Professional Documents
Culture Documents
C
9
H A P T E R
Radiographic Film
Development1
More information on the radiographic examination, the silver of this image was
latent image, its formation and processing also discovered to be localized at certain
are available elsewhere.1-4 discrete areas of the grain (Fig. 2), just as
the latent image.
Thus, the process that made an
exposed photographic grain capable of
Introduction transformation into metallic silver (by the
Throughout much of photography’s mild reducing action of a developer)
history, the nature of the latent image was involved a concentration of silver atoms
unknown. The first public announcement
of Daguerre’s photographic process was
made in 1839 but it was not until 1938
that a satisfactory and coherent theory of FIGURE 1. Localized sites on grains.
photographic latent image formation was
proposed.5 That theory has been
undergoing refinement and modification
ever since.
Some of the investigational difficulty
arose because latent image formation is
actually a very subtle change in the silver
halide grain. The process may involve the
absorption of only one or, at most, a few
photons of radiation and this may affect
only a few atoms out of some 109 or 1010
atoms in a typical photographic grain.
Formation of the latent image, therefore,
cannot be detected by direct physical or
analytical chemical means.
A good deal was known about the FIGURE 2. Localized silver in printout image.
latent image’s physical nature. It was This is a T grain, a form used with soft
understood, for example, that the latent intensifying screens.
image was localized at certain discrete
sites on the silver halide grain. If a
photographic emulsion was exposed to
light, developed, fixed and then examined
under a microscope (Fig. 1), the change of
silver halide to metallic silver was visible
at only a limited number places on the
crystal. Because small amounts of silver
sulfide on the surface of the grain were
known to be necessary for high
photographic sensitivity, it seemed likely
that the spots where the latent image
formed were also concentrations of silver
sulfide.
It was further known that the material
of the latent image was probably silver.
For one thing, chemical reactions that
oxidized the silver also destroyed the
latent image. It was also a common
observation that photographic materials
given prolonged exposure to light
darkened spontaneously, without the
need for development. This darkening was
known as the printout image. The
printout image contained enough
material to be identified chemically as
metallic silver. By microscopic
Legend
= silver (Ag+) ion
Legend = bromine (Br –) ion
= silver (Ag+) ion = interstitial silver ion
= bromine (Br –) ion = latent image site
4.0
FIGURE 6. Characteristic curves plotted from
data of Fig. 9.
3.5
4.0
3.0
3.5
2.5
Net density
3.0
Fast film
2.0
Net density
2.5
Fast film
1.5
2.0
1.0 1.5
Slow film Slow film
0.5 1.0
0.5
1.5 0 0.5 1.0 1.5 2.0 2.5
0 10 20 30 40 50 60
Log exposure
Exposure (s)
Contrast
The slope of the characteristic curve for
film can change continuously along its
length. It has been shown qualitatively
that a density difference, corresponding to
a difference in specimen thickness,
depends on the region of the
characteristic curve where the exposure
falls. The steeper the slope of the curve in
this region, the greater the density
difference and hence the greater the
visibility of detail (assuming an
illuminator bright enough so that a
3.0 8.0
A Film Y
2.5 7.0
B
Film X
Density
2.0 6.0
1.5 5.0
Gradient
1.0 4.0
C
D
0.5 3.0
E
0 Film Z
2.0
0 0.5 1.0 1.5 2.0 2.5 3.0
b’ kilovoltage).
2.0
Average
gradient = a·b –1 = 2.3
1.5
a TABLE 2. Densities obtained through 13 to 16 mm (0.5 to
0.6 in.) steel sections by using exposure of 8 mA·min.
1.0
Relative
Energy Density
_________________ Radiographic Radiographic
0.5
b (kV) DB DA Contrast Contrast
Density = 1.5
_____________________ Density = 2.5
____________________
Relative Relative Relative Relative 0.5
Film Speed Exposure Speed Exposure
0
X 100 1.0 100 1.0 0 10 20 30 40 50 60
Y 24 4.2 26 3.9
Relative exposure
Z 250 0.4 150 0.7
may even decrease. In this case, fog than in the faster film to produce a
increases and contrast may decrease. particular density.
The increase in graininess with
increasing kilovoltage can also be
understood on this basis. At low
Graininess kilovoltages, each absorbed photon
Graininess is defined as the visual exposes one photographic grain; at high
impression of nonuniformity in the kilovoltages, one photon will expose
density of a radiographic (or many grains. At high kilovoltages, then,
photographic) image. With fast films fewer absorption events are required to
exposed to high kilovoltage radiation, produce a given density. Fewer absorption
graininess is easily visible with unaided events, in turn, mean a greater relative
vision; with slow films exposed to low deviation from the average and hence
kilovoltage X-rays, moderate greater graininess.
magnification may be needed. In general,
graininess increases with increasing film Screens
speed and with increasing radiation
energy. The above discussion of graininess applies
The clumps of developed silver also to exposures made with lead screens.
responsible for the impression of As stated earlier, the grains in a film
graininess do not each arise from a single emulsion are exposed by high speed
developed photographic grain. The electrons. Silver bromide cannot
particle of black metallic silver caused by distinguish between electrons from an
the development of a single photographic absorption event within the film
grain in an industrial radiographic film is emulsion and those from a lead screen.
rarely larger than 1.0 µm (4 × 10–5 in.) The quantum mottle observed in
and is usually less. The unaided human radiographs made with fluorescent
eye cannot see an individual grain. intensifying screens has a statistical origin
The visual impression of graininess is similar to that of film graininess. In this
caused by the random, statistical grouping case, however, the number of photons
of these individual silver particles. Each absorbed in the screens is significant. The
quantum (photon) of X-radiation or grain size of a fluorescent crystal is greater
gamma radiation absorbed in the film than that of silver bromide, so a spread
emulsion exposes one or more tiny function also contributes to
crystals of silver bromide. These nonuniformity.
absorption events occur at random. Even
in a uniform radiographic beam, the
number of absorption events will differ X-Ray Spectral Sensitivity
from one small area of the film to the
next, for purely statistical reasons. Thus, The shape of the characteristic curve of a
the exposed grains will be randomly radiographic film is unaffected, for
distributed and their numbers will have a practical purposes, by the wavelength of
statistical variation from one area to the the exposing X-rays or gamma rays.
next. However, the sensitivity of the film (the
With a very slow film, it might be number of coulombs per kilogram, or
necessary for 10 000 photons to be roentgens, required to produce a given
absorbed in a small area to produce a density) is strongly affected by the
density of, for example, 1.0. With an wavelength of the exposing radiation.
extremely fast film it might require only Figure 12 shows the number of
100 photons in the same area to produce roentgens needed to produce a density of
the same density. When only a few 1.0, for a particular radiographic film and
photons are required to produce the specific processing conditions (exposures
density, the random positions of the were made without screens).
absorption events become visible in the The spectral sensitivity curves for all
processed film as film graininess. On the radiographic films have roughly the same
other hand, the more X-ray photons that features as the curves shown in Fig. 12.
are required, the less noticeable the Details, among them the ratio of
graininess in the radiographic image, maximum to minimum sensitivity, differ
when all other conditions are equal. with film type.
In general, the silver bromide crystals The spectral sensitivity of a film or
in a slow film are smaller than those in a differences in spectral sensitivity between
fast film and thus will produce less light two films, need rarely be considered in
absorbing silver when they are exposed industrial radiography. Usually such
and developed. At low kilovoltages, one changes in sensitivity are automatically
absorbed photon will expose one grain, of taken into account in the preparation of
whatever size. Thus, more photons will exposure charts and tables of relative film
have to be absorbed in the slower film speeds. The spectral sensitivity of a film is
very important in radiation monitoring,
because here an evaluation of the number ISO 11699-19 are examples of two film
of roentgens incident on the film is classifications. Table 4 compares films
required. listed according to the ASTM classification
with the corresponding ISO classification
level.7
Film Classification
Radiographic film systems can be
classified on the basis of their image Reciprocity Law Failure
quality performance. The classification of The Bunsen-Roscoe reciprocity law states
films provides a means of specifying that the density of a photochemical
radiographic film and film systems reaction depends only on the product of
without mentioning film brand names. the radiation intensity and the duration
This specifying of film is according to of the exposure and is independent of the
measurable physical characteristics such absolute values of either quantity. Applied
as the minimum film gradient at film to radiography, this means that the
density 2.0, minimum film gradient at developed density in a film depends only
film density 4.0, maximum granularity on the product of X-ray or gamma ray
and the minimum ratio of film gradient intensity reaching the film and the time
to granularity.7 ASTM E 18158 and of exposure.
FIGURE 12. Typical X-ray spectral sensitivity curve of radiographic film, showing radiation
required to produce density of 1.0 for various radiation qualities.
516 (2.00)
258 (1.00)
Radiation for density of 1.0, MC·kg-1 (R)
206 (0.80)
155 (0.60)
Heavy filtration
103 (0.40) Light filtration
52 (0.20)
26 (0.10)
21 (0.08)
15 (0.06)
10 (0.04)
8 (0.03)
10 20 30 40 60 80 100 200 400 600 800 1000 2000
(I × t)L
(I × t)H
Log (I × t)
(I × t)0
IL I0 IH
Log I
Legend
I = intensity of light or electromagnetic radiation
0 = subscript denoting particular value
H = subscript denoting higher value
L = subscript denoting lower value
t = exposure duration
Dmax and Dmin for optimum contrast. chemicals should be changed periodically
Deviation in either direction because of to eliminate particulates and
temperature change will generally result contaminants from accumulating.
in lower contrast (see Fig. 14). Changing out chemistry once a month or
Agitation. Agitation increases both the every two to six weeks should be by
rate of development and the rate of choice and convenience and never
reduction. To clarify, development rate is because the activity has been lost.
increased because agitation permits a Starter Solution. This is an acid solution
constant mixing of the solution and aids (pH 2 approximately) containing
in washing bromine and the oxidized bromides that is added to fresh developer
developers out of the emulsion. Agitation each time the automatic processor is
aids reduction by constantly swirling the filled. Between 20 and 25 mL·L–1 (2.5 and
reducing agents in and around the silver 3.2 oz per 1 gal) of developer are added to
halide crystal lattice. When replenishment the processor, depending on the
systems are used, agitation helps keep the manufacturer. Each manufacturer’s brand
stronger replenishment solution properly of starter should be used with the
mixed into the working solution. corresponding brand of developer. Starter
Agitation also helps in the filtration of is not normally added to the
reaction byproducts, mostly gelatin, by replenishment chemistry.
circulating the chemicals through a filter. Starter gets its name from the fact it is
Finally, agitation keeps the temperature used when a fresh batch of developer is
uniform. first used. Its acid nature primarily
Replenishment. Chemically defined, deactivates the developer to help control
replenishment is only a replacement of fog. Its bromides are added to simulate
quantity, of volume, a maintenance of a used developer and thereby provide
preset amount. Regeneration is the second consistent, reproducible quality from
function of an adequate replenishment batch to batch. For Class I films, which
system and its job is to ensure consistent benefit from higher bromide levels, the
activity by a replacement of spent starter both lowers pH and increases the
chemicals. It is the purpose of developer development rate.
regeneration to ensure that the The developer chemistry manufacturer
characteristics of the finished radiograph provides guidance on the amount of time
— its speed, contrast level, fog level and and temperature to use with its product.
maximum density — remain substantially The manufacturer’s recommendations are
constant. based on the assumption that all
A good replenishment and instructions have been carefully followed,
regeneration system will prolong the life including the addition of the correct
of chemistries, aid in the maintenance of amount of the correct brand of starter.
consistent quality and may lead to Faults from Developer. Types of faults due
improved sensitometric quality. The to developer include too much or too
proper replenishment or regeneration little density (toe, straight line, shoulder
system means that chemicals need to be areas), too much contrast or too little
changed less often. Although contrast. See Table 6 for faults related to
replenishment keeps the chemical the developer.
conservation constant the system Automatic versus Manual Processing and
Chemistry. Automatic developers contain
gluteraldehyde as a hardening agent to
control emulsion (gelatin) swelling.
FIGURE 14. Gamma versus temperature Because manual developers have no
response curve.
Dmax
Dmax
TABLE 6. Faults from developer.
Optimal Processing Action Underdevelopment Overdevelopment
Contrast
Speed Temperature low high
Contrast Transport rate fast slow
Speed Solution level low not applicable
Base
and fog Agitation low not applicable
Base
and fog Chemical reaction; oxidation high not applicable
Contamination high high
Class I film replenishment over under
28 30 32 34 Class II film replenishment under over
(82) (86) (90) (93 ) Class III film replenishment under over
Developer temperature, °C (°F) Class IV film replenishment under over
hardeners the gelatin carries out more of ammonium salts, is the usual fixing agent.
the developer volume. In automatic Sodium thiosulfate is best known as hypo.
processing, in addition to developer However, all of the terms hypo, fixer,
hardener, the processor uses squeegee clearing agent, fixing agent and thiosulfate
rollers to remove excess developer and an are generally synonymous. The basic
automatic replenishment system to reaction between thiosulfate and silver
sustain both volume and activity levels of halide is that of dissolving and carrying
all chemicals. Of course the processor has away the undeveloped silver. Thiosulfate
no short stop and this reduces the overall can, however, attack the developed silver
size by about 20 percent. Automatic if the pH is decreased (moved toward a
developers can generally operate at higher neutral or basic pH). Thus, replenishment
temperatures than manual developers. is important to the fixer in regeneration
Manual processing uses a short stop, of chemical strengths. The developer
first rinse or acid bath between the carryover into the fixer replaces what fixer
developer and fixer to stop development is carried out but also reduces the pH
or prevent excess developer from carrying slightly. If left within the emulsion,
into the fixer and diluting or thiosulfate reacts with silver particles to
contaminating it (to prolong the fixer form silver sulfide (Ag2S), which has a
life). Fixers are generally the same for characteristic objectionable yellow brown
both automatic and manual processing. stain. This is referred to as residual hypo or
Manual Acid Stop Bath. The acid stop hypo retention.
bath, normally 2 to 3 percent acetic acid Hardener. The hardener shrinks and
solution, functions in several ways: it hardens the emulsion. Aluminum
neutralizes alkaline developer by rapidly chloride is frequently used but any
lowering the pH to the point where aluminum compound, such as potassium
development stops; it helps prevent aerial alum or chrome alum, will work. The
oxidation of the developer agent, which hardener has several functions: (1) to
otherwise could form staining products; it increase resistance to abrasion; (2) to
dissolves or retards the formation of minimize water absorption by the gelatin
calcium scum and preserves the acidity of (this reduces drying time); and (3) to
the fixer and helps control gelatin reduce swelling to permit roller transport.
swelling. Some commonly used agents are Activator. Acetic acid provides acid media
acetic acid, citric acid, diglycolic acid and of about pH 4.0 and aids in the hardening
sodium bisulfite. of the emulsion. However, the most
The rate of neutralization for the acid important function is the neutralizing of
stop bath of the fixer depends on developer carryover and of the developer
(1) nature and thickness of emulsion; trapped within the emulsion. The
(2) pH value of stop bath, fixer or both; reducers of the developer require high
(3) total acidity of the stop bath and fixer; basic or alkaline media in which to react
(4) agitation; (5) developer alkalinity; and they will continue to react, even after
(6) developer pH; (7) type of developer the film is removed from the developer
agents used; (8) age, a function of solution, until they are neutralized.
replenishment; and (9) temperature. Because a very small part of the fixer
(acid) will neutralize or at least lower the
Fixer pH or a larger volume of developer,
greater care is required when mixing
Standard fixers are composed of chemicals chemistries so that contamination of the
listed in Table 7. developer with fixer does not occur.
Fixing Agent. The function of the fixing Acetic acid is usually used because it is
agent is to form soluble stable complexes a weak acid. It achieves good buffering
of silver salts that can be removed readily and a slightly acid medium permits
from the emulsion. Fixing agents should aluminum hardeners.
have no effect on the emulsion binder or Preservative. Sodium sulphite is also the
on the already developed silver. preservative for the fixer but its general
Thiosulfate, in the form of sodium or function is to prolong the life of
Ammonium thiosulfate clearing agent clears away unexposed, undeveloped silver bromide
Aluminum chloride hardener shrinks and hardens emulsion
Acetic acid activator acid media that neutralizes developer
Sodium sulfite preservative maintains chemical balance
Water solvent dissolves chemicals
PART 3. Darkroom3
artifacts from static electricity, handling or diffuser and wash it. Then paint the
moisture. In addition, adequate inside with a good, durable white enamel
ventilation is needed to keep harmful finish.
chemical fumes from accumulating and
affecting either the darkroom personnel Avoiding Static
or the radiographic film.
There are two ways to avoid markings on
X-ray films. One is to prevent the
Cleaning Tanks generation of static electricity; the other is
Corrosion seldom occurs when the tanks to cause such charges, once generated, to
are full of normal chemical solutions and leak off gradually rather than to discharge
are kept clean. rapidly, which is what causes damage.
Deposits often form on the walls of the The most successful procedure is to
developer tanks because of the action keep a high relative humidity in the
between mineral salts dissolved in the surrounding atmosphere. An accurate
water and carbonate in the developing instrument for measuring relative
solutions. These deposits can be removed humidity, called a psychrometer, is a
by using commercially prepared stainless valuable addition to any radiographic
steel tank cleaner. Follow the directions of darkroom. Periodic checks on prevailing
the manufacturer, being sure to rinse the darkroom humidity enable one to take
tank walls with fresh water. Wipe the tank special precautions necessary to minimize
out with a clean cloth or cellulose sponge. the generation and discharge of static
Clean the exterior stainless steel before electricity. The relative humidity in the
any deposits can attack the surface. Wipe darkroom should be between 40 and
with a cloth and warm soapy water and 60 percent.
then rinse, making sure no soap deposits The following precautions will be of
get into the chemistry. Once a week, use a assistance in overcoming the most
stainless steel cleaner according to the common causes of static:
directions on the label. 1. If using X-ray film that has
Always give special attention when interleaving paper, handle film gently.
cleaning welds and corners where deposits Let the interleaving paper fall away
can cling. If deposits are heavy, remove from the film and place film in the
the worst of them with fiber brush or cassette gently, without sliding it over
plastic cleaning pad, then polish with a the screen.
stainless steel cleaner. If an abrasive is 2. Following X-ray exposure, the cassette
required, use a very fine sandpaper. or holder should be opened slowly
Use caution never to use metallic and the film removed carefully. The
abrasives, steel wool or wire brushes, as reason for careful handling is that the
they can contaminate the surface of film is more than twice as susceptible
stainless steel. Any foreign metal particles to an energy source once it is exposed.
will cause corrosion and may contaminate Thus, a film will react to much smaller
chemicals. Do not use commercial steel electrical discharges after exposure.
wool pads or strong detergents, because 3. Move slowly when handling the film.
these are hard on the stainless steel and 4. Make sure everything is grounded.
could react unfavorably with the 5. Use X-ray antistatic salt screen cleaner
chemistry. regularly.
6. Avoid static generating synthetic
Cleaning Illuminators clothing.
Quite frequently, good radiographs will Grounding. Electrically ground the metal
appear dull because they are viewed on top of the film loading bench, film bin,
faulty illuminators. Illuminators are faulty X-ray table, pass boxes and other
when the glass plate is dirty or bulbs of equipment such as processors. In the
different wattage, age, color or size are darkroom, avoid nonconductive floor
used. Old interior paint that is dull or covering (rubber tile), hard floor waxes on
dusty will cause the radiograph to appear concrete, rubber soled and plastic soled
dull. Use a regular photographic exposure shoes, intensifying salt screens with worn
meter to test the illuminator. Identical surfaces and using a dry cloth to clean
radiographic studies should be viewed at intensifying screens. A camel hair brush
the same intensity. An amperage meter or vacuum cleaner should be used for
can be installed to control uniform dusting and a lintfree cloth and screen
output. cleaner with antistatic solution should be
Wash the outside of the viewing glass used for washing intensifying salt screen
plate every day. Once a month wash out surfaces. In periods of low ambient
the inside of the view box. Always use humidity (winter time or northern
bulbs with similar intensities. When climates), when static is prone to occur,
conditions indicate paint deterioration, antistatic solution can be applied to
unplug the illuminator, remove the front
intensifying screens as added protection people use is to write on each box the
against static discharges. date when it is received. Whatever system
Common Static Marks. Three kinds of is used, be sure to keep records and always
static markings are illustrated in Fig. 15. use the oldest film first. Film must be
Crown and tree are considered to be results stored and inventoried by expiration date,
of heavier electric discharges. They can be film brand, type (class) and speed.
generated by very rapid motions, such as Film boxes should be stored on their
occur when film is removed from edges. This distributes the weight and
interleaving paper, when interleaving helps protect the film from pressure
paper breaks contact with the film or marks.
when the film is touched by fingers. When using one box at a time or when
Smudge static markings may result from there is no film bin, always be sure to fold
photographic exposure to visible light over or close the bag and to replace the
produced by sparks in air next to the film. lid after each film is removed. Film is
Smudge static is produced when relatively packaged in hermetically sealed light tight
low potential discharges occur over a large and moisture proof pouches — for
area. example, black polyethylene or
aluminized plastic. Once the pouch is
opened to expose the fresh film there is
Color Conditioning still enough bag remaining to fold over to
Surfaces above the working area should be make the bag light tight again. This
finished with a dark matte paint to feature is not to obviate the box top or
minimize reflected light. film bin but is an added safeguard.
Cassettes should not be stored for
Storing X-Ray Film prolonged periods of time loaded with
film. Load cassettes and holders with fresh
Recommended storage conditions for all film before each use.
types of X-ray films are temperatures
between 18 and 24 °C (65 and 75 °F) and
40 to 60 percent relative humidity.
Usually most radiographic facilities will
have two storage areas. One area is for
long term supplies and another, usually
the darkroom, is for short term needs. In
either case, it is required that a stock
rotation plan be instituted. The plan is
quite simple in that as new film arrives, it
is placed on the right side of the supply.
As the film is needed, it is removed from
the left side. This is called a FIFO system:
first in, first out. To assist in rotating film
boxes, remember that all film boxes have
an emulsion number and an expiration
date on the end label. A system many
(a)
(b)
(c)
2.0
Straight line A B
2.0
portion
1.0
1.0 Toe
1 2 3 4 5 1 2 3 4 5
the same contrast. The separation of the then proceeds to form soluble compounds
films is a measure of the speed difference. that are removed by washing.
Film A requires less exposure to achieve a Inadequate fixing leaves small amounts
speed point of density 2.00 above base of these complexes that cannot be
plus fog. Film B requires more exposure, removed by normal washing. In time,
so it is said to respond slower or be less these residues break down and react with
fast. The convenience of using relative the silver image to form silver sulfide
exposure also applies to speed. The speed (Ag2S). This sulfide is usually yellow but
of one film can be expressed on a basis may range from pink to brown. It cannot
relative to another film when one is made be easily removed from the radiograph.
the standard of comparison. This Extremely minute amounts of residual
reference film can be assigned any thiosulfates are sufficient to cause serious
arbitrary speed value, such as 100. If deterioration. Therefore, the radiograph
another film requires only half the may appear transparent and at the same
exposure to reach the same density as the time be inadequately fixed. For this
reference film, the faster film will have a reason, close attention to recommended
relative speed of 200. A density of 2.00 fixing technique is mandatory.
above base plus fog has been designated
as the density to compute film speed. This Washing
density has been chosen because it
represents the minimum useful density Radiographs must be washed thoroughly
range for much of radiography. For lower to prevent discoloration with age and to
density films, speeds are often calculated ensure preservation of the image.
at the lower density of 1.00.
Drying
Fixer Avoid temperatures above 49 °C (120 °F)
Failure to fix a film sufficiently results in in drying and maintain a steady volume
its discoloring with age. A good rule of of air across the film surface. An
thumb for determining the minimum excessively low relative humidity within
fixing time is as follows. After a film has the dryer may cause water spot drying
cleared, leave it in the fixer an additional marks and streaks on the film surface.
two times as long as it took to clear. For
example, if it takes 1 min for a film to
clear, then it should be left in the fixer at
least an additional 2 min or a total period Safety
of 3 min. Developer, with its hydroquinone and
To prepare fixing baths from liquid alkalinity, forms a very hazardous
concentrate fixer, it is essential that solution. Always use good ventilation
directions on the container be followed. when mixing chemicals. A nose filter or
Attention is paid only to the temperature respirator is suggested when powdered
of the developer, because fixer chemicals are mixed. Goggles are required
temperature is less critical; however, at by the United States Occupational Safety
high temperatures, it is important that the and Health Administration to protect the
processing baths be maintained at about eyes.10 In addition an eye wash station is
the same temperature. Changes in required. If developer gets into the eye the
temperature cause the gelatin of the worker must begin washing within 15 s
emulsion to swell and contract. When the and continue washing for 15 min
temperatures of two baths differ minimum. Washing for 1 h is preferred.
excessively, this change in the gelatin The developer is most hazardous but
takes place so abruptly that unevenness is care must be taken when working with
likely to result. The effect produced on fixers also. In addition to the above,
the film is known as reticulation — the rubber gloves will protect the skin and
gelatin breaks, producing fine cracks. keep chemicals out of cuts; rubberized or
Archival Quality. The importance of plastic aprons will protect the worker and
adequate fixing to archival keeping clothing.
quality cannot be overemphasized. Poorly
fixed films will not deteriorate until many Clearing Film Base
months after processing. The lighter
portion of a radiograph may become Liquid laundry bleach will dissolve the
yellow and the image may tend to fade. gelatin and produce a clean, clear sheet of
This delayed action may be traced back to base plastic. Warm solutions work faster.
the nature of the fixing process, which is Enzymes could be used but are very
believed to consist of two steps. hazardous.
Undeveloped silver halide is first
converted to silver thiosulfate complexes
that are only slightly soluble. The reaction
Processing Chemicals
Processing chemicals for automatic
equipment differ from deep tank
chemicals. Never attempt to use
conventional chemicals in automatic
processors. Although the basic chemical
reactions are similar, automatic processing
chemicals are especially formulated for
high speed roller operation. Modern
automatic processing chemicals make
possible maximum ease of use and
uniformity of finished radiographs,
regardless of the make of processor.
Although all chemicals operate
efficiently with any make or speed of film,
they generally provide best results with
the same brand of film. In addition, using
companion products causes fewer
variables and is better understood by a
manufacturer.
All ingredients, plus full instructions,
are included in each package of developer,
fixer and starter solution. These chemicals
are prepared in strict conformity with
basic formulas known to produce
excellent results. When properly mixed
and cared for, they operate efficiently over
long periods of time.
Finally, clean the work area and
equipment and put the equipment away.
Then install the crossovers and prepare
the processor for operation. This
installation of fresh chemistry should be
entered in the processor log book.
Solution Services
In many localities, specialized
organizations handle the mixing and
maintenance of chemicals for automatic
processing equipment. These services
include routine inspection, cleaning and
refilling. Although the services are
provided by professionals, it is still good
policy for each X-ray department to have
its own in-house specialists and to
perform routine, daily inspection.
The scarcity and high price of silver make safe but require constant pH adjustment
its recovery from fixing baths important by a technician.
for ecological, environmental and Electrolytic Recovery. Silver is recovered
economic reasons. Fixer contains about by passing an electric current (direct
40 percent of the original silver in the current) through the silver laden fixer.
film. Laws and standards in the United Electrolytic recovery systems (called cells)
States establish limits for dissolved silver are classified as agitated or nonagitated,
and forbid its disposal in drains.10,11 This high or low current density units. The two
consideration is in addition to silver’s terminals (electrodes) are a positively
scarcity and value. Even the smallest user charged anode (usually graphite) and a
should treat the used fixer to remove negatively charged cathode of stainless
silver — the money recovered will pay for steel. Positively charged silver ions (Ag+),
the effort. are attracted to the cathode, where they
plate out as metallic silver (Ag°), called
Recovery Techniques flake. This is the most efficient technique
for medium or larger installations.
The techniques of recovery are chemical
(which includes precipitation and metallic
replacement) and electrolytic. Metallic Efficiency of Silver Recovery
replacement is simplest but requires a low Equipment
volume continuously. Electrolysis is During processing, the developer converts
recommended for higher volume the latent image bearing silver halide
processors. crystals into a visible image — black
Metallic Replacement. If a fixer solution metallic silver. Those crystals exposed but
containing silver ions is brought into not developed and those not exposed are
contact with a metal, the less noble metal washed out in the fixer. In industrial
(such as steel wool, zinc, copper or steel radiography the silver is dissolved into the
turnings) is replaced by the silver. The fixer in a ratio of more than 6 g·L–1
silverless fixer cannot be reused. (1 troy oz·gal–1) of fixer. Silver recovery or
Metallic replacement units, also called reclamation is the process of converting
buckets or cartridges, contain steel wool the silver to metallic silver. Understanding
or zinc screen and are usually used in the factors that control the efficiency of
tandem. As the acid fixer breaks down the this operation will help in understanding
less noble steel, the more noble silver and upgrading existing systems or in
metal precipitates as metallic silver. A generating specifications for new systems.
sludge of iron oxide and silver forms in Dwell Time. Sufficient time is required for
the bottom of the container. the reaction to occur. Electrolytic units
This technique is both inexpensive and (cells) are rated in troy ounce per hour
efficient. It can remove 60 to 95 percent capacity. Buckets are rated in cubic
of the silver: 1 kg (2.2 lbm) of steel wool centimeters per minute or gallons per
will collect 3 to 4 kg (6 to 9 lbm) of silver. hour of fixer flow. Exceeding these limits
However, efficiency is based on a slow, will result in silver going through the unit
steady, continuous flow of silver laden and down the drain.
fixer. Efficiency is about perfect for the
first 25 percent of the unit’s life Agitation. Buckets provide agitation by
expectancy, 400 to 800 L (100 to 200 gal) flowing the fixer over the many wire
of silver laden fixer and then often filaments. Electrolytic cells use pumps or
becomes only 30 percent effective. Also, impellers. Greater physical agitation
the sludge produced by this technique is increases the unit’s efficiency in
expensive to ship and refine. producing metallic silver and allows
higher plating currents to be used.
Chemical Precipitation. Precipitation is a However, agitation should not be so
chemical reaction that separates the silver violent as to cause splatter, spillover or
from the solution in an insoluble, solid excess evaporation.
form. This type of unit, particularly those
that use sodium sulfide and zinc chloride, Surface Area. The larger the surface area
produces toxic and volatile fumes and so the higher the plating current can be in a
should be avoided. The units using cell. In any unit, increased area increases
sodium borohydride are very efficient and recovery rate.
Edge Effect. The edge effect is related to reduces fixer consumption and silver lost
surface area and electron flow efficiency; into the wash. The major disadvantages
the more edges or surface area, the better are increased cost of the cell and
the efficiency. potentially higher hypo retention levels
Electrolysis. As the acid fixer enters the because of reduced fixer efficiency. Fixer
bucket the steel wool is attacked, can only be recirculated from cells —
producing chemical electrolysis. The steel never from buckets or precipitation.
becomes oxidized to iron oxide and the The following precautions should be
silver in solution becomes metal. Cells observed when installing or operating any
contain an anode and a cathode, a silver recovery units:
rectifier and a transformer to pass direct 1. Make sure than fixer overflow to the
current through the solution. The higher silver recovery unit is a continuous
the current, the greater the efficiency. Too downward flow.
high a current (usually when little silver is 2. Clean the standpipe on metal
present) can break down the fixer, this is exchange units regularly.
called sulfurization and is to be avoided. 3. Ensure against air locks in electrolytic
Maintenance. All units require records, units.
regular inspection and regular 4. Be sure there is an air break
maintenance to ensure proper use. (electrolytic break) between the
Buckets can clog, back up or leak. Cells solution in the electrolytic recovery
can become too loaded with silver, short unit and the incoming fixer. Without
out, blow a fuse or burn up an agitation one, there is danger of plating silver in
pump. The amperage should be the processor fixer tank.
automatically or manually adjusted 5. Use the highest amperage possible for
according to the film (silver) volume optimal recovery — but keep it short
during the day. of sulphurization (characterized by
Centralization. Centralized recovery is the yellow color and smell of rotten eggs).
most efficient system where three or more High amperage produces soft, black
units are involved. A holding tank feeds a silver; low amperage produces hard,
single cell a continuous supply for shiny silver but of approximately
optimal efficiency. equal quality. The higher amperage
helps to ensure removal of most of the
silver.
Silver Estimation
Silver estimating paper, which indicates
the relative amount of silver per gallon of Scrap Film
used fixer, can be readily purchased from Films that are to be discarded also have
most silver reclaimers. The test strips are value for their silver content. About
used just like pH paper strips. Industrial 60 percent of the original silver remains
radiography usually operates at a level of in the film to form the visible image. Both
about 10 g·L–1 (1.2 ± 0.2 troy oz·gal–1). waste film and outdated records are
An even simpler technique of valuable and should be sold for their
determining if there is any silver in the silver content.
solution is to put a brightly polished
copper tube in the solution. Any silver in
the used fixer will quickly adhere to the Security and Selling Recovered
copper tube and give it a gray color. Silver
Purchase tailing or central electrolytic In radiography, business economics
units according to the calculated capacity revolve around the cost of producing a
in gram per second. Collect the fixer in visible image versus the value or price of
the processor at cleaning time. If a silver the product. The single largest budget
recovery unit malfunctions, disconnect it item is manpower, which must be used
or isolate it so that it cannot ruin efficiently.
radiographs being processed. Radiographic film is expensive. It must
be kept in the best possible condition and
Silver Recovery Installations protected against abuse and theft,
whether in its fresh or used form, through
Tailing units may be buckets or cells inventory controls such as records, policy
placed individually or in tandem on each procedures and security. Such programs
processor. Usually buckets are used in will pay for themselves in improved
tandem, with the second becoming the earnings.
first after every 400 to 1300 L (100 to Silver should be recovered from used
300 gal) of fixer, depending on brand and fixer and used films to reduce the cost of
size. Properly sized cells should not the original fresh stock. Because a
require a tailing unit. substantial value is represented by
Recirculation cells take the fixer from recovered silver, it is important to impose
the processor, remove most of the silver inventory and security controls.
and return the fixer to the processor. This
Specific Suggestions
Silver flake is derived from scraping off the
collection plate of an electrolytic silver
recovery unit (cell). There is little
significance whether it is silver colored
and hard (result of low current levels) or
black and soft (result of high current
levels). A properly sized cell will collect 90
to 95 percent of the silver, which will be
95 percent pure. Dry the silver before
weighing. (If weighed wet, deduct five
percent of the weight for trapped
moisture).
Silver sludge from buckets should be
shipped in solution in the bucket.
Draining the fixer exposes the sludge to
air and an exothermic reaction produces
heat. If possible, the sludge should be
dried in a large open pan and then
shipped. The sludge damages the refiner’s
crucibles and the refiner charges more to
handle sludge than flake.
References
1. Rivkin, W.B. and G. Wicks. Sec. 4, 11. 2001 TLVs® and BEIs®. Cincinnati,
“Radiation Detection and Recording.” OH: American Conference of
Nondestructive Testing Handbook, Governmental Industrial Hygienists
second edition: Vol. 3, Radiography and (2001).
Radiation Testing. Columbus, OH:
American Society for Nondestructive
Testing (1985): p 174-185.
2. Quinn, R.A. and C.C. Sigl, eds. Bibliography
Radiography in Modern Industry,
fourth edition. Rochester, NY: Eastman
Kodak Company (1980). Books
3. McKinney, W.E.J. Sec. 7, “Radiographic Bruce, H.F. Your Guide to Photography,
Latent Image Processing.” second edition. New York, NY: Barnes
Nondestructive Testing Handbook, and Noble (1974).
second edition: Vol. 3, Radiography and Bunting, R.K. The Chemistry of
Radiation Testing. Columbus, OH: Photography. Normal, IL: Photoglass
American Society for Nondestructive Press (1987).
Testing (1985): p 299-376. Control Techniques in Film Processing. New
4. McKinney, W.E.J. Radiographic York, NY: Society of Motion Picture
Processing and Quality Control. and Television Engineers (1960).
Philadelphia, PA: J.B. Lippincott Eaton, G.T. Photographic Chemistry in
Company (1988). Black-and-White and Color Photography,
5. Gurney, R.W. and N.F. Mott. fourth edition, revised. Dobbs Ferry,
Proceedings of the Royal Society of NY: Morgan and Morgan (1986).
London: Series A, Mathematical and Feininger, A. Darkroom Techniques. Garden
Physical Sciences. Vol. 164. London, City, NY: Amphoto (1974).
United Kingdom: Royal Society (1938): Folts, J.A., R.P. Lovell and F.C. Zwahlen, Jr.
p 151. Handbook of Photography, fifth edition.
6. Gurney, R.W. and N.F. Mott. Electronic Albany, NY: Delmar Thomson
Processes in Ionic Crystals. Oxford, Learning (2001).
United Kingdom: Clarendon Press Gray, J.E. Photographic Processing, Quality
(1940). Control and the Evaluation of
7. Marstbloom, K., R. Kochakian, Photographic Materials, Vol. 2. HEW
B. Vaessen and P. Willems. “Analog Publication 77-8018. Rockville MD:
Film Radiography Technology Bureau of Radiologic Health (1977).
Advancements.” ASNT Fall Conference Gray, R.H., ed. Applied Processing: Practice
and Quality Testing Show 2001 Paper and Techniques. Washington, DC:
Summaries Book. Conference [Columbus, Society of Photographic Science and
OH, October 2001]. Columbus, OH: Engineering (1968).
American Society for Nondestructive Gregg, D.C. Principles of Chemistry, third
Testing (2001): p 165-167. edition. Boston, MA: Allyn and Bacon
8. ASTM E 1815-96, Standard Test Method (1968).
for Classification of Film Systems for Haist, G. Modern Photographic Processing,
Industrial Radiography. West Vols. 1 and 2. New York, NY: Wiley
Conshohocken, PA: ASTM Interscience (1979).
International (2001). Herz, R.H. The Photographic Action of
9. ISO 11699-1, Non-Destructive Testing — Ionizing Radiations in Dosimetry and
Industrial Radiographic Films. Geneva, Medical, Industrial, Neutron, Auto- and
Switzerland: International Microradiography. New York, NY: Wiley
Organization for Standardization Interscience (1969).
(1998). James, T.H., ed. The Theory of the
10. 29 CFR 1210, Occupational Safety and Photographic Process, fourth edition.
Health Standards. Code of Federal New York, NY: Macmillan (1977).
Regulations: Title 29, Labor. James, T.H. and G.C. Higgins.
Washington, DC: United States Fundamentals of Photographic Theory,
Department of Labor, Occupational second edition. New York, NY: Morgan
Safety and Health Administration; and Morgan (1960).
Government Printing Office.