DPXUPDAT Version 4.

7e
DPXUPDAT Version 4.7e
Thu Jan 20 14:30:01 2005

SYSTEM FILE CONTENTS

CUSTOMER VALUES

Customer Serial Number = 5803

Hardware Scanner Type = DPXIQ240

Scanner Serial Number = 5803

AP Spine Scan Type Available? Y

Lateral Spine Scan Type Available? Y

Femur Scan Type Available? Y

Dual Femur Scan Type Available? N

Forearm Scan Type Available? N

Total Body Scan Type Available? Y

Pediatric Scan Type Available? N

Orthopedic Hip Scan Type Available? N

Small Animal Total Body Scan Type Available? N

Small Animal Appendicular Scan Type Available? N

Hand Scan Type Available? N

Research Scan Type available? N


Serial Port Number = COMM2

Serial Source = Comm Port

Serial Role = Receiver

Serial Baudrate = 19200

Serial Parity = Even

Serial Stop Bits = 1

Serial Data Bits = 8

IQ-NET Enabled = N

5ma Power Supply

DPXIQ240

CORRECTION VALUES

Low channel deadtime = 0.00000075

High channel deadtime = 0.00000075

Low channel thickness correction = 0.000000

High channel thickness correction = 0.000000

DPXIQ240

ACTUAL STANDARD MASS VALUES

Large chamber actual mass = 5.785000

Medium chamber actual mass = 4.300000

Small chamber actual mass = 3.050000


DPXIQ240

OBSERVED STANDARD MASS VALUES

Large chamber observed mass = 272.700012

Medium chamber observed mass = 203.000000

Small chamber observed mass = 144.199997

DPXIQ240

ACTUAL STANDARD WIDTH VALUES

Large chamber actual width = 55.000000

Medium chamber actual width = 47.500000

Small chamber actual width = 40.000000

DPXIQ240

OBSERVED STANDARD WIDTH VALUES

Large chamber observed width = 498.000000

Medium chamber observed width = 433.000000

Small chamber observed width = 361.000000


DPXIQ240

CALIBRATION VALUE SETTINGS

Source type = Ce X-ray

Amplifier current setting = 150.000000

Tube head voltage setting = 76.000000

Low keV attenuation coefficient = 9.950000

High keV attenuation coefficient = 6.200000

Detector collimation = 1.680000

DPXIQ240

CALIBRATION VALUE LIMITS

Low keV channel background test upper limit = 50

High keV channel background test upper limit = 50

Spillover test lower limit = 0.063000

Spillover test upper limit = 0.100000

Chi square test lower limit = 0.0

Chi square test upper limit = 40.0

Low keV channel air value lower limit = 450000.0

Low keV channel air value upper limit = 900000.0

High keV channel air value lower limit = 270000.0

High keV channel air value upper limit = 630000.0

Pseudo air ratio lower limit = 0.500


Pseudo air ratio upper limit = 0.700

True air ratio lower limit = 0.700

True air ratio upper limit = 0.900

Longitudinal steps lower limit = 19700

Longitudinal steps upper limit = 19800

Transverse steps lower limit = 12350

Transverse steps upper limit = 12550

Tissue value lower limit = 1.302

Tissue value upper limit = 1.320

Collimation ratio lower limit = 3.500

Collimation ratio upper limit = 4.500

Standard mean range criteria = 5.000

Standard covariance range criteria = 2.000

DPXIQ240

IN-TABLE CALIBRATION VALUES

Bone Value lower limit = 0.500

Bone Value upper limit = 0.700

Bone Value actual mass = 1.000

DPXIQ240
SCAN DISTANCE VALUES

Number of longitudinal steps on table = 19669

Number of transverse steps on table = 12288

Longitudinal distance per step (mm) = 0.10000

Transverse distance per step (mm) = 0.05000

DPXIQ240

SYSTEM POWER LIMIT VALUES

Maximum system current value = 5000

Maximum system voltage value = 100

FIRMVER v4.0 - SBC-03E Firmware Version Verification Program


(c)1990 - 2000 Lunar Corporation

The scanner now contains the following version of Boot firmware:

DPL BOOT 2.4

The scanner now contains the following version of firmware:

VER DPX 5.26