Professional Documents
Culture Documents
Depiction of Bias
and Unbiased
Measurements
Unbiased measurements relative to the
target
Page 2 of 7
Calibration
CALIBRATION
Instruments which The instrument reads in the same units as the reference
require Correction standards. The purpose of the calibration is to identify
for Bias and eliminate any bias in the instrument relative to the
defined unit of measurement.
Page 3 of 7
Calibration
Basic Steps for The calibration method requires the following basic
Correcting the steps:
Instrument for • Selection of reference standards with known values
Bias to cover the range of interest.
• Measurements on the reference standards with the
instrument to be calibrated.
• Functional relationship between the measured and
known values of the reference standards (usually a
least-squares fit to the data) called a calibration
curve.
• Correction of all measurements by the inverse of the
calibration curve.
Possible Form for There are several models for calibration curves that can
Calibration Curves be considered for instrument calibration. The simplest
one is the following which is linear:
Page 4 of 7
Calibration
Page 5 of 7
Calibration
Outliers in the Outliers in the calibration data can seriously distort the
Calibration Data calibration system. Isolated outliers should be deleted
from the calibration data.
Page 6 of 7
Calibration
Page 7 of 7