Professional Documents
Culture Documents
C
11 H A P T E R
Digital Radiographic
Imaging
The detection devices that support the amorphous silicon, amorphous selenium
larger imaging systems already mentioned and charge coupled device technologies
are the following: (1) phosphors deposited are described below.
on amorphous silicon thin film transistor Each of these devices can be used to
diodes; (2) photoconductors such as replace film radiographic techniques
amorphous selenium deposited on thin depending on the size of the application
film transistors; (3) phosphors deposited and on the spatial resolution, image
or coupled through fiber optic lenses onto contrast and speed required. As noted in
charge coupled device based detectors and Table 1, the detectors have variable modes
complementary metal oxide silicon based of operation or are available in different
detectors; (4) photostimulable storage architectures to address diverse
phosphors; (5) phosphors deposited on applications. There are numerous pixel
linear array systems; and (6) X-ray architectures of amorphous silicon
scanning source reversed geometry detectors but it is important to note that
detectors. Further details are provided in currently not all detector choices allow
Table 1. real time operation of 30 frames per
Each of these devices has an X-ray second.
capture material as its primary means for
detecting X-rays. This material is either an
X-ray phosphor material combined with a
photoelectric device (diode, Charge Coupled Devices
photomultiplier tube or charge coupled Scientific charge coupled devices,
device) or is an X-ray photoconductor although they are typically small in size,
material that is then followed by an have been made with high pixel densities.
electronic readout device. The most The fields of photography, astronomy and
common of these detection systems in microscopy have demanded this and the
operation today are the flat panel nondestructive testing industry has been a
detection systems based on amorphous beneficiary of these developments. Table 1
silicon and amorphous selenium illustrates these small pixel dimensions
structures, the camera systems based on (9 to 50 µm pixels).
charge coupled device technology and the Charge coupled devices have not been
storage phosphor systems. The fabricated into larger arrays because the
Amorphous 200 × 200, 1024 x 1024, 100, 127 or 200 < 1 (real time) thallium activated cesium 55/53 or 64
silicon 230 × 190, 2304 × 1920, iodide or terbium
410 × 410 or 2048 × 2048 or activated gadolinium
280 × 410 2304 × 3200 oxysulfide
Amorphous 350 × 430 to 2560 × 3072 139 < 1.0 to 1.5 selenium 34
selenium
Charge coupled small to 100 × 100; to 4096 × 4096 9 to 50; effectively < 1 (real time) thallium activated cesium 55/53 or 64
devices larger with lenses larger with other iodide or terbium activated
optics gadolinium oxysulfide
Storage small to 1550 × 430 to 15 500 × 4300 25 to 250 < 1 to 4 europium activated 56/35
phosphors barium fluorobromide
Linear arrays small to 500 to 4096 10 to 200 < 1 (real time) thallium activated cesium 55/53 or 64
iodide or terbium activated
gadolinium oxysulfide
Reversed small to 450 × 450 to 2048 × 2048 25 to 200 < 1 (real time) thallium activated sodium 11/126
geometry iodide
quality image, the dynamic range and the imaging detector and the source is about
corresponding benefits of that digital the size of a point source. The data
image file, such as easy archival and acquisition computer also controls the
retrieval. rastering of the electron beam. By
acquiring the output of the detector as a
function of electron beam position, the
computer can generate a real time
Linear Arrays radiograph of the specimen under
Linear array detectors are much like examination.
charge coupled devices, except that they Because a single small area detector is
typically only have pixels in 1 dimension used and the object is placed at the
or they may be composed of a small source, not at the detector, the X-ray
rectangular array such as a 32 × 1024 pixel scatter from the object is essentially nil.
array. The advantage of linear arrays is The disadvantage of this approach is that,
their intrinsic scatter rejection capability. because it is reversed geometry, the
X-ray scatter exiting a specimen under effective focal spot size is that of the
examination can be a large contribution detector size. The detector size is typically
to the degradation of the contrast in the much larger than a typical industrial
image. The linear array system acquires its X-ray focal spot. So that any specimen
image by being scanned one line (or a that has some thickness will show
group of lines) at a time across an significant unsharpness as the feature of
object.6,7 The key is that the radiation interest moves away from the X-ray
beam is masked or collimated to match source.
the size of the detector. This dramatically
decreases the object’s scatter field. The
scatter detected at each of those lines is
substantially less than that of individual Detection Efficiency
lines in an area array. Linear arrays have With the exception of the
been successfully used in computed photoconductive selenium based detector,
tomography applications and have also all detectors listed use a phosphor layer of
been found to be effective for digital one sort or another to capture and
radiographs. convert the X-ray intensity. The selection
of the phosphor or photoconductive
material, its thickness and effective
atomic number will impact the total
Scanning Beam, Reversed number of X-rays absorbed in the
Geometry conversion material. Once energy is
absorbed each material, phosphor or
The reversed geometry system8-10 goes one photoconductor, has its own efficiencies
step further in reducing X-ray scatter in for conversion of this energy into either
the examination. In this case the data are light or charge carriers. There are other
acquired with a small thallium activated coupling steps following this to transfer
sodium iodide (NaI:Tl) scintillator coupled the signal onto the pixelized readout
to a photomultiplier tube. A large scanned circuitry. The performance of the X-ray
X-ray source with a target diameter of detector to convey the information in the
about 250 mm (10 in.) is used to define radiation beam is then dependent on the
the image. The X-ray source operates in a efficiency of each step in the X-ray
manner similar to a video monitor. An conversion process leading to an
electron beam is electronically rastered electronic signal. The signal-to-noise ratio
over the inner surface of the front of the of the detector and thus the image
X-ray source. Where the electrons collide contrast are therefore dependent on the
with the inner surface of the tube, X-rays transfer of information along the imaging
are generated. By electronically scanning chain. Digital imaging chain statistics and
the electron beam, the instantaneous the relation with image contrast are
position of the X-ray source is scanned discussed immediately below.
over an area of the front surface of the
tube. The size and location of the scanned
region is user definable, variable from
0.25 to 16 s. The acceleration voltage is
also user definable from 55 to 160 kV
with an electron beam current up to
about 0.5 mA. The diameter of the
electron beam spot at the inner surface of
the tube is about 25 µm (0.001 in.).
The specimen under examination is
placed on top of the X-ray source. This is
the opposite of conventional radiography
where the object is placed near the
The transmitted X-ray beam signal phosphor layer typically creates a large
propagates through various energy gain factor at this point. Following this,
conversion stages of an imaging system, any subsequent inefficiencies in emitting
as discussed elsewhere.11,12 In Fig. 1, N0 the light and capturing it by the
quanta are incident on a specified area of photodiode will result in losses and
the detector surface (stage 0). A fraction of additional sources of noise. If the number
these, given by the absorption efficiency of quanta falls below the primary
(quantum efficiency) of the phosphor quantum sink, then a secondary quantum
material, interact (stage 1). Here it is sink will be formed and becomes an
important that the absorption efficiency additional important noise source.
be high, or a larger X-ray dose would be For most detection systems discussed
needed to arrive at a desired signal level. here, where the phosphor is in direct
The mean number N1 of quanta contact with the diode as in the flat panel
interacting represents the primary detectors, the limiting source of noise is
quantum sink of the detector. The the quantum efficiency of the X-ray
fluctuation about N1 is σN1 = √(N1). This conversion material. Noise characteristics
defines the signal-to-noise ratio of the of digital detectors are discussed
imaging system, which increases as the elsewhere.12
square root of the number of quanta In efficient systems, because the noise
interacting with the detector. Regardless is related to the square root of the number
of the value of the X-ray quantum of X-ray quanta absorbed, it is crucial to
efficiency, the maximum signal-to-noise have a sufficient signal level to avoid
ratio of the system will occur at this quantum mottling. Quantum mottling
point. If the signal-to-noise ratio of the makes detection of smaller features more
imaging system is essentially determined difficult. In medical imaging, regulations
there, the system is said to be X-ray allow a certain maximum dose to the
quantum limited in performance. The patient and optimal signal levels may not
be obtainable. In this scenario, it is critical
to absorb as many X-ray photons as
possible and not to allow secondary
FIGURE 1. Quantum statistics of X-ray imager.
quantum sinks. In nondestructive testing,
it may be possible to increase signal levels
105 Scintillating gain to light photons by selecting any or all of the following:
Exitance efficiency a longer exposure time, a higher beam
flux, a higher radiation beam energy
104 g2 (assuming absorption is still high at those
Quantity of quanta or electrons
1. Cesium iodide can be formed into phosphor material on these same detector
needles on top of the diode structure structures. The fiber optic scintillator
to direct the light to the photodiodes structure and the hybrid analogs (Fig. 3)
without significant light scatter. The provide a similar light guiding capability,
needles are separated by small air gaps provide high X-ray absorptivity without
and the high refractive index sensitivity to moisture and permit control
mismatch between the cesium iodide of afterglow17-19 but are not as bright an
needles and air ensures a high total X-ray converter as is the cesium iodide.
internal reflection of the light as it is Because the gain in the fiber optic
directed to the photodiode. scintillators materials is not as high as in
2. The needle structure enables thick cesium iodide, a potential secondary
phosphor layers, which improves quantum sink is possible especially in a
X-ray absorption. lens based charge coupled device system.
3. The cesium iodide has a high effective This lower gain will clearly lead to a
atomic number (Z) which also higher noise level for a given exposure or
contributes to good X-ray absorption to a much longer exposure time to
efficiency. generate similar statistics. If the detection
One disadvantage of the thallium system used leads to a strictly quantum
activated cesium iodide material is that it limited configuration for any of the
is sensitive to moisture and this phosphor systems mentioned, then they
hygroscopic nature can degrade the all should provide similar image quality,
spatial resolution of the phosphor and assuming the resulting signal levels are
therefore the device over time if allowed high enough above the noise floor of the
to be in contact with ambient humid detection system to render the detector
conditions. Manufacturers of cesium noise negligible.
iodide based systems provide sealed On the other hand, if a phosphor is
enclosures. A sealed enclosure may too bright under a range of X-ray
typically add 20 percent or more to the conditions, then it is possible for the
cost relative to other phosphors and may diodes to be filled too quickly and a
reduce the robustness of the system. mottled image can result simply because
Another disadvantage is that the of the low number of X-ray photons
phosphor is prone to afterglow and actually transmitting the part. One way to
potential variations in light output as a solve this problem is to use a lower gain
function of increasing X-ray dose. phosphor or the fiber optic scintillators
Nevertheless, the thallium activated and use an extended exposure in a single
cesium iodide phosphor has shown frame. Another way is to average multiple
increasing use on the amorphous silicon frames with the faster phosphor until the
detector,13 the charge coupled device desired image quality is produced.
detector and the linear array detector. To summarize, the selection of the
Other inorganic phosphor materials, such phosphor is as important as the selection
as terbium activated gadolinium of the readout electronics and image
oxysulfide (Gd2O2S:Tb),14 high density acquisition software. All three have to be
glass fiber optic scintillators (FOSs)15 and considered together in the design and
hybrid combinations of gadolinium purchase of a system, as well as in the
oxysulfide and fiber optic scintillators16 operation of the system.
have also been used successfully as X-ray The selenium photoconductive
material has been the photoconductive
material of choice as a direct means of
converting X-rays directly into charged
FIGURE 2. Photograph of cesium iodide carriers and avoiding the production of
fibers grown onto a photodiode array. light.3 This is described in more detail
elsewhere. The obvious advantage here is
that the image forming carriers can be
more effectively and efficiently directed to
the electrode structure than is possible
with light. The image sharpness and speed
that results can be very high. The
modulation transfer function can in
principle be higher than that of phosphor
based systems for a given pixel pitch. The
disadvantage of the selenium material,
however, is that it does not have as high
an atomic number as cesium iodide or
gadolinium oxysulfide. For nondestructive
testing applications where the X-ray
energy is typically above 50 kV, to obtain
similar X-ray sensitivity to the higher Z
phosphor materials, the selenium layer
needs to be substantially thicker than the signal-to-noise ratio. This problem has
phosphor material. This means that the been corrected through reduction of gain
X-ray energy will be deposited in a thicker amplification during electronic readout.
region than in these phosphor Clearly, the debate of photoconductor
configurations. The downfall here is that versus phosphor will remain an
X-ray scatter and the diverging X-ray interesting topic for discussion and both
beam can reduce the resulting spatial approaches will continue to be used
resolution in the image, reducing some of successfully. If system noise or other
the benefit of the electrostatic transfer. artifacts in the system begin to compete
A problem for the selenium approach with statistical noise, then in addition to
in the 1990s was that the speed of the proper selection of the converter, the
selenium process was too high in some detection system electronics needs to be
circumstances and the pixels were filled carefully evaluated. The systems discussed
with charge at a relatively low X-ray here have a wide range of system noise
exposure. This overcharging resulted in values, some from different sources and
quantum mottling and a reduced some not necessarily correctable. As a
(a)
Scintillating fiber optics
Phosphor layer
Coupling fluid
(b)
Radiation (X-rays) Phosphor
particles
Phosphor
layer
Scintillating
fiber optics
Cladding
material
Fibers
Light
final check, it is best to test the system for selected goes down, the percentage of
the application at hand. X-ray information in the radiation beam
will diminish and poor image contrast
will result.
To compensate for this image
Spatial Resolution degradation, the conversion material can
Considerations be made thicker or, if time permits, a
longer total exposure time can be selected
The efficiency of the energy conversion to capture more X-ray quanta. Making the
process relates to the speed of the test, the conversion layer thicker can impact the
throughput and the tradeoff with contrast spatial resolving power of the device
sensitivity (the ability to detect a small because both X-ray cross talk and signal
change in thickness or density). (light or electron hole carriers) cross talk
will yield a breakdown in the modulation
Detector Resolution of the signal somewhere in the spatial
The spatial resolution of the detector frequency range of the detector.
determines if features in the object are
detectable from a pixel sampling Modulation Transfer Function
consideration. The selection of the spatial
resolution of the detector is also A good measure of the spatial resolution
important in designing or selecting a therefore is the modulation transfer
detection system. From the aspect of function (MTF). The modulation transfer
image contrast and spatial resolution, it is function measures the signal modulation
desirable to have the largest pixel that will as a function of spatial frequency and is
allow detection of the features of interest typically computed using a fourier
in the radiographic examination. For transform of a line spread function
example, it is not necessary to select a acquired on an angled tungsten edge
39 µm pixel pitch if the application is for placed directly on the detector.13 Figure 4
the detection of large foreign objects left shows the power of a modulation transfer
behind in an engine nacelle. Similarly, function for revealing a breakdown in
fatigue crack detection is probably not spatial resolution throughout the spatial
going to be too successful with a pixel frequency regime of the detector. If the
pitch of 200 µm or larger. spatial resolution drops near the 0 line
pairs per millimeter spatial frequency
regime, this drop can be interpreted as a
Pixel Pitch severe degradation in image contrast and
The predominant factor that governs the will result in poor density discrimination.
spatial resolution of a detector is the pixel If the modulation transfer function is low
pitch. The pixel size of a number of digital at high spatial frequencies, near the
detectors is provided in Table 1. The sampling limit of the detector, this
selection of the X-ray conversion screen
then becomes important. Here the
architecture of the X-ray conversion FIGURE 4. Examples of modulation transfer function curves,
material will dictate to what degree the showing localized variations in modulation transfer function
full spatial resolution of the detector can curve in some regions of spatial frequency domain.
be realized. Maximum spatial resolution for detector is 10 line pairs per
As the pixel pitch is reduced to increase millimeter.
resolution, the total number of pixels in
the image increases for a constant field of
view. The file sizes for typical images run
1
from 2 to 8 megabytes. However, for Scintillating fiber
Modulation transfer function
indicates that the conversion material is damage is a general term that can refer to
not a good choice for detection of the fine any range of damage to a component in
features that the system was designed to the detection chain. The damage can lead
detect. Another choice should be selected. to subtle changes in performance, all the
Balancing the spatial resolving powers of way to failure. Most digital detectors are
a conversion material with its quantum designed so that the electronic
efficiency has been an active area of components behind the X-ray conversion
research and development in digital material are either shielded from the
radiography since 1985. X-rays (for example, by the conversion
material itself or by fiber optic transfer
Gain and Offset Correction components behind it) or are sufficiently
thin to absorb only a small portion of the
Imagery from digital detectors are X-rays that impinge on the component.
frequently normalized for pixel-to-pixel The damage that occurs in the electronic
gain variations and also adjusted to circuitry can result in an increase in the
subtract out the background or offset. The electronic noise of the device and
offset or background signal is usually a eventually to failure as the accumulated
small percentage of the maximum signal dose to the component increases. Each
and is common to all digital detectors. It manufacturer uses proprietary circuitry
is important to subtract this background and various forms of shielding elements
signal to provide a wider linear range and to prevent these effects. Each system is
to subtract any latent images on the different, so the reader is referred to a
detector. In performing a gain correction, general text on radiation effects on silicon
not only are pixel-to-pixel variations circuitry.20
reduced but also variations in the optical The X-ray conversion material, being
components feeding these pixels will be the primary X-ray absorption component,
diminished. Performing this gain is exposed to the highest levels of
correction can also be used to flatten the radiation within the imaging chain.
radiation intensity distribution across the Phosphors such as cesium iodide and
detector panel. Making the radiation photoconductive materials such as
beam intensity more uniform across the selenium have discontinuity centers
detector can result in wider latitude within their band structures that will trap
(viewable thickness range) in the image. electron and hole carriers produced by the
This normalization is really not possible ionizing radiation. In many
with film radiography. circumstances, thermally released carriers
The gain correction is accomplished by from these traps will yield a delayed
taking an image with a radiation luminescence or a delayed release of
technique similar to that planned for charge. This form of radiation damage
production but without an object in the known as afterglow or lag usually increases
beam (an air image) and with a much as a function of radiation dose until an
reduced X-ray intensity. By simply equilibrium occurs where the number of
performing an image division by the gain carriers being trapped equals the number
factor on a pixel-by-pixel basis, the offset being thermally released.
corrected air image is then used to correct Another form of radiation damage to
each subsequent image of an object. X-ray conversion materials that occurs is
Following gain and offset correction, when the carriers are permanently
detection sensitivity improves in relation trapped in deep centers within the band
to an image that does not have this gap. This trapping is sometimes associated
correction. For the air image, it is critical with a darkening of the conversion
that the image be free of transient latent material and usually results in a rapid
images, have the correct intensity and decrease in signal that can only be healed
also not contain an object of any sort by heat annealing of the material or by
(such as a fixture) in the beam. If any of slow thermal release at room temperature.
these occur, then every subsequent This form of damage is known as a gain
corrected object image will contain decrease. In other materials, it is possible
artifacts and the correction will do more to observe a rapid signal gain increase as a
damage than good. function of increased radiation dose.
Although the mechanism of gain decrease
is not widely understood, both gain
changes can impart spatial artifacts into a
Radiation Damage current image created by the variation in
In digital imaging devices, there are radiation intensity across a prior specimen
numerous elements of the detector image. In most cases these gain changes
assembly that can be damaged by the are not long term or permanent. If the
ionizing radiation. Every component in system is prone to these radiation induced
the imaging chain not shielded gain changes, it is important to
appropriately from X-rays or gamma rays continually update gain and offset data,
can be damaged. The term radiation even if the actual examination is not
changing, so that these artifacts can be possible because larger pixels can produce
reduced. If the problem becomes severe it a higher signal-to-noise ratio for a given
might warrant a new phosphor. X-ray exposure. Larger pixels will also
The storage phosphor used in allow a lower exposure for a constant
computed radiography systems, europium signal-to-noise ratio. Larger pixels permit
activated barium fluorobromide thicker X-ray conversion materials, again
(BaFBr:Eu), inherently has discontinuity potentially adding speed to the test.
centers when prepared under certain Finally larger pixels will result in a larger
reducing conditions in the presence of a overall field of view (larger throughput).
partial pressure of hydrogen (H2) gas in an For example, a four million pixel array of
otherwise inert atmosphere.4,21,22 In this 200 µm pixels will have 16× the field of
phosphor, this radiation damage has been view of a four million pixel array of
used in a novel way by storing these 50 µm pixels. As mentioned earlier, the
charge carriers in the phosphor material size of the detector and the size of the
and then later reexciting those carriers pixel still go hand in hand using today’s
(with red light emitting diodes or a technology. It is be possible to have a
helium neon laser) to produce a delayed 10 000 × 10 000 pixel array of 25 µm
luminescence. In storage phosphors, this pixels resulting in a 250 × 250 mm device.
radiation damage is beneficial but, in That said, if a smaller pixel device is
promptly emitting materials such as selected, it might be possible to average
cesium iodide, is to be avoided if possible. pixels into larger superpixels to enhance
speed and part throughput. The minor
drawbacks of such superpixels is that the
X-ray conversion material may not be of
Selection of Systems to optimal thickness for the larger size pixel
Match Application and the percent of active pixel (because
the amorphous silicon approach may be
Some of the key characteristics that might summing four field effect transistors) may
be considered in the selection of a digital not be as great as if the pixel were
radiographic imaging system are the designed with a single set of readout
following: (1) detection precision and circuitry. Finally, the noise of averaging
accuracy; (2) system speed to match that four pixels is a little higher than the noise
of manufacturing and test processes; of a similar detector element of the same
(3) area of the detector to match size.
manufacturing throughput needs; For tight locations, small detectors
(4) volume of the device for access to based on charge coupled device or
tight locations in an assembly; complementary metal oxide silicon
(5) presence of artifacts that can impact technology can be used. Some of these
detection capability. devices are being used for dental
If a large area detector is needed and radiography and they are beginning to
there is a requirement to work at real time find application in nondestructive testing.
frame rates of 30 frames per second, then Where the requirement is to simply
an amorphous silicon detector or charge replace film in favor of a lower cost digital
coupled device based detector should be solution, then storage phosphors can be
selected.23 Note that technology in 2002 used quite successfully. However, if access
may limit digital radioscopy frames to is not an issue, then the other digital
about one million pixels. If static imaging approaches may be more cost effective
is required but the highest spatial over the long term because they are more
resolution is needed and the object size is amenable to high speed mechanized
not large, then a system using a low noise automation of the detector and X-ray tube
phosphor or charge coupled device should to scan about a part or conversely for the
be selected. For this same application, a part to be scanned through the stationary
large area flat panel detector operating in tube and detector configuration.
static mode can also be selected if used in Linear arrays can be used in an
combination with a microfocus X-ray tube assembly line configuration, as can the
but only if the application can withstand real time flat panel and charge coupled
the longer exposure times associated with device detector based systems. Line
magnification radiography.24 scanners offer the advantage of reduced
If super high resolution is required, for sensitivity to X-ray scatter in relation to
example, very tight small crack detection, area array systems.22
then magnification may be required with The scanning beam, reversed geometry
the high resolution charge coupled device system has shown promise in reduced
devices. access applications. This detector is
As mentioned above, it is important to natural because the detector module is
have the largest pixel that can be accepted quite small. The reversed geometry system
from a feature detection (spatial is probably the best system for reduced
resolution) standpoint. This parameter sensitivity to X-ray scatter because the
then provides the highest throughput detector is essentially a point based
Scintillator
Photodiode
Transistor array
Glass
substrate One pixel
(0.004 × 0.004 in.) to 400 × 400 µm each pixel and the connections used to
(0.016 × 0.016 in.). A typical pixel is get the image data out.
shown in Fig. 7. The high resolution view of an
Figure 8 shows a cross sectional view of aluminum tube weld in Fig. 10 was
an amorphous silicon receptor that uses a obtained with an amorphous silicon panel
cesium iodide scintillator. This view originally acquired at 4× geometric
suggests the path taken by the X-ray beam magnification with a microfocus X-ray
as it exits the object being tested and source. In the lower portion of the image
enters the input of the receptor. Figure 9 is the placement of a 75 µm (0.003 in.)
shows an image receptor assembly with thick ASTM aluminum plaque image
the electronics folded out from behind quality indicator. The central hole in this
the panel during assembly and testing. plaque is 250 µm (0.010 in.) in diameter
The large dark area at the right center of and is clearly visible through the thin
the picture is the amorphous silicon array. aluminum walls, about 1.25 mm
Around the edges of the array are all the (0.050 in.) thick. This image illustrates
electronics required to control and read that very small diameter porosity can be
out the image data. detected in these structures: comparing
The micrograph in Fig. 7 shows the the gray scale and size of the pores in the
mechanical makeup of the amorphous weld with the holes of the image quality
silicon layer. The bias and data lines indicator reveals porosity much smaller in
provide the ability to properly control diameter than the 1T (250 µm [0.010 in.])
hole. Figure 10b provides a high pass filter
rendition of this image. Once filtered,
contrast may be added to the image so
FIGURE 8. Schematic cross section of photodiode X-ray that high contrast can be observed across
detector using amorphous silicon receptor with cesium the entire thickness range of the object.
iodide scintillator. This now provides information on the
weld almost to the tangent point and
X-rays assists the operator in identifying
discontinuities over a wider range of
Cesium thickness in a single view.
Visible photons
iodide Most flat panel receptors available
Bias today are designed to provide
Output radiographic acquisition capability at a
rate of one image about every 5 to 10 s.
Some designs take more or less time to
put the image on the monitor but most
Photodiode fall into this range. This speed is certainly
Row select
much faster than what can be achieved field lines. Because the field lines are
with film cassettes. This technology parallel to the incident X-ray beam (other
represents an advance in practicality since than for oblique angles), the field
the 1990s. prevents the charge from lateral scattering
and thus there is virtually no blur.
Intuitively, this would seem to suggest
that the amorphous selenium conversion
Amorphous Selenium layer (excluding the pixel electrodes)
Detectors should exhibit extremely high resolution.
In fact, measurements prove this to be the
In flat panel arrays using an amorphous case.
selenium converter (or other
photoconductors), the X-ray to electrical
charge conversion process is referred to as
direct because no intermediate steps are Charge Coupled Device
required. As shown in Fig. 11, the high
voltage bias field applied to an Radiographic Systems
amorphous selenium layer creates vertical Charge coupled devices are used in X-ray
imaging systems in combination with
X-ray phosphors or scintillators without
FIGURE 11. Schematic cross section of amorphous selenium the need for electronic image
X-ray detector.12 intensification. A charge coupled device is
an integrated circuit formed by depositing
Top bias electrode Positive voltage a series of electrodes, called gates on a
semiconductor substrate to form an array
of metal oxide semiconductor (MOS)
capacitors. By applying voltages to the
Amorphous selenium
gates, the material below is depleted to
Field lines form charge storage wells. These store
charge injected into the charge coupled
device or generated within the
Trapped holes
semiconductor by photoelectric
Pixel electrode Pixel electrode absorption of optical quanta. If the
voltages over adjacent gates are varied
++++ appropriately, the charge can be
Insulator
Drain Source Drain Source transferred from well to well under the
gates, much in the way that boats will
Gate Glass substrate Gate move through sets of locks as the
potential (water heights) are adjusted.12
FIGURE 12. Charge couple device based X-ray detector: (a) X-rays directly excite charge
coupled device through phosphor (phosphor does not provide enough shielding); (b) fiber
optic scintillator coupled directly to charge couple device provides shielding to sensor.
(a)
Phosphor layer, 0.05 to 0.20 mm
(0.002 to 008 in.) thick
(b)
Fiber optic protection plate or fiber optic
scintillator, 1 to 25 mm (0.04 to 1.0 in.) thick X-rays
In the simplest charge coupled device surface. Fiber optic tapers thereby increase
systems, the charge coupled device is the field of view, provide efficient light
rapidly scanned to provide television collection (with respect to a lens), offer
frame rates with typical exposures per shielding of the charge coupled device
frame of 33 ms. In this mode, the signal from direct X-ray hits and can yield a
captured can be very low and the compact, light weight rigid design. Fiber
resulting signal-to-noise ratio will optic tapers have now been incorporated
therefore also be low because of the small with a 100 × 100 mm (4.0 × 4.0 in.) active
number of photons impinging on the area.25
phosphor in the time allotted and the A lens as an optical coupling device has
high noise level of the charge coupled the drawback that it is a very inefficient
device. The noise of the device increases light collection device. Relative to a fiber
as a function of the square root of the optic taper a lens system is less efficient
readout speed and is quite high at real by a factor roughly of ten or more. This
time frame rates. The image quality can inefficiency can lead to secondary
be improved by averaging multiple frames quantum sinks and additional noise in
in a digital processor but the high noise of the image. Secondly, the lens does not
the device operating at these speeds does provide adequate shielding to the charge
not provide film quality images. coupled device, so an additional shielding
The better way to improve glass is needed directly in front of the
signal-to-noise ratio using charge coupled charge coupled device to reduce direct
devices, is to integrate the charge X-ray hits on the device. In addition, a
produced by light from the phosphor mirror can be used to move the camera
directly on the charge coupled device out of the radiation beam. The charge
cells. The wells generated by the readout coupled device can then be shrouded in
approach can be sufficiently deep to lead to reduce excitation by tangentially
capture three to four orders of magnitude scattered X-rays. One advantage of a lens
in equivalent light levels. Because the is the increased flexibility it offers to
exposure times are now much slower than
the real time rates of traditional charge
coupled device video cameras, the readout
FIGURE 13. Coupling of light from phosphor to charge couple
speed can be reduced to obtain lower
device in X-ray detector system: (a) lens coupling; (b) fiber
camera noise levels. On a frame-by-frame
optic coupling.
basis, the signal levels have been
increased while the additive noise from (a)
the camera has been decreased. In this Fiber optic scintillator or
mode, further electrostatic image phosphor or both
intensification is not needed.
Charge coupled devices are now
available with image formats as large as
4096 × 4096 pixels and 16 bits. Some
devices have been made as large as
60 × 60 mm (2.4 × 2.4 in.). A phosphor
screen can be coupled directly to the
charge coupled device itself but even if
the phosphor has good X-ray quantum
efficiency, those X-ray photons not
absorbed by the phosphor (even if it is a Lens
small percentage) can still be absorbed in
Shielding glass Cooled charge coupled
the silicon layer of the charge coupled device camera
device and yield a significant direct
excitation speckle noise in the image. To (b)
avoid this noise a fiber optic image Scintillating
transfer plate or a scintillating fiber optic fiber optics
plate may be used16 to absorb the X-ray shield
transmitted X-rays before being absorbed
in the silicon (see Fig. 12).
The field of view of the charge coupled
device based X-ray systems can be
Charge
expanded with a fiber optic taper or a lens Fiber optic coupled
system. These configurations are shown in taper device
Fig. 13. camera
Fiber optic tapers are fiber optic face
plates in which the size of each fiber in
the face plate is reduced so that an image
deposited at the input surface may be Phosphor layer
transferred to a smaller device such as a X-ray and light shield
charge coupled device at the output
References