OSA3 DTV PRE-FILTER CLEANING AND REPLACEMENT

Page 1 of 2

OSA3 DTV Pre-Filter Maintenance


*** Make sure to SAVE ALL REMOVING VISCOMETER Pre-Filter
HARDWARE throughout the course of 4. Disconnect the Plumbing at the
this procedure for use during output of the large pump by simply
Re-assembly *** unscrewing the white nut. (Counter
The Pre Filter can be reached from the back clock wise).
DTV Pre
of the Analyzer… Picture # 3 Filter
1. On the back of the analyzer, remove the
6 Screws (10-32 x 3/8”) from the Back
Panel using a Torx Screw Driver
(T-25). Large pump
output

Picture # 1

DTV 5. Unscrew the Black nut at the “Output”


of the Black Filter. Hold the Middle
Piece of the Filter while you unscrew
DTV Pre
the black nut. (Counter clock wise)
Filter
Picture # 4

2. Disconnect the power wire for the 2


Fans behind the Back Panel. It will be
located on the Left side. It is a White 4
pin connection with Yellow and Black
wires. To disconnect; pull each side
apart from each other.
Picture # 2 Black Nut at
OUTPUT

Cleaning the inside of the Pre-Filter


6. Using a 3/32 Allen Key, Remove the 4
screws from the top Black piece
(Marked by 4 Blue arrows on Picture
#5) so you can get at the Filter screen.
3. Set the panel aside. Picture # 5
OSA3 DTV PRE-FILTER CLEANING AND REPLACEMENT
Page 2 of 2
7. Remove and clean the small metal 13. Screw in the Black Nut on the right side of
Filter. Use Q-tips and if needed place the Pre-Filter. (Clock wise)
the filter inside a sample bottle with 14. Make sure the 2 black nuts on both sides are
CheckFlush, close the bottle and shake screwed in all the way, and are tightened so
it until the filter is clean. there are no air leaks.
15. Make sure that the White Nut is tightened all
Small Black the way so there are no air leaks.
Metal Picture # 6 Sealing 16. Run a sample through the viscometer and
Filter Gasket watch the pre-Filter tube to make sure that
there are no air bubbles or leaks.
Service@on-siteanalysis.com
international@on-siteanalysis.com
Service Line: 1-877-685-4357

3-32”
Allen
Bit

Re-Installing the Pre-Filter into the


Viscometer
8. Place the Small metal filter back into
the longer side of the Pre-Filter.
9. Place the Black Sealing Gasket into the
shorter side of the Pre-Filter.
10. Place both Sides together making sure
that the Sealing gasket and the metal
filter are still in place.
11. Re-insert the 4 screws and make sure
that they are tightened.
12. Pull the White nut back so you can
make sure the clear tubing is inserted
into the pump all the way until it
bottoms out and then tighten the white
nut. (Clock wise).

Picture # 7

Space to ensure tube


will be pushed in all
the way.