Service Manual of FUS-100 Urine Sediment Analyzer
Preface
Introduction:
♦ Please read the User Manual and Service Manual carefully after unpacking.
♦ Installation, debugging and maintenance should be executed in strict accordance with the
Manual.
♦ Keep the Manual safely for your any time reference.
♦ Incorrect use may affect instrument accuracy and precision or endanger personal safety.
Note:
♦ Instrument should be operated by medical inspection specialist, physician, nurse or lab
assistant whom are specially trained.
♦ Instrument should be controlled by special software. Please install the software that is
appointed by our company. Installation of other software/hardware may interfere normal
working. Don’t run other software when instrument is working.
♦ Dust may accumulate on the surface of instrument after long time storage. Clean soft cloth or
gauze can be used for cleaning work, and a little detergent can be used if necessary. Please
cut off the power supply before cleaning.
♦ As to the use and storage method of Control, Standard Solution and Focus, please refer to the
relevant instruction.
♦ Sample, Controls, Standard Solution, Focus and waste solution, etc. have potential
biochemical infectivity, the detergents are corrosive that may hurt eyes, skin and mucosa.
Operator should refer to the safety regulation for lab operation. Protective measure should be
taken (Such as lab protective clothes and gloves).
♦ Avoid contact with eyes and skin, in case of skin contact, rinse immediately with plenty of
water and seek medical advice.
♦ Operator should comply with the local regulation when draining and dealing with reagent,
waste solution, waste sample, consumable, etc. Please dispose the waste solution and
instrument consumable according to the regulation of medical waste, infective waste and
industrial waste.
Service Manual of FUS-100 Urine Sediment Analyzer
Warning:
♦ Instrument should be operated in a good ground condition, and an independent power supply
is a must, the input power should conform to instrument requirement.
♦ Don’t pull the electrical wire with wet hand, or there is a risk of electrical shock.
♦ Don't stamp, twist or drag the wire and cable, or it may cause a fire.
♦ Please don’t open the back and side cover board before cutting the general power supply
except DIRUI service staff.
♦ If liquid occurs inside instrument or there is an internal pipeline leakage, please cut off the
general power supply immediately, and contact DIRUI service dept.
♦ Periodic maintenance should be executed strictly according to the User Manual and Service
Manual. Or it may cause instrument malfunction, or affect the accuracy and precision of test
results.
♦ Make sure that the Urine Sediment Analyzer is operated according to the User Manual, or the
test result will not be reliable, and the damage on instrument may endanger personal safety.
♦ Please don’t place combustible material around the instrument.
Service Manual of FUS-100 Urine Sediment Analyzer
Content
Chapter Ⅰ Introduction .................................................................................................................. 6
1.1 Urine Sediment Analyzer Overview ................................................................................... 6
1.2 Test Principle ....................................................................................................................... 6
1.3 System Principle Diagram................................................................................................... 6
1.3.1 Instrument Compose .............................................................................................. 6
1.3.1.1 Front View ..................................................................................................... 6
1.3.1.2 Rear View ...................................................................................................... 8
1.3.1.3 Right Side View ............................................................................................ 8
1.3.2 Air and Liquid Principle Diagram ............................................................................ 9
1.3.3 Electric Principle Diagram ....................................................................................... 9
1.3.4 Optics Principles Diagram ..................................................................................... 11
1.4 Test Item and Image .......................................................................................................... 12
1.5 Marks ................................................................................................................................ 13
Chapter Ⅱ Installation .................................................................................................................. 14
2.1 Installation Instruction ...................................................................................................... 14
2.2 External Device Connection ............................................................................................. 14
2.2.1 Offline Mode .......................................................................................................... 14
2.2.2 Online Mode .......................................................................................................... 14
2.2.3 Software Installation .............................................................................................. 14
Chapter Ⅲ Focus and Calibration ................................................................................................. 15
3.1 Focus ................................................................................................................................. 15
3.2 Calibration......................................................................................................................... 15
Chapter Ⅳ QC .............................................................................................................................. 17
4.1 Control .............................................................................................................................. 17
4.2 QC Register and QC Setting ............................................................................................. 17
4.3 QC Test ............................................................................................................................. 19
4.4 QC Review ........................................................................................................................ 20
4.4.1 QC Result Inquiry .................................................................................................. 21
4.4.2 QC Result Modification ......................................................................................... 25
4.4.3 Add QC Result ....................................................................................................... 26
4.4.4 Delete QC Result.................................................................................................... 26
4.5 QC Stat. ............................................................................................................................. 26
4.5.1 QC Chart ................................................................................................................ 27
4.5.2 About QC ............................................................................................................... 28
Chapter Ⅴ Sample Test ................................................................................................................ 30
5.1 Test Preparation ................................................................................................................. 30
5.1.1 Note for Sample Preparation .................................................................................. 30
5.1.2 Sample Volume ...................................................................................................... 30
5.1.3 Tube Requirement .................................................................................................. 30
5.1.4 Barcode Using Requirement .................................................................................. 30
5.1.5 Check Waste Liquid, Sheath and Printer ................................................................ 32
Service Manual of FUS-100 Urine Sediment Analyzer
Chapter Ⅰ Introduction
6
Service Manual of FUS-100 Urine Sediment Analyzer
Figure 1-1
“Start” button is used for upgrade program, its usage method is as follow:
Switch off the power switch under standby, press “Start” button, switch on power switch, release
“Start” button. Open the upgrade software within the computer, select the document need to be
upgraded. Start upgrading, switch off power and restart after upgrading.
7
Service Manual of FUS-100 Urine Sediment Analyzer
Figure 1-2
Figure 1-3
Note:the fixing baffle should be removed when connecting with Urine Analyzer.
8
Service Manual of FUS-100 Urine Sediment Analyzer
9
Service Manual of FUS-100 Urine Sediment Analyzer
10
Service Manual of FUS-100 Urine Sediment Analyzer
11
Service Manual of FUS-100 Urine Sediment Analyzer
Squamous Epithelial
SQEP
Cell
Non-Squamous
NSE
Epithelial Cell
Bacteria BACT
Crystal CRYS
Yeast BYST
Sperm SPRM
Mucus MUCS
Note:The disk equipped with the instrument contains classic cell images.
12
Service Manual of FUS-100 Urine Sediment Analyzer
1.5 Marks
Marks Meaning
AC mark
Storage temperature
Lot No.
Warranty date
Series No.
Date of manufacture
Manufacturer
Grounding mark
13
Service Manual of FUS-100 Urine Sediment Analyzer
Chapter Ⅱ Installation
The instrument installation shall be executed by professional. And the space, power supply and
environment requirement should be met. If the installation is conducted by the user, User Manual
should be read carefully beforehand.
Take down the tie wire used for fixing sample probe after sample carrier installation, otherwise,
the instrument can not work.
The connection should be conducted in strict accordance with the User Manual.
The online software will be installed by professional, the user shall not uninstall the software
except particular condition. If the software needs to be reinstalled or uninstalled, follow the user
manual.
14
Service Manual of FUS-100 Urine Sediment Analyzer
3.1 Focus
In order to make sure the visible component in urine is captured clearly when passing through
focusing plane of microscope, daily focus should be conducted before test.
Note:
2. In order to capture the clear images, the Focus specified by DIRUI should be used.
3.2 Calibration
(a) Standard Solution Storage
♦ Standard Solution is the solution containing RBC, refer to instruction for cell volume.
♦ Standard Solution should be stored at 2--8℃. It can be used after placing at room temperature
for a while.
(b) Standard Solution Usage
♦ Shake before use:inverse the Standard Solution bottle, shake back and forth for 5 times, and
shake the bottle for 5 times, it can be used when no air bubble exists.
♦ The standard solution should be used immediately after unseal. Each bottle is for single time
use, the left can not be used. Standard solution mixture with different lot No. can not be used.
(c) Calibration Frequency
♦ Monthly FUS-100 Urine Sediment Analyzer calibration should be conducted, focus should be
conducted before calibration.
(d) Calibration Preparation
♦ Take 10 glass test tubes, inject 4ml Standard Solution in each tube.
♦ Place 10 test tubes in the test tube rack.
(e) Calibration Note
♦ Place the test tube rack on the right of the sample carrier, click “Calib” in figure 3-1, refer to
15
Service Manual of FUS-100 Urine Sediment Analyzer
Figure 3-2
♦ Input lot No. and Mean Value of Standard Solution in figure 3-2, click “Calib”, sampling
probe will conduct sampling automatically. The calibration status will be displayed in the
interface after 10 test tube test has been finished.(meanwhile, “System Monitor” interface
will display calibration time and calibration coefficient.). The “Status” column will display
“Pass” if calibration succeed. The instrument will enter into standby automatically.
♦ The “Status” column will display “Fail” if calibration failed. Conduct calibration again after
standard solution replacement (with another lot No.) . Contact the after-sale service dept. or
distributor of DIRUI if calibration still failed.
16
Service Manual of FUS-100 Urine Sediment Analyzer
Chapter Ⅳ QC
QC function is to check and test analysis accuracy through testing urine control and collecting QC
data.
4.1 Control
(a) About Control
Control is used for QC of dimming and test result. Positive Control is used for abnormal QC,
negative Control is used for normal QC, both positive Control and negative Control are used for
checking whether the instrument can conduct counting accurately.
Positive Control is the ionic equilibrium solution containing fixed volume RBCs. Refer to the
User Manual for cell volume.
Negative Control does not contain particle.
Positive Control and Negative Control must be refrigerated and placed at room temperature for a
while before use.
(b)QC Frequency
Positive and negative QC should be conducted at least once a day, or according to the standard of
lab QC procedure manual. Focus should be conducted before QC.
17
Service Manual of FUS-100 Urine Sediment Analyzer
Figure 4-1
(a)Add QC Item
Input corresponding parameter in the input box behind “Model”, “Mean Value” and “Name”.Click
the drop-down menu behind “Validity”, select the validity marked on the QC instruction, and click
“Add”.
(b)Delete QC Item
Select the QC item need to be deleted in QC register list(as figure 4-1 shows), click “Delete”, as
figure 4-3 shows:
Figure 4-3
(c)QC Setting
18
Service Manual of FUS-100 Urine Sediment Analyzer
Figure 4-3
1SD、2SD、3SD means 1 time, 2 times and 3 times standard deviation for test result. If the
【Exceed 3SD】alarm range is selected, the instrument will prompt alarm when the QC result
exceeds 3SD.
Figure 4-4
4.3 QC Test
(a)Control Usage
Shake before use: inverse the Positive Control bottle, shake back and forth for 5 times, and shake
the bottle for 5 times, it can be used when no air bubble exists.
Note: Negative Control does not contain particle, shaking is not needed.
19
Service Manual of FUS-100 Urine Sediment Analyzer
♦ Inject 3ml Positive Control, 3ml Negative Control into two test tubes.
♦ Place the 2 test tubes in the test tube rack according to the sequence of QC register.
(d)QC Test
Figure 4-5
♦ Place the test tube rack on the right of the sample carrier. Select the box in front of the lot
♦ The “Status” column will display “Fail” if QC failed. Conduct QC again after Control
replacement (with another lot No.). Contact the after-sale service dept. or distributor of
DIRUI if QC still failed.
4.4 QC Review
20
Service Manual of FUS-100 Urine Sediment Analyzer
Click “QC Review”in figure 4-5, as figure 4-6 shows, the tested QC information can be inquired.
Figure 4-6
Figure 4-7
21
Service Manual of FUS-100 Urine Sediment Analyzer
Select the lot number (for example select lot number 000001) in the box under lot number and
Figure 4-8
All QC result with lot number 000001 will be displayed on the screen, click to get
the preview for QC result , shown as figure 4-9.
22
Service Manual of FUS-100 Urine Sediment Analyzer
Figure 4-9
Select the QC name in the list of name(for example select negative) in the box under name in
23
Service Manual of FUS-100 Urine Sediment Analyzer
Figure 4-10
All negative QC result will be displayed on the screen, click button to get the
preview for QC result as shown in figure 4-9. Click button for the print of the QC result.
Select the status after QC test (for example select pass) in the box under status in figure 4-7 and
Figure 4-11
All passed QC result will be displayed on the screen, click button to get the preview
for QC result as shown in figure 4-9. Click button for the print of the QC result.
Select the box in front of【use date –time】in figure 4-7 , choose the start & end date, time and
24
Service Manual of FUS-100 Urine Sediment Analyzer
Figure 4-12
All QC result within the time will be displayed on the screen, click button to get the
preview for QC result as shown in figure 4-9. Click button for the print of the QC result.
Figure 4-13
25
Service Manual of FUS-100 Urine Sediment Analyzer
Select the lot number, name and date and time of QC test need to be modified and input QC
Note: the low value, high value and mean value of QC test can not be modified.
Figure 4-14
Select the QC lot No., name and date, time of QC test need to be added. Input the low value, high
4.5 QC Stat.
Click in figure 4-6, as figure 4-16 shows to review and print QC chart.
26
Service Manual of FUS-100 Urine Sediment Analyzer
Figure 4-16
4.5.1 QC Chart
(a)About QC Chart
The abscissa of the QC chart is the date of the QC, ordinate is the QC line.
QC chart can display 31 points at most.
Each point correspond to 1 QC result, the points are connected by segment.
(b)QC Chart Review and Print
Click the drop-down list behind QC date, select the QC date; select the QC lot No. in the
drop-down list behind QC lot No., and select the QC name in the drop-down list of QC name, the
inquired result will be displayed in the QC chart.
27
Service Manual of FUS-100 Urine Sediment Analyzer
Figure 4-17
x (mean value):the mean value for a group of number, and calculate with the following formula:
∑x
i =1
i
x=
n
SD(standard deviation):a statistics method to describe the dispersion degree for a group of test
∑ (x
i =1
i − xi ) 2
SD=
n −1
CV valve (coefficient of variation): the ratio between standard deviation and mean value called
28
Service Manual of FUS-100 Urine Sediment Analyzer
CV value, it is a statistics reflect the variation degree for each test result, and calculate with the
following formula:
SD
CV = × 100%
x
29
Service Manual of FUS-100 Urine Sediment Analyzer
The minimum urine volume is 4.0 ml(Use FUS -100 alone and connecting with DIRUI H-800).
Note:
Do NOT fill the test tube at any time, because auto sample mixing will be conducted before
sampling, sample overflow should be avoided.
5.1.3 Tube Requirement
(a)Sample tube standard: 16mm*100mm tube, the tube mouth must be regular without
deformation.
(b)Sample tube rack standard: use the supplied sample tube rack.
5.1.4 Barcode Using Requirement
(b)Print width for barcode should between 8~12mm. And the valid length for barcode should
30
Service Manual of FUS-100 Urine Sediment Analyzer
be less than 40mm.The start blank and finish blank on the barcode should be more than 3mm.
And the number on barcode should be more than 4 bits and smaller than 18 bits. As shown in
figure 5-1:
Figure 5-1
Figure 5-2
3. When the barcode is CODE39 small letter, the scanned ID number on screen and print report
4. For example:ID number is 36Fa, and the scanned number will be 36F+A,“+” take one place
too.
31
Service Manual of FUS-100 Urine Sediment Analyzer
Code 39 5~13
Code 93 5~15
CODABAR 11
Figure 5-1
(a) Turn on the computer, and the power supply switch for display;
(b) Turn on the power supply for printer and place the print paper;
(c) Turn on the power supply for FUS-100 urine sediment analyzer;
(d) Register FUS-100 urine sediment analyzer software and the instrument enter standby status.
Note: Switch on the power supply of FUS-100 urine sediment analyzer, and then login the
software.
32
Service Manual of FUS-100 Urine Sediment Analyzer
Figure 5-3
Urine Sediment Analyzer Mode: confirm “Urine Sediment Analyzer Mode” is on “Offline Mode”.
Urine Analyzer Status:make sure “Urine Analyzer Status” is on “Offline mode”.
LIS System: if LIS system is used, connect data cable, click” Check LIS”, make sure LIS system
is on “Online”.
Offline test can be conducted when above status are confirmed.
(b) Warning Tone for Alarm
The instrument has both alarm mark and warning tone during alarm. And you can select
Figure 5-4
So the instrument will give double remind for alarm.
(d)Focus: conduct once a day after booting according to User Manual.
(e)Calibration: conduct once a month according to User Manual.
(f)QC: conduct once a day after booting according to User Manual.
Note: the instrument should conduct focus, calibration and QC by turn, and conduct focus
and QC by turn when no need to do calibration.
5.2.3 Offline Sample Test
(a)Place the test tube with barcode outside and urine sample inside in the test tube rack, the
direction of the test tube should be straight.
(b)Place the test tube rack on the right of sample carrier, click “Start test” in “Sample Register”
33
Service Manual of FUS-100 Urine Sediment Analyzer
interface.
♦ Click , set ID “ON”, check whether “Test Tube Rack Load Mode”is “Online”,
contact DIRUI after-sale service personnel if it is “Offline”.
♦ Refer to Chapter Ⅲ of H-800 Urine Analyzer Manual for other setting.
5.3.3 Online Sample Test
(a) Place the tubes with barcode on the tube rack vertically.
(b) Place the tube rack on the right side of the H-800 urine analyzer sample carrier as figure 5-5
shows:
34
Service Manual of FUS-100 Urine Sediment Analyzer
Figure 5-5
Click the button in front of【start】on the screen and the instrument will feed in the tube
rack to test the sample one by one. The test result will be transmitted to the image working station
software, and display the sample NO., tube NO. and rack NO., meanwhile the current status will
be displayed in “Dry Chemistry” and “Online” line( appear in “Dry Chemistry” line means
the dry chemistry test has been finished, appear in “Online” column means the current
status is connected), as figure 5-6 shows:
Figure 5-6
The test tube rack will conduct feed-in after Urine Analyzer test has been finished, and the rack
Note: the number on the right upper corner of the urine analyzer screen, under the key
button is the reagent strips can be tested and the rack no. and tube no. of the sample to be
tested.
(c)After the test of urine analyzer, click “Sample Register” of main menu area of Urine Sediment
Image Workstation, click “Start Test” or press “Start” at the left upper corner of urine sediment
analyzer, the test tube rack will be transmitted to the sampling position of urine sediment analyzer
automatically, the following box will pop up meanwhile:
35
Service Manual of FUS-100 Urine Sediment Analyzer
Figure 5-7
Input the number after “Code” of main screen before test in figure 5-7, click “OK”, urine sediment
(d)If connecting bridge failure happens, set the urine analyzer to be “Online” in urine sediment
image working station software(“Urine Analyzer Status” in figure 5-8). Set the “Test tube rack
transmitting mode” of urine analyzer to be “Off-line”. The data will be transmitted to the urine
sediment analyzer after each sample test. The test tube rack will be transmitted to the left of the
Place the test tube rack on the right of auto-sampler of urine sediment analyzer after sample test.
Figure 5-8
Click “OFF” after “Urine Sediment Analyzer Online Mode” in figure 5-8, the urine sediment
analyzer is on “Off-line” status, and click sample register of main menu of urine sediment
analyzer image working station software, click “Start Test” or press “Start” at the left upper corner
of urine sediment analyzer, the test tube rack will be transmitted to the sampling position of urine
sediment analyzer, figure 5-7 will pop up, input the start sample NO. for test, click OK, the urine
sediment analyzer will start test from the input sample NO.
36
Service Manual of FUS-100 Urine Sediment Analyzer
Note:
Patient information editing can be conducted before, during and after sample test(before
Figure 5-9
Input sample number in “sam. No.” box and it can be selected through “previous sample” button
or “next sample” button.
Input pat. name, age, ID, bed No., detailed information and select patient sex, age, patient type,
sample type, date for sending, department for sending, doctor for sending, test doctor, dilution
times, etc. Memoni can be used in register.
Click “Modify” after editing, patient information will be displayed in browser.
Note: input patient name in patient information editing, otherwise, patient information
register can not be conducted.
Age: input age in input box, and select the age unit in the drop down list. This function can be set
in “System Management”, refer to User Manual for detail operation.
Patient type: select a right type in drop down list. This function should be set in “system
management” interface of main key area, refer to User Manual for detailed operation.
Sending department: select the department type for the sending sample. This function should be
37
Service Manual of FUS-100 Urine Sediment Analyzer
set in “system management” interface of main key area, refer to User Manual for detail operation.
Sending doctor: select the doctor name for the sending sample. This function should be set in
“system management” interface of main key area, refer to User Manual for detail operation.
Sample type: select the suitable type. This function should be set in “system management”
interface of main key area, refer to User Manual for detail operation.
Dilution times: select the suitable type. This function should be set in “system management”
interface of main key area, refer to User Manual for detail operation.
5.4.2 Alarm Information Confirm
If failure occur during test, the software will prompt alarm remind and you can see the alarm mark
Click the mark to display alarm code, alarm degree, alarm information, etc, as shown in
figure 5-10:
Figure 5-10
If the detail information need to be checked, click the information line, the detail alarm
information and the detail solution will be displayed, as figure 5-11 shows:
Figure 5-11
Refer to the failure solution, please contact DIRUI after-sales personnel if trouble can not be
solved.
38
Service Manual of FUS-100 Urine Sediment Analyzer
Select the alarm information need to be deleted and click button under it. Click the
button under it and all the selected record will be deleted. Click button and all the
alarm record will be deleted.
If there is any sample need to be reviewed after first time test, select the sample number in “test
result” list, which is , click【review】button in figure 5-9, and shown in figure 5-12 :
Figure 5-12
Put the review samples on another tube rack and put the tube rack on the right side of the FUS-100
urine sediment analyzer sample carrier.
Re-select the sample number in figure 5-12 and click “Start Review” button for the review of the
sample.
The review result will be displayed in result column after sample review.
Note:
Only the test result of the day can be reviewed.
Dry chemistry result can not be reviewed, conduct re-test if review is needed.
Review sample test must be realized by clicking【start review】 button to send the order for
review.
Double click the tested sample information to check the test result in “Test Result” list, as figure
5-11 shows:
Figure 5-11
The urine dry chemistry result is displayed on the right side of the interface.
The urine sediment result is displayed on the left side of the interface, each visual component and
standard valve value can be checked. The result within normal range will be green, the result
exceeds normal range will be red, and the exceed area will be red.
If there are to many visual components in the sample test result, click button on the
lower left place of the screen to review next page, and click button to read previous
page.
Double click the button with cell mark for urine sediment result and the enlarged image will
40
Service Manual of FUS-100 Urine Sediment Analyzer
Figure 5-12
Click button or button on the right side of the screen to see the image for
Select the image need to be added in figure 5-12, right click the mouse,
will pop up, select “Send to the print report ”as figure 5-13 shows, left click the mouse to send the
image to the remark information column of the report.
Figure 5-13
Report Image Deleting
If the image send to the printing report is found to be wrong, select “Delete from printing report”
41
Service Manual of FUS-100 Urine Sediment Analyzer
If the image is not consistent with the test item name, the images can be reclassified, and the
operation will be as follows:
Select the image in Figure 5-12(inconsistent with name), press the correspond button on the right
of the screen to reclassify. Click “Result” and return to 5-11 interface, click “Accept” to save
and confirm the above modification.
Click button or button on the right side of the screen to check other test
item. Repeat the above steps until all the image need to be modified have been finished.
In the above procedure , if mistake in operation has been found, press【recover】button in figure
5-11 to cancel the operation and conduct reclassification.
WBC, crystal, WBC Clumps, Bacteria, RBC, Yeast, Squamous Epithelial Cells, sperm,
Non-Squamous Epithelial Cells, Mucus, Hyaline Casts, Pathological Casts.
42
Service Manual of FUS-100 Urine Sediment Analyzer
Select the sample and click for a single sample record information in Figure 5-7,
and the corresponding item will have status mark , sample examine has been finished.
(b)Lot Sample Examine
Click for lot samples record information in Figure 5-7, and input the sample
number for start and end samples in “sample number” box. Select the examiner in the drop-down
list as figure 5-14:
Figure 5-14
Click “OK”, close “Lot Examine”window, the examine record number will be displayed.
Note:
Examine can only be conducted for the sample with result and name.
The sample only with dry chemistry result can not be examined.
The start sample number for lot examine must be smaller or same as the end
sample number.
User with operator permission can only modify and delete the unexamined samples
of the day. User with administrator permission can modify the examined and
unexamined sample of the day.
43
Service Manual of FUS-100 Urine Sediment Analyzer
Figure 5-16
Figure 5-17
Note: the start sample number must be smaller or same as the end sample number.
Click in figure 5-7 if any sample result need to be deleted, as figure 5-18 shows:
44
Service Manual of FUS-100 Urine Sediment Analyzer
Figure 5-18
Input the start No. and the end No. of the sample need to be deleted in the input box of “Sample
No.”, click “OK” to delete the sample results.
Note:
The start sample No. must not be bigger than end sample No.
Periodical database backup is recommended to avoid accidental data loss. Database backup and
recovery should be conducted under standby. Database recovery should be conducted with caution
for it may overwrite the current database.
45
Service Manual of FUS-100 Urine Sediment Analyzer
Chapter Ⅶ Troubleshooting
If failure alarm is prompted during test, test must be carried out after failure is cleared.
46
Service Manual of FUS-100 Urine Sediment Analyzer
“M5 Motor Zero Position Optocoupler Error” or “M4 Motor Zero Position Optocoupler Error” or
“M3 Motor Zero Position Optocoupler Error” is prompted during software test.
2. Solution
Dust may accumulated in the instrument after long time using. Alarm will be prompted, wipe the
optocoupler with cotton cloth in this condition. If the same optocoupler is wiped for continuous 2
times, contact the after-sale service personnel at the 2nd time.
Note: reset is needed after wiping.
7.1.4 Electric Level Error
1.Failure
“Vertical motor electric error”or “Horizontal motor electric error” is prompted during test.
2. Solution
Power cord YJ1 or TJ1 of sampling system may be touched during packaging, transportation or
debugging. Replug YJ1 into TJ1.
47
Service Manual of FUS-100 Urine Sediment Analyzer
48
Service Manual of FUS-100 Urine Sediment Analyzer
Emargination
major grove
ball bearing
F604
49
Service Manual of FUS-100 Urine Sediment Analyzer
1.Failure
For the long working period of the instrument, the 3 syringe pump at the back may make sound
during test.
2. Solution
Take down the emargination major grove ball bearing.
50