Professional Documents
Culture Documents
Page 2 of 232
Page 3 of 232
Page 4 of 232
ecam BOOM
Gantry ARM
HAND
CONTROL CAUDAL TILT SWITCH
E STOP
PPM
esoft
DET 1 Workstation
E STOP
C-FRAME
TOWER
PALLET
PALLET CLUTCH
RELEASE
SWITCH_LEFT HORIZ
VERTICAL ACME MOTOR ASM.
SCREW COVERS
REAR DOCKERS
BED
LEFT D
COVER FRONT
BED
LEFT CABLES
FRONT
ROLLER DET 2
FRONT BED
Page 5 of 232
E.CAM SYSTEM
YOKE MOTOR
SHAFT BRAKE
90VDC MED
ENCODER 24VDC
E.SOFT
MDB YOKE 1 HOME
10K POT CAUDAUL TILT
CAUDAL TILT CLUTCH 24VDC
YOKE 1
WITH 10K POT
TUBE GEAR
STRIP
RAIL
DETECTOR
ACQUISITION DETECTOR 1 CART 10K POT
1 SHOWN
E.SOFT WORKSTATION DETECT
SNAC VI DEO C-FRAME
MONITOR COLLIMATOR WITH EEPROM MOTOR
ETHERNET(EXT)
SEATED 24VDC
ETHERNET(SNAC) ETHERNET
BLANK
MED 2 ( NOT SHOWN )
TUBE GEAR
TICKS B
BRAIN MAGNET
SCL 38. 4K BAUD RATE
RS-485 COM. A
RAIL
POSITION RIGHT ARM REST SWITCH
P/S5V&12V
J4 BELT DRIVE
ESB/SOLENOID J12
J14 MDB YOKE 2 HOME
P/S 24V
AC POWER H2 180
J15 ON HORIZONTAL CLUTCH 24VDC RE-CONFIGURE SOLENOID 165VDC
PPM 12VDC J11 SWITCH
FAN1 MEB FAN2 MDB OUT HOME DETECTOR 2 CONNECTS TO J12
HAND CONTROLLOR
J10
HORIZONTAL MOTOR
J1
J2
J3
J5
J6
J9
J8
HOME
90VDC
OUT ER R I NG
DETECTOR 2
I NNER RI NG
RS- 485 COMM.
CLEAN BOX
ROTATE BRAKE 24VDC
ROTATE MOTOR 109VDC
TOWER E
200 - 240 VAC 30 AMP PPM
1 C ROTATE SHAFT
A ENCODER
HD3 DET ECT OR M
TO DETECTOR 1
T AXI MDB
ROTATE / DETECTOR
DETECTORP/S 1 HOME
DTPS BOARD
FLOOR
PCALCBOARD MAGNET
TAXI DOCKING
AIMD BOARD DOCKING FRONT 2
OUTSIDE TUB 00 REAR
INSIDE TUB PI59 BD
DD59 BOARD
PRE-AMP PALLET
DA C CARRIAGE
SHAF T EN CODER
PMT ’ s UP
10K POT BASE
DETECTOR 2 SHOWN MDB 10K POT
SHAF T CARRIAGE DOWN 1 0 K POT
E NCODE R HOME
FRONT UP / DOWN HOME SWITCH REAR HOME
UP / DOWN MOTOR 90VDC MDB
NOTE: AMP BOARD FUNCTION SWITCH
FLOOR
AMP 1 = ROTATE MOTOR LEFT / RIGHT MOTOR DRIVE 90VDC
AMP 2 = RADIAL 1 / PATIENT BED REAR MDB REAR UP / DOWN
AMP 3 = RADIAL 2 / PATIENT BED UP / DN MOTOR 90VDC
AMP 4 = GANTRY L / R - PATIENT BED IN / OUT
Page 6 of 232
Page 7 of 232
PPM
E STOP
TOWER
HAND
CONTROL
Page 8 of 232
1. STARTS OR STOPS IMAGE ACQUISTIONS.
1 2 2. CLEARS THE PATIENT POSITIONING MONITOR
(PPM) SCREEN SHOWN ON THE DISPLAY PANEL.
3. MOVES THE PATIENT BED TO A PREDETERMINED
SETTING FOR LOADING/UNLOADING WHEELCHAIR
3 4 PATIENTS.
5 4. MOVES THE PATIENT BED TO A PREDETERMINED
SETTING FOR LOADING/UNLOADING GURNEY
7 6 PATIENTS.
5. MOVES THE PATIENT BED IN AN UPWARD
8 DIRECTION.
NOTE: THIS BUTTON IS ALSO USED FOR AUTO
9 11 HEIGHT POSITIONING.
6. MOVES THE PATIENT BED AWAY FROM THE
10 12 GANTRY.
7. MOVES THE PATIENT BED IN TOWARDS THE
GANTRY.
8. MOVES THE PATIENT BED IN A DOWNWARD
14 13 DIRECTION.
9. MOVES DETECTOR 1 OUTWARD (RADIALLY).
10. MOVES DETECTOR 1 INWARD (RADIALLY).
15 16 11. MOVES DETECTOR 2 OUTWARD (RADIALLY).
12. MOVES DETECTOR 2 INWARD (RADIALLY).
NOTE: NUMBERS 11 AND 12 ARE NOT APPLICABLE
TO A SINGLE-DETECTOR SYSTEM.
13. ROTATES THE GANTRY IN A CLOCKWISE
DIRECTION.
14. ROTATES THE GANTRY IN A COUNTERCLOCKWISE
DIRECTION.
15. MOVES GANTRY TO LEFT DIRECTION.
16. MOVES GANTRY TO RIGHT DIRECTION.
NOTE: NUMBERS 15 AND 16 ARE NOT APPLICABLE
TO A SINGLE-DETECTOR SYSTEM.
Page 9 of 232
Page 10 of 232
CAUDAL TILT CLUTCH
GEAR TEETH
Page 11 of 232
Page 12 of 232
DET. 1 RADIUS
DRIVE ASM.
E STOP
HAND
CONTROL
RADIUS
HOME FLAG
RADIUS
POT GEAR
RADIUS
POT
Page 13 of 232
RECONFIGURE
FLAG
DETECTOR
RADIUS HOME
FLAG
FLAG SENSORS
GEAR STRIP
DETECTOR
RADIUS
DETECTOR
CALIBRATE POT
RADIUS
ADJUST
CALIBRATE POT
DETECTOR
RADIUS
CALIBRATE POT
GEAR
Page 14 of 232
Page 15 of 232
DETECTOR 1
RADIUS DRIVE
ASSEMBLY
Page 16 of 232
Page 17 of 232
GANTRY HEIGHT
ADJUSTMENT
BOLT
LEFT SIDE
GANTRY FRONT
WHEEL
STOP BOLT
LOCKING BOLT
FRONT WHEEL
WEAR PLATE
Page 18 of 232
RADIAL DRIVE
SPUR GEAR
(WORM GEAR SIDE)
RADIAL DRIVE
SPUR GEAR
(MOTOR SIDE)
Page 19 of 232
Page 20 of 232
Page 21 of 232
OUTER RING
RECONFIGURE
SOLENOID INNER RING
E STOP
A FRAME
CABLE TAKE-UP
ROLLERS HAND
CONTROL
REAR BED
ROTATE
HOME FLAG
ROTATE
MOTOR
Page 22 of 232
ROTATE
ROTATE MOTOR
HOME FLAG MOUNTING
BRACKET
ROTATE
HOME
SENSOR
ROTATE
MOTOR
Page 23 of 232
ROTATE
HOME FLAG
ROTATE
HOME
SENSOR
Page 24 of 232
ROTATE MOTOR
DRIVE GEAR
ROTATE HOME
ADJUSTMENT POT
Page 25 of 232
OUTER RING
INNER RING
Page 26 of 232
RECONFIGURE
SOLENOID ASSEMBLY
OUTER RING
INNER RING
A FRAME
Page 27 of 232
RECONFIGURE
SOLENIOD STATUS
SWITCH/SENSOR
RECONFIGURE
LATCH
RECONFIGURE
SOLENIOD
RECONFIGURE
SENSOR
A FRAME
Page 28 of 232
RECONFIGURE
LATCH
RECONFIGURE
FLAG
RECONFIGURE
SOLENOID
RECONFIGURE
A FRAME
SENSOR
Page 29 of 232
DETECTOR CABLE
TAKE-UP ROLLERS
Page 30 of 232
A FRAME
REEL SPRING
MDB
REEL SPRING
STEEL CABLE
DETECTOR
CABLES
Page 31 of 232
STEEL REEL CABLE
Page 32 of 232
E-STOP-(GANTRY
LEFT SIDE-OUTSIDE)
E-STOP-(GANTRY
LEFT SIDE-INSIDE)
Page 33 of 232
E-STOP
(VIEWED FROM INSIDE
GANTRY REAR)
Page 34 of 232
RING LOCK
LEVER
Page 35 of 232
RING LOCK LEVER
SPRING
Page 36 of 232
OUTER RING
RING LOCK
BLOCK
INNER RING
Page 37 of 232
Page 38 of 232
FRONT
BED
Page 39 of 232
ROLLER REAR ROLLER
ASM. ASM.
RIGHT LEFT
LEF
RES
TA
ARM
RM
HT
RES
RIG
T
GROUND
ROLLER
FRONT
Page 40 of 232
CUT-OUT IN GUIDE
ROD FOR BRAIN
POSITION SWITCH
BRAIN
POSITION
SWITCH
TRAY FOR
PALLET
IN/OUT DRIVE
BELT
OUTSIDE
GUIDE ROD
BRASS NUT
ACTUATOR
FOR SAFETY
SWITCH
WEAR
INDICATOR
PLATE
Page 43 of 232
BOLT FOR FRONT
COLUMN COVER
VERTICAL
MECHANICAL
STOP BOLT
VERTICAL GUIDE
RAIL
Page 44 of 232
VERTICAL
Page 45 of 232
VERTICAL
DRIVE
PULLEYS
VERTICAL
DRIVE BELT
Page 46 of 232
FRONT BED IN/OUT (HORIZONTAL) DRIVE ASSEMBLY
Page 47 of 232
FRONT BED IN/OUT (HORIZONTAL) DRIVE ASSEMBLY - LEFT SIDE
Page 49 of 232
FRONT BED DOCKING ASSEMBLY
Page 50 of 232
NEW STYLE DOCKING ASSEMBLY
Page 51 of 232
SENSOR
MAGNET
Page 52 of 232
Page 53 of 232
Page 54 of 232
LOWER LIMIT
VERTICAL SENSOR
MOTOR
MDB
Page 55 of 232
REAR BED
IN
DOWN
POSITION
Page 56 of 232
Page 57 of 232
ECAM TOWER POWER SUPPLIES/AMPLIFIERS
MECP – (Motion Electronics Control Processor) – Control processor for all motions.
• +5 VDC – All logic and remote sensors (opto home sensors, magnetic reed sensors,
microswitches).
• +12VDC – Circuits related to analog and differential signals.
Page 58 of 232
MOTOR
POWER +5V
SUPPLY +/-12V
POWER
SUPPLY
+24V
POWER
SUPPLY
Page 59 of 232
J7-DETECTOR 2
PMAC BOARD
J4-DETECTOR 1
MEB
Page 60 of 232
J7 J4
PMAC
MEB
Page 61 of 232
Page 62 of 232
DETECTOR PROCESSOR
COVER
Page 63 of 232
DETECTOR
PROCESSOR
ASSEMBLY
Page 64 of 232
Page 65 of 232
DEAD
TIME
AIMD
TAXI
PCALC
Page 66 of 232
DTPS
Page 67 of 232
Page 68 of 232
SIEMENS ecam ESOFT NORMAL
POWER-UP / POWER-DOWN SEQUENCE
Power Up Sequence
Page 69 of 232
ECAM NORMAL POWER UP
COMPUTER BREAKER
COMPONENT OBSERVE
SNAC computer begins SNAC green LED turns on. Hard disk access amber
to boot. LED turns on.
CAMERA BREAKER
COMPONENT OBSERVE
MEB fan #2 turns on. Fan on left side of MEB Base turns on.
MEB power on. MEB Backplane green power LED's light; (+12V, -
12V, +5v, +5V MECP, +5V SENSOR). Axes
connector LED's light; (D20 thru D26)
MECP boots. Red LED on MECP flashes once.
Red & Green LED's on MECP ON for 2 seconds.
Red & Green LED's on MECP OFF.
Green LED's on MECP ON steady.
PMAC boots. Red LED very dim while MECP is booting.
Yellow and Green LED's on after successful MECP
boot.
Green Amplifier LED's light.
MOTION BREAKER
COMPONENT OBSERVE
+24V Switched turns on. Patient bed pallet clutch engages.
Page 70 of 232
ECAM TOWER / CLEAN BOX CIRCUIT BREAKERS
FAN #2 FAN #1
MAIN
INPUT MOTION COMPUTER
AC INPUT
ECG CAMERA
MOTION COMPUTER
MAIN INPUT
AC
INPUT
ECG CAMERA
VOLTAGE SELECTION
200v
208v
AUX AUX AUX AUX 220v
230v
240v
ECG INPUT
AC INPUT 0.5A 120V Page 71 of 232
ECAM esoft SYSTEM POWER UP ACQUISITION esoft
FRONT REAR
MAIN CIRCUIT
POWER e.soft
BREAKER - 30 AMP
FD
( 200 - 240 VAC )
VOLTAGE
HD MONITOR 19"
TAP HERE PRESS SOFT START
HARD DRIVE TO POWER UP ICON
CD
PPM
CLEAN BOX
ETHERNET PLUG
HOME
12VDC RESET
GREEN RED
LEDs
ETHERNET PLUG POWER
CAMERA BREAKER
SWITCH
POWER TO DETECTOR S1
ASSEMBLIES & MOTION
MEB BOOTING MEB CARD CAGE POWER STEP 1
ELECTRONICS CARD
GREEN
PS 5 / 12VDC
CAGE - LOGIC PMAC DETECTOR POWER 120 VAC
RED
ON
MOTOR PS
MOTION BREAKER OEM
+5 MEB POWER SUPPLY
POWER TO MOTION
AMP 4
AMP 3
AMP 2
AMP 1
+5 MECP +/- 5 & 20 VDC
ELECTRONICS CARD
MECP
PS 24VDC
CAGE - MOTOR POWER +5 SEN
15 VDC +12 VDC
OFF
-12 VDC
ECG BREAKER
DTPS -20VDC
-5VDC
POWER TO ECG
PLUG PATIENT BED +5VDC
-15VDC +20VDC
FAN #1 FAN #2
-12VDC
ON OFF +12VDC HV
+15VDC
MEB CARD CAGE POWER STEP 2
PS 5 / 12VDC
EMERGENCY PCALC AIMD
STOP BUTTONS
ON
MOTOR PS
CPU
AMP 4
AMP 3
AMP 2
AMP 1
MECP
PS 24VDC
ECG
ON
-15VDC
+15VDC
-5.2VDC
+5VDC
+5VDC
ON ON
Page 72 of 232
Page 73 of 232
ecam Tower AC Voltage Power Checks
MOTION COMPUTER
MAIN INPUT
AC
INPUT
ECG CAMERA
Make sure all circuit breakers are set to “ON”. Use the following table to verify
that the Tower AC voltages are within specifications.
VOLTAGE SELECTION
200v
208v
AUX AUX AUX AUX 220v
230v
240v
ECG INPUT
AC INPUT 0.5A 120V
Page 74 of 232
Page 75 of 232
TROUBLESHOOTING ecam +5V/+12V POWER SUPPLY
NOTE: If adjustment is required, it will be necessary to remove the Motor Power Supply
board and maybe the Rotate Amp board to access adjustment pot. Before removing this
or any board(s), properly shutdown ecam system, remove board(s), then properly apply
power to ecam system.
01. With power applied to the ecam system, measure with DVM the outputs of the
power supply. Adjust if Clean Box tapping matches input voltage.
Page 76 of 232
a. +5V +0.25V / -0.00V at TP31 using TP19 for reference. Adjust with
V1ADJ pot on power supply; (CW increases voltage).
b. +12V +0.6V at E5 using TP19 for reference. Adjust with V2ADJ pot on
power supply; (CW increases voltage).
V1ADJ (+5V)
+5/+12V Power Supply
V2ADJ (12V)
MEB Backplane
V1ADJ (+24V)
+24V Power Supply
Page 77 of 232
MOTOR
POWER +5V
SUPPLY +/-12V
POWER
SUPPLY
+24V
POWER
SUPPLY
Page 78 of 232
Page 79 of 232
TROUBLESHOOTING ecam +24V POWER SUPPLY
NOTE: If adjustment is required, it will be necessary to remove the Motor Power Supply
board and maybe the Rotate Amp board to access adjustment pot. Before removing this
or any board(s), properly shutdown ecam system, remove board(s), then properly apply
power to ecam system.
01. With power applied to the ecam system, measure with DVM the output of the
power supply. Adjust if Clean Box tapping matches input voltage.
a. +24V +1.2V between E15 or E16 (GRY) to E19 or E20 (BRN). Adjust
V1ADJ pot on power supply; (CW increases voltage).
V1ADJ (+5V)
+5/+12V Power Supply
V2ADJ (12V)
MEB Backplane
V1ADJ (+24V)
+24V Power Supply
Page 80 of 232
MOTOR
POWER +5V
SUPPLY +/-12V
POWER
SUPPLY
+24V
POWER
SUPPLY
Page 81 of 232
Page 82 of 232
ECAM HD3 DETECTOR OEM POWER SUPPLY VOLTAGE
VERIFICATION
Use this procedure to verify the detector OEM power supply voltages.
03. Set the oscilloscope time base at 5msec/div and check the ripple voltage on the
+5.15V and –5.2V supplies. The ripple voltage for these supplies must not
exceed 50mV.
04. There is an in-line fuse which comes off the +5.15V terminal on the OEM Power
Supply. Check this fuse if there is no +5.15V to the detector PC boards.
OEM R59
R7
R405
-20V
TP22
-5V
+5V
TP2
+20V
HV
DTPS S1
-12V
+15V
+12V
Page 83 of 232
OEM
POWER
SUPPLY
Page 84 of 232
Page 85 of 232
ECAM HD3 DETECTOR DTPS (DETECTOR POWER SUPPLY)
VOLTAGES VERIFICATION
Using a DVM and oscilloscope, verify the following DTPS output voltages and adjust if
necessary.
Page 86 of 232
DTPS Oscillator Verification
Page 87 of 232
Page 88 of 232
Page 89 of 232
ECAM HD3R/HD4 DETECTOR VOLTAGE CHECKS
The new detector assembly for the Siemens ecam system is called the HD3R or HD4.
Currently, we have little information concerning this new detector, however this does not
mean there is nothing we can do. We are still limited due to not having access to
“Standard Plus” diagnostics. Operational speaking, both the HD3 and HD3R detectors
function the same.
• The PI59 and the PCALC boards have been combined into the PI59R board.
(This PI59R board is located inside the detector tub.)
• The AIMD and the DTPS boards have been combined into the AIMDR board.
• The OEM power supply has been changed. (Still referred to as the OEM.)
• Both the AIMDR and the PI59R boards are “digitally” calibrated using
diagnostics which are currently not available to us.
-100V
• Can be measured on the AIMDR board, TP116 (J15) using TP63 (Z15) as
reference. (+100mV)
-200V
• Can be measured on the AIMDR board, TP115 (J16) using TP63 (Z15) as
reference. (+100mV)
HV S6
• The detector HV ON/OFF Switch S6 is located on the AIMDR board (C13).
HV Reference
• The HV Reference voltage can be measured on the AIMDR at TP109 (D12) using
TP10 (H12) as reference. The HV reference voltage adjustment is R918 (F13).
Page 90 of 232
ECAM HD3R DETECTOR OEM POWER SUPPLY VOLTAGE CHECKS
R41
Page 91 of 232
Page 92 of 232
Page 93 of 232
ECAM HD3 DETECTOR GAMMA EVENT FLOW
1. Gamma ray strikes the crystal. The energy from the gamma ray is transferred to the
sodium iodide molecules in the crystal causing the excited molecules to emit light
photons which in turn strike the PMTs.
2. Energy from the photons is transferred to the molecules in the light-sensitive coating that
covers the inside base of the PMT. These exited molecules emit electrons. A focusing
electrode attracts the electrons to the first of a series of dynode plates inside the PMT.
Each plate is set to a decreasing voltage potential, the first is at a potential of
approximately –800V, the second approximately –700V and continues until the last plate
at -100V. Each successive plate more and more electrons are released multiplying the
few that strike the first plate to millions at the last plate. The result is a small but
measurable current at the output of the PMT.
3. The PMT current passes to the preamp resulting in the signal PAOUT. The signal
PAOUT represents the energy of the gamma ray event as seen by each PMT. At the
preamp, PAOUT is amplified 10 times then sent to the DD59(C) or (CF) board.
4. On the DD59, each PAOUT signal is summed through a 2K resistor network resulting in
the signal ES which is then sent to the PI59 board.
5. On the PI59, ES is now called SUM_E_MTX and applied to negative input of Energy
Sum Amp. Output of the Energy Sum Amp is not called SUM_E_AMP and used in two
places:
a. Applied to an amplifier to energy FLASH ADC on PI59 where it is digitized
representing the raw energy value of the gamma event;
b. Passes through a 49.93/4 resistor on the PI59 and renamed SUM_E_OUT and goes
to an amplifier on the AIMD. Output of the amplifier signal is now called
SUM_E.
6. Each of the 59 PAOUT signals are passed through the DD59 to the PI59 board where
they are fed into on of the 59 Flash ADCs which convert the analog signal from each of
the PMTs to a digital signal. These signals are then integrated. Integrations is
accomplished by the ASIC. These integrals represent the total energy of the detected
gamma ray and the portion seen by each PMT. (Each ASIC is connected to 12
Amplifier/FLASH ADC circuits, therefore five ASIC devices are required on the PI59
board.)
Page 94 of 232
7. While the event is being integrated on the PI59 board, the AIMD board is examining the
SUM_E signal for that same event. The analog SUM_E signal on the AIMD board is
used to drive several circuits.
a. INDET – The signal SUM_E is fed through an amplifier with a threshold detector
resulting in the output signal INDET; (Incident Detection). The signal INDET
starts the Flash ADC clock which in turn starts the integration process.
b. PILE-UP DETECT – If INDET stays high after the rising edge of PILE-UP
DETECT, then a pile-up condition exists. (In a PILE-UP, a new event has started
integrating before the previous event has completed integration.)
c. PILE-UP AMPLITUDE – This amplifier measures the amplitude of the pile-up
event. If the level is below a certain threshold, then the pile-up event is rejected.
d. EARLY DUMP – The signal SUM_E is compared to a coarse upper and lower
energy window. If a pile-up event does not fall in this range, then the EARLY
DUMP signal is sent to the PI59 board to stop integration.
e. COSMIC OVERLOAD – If SUM_E goes above +3.6V, the circuit generates a
signal which dumps the event. An OVERLOAD DELAY signal is also generated
to prevent a new event from starting right away.
f. AFTERGLOW – The afterglow circuit compensates for the baseline shift
encountered at high count rates.
8. AIMD –
a. Determines if the event is accepted or dumped.
b. Uses the signal INDET to feed the dead time circuit via a resistor/capacitor
network which is charged and discharged. This voltage is fed to an ADC in the
microprocessor controller where 5V is equal to 64% dead time and 0.3V is equal
to <4% dead time.
c. There is a 120MHz clock and phase circuitry which controls event integration on
the PI59 board.
d. There is also a microprocessor which acts as a board controller controlling board
functions such as High Count Rate, Temperature Compensation and
communicating with the DPC board in the SNAC to load the PMT gain DAC data
for tuning.
9. After the PI59 has completed the PMT integration process, this data is sent to a FIFO on
the PI59 which is connected to a FIFO on the PCALC board. From this FIFO, the data if
fed through buffers to the DSP, (Digital Signal Processing), nodes of which there are
eight. (One DSP acts as master node, the other seven are general processing nodes.)
Each node processes a full block before outputting the data to the TAXI interface and
accepting a new input block. The following events take place in the DSP nodes:
a. Correct for short integration or pile-up
b. Correct Zi magnitudes for offset errors
c. Process PMT data for LRF (Light Response Function) corrections (optical
distortion)
d. Calculate the event centroid and correct for energy
e. Scale and position the event.
Page 95 of 232
10. Event information is now sent from the PCALC board to the DPC board in the SNAC
computer via the TAXI, (Transparent Asynchronous Xmitter-Receiver Interface), which
is a high speed serial link having a transfer rate of 140Mbps. The TAXI can be thought
of as a pseudo-parallel register. It load data into one side and outputs it on the other, but
in this case the “other” side is separated by a long serial link utilizing 503/4 ohm coaxial
cable.
11. The TAXI link, Formatter and Input FIFO make up the data input function of the DPC.
a. TAXI Link – Receives the high speed serial data from the PCALC board and is
converted to a parallel format in the TAXI circuitry. The CTS, (Clear To Send)
signal is generated by the TAXI circuit to control data transfer speed between the
PCALC board and the DPC.
b. Formatter – The event data is received by the formatter in three 16-bit words
(SPECT mode) or four 16-bit words (COIN mode). The formatter rearranges the
event data into three 18-bit words (SPECT mode) or four 18-bit words (COIN
mode). This is done so the data is in a form that is easily used by the Correction
Processor.
c. Input FIFO – The purpose of the FIFO is to decouple the TAXI serial input from
the Correction Processor. Writing to the FIFO is done on the TAXI clock while
reading from the FIFO is done on the Correction Processor clock.
12. Correction Processor (COP) has two major functions:
a. Correction Processor Control (COPC) controls functions such as synchronous
state machines which keep the event processing synchronized.
b. Correction Processor Data Path (COPD) contains adders, multipliers and LUTs
that perform actual data calculations.
13. The event passes through four stages on the Correction Processor (COP)
a. Stage 1 – Zmap Coefficients are applied to the event data. The Zmap coefficients
are stored on the flash memory on the DPC board.
b. Stage 2 – Data is compared to upper and lower energy boundaries to determine if
the event falls within a certain energy window. If it is, the image event is passed
on to Stage 3. If it is tuning or peaking event, it is read out by the SNAC
computer at this point and no further processing is performed on the data.
c. Stage 3 – Performs the Linearity Correction algorithm on the event data. The
coefficients used in this process are unique to each detector and stored on flash
memory on the DPC board.
Page 96 of 232
d. Stage 4 – The event data is prepared for display.
i. The X and Y data is compared against upper and lower X/Y limits and if
within these limits, the event is considered to be within the Fine FOV.
Only events within the Fine FOV are processed further.
ii. The event is oriented for display. The X and Y values can be swapped
and/or flipped if desired.
iii. Following orientation, image panning is performed which is a function
that adds a pan offset to the detector coordinate system data.
iv. The magnify function is performed next. The algorithm used allow only
certain magnification factors.
v. Matrix sizing is now implemented. Matrix sizing allows viewing of the
image with various resolutions.
vi. The Scaler ID is determined. The Scaler ID is a function of the detector
and the associated analyzer.
vii. Counters are activated:
1. Scaler Counters – Four 32-bit counters that count events in the
Fine FOV for an associated head and analyzer.
2. Count Rate Counters – There 24-bit counters that count gamma
events.
3. Time Tick Counters – A 26-bit down counter counts COP clock
cycles and when zero is reached, a “Time Tick” is generated and
the counter is reset to a value determined by software.
14. Data from the COP is outputted to the SNAC computer by the Output FIFO, PCI
Interface and DMA Controller.
15. There types of communication between SNAC and ICON/esoft workstation:
a. Command Control – Sets up and controls motion and electronic parameters for
camera operation.
b. Growing Image Server – Sends an image in progress to the workstation for
display during acquisition.
c. Image File Server – Sends the final image to the workstation. (During acquisition,
image data is stored on the SNAC computer until acquisition is complete, then
entire acquisition is sent to the workstation.)
NOTE – All boards referenced such as the AIMD also include the AIMD(C)(CF), just as
the DD59(C)(CF) board, however does not include any boards contained in the HD3R
detectors.
Page 97 of 232
Processing Computer
DAC Control
J17
DD59
HV
ADC . Scale (COP)
. Position
PAOUT
DSP
2K X, Y, E
Det. ID
TAXI Z MAP
Ethernet
TAXI
FLASH Stage 1
J16
FIFO
FIFO
DSP
J5
ASIC ASIC (External Network)
CTS
TUNING PEAKING - PMT Data
FLASH
PAOUT ASIC - (Tuning Peaking)
2K ADC DSP
J9
-
Stage 3
J8
ASIC + + Ethernet
- Linearity Correction
DUMP/ MASTER
LINEARITY
CORRECTION (External Network)
R112 R127 FLASH INTEG
SUM_E DSP
ADC CONTROL NODE - LUT
HV DIAG. Stage 4
BASELINE
LOOPBACK (Flash - Fine FOV
SUM E
Memory)
OFFSET OFFSET
SCL SCL +
- Orientation ICON /
- Panning
PAOUT
2K INTERFACE INTERFACE - Magnify esoft Workstation
J3 J14 J3 J1 J2 J3 J13 J12 - Matrix Sizing
Controller
- Scaler ID
- Output Format Computer
To Detector 1 or (Acquisition)
DMA
Terminator . CTS
(If single head.)
(CLEAR TO
SEND)
J3 J5 J3 J1 J2 J3
EVENT CONTROL
SCL . Pile-Up
. Phase
INTERFACE . Indet
. Dump PCI Output FIFO
OVERLOAD
. Clock
DELAY
+ 3.6V
OVERLOAD
Interface 512x 18 Bits
4us
Ethernet
E OFFSET
BASELINE
One
Shot DUMP
PA9 EARLY
EARLY DUMP LOWER LIMIT
DUMP DISABLE
120nS 0.0V
DEADTIME
One
EARLY DUMP UPPER LIMIT Shot PA8 EARLY
INDET DUMP ENABLE DTAB
OFF ADJ.
Serial Communications Link
SUM E OUT
INDET
SERIAL I/O
INDET INDET
DC OFFSET
SUM Z 180-280nS
END PCI BUS
DACs
PILEUP
ON ADJ.
AIMD
ISA BUS
Ethernet
SCL INTERFACE
SLR + / -
SLT + / -
Page 98 of 232
SNAC COMPUTER
Page 99 of 232
ECAM – MOTION AXES – (DUAL HEAD)
Each axis, the electrical interface is almost identical. Each axis consists of a motor, an encoder
feedback, (except Up/Down rear bed (PHS) which has no encoder), a position potentiometer,
(used at power up and with the home sensor for position verification), and a home flag sensor.
Three axes have a brake. The rotate brake is integrated into the rotate motor. The two other
brakes are on the output shaft of the radial 1 In/Out drive and the radial 2 In/Out drive. All three
brakes are controlled by the PMAC.
Other sensors such as Rear PHS carriage tilt sensors, Front Patient Bed docker sensors, etc. are
carried to the MEB for the MECP/MECP2 via the axis cable but do not report to the PMAC.
Each axis has one cable from the MEB to the axis. At each axis is a Motor Distribution Board
(MDB) which serves as a harness to break out the various signals to their individual components.
PREREQUISITES
SEQUENCE – 180o
SEQUENCE – 90o
MECP – (Motion Electronics Control Processor) – Control processor for all motions.
• +5 VDC – All logic and remote sensors (opto home sensors, magnetic reed sensors,
microswitches).
• +12VDC – Circuits related to analog and differential signals.
TOWER
Acquisition Computer
MEB
MDB
AXIS MOTOR
NOTES:
VIEW
FLAG
DETECTOR 1 at 360o
REAR VIEW
SAFE TO PRESS
CCW ROTATE
BUTTON ONLY
CAUTION:
PHYSICAL STOP
ECAM diagnostic tests for esoft workstations are “CD_ROM” based. The
diagnostics are presented on two levels.
The STANDARD level contains all the diagnostic test necessary to calibrate the
system. You can always run the STANDARD diagnostics by leaving the
password fields blank on the “Service Software Authentication” page.
The STANDARD PLUS level contains all the STANDARD level test, plus many
additional tests, which can be helpful in system troubleshooting.
NOTE:
• For troubleshooting a detector problem, make sure you are using the
STANDARD PLUS level of diagnostics.
DETECTOR DIAGNOSTICS
DETECTOR - TEST - PCALC: SELF (DETECTOR 1)
DETECTOR - TEST - PCALC: SELF (DETECTOR 2)
DETECTOR - TEST - AIMD: SELF TEST (DETECTOR 1)
DETECTOR - TEST - AIMD: SELF TEST (DETECTOR 2)
DETECTOR - TEST - PI: SELF TEST (DETECTOR 1)
DETECTOR - TEST - PI: SELF TEST (DETECTOR 2)
DETECTOR - TEST - PI: ADC RAMP (DETECTOR 1)
DETECTOR - TEST - PI: ADC RAMP (DETECTOR 2)
DETECTOR - TEST - COIN: SELF TEST **
DETECTOR - TEST - EVENT CONTROLLER (DETECTOR 1)
DETECTOR - TEST - EVENT CONTROLLER (DETECTOR 2)
GANTRY DIAGNOSTICS
GANTRY - TEST - MECP: SELF TEST
GANTRY - TEST - MED: SELF TEST (DETECTOR 1)
GANTRY - TEST - MED: SELF TEST (DETECTOR 2)
Follow the steps below to start a “Standard Diagnostic” session on an ecam Macintosh based
system.
2. Open the ICON 8.5 CD and then open the following folders in the order listed:
• ecam
• ICONF
• Service
• Tools
4. Select New Diagnostic Session from beneath the “File” menu item.
01. Insert the ecam STANDARD PLUS CD-ROM into the SNAC computer
CD-ROM drive. (Make sure it is the correct revision) CD needed if
running “Standard Plus” diagnostics.
02. Verify that the esoft application is running by:
• The first way is to double click on the Internet Explorer icon on your
desktop. This will take you to the “Service Software Authentication”
window if it is still setup from the factory. Now you can enter in the
passwords you received from the Uptime Center. If someone has changed
the configuration of the Internet Explorer then you will have to use the
second way of launching the diagnostics.
• The second way of launching the diagnostics is thru the esoft application
software. Follow the steps listed below to get to the “Service Software
Authentication” window:
o You must start the esoft application in order to gain access to the
diagnostic package. Launch the “Start E.Soft” application using the
shortcut located on the esoft Workstation desktop.
05. This is the step that requires you to enter a password to continue.
11. You can now scroll down through the diagnostic test listing until you find
the one you wish to run. Click once on the diagnostic you wish to run to
highlight it, the click on “Execute”. Make sure you follow the
instructions and read the warnings for the diagnostic you are running.
FEATURES
DETECTOR #1
This procedure allows the calibration of the Collimator C-Frame on detector #1.
CAUTION:
• When a new position pot is installed, or in any case when the pot position
is unknown, leave the gear set screw loose on the pot, so the pot will not
be damaged. This is a ten turn pot and 9.5 turns are used in the range of
the C-Frame motion from open to closed. If the gear set screw is tight on
the pot, the pot could easily be damaged by the running past its end point.
The Calibrate C-Frame diagnostic does not depend on the pot value for
opening and closing the C-Frame.
NOTE:
• If Calibrate C-Frame initially fails with the error message “Collimator
partially installed” it may be because the Collimator Cleared Switch is
activated by the tub. It will be necessary in this case to open the C-Frame
enough to deactivate the Collimator Cleared Switch. Open the C-Frame
on one of two ways:
o Run Exercise C-Frame. The C-Frame opens and the diagnostics
pause. At this point, click the Close box in the upper left corner of
the diagnostic window. The diagnostics will exit with the C-Frame
open. Then run Calibrate C-Frame.
o Use the MEDP switches (see MEDP service map) to open the
C-Frame enough to deactivate the Collimator Cleared Switch.
Then run Calibrate C-Frame.
• Prompt
i. Calibrate and lock down the potentiometer.
ii. Press return to continue.
11. Adjust the potentiometer with a 1/16” allen wrench to the expected value,
lock it down and press return.
12. C-Frame opens.
• Prompt
i. Verify the C-Frame is in the open position.
ii. Enter Y or N
NOTE:
• When you have completed with diagnostics, you must EXIT and CLOSE
the diagnostic session.
WARNING!
When you enter into and while you are running diagnostics,
ALL SYSTEM FAIL SAFES are DISABLED!
Purpose:
• The following steps and procedures will assist you in setting up the ecam camera system for
preparing the source for Peaking, Tuning and performing an Intrinsic Flood. Always have an
authorized site person prepare the source for you, but you can request the way it should be
prepared and the amount of activity needed.
NOTE:
• Have authorized site personnel prepare the source as described below. Use as little liquid as
possible and ensure that the point source’s activity is correct using the attached table.
01. Place a small amount of cotton into the bottom of the plastic vial. Be sure not to use too much
cotton. Cotton should cover the “bottom” of the vial only and not fill the vial.
99
02. Place the drops of Tc on the cotton. Do not exceed the cotton’s saturation capacity.
99
03. Use the chart below to determine the amount of Tc activity needed for calibration to be
performed.
NOTE:
• A deadtime between 3% and 9% is recommended.
NOTE:
• Before performing step 6, you might wish to cut a finger out of a latex glove and put your prepared
source inside the finger. This will help prevent possible leakage and crystal contamination.
06. Clamp the capped end of the vial into the source holder and ensure that the cotton tip with the
activity is approximately centered above the detector.
07. For a Dual-Detector system, raise or lower the bed until the source is centered between the
detectors. The count rates shown on the PPM display should be approximately equal.
NOTE:
• Make sure there is no stray or scatter radioactivity in the department or system room. (With
collimator(s) removed and no source, you should see no more than 400cps.) An injected patient or
an unshielded dose, even if they are in different rooms, may produce a sufficient amount of scatter
activity to cause peaking and tuning problems. (More likely to present tuning issues.) Without
collimator(s) installed, the camera is significantly more sensitive than with the collimator(s)
installed. A typical patient is injected with several mci of activity, which is 10-100 times more than
the amount of activity used for peaking and tuning. Simply put, an injected patient in another room
or if that patient even just walks by the system room, can have a negative effect on peaking and will
negatively effect tuning causing tuning problems or tune failure.
NOTE:
• The Spectrum display is located or shown on the Analyzer tab.
NOTE:
• The spectrum display will stop growing when it reaches approximately 32,000 counts. At this point
you will need to press the “Stop” button as it does not automatically become disabled. Press the
“Start” button to restart the spectrum.
NOTE:
• When you peak the detector, you are asking the system to slightly shift the analyzer’s window so
that it is centered around the energy peak as detected by the camera. Only one detector is peaked
at a time. If you have a Dual-Detector system and both Detector 1 and Detector 2 are activated, it
will peak Detector 1 then peak Detector 2. During peaking, you will notice a loss of activity on the
PPM display for the detector being peaked.
NOTE:
•
99 57
It is recommended to “Fine Tune” the detector(s) if the Tc or Co shift is larger than +3.0%. This
will ensure the stored Peak Shifts for other isotopes remains accurate.
Tuning the detector helps you tune the photomultiplier tubes, (PMTs), automatically inside each detector.
“Fine Tune” adjust the PMT gains slightly, while a “Coarse Tune” provides major adjustments of PMT gains
and is not normally required. The available gain range for tuning each PMT is –512 to +511 and the HV
range is –128 to +127. Only one detector can be tuned at a time.
Fine Tune:
NOTE:
• The processor will tune the PMTs in six or less steps which are referred to as “passes” in the
DAC/PMT window’s status field. These passes begin at six and work their way down to one. The
passes then start over at six and then count a second time. During these passes, the processor
adjusts the PMT gains so that the tune values are less than 7, (read 0.7%), for the center PMTs
and 15, (read 1.5%), for the PMTs on the edges. Tune values represent the amount of change
required to perform optimally.
06. The system will stop tuning over the course of its maximum of 12 passes.
07. To display the tune values, click on the “Display Tune Values” button. The value assigned to each
PMT after the current pass is displayed in the field above each PMT number.
NOTES:
• When all the tune values are less than 7, (read 0.7%), for the center PMTs and 15, (read 1.5%), for
the PMTs on the edges, tuning has completed successfully.
• The acquisition controller displays an error message if the system is unable to complete tuning
successfully. (If so, follow the instructions given in the Tuning window’s message field.) If tuning
should fail, it may be necessary to exit the Tuning page, reopen it, then redo the tuning procedure.
Coarse Tune:
01. Proceed as instructed for “Fine Tuning” the detector with the exception of step 5; click on the “Start
Coarse Tuning” button.
02. Click on the “Display Gain Values” button to display the gain values assigned to each PMT after the
current tune pass. The “High Voltage DAC” value is displayed which depicts how much DAC
output is needed to tune the detector. The gain value, the voltage applied to each PMT to bring it
into tune, appears in the field above each PMT.
03. After a Coarse Tune has been performed successfully, always perform a Fine Tune.
NOTES:
• If a “Coarse Tune” is performed or the DACs are cleared, all analyzer isotope presets will have to
be repeaked.
• If the PMT DACs are cleared, a “Coarse” and “Fine” Tune must be performed as described.
Dual Rod
System
Source Vial
Rear Bed
Single
Rod
Source Vial
Tc99
Vial
Cotton or Guaze
Source Preparation
Frequency/Problem Calibration
99 57
Daily Peaking for Tc or Co
Fine tune, if peaking value is larger than +3.0%, then re-peak.
(NOTE: This value becomes +2.0% if on a Profile system.)
Daily Intrinsic Flood QC
Monthly Peak
Tune
Peak
MHR/COR for each collimator set used for tomo.
NCO Calibration (dual-detector variable-angle and cardic systems only)
Monthly Intrinsic Flood >200 million counts
Whenever Collimator Peak
Damage Occurs Tune
Peak
Monthly Intrinsic Flood >200 million counts
Monthly Extrinsic Flood for the damaged collimator.
Calibration Sequence - The proper sequencing of calibrations is as follows:
1. MHR/COR
2. NCO (Dual-Detector variable-angle and cardiac systems only)
3. Intrinsic Flood (Daily/Monthly)
4. Extrinsic Flood (if collimator damage occurs)
03. There are four mandatory fields, (in bold), which must be completed or filled out.
a) Last name (For our purpose, we used “ECAMTEST”)
c) Date of Birth (For our purpose, we just entered the current date.)
e) All other information can be left blank for running test images.
04. If you are now ready to perform test study, click on “Exam”.
Camera-Patient Setup_01
Camera-Patient Setup-02
DETECTOR 1 AT 00
ARMREST ENGAGED
PATIENT
PALLE
T IN
REAR BED DOWN
DET
ECT
OR 2
AT 1
80 0
Select Autocontour
VERIFY BOTH
VERIFY BOTH
DETECTORS
DETECTORS
ARE ACTIVE
ARE PEAKED
WBA Scanning_01
BOTH DETECTOR
PATIENT RADIUS MOVED PALLET
IN TOWARD SCANNING OUT
CENTER
LEFT
LOOK AHEAD
DETECTOR 1
LEFT
LOOK
AHEAD
PATIENT
WBA Scanning_03
DETECTOR 1
COLLIMATOR
DETECTOR 1
COLLIMATOR
PATIENT
WBA Scanning_05
WINDOWING
BAR
DETECTOR 2
DETECTOR 1 IMAGE
IMAGE
Workflow Stopped
This procedure will assist you in removing the C-Frame assembly. (It will be necessary to
remove the “C-Frame” if you have a need to put the detector into the “Detector Service
Position”.) This procedure assumes the person performing it obtains the basic knowledge to do
so. It is recommended you thoroughly read and understand this procedure before actually
performing this procedure.
While performing this procedure, you must understand, follow and adhere to all safety
guidelines and procedures concerning, however not limited to the following: personal,
equipment, environmental, electrical and radioactive.
Items Needed:
• Your service tool kit.
• C-Frame Closed Jumper (Siemens part number 5223248) (See Figure: 1)
• CAUTION: While system is in “diagnostic mode”, there are “NO” system fail-safes.
• NOTE: System must be “homed” and in the 180o configuration to perform this procedure.
(See Figure: 2)
• NOTE: This procedure assumes “Standard” diagnostics is used for some removals as
stated in this procedure.
01. Remove the detector processor cover from the detector being serviced to gain access to
the C-Frame. (See Figure: 3)
02. Rotate the gantry so the detector being serviced is at 90o or 270o, or the detectors are in
the collimator change positions. (See Figure: 4)
03. Remove the collimator from the detector being serviced.
• CAUTION: When removing the 5mm hex bolt and stop washers from the C-Frame
bearing posts in Step 4, you must take note and record any washers which are
removed. Some C-Frames have “shims”, (large and small), used for C-Frame
alignment. These “shims”, (washers), “must” be replaced on the same posts from
which they were removed.
04. Use a 5mm hex wrench to remove the stop washers from the C-Frame bearing posts.
(See Figure: 5)
05. Enter into a “Standard” diagnostic session.
06. Perform diagnostic “Exercise C-Frame”, (for the detector or C-Frame being serviced), to
“OPEN” the C-Frame.
07. Disconnect J10, J11, J12 and J13 ribbon cables from the MED/U board. Take care
when disconnecting these cables, the connection is “key locked” and make sure these
cables are not even slightly bent or twisted; these cables are very easily damaged.
(These are the ribbon cables from the “Light Rails” and can be disconnected with power
applied.) (See Figure: 5)
08. Disconnect J4 from the MED/U. Take care when disconnecting this cable, the
connection is “key locked”. (Connection J4 can be disconnected with power applied.)
09. Loosen the tie wraps securing the ribbon cables to the C-Frame motor.
10. Use a 2.5mm hex wrench to remove the C-Frame “lock plate” from the rear of the
C-Frame. (See Figure: 6)
• CAUTION: Carefully observe that: all cables clear without getting pinched; if
removed, the CIDI cable does not get caught on the detector flange; and the C-
Frame motor extension arms feed into the holes of the detector flange without
binding.
END
This procedure will assist you in putting the ecam detector into the “Detector Service Position”.
In the “Detector Service Position”, you are removing the inner workings of the detector, (crystal,
optical spacer, PMTs, preamps and inner cables from the detector tub. It is necessary to place
the detector in the “Detector Service Position”, as an example, if it is necessary to remove and
replace a PMT. This procedure assumes the person performing it obtains the basic knowledge
to do so. It is recommended you thoroughly read and understand this procedure before actually
performing this procedure.
While performing this procedure, you must understand, follow and adhere to all safety
guidelines and procedures concerning, however not limited to the following: personal,
equipment, environmental, electrical and radioactive.
Items Needed:
• Your service tool kit.
• Four pieces of “2x4” straight lumber cut to approximately 24” to 27” lengths. (Two
pieces of “4x8” can be used if that is all, which is available). A clearance of 4” to 6” is
required.
• C-Frame Closed Jumper (Siemens part number 5223248)
• CAUTION: While system is in “diagnostic mode”, there are “NO” system fail-safes.
• NOTE: System must be “homed” and in the 180o configuration to perform this procedure.
• NOTE: It is recommended before beginning with this procedure if not already, the
lightest weight collimators available on site be installed.
• NOTE: This procedure assumes “Standard” diagnostics is used for some removals as
stated in this procedure.
01. Remove the “detector processor” cover from the detector being serviced.
(See Figure: 1)
02. Rotate the detector being serviced to the 90o position as facing the front of the gantry.
(See Figure: 2)
03. Start a “Standard” diagnostics session.
04. Perform the diagnostic “Exercise C-Frame” and open the C-Frame for the detector being
serviced.
05. Remove the collimator for the detector being serviced. (See Figure: 3)
06. With C-Frame still open and in diagnostics, remove the C-Frame assembly.
(See Figure: 4 & Figure: 5)
07. With C-Frame removed and the C-Frame drive mechanism still open; install the “Closed
Jumper”, (Siemens #5223248), on the MED/U board at J4. (You will have to remove the
connector already installed. Safe to do this under power.) (See Figure: 6 & Figure: 7)
10. At this point, you should have motion from the hand controller. If safe, check to see by
attempting to rotate a couple of degrees and by moving radius of detector being serviced.
11. Turn off detector voltage. (See Figure: 9)
12. Turn off detector HV. (See Figure: 10)
13. Disconnect the HV BNC cable. (See Figure: 11)
14. Disconnect the ribbon cable from the CIDI board. (See Figure: 12)
15. Rotate the detector being serviced to the “top”, or at the 0o position. (See Figure: 13)
16. Remove the “pallet” from the “Front Bed”.
17. Lock down the “Front Bed” docker micro switches. (This can be done with tape or by
using a tie-wrap taking care not to damage the switch.) This must be done in order to
fake or make the system think or recognize the “Front Bed” is docked. No motion is
allowed if the system finds the “Front Bed” is not docked.
• CAUTION: When using the “Front Bed” or “PHS” to perform system servicing
procedures, you must not exceed 600 lbs. of weight or force on the “Front Bed” or
“PHS”. Reaching or exceeding this limit will cause damage to the “Front Bed” or
“PHS”.
• CAUTION: When placing or moving the “Front Bed” parallel to the gantry, be
aware the bottom part of the “Front Bed” is not the lowest part, it is the “support
frames” on either side underneath which are much lower. If necessary, use the
“Hand Control” to adjust the height of the “Front Bed” so it does not contact the
bottom or detector not being serviced.
18. Place the “Front Bed” parallel to the gantry. (See Figure: 14)
19. Slowly push the “Front Bed” between the detectors until the “Front Bed” stops travel.
(Travel will stop when the bottom front feet, (left and right side), of the “Front Bed” come
in contact with the front feet of the gantry, (left and right side).
20. Place two stacks of lumber, (each containing two “2x4”), on the “Front Bed”.
(See Figure: 15)
21. Raise the “Front Bed” to within 1” of maximum height or until lumber is within just less
than ½” of the detector tub. (If necessary, you may have to radius “IN” the detector
being serviced.)
22. Align the lumber so as the edges of the detector tub when lowered, will be resting on the
lumber leaving room to remove the 3 hex bolts on each side mounting the detector to the
detector tub. (See Figure: 16 & Figure: 17)
23. Before placing the detector on the “Front Bed” or lumber; perform the following:
• Lower the detector being serviced to within just less than ½” of the lumber blocks.
• Six M4 cap screws mount the detector to the tub. Use a 3mm hex wrench and
evenly loosen the six cap screws 3 turns CCW, starting first with the left side,
then the right side. This is to create a 2mm air gap between the detector and the
tub.
• NOTE: “Caudal Tilt” function should be active. If for some reason “Caudal Tilt” function
is not active at this time, on the MED/U board, set SW2 positions 6 & 7 to “ON”, press
the MED/U reset switch and wait for tone. When tone is heard, “Caudal Tilt” function
should be active. (See Figure: 18)
• CAUTION: To prevent damage to the detector and detector flange, make sure and
observe while lowering the detector from the tub, the detector does not “hang up”
or “bind”, and the detector and detector flange separates evenly “front-to-back”
and “side-to-side”.
• NOTE: It is recommended, before disconnecting any cables, you make at least a mental
note of placement and folds of all the cables. Also recommended, before disconnecting
any cable, if not marked or tagged where goes, you do so before disconnecting as well
as marking or tagging any cables that might have a “tie-warp” so these can be replaced
as found before reinstalling the detector. (You also might wish to note the “number” of
cables disconnected.) To avoid cable damage and/or system damage, it is critical
all cables be properly routed, tie-wrapped and connected before reinstallation of
detector.
29. Using the “Hand Control”, continue lowering detector until there is enough clearance,
(making sure you do not lower too far or far enough to “stress” any cable or connector),
to access inner detector cables so they can be disconnected. This would include, not
limited to, all board cable connections, board coax connections and fan cable
connections.
30. Observe that the disconnected CIDI cable, (at rear of detector tub), does not get caught
or bound on the detector flange openings.
• CAUTION: While lowering the inner detector assembly for the detector tub,
observe that you have plenty room between the “Front Bed” and the bottom
detector.
31. Once you have verified all inner head cables have been disconnected and will not bind,
slowly and carefully continue to lower the inner detector until it clears the detector tub.
32. When you have verified proper clearance between the inner detector and the detector
tub, you may now slowly and carefully move the “Front Bed” out to gain access to the
inner detector assembly for service.
33. Once you have completed necessary service, reverse this sequence to reinstall the inner
detector into the detector tub.
• CAUTION: While reinstalling inner detector into detector tub, be sure to observe at
all times, all cable clear without getting pinched, the CIDI cable does not get
caught on the detector flange and the C-Frame motor extension arms feed into the
holes of the detector flange without binding. (See Figure: 19)
34. Using the “Hand Control”, move the inner detector assembly into the detector tub.
“STOP” all motion when the inner detector is within 2mm of the detector tub.
35. Using a 3mm hex wrench, tighten the six, (three each side), M4 cap screws, first the left
side and then the right side evenly until all screws are tight.
• NOTE: If you had to activate “Caudal Tilt” using SW2 on the MED/U board, then now on
the MED/U board, set SW2 positions 6 & 7 to “OFF”, press the MED/U reset button and
wait for the tone. When you hear the tone, continue with procedure.
• NOTE: If at this point you “do” observe a “hole” in the flood, this generally points to either
an HV wire or data cable is either improperly connected or not connected. This requires
putting the detector back into the “Detector Service Position” and verifying connections.
• NOTE: If at this point you observe “no” counts or activity from the detector, which was
serviced, and the detector power is on and supplies have been verified, this generally is
due to a loss of “detector communication”. To attempt to reestablish or restore detector
communications, first attempt a “soft reboot” of the SNAC computer from the PPM. If
this does not work, properly shutdown the ecam system and then properly re-apply
power to the ecam system. After reestablishing detector communications and you have
counts or activity, make sure no objects are in the way of the gantry and “home” the
system; then continue with procedure.
41. Remove the tape or tie-wrap or what was used to close the “Front Bed” docker
assembly’s micro switches, then store or dock the “Front Bed” in the “storage” position.
42. Reinstall the C-Frame assembly, (reassemble in reverse order).
43. With the C-Frame Closed Jumper removed from J4 on the MED/U board and the original
connector install at J4, press the MED/U reset button.
44. “Home” the system.
45. Perform the “Fan Current” check procedure to insure all fans are running inside the
detector.
46. Verify proper collimator change sequence.
END
This procedure explains the basic steps required to remove a “gelled” tube from a Siemens
ecam detector. This procedure applies to both 2” and 3” PMTs. This procedure also applies to
all digital detectors using gel to bond the PMTs to the optical spacer. Also included, are basic
cautions required when performing this procedure. It is suggested you completely read and
understand this procedure before beginning.
While performing this procedure, you must understand, follow and adhere to all safety
guidelines and procedures concerning, however not limited to the following: personal,
equipment, environmental, electrical and radioactive.
NOTE
• Some tubes, ( 1,5,8,28,32,52,55 and 59 ) have capacitors attached from the PMT
preamp to the shield. Use a grounded soldering iron to remove the wire from the PMT
preamp. The wire is tacked to one leg of C103 or C4. (Refer to diagram below)
SOLDER
C103 or C4
CAPACITOR
( CERAMIC DISK 4700pF +80/-20%
2KV )
PMT
SHIELD
C1
D1
03
R112 R127
R112 R127
D8
D2 D8
D2
1
D7 1
D7
D6 D5 D4 D3
D6 D5 D4 D3
FF5 PREAMP BOARD
FF6 / FF7 PREAMP BOARD
FE
U4 AMP K
C4
D1
U1 DAC
R13
D8
D2
1
D7
HV
D6 D5 D4 D3
08. Apply pressure to the PMT in a forward direction to begin to release the gel seal. Then
apply reverse pressure to the PMT to begin to release the gel seal. Now apply pressure
to the right and then pressure to the left as in a “twisting motion”. Repeat as necessary
until the gel bond starts to release. (See diagram below)
10. Twist and pull up on the PMT, gripping the glass, until it is out of the shield.
11. Using a plastic scraper, very carefully remove all visible signs of the gel residue from the
optical spacer. Clean the optical spacer thoroughly, but do not use any solvents or
alcohol prep pads, which will prevent the new tube from bonding properly.
NOTE
• The detector must be level to ensure that the gel, after it is distributed, stays
evenly under the PMT.
END
This procedure assumes you have followed the proper processes/procedures to include all
safety procedures for removing the bad or suspected PMT and are now ready to install the new
or replacement PMT. This being the case, proceed with the PMT installation process.
While performing this procedure, you must understand, follow and adhere to all safety
guidelines and procedures concerning, however not limited to the following: personal,
equipment, environmental, electrical and radioactive.
Normal Process
01. Clean the new PMT face with a dry soft cloth. “DO NOT” use any solvents or cleaners
as these can cause improper bonding issues.
02. Remove the divider bar from one packet of gel and squeeze the contents in the upper
compartment into the bottom one. (Make sure you are using the proper gel packet for
the size PMT you are installing.) Gently knead the packet for approximately 60 to 90
seconds to thoroughly mix the contents. (While mixing gel packet contents, attempt to
eliminate at many air bubbles as possible.)
03. Pour the entire contents of the gel pack mixture onto the “center” of the optical spacer
pad. While pouring the gel onto the center area, “DO NOT” attempt to spread the gel,
let the gel spread itself.
04. Wait approximately 3 to 5 minutes for the air bubbles to escape.
05. Immediately distribute evenly as possible approximately 15 to 20 1mm glass beads in
the gel.
Page 203 of 232
06. Correctly orient and align the PMT just above the gel and set it onto the gel puddle. “DO
NOT ROTATE THE PMT!” Rotation of the PMT may cause the glass beads to move
out from under the PMT.
07. Place the proper weight base and weight on the PMT. (The weight kit includes bases for
round and hex preamps and weights for 2” and 3” PMTs. Both weight base and weight
“must” match the shape preamp and size PMT you are installing for proper bonding.)
Allow the PMT to set with weight for at least 4 hours at normal room temperature. If
room temperature is below 70oF, let the PMT set with weight for a longer time period;
approximately twenty minutes per degree for every degree below 70oF. After 4 hours,
(time could be longer depending on room temperature), remove the weight and carefully
begin re-assembly of the detector. (See Figure: 1 & Figure: 2) DO NOT INSTALL
DETECTOR IN TUB AT THIS TIME!
08. Certain PMTs, (1, 5, 8, 28, 32, 52, 55 and 59), have capacitors attached from the
preamp to the PMT shield or “MU Shield”. Solder the wire to the capacitor on the
preamp board if this is the case. (If you are replacing on of the above stated PMTs, this
capacitor wire should have been removed before removing the PMT.) (See Figure: 3)
09. Reassemble and re-cable the detector.
• DO NOT PUT THE REASSEMBLED DETECTOR BACK IN THE TUB AT THIS TIME
• Re-install HV wires by grasping the connector, not the wire. Make sure you install
HV wires as you removed them, oriented correctly if the connecting posts are
square.
10. Make sure HV is off and the HV coax cable is disconnected. Turn the detector on; apply
only “low voltages” to detector. Let the detector warm up for 10 minutes.
11. Using a DVM, attach the positive lead to the test point on the DD59(C)(CF)(R) board for
the replaced preamp. (These test points are number as the PMTs; TP1 thru TP59.)
Attach the negative lead to TP67.
12. Adjust R112, Preamp Zero Adjust, (located on the preamp), for 0V, +500uV. (This
adjustment is only available on the FF6 and FF7 preamps.
13. Power down the detector.
• Once you have attached all HV wires, Data cables and connectors, it is
recommended you “double check” all HV wire, Data cables and connectors for
proper placement and secure connection. Also make sure all cables are “tie-
wrapped” as you found them and cables are positioned correctly for proper
“folding” when detector is placed inside the tub.
• If you see no counts or activity from detector in question, at this point, it may be
necessary to perform a “Soft Boot” of the SNAC from the PPM to reestablish
detector communications.
17. At this time, using a source, (15uci to 20uci), check to make sure you have no “holes” in
the flood.
• When the PMT Gain DACs are “cleared” or a “Coarse Tune” is performed, the
system no longer retains “peaking data”! As a result, it will be necessary for the
customer to “re-peak” all the isotopes they use. Once an isotope has been
properly peaked, the system will retain the peaking data for this isotope. It is vital
you make your customer aware of this.
21. Select the proper isotope preset. (You should be using 99Tc)
22. Peak the analyzer. (Peak shift percentage should be +3.0%)
23. Perform a Coarse Tune. When the Coarse Tune is completed, view the Gain Values
display map.
24. Inspect the PMT gain values for any value greater than 200 from the default value. (The
“default” value can be found in the “Detector Service Log Book” or the PMT gain value
all PMTs displayed after the DACs were cleared.)
• There are several different default values which is factory determined depending
on the gain of the PMTs used in the detector. An example default value would be
–209.
25. If any PMT gain value is found to be greater than 200 from the default value, the gain of
these PMTs will need to be adjusted.
• Example: If your PMT DAC gain default value is –209, given a 200 gain range value,
your PMT DAC gain value range after Coarse Tune would be –009 to –409.
26. If PMT DAC Gain adjustments where necessary, when completed, allow detector to
warm up for at least 30 minutes before proceeding.
27. Once detector has warmed up, clear the DACs, re-peak the system and perform Coarse
Tune.
28. If Coarse Tune passes and all PMT DAC Gain values are within specifications, continue.
If not, then go back to Step 20.
29. Perform Fine Tune.
30. Once Coarse Tune and Fine Tune have been performed and passed with all values in
specifications, perform a uniformity flood.
31. Perform “Post Repair Detector Testing”.
3528978 3529000
BASE FOR BASE FOR
NON-ASIC HD3 / ASIC
PREAMPS PREAMPS
BOLT ATTACHES
WEIGHT TO BASE
WEIGHT
3529703
EYE BOLT
PREAMP
TUBE
BASE
FITS OVER
PREAMP
Figure: 1
PMT WEIGHT
BASE
INSTALL NEW PMT, (WITH PREAMP), MAKING SURE PREAMP ORIENTATION IS THE
SAME AS ALL OTHER PMT’S AND PREAMP’S. ONCE PMT HAS REACHED THE GEL,
VERY SLIGHTLY APPLY DOWNWARD PRESSURE AND TWIST, (NOT MORE THAN ¼"
EITHER DIRECTION), JUST ENOUGH TO SEAT THE PMT IN THE GEL. (CAUTION: DO NOT
APPLY ENOUGH PRESSURE OR TWIST IN AN ATTEMPT TO SPREAD THE GEL. IF SO,
PMT WILL NOT BOND PROPERLY!) ONCE PMT IS SEATED, PLACE WEIGHT AND BASE,
(ASSEMBLY REQUIRED), ON TOP OF PMT/PREAMP MAKING SURE IT IS LEVEL.
(CAUTION: IF WEIGHT KIT NOT LEVEL, WILL CAUSE IMPROPER BONDING!) LEAVE
WEIGHT KIT ON FOR AT LEAST 3 TO 4 HOURS UNDISTURBED ALLOWING FOR INITIAL
SETUP TIME. AFTER THIS INITIAL SETUP TIME, THE WEIGHT CAN BE REMOVED AND
REASSEMBLY OF THE DETECTOR CAN BEGIN.
CAUTION: DO NOT ROTATE OR FLIP DETECTOR FOR AT LEAST 24 HOURS FROM THE
TIME THE WEIGHT WAS FIRST PLACED ON PMT!
(BE AWARE THERE IS A DIFFERENT WEIGHT FOR BOTH 2" AND 3" TUBES, AND A BASE
FOR BOTH HEX AND ROUND PREAMPS.)
Figure: 2
C1
D1
03
R112 R127
R112 R127
D8
D2 D8
D2
1
D7 1
D7
D6 D5 D4 D3
D6 D5 D4 D3
FF5 PREAMP BOARD
FF6 / FF7 PREAMP BOARD
FE
U4 AMP K
C4
D1
U1 DAC
R13
D8
D2
1
D7
HV
D6 D5 D4 D3
Figure: 3
Tuning
Uniformity Flood
01. For each detector, acquire a 30 million count uniformity flood. Choose the appropriate
section:
a. ecam Systems with ICON Computer
b. ecam Systems with esoft Workstation Computer
• For the ecam Duet system, there will be three additional intrinsic floods to be performed
unless the customer is not licensed for any isotopes within a particular energy range.
These energy ranges and isotopes are:
o Very Low Energy – Thallium – Tl-201 – (70keV)
o Medium Energy – Indium – In-111 – (274keV)
o High Energy – Iodine – I-131 – (364keV)
• The calibration floods patient folder is now created. Select this patient for each
calibration performed in this procedure.
• When replacing or regenerating a Duet detector, all numbers should be compared to the
“Service Log” specifications. For all other servicing, intrinsic flood should meet the
uniformity specifications for HD3CF – (Deut only).
• If “Uniformity on the Fly” feature is used, (monthly 200-million count correction flood),
perform the following steps:
END
57
Co or 99Tc CENTRAL FOV USEFUL FOV
TI-201 (VERY LOW) CENTRAL FOV USEFUL FOV TC-99m (LOW) CENTRAL FOV USEFUL FOV
Co57 (LOW) CENTRAL FOV USEFUL FOV In-111 CENTRAL FOV USEFUL FOV
(MEDIUM)
ITEM FUNCTION
AIMD Board Analog Interface Microprocessor-Digital. Provides the PI59 with data about
the event to aid in integration. The board has two main circuits; Event
Controller and Micro-processor Controller.
Analog Amplifier
Four analog amplifiers convert digital velocity data received from the PMAC and
supply current to each of the axis motors. The four amplifiers are identical except
the rotate amplifier can drive more current. Each of the amplifiers except the
rotate amplifier dirve 2 axes through PMAC controleed relay switching.
Clean Box Assembly with a toroidal isolation transformer, voltage taps, built in noise
filters and relay circuitry. Converts 230VAC input to filtered and isolated
125VAC. Also has relay circuitry which provides surge protection on
power up.
Coarse Tune Process to adjust the HV and gain of each PMT until the energy from that
PMT fall in the center of the tune window. Edge PMT's need to be within
2.3%, all others must be within 1.4%.
COP Correction Processor. Name for a collection of circuits on the DPC board
which take the data from the HD3 detector and prepare it for storage and
display.
COPC Correction Processor Control. Part of the Correction Processor which makes
sure all processes operate synchronously so that no events are lost or jumbled.
COPD Correction Processor Datapath. Part of the Correction Processor which performs
actual operations on the event data.
Dark Current When a PMT sends a signal with no source present, HV turned on and no light
from the crystal. "Unwanted Noise" which can effect image uniformity.
Detector P/S - (OEM) Provides +5VDC and +20VDC to the DTPS board. Voltages are used by the
detector PCB's and the detector. Also referred to as "OEM" supply.
DD59 Detector Master Board. Functions as a cable harness to connect the PMT
preamp outputs to the PI59 board. Also provides voltage to the PMT's and
biases the PMT MU-metal shield to the HV. Temperature sensors are located on
this board.
DPC Detector Processor Card. Installed in the SNAC computer, which performs
gamma event correction processing.
DTPS Detector Tube Power Supply. DTPS provides power to the detector boards;
(AIMD, PI59 and DD59). Also HV to the DD59 which goes to the preamps and
PMT's.
EVE Key Enables only those parts of the ICON software that the customer has purchased.
To see what options the customer has purchased, run a program called EVE
READ found in the utilities folder on the ICON software CD.
ITEM FUNCTION
LFR Light Response Function. The typical response of a PMT to an event introduces
some distortions. These Light Response Functions are corrected for in the
PCALC board.
MDB Motor Distribution Board. At each axis is an MDB which serves has a harness to
break out the various signals to their individual components.
MEB Motion Electronics Base. Both the MECP/MECP2 and the PMAC interface to the
MEB directly through connectors. Power supplies and each axis are also
connected to the MEB.
MECP Motion Electronic Control Processor. A single board computer which is the
"master" controller of all motions. The MECP/MECP2 communicates to the
SNAC, PMAC, MED/MEDU over separate serial RS485 connections. The
MECP/MECP2 monitors the system status via various sensors at all times and
decides whether the motion is allowed or not. (MECP2 is downward compatible
with systems running 2.1 and higher software).
MED/MEDU Motion Electronics Detector. One for each detector located under the detector
cover, are responsible for Caudal Tilt, C-Frame operation and auto contour light
curtains and look aheads. Both MED/MEDU are communicating to the
MECP/MECP2 via RS485 and are connected to the MEB. The MECP/MECP2
can override the MED/MEDU.
PCALC Position Calculator. Employs parallel signal processing to perform the following
functions: Pile-Up correction, LRF corrections, centroid calculation, correction for
offset errors, and scaling and positioning of the event.
Pedestal Input level to each of the FLASH ADC inputs on the PI59 board. This level is set
so that the input is just above the noise floor (inherent noise level of the system at
rest)Input level to each of the FLASH ADC inputs on the PI59 board. This level is
set so that the input is just above the noise floor (inherent noise level of the
system at rest).
PI59 PMT Integrator. Contains the Flash ADC's that digitize the preamp outputs.
PMAC Programmable Multi-Axis Controller. Dedicated real-time motion controller. It
directly controls the motor current to each axis through four analog amplifiers with
encoder and/or potentiometer feedback. When a command for motion is
received from the MECP/MECP2, the PMAC will switch the amplifier to the axis,
enable the amplifier and supply a digital value which is converted on the amplifier
to an analog voltage representative of the desired axis velocity.
POST Power On Self Test. Series of test performed by the computer when power is
switched on.
PPM Patient Positioning Monitor. The ecam flat panel display, which is installed on the
end of a moveable boom arm. It allows the nuclear technician to properly
position the patient for image acquisition. The PPM also provides gantry status,
collimator information and a diagnostic interface.
SCL Serial Command Link. Mode of communication between detector components
and the console. The SCL uses a RS485 at a baud rate of 38.4KHz.
SCSI Small Computer System Interface. High speed data bus used for connecting
hard disks, tape drives, and other accessories to the computer.
Serial I/O Board The communications interface board installed in the SNAC computer. Channel
A (RS232) communicates with the MEB and gantry. Channel B (RS485)
provides SCL communications to the detector(s).
ITEM FUNCTION
SNAC Siemens Nuclear Acquisition Computer. This computer, along with the ICON
computer, provides control of the ecam system.
SPECT Single Photon Emission Computed Tomagraphy. Rotational study around the
patient in which a series of images at different angles are taken. These views
can be reconstructed later.
TAXI Transparent Asynchronous Xmitter-Receiver Interface. High speed data link from
the PCALC to the DPC board in the SNAC.
Tower The ecam electronics tower contains the clean box, Motion Electronics Base,
(MEB), SNAC computer and ICON computer. The ICON monitor and keyboard
may sit on top of the tower.
Tune Map Found under ICON Camera Tools (DAC/PMT). This is a display of how close
each PMT photopeak is to the keV that is used for tuning. When the detector is
tuned correctly, both coarse and fine tune are less than 1.4% and inside PMT's
are less than 0.&% of keV used for tuning.
SIEMENS
DESCRIPTION PART NUMBER WHERE USED INTERVAL
CHRISTO LUBE MCG 200 GREASE 5224634 PATIENT FRONT BED ACME SCREWS BI-ANNUALLY
WAX STICK LUBRICANT (GS-20-1.6 OZ. STICK) 5224030 COLLIMATORS & SLIDE RAILS AS NEEDED
M-24 GREASE (2 OZ. BOTTLE) 5536867 YOKE BALL SCREWS ANNUALLY
GANTRY LINEAR BEARINGS ANNUALLY
PATIENT BED FRONT LINEAR BEARING BLOCKS ANNUALLY
C-FRME DRIVE ASSEMBLY ACME SCREW BI-ANNUALLY
C-FRAME DRIVE ASSEMBLY SPUR GEAR BI-ANNUALLY
REAR BED LINEAR BEARING BLOCKS ANNUALLY
OPEN GEAR 936 GREASE (IN CAULK TUBE) 569-000520 ROTATE RING GEAR ANNUALLY
RE-CONFIGURE LEVER AS NEEDED
BEVEL GEARS ANNUALLY
CP-28 GREASE (2 OZ. TUBE) 1242432 LEFT/RIGHT MOTOR ACME SCREW ANNUALLY
WD-40 OIL (12 OZ. SPRAY CAN) 915-000042 PATIENT BED FRONT DOCKER PINS AS NEEDED
COLLIMATOR CART DOCKER PINS AS NEEDED
C-FRAME BALL BUSHINGS (ON TUB) AS NEEDED
SIEMENS
DESCRIPTION PART NUMBER WHERE USED INTERVAL
PAINT - MED. WHITE (0.6 FL. OZ. BOTTLE) 2997562 GANTRY & PATIENT BED AS NEEDED
PAINT - MED. GRAY (0.6 FL. OZ. BOTTLE) 4372848 CABINETS - TOWER AS NEEDED
PAINT - MED. BLUE (0.6 FL. OZ. BOTTLE) 4387838 YOKE ASSEMBLY - GANTRY AS NEEDED
ACCESSORY BOARDS
After the completion of the tuning for normal mode narrow window, the system will
automatically tune for the extended range. The extended range is used for isotopes that
require any part of the window setting to be above 324kev. The HV value for the detector is
lowered by approximately 10%. This can be measured with a meter at TP8 on the DTPS
board. The reason the HV must be lowered is that the range of the 9 bit flash ADC's is
limited to a 2v signal. Isotopes above 245kev normally exceed this value. This results in a
need for a completely separate PMT DAC gain map.
COARSE
FINE
UFOV CFOV
3.74% 2.94% INTEGRAL
2.74% 2.54% DIFFERENTIAL
3528986 450-009202
LARGE WEIGHT SMALL WEIGHT
FOR 3" TUBES FOR 2" TUBES
3528978 3529000
BASE FOR BASE FOR
NON-ASIC HD3 / ASIC
PREAMPS PREAMPS
BOLT ATTACHES
WEIGHT TO BASE
WEIGHT
3529703
EYE BOLT
PREAMP
TUBE
BASE
FITS OVER
PREAMP
This procedure will assist you in putting the ecam system into “Manual Mode”. This procedure
assumes the person performing it obtains the basic knowledge to do so. It is recommended you
thoroughly read and understand this procedure before actually performing this procedure.
While performing this procedure, you must understand, follow and adhere to all safety
guidelines and procedures concerning, however not limited to the following: personal,
equipment, environmental, electrical and radioactive.
Items Needed:
CAUTION: While system is in “Manual Mode”, there are “NO” system fail-safes.
CAUTION: While system is in “Manual Mode”, “DO NOT” move Detector 1 radius
“out”. (If you move Detector 1 radius “out”, it may cause the “inner ring” to
disengage from the “outer ring” which could cause serious body injury and
equipment damage.)
01. When you see the message below displayed on the screen, press the keys on the
keyboard “Ctrl + Alt + Delete”.
02. When you see the “Logon Information” window display, for “User name” type “esoft”.
(Without quotations and in lowercase.)
04. Once the “User name” and “Password” have been entered correctly, you should see the
esoft boot screen displayed something similar to below.
01. When you see the message below displayed on the screen, press the keys on the
keyboard “Ctrl + Alt + Delete”.
02. When you see the “Logon Information” window display, for “User name” type
“Administrator”. (Without quotations.)
04. Once the “User name” and “Password” have been entered correctly, you should see the
esoft boot screen displayed something similar to below.