagnostics,
ALL SYSTEM FAIL SAFES are DISABLED!

Page 120 of 232


Page 121 of 232
SETTING UP EQUIPMENT and SOURCE MATERIAL for

PEAKING, TUNING AND

INTRINSIC FLOOD on ecam CAMERA SYSTEM

Purpose:
• The following steps and procedures will assist you in setting up the ecam camera system for
preparing the source for Peaking, Tuning and performing an Intrinsic Flood. Always have an
authorized site person prepare the source for you, but you can request the way it should be
prepared and the amount of activity needed.

Preparing the Source:

NOTE:
• Have authorized site personnel prepare the source as described below. Use as little liquid as
possible and ensure that the point source’s activity is correct using the attached table.

01. Place a small amount of cotton into the bottom of the plastic vial. Be sure not to use too much
cotton. Cotton should cover the “bottom” of the vial only and not fill the vial.
99
02. Place the drops of Tc on the cotton. Do not exceed the cotton’s saturation capacity.
99
03. Use the chart below to determine the amount of Tc activity needed for calibration to be
performed.

SOURCE ACTIVITY USAGE FOR ecam


CALIBRATIONS
CALIBRATION ACTIVITY (uci)
Peaking 15 - 20
Tuning 20
MHR/COR 900 - 1100
NCO 900 - 1100
Intrinsic Flood 15 - 20

Setup Point Source:

NOTE:
• A deadtime between 3% and 9% is recommended.

01. Remove the collimator(s).


02. Fully retract the detector(s).
03. For a Dual-Detector system, rotate the gantry so Detector 1 is at 0o and Detector two is at 180o.
o
For a Single-Detector system, rotate the gantry so detector is at 0 .
04. Extend the integrated source holder from its storage position on the rear bed and pull out until the
source holder is centered.
05. Have the point source prepared as described.

NOTE:
• Before performing step 6, you might wish to cut a finger out of a latex glove and put your prepared
source inside the finger. This will help prevent possible leakage and crystal contamination.

06. Clamp the capped end of the vial into the source holder and ensure that the cotton tip with the
activity is approximately centered above the detector.
07. For a Dual-Detector system, raise or lower the bed until the source is centered between the
detectors. The count rates shown on the PPM display should be approximately equal.

Page 122 of 232


Using the Spectrum:

NOTE:
• Make sure there is no stray or scatter radioactivity in the department or system room. (With
collimator(s) removed and no source, you should see no more than 400cps.) An injected patient or
an unshielded dose, even if they are in different rooms, may produce a sufficient amount of scatter
activity to cause peaking and tuning problems. (More likely to present tuning issues.) Without
collimator(s) installed, the camera is significantly more sensitive than with the collimator(s)
installed. A typical patient is injected with several mci of activity, which is 10-100 times more than
the amount of activity used for peaking and tuning. Simply put, an injected patient in another room
or if that patient even just walks by the system room, can have a negative effect on peaking and will
negatively effect tuning causing tuning problems or tune failure.

NOTE:
• The Spectrum display is located or shown on the Analyzer tab.

01. Select the particular detector to view its spectrum.


02. Click on the “Start” button in the Spectrum field.
• A graph is displayed and shows:
i. Range of energy values, X axis.
ii. The number of gamma rays falling within the given energy value Y axis.
iii. The width and center of selected energy windows.
03. To stop recording events, click on the “Stop” button.
04. To clear the Spectrum graph or display, click on the “Clear” button.

NOTE:
• The spectrum display will stop growing when it reaches approximately 32,000 counts. At this point
you will need to press the “Stop” button as it does not automatically become disabled. Press the
“Start” button to restart the spectrum.

Peaking the Detector(s)

NOTE:
• When you peak the detector, you are asking the system to slightly shift the analyzer’s window so
that it is centered around the energy peak as detected by the camera. Only one detector is peaked
at a time. If you have a Dual-Detector system and both Detector 1 and Detector 2 are activated, it
will peak Detector 1 then peak Detector 2. During peaking, you will notice a loss of activity on the
PPM display for the detector being peaked.

01. Set up the source as described.


02. Select the isotope being peaked.
03. Click on the “Begin Peaking” button on the Analyzer tab. The system peaks the activated
detector(s) according to the Analyzer values.
04. When the system completes the peaking process, it automatically enters the appropriate status in
the “Peak Status” fields and peak shift values in the “Peak Shifts” (%) fields.
• Peak Status – this field indicates whether the peaking process for the detector was
completed successfully; (field reads “PEAKED”); or not completed successfully; (field
reads “NOT PEAKED”).
• Peak Shifts (%) – this percentage represents how far the system had to shift the detected
isotope photopeak to make it comply with established analyzer values.
• Peak Shifts smaller than +3.0% are acceptable. The detector(s) are capable of working
correctly with shifts up to +12.8%.

NOTE:

99 57
It is recommended to “Fine Tune” the detector(s) if the Tc or Co shift is larger than +3.0%. This
will ensure the stored Peak Shifts for other isotopes remains accurate.

Page 123 of 232


Tuning the Camera

Tuning the detector helps you tune the photomultiplier tubes, (PMTs), automatically inside each detector.
“Fine Tune” adjust the PMT gains slightly, while a “Coarse Tune” provides major adjustments of PMT gains
and is not normally required. The available gain range for tuning each PMT is –512 to +511 and the HV
range is –128 to +127. Only one detector can be tuned at a time.

Fine Tune:

01. Set up the source as described.


02. Peak the system, (target detector), for 99Tc as described.
03. Open the “Tuning Workflow” template and examine the Tuning page.
04. Select the detector being tuned.
05. Click on the “Start Fine Tuning” button to begin the tuning process.

NOTE:
• The processor will tune the PMTs in six or less steps which are referred to as “passes” in the
DAC/PMT window’s status field. These passes begin at six and work their way down to one. The
passes then start over at six and then count a second time. During these passes, the processor
adjusts the PMT gains so that the tune values are less than 7, (read 0.7%), for the center PMTs
and 15, (read 1.5%), for the PMTs on the edges. Tune values represent the amount of change
required to perform optimally.

06. The system will stop tuning over the course of its maximum of 12 passes.
07. To display the tune values, click on the “Display Tune Values” button. The value assigned to each
PMT after the current pass is displayed in the field above each PMT number.

NOTES:
• When all the tune values are less than 7, (read 0.7%), for the center PMTs and 15, (read 1.5%), for
the PMTs on the edges, tuning has completed successfully.
• The acquisition controller displays an error message if the system is unable to complete tuning
successfully. (If so, follow the instructions given in the Tuning window’s message field.) If tuning
should fail, it may be necessary to exit the Tuning page, reopen it, then redo the tuning procedure.

08. Peak the system once again as described.


09. When you have completed viewing or saving information, click the “Activity Done” button to end the
Tuning activity and continue with the workflow.

Coarse Tune:

01. Proceed as instructed for “Fine Tuning” the detector with the exception of step 5; click on the “Start
Coarse Tuning” button.
02. Click on the “Display Gain Values” button to display the gain values assigned to each PMT after the
current tune pass. The “High Voltage DAC” value is displayed which depicts how much DAC
output is needed to tune the detector. The gain value, the voltage applied to each PMT to bring it
into tune, appears in the field above each PMT.
03. After a Coarse Tune has been performed successfully, always perform a Fine Tune.

NOTES:
• If a “Coarse Tune” is performed or the DACs are cleared, all analyzer isotope presets will have to
be repeaked.
• If the PMT DACs are cleared, a “Coarse” and “Fine” Tune must be performed as described.

Page 124 of 232


Rear Bed

Dual Rod
System

Source Vial

Old Style ecam Source Holder

Rear Bed

Single
Rod

Source Vial

New Style ecam Source Holder

Page 125 of 232


Vial Cap

Tc99

Vial

Cotton or Guaze

Source Preparation

Quality Control & Calibrations Schedule

Frequency/Problem Calibration
99 57
Daily Peaking for Tc or Co
Fine tune, if peaking value is larger than +3.0%, then re-peak.
(NOTE: This value becomes +2.0% if on a Profile system.)
Daily Intrinsic Flood QC
Monthly Peak
Tune
Peak
MHR/COR for each collimator set used for tomo.
NCO Calibration (dual-detector variable-angle and cardic systems only)
Monthly Intrinsic Flood >200 million counts
Whenever Collimator Peak
Damage Occurs Tune
Peak
Monthly Intrinsic Flood >200 million counts
Monthly Extrinsic Flood for the damaged collimator.
Calibration Sequence - The proper sequencing of calibrations is as follows:
1. MHR/COR
2. NCO (Dual-Detector variable-angle and cardiac systems only)
3. Intrinsic Flood (Daily/Monthly)
4. Extrinsic Flood (if collimator damage occurs)

Page 126 of 232


Page 127 of 232
Esoft WORKSTATION

REGISTER/CREATE PATIENT PROCEDURE

01. With esoft running, under “Patient”, select “Register”.

Page 128 of 232


02. “Patient Registration” page is displayed.

03. There are four mandatory fields, (in bold), which must be completed or filled out.
a) Last name (For our purpose, we used “ECAMTEST”)

Page 129 of 232


b) Patient ID (For our purpose, we used “001”)

c) Date of Birth (For our purpose, we just entered the current date.)

Page 130 of 232


d) Sex (For our purpose, we just entered “Male”.)

e) All other information can be left blank for running test images.

04. If you are now ready to perform test study, click on “Exam”.

Page 131 of 232


05. This will return you to the “esoft” page with your created patient already selected
where you can now select the type of study you would like to perform.

Page 132 of 232


Page 133 of 232
ECAM – EXAMPLE WBA STUDY SETUP PROCEDURE

Camera-Patient Setup_01

Camera-Patient Setup-02

DETECTOR 1 AT 00

ARMREST ENGAGED

PATIENT

PALLE
T IN
REAR BED DOWN

DET
ECT
OR 2
AT 1
80 0

Page 134 of 232


Select Patient

Select Study Category

Page 135 of 232


Select Wholebody

Select Example Wholebody

Page 136 of 232


Check Patient Name

Select Stop Conditions

Page 137 of 232


Select Enter Scan Length

Enter Scan Speed

Page 138 of 232


Scan Speed Entered

Select Autocontour

Page 139 of 232


Select Passes Equal One

Select Analyzer and Isotope

Page 140 of 232


Verify Detectors and Peaking

VERIFY BOTH
VERIFY BOTH
DETECTORS
DETECTORS
ARE ACTIVE
ARE PEAKED

Select Prepare Acquisition

Page 141 of 232


Select Acquisition Start

WBA Scanning_01

BOTH DETECTOR
PATIENT RADIUS MOVED PALLET
IN TOWARD SCANNING OUT
CENTER

LEFT
LOOK AHEAD

Page 142 of 232


WBA Scanning_02

DETECTOR 1

LEFT
LOOK
AHEAD

PATIENT

WBA Scanning_03

DETECTOR 1

C-FRAME LEFT RIGHT


LIGHT LIGHT
RAIL PATIENT RAIL

COLLIMATOR

Page 143 of 232


WBA Scanning_04

DETECTOR 1

COLLIMATOR

PATIENT

WBA Scanning_05

Page 144 of 232


Scanned Image View

WINDOWING
BAR

DETECTOR 2
DETECTOR 1 IMAGE
IMAGE

Scan Compete – Select D/A (Display Analysis)

Page 145 of 232


Select Pause Workflow

Workflow Stopped

Page 146 of 232


Close Window

Exit Workflow – “YES”

Page 147 of 232


Back to Patient Database

Page 148 of 232


Page 149 of 232
Page 150 of 232
ECAM – HD3 C-FRAME REMOVAL PROCEDURE

This procedure will assist you in removing the C-Frame assembly. (It will be necessary to
remove the “C-Frame” if you have a need to put the detector into the “Detector Service
Position”.) This procedure assumes the person performing it obtains the basic knowledge to do
so. It is recommended you thoroughly read and understand this procedure before actually
performing this procedure.

While performing this procedure, you must understand, follow and adhere to all safety
guidelines and procedures concerning, however not limited to the following: personal,
equipment, environmental, electrical and radioactive.

Items Needed:
• Your service tool kit.
• C-Frame Closed Jumper (Siemens part number 5223248) (See Figure: 1)

• CAUTION: While system is in “diagnostic mode”, there are “NO” system fail-safes.
• NOTE: System must be “homed” and in the 180o configuration to perform this procedure.
(See Figure: 2)
• NOTE: This procedure assumes “Standard” diagnostics is used for some removals as
stated in this procedure.

01. Remove the detector processor cover from the detector being serviced to gain access to
the C-Frame. (See Figure: 3)
02. Rotate the gantry so the detector being serviced is at 90o or 270o, or the detectors are in
the collimator change positions. (See Figure: 4)
03. Remove the collimator from the detector being serviced.

• CAUTION: When removing the 5mm hex bolt and stop washers from the C-Frame
bearing posts in Step 4, you must take note and record any washers which are
removed. Some C-Frames have “shims”, (large and small), used for C-Frame
alignment. These “shims”, (washers), “must” be replaced on the same posts from
which they were removed.

04. Use a 5mm hex wrench to remove the stop washers from the C-Frame bearing posts.
(See Figure: 5)
05. Enter into a “Standard” diagnostic session.
06. Perform diagnostic “Exercise C-Frame”, (for the detector or C-Frame being serviced), to
“OPEN” the C-Frame.
07. Disconnect J10, J11, J12 and J13 ribbon cables from the MED/U board. Take care
when disconnecting these cables, the connection is “key locked” and make sure these
cables are not even slightly bent or twisted; these cables are very easily damaged.
(These are the ribbon cables from the “Light Rails” and can be disconnected with power
applied.) (See Figure: 5)
08. Disconnect J4 from the MED/U. Take care when disconnecting this cable, the
connection is “key locked”. (Connection J4 can be disconnected with power applied.)
09. Loosen the tie wraps securing the ribbon cables to the C-Frame motor.
10. Use a 2.5mm hex wrench to remove the C-Frame “lock plate” from the rear of the
C-Frame. (See Figure: 6)

Page 151 of 232


11. Using extreme caution not to damage any cables, carefully remove the C-Frame
assembly. (See Figure: 7)
12. If motion is required after the C-Frame has been removed, perform the following:
• Install the “C-Frame Closed Jumper” at J4 on the MED/U board. (See Figure: 1
& Figure: 8)
• Properly exit the “Standard” diagnostic session.
• Press the “Reset” button on the MED/U board. (See Figure: 9)

• CAUTION: Carefully observe that: all cables clear without getting pinched; if
removed, the CIDI cable does not get caught on the detector flange; and the C-
Frame motor extension arms feed into the holes of the detector flange without
binding.

13. Reassemble in reverse order.


14. Perform a collimator change.
15. Verify the system can perform a 90o “auto contour” SPECT study.

END

Page 152 of 232


Page 153 of 232
Page 154 of 232
Page 155 of 232
Page 156 of 232
Page 157 of 232
Page 158 of 232
Page 159 of 232
Page 160 of 232
Page 161 of 232
Page 162 of 232
Page 163 of 232
Page 164 of 232
Page 165 of 232
Page 166 of 232
Page 167 of 232
Page 168 of 232
Page 169 of 232
Page 170 of 232
Page 171 of 232
Page 172 of 232
Page 173 of 232
Page 174 of 232
Page 175 of 232
Page 176 of 232
Page 177 of 232
Page 178 of 232
Page 179 of 232
ECAM – HD3 DETECTOR SERVICE POSITION PROCEDURE

This procedure will assist you in putting the ecam detector into the “Detector Service Position”.
In the “Detector Service Position”, you are removing the inner workings of the detector, (crystal,
optical spacer, PMTs, preamps and inner cables from the detector tub. It is necessary to place
the detector in the “Detector Service Position”, as an example, if it is necessary to remove and
replace a PMT. This procedure assumes the person performing it obtains the basic knowledge
to do so. It is recommended you thoroughly read and understand this procedure before actually
performing this procedure.

While performing this procedure, you must understand, follow and adhere to all safety
guidelines and procedures concerning, however not limited to the following: personal,
equipment, environmental, electrical and radioactive.

Items Needed:
• Your service tool kit.
• Four pieces of “2x4” straight lumber cut to approximately 24” to 27” lengths. (Two
pieces of “4x8” can be used if that is all, which is available). A clearance of 4” to 6” is
required.
• C-Frame Closed Jumper (Siemens part number 5223248)

• CAUTION: While system is in “diagnostic mode”, there are “NO” system fail-safes.
• NOTE: System must be “homed” and in the 180o configuration to perform this procedure.
• NOTE: It is recommended before beginning with this procedure if not already, the
lightest weight collimators available on site be installed.
• NOTE: This procedure assumes “Standard” diagnostics is used for some removals as
stated in this procedure.

01. Remove the “detector processor” cover from the detector being serviced.
(See Figure: 1)
02. Rotate the detector being serviced to the 90o position as facing the front of the gantry.
(See Figure: 2)
03. Start a “Standard” diagnostics session.
04. Perform the diagnostic “Exercise C-Frame” and open the C-Frame for the detector being
serviced.
05. Remove the collimator for the detector being serviced. (See Figure: 3)
06. With C-Frame still open and in diagnostics, remove the C-Frame assembly.
(See Figure: 4 & Figure: 5)
07. With C-Frame removed and the C-Frame drive mechanism still open; install the “Closed
Jumper”, (Siemens #5223248), on the MED/U board at J4. (You will have to remove the
connector already installed. Safe to do this under power.) (See Figure: 6 & Figure: 7)

• NOTE: By installing the “C-Frame Closed Jumper” at J4 on the MED/U


board simply fakes the system to think or recognize that the C-Frame is closed
when it is open.

08. Properly quit the diagnostics.


09. Press the “reset” button on the MED/U. (See Figure: 8)

Page 180 of 232


• NOTE: In extreme cases, where the procedure has not be followed as stated, it may be
necessary to press the “Reset” button on the MECP board to gain motion from
the hand controller.

10. At this point, you should have motion from the hand controller. If safe, check to see by
attempting to rotate a couple of degrees and by moving radius of detector being serviced.
11. Turn off detector voltage. (See Figure: 9)
12. Turn off detector HV. (See Figure: 10)
13. Disconnect the HV BNC cable. (See Figure: 11)
14. Disconnect the ribbon cable from the CIDI board. (See Figure: 12)
15. Rotate the detector being serviced to the “top”, or at the 0o position. (See Figure: 13)
16. Remove the “pallet” from the “Front Bed”.
17. Lock down the “Front Bed” docker micro switches. (This can be done with tape or by
using a tie-wrap taking care not to damage the switch.) This must be done in order to
fake or make the system think or recognize the “Front Bed” is docked. No motion is
allowed if the system finds the “Front Bed” is not docked.

• CAUTION: When using the “Front Bed” or “PHS” to perform system servicing
procedures, you must not exceed 600 lbs. of weight or force on the “Front Bed” or
“PHS”. Reaching or exceeding this limit will cause damage to the “Front Bed” or
“PHS”.

• CAUTION: When placing or moving the “Front Bed” parallel to the gantry, be
aware the bottom part of the “Front Bed” is not the lowest part, it is the “support
frames” on either side underneath which are much lower. If necessary, use the
“Hand Control” to adjust the height of the “Front Bed” so it does not contact the
bottom or detector not being serviced.

18. Place the “Front Bed” parallel to the gantry. (See Figure: 14)
19. Slowly push the “Front Bed” between the detectors until the “Front Bed” stops travel.
(Travel will stop when the bottom front feet, (left and right side), of the “Front Bed” come
in contact with the front feet of the gantry, (left and right side).
20. Place two stacks of lumber, (each containing two “2x4”), on the “Front Bed”.
(See Figure: 15)
21. Raise the “Front Bed” to within 1” of maximum height or until lumber is within just less
than ½” of the detector tub. (If necessary, you may have to radius “IN” the detector
being serviced.)
22. Align the lumber so as the edges of the detector tub when lowered, will be resting on the
lumber leaving room to remove the 3 hex bolts on each side mounting the detector to the
detector tub. (See Figure: 16 & Figure: 17)
23. Before placing the detector on the “Front Bed” or lumber; perform the following:
• Lower the detector being serviced to within just less than ½” of the lumber blocks.
• Six M4 cap screws mount the detector to the tub. Use a 3mm hex wrench and
evenly loosen the six cap screws 3 turns CCW, starting first with the left side,
then the right side. This is to create a 2mm air gap between the detector and the
tub.

• NOTE: “Caudal Tilt” function should be active. If for some reason “Caudal Tilt” function
is not active at this time, on the MED/U board, set SW2 positions 6 & 7 to “ON”, press
the MED/U reset switch and wait for tone. When tone is heard, “Caudal Tilt” function
should be active. (See Figure: 18)

Page 181 of 232


24. If necessary, use the “Caudal Tilt” switch to tilt the detector “front-to-back” so that the
detector is parallel with lumber placed on the “Front Bed”.
25. If necessary, use the “Hand Control” to rotate the detector so that it is “level” with the
lumber from “left-to-right”.
26. Double check and make sure the detector is “square”, “level” and “parallel” with the
lumber before proceeding. Make necessary adjustments.
27. Using the “Hand Control”, (by either lowering the detector or raising the “Front Bed”), so
the detector is parallel with the lumber and resting on the lumber until the “air gap” just
begins to close. “STOP” all motion when the “air gap” just begins to close.
28. Remove the six, (three each side of detector), M4 3mm cap screws. (These screws
require approximately 15 turns CCW to completely remove.)

• CAUTION: To prevent damage to the detector and detector flange, make sure and
observe while lowering the detector from the tub, the detector does not “hang up”
or “bind”, and the detector and detector flange separates evenly “front-to-back”
and “side-to-side”.

• NOTE: It is recommended, before disconnecting any cables, you make at least a mental
note of placement and folds of all the cables. Also recommended, before disconnecting
any cable, if not marked or tagged where goes, you do so before disconnecting as well
as marking or tagging any cables that might have a “tie-warp” so these can be replaced
as found before reinstalling the detector. (You also might wish to note the “number” of
cables disconnected.) To avoid cable damage and/or system damage, it is critical
all cables be properly routed, tie-wrapped and connected before reinstallation of
detector.

29. Using the “Hand Control”, continue lowering detector until there is enough clearance,
(making sure you do not lower too far or far enough to “stress” any cable or connector),
to access inner detector cables so they can be disconnected. This would include, not
limited to, all board cable connections, board coax connections and fan cable
connections.
30. Observe that the disconnected CIDI cable, (at rear of detector tub), does not get caught
or bound on the detector flange openings.

• CAUTION: While lowering the inner detector assembly for the detector tub,
observe that you have plenty room between the “Front Bed” and the bottom
detector.

31. Once you have verified all inner head cables have been disconnected and will not bind,
slowly and carefully continue to lower the inner detector until it clears the detector tub.
32. When you have verified proper clearance between the inner detector and the detector
tub, you may now slowly and carefully move the “Front Bed” out to gain access to the
inner detector assembly for service.
33. Once you have completed necessary service, reverse this sequence to reinstall the inner
detector into the detector tub.

• CAUTION: While reinstalling inner detector into detector tub, be sure to observe at
all times, all cable clear without getting pinched, the CIDI cable does not get
caught on the detector flange and the C-Frame motor extension arms feed into the
holes of the detector flange without binding. (See Figure: 19)

Page 182 of 232


• NOTE: Before reinstalling the inner detector assembly into the detector tub, double
check and verify all cables and connections are routed correctly and properly connected.
This includes the 59 HV wires and the 59 data cables. (See Figure: 20)

34. Using the “Hand Control”, move the inner detector assembly into the detector tub.
“STOP” all motion when the inner detector is within 2mm of the detector tub.
35. Using a 3mm hex wrench, tighten the six, (three each side), M4 cap screws, first the left
side and then the right side evenly until all screws are tight.
• NOTE: If you had to activate “Caudal Tilt” using SW2 on the MED/U board, then now on
the MED/U board, set SW2 positions 6 & 7 to “OFF”, press the MED/U reset button and
wait for the tone. When you hear the tone, continue with procedure.

36. Remove the “Front Bed” from in front of the gantry.


37. Reconnect the CIDI board cable and HV cable.
38. Turn the HV back on.
39. Turn the detector power back on.
40. At this point, it is recommended you flood the detector serviced using a point source and
observe there are no “holes” in the flood.

• NOTE: If at this point you “do” observe a “hole” in the flood, this generally points to either
an HV wire or data cable is either improperly connected or not connected. This requires
putting the detector back into the “Detector Service Position” and verifying connections.

• NOTE: If at this point you observe “no” counts or activity from the detector, which was
serviced, and the detector power is on and supplies have been verified, this generally is
due to a loss of “detector communication”. To attempt to reestablish or restore detector
communications, first attempt a “soft reboot” of the SNAC computer from the PPM. If
this does not work, properly shutdown the ecam system and then properly re-apply
power to the ecam system. After reestablishing detector communications and you have
counts or activity, make sure no objects are in the way of the gantry and “home” the
system; then continue with procedure.

41. Remove the tape or tie-wrap or what was used to close the “Front Bed” docker
assembly’s micro switches, then store or dock the “Front Bed” in the “storage” position.
42. Reinstall the C-Frame assembly, (reassemble in reverse order).
43. With the C-Frame Closed Jumper removed from J4 on the MED/U board and the original
connector install at J4, press the MED/U reset button.
44. “Home” the system.
45. Perform the “Fan Current” check procedure to insure all fans are running inside the
detector.
46. Verify proper collimator change sequence.

END

Page 183 of 232


Page 184 of 232
Page 185 of 232
Page 186 of 232
Page 187 of 232
Page 188 of 232
Page 189 of 232
Page 190 of 232
Page 191 of 232
Page 192 of 232
Page 193 of 232
Page 194 of 232
Page 195 of 232
Page 196 of 232
ECAM – REMOVING THE GELLED TUBE

This procedure explains the basic steps required to remove a “gelled” tube from a Siemens
ecam detector. This procedure applies to both 2” and 3” PMTs. This procedure also applies to
all digital detectors using gel to bond the PMTs to the optical spacer. Also included, are basic
cautions required when performing this procedure. It is suggested you completely read and
understand this procedure before beginning.

While performing this procedure, you must understand, follow and adhere to all safety
guidelines and procedures concerning, however not limited to the following: personal,
equipment, environmental, electrical and radioactive.

The following item are necessary to replace PMT:

ITEM PART NUMBER


1mm Glass Beads 1784813
Teflon Scraper 567-000743
Blade Scraper 4358920
HD3/ASIC PMT Coupling Weight Kit 3529711
Gel for 2" PMT (6 months shelf life) 782-001004 (9489287)
Gel for 3" PMT (6 months shelf life) 782-001044 (1789960)
C-Frame Closed Jumper 5223248
New PMT Check Detector Log Book
or Detector Assembly
Parts list.
Lumber You will need four pieces
of 2x4", each cut
approximately 27" long.
NOTE: Gel packets are not inter-changeable. If a three inch packet
is used on a two inch tube, the tube cannot be removed and the
detector will have to be returned to the factory.

Removing the Gelled Tube

01. Place the detector in the “Service Position”.


02. Level the detector from “side to side” and “front to back”. (It is important the detector be
level).
03. Disconnect all detector head cable connected to the PI59(C)(CF)(R) and
DD59(C)(CF)(R) boards.
04. Remove the three Phillips screws holding the PI59(C)(CF)(R) board to the
DD59(C)(CF)(R) board. This will allow the PI59(C)(CF) board to pivot on its hinges.
05. Remove the four fans taking note the flow direction for re-assembly.
06. Remove the brackets around the DD59(C)(CF)(R) board.
07. Disconnect all detector cabling connection the tubes to the DD59(C)(CF)(R) board.
Remove the brackets holding the DD59(C)(CF)(R) board to the standoffs.

Page 197 of 232


CAUTION
• There are no longer separate tube shields. They are now spot-welded together as a
“nest”. Be careful not to damage a shield as this will result in detector
replacement.

NOTE
• Some tubes, ( 1,5,8,28,32,52,55 and 59 ) have capacitors attached from the PMT
preamp to the shield. Use a grounded soldering iron to remove the wire from the PMT
preamp. The wire is tacked to one leg of C103 or C4. (Refer to diagram below)

SOLDER

WIRE & SHRINK TUBING

C103 or C4
CAPACITOR
( CERAMIC DISK 4700pF +80/-20%
2KV )

PMT

SHIELD

Page 198 of 232


CAPACITOR CONNECTION PMT SHIELD / PREAMP
( TUBES 1,5,8,28,32,52,55 & 59 ONLY ) CAPACITOR CONNECTION
( TUBES 1,5,8,28,32,52,55 & 59 ONLY )
A
A
FE
ASIC FE
K ASIC
K
03
C1 D1

C1
D1

03
R112 R127
R112 R127

D8
D2 D8
D2
1
D7 1
D7

D6 D5 D4 D3
D6 D5 D4 D3
FF5 PREAMP BOARD
FF6 / FF7 PREAMP BOARD

PMT SHIELD / PREAMP


CAPACITOR CONNECTION
( TUBES 1,5,8,28,32,52,55 & 59 ONLY )

FE

U4 AMP K

C4
D1
U1 DAC
R13

D8
D2

1
D7
HV

D6 D5 D4 D3

PAD1 PREAMP BOARD

08. Apply pressure to the PMT in a forward direction to begin to release the gel seal. Then
apply reverse pressure to the PMT to begin to release the gel seal. Now apply pressure
to the right and then pressure to the left as in a “twisting motion”. Repeat as necessary
until the gel bond starts to release. (See diagram below)

Page 199 of 232


09. Slowly rotate and rock the PMT. Repeat until the gel bond is released. (See diagram
below)

10. Twist and pull up on the PMT, gripping the glass, until it is out of the shield.
11. Using a plastic scraper, very carefully remove all visible signs of the gel residue from the
optical spacer. Clean the optical spacer thoroughly, but do not use any solvents or
alcohol prep pads, which will prevent the new tube from bonding properly.

NOTE
• The detector must be level to ensure that the gel, after it is distributed, stays
evenly under the PMT.

END

Page 200 of 232


Page 201 of 232
Page 202 of 232
ECAM HD3 – INSTALLING NEW PMT

This procedure assumes you have followed the proper processes/procedures to include all
safety procedures for removing the bad or suspected PMT and are now ready to install the new
or replacement PMT. This being the case, proceed with the PMT installation process.

While performing this procedure, you must understand, follow and adhere to all safety
guidelines and procedures concerning, however not limited to the following: personal,
equipment, environmental, electrical and radioactive.

• IT IS RECOMMENDED YOU READ AND UNDERSTAND THIS PROCEDURE


BEFORE PERFOMRING!

Some items you will need:

ITEM PART NUMBER


1mm Glass Beads 1784813
Teflon Scraper 567-000743
Blade Scraper 4358920
HD3/ASIC PMT Coupling Weight Kit 3529711
Gel for 2" PMT (6 months shelf life) 782-001004 (9489287)
Gel for 3" PMT (6 months shelf life) 782-001044 (1789960)
C-Frame Closed Jumper 5223248
New PMT Check Detector Log Book
or Detector Assembly
Parts list.
Lumber You will need four pieces
of 2x4", each cut
approximately 27" long.
NOTE: Gel packets are not inter-changeable. If a three inch packet
is used on a two inch tube, the tube cannot be removed and the
detector will have to be returned to the factory.

Normal Process

01. Clean the new PMT face with a dry soft cloth. “DO NOT” use any solvents or cleaners
as these can cause improper bonding issues.
02. Remove the divider bar from one packet of gel and squeeze the contents in the upper
compartment into the bottom one. (Make sure you are using the proper gel packet for
the size PMT you are installing.) Gently knead the packet for approximately 60 to 90
seconds to thoroughly mix the contents. (While mixing gel packet contents, attempt to
eliminate at many air bubbles as possible.)
03. Pour the entire contents of the gel pack mixture onto the “center” of the optical spacer
pad. While pouring the gel onto the center area, “DO NOT” attempt to spread the gel,
let the gel spread itself.
04. Wait approximately 3 to 5 minutes for the air bubbles to escape.
05. Immediately distribute evenly as possible approximately 15 to 20 1mm glass beads in
the gel.
Page 203 of 232
06. Correctly orient and align the PMT just above the gel and set it onto the gel puddle. “DO
NOT ROTATE THE PMT!” Rotation of the PMT may cause the glass beads to move
out from under the PMT.
07. Place the proper weight base and weight on the PMT. (The weight kit includes bases for
round and hex preamps and weights for 2” and 3” PMTs. Both weight base and weight
“must” match the shape preamp and size PMT you are installing for proper bonding.)
Allow the PMT to set with weight for at least 4 hours at normal room temperature. If
room temperature is below 70oF, let the PMT set with weight for a longer time period;
approximately twenty minutes per degree for every degree below 70oF. After 4 hours,
(time could be longer depending on room temperature), remove the weight and carefully
begin re-assembly of the detector. (See Figure: 1 & Figure: 2) DO NOT INSTALL
DETECTOR IN TUB AT THIS TIME!
08. Certain PMTs, (1, 5, 8, 28, 32, 52, 55 and 59), have capacitors attached from the
preamp to the PMT shield or “MU Shield”. Solder the wire to the capacitor on the
preamp board if this is the case. (If you are replacing on of the above stated PMTs, this
capacitor wire should have been removed before removing the PMT.) (See Figure: 3)
09. Reassemble and re-cable the detector.

• DO NOT PUT THE REASSEMBLED DETECTOR BACK IN THE TUB AT THIS TIME

• Re-install HV wires by grasping the connector, not the wire. Make sure you install
HV wires as you removed them, oriented correctly if the connecting posts are
square.

10. Make sure HV is off and the HV coax cable is disconnected. Turn the detector on; apply
only “low voltages” to detector. Let the detector warm up for 10 minutes.

• There is no Preamp Zero Adjust for the PAD1 preamps.

11. Using a DVM, attach the positive lead to the test point on the DD59(C)(CF)(R) board for
the replaced preamp. (These test points are number as the PMTs; TP1 thru TP59.)
Attach the negative lead to TP67.
12. Adjust R112, Preamp Zero Adjust, (located on the preamp), for 0V, +500uV. (This
adjustment is only available on the FF6 and FF7 preamps.
13. Power down the detector.

• Once you have attached all HV wires, Data cables and connectors, it is
recommended you “double check” all HV wire, Data cables and connectors for
proper placement and secure connection. Also make sure all cables are “tie-
wrapped” as you found them and cables are positioned correctly for proper
“folding” when detector is placed inside the tub.

14. Replace the detector into the tub.


15. Reconnect the HV coax cable and turn HV on.
16. Apply power to the detector.

• If you see no counts or activity from detector in question, at this point, it may be
necessary to perform a “Soft Boot” of the SNAC from the PPM to reestablish
detector communications.

17. At this time, using a source, (15uci to 20uci), check to make sure you have no “holes” in
the flood.

Page 204 of 232


• If you see a hole in the area where you just replaced the PMT, this could be
caused by “invalid” DAC value. Try clearing the PMT DACs.
• If you see a hole in the area where you did not replace the PMT, this could be
caused either by a loose or uninstalled HV wire or data cable. At this point, you
would have to put the detector in the “Service Position” and check all HV wires
and Data cables.

18. Let the detector warm up for at least 60 minutes.


19. Perform Light Leak Check.
20. Launch the ICON / esoft software. Clear the PMT DACs.

• When the PMT Gain DACs are “cleared” or a “Coarse Tune” is performed, the
system no longer retains “peaking data”! As a result, it will be necessary for the
customer to “re-peak” all the isotopes they use. Once an isotope has been
properly peaked, the system will retain the peaking data for this isotope. It is vital
you make your customer aware of this.

21. Select the proper isotope preset. (You should be using 99Tc)
22. Peak the analyzer. (Peak shift percentage should be +3.0%)
23. Perform a Coarse Tune. When the Coarse Tune is completed, view the Gain Values
display map.
24. Inspect the PMT gain values for any value greater than 200 from the default value. (The
“default” value can be found in the “Detector Service Log Book” or the PMT gain value
all PMTs displayed after the DACs were cleared.)

• There are several different default values which is factory determined depending
on the gain of the PMTs used in the detector. An example default value would be
–209.

25. If any PMT gain value is found to be greater than 200 from the default value, the gain of
these PMTs will need to be adjusted.

• If PMT gain value adjustments are necessary, make adjustments accordingly to


achieve whatever default PMT DAC gain value your detector has.

• Example: If your PMT DAC gain default value is –209, given a 200 gain range value,
your PMT DAC gain value range after Coarse Tune would be –009 to –409.

26. If PMT DAC Gain adjustments where necessary, when completed, allow detector to
warm up for at least 30 minutes before proceeding.
27. Once detector has warmed up, clear the DACs, re-peak the system and perform Coarse
Tune.
28. If Coarse Tune passes and all PMT DAC Gain values are within specifications, continue.
If not, then go back to Step 20.
29. Perform Fine Tune.
30. Once Coarse Tune and Fine Tune have been performed and passed with all values in
specifications, perform a uniformity flood.
31. Perform “Post Repair Detector Testing”.

Page 205 of 232


3528986 450-009202
LARGE WEIGHT SMALL WEIGHT
FOR 3" TUBES FOR 2" TUBES

3528978 3529000
BASE FOR BASE FOR
NON-ASIC HD3 / ASIC
PREAMPS PREAMPS

BOLT ATTACHES
WEIGHT TO BASE

WEIGHT

3529703
EYE BOLT

PREAMP
TUBE

BASE
FITS OVER
PREAMP

Figure: 1

Page 206 of 232


BOLT
ATTACHES
WEIGHT TO
BASE

PMT WEIGHT

BASE

INSTALL NEW PMT, (WITH PREAMP), MAKING SURE PREAMP ORIENTATION IS THE
SAME AS ALL OTHER PMT’S AND PREAMP’S. ONCE PMT HAS REACHED THE GEL,
VERY SLIGHTLY APPLY DOWNWARD PRESSURE AND TWIST, (NOT MORE THAN ¼"
EITHER DIRECTION), JUST ENOUGH TO SEAT THE PMT IN THE GEL. (CAUTION: DO NOT
APPLY ENOUGH PRESSURE OR TWIST IN AN ATTEMPT TO SPREAD THE GEL. IF SO,
PMT WILL NOT BOND PROPERLY!) ONCE PMT IS SEATED, PLACE WEIGHT AND BASE,
(ASSEMBLY REQUIRED), ON TOP OF PMT/PREAMP MAKING SURE IT IS LEVEL.
(CAUTION: IF WEIGHT KIT NOT LEVEL, WILL CAUSE IMPROPER BONDING!) LEAVE
WEIGHT KIT ON FOR AT LEAST 3 TO 4 HOURS UNDISTURBED ALLOWING FOR INITIAL
SETUP TIME. AFTER THIS INITIAL SETUP TIME, THE WEIGHT CAN BE REMOVED AND
REASSEMBLY OF THE DETECTOR CAN BEGIN.
CAUTION: DO NOT ROTATE OR FLIP DETECTOR FOR AT LEAST 24 HOURS FROM THE
TIME THE WEIGHT WAS FIRST PLACED ON PMT!
(BE AWARE THERE IS A DIFFERENT WEIGHT FOR BOTH 2" AND 3" TUBES, AND A BASE
FOR BOTH HEX AND ROUND PREAMPS.)

Figure: 2

Page 207 of 232


CAPACITOR CONNECTION PMT SHIELD / PREAMP
( TUBES 1,5,8,28,32,52,55 & 59 ONLY ) CAPACITOR CONNECTION
( TUBES 1,5,8,28,32,52,55 & 59 ONLY )
A
A
FE
ASIC FE
K ASIC
K
03
C1 D1

C1
D1

03
R112 R127
R112 R127

D8
D2 D8
D2
1
D7 1
D7

D6 D5 D4 D3
D6 D5 D4 D3
FF5 PREAMP BOARD
FF6 / FF7 PREAMP BOARD

PMT SHIELD / PREAMP


CAPACITOR CONNECTION
( TUBES 1,5,8,28,32,52,55 & 59 ONLY )

FE

U4 AMP K
C4

D1
U1 DAC
R13

D8
D2

1
D7
HV

D6 D5 D4 D3

PAD1 PREAMP BOARD

Figure: 3

Page 208 of 232


Page 209 of 232
Page 210 of 232
ECAM – POST REPAIR DETECTOR TESTING

• It is recommended you first read and understand this procedure before


performing. Also remember to adhere to all personal, equipment, environmental
and radiation safety guidelines.

Tuning

01. Tune the detector(s). (Peak, Fine Tune, Peak)

Uniformity Flood

01. For each detector, acquire a 30 million count uniformity flood. Choose the appropriate
section:
a. ecam Systems with ICON Computer
b. ecam Systems with esoft Workstation Computer

ecam Systems with ICON Computer

01. Remove the collimators.


02. Place the detectors in the “collimator change” position, (Detector 1 at 90o and Detector 2
at 270o).
03. Using “Caudal Tilt”, turn the detectors so that the crystals face away from the gantry or
out.
04. Use a 57Co or 99Tc source, (you will probably be using 99Tc which is just fine), that
produces a dead time of 3%-6%. Make sure that the source is 5 UFOV lengths from the
detector, (130 to 136 inches from the crystal face).
05. Place the selected source onto a source stand. Adjust the height of the source so it is
the same distance from the floor as the center of the crystal in question.
06. Center the source “Left-to-Right” on the crystal.
07. Flood the crystal.
08. From the ICON Desktop, double click on the “Acquisition Protocols” icon. The
Acquisition Protocols dialog box will be displayed. Click on the “Static” button. Select
“Example Static”, and then click “Open”. The “Patient Database” will be displayed.
09. Pull down the “Patient” menu item and select “New Patient”.
10. Select the ICON destination volume, (usually the computer hard drive); by clicking “OK”
button.
11. Enter the “Intrinsic Flood” for the patient name.
12. Enter the date for the “Patient ID”.
13. Click “Accept” box.
14. The “Patient Database” will display, click “OK”.
15. The “Acquisition Predefined” window will be displayed.
16. On the “Acquisition Predefined” window, click the “Parameters” button in the upper left
corner.
17. Select the “Camera Parameters” page and set “Zoom” to 1.00. (If the “Uniformity on the
Fly” feature is available, turn uniformity correction off.)
18. Select the detector in question. (De-activate the other detector.)
19. Collimator type should be set to “None”. (“Nothing” or “Virtual” will be displayed for
collimator type.)
20. Set the “Camera Preset” to isotope being used. (Usually 99Tc)

Page 211 of 232


21. Select the “Acquisition Parameters” page.
22. Set “Matrix Size” to 256x256.
23. Set the “Stop Conditions” for 30,000,000 (30-million) counts.
24. Set the “Number of Statics” to 1.
25. Click the “Enable” button on the “Acquisition” window. The “Start” button should
highlight.
26. Click the “Start” button on the “Acquisition” control panel. The flood acquisition will begin.
27. When 30,000,000 counts have been accumulated, close the “Acquisition” window and
quit the “Acquisition Protocols” application.
28. Double click on the “Processing Protocols” icon to open the “Processing” application.
29. Make sure “D/A” button is selected.
30. Select “Flood Uniformity” from the list of pre-defines. Click the “Open” button.
31. Select the “Intrinsic Flood” name from the patient names list.
32. Click “Select”, click “OK”.
33. The message “Detector Shape” will appear. Make sure “rectangular” is selected and
then click “OK”.
34. The intrinsic flood picture will be displayed along with the intrinsic uniformity calculations.
35. Visual Analysis:
a. Zoom and window the image. Use a “Top” window level of 1000 and a “Bottom”
window level of 750.
b. Visually examine the image for proper uniformity.
36. Calculated Analysis:
a. Examine the uniformity numbers calculated to see if they meet or surpass the
published uniformity values for the detector.
37. If the numbers “exceed” specifications:
a. Close the “Processing” window and quit the “Processing” application program.
b. Proceed to detector troubleshooting to determine and resolve non-uniformity
issue.
c. Perform procedure until calculated numbers are within specifications.
38. If the numbers are “within” specifications:
a. Take a “screen snapshot” of these numbers by holding down the keys
“Command+Shift+3”. (This data may be used at a later date.)
b. Close the “Processing” window and quit the “Processing” application program.

Page 212 of 232


ecam Systems with esoft Workstation Computer

• For the ecam Duet system, there will be three additional intrinsic floods to be performed
unless the customer is not licensed for any isotopes within a particular energy range.
These energy ranges and isotopes are:
o Very Low Energy – Thallium – Tl-201 – (70keV)
o Medium Energy – Indium – In-111 – (274keV)
o High Energy – Iodine – I-131 – (364keV)

01. Remove the collimators.


02. Place the detectors in the “collimator change” position, (Detector 1 at 90o and Detector 2
at 270o).
03. Using “Caudal Tilt”, turn the detectors so that the crystals face away from the gantry or
out.
04. Use a 57Co or 99Tc source, (you will probably be using 99Tc which is just fine), that
produces a dead time of 3%-6%. Make sure that the source is 5 UFOV lengths from the
detector, (130 to 136 inches from the crystal face).
05. Place the selected source onto a source stand. Adjust the height of the source so it is
the same distance from the floor as the center of the crystal in question.
06. Center the source “Left-to-Right” on the crystal.
07. Flood the crystal.
08. From the Patient menu, select “Register”.
09. Enter the following information:
a. For the “Last Name”, enter: <Camera Name> QC
b. For “Patient ID”, enter: <Today’s Date> (Month/Day/Year)
c. For “Date of Birth”, enter: <Today’s Date> (Month/Day/Year)
d. For “Sex”, enter: Unknown
10. Click on the “Exam” button. (The new patient will appear in the patient database.

• The calibration floods patient folder is now created. Select this patient for each
calibration performed in this procedure.

11. Above “Category”, confirm that the “Local” checkbox is checked.


12. From the “Category” pull down menu box, select “Service” workflows.
13. From the workflow list, select the “Five FOV Intrinsic Flood QC” workflow by double
clicking on it.
14. On the “Acquisition” page, select the “Series Information” tab. Change the series name
to “Intrinsic Flood <isotope> <date>”.
15. Select the “Camera Parameters” tab. Confirm that the correct isotope has been selected
and that default is set to “Both Detectors”.
16. Select the “Stop Conditions” tab. Confirm that “Stop Conditions” is set to 30-million
counts 57Co or 99Tc; or 50-million counts for the other Duet isotopes.
17. Press the “Update View” button to apply changes made.
18. Click on the “Prepare Acquisition” button. If all system parameters are correct, the
“Start” button will highlight.
19. Click on the “Start” button. The 30-million or (50-million) count intrinsic flood will begin.
20. When the “PPM” shows that the acquisition has been completed, click on “Done”.
21. Click on the “Intrinsic Calculation” activity icon. (The intrinsic and curvature corrected
floods will be displayed. The uniformity numbers will also be displayed.)

Page 213 of 232


22. Compare all values with the values listed in the “Service Log” for that detector. If the
numbers are not as good as the values in the “Service Log”, then compare them with the
following minimum HD3/HD3C/HD3CF Detector specifications.

• When replacing or regenerating a Duet detector, all numbers should be compared to the
“Service Log” specifications. For all other servicing, intrinsic flood should meet the
uniformity specifications for HD3CF – (Deut only).

23. If the numbers “exceed” specifications:


a. Exit the workflow by clicking on the “Complete” box.
b. Proceed to detector troubleshooting to determine and resolve.
c. Repeat “QC” procedure.
24. If the numbers are “within” specifications:
a. Exit the workflow by clicking on the “Complete” box.
b. Continue this procedure.

Set-up new QC Data – Uniformity on the Fly Feature

• If “Uniformity on the Fly” feature is used, (monthly 200-million count correction flood),
perform the following steps:

01. Perform monthly QC flood procedure following manual instructions.


02. Check for acceptable uniformity values. If acceptable, save results.
03. Perform daily QC procedure.
04. Visually inspect the uniformity following correction. If acceptable, save results and new
reference image.

Other Post Repair Activities

01. Update the “Detector Service Log”, if necessary.


02. Create a new preserve directory floppy disk, if necessary.

END

Page 214 of 232


UNIFORMITY SPECIFICATIONS FOR HD3/HD3C DETECTORS

57
Co or 99Tc CENTRAL FOV USEFUL FOV

INTEGRAL <2.94% <3.74%

DIFFERENTIAL <2.54% <2.74%

UNIFORMITY SPECIFICATIONS FOR HD3CF DETECTORS (DUET ONLY)

TI-201 (VERY LOW) CENTRAL FOV USEFUL FOV TC-99m (LOW) CENTRAL FOV USEFUL FOV

INTEGRAL <7.20% <9.60% INTEGRAL <3.80% <5.80%

DIFFERENTIAL <4.30% <7.30% DIFFERENTIAL <3.10% <4.30%

Co57 (LOW) CENTRAL FOV USEFUL FOV In-111 CENTRAL FOV USEFUL FOV
(MEDIUM)

INTEGRAL <6.70% <10.80% INTEGRAL <10.40% <12.00%

DIFFERENTIAL <4.70% <7.60% DIFFERENTIAL <8.00% <10.40%

I-131 (HIGH) CENTRAL FOV USEFUL FOV

INTEGRAL <8.60% <12.00%

DIFFERENTIAL <3.80% <7.40%

Page 215 of 232


Page 216 of 232
ECAM FUNCTIONS

ITEM FUNCTION
AIMD Board Analog Interface Microprocessor-Digital. Provides the PI59 with data about
the event to aid in integration. The board has two main circuits; Event
Controller and Micro-processor Controller.
Analog Amplifier
Four analog amplifiers convert digital velocity data received from the PMAC and
supply current to each of the axis motors. The four amplifiers are identical except
the rotate amplifier can drive more current. Each of the amplifiers except the
rotate amplifier dirve 2 axes through PMAC controleed relay switching.
Clean Box Assembly with a toroidal isolation transformer, voltage taps, built in noise
filters and relay circuitry. Converts 230VAC input to filtered and isolated
125VAC. Also has relay circuitry which provides surge protection on
power up.
Coarse Tune Process to adjust the HV and gain of each PMT until the energy from that
PMT fall in the center of the tune window. Edge PMT's need to be within
2.3%, all others must be within 1.4%.
COP Correction Processor. Name for a collection of circuits on the DPC board
which take the data from the HD3 detector and prepare it for storage and
display.
COPC Correction Processor Control. Part of the Correction Processor which makes
sure all processes operate synchronously so that no events are lost or jumbled.

COPD Correction Processor Datapath. Part of the Correction Processor which performs
actual operations on the event data.
Dark Current When a PMT sends a signal with no source present, HV turned on and no light
from the crystal. "Unwanted Noise" which can effect image uniformity.
Detector P/S - (OEM) Provides +5VDC and +20VDC to the DTPS board. Voltages are used by the
detector PCB's and the detector. Also referred to as "OEM" supply.
DD59 Detector Master Board. Functions as a cable harness to connect the PMT
preamp outputs to the PI59 board. Also provides voltage to the PMT's and
biases the PMT MU-metal shield to the HV. Temperature sensors are located on
this board.
DPC Detector Processor Card. Installed in the SNAC computer, which performs
gamma event correction processing.
DTPS Detector Tube Power Supply. DTPS provides power to the detector boards;
(AIMD, PI59 and DD59). Also HV to the DD59 which goes to the preamps and
PMT's.
EVE Key Enables only those parts of the ICON software that the customer has purchased.
To see what options the customer has purchased, run a program called EVE
READ found in the utilities folder on the ICON software CD.

Extrinsic Flood Flood acquired with a collimator attached to the detector.


Fine Tune Process to adjust only the gain of each PMT until the energy from that PMT fails
into the center of the tune window. When a fine tune completes, edge PMT's
need to be within 1.4%; all other PMT's must be within 0.7%.
Framing Process in which image and physiological or timing data are sorted into correct
pixel locations within a matrix.
Gain Map Two sets of PMT values above and below 324kev.
Intrinsic Flood Flood acquired without a collimator attached to the detector.

Page 217 of 232


ECAM FUNCTIONS - CONTINUED

ITEM FUNCTION
LFR Light Response Function. The typical response of a PMT to an event introduces
some distortions. These Light Response Functions are corrected for in the
PCALC board.
MDB Motor Distribution Board. At each axis is an MDB which serves has a harness to
break out the various signals to their individual components.
MEB Motion Electronics Base. Both the MECP/MECP2 and the PMAC interface to the
MEB directly through connectors. Power supplies and each axis are also
connected to the MEB.
MECP Motion Electronic Control Processor. A single board computer which is the
"master" controller of all motions. The MECP/MECP2 communicates to the
SNAC, PMAC, MED/MEDU over separate serial RS485 connections. The
MECP/MECP2 monitors the system status via various sensors at all times and
decides whether the motion is allowed or not. (MECP2 is downward compatible
with systems running 2.1 and higher software).
MED/MEDU Motion Electronics Detector. One for each detector located under the detector
cover, are responsible for Caudal Tilt, C-Frame operation and auto contour light
curtains and look aheads. Both MED/MEDU are communicating to the
MECP/MECP2 via RS485 and are connected to the MEB. The MECP/MECP2
can override the MED/MEDU.
PCALC Position Calculator. Employs parallel signal processing to perform the following
functions: Pile-Up correction, LRF corrections, centroid calculation, correction for
offset errors, and scaling and positioning of the event.
Pedestal Input level to each of the FLASH ADC inputs on the PI59 board. This level is set
so that the input is just above the noise floor (inherent noise level of the system at
rest)Input level to each of the FLASH ADC inputs on the PI59 board. This level is
set so that the input is just above the noise floor (inherent noise level of the
system at rest).
PI59 PMT Integrator. Contains the Flash ADC's that digitize the preamp outputs.
PMAC Programmable Multi-Axis Controller. Dedicated real-time motion controller. It
directly controls the motor current to each axis through four analog amplifiers with
encoder and/or potentiometer feedback. When a command for motion is
received from the MECP/MECP2, the PMAC will switch the amplifier to the axis,
enable the amplifier and supply a digital value which is converted on the amplifier
to an analog voltage representative of the desired axis velocity.
POST Power On Self Test. Series of test performed by the computer when power is
switched on.
PPM Patient Positioning Monitor. The ecam flat panel display, which is installed on the
end of a moveable boom arm. It allows the nuclear technician to properly
position the patient for image acquisition. The PPM also provides gantry status,
collimator information and a diagnostic interface.
SCL Serial Command Link. Mode of communication between detector components
and the console. The SCL uses a RS485 at a baud rate of 38.4KHz.
SCSI Small Computer System Interface. High speed data bus used for connecting
hard disks, tape drives, and other accessories to the computer.
Serial I/O Board The communications interface board installed in the SNAC computer. Channel
A (RS232) communicates with the MEB and gantry. Channel B (RS485)
provides SCL communications to the detector(s).

Page 218 of 232


ECAM FUNCTIONS - CONTINUED

ITEM FUNCTION
SNAC Siemens Nuclear Acquisition Computer. This computer, along with the ICON
computer, provides control of the ecam system.
SPECT Single Photon Emission Computed Tomagraphy. Rotational study around the
patient in which a series of images at different angles are taken. These views
can be reconstructed later.
TAXI Transparent Asynchronous Xmitter-Receiver Interface. High speed data link from
the PCALC to the DPC board in the SNAC.
Tower The ecam electronics tower contains the clean box, Motion Electronics Base,
(MEB), SNAC computer and ICON computer. The ICON monitor and keyboard
may sit on top of the tower.
Tune Map Found under ICON Camera Tools (DAC/PMT). This is a display of how close
each PMT photopeak is to the keV that is used for tuning. When the detector is
tuned correctly, both coarse and fine tune are less than 1.4% and inside PMT's
are less than 0.&% of keV used for tuning.

Page 219 of 232


Page 220 of 232
SIEMENS ECAM LUBRICATION SCHEDULE

SIEMENS
DESCRIPTION PART NUMBER WHERE USED INTERVAL
CHRISTO LUBE MCG 200 GREASE 5224634 PATIENT FRONT BED ACME SCREWS BI-ANNUALLY
WAX STICK LUBRICANT (GS-20-1.6 OZ. STICK) 5224030 COLLIMATORS & SLIDE RAILS AS NEEDED
M-24 GREASE (2 OZ. BOTTLE) 5536867 YOKE BALL SCREWS ANNUALLY
GANTRY LINEAR BEARINGS ANNUALLY
PATIENT BED FRONT LINEAR BEARING BLOCKS ANNUALLY
C-FRME DRIVE ASSEMBLY ACME SCREW BI-ANNUALLY
C-FRAME DRIVE ASSEMBLY SPUR GEAR BI-ANNUALLY
REAR BED LINEAR BEARING BLOCKS ANNUALLY
OPEN GEAR 936 GREASE (IN CAULK TUBE) 569-000520 ROTATE RING GEAR ANNUALLY
RE-CONFIGURE LEVER AS NEEDED
BEVEL GEARS ANNUALLY
CP-28 GREASE (2 OZ. TUBE) 1242432 LEFT/RIGHT MOTOR ACME SCREW ANNUALLY
WD-40 OIL (12 OZ. SPRAY CAN) 915-000042 PATIENT BED FRONT DOCKER PINS AS NEEDED
COLLIMATOR CART DOCKER PINS AS NEEDED
C-FRAME BALL BUSHINGS (ON TUB) AS NEEDED

SIEMENS ECAM PAINTS

SIEMENS
DESCRIPTION PART NUMBER WHERE USED INTERVAL
PAINT - MED. WHITE (0.6 FL. OZ. BOTTLE) 2997562 GANTRY & PATIENT BED AS NEEDED
PAINT - MED. GRAY (0.6 FL. OZ. BOTTLE) 4372848 CABINETS - TOWER AS NEEDED
PAINT - MED. BLUE (0.6 FL. OZ. BOTTLE) 4387838 YOKE ASSEMBLY - GANTRY AS NEEDED

Page 221 of 232


SNAC

(SIEMENS NUCLEAR ACQUISITION COMPUTER)

ACCESSORY BOARDS

There are five accessory board installed in the SNAC computer.

1. Serial I/O (SCL – Serial Command Link – communications.


a. Two DB9 connectors:
i. One provides gantry communications
ii. One provides detector communications.
2. Parallel Digital I/O
a. Provides gantry motion and PPM switches communications.
3. DPC (Detector Processor Card)
a. Provides gamma event correction processing.
b. DPC has four LEDs primarily for debug purposes. (Listed in order from
left to right when viewing front, (component side), of board.
i. +5V Power
ii. Diagnostic Mode – TAXI interface is in diagnostic mode
iii. Violation Flag – A TAXI communications violation has been
detected by the TAXI circuitry.
iv. Hold – The input FIFO is almost full or hold has been asserted
through the TAXI Command register which is equivalent to the
clear-to-send output.
4. Ethernet
a. Interface (communication link) between the SNAC computer and the
ICON/esoft workstation.
5. Video
a. Provides system data to PPM for display by PPM.
b. PPM is also used as display monitor for SNAC computer.

Page 222 of 232


NORMAL / EXTENDED RANGE TUNING

After the completion of the tuning for normal mode narrow window, the system will
automatically tune for the extended range. The extended range is used for isotopes that
require any part of the window setting to be above 324kev. The HV value for the detector is
lowered by approximately 10%. This can be measured with a meter at TP8 on the DTPS
board. The reason the HV must be lowered is that the range of the 9 bit flash ADC's is
limited to a 2v signal. Isotopes above 245kev normally exceed this value. This results in a
need for a completely separate PMT DAC gain map.

CYCLE WINDOW TOLERANCE


TUNE MODE
COUNT INNER PMT'S OUTER PMT'S

COARSE

1ST PASS 13 WIDE 2.90% 3.90%

AFTER HV ADJUST 12 WIDE 2.90% 3.90%

AFTER MEETING WIDE


6 NARROW 1.40% 2.10%
WINDOW TOLERANCE

EXTENDED ENERGY 10 WIDE 2.90% 3.90%

AFTER HV ADJUST 12 WIDE 2.90% 3.90%

AFTER MEETING WIDE


6 NARROW 1.40% 2.10%
WINDOW TOLERANCE

FINE

NORMAL 6 NARROW 0.60% 1.40%

EXTENDED 6 NARROW 0.60% 1.40%

UNIFORMITY SPECIFICATIONS (30 MILLION COUNT FLOOD)

UFOV CFOV
3.74% 2.94% INTEGRAL
2.74% 2.54% DIFFERENTIAL

Page 223 of 232


TUBE WEIGHT KIT
( GE PART NUMBER 2326539 )
( SIEMENS PART NUMBER 3529711 )

3528986 450-009202
LARGE WEIGHT SMALL WEIGHT
FOR 3" TUBES FOR 2" TUBES

3528978 3529000
BASE FOR BASE FOR
NON-ASIC HD3 / ASIC
PREAMPS PREAMPS

BOLT ATTACHES
WEIGHT TO BASE

WEIGHT

3529703
EYE BOLT

PREAMP
TUBE

BASE
FITS OVER
PREAMP

Page 224 of 232


Page 225 of 232
ECAM ENTER/EXIT “MANUAL MODE” PROCEDURE

This procedure will assist you in putting the ecam system into “Manual Mode”. This procedure
assumes the person performing it obtains the basic knowledge to do so. It is recommended you
thoroughly read and understand this procedure before actually performing this procedure.

While performing this procedure, you must understand, follow and adhere to all safety
guidelines and procedures concerning, however not limited to the following: personal,
equipment, environmental, electrical and radioactive.

Items Needed:

• Your service tool kit.


• Two MED “Loopback Jumper”, (Siemens part number 5233122) (See Figure: 1)

CAUTION: While system is in “Manual Mode”, there are “NO” system fail-safes.

CAUTION: While system is in “Manual Mode”, “DO NOT” move Detector 1 radius
“out”. (If you move Detector 1 radius “out”, it may cause the “inner ring” to
disengage from the “outer ring” which could cause serious body injury and
equipment damage.)

CAUTION: While system is in “Manual Mode”, if “rotation” motion is desired or


required, first visually check to make sure the “Reconfigure Solenoid Lever” is not
engaged.

Enter Manual Mode

01. Power down the ecam system.


02. Remove the tower cover allowing access to the MEB board.
03. Locate MED “loopback jumpers”, (two), J7 and J4.
04. Remove connectors J7 and J4 from the MEB backplane. (See Figure: 2)
05. Insert MED “loopback jumpers” into the MEB backplane connectors J7 and J4.
06. Power up the ecam system.
07. Observe the MECP “Red” and “Green” LEDs flashing simultaneously and that the 4
“amplifier” LEDs are on.
08. Press the MECP “Diagnostic” switch, (SW2 – lower switch).
09. The MECP “Green” LED should flash rapidly, the AC fan comes up and the 165VDC is
available.

Exit Manual Mode

01. Power down the ecam system.


02. Remove the MED “loopback jumpers” at J7 and J4 from the MEB backplane.
(See Figure: 2)
03. Reconnect the original cables at J7 and J4 onto the MEB backplane.
04. Power up the ecam system.
05. When prompted, “Home” the system and verify proper axes motion.
06. Reinstall any removed covers.

Page 226 of 232


Page 227 of 232
Page 228 of 232
ECAM esoft LOGIN PROCEDURE

Use this procedure to login on the esoft workstation as “esoft” user.

01. When you see the message below displayed on the screen, press the keys on the
keyboard “Ctrl + Alt + Delete”.

02. When you see the “Logon Information” window display, for “User name” type “esoft”.
(Without quotations and in lowercase.)

Page 229 of 232


03. Once you have entered the “User name”, enter the “Password” using “esoft2000”.
(Without quotations and in lowercase.)

04. Once the “User name” and “Password” have been entered correctly, you should see the
esoft boot screen displayed something similar to below.

Page 230 of 232


ECAM esoft LOGIN PROCEDURE

Use this procedure to login on the esoft workstation as “Administrator” user.

01. When you see the message below displayed on the screen, press the keys on the
keyboard “Ctrl + Alt + Delete”.

02. When you see the “Logon Information” window display, for “User name” type
“Administrator”. (Without quotations.)

Page 231 of 232


03. Once you have entered the “User name”, enter the “Password” using “e.inside”.
(Without quotations and in lowercase.)

04. Once the “User name” and “Password” have been entered correctly, you should see the
esoft boot screen displayed something similar to below.

Page 232 of 232

You might also like