CONTENT
CHAPTER 1 OVERVIEW.............................................................................................................................................- 6 -
1.1 FEATURES................................................................................................................................................................ - 6 -
1.2 SCOPE OF APPLICATION........................................................................................................................................... - 7 -
1.3 CONTRAINDICATIONS...............................................................................................................................................- 7 -
1.4 SAFETY REQUIREMENT............................................................................................................................................ - 7 -
1.5 ENVIRONMENTAL TESTING REQUIREMENTS............................................................................................................. - 7 -
1.6 SOUND OUTPUT DATA PUBLISHED............................................................................................................................- 7 -
1.7 SECURITY ID............................................................................................................................................................ - 8 -
CHAPTER 2 FUNCTIONS AND SPECIFICATIONS................................................................................................- 9 -
2.1 MAIN FUNCTIONS.................................................................................................................................................... - 9 -
2.2 MAIN SPECIFICATIONS.............................................................................................................................................- 9 -
CHAPTER 3 ENVIRONMENTAL AND ELECTROMAGNETIC COMPATIBILITY REQUIREMENTS.......- 10 -
3.1 ENVIRONMENTAL REQUIREMENTS......................................................................................................................... - 10 -
3.2 TRANSPORT AND STORAGE ENVIRONMENT REQUIREMENTS................................................................................... - 10 -
3.3 ELECTROMAGNETIC TABLE.................................................................................................................................... - 10 -
CHAPTER 4 APPARATUS INSTALLATION........................................................................................................... - 15 -
4.1 OUT OF BOX AUDIT................................................................................................................................................ - 15 -
4.2 PROBE INSTALLATION............................................................................................................................................ - 16 -
4.3 INSTALLATION OF EXTERNAL VIDEO DEVICE.......................................................................................................... - 17 -
4.4 PROTECTIVE GROUND WIRE CONNECTION.............................................................................................................. - 17 -
4.5 POWER CONNECTION..............................................................................................................................................- 17 -
CHAPTER 5 OPERATING INSTRUCTIONS...........................................................................................................- 18 -
5.1 SWITCH ON............................................................................................................................................................. - 18 -
5.2 SWITCH OFF............................................................................................................................................................- 18 -
5.3 KEYBOARD LAYOUT:.............................................................................................................................................. - 18 -
5.4 FUNCTION KEY DESCRIPTION................................................................................................................................. - 19 -
5.5 COMBINATION KEY FUNCTION DESCRIPTION..........................................................................................................- 21 -
5.6 SPECIFIC OPERATION INTRODUCTION.................................................................................................................... - 21 -
5.7 SPECIFIC OPERATION INSTRUCTION........................................................................................................................- 23 -
CHAPTER 6 DISPLAY MODE................................................................................................................................... - 31 -
6.1 B MODE(DEFAULT MODE).................................................................................................................................- 31 -
6.2 B/B MODE.............................................................................................................................................................. - 32 -
6.3 4B MODE................................................................................................................................................................- 33 -
6.4 B/M MODE............................................................................................................................................................. - 34 -
6.5 M MODE.................................................................................................................................................................- 35 -
6.6 B+C MODE(COLOR DOPPLER MODE)............................................................................................................... - 36 -
6.7 PDI MODE.............................................................................................................................................................. - 37 -
6.8 B+D MODE.............................................................................................................................................................- 38 -
6.9 B+C+D MODE........................................................................................................................................................- 39 -
6.10 PWD MODE..........................................................................................................................................................- 40 -
6.11 CWD MODE/ CONTINUOUS WAVE DOPPLER (PRESET)........................................................................................ - 41 -
CHAPTER 7 OPTIMIZE B MODE IMAGES........................................................................................................... - 42 -
7.1 USAGE OF B MODE................................................................................................................................................ - 42 -
7.2 CONVENTIONAL B MODE CHECK PROGRAM........................................................................................................... - 42 -
7.3 B MODE TOP / SECONDARY MENU.......................................................................................................................... - 43 -
7.4 B MODE FUNCTION EXPLANATION AND ADJUSTMENT........................................................................................... - 43 -
7.5 OPTIMIZE B MODE................................................................................................................................................. - 46 -
7.5.1 Depth adjustment................................................................................................................................... - 46 -
7.5.2 Gain Control.......................................................................................................................................... - 46 -
7.5.3 Focus number........................................................................................................................................ - 47 -
7.5.4 Frequency.............................................................................................................................................. - 48 -
7.5.5 THI.........................................................................................................................................................- 48 -
7.5.6 Dynamic range.......................................................................................................................................- 49 -
7.5.7 Scan density...........................................................................................................................................- 50 -
7.5.8 Grayscale............................................................................................................................................... - 51 -
7.6 B/M MODE ADJUSTMENT....................................................................................................................................... - 51 -
7.7 M MODE ADJUSTMENT...........................................................................................................................................- 52 -
CHAPTER 8 COLOR MODE IMAGE OPTIMIZATION....................................................................................... - 53 -
-1-
Color Ultrasonic Diagnostic Apparatus User Manual
8.1
USAGE....................................................................................................................................................................- 53 -
8.2
INTRODUCTION.......................................................................................................................................................- 53 -
8.3
COLOR FLOW MODE (CFM) PARAMETERS ADJUSTMENT...................................................................................... - 53 -
8.4
OPTIMIZE COLOR FLOW.........................................................................................................................................- 57 -
8.4.1 Activate color flow mode...................................................................................................................... - 57 -
8.4.2 C Gain....................................................................................................................................................- 58 -
8.4.3 PRF........................................................................................................................................................ - 58 -
8.4.4 Wall Filter.............................................................................................................................................. - 59 -
8.4.5 Color scanning area............................................................................................................................... - 59 -
8.4.6 Color invert............................................................................................................................................- 60 -
8.4.7 C Speed.....................................................................................................................................................- 60 -
8.4.8 Steer....................................................................................................................................................... - 61 -
8.4.9 Spatial Filter.......................................................................................................................................... - 61 -
8.4.10 Ensemble............................................................................................................................................. - 62 -
8.4.11 C Persist............................................................................................................................................... - 62 -
8.5 POWER DOPPLER IMAGING (PDI MODE).........................................................................................................- 63 -
8.6 OPTIMIZE PW MODE............................................................................................................................................. - 63 -
8.6.1 PW mode parameter adjustment............................................................................................................- 64 -
CHAPTER 9 OPTIMIZE COLOR DOPPLER IMAGES........................................................................................ - 67 -
9.1 USAGE....................................................................................................................................................................- 67 -
9.2 TYPICAL APPLICATIONS - PW (PULSE WAVE) DOPPLER......................................................................................... - 67 -
9.3 TYPICAL INSPECTION PROGRAM............................................................................................................................. - 67 -
9.4 SPECTRUM DOPPLER DISPLAY................................................................................................................................- 68 -
9.4.1 Spectrum mode...................................................................................................................................... - 69 -
9.4.2 Doppler sampling volume gate positioning...........................................................................................- 69 -
9.4.3 PRF........................................................................................................................................................ - 70 -
9.4.4 Angle correction.................................................................................................................................... - 70 -
9.4.5 Steer....................................................................................................................................................... - 71 -
9.4.6 Wall Filter.............................................................................................................................................. - 71 -
9.4.7 Baseline................................................................................................................................................. - 72 -
9.4.8 Wave Invert............................................................................................................................................- 72 -
9.4.9 Dual Synchronous / Triplex Synchronous............................................................................................. - 73 -
9.4.10 Volume................................................................................................................................................. - 73 -
9.4.11 D Velocity............................................................................................................................................ - 73 -
9.4.12 D Gain..................................................................................................................................................- 74 -
9.4.13 Frequency............................................................................................................................................ - 74 -
CHAPTER 10 GENERAL MEASUREMENT AND VALCULATION................................................................... - 75 -
10.1 DESCRIPTION........................................................................................................................................................- 75 -
10.2 OVERVIEW............................................................................................................................................................- 75 -
10.3 CHECK WORKFLOW.............................................................................................................................................. - 75 -
10.4 GENERAL MEASUREMENT................................................................................................................................... - 75 -
10.4.1 B mode general measurement..............................................................................................................- 75 -
10.4.2 Corresponding key and menu operating instructions:......................................................................... - 76 -
10.4.3 Angle measurement............................................................................................................................. - 76 -
10.4.4 Volume measurement...........................................................................................................................- 77 -
10.4.5 B/M, M mode general measurement................................................................................................... - 78 -
10.4.6 Heart Rate measurement......................................................................................................................- 79 -
10.4.7 Valve Velocity Measurement............................................................................................................... - 79 -
10.4.8 Corresponding key and menu operating instructions:......................................................................... - 80 -
10.4.9 Spectrum / D General Measurement................................................................................................... - 81 -
10.4.10 Time Measurement............................................................................................................................ - 81 -
10.4.11 HR Measurement............................................................................................................................... - 82 -
10.4.12 Velocity Measurement....................................................................................................................... - 82 -
10.4.13 Acceleration Measurement................................................................................................................ - 83 -
10.4.14 D General Measurement.................................................................................................................... - 83 -
10.4.15 Doppler Trace.................................................................................................................................... - 84 -
10.4.16 Doppler Automatic Measurement......................................................................................................- 85 -
10.5 SPECIALIST MEASUREMENT.................................................................................................................................- 86 -
10.5.1 Abdomen, General Measurement........................................................................................................ - 86 -
10.5.2 Urology Measurement Operation........................................................................................................ - 87 -
-2-
Color Ultrasonic Diagnostic Apparatus User Manual
-3-
Color Ultrasonic Diagnostic Apparatus User Manual
-4-
Color Ultrasonic Diagnostic Apparatus User Manual
-5-
Color Ultrasonic Diagnostic Apparatus User Manual
Chapter 1 Overview
The full digital color Doppler ultrasonic diagnostic apparatus is a professional ultrasonic
diagnostic system.
This manual provides operator information for ultrasound systems, readers of ultrasound
imaging technology, but does not provide training in ultrasound or clinical practice. Operators must be
trained in ultrasound before using the apparatus.
Compliance with GB9706.1-2007 Safety General Requirements, GB9706.15 Safety
Requirements Common Standards: Medical Electrical System Safety Requirements and
GB9706.9-2008 Medical Ultrasound diagnostic and Monitoring Equipment for Special Safety
Requirements.
This ultrasonic apparatus is used in patients with electric shock protection for Class I,
medical product of BF type medical diagnostic applications.
This color Doppler ultrasonic diagnostic apparatus is researched and developed by the supplier
For more information, please contact the supplier.
Note: This manual is not a maintenance manual, qualified service personnel can contact
our maintenance department.
1.1 Features
This color Doppler ultrasonic diagnostic apparatus adopts the use of foreign advanced
technology, is based on computer and ultrasound front end of a variety of advanced image processing
technology at both domestic and international combined ultrasound imaging system.
This color Doppler ultrasonic diagnostic apparatus is not only for a variety of conventional B/W
images of ultrasonic diagnosis, but also on the cardiovascular disease and other diseases for early high
image quality diagnosis based on color Doppler field.
In function, it has not only the general ultrasonic image scanning, measurement, calculation,
display, query, body mark, comment, print, medical records etc., but also can edit the corresponding
inspection report and system settings.Support the DICOM (Digital Imaging and Communication in
Medicine) protocol, which is internationally accepted medical imaging standards, accessible and easy
access to data and information exchange, and can be used as terminal access PACS (Picture Archiving
and Communication System) network, PACS is a medical image archiving and communication system,
which greatly facilitates the hospital's modern network management system, and also provides
convenient for the expert to realize the remote diagnosis of the network.
The apparatus is multi-functional, easy to operate, support B, B / B, 4B, B / M, M, B / PWD,
-6-
Color Ultrasonic Diagnostic Apparatus User Manual
In the image display, we use medical large-screen LCD color display for image display,
flicker-free, high image sensitivity, high resolution, rich gray scale, clear level, and environmental
protection without radiation. At the same time, the display can also be: up, down, left, right, front and
rear and rotation adjustment, greatly facilitate the user.
1.3 Contraindications
This apparatus does not apply to the gas organs, such as: gastrointestinal, lung and other organs
of the examination. Burning, scalding or human body surface damage, this apparatus can not be used;
other precautions see the product manual.
-7-
Color Ultrasonic Diagnostic Apparatus User Manual
1.7 Security ID
Apparatus identification instructions:
-8-
Color Ultrasonic Diagnostic Apparatus User Manual
-9-
Color Ultrasonic Diagnostic Apparatus User Manual
Electromagnetic radiation
Full-digital color Doppler ultrasonic diagnostic apparatus should be used in the electromagnetic
environment specified in the table below.
Customers or users of this Full-digital Color Doppler Ultrasonic Diagnostics Apparatus should ensure
that the apparatus is used in such an environment.
- 10 -
Color Ultrasonic Diagnostic Apparatus User Manual
Electromagnetic immunity
All apparatuses should be used in the electromagnetic environment specified in the table below.
Customers or users should ensure that the apparatus is used in such an environment.
- 11 -
Color Ultrasonic Diagnostic Apparatus User Manual
GB/T17626.6
Radiation [E1]=3V/m 3 .5
d=[ ] P 80MHz-800MHz
frequency E1
GB/T17626.3 7
d=[ ] P 80MHz-2.5GHz
E1
- 12 -
Color Ultrasonic Diagnostic Apparatus User Manual
Interval distance
The Full-digital color Doppler ultrasonic diagnostic apparatus is suitable for electromagnetic
interference controlled by radiation interference. The customers or users can help prevent
electromagnetic interference by keeping the minimum distance suggested below between the portable
mobile radio communication device and the Full-digital color Doppler ultrasonic diagnostic apparatus,
which is calculated from the maximum output power of the communication device.
- 13 -
Color Ultrasonic Diagnostic Apparatus User Manual
Max output power of According to the transmitter frequency to determine the distance between
the transmitter is watts the meters(m)
(W) 150kHz-80MHz 80MHz-800MHz 800MHz-2.5GHz
3 .5 P 3 .5 P 7 P
=d =d =d
V1 E1 E1
0.01 0.12m 0.12m 0.23m
0.1 0.38m 0.38m 0.73m
1 1.2m 1.2m 2.3m
10 3.8m 3.8m 7.3m
100 12m 12m 23m
For the max rated output power at the transmitter listed above, the recommended interval distance d in
meters (m) can be estimated by the applicable equation for the transmitter frequency, where P is the
transmitter's max rated output power declared by the transmitter manufacturer , Unit watts(W).
Note: In the 80MHz-800MHz band, apply a higher frequency range of the distance.
Note: These guidance are not common in all cases. Electromagnetic propagation is affected by the
absorption and reflection of structures, objects, and the human body.
Note: Adapter (Shanghai Biaojun Electronic Technology Co., Ltd., Model: BJE1M-0150-F301) and
power supply length of 3.8 meters.
Convex probe (F3-1 / R50 / 3.5MHz), Trans-vaginal probe (F3-2 / R10 / 6.5MHz), linear probe (F3-3
/ L40 / 7.5MHz), Phased array probe (F3-4 / PA16 / 3.0MHz) , probe plug is metal socket shielded
ground processed, probe cable length is 2.5 meters, the line to do magnetic ring shielding treatment.
Warning: In addition to the components and cables to be sold by the apparatus manufacturer, the use
of out of specified accessories, transducers and cables may result in increased equipment or system
emissions or reduced immunity.
Warning: Apparatus or systems should not be used or stacked with other devices. If you have to be
close or stacked, you should observe that the verification works under normal configuration.
Note: Based on the characteristics of the ultrasonic probe, the conduction immunity level is set to 1V /
m.
- 14 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 15 -
Color Ultrasonic Diagnostic Apparatus User Manual
Probe socket
1. Insert the probe into the probe port 2. Lift the lock knob to fix
- 16 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 17 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 18 -
Color Ultrasonic Diagnostic Apparatus User Manual
Filter adjustment
Probe switch
- 19 -
Color Ultrasonic Diagnostic Apparatus User Manual
Obstetric measurement
Measurement menu
Freeze/Unfreeze key
- 20 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 21 -
Color Ultrasonic Diagnostic Apparatus User Manual
Use the cursor to click on the drop-down menu to set the system language, date, and time. Enter
the name of the hospital and the name of the doctor in the blank of the hospital and the doctor's name .
Select the number of images to be inserted into the report, select between 1 and 4, and select the
print size according to the actual needs of the print.
Choose a different obstetric table as needed or edit the obstetric table according to your
physician's actual experience.
Also for the cine loop frame number, cine loop speed with the drop-down menu to set the default
cine loop frame number of 200 frames, up to 1200 frames, cine loop speed 5 files optional.
Set the apparatus to automatically freeze the time, without any operation in a certain period of
time, the apparatus will automatically freeze.
Upon completion of the language conversion, the software will end into the desktop, then
double-click the desktop's ultrasound icon “SCUS” to open. After all settings completed, click OK to
save all settings.
Image Note
Press the "Func" button in the real-time state, press the "Enter" key to finish the comment; press
the "Func" button to finish the comment. After the click on the "set" key appears green frame, the
removable trackball placed in the specified location, press the "set" button to complete.
- 22 -
Color Ultrasonic Diagnostic Apparatus User Manual
4B mode
In the real time state, press "Func" key and make it light, press “Q” key, to enter into 4B mode,
press “Q” continuously, four images will show in turn. Press “Func” key again to make the light off,
press “B” knob to enter into B mode.
Puncture guide
In the real time state, press "Func" key, and then press the letter "X" to show the puncture guide
line. Press "Angle" key, when light on, rotate the "encoder Adjust" knob, you can adjust the puncture
line angle, move the trackball to adjust the puncture line position, press the "SET" key to determine
the puncture line position. Press "Func" key to make the light off and then press "Clear" or "Exit" to
cancel the puncture guide line.
Body mark
In the real time state, press “Func” key, to make it light, press letter “B”, you can select body
mark. Select corresponding icon and click (in the same icon click once only), click "close" to exit.
Press "Func" key to make the light off, and other do other operations.
Set Applications
Save the current parameter combination, press "Func" in real time state, press "S", select the
corresponding application, click "Save Settings" or 5 seconds to exit automatically. When complete,
press the "Func" key to make the light off.
- 23 -
Color Ultrasonic Diagnostic Apparatus User Manual
Doctors can move the trackball to the check type under the display probe. Press the "Set" key to
select the check status. You can also use the "Probe" key during the use of the probe selection
interface.
Patient information
It is also possible to check the work without entering patient information. In order to avoid data
confusion, it is advisable to enter the patient information before the check. Establish the patient's
information base in the system to better manage the patient data.
Press the "patient" key on the operation panel to bring up the new case interface; register as
needed.
- 24 -
Color Ultrasonic Diagnostic Apparatus User Manual
Application settings, in real-time state, press "Func" key, make the light on and then press the "S"
key, screen shows as follow:
Users can personalize set according to different needs, each probe can be set to 14 different
parameters, easy to use in different circumstances. After ajustment of parameters, save the settings by
pressing "Func" key, press "S" key, click "Save Settings" button, and then click the current red button
- 25 -
Color Ultrasonic Diagnostic Apparatus User Manual
to save the settings. Click on the blue button to select a different setting if you need to switch directly.
Note, press "Func" in the frozen state, make the light on and then press the "C" key, the screen
shows as follows:
You can enter letters, numbers or Chinese, move the trackball and make the cursor to the
character input box, press "Set" key after the characters are displayed on the screen, press "Clear" to
delete the characters.
Memory function, press the "Save" button in real time, the current image is saved and displayed
at the bottom of the screen, the screen displays as follows:
- 26 -
Color Ultrasonic Diagnostic Apparatus User Manual
To recall the saved image, press the "Cine loop" key in the frozen state, all the saved images
appear at the bottom of the screen, move the cursor to the desired image and press "Set" key to display
on the screen the corresponding image, to delete as long as the cursor to the top right of the icon "×"
by "Set" can delete the image.
Local zoom, press the "ZOOM" button in real time, the screen is displayed as follows:
In the image appears in the partial magnification sampling box, in the upper left corner shows the
- 27 -
Color Ultrasonic Diagnostic Apparatus User Manual
local magnification, the default magnification is 2, you can rotate the "encoder Adjust" to change the
magnification, clockwise increase, up to 10, Minimum of 1. Select the appropriate magnification,
press the encoder, the enlarged image displayed on screen, press the "Encoder Adjust" again to exit the
local zoom state.
Body mark, press the "Func" key in real time, when its light on, press letter "B" key, the screen
displays as follows:
Move the trackball and make the cursor to the desired body mark, press "Set" key to select, and
then click "Close" button. You can also flip through the flip button on the upper right of the body logo
group to select the next page of body markers.
Cine loop, the image will automatically freeze, scroll bar of the cine loop is in the lower left of
the screen, the screen displays as follows:
- 28 -
Color Ultrasonic Diagnostic Apparatus User Manual
Press the "Cine loop" button to automatically playback images, touch the trackball can playback
manually, in the playback process, you can store the satisfied images by press "Save" key.
Report function, in the frozen state press "Report" button, select the corresponding report card,
the screen displays as follows:
- 29 -
Color Ultrasonic Diagnostic Apparatus User Manual
There is a "Diagnosis Module" option in the bottom right of the report display interface, open it,
there are different diagnostic templates for different parts, as shown below:
Fill in the relevant content and press "Preview" to view the effect, the screen displays as follows,
press "Print" to print the report, or press "Close" to exit the report.
Move the trackball and make the cursor to "Print" and press the "Set" button to print directly.
Press “Back” to exit the report function.
- 30 -
Color Ultrasonic Diagnostic Apparatus User Manual
Description
Modify the corresponding data on the left side of the screen through the keyboard arrow keys,
- 31 -
Color Ultrasonic Diagnostic Apparatus User Manual
Description
Press "B / B" to enter 2B mode, display the left frame B, and then press the "B / B" again, display
the right frame, at the same time, left frame froze. Press "knob B" to exit 2B mode.
Menu function
Modify the corresponding data on the left side of the screen through the keyboard arrow keys,
- 32 -
Color Ultrasonic Diagnostic Apparatus User Manual
6.3 4B mode
Description
Press the "Func" key to activate it, then press "Q" key to enter 4B mode, display the first picture
on the upper left, press "Q" key to display the upper right image and freeze the left picture. The third
four maps the same token, press the "knob B", exit 4B mode and back into the B mode.
Menu function
Modify the corresponding data on the left side of the screen through the keyboard arrow keys,
- 33 -
Color Ultrasonic Diagnostic Apparatus User Manual
Description
Press “B/M” key, the screen shows B/M image. Do various measurement in freeze state.
Press”Knob B” encoder to exit B/M mode.
Menu function
Modify the corresponding data on the left side of the screen through the keyboard arrow keys,
- 34 -
Color Ultrasonic Diagnostic Apparatus User Manual
6.5 M mode
Description
Press B/M key, the screen shows B/W image, press B/M again, enter into M mode. Press
“Knob B” encoder, to exit M mode.
Menu function
Modify the corresponding data on the left side of the screen through the keyboard arrow keys,
- 35 -
Color Ultrasonic Diagnostic Apparatus User Manual
Description
Press “Knob C” encoder under B mode, the screen shows B+C image mode, that is, blood flow
and B/W images superimposed. Change the position and size of the sampling box by trackball and
“SET” key. Do various measurement in freeze state. Press “Knob C” encoder again to exit B+C mode.
Menu function
Modify the corresponding data on the left side of the screen through the keyboard arrow keys,
- 36 -
Color Ultrasonic Diagnostic Apparatus User Manual
Description
Press “Knob C” encoder under B mode, the screen shows B+C image mode, that is, blood flow
and B/W images superimposed. Change the position and size of the sampling box by trackball and
“SET” key. Change the display mode into “POWER” mode by adjust the keyboard arrow keys at the
left side of the menu to enter Power Doppler Image mode. Press “Knob C” encoder again to exit PDI
mode.
Menu function
To modify the corresponding data on the left side of the screen through the keyboard arrow keys ,
rotate the "C" encoder can modify the blood flow gain, clockwise increase, counterclockwise decrease,
modify the image parameters in real time state.
- 37 -
Color Ultrasonic Diagnostic Apparatus User Manual
Description
Press “PW” encoder under B mode, the screen shows B+D mode. Change the position and size
of the sampling box by trackball and “SET” key. Press “PW” encoder again to exit B+D mode.
Menu function
To modify the corresponding data on the left side of the screen through the keyboard arrow keys ,
rotate the "PW" encoder can modify the pulse wave gain, clockwise increase, counterclockwise
decrease, modify the image parameters in real time state.
- 38 -
Color Ultrasonic Diagnostic Apparatus User Manual
Description
Press “C” encoder under B mode, the screen shows B+C mode, press “PW” encoder, it
shows B+C+D mode. Change the position and size of the sampling line and sampling box by
trackball and “SET” key, change the sampling angle by “Angle” key and “Ajust” knob. Press “PW”
encoder to exit B+C+D mode, and back to B+C mode. Press “C” encoder, to back to B mode.
Menu function
To modify the corresponding data on the left side of the screen through the keyboard arrow keys ,
modify the image parameters in real time state.
- 39 -
Color Ultrasonic Diagnostic Apparatus User Manual
Description
Press “Update” key under B+C+D mode, Or change the "display mode" column from
"synchronization" to "refresh" through the arrow keys in the left side of the screen, to enter pulse
wave. Change the position and size of the sampling line and sampling box by trackball and “SET” key.
Do various measurement in freeze state.
Menu function
To modify the corresponding data on the left side of the screen through the keyboard arrow keys ,
rotate “PW” encoder to adjust gain, clockwise increase, counterclockwise decrease. Modify the image
parameters in real time state.
- 40 -
Color Ultrasonic Diagnostic Apparatus User Manual
Description
Press “CW” key under B mode, move the sampling volume to the measurement point, press
“Update” to show pulse wave. Change the position and size of the sampling line and sampling box by
trackball and “SET” key.
Menu function
Modify the corresponding data on the left side of the screen through the keyboard arrow keys,
- 41 -
Color Ultrasonic Diagnostic Apparatus User Manual
function state
B mode diagram
7.2 Conventional B mode check program
The use of B mode check is generally carried out as follows:
1. Record the patient information associated with the check, verify the system settings (probe and
preset).
2. Allow the patient to be in the proper position and fix the device, to facilitate the operator to
operate, and to consider whether the patient is comfortable; and do the scan.
3. Collect all the data to complete the study.
4. Press “B” encoder o display the B mode. Modify the image parameters in real time By adjusting
the function of the menu at the bottom of the image. Do various measurement in freeze state.
- 42 -
Color Ultrasonic Diagnostic Apparatus User Manual
Press direction key “ ” or “ ”once, the screen for the encoder position of the menu will
change color, the color becomes gray green selected. Press direction key“ ” or “ ” once, the
corresponding data in the image changes. As shown above: current selected "dynamic range" is
adjustable.
Power
Ultrasonic transmit power, digital change is 0-15, press “ ” key to decrease the value, press
“ ” key to increase the value, real-time ultrasonic image changes with the value.
Frequency
Probe transmit frequency, different probe configuration, the transmit frequency is different. Press
the “ ” key to decrease the value, press the “ ” key to increase the value, the real-time
ultrasonic image changes with the value.
3.5MHz frequency range: 2.0MHz, 3.0MHz, 3.5MHz, 4.0MHz, 5.5MHz。
Line density
Line density, the value changes to "256" and "128" two options, “ ” or “ ”can be
adjusted. At the same time, in the image on the right side of the information display area "FPS XHz"
frame frequency display (X is with the scanning density changes when the frame rate changes).
Dynamic range
The max distortion-free output power and static system noise output power ratio of the
logarithmic value, the digital change is 0dB-270dB, press the “ ” key to decrease the value,
- 43 -
Color Ultrasonic Diagnostic Apparatus User Manual
press “ ” key to increase the value, the dynamic range controls how the echo intensity is
converted to the gray scale, thus increasing the adjustable range of the contrast. At the same time,
reduce the dynamic range through the gain adjustment compensation. The system has a reasonable
dynamic range value set before deliver out from factory.
Smooth
For image noise processing, the number changes to 0-7, press “ ” key to decrease the value,
Persist
Frame average, the number changes to 0-7, press “ ” key to decrease the value, press “ ”
key to increase the value.
THI
Focus #
The number of focus, the figure changes to 1-4, press “ ” key to reduce the number of keys,
press “ ” key to increase the number of points. The focus will appear on the edge of the image
scale with the rotation of the encoder, and the "FPSXHz" frame frequency display in the information
display area on the right side of the image (X is the number of changes in the frame rate as the focus
changes).
Sector
Sector width, the figure changes from 5% -100%, press “ ” key to decrease, press “ ”
key to increase. At the same time in the image on the right side of the information display area
- 44 -
Color Ultrasonic Diagnostic Apparatus User Manual
"FPSXHz" frame frequency display (X is the scan range changes with the frame rate changes in the
number).
iClear
For the image signal post-processing, the figure changes from 1-6, press “ ” key to decrease,
Enhance
Enhance the edge of the organ obvious degree, the figure changes from 0-7, press “ ” key to
Gray Map
The gray scale map refers to suppressing an image signal below a certain gray scale. The figure
Horizontal
The image turns left and right, the figure changes to "OFF" or "ON" two options, “ ” or
“ ”can be adjusted. At the same time in the center of the image display area (sector) with “ ”
Vertical
The image turns up and down, the figure changes to "OFF" or "ON" two options, “ ” or
“ ” can be adjusted. At the same time, in the center of the image display area "sector" with
- 45 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 46 -
Color Ultrasonic Diagnostic Apparatus User Manual
knob to display the value. The gain value varies depending on the probe, application, and frequency
settings. It is suggested that the maximum gain is the optimal gain preset at the factory and can
eliminate the noise when the image is displayed. If you change the following, the gain value will
return to the factory settings or user defaults: probe, check category / application, or new patient.
Advantages
Gain allows you to balance the echo contrast so that the capsule structure does not appear echoes
and reflects its internal organization. Effects on other controls after you output energy, you need to
adjust the gain. In general, if you increase the output energy, you need to reduce the gain; if you
reduce the output energy, you need to increase the gain. Gain and TGC together will affect each
other.The biological effect gain does not affect the output energy. But increasing the gain typically
reduces the output energy level to produce the same image quality.
Note
Always optimize the gain and then to decide whether to increase the output energy.
- 47 -
Color Ultrasonic Diagnostic Apparatus User Manual
frame rate.
Biological effects
The number of focus changes the heat index and / or the mechanical index. Note about any
affected output display.
7.5.4 Frequency
Description
Multi-frequency mode allows you to switch to the lower frequency of the current frequency of the
probe or a higher frequency. In the frequency conversion mode, select the low frequency mode of
operation, to meet the fat of the ultrasound diagnosis. Select the high frequency mode of operation, to
adapt to the clinical diagnosis of children.
Adjustment
1. Select the "Frequency" option on the left side of the screen.
2. Select the desired frequency through the button “ ” and “ ” on the keyboard.
Note
When the image is frozen, the frequency change can not be activated. The selected frequency is
displayed in the upper right corner of the screen information area.
Value
The frequency changes due to the probe and application. If you change the following, the
frequency value will return to the factory settings or user defaults: probe, check category, check
calculation, or new patient.
Advantages
In the current working probe mode, according to the characteristics of the human body to change
the frequency of scanning, to fat patient, you can choose the trend of low frequency, to thin patient,
you can choose the high frequency trend.
Biological effects
Activating the multi-frequency mode will change heat index and / or mechanical index. Note
about any affected output display.
7.5.5 THI
Description
Harmonics not only enhance far field penetration, but also enhance near-field resolution, improve
small organ imaging.
Adjustment
- 48 -
Color Ultrasonic Diagnostic Apparatus User Manual
To activate the harmonic imaging: select the "THI" button on the left side of the screen, press
“ ” and “ ” button on the keyboard, generate harmonic function, press again, return to
normal frequency mode. Selecting harmonics can be switched by pressing the “THI” button on the
keyboard.
Note: When the image is frozen, THI can not be activated.
Advantages
Harmonics can reduce the frequency of low amplitude of the large noise, improve the imaging of
patients with technical difficulties. The advantages of harmonics are particularly significant when
imaging the anatomical structures of the shallow anatomical structures of the breast, the liver and the
difficult imaging of the fetal anatomy. Harmonics can improve the image quality of the B mode
without the use of contrast agents.
Impact on other controls
You may need to change the field of view (depth). Do not use contrast agents when organizing
harmonic imaging.
Biological effects
Activating the THI mode will change the heat index and / or mechanical index. Note about any
affected output display.
“ ” or “ ” button on the keyboard. The value is set between 0dB and 270dB. The current
value is displayed on the screen. The dynamic range value varies depending on the probe, application,
and frequency settings.
Note
If you change the following, the dynamic range level will return to the factory settings or user
defaults: probe, check category, check calculation, new patient, or multi frequency.
Advantages
The dynamic range helps to optimize the texture of different anatomical structures. The dynamic
- 49 -
Color Ultrasonic Diagnostic Apparatus User Manual
range is selected for the best echo signal in the "0" mode of operation, and the higher the dynamic
range value is selected, the smaller the loss of the echo signal. The dynamic range should be adjusted
so that the edges with the largest amplitude are displayed as white, and the smallest amplitude (eg,
blood flow) is shown to be just visible.
Impact on other controls
The dynamic range is available in real-time mode. It also affects the gain.
the screen, press “ ” or “ ” button on the keyboard, it changes to a high frame rate state, then
Value
- 50 -
Color Ultrasonic Diagnostic Apparatus User Manual
7.5.8 Grayscale
Description
The apparatus provides B, 2B, M and Doppler mode system diagrams.
Adjustment
Select the "Grayscale" on the left side of the screen, press “ ” or “ ” on the keyboard,
you can get a different gray-scale images. The ultrasound image on the screen changes by pressing
“ ” or “ ”.
Value
Grayscale is different from the smallest or the most gentle transition to the dark and the most
intense. The value of the grayscale varies depending on the probe, application, and multi-frequency
settings. If you change the following, the value of the grayscale will return to the factory settings or
user defaults: probe, check category, check calculation, or new patient.
- 51 -
Color Ultrasonic Diagnostic Apparatus User Manual
Description
The pattern is generally used in cardiac mode. The menu operation of this mode is basically
equivalent to the menu function operation in B mode.
Adjustment
After entering the B / M mode, move the trackball and make the sampling line to the site of interest,
and for the M-scan, select "M-line speed", press “ ” or “ ” to adjust the speed. The selection
of speed is: 0-3. If necessary, adjust the dynamic range, TGC, gain, sound power and focus position
under B / M mode.
Advantages
You can speed up or slow down the timeline to observe events that occur more or less times over a
period of time.
Value
Different scanning speed, response different under M mode.
Note
Scanning speed and observation of human parts related to the need to observe a few peaks at a
time, the scanning speed can be slowed down.
- 52 -
Color Ultrasonic Diagnostic Apparatus User Manual
8.2 Introduction
The use of color flow pattern check is generally carried out as follows:
1. The procedure described in the B mode is used to locate the specified anatomical region.
2. After optimizing the B mode image, add colored blood flow.
3. Move the color flow area to be viewed as close as possible to the center of the image.
4. Optimize the color flow parameters to achieve a higher frame rate and display the appropriate
blood flow velocity.
5. If you need more restrictive information about blood flow, use the procedure described in
Doppler mode.
6. The following functional description, the working state should be under real-time.
- 53 -
Color Ultrasonic Diagnostic Apparatus User Manual
side of the display, press “ ” key to decrease, press “ ” key to increase, real-time ultrasound
image changes with the value.
C frequency
Select "C frequency" option on the left side of the display, press “ ” key to decrease, press
“ ” key to increase. The "C frequency" value on the right side of the screen image information
area changes as per the encoder.
Display mode
It displays according to different brightness, color, contrast and other parameters, select "display
mode" option on the left side of the display, press “ ” or “ ” key, the value can be changed to
V, T, POW.
Steer
Select "Steer" option on the left side of the display, press “ ” key to decrease, press“ ” to
- 54 -
Color Ultrasonic Diagnostic Apparatus User Manual
PRF
Repeat the scan pulse speed, select “PRF” option on the left side of the display, press “ ” key to
decrease, press “ ” key to increase. "FPS: XHz, PRF XKHz" on the image information area at the
right side of the screen changes as per the encoder. The speed range / pulse repetition rate (PRF)
determines the maximum speed that can be observed. If the blood flow velocity is adjusted too low,
the spectral image will be out of range. If the speed is set too high, the spectral image is compressed
too much on the longitudinal axis, and the accuracy will decrease. In high frequency mode should
choose to reduce the PRF.
Baseline
Select "Baseline" option on the left side of the display, press “ ” key to decrease, orange
color gray scale at the right side of the image changes from +8 ~ +53, blue color bar changes from -53
~ -8, press “ ”key to increase. Baseline adjustment is used to increase the range of Doppler flow
velocity in one direction. It is mainly used to reduce the direction of the spectrum confusion.
Threshold
Image grayscale and contrast threshold, select "Threshold" option on the left side of the display ,
press “ ” key to decrease, press “ ” key to increase, the value range 0 ~ 3, it is mainly used to
enhance the image contrast function.
Wall Filter
Select "Wall Filter" option on the left side of the display, press “ ” key to change the number
from 3-0, press “ ” key to change the number from 0-3. The real-time ultrasound image changes
as per pressing “ ” or “ ”. The "Wall Filter" on the image information area at the right side
- 55 -
Color Ultrasonic Diagnostic Apparatus User Manual
C Persist
Select "C Persist" option on the left side of the display, press “ ” key to change the number
from 7-0, press “ ” key to change the number from 0-7. The "C Persist" on the image information
Ensemble
Select "Ensemble" option on the left side of the display, press “ ” key to change the number
from 15-8, press “ ” key to change the number from 8-15. The "FPS XHz" on the right side of the
C Speed
Select "C Speed" option on the left side of the display, press “ ” or “ ” key, the blood
flow speed changes from "high" or "low", the color fringe value in the screen changes from 60 or 30
Tip: Increase or decrease the speed range according to the speed of blood flow. The speed range /
pulse repetition rate (PRF) determines the maximum speed that can be observed. If the blood flow
velocity is adjusted too low, the spectral image will be out of range. If the speed is set too high, the
spectral image is compressed too much on the longitudinal axis, and the accuracy will decrease.
Spatial Filter
Select "Spatial Filter" option on the left side of the display, press “ ” key to change the
number from 3-0, press “ ” key to change the number from 0-3.
C Gain
Select "C Gain" option on the left side of the display, press “ ” key to change the number
from 127-0, press “ ” key to change the number from 0-127. The "C gain" value on the right side
- 56 -
Color Ultrasonic Diagnostic Apparatus User Manual
Color Invert
Select "Color Invert" option on the left side of the display, press “ ” or “ ” key, the word
changes between OFF and ON, while the color of the screen inverted.
- 57 -
Color Ultrasonic Diagnostic Apparatus User Manual
8.4.2 C Gain
Description
Increase the gain adjustment to increase the color brightness, easy to observe, but while the gain
increases, the noise signal is also amplified, interference on the color flow signal observation. For
low-speed low-flow blood flow testing, the gain should be increased so that these blood flow can be
clearly displayed, but at the same time should pay attention to avoid excessive gain and noise signal,
affecting the observation of blood flow, and even cause illusion. The gain amplifies the overall
strength of the echo in the color flow window or the spectral Doppler time line.
Adjustment
The gain values vary depending on the probe and the application; they are independent of the
exact position of the Gain knob. To decrease / increase the gain, adjust the knob to the left / right.
Value
Color gain values vary depending on the probe, application, and multi-frequency settings. The
gain is displayed as %. If you change the following, the gain value will return to the factory settings or
user defaults: probe, check category, check calculation, new patient, or multi frequency.
Advantage
Allows control of the number of intravascular colors, or the addition or removal of spectral
information.
Biological effects
The gain does not affect the output energy. However, increasing the gain typically reduces the
output energy level to produce the same image quality.
8.4.3 PRF
Description
The speed range / pulse repetition rate (PRF) determines the maximum speed that can be
observed.If the blood flow velocity is adjusted too low, the spectral image will be out of range. If the
speed is set too high, the spectral image is compressed too much on the longitudinal axis, and the
accuracy will decrease. In high frequency mode, select to reduce the vale of PRF.
Adjustment
Select ”PRF” on the left side of the screen, press “ ” or “ ” key, to carry out the PRF
value of the adjustment, the value changes from 2KHz --- 10KHz, if the value is too large, it will
affect the blood flow frame rate. By adjusting the image depth also changes the PRF.
Value
PRF in kHz. If you change the following, the PRF value will return to the factory settings or user
- 58 -
Color Ultrasonic Diagnostic Apparatus User Manual
defaults: probe, check category, calculation, or new patient. Note that changing the pulse repetition
rate (PRF) affects the output energy, the frame rate, and the wall filter.
Biological effects
Changing the pulse repetition frequency (PRF) range will change the heat index and / or the
mechanical index. Note about any affected output display.
digital display in the display information area as per pressing “ ” or “ ” key. The wall filter
value varies depending on the probe and application. If you change the following, the factory settings
or user defaults are returned: probe, check category, check calculation, or new patient.
Advantages
Removes unnecessary low frequency signals caused by motion.
- 59 -
Color Ultrasonic Diagnostic Apparatus User Manual
To invert the color flow, select "Color Invert" option on the left side of the screen, press “ ” or
“ ” key, color gray scale and color flow on the image invert changes. In synchronous mode, both
the color flow and the Doppler mode speed scale are reversed.
Value
Reversal and non-reversal. Values vary by probe and application. If you change the following, the
reversal value will return to the factory settings or user defaults: probe, check category, check
calculation, or new patient.
Advantages
Allows you to watch your bloodstream according to your individual needs without turning the probe.
8.4.7 C Speed
Description
Optimize the color flow frame rate or spatial resolution to get the best color image.
Adjustment
To adjust the line density, select the "C Speed" on the left side of the screen and press the “ ”
or “ ” key.
Value
The current frame rate is displayed on the color bar. The frame rate / resolution value varies
depending on the probe, check category, check calculation, new patient, and frequency. If you change
any of the above, the setting value will return to the factory preset or user preset.
- 60 -
Color Ultrasonic Diagnostic Apparatus User Manual
Advantages
Low line density is helpful for fetal heartbeats, adult cardiac applications, and clinical radiology
applications requiring higher frame rates. High resolution is helpful in imaging small organs, such as
thyroid, testes. Impact on other controls line density changes vector density and frame rate.
Biological effects
Modifying the linear density changes the heat index and / or the mechanical index. Note about any
affected output display.
8.4.8 Steer
Description
The changes of angle can only be used in linear probe. You can zoom in and out to change the
color flow sampling frame angle to obtain a color flow linear image to get more information without
moving the probe.
Adjustment
Select the "Steer" option on the left side of the screen, to adjust the angle of the left / right slanted
image. Press “ ” or “ ” key, each time pressed, the frame angle changes once.
Value
The linear probe can be steered to the left (20 degrees), middle or right (20 degrees). If you change
the following, the steer value will return to the factory settings or user defaults: probe, check category,
check calculation, or new patient.
Advantages
It provides a Doppler sampling line angle, which is suitable for linear probe orientation. It is
helpful in carotid artery imaging in peripheral vascular applications.
Biological effects
Activating the steer will change the heat index and / or the mechanical index. Note about any
affected utput display.
Select "Spatial Filter" option on the left side of the screen, press “ ” or “ ” to change the
spatial filtering from 0-3 or 3-0.
- 61 -
Color Ultrasonic Diagnostic Apparatus User Manual
Value 0-3.
Advantages
Make the image smooth.
8.4.10 Ensemble
Description
Controls the number of samples of a single color flow vector.
Adjustment
To increase / decrease the size of the sample packetage, select the "Ensemble" on the left side of
the screen. Press “ ” or “ ” to change its value from 8-15 or 15-8. During the adjustment of
the encoder, the vale of FPS on the image at the right side of the information area changes, the
package value is small, FPS value is large, otherwise small.
Value
Sampling package size varies depending on the probe and application, if you change the probe,
check category, check calculation or new patient, the value will be return to factory settings or user
default.
Advantages
It allows you to increase the accuracy of color sensitivity and color averaging (increase the size of
the sample packet) or the frame rate (decrease the size of the sample packet) as needed. Impact on
other controls, decrease the size of the samplimg package size will increase the frame rate while
reducing image quality. Increase the size of the sampling package size will increase image quality
while reducing frame rate.
Biological effects
Changing the sampling package size changes the heat index and / or the mechanical index. Note
about any affected output display.
8.4.11 C Persist
Description
Use this technology can provide continuous flow information in real-time state. Real-time display
maximum flow rate, the performance of pulsatile blood flow. Persist in the 2D ultrasound imaging is
the frame (image) overlap, used in color Doppler flow imaging, low speed, low flow of blood, can
make it clear and easy to observe.
Adjustment
- 62 -
Color Ultrasonic Diagnostic Apparatus User Manual
Select "C Persist" on the left side of the screen, press “ ” or “ ” key to obtain C Persist
changes from 0-7 or 7-0. Corresponding values at the right side of the information display area
changes.
Adjustment
Press the "C" knob. The color flow window appears in the B mode image. Press the "Set" button,
move the trackball to move the size of the blood flow frame. Adjust the "Display Mode" on the left
side of the screen to the "T" state. To exit, press the "B" knob.
Advantages
Since PDI does not display speed, it does not appear aliasing. Impact on other controls, after
activating PDI, the following controls are adjusted: PRF is set to low value. Color Map is set to
energy diagram. Adjust Line Density. Wall Filter is set to Low. Spatial Filter to “0”. Adjust Sampling
Package Size. After getting a good B/W image, press “C" knob to enter the blood flow mode.
Note
When changing the PDI image, it may be necessary to adjust the gain setting.
- 63 -
Color Ultrasonic Diagnostic Apparatus User Manual
Ultrasonic transmit power, digital changes from 0-15, press “ ” key to decrease, press
different. Press “ ” key to decrease, press “ ” key to increase, and the real-time ultrasonic
image will change with the numerical value.
D Gain
In PW mode, adjust the spectral color brightness, spectral sound size. Press “ ” key to
decrease, press “ ” key to increase, in real-time ultrasonic spectrum image, the spectral sound size
changes with the value.
Display mode
In the real-time state, select "Synchro", you will activate "B" mode. Select "Refresh" mode, "B"
image freeze, the spectrum enter into the real-time mode of operation.
Steer
Linear probe frame angle adjustment, the value changes from fixed several values among -20-20,
press “ ” key to decrease, press “ ” key to increase. “Steer” at the right side of the screen
changes.
PRF
On real-time state, select “PRF”, press “ ” key to decrease, the frequency changes from
29.43khz-2khz. Press “ ” key to increase, the frequency changes from 2khz-29.43khz. Note:
different probe adjustment range is different. Note: different probe adjusted PRF range adjustment is
different.
- 64 -
Color Ultrasonic Diagnostic Apparatus User Manual
Volume
In real-time state, select “Volume”, adjust the Doppler volume, press “ ” key, the volume changes
from "255--0%", the Doppler volume decreases. Press “ ” key, the volume changes from
"0--255%", the Doppler volume increases. The audio figures on the right side of the image area
changes.
D Velocity
In real-time state, select “D Velocity”, press “ ” key, the scanning speed changes from 0-2.
Press “ ” key to change the scanning speed from 2-0. The "Sweep" English letter on the right side
of the information area changes.
Triplex
In real-time state, select ”Triplex”, press“ ” or “ ”key, you can achieve B mode, C
mode, D mode triplex synchronous work.
Enhance
Enhance the organ edge of the degree of brightness, the digital changes from 0-7, press “ ”
Wall Filter
In real-time state, select “Wall Filter”, press “ ” key, the number changes from 3-0, press
“ ” key, the number changes from 0-3. The “WFD” figure on the right side of the information
area changes.
Baseline
In real-time state, select “Baseline”, press “ ” key, the number changes from 6-0, press
“ ” key, the number changes from 0-6. The spectral curve moves correspondingly to the baseline.
- 65 -
Color Ultrasonic Diagnostic Apparatus User Manual
Wave Invert
In real-time state, select “Wave Invert”, “ ” or “ ”key, the letters change between OFF
- 66 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 67 -
Color Ultrasonic Diagnostic Apparatus User Manual
the left side of the screen and press “ ” or “ ” key to turn on the Triplex mode. The Doppler
spectrum is displayed along with the color flow and B mode images. The cursor changes to a Doppler
sampling line. You can now determine the position and size of the sampling volume gate to get a
certain speed. Use Doppler audio to determine whether the sampling volume gate is located in the
blood flow area. Use the Triplex mode to check blood flow information. To exit the PW Doppler mode,
press the “PW” knob, and then press the “B” knob.
- 68 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 69 -
Color Ultrasonic Diagnostic Apparatus User Manual
9.4.3 PRF
Description
Adjust the speed scale to accommodate faster / slower blood flow rates. The speed scale
determines the pulse repetition frequency. The pulse repetition frequency (PRF) is displayed on the
left side of the spectrum, expressed in units of frames per second.
Note
Make sure that only one gate is covered on the blood vessel at a time. Otherwise, signals from
multiple bloodstream regions will overlap.
Adjustment
To increase / decrease the PRF, select the "PRF" option, press “ ” or “ ” to change the
value from 2KHz-14.08KHz. By adjusting the image depth, PRF also changes. When the encoder is
adjusted, the system will update the speed scale parameter display synchronously.
Value
The speed scale value varies depending on the probe and application. In Triplex mode, if you
change the speed scale of the color flow, the speed scale of the Doppler mode will also be refreshed.
- 70 -
Color Ultrasonic Diagnostic Apparatus User Manual
9.4.5 Steer
Description
Under Doppler mode, steer function is only applicable to linear probe
Adjustment
To center the left / right slanted image, select “Steer” option and press “ ” or “ ” key. The
linear probe can be steered to the left (20 degrees) and right (20 degrees). If you change the following,
the steer value will return to the factory settings or user defaults: probe, check category, check
calculation, or new patient.
Advantages
Provides a Doppler sampling line angle, which is suitable for linear probe orientation. It is
helpful in carotid artery imaging in peripheral vascular applications.
Biological effects
Steer changes the heatl index and / or mechanical index. Note about any affected output display.
To increase / decrease the wall filter, select "Wall Filter" option, press “ ” or “ ” to
select 0-3 filter.
Value
Wall Filter values vary by probe and application. If you change the following, the wall filter
value will return to the factory settings or user defaults: probe, check category, check calculation, or
new patient.
Advantages
Remove unnecessary information. Remove the low level noise above and below the baseline so
that you can not see them on the spectrum.
- 71 -
Color Ultrasonic Diagnostic Apparatus User Manual
9.4.7 Baseline
Description
Adjust baseline to accommodate faster and slower blood flow to eliminate aliasing.
Adjustment
Baseline adjusts the point where the velocity trajectory is zero in the spectrum. The default
baseline is at the midpoint of the spectrum. To move the baseline, select “Baseline” and press“ ”
or “ ” key to adjust it. The baseline position varies with the baseline adjustment bar.
Note
The baseline shows a solid line passing through the spectrum. Baselines increase and decrease in
equal increments, depending on the current Doppler scale factor. If the maximum baseline offset
(either direction) is reached, the control will not work.
Value
50% is the center of the screen, + 95% is the top edge of the screen, and 5% is the bottom edge
of the screen. Baseline values vary depending on the probe and application. If you change the
following, the factory settings or user defaults are returned: probe, check category, check calculation,
or new patient.
Advantages
Eliminate aliasing. Raise the ruler range without changing the speed scale. Correct the positive
and negative speed limits without changing the entire speed range.
To invert the spectrum curve, select “Wave Invert” and press “ ” or “ ” key. After
inverting the spectrum, the positive sign (+) and the minus sign (-) on the speed scale will be inverted
accordingly. Positive speed is shown below the baseline.
Advantages
If you change the angle of the probe to adapt to the anatomical structure, the blood flow will still
move in the same direction, but the Doppler information is inverted. In this case, the Wave Invert
function is easier than changing the direction of the probe.
- 72 -
Color Ultrasonic Diagnostic Apparatus User Manual
9.4.10 Volume
Description
Under Doppler mode, the Doppler can be adjusted to the best image by adjusting the volume, get
a good spectrum of sound effects.
Adjustment
Select “Volume” and press “ ” or “ ” key to adjust the volume level. The corresponding
change value is displayed in the information area at the right side of the image.
9.4.11 D Velocity
Description
Refers to the horizontal movement of the M-mode image within a second distance, where it
refers to the time it takes an image to sweep from the left to the right. By adjusting the M mode
scanning speed, can be recorded in a fixed sampling line, the tissues and organs with time changes in
the longitudinal movement of the method.
Adjustment
Select "D Velocity" option, press “ ” or “ ” key, you can debug M mode scanning speed
of low speed, medium speed, high speed changes. When adjusting the scanning speed, there is a speed
change display in the information area at the right side of the image. At the same time, M speed scale
also changes.
- 73 -
Color Ultrasonic Diagnostic Apparatus User Manual
9.4.12 D Gain
Description
By changing the gain value, you can increase the Doppler mode to obtain remote effects. But the
gain should not be too large, too large will increase the image interference.
Adjustment
Select "D Gain" option, press “ ” or “ ” key, you can adjust the value of the gain, while
value changes at the right side of the image information area.
9.4.13 Frequency
Description
In Doppler mode, select “Frequency”. By selecting the frequency change, the speed change is
displayed on the color gray scale, and the D line scale speed also changes with the frequency to meet
the actual spectrum measurement needs.
Adjustment
Select "Frequency" option, press the “ ” or “ ” key, the frequency changes from
2.0MHz-5.0MHz, 6.0 MHz -12 MHz.
- 74 -
Color Ultrasonic Diagnostic Apparatus User Manual
In clinical diagnosis, the acquisition of ultrasound images in the measurement and calculation
results for the doctor is directly on the lesion analysis measures and results. The measurement
accuracy depends not only on the system accuracy, but also on the user's impact on the use of
appropriate medical programs. If necessary, be sure to indicate a medical plan related to measurement
and calculation. Be sure to refer to the original paper related to the program, as it explains the clinical
practice recommended by the researcher.
10.2 Overview
This chapter provides information on how to make measurements and describes the calculations
available in each mode.
Content includes
Check workflow Operation process
Measure component location Measurement description
General measurement list General information on how to make measurements
Mode Measurements: Step-by-step instructions for special measurements
- 75 -
Color Ultrasonic Diagnostic Apparatus User Manual
All operations must be carried out in a frozen state. B mode general measurement with the function:
distance, diameter narrow rate, angle, circumference area, area narrow rate, volume and so on.
- 76 -
Color Ultrasonic Diagnostic Apparatus User Manual
ultrasound area appears "+" cursor, enter into the angle measurement state;
2. Use the trackball to move the cursor to the O point and press【Set】 key to fix the point;
3. Use the trackball to move the cursor to point A and press 【Set】key to fix the point;
4. Use the trackball to move the cursor to point B and press 【Set】key to fix the point;
5. Use the trackball to move the cursor to point C and press 【Set】key to fix the point;
6. You will obtain the angle of OAB (OA line and OB line) and the angle of OAC (OA line and
OC line) ;
Repeat step 1-5 for next measurement.
Press 【Clear】 key to clear the measurement.
Press 【Freeze】 key to exit the measurement during the measurement.
- 77 -
Color Ultrasonic Diagnostic Apparatus User Manual
cursor to the "Volume" option with the trackball, press 【Set】 Key to move the cursor to one
end of the elliptical axis, press 【Set】 key to fix the starting point of the shaft;
2. Use the trackball to move the cursor to the other end of the elliptical axis and press 【Set】
key to fix the end of the axis;
3. Move the trackball to change the length of the other axis of the ellipse;
4. Use the trackball to move the cursor to the start point of the distance and press 【Set】key to
fix the point;
5. Use the trackball to move the cursor to the end of the distance and press 【Set】key to fix the
point;
Repeat step 1-5 for the next measurement.
Press 【Clear】 key to clear the measurement.
Press 【Freeze】 key to exit the measurement during the measurement.
All operations must be carried out in freeze state. M mode general measurement with the
function: heart rate, time, distance, valve speed selection. Choose blue button
to enter the following menu to operate.
- 78 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 79 -
Color Ultrasonic Diagnostic Apparatus User Manual
Clear measurement
【Clear】 With measurement needs
data
Start and End key of
【Set】 With measurement needs
measurement
Press this key directly
【Freeze】 to enter general Freeze image
measurement
M General Display M General See above
【M General】
measurement menu measurement menu diagram
D General Display D General
【D General】
measurement menu measurement menu
Move the cursor to the
【ET】 Time measurement desired measurement item
Heart Rate via the trackball and press
【HR】
measurement "SET" key. Move the cursor
to the measurement start
Distance point and press "SET" key to
【Distance】
measurement move the cursor to the
measurement end and press
"SET" key. At this point the
measurement data shows the
Valvular velocity bottom of the image. Press
【Vel】
measurement "Clear" key to clear the
measured value, press
"Freeze" key to exit this
measurement.
- 80 -
Color Ultrasonic Diagnostic Apparatus User Manual
Operation steps:
1. In D mode freeze state, use the trackball to move the cursor to the "Measure" option of the
measurement menu, press 【Set】 key;
2. Move the trackball to the "D General" option, press 【 Set 】 key, pop up the D General
Measurement sub-menu;
3. Use the trackball to move the cursor to "ET" option, press 【Set】 key to enter the measurement
state;
4. Use the trackball to move the cursor to the start point of the measurement, press 【Set】 key to
fix the start point of measurement;
5. Use the trackball to move the cursor to the end of the measurement and press 【Set】 key to fix
- 81 -
Color Ultrasonic Diagnostic Apparatus User Manual
10.4.11 HR Measurement
Measure the time interval between two heartbeat cycles on the ultrasound image in Doppler
mode and calculate the heart rate.
Operation steps:
1. In D mode freeze state, use the trackball to move the cursor to the "Measure" option of the
measurement menu, press 【Set】 key;
2. Move the trackball to the "D General" option, press 【 Set 】 key, pop up the D general
measurement sub-menu;
3. Use the trackball to move the cursor to the "HR" option, press 【 Set 】 key to enter the
measurement state;
4. Use the trackball to move the cursor to the start point of the measurement, press 【Set】 key to
fix the start point of measurement;
6. After measurement completed, the results are displayed at the bottom right of the ultrasound
zone. Repeat steps 2-5 to make the next measurement.
Press 【Clear】 key to clear the measurement data and symbol, press 【Freeze】 key to exit the
measurement.
- 82 -
Color Ultrasonic Diagnostic Apparatus User Manual
4. Use the trackball to move the cursor to the start point of the measurement, press 【Set】 key to
fix the start point of measurement;
6. After measurement completed, the results are displayed at the bottom right of the ultrasound
zone. Repeat steps 2-5 to make the next measurement.
Press 【Clear】 key to clear the measurement data and symbol, press 【Freeze】 key to exit the
measurement.
Operation steps:
- 83 -
Color Ultrasonic Diagnostic Apparatus User Manual
1. In D mode freeze state, use the trackball to move the cursor to the "Measure" option of the
measurement menu, press 【Set】 key;
2. Move the trackball to the "D General" option, press 【 Set 】 key, pop up the D normal
measurement sub-menu;
3. Use the trackball to move the cursor to the "Common" option, press 【Set】 key to enter the
measurement state;
4. Use the trackball to move the cursor to the measuring point, press 【 Set 】 key to fix the
measuring point PS;
5. Use the trackball to move the cursor to the measuring point, press 【 Set 】 key to fix the
measuring point MD;
6. Use the trackball to move the cursor to the measuring point, press 【 Set 】 key to fix the
measuring point ED;
7. After measurement completed, the results are displayed at the bottom right of the ultrasound
zone. Repeat steps 2-6 to make the next measurement.
Press 【Clear】 key to clear the measurement data and symbol, press 【Freeze】 key to exit the
measurement.
Operation steps:
1. In D mode freeze state, use the trackball to move the cursor to the "Measure" option of the
measurement menu, press 【Set】 key;
2. Move the trackball to the "D General" option, press 【 Set 】 key, pop up the D normal
measurement sub-menu;
3. Use the trackball to move the cursor to the "Trace" option, press 【 Set 】 key to enter the
measurement state;
- 84 -
Color Ultrasonic Diagnostic Apparatus User Manual
4. Use the trackball to move the cursor to the start point of the measurement, press 【Set】 key to
fix the start point of measurement;
6. After measurement completed, the results are displayed at the bottom right of the ultrasound
zone. Repeat steps 2-5 to make the next measurement.
Press 【Clear】 key to clear the measurement data and symbol, press 【Freeze】 key to exit the
measurement.
- 85 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 86 -
Color Ultrasonic Diagnostic Apparatus User Manual
Operation
Under B or B / B mode, press "DIS" key to enter the measurement menu.
1. Move the cursor to "Abdomen" or "Measure" and press "SET" key;
2. Move the cursor to "Liver" and press "SET" key;
3. Move the cursor, combined with "SET" key, do the measurement;
4. Repeat steps 2-3 can be measured multiple times; the measurement results are displayed at the
bottom right of the image.
- 88 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 89 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 90 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 91 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 92 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 93 -
Color Ultrasonic Diagnostic Apparatus User Manual
5. Repeat steps 3-4 to measure multiple times; the measurement results are displayed at the bottom
right of the image.
Ureter connected to the renal pelvis, the next bladder, is a pair of slender pipes, was flat
cylindrical, diameter average of 0.5 to 0.7 cm. Adult ureteral length of 25 to 35 cm, located in the
retroperitoneal, along the medial side of the psoas tendon straight down into the pelvis. The ureter has
three narrow parts: one at the transition of the renal pelvis and the ureter (at the beginning of the
ureter); one at the entrance to the small pelvis; the last one inside the bladder wall.
Operation
Under B or B / B mode, press "DIS" key to enter the measurement menu.
1. Move the cursor to "Abdomen" or "General" and press "SET" key;
- 94 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 95 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 96 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 97 -
Color Ultrasonic Diagnostic Apparatus User Manual
AFI Measurement:
Main Content
◆ Fixed database version selection
◆ Customize database edit
◆ Calculate GA and EDC
◆ Calculate FW
◆ Calculate GA and EDC by LMP
◆ Calculate GA and EDC by BBT
◆ Fetal physiology score
◆ Fetal growth curve
- 98 -
Color Ultrasonic Diagnostic Apparatus User Manual
Users can also measure AC/BPD, AC/FL or AC/FL/HC. The data will be automatically stored in
memory. When the obstetric report is called, the system will automatically calculate the result
according to the multi-parameter calculation.
Operation steps:
Move the trackball to select the basic information page of the obstetric report. Click the
drop-down button in the "LMP" input field, select the date, and then click on the input field after
pregnancy. The GA will be displayed in the input field.
- 99 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 100 -
Color Ultrasonic Diagnostic Apparatus User Manual
from severe asphyxia. Most of the newborns score among 7-10 points, the doctor will be based on the
child's score to be dealt with accordingly. Mild asphyxia neonates generally cleaned up the respiratory
tract, oxygen and other measures will soon improve, the prognosis is good.
Clinical significance: usually considered the level of fetal score and neonatal Apgra score, fetal
distress during childbirth, prenatal and perinatal mortality are highly correlated. The lower the score,
the higher the correlation between the two. It is reported that when all scoring items are normal, the
fetal physiological score is "10", the perinatal mortality rate is 0; the other hand, if all parameters are
abnormal, fetal physiological score of "0", the fetal mortality rate of 40%.
The system uses the Vintzileos scoring method to rate the following items:
Reactive fetal heart rate accelerate Fetal movement Fetal respiratory activity Fetal
muscle tension AFI
Operation steps:
Move the trackball, select the measurement page of the obstetric report, select page 3, and the
report menu shows the above functions, and the corresponding selection is made.
10.6.11 GS Measurement
Description
Gestational Sac is the form of the initial embryo of pregnancy. The second week of embryos
implanted in the endometrium, developed into the two germ layers, in the ventral bloody out of a large
cavity called the primary yolk sac, yolk sac in the process of embryonic development continue to
shrink and degenerate, but the egg yolk wall The outer mesoderm is the earliest birthplace of blood
cells, blood vessels and primordial germ cells - the fetus develops in the future.
Note
To calculate GS, three distance measurements can be made in two scanning planes. To display two
scan planes, press the L or R key. Get an image in each scan plane and press Freeze.
Operation
1. Select GS; an active caliper is displayed on the screen.
a. Move the cursor to the starting point and press SET key.
b. Move the cursor to the end point and press SET key.
- 101 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 102 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 103 -
Color Ultrasonic Diagnostic Apparatus User Manual
10.6.15 FL Measurement
Description
The femur is the largest long bone of the human body, referring to the femur, FL is the length of
the femur. FL is a measure of fetal body mass.
Operation
To measure single distance to measure FL:
1. Select FL; an active caliper is displayed on the screen.
a. Move the cursor to the starting point and press SET.
b. Move the cursor to the end point and press SET.
- 104 -
Color Ultrasonic Diagnostic Apparatus User Manual
10.6.16 HL Measurement
Description
Humer length (HL) is that doctors use B-ultrasound (or color Doppler ultrasound) to pregnant
women for pregnancy check, use it to observe the fetal development of the commonly used indicators.
The humerus is a typical long tubular bone in the human body and can be divided into two ends.
Humerus is located in the upper arm, also called the upper arm bone. The FL refers to the length of the
upper arm. Fetal humerus length = -5.4282 + 0.7542 × biparietal diameter. According to the study, the
fetal abdomen area, abdominal circumference and fetal humerus length is a more sensitive index of
fetal body mass, fetal humerus long in predicting the quality of the fetus is better than the femur
length
Operation
To measure single distance to measure the HL:
1. Select HL; an active caliper is displayed on the screen.
a. Move the cursor to the starting point and press SET.
b. Move the cursor to the end point and press SET.
c. The measured data are displayed at the bottom of the image.
2. To perform the second and third measurements, repeat the steps above.
- 105 -
Color Ultrasonic Diagnostic Apparatus User Manual
10.6.18 LV Measurement
Description
Spine also known as the spine, spine bone, by the shape of the special vertebrae and
intervertebral discs connected, located in the back of the middle, with the skull, the middle and the
ribs connected to the bottom and hip composition pelvis. There were 7 cervical spine, 12 thoracic
vertebrae, 5 lumbar vertebrae, 1 sacrum (synthesized by 5 sacral vertebrae) and 1 tail spine (composed
of 4 caudal vertebrae) 24 (adult) independent Of the vertebrae. Inside the vertebrae from top to bottom
to form a longitudinal ridge, there are spinal cord.
Operation
To measure single distance to measure the LV:
1. Select LV; an active caliper is displayed on the screen.
a. Move the cursor to the starting point and press SET.
b. Move the cursor to the end point and press SET.
c. The measured data are displayed at the bottom of the image.
2. To perform the second and third measurements, repeat the steps above.
10.6.20 AC Measurement
Description
Ultrasound measurement of fetal abdominal circumference can predict the risk of pregnancy
caused by the great children. Fetal abdominal measurement, mainly to see the fetal development of the
abdomen. To calculate the head circumference, you can make an ellipse measurement.
- 106 -
Color Ultrasonic Diagnostic Apparatus User Manual
Operation
Ellipse
1. Select AC; an active caliper will appear on the screen.
a. Move the cursor to the starting point and press SET.
b. Move the cursor to the end point and press SET.
c. Move the trackball to confirm the trace of AC
d. The measured data are displayed at the bottom of the image.
2. To perform the second and third measurements, repeat the steps above.
10.6.21 HC Measurement
Description
Head circumference refers to the maximum length of the circle around the fetal head. The head
of the fetus from the front to the back of the longest part, usually from the "forehead of the nose" to
"the back of the occipital bone carina," the longest distance, so the general head circumference is from
the "forehead of the nose" to "Back of the brain occipital protuberance" around the length of the
circle.
Operation
Ellipse
1. Select HC; an active caliper will appear on the screen.
a. Move the cursor to the starting point and press SET.
b. Move the cursor to the end point and press SET.
c. Move the trackball to confirm the trace of HC
d. The measured data are displayed at the bottom of the image.
2. To perform the second and third measurements, repeat the steps above.
- 107 -
Color Ultrasonic Diagnostic Apparatus User Manual
fetal urine excretion of the place. It can be said that amniotic fluid is one of the essential living
conditions in fetal growth.
Operation
To measure single distanceto measure AFI:
1. Select AFI; an active caliper is displayed on the screen.
a. Move the cursor to the starting point and press SET.
b. Move the cursor to the end point and press SET.
c. The measured data are displayed at the bottom of the image.
2. To perform the second and third measurements, repeat the steps above.
10.6.23 NT Measurement
Description
Neck transparent layer refers to the early pregnancy using ultrasound to observe the fetal neck
after subcutaneous water, which is in early pregnancy, especially in early pregnancy all the fetus can
appear an ultrasound phenomenon, its thickness and chromosomes and other lesions are closely
related, In other words, the thickness of the transparent layer of the neck can help identify other
chromosomal abnormalities and various structures, in addition to the risk of assessing the down
syndrome, in addition to the thickening of the neck, deformity and genetic disease. Because about
down syndrome, about 75% of the neck transparent layer will appear thicker.Pregnant to the
pregnancy, the transparent layer will usually fade, but in a small number of cases will become cervical
edema or water cysts. The significance of transparent layer thickness detection.
Operation
To measure single distance to measure the NT:
1. Select NT; an active caliper is displayed on the screen.
a. Move the cursor to the starting point and press SET.
b. Move the cursor to the end point and press SET.
c. The measured data are displayed at the bottom of the image.
2. To perform the second and third measurements, repeat the steps above.
- 108 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 109 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 110 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 111 -
Color Ultrasonic Diagnostic Apparatus User Manual
advantage of the young do a good job of Ovary health care and conservation, escort for their own, to
extend the freshness and vitality.
Note
In the B mode, the left / right ovary for ultrasound scanning positioning can be measured. The
Ovary volume can be calculated from the three values of length, widtch and height.
Operation
1. Under B or B / B mode, press the "DIS" key to enter the measurement menu. Move the cursor
to the "Pelvic" item in the "Application" of the left menu and press "SET";
2. Enter the measurement menu, move the cursor to the "Lt / Rt Ovary", press "SET";
3. Move the cursor to "Volume", press "SET";
4. Move the cursor, combined with the "SET" key, do the length, height, widtch and volume
measurements ("LHWV").
5. The measurement results are displayed at the bottom right of the image.
- 112 -
Color Ultrasonic Diagnostic Apparatus User Manual
cavity was inverted triangular, about 6cm deep, the top of the two horns for the "uterine horn", leading
to the fallopian tube. The lower end of the narrow "isthmus", about 1cm. The isthmus gradually
expanded during pregnancy, labor formation of the lower uterine segment. Uterus and cervix ratio
varies according to age, infancy is 1: 2, adolescence is 1: 1, growth period is 2: 1. The uterus is
slightly curved forward, the front wall prone to the bladder, and the vagina almost at right angles, the
location can be varying with the degree of bladder rectification and change. Uterine wall from the
outside to the serous, myometrial and mucous membrane (ie, endometrial) three.
- 113 -
Color Ultrasonic Diagnostic Apparatus User Manual
and swelling of the symptoms. If the occurrence of torsion or other circumstances can cause pain,
multiple common uterine fibroids are common.
Operation
Under B or B / B mode, press the "DIS" key to enter the measurement menu.
1. Move the cursor to the "Pelvic" item in the "Application" of the left menu and press "SET";
2. Move the cursor to the "Uterus", press "SET";
3. Move the cursor to "Fibroid Volume", press "SET";
4. Move the cursor, combined with "SET" button, do the uterine fibroids measured.
The measurement results are displayed at the bottom right of the image.
- 114 -
Color Ultrasonic Diagnostic Apparatus User Manual
10.8.1 Description
Skeletal muscle, also known as striated muscle, a muscle. Human body has about 600 pieces of
skeletal muscle.
1. Bone marrow is a contraction of the muscle cells (due to its shape of long fibrous growth, so
also known as muscle fibers) formed. Any physical activity and physical activity, are the completion
of skeletal muscle contraction, a direct impact on the body's strength and endurance.
2. Skeletal muscle is also the main storage site for human sugar, commitment to more than 70% of
sugar storage, the body's blood sugar balance has an extremely important buffer effect.
3. Skeletal muscle can be free exercise, the muscles have a certain flexibility, is elongated, when
the lift force, can automatically return to the original level, this elasticity can slow down the impact of
external forces on the human body.
- 115 -
Color Ultrasonic Diagnostic Apparatus User Manual
Clear measurement
【Clear】 With measurement needs
data
Measurement start
【Set】 With measurement needs
and end point
Distance Under B or B / B mode, adjust the image, press
Distance
measurement "DIS" button, use the trackball to move the
cursor to the needed point to measure, press
Area Area measurement "SET". Move the cursor to the measurement
start point and press "SET", move the cursor to
the measurement end and press "SET". The
Hip angle measurement data shows the at the bottom of
Hip Angle the image. Press "Clear" key to clear the
measurement
measured data, press "Freeze" to exit this
measurement.
10.8.3 Distance Measurement
Operation
Under B or B / B mode, press the "DIS" key to enter the measurement menu.
1. Move the cursor to "Application" and press "SET";
2. Move the cursor to "MSK" and press "SET";
3. Move the cursor to "Distance" and press "SET";
4. Move the cursor, combined with "SET" button, do the skeletal muscle distance measurement.
5. The measurement results are displayed at the bottom right of the image.
10.8.4 Area Measurement
Operation
Under B or B / B mode, press the "DIS" key to enter the measurement menu.
1. Move the cursor to "Application" and press "SET";
2. Move the cursor to "MSK" and press "SET";
3. Move the cursor to "Area" and press "SET";
4. Move the cursor, combined with "SET" key, do the skeletal muscle area measurement.
5. The measurement results are displayed at the bottom right of the image.
10.8.5 Hip Angle Measurement
Operation
Under B or B / B mode, press the "DIS" key to enter the measurement menu.
1. Move the cursor to "Application" and press "SET";
2. Move the cursor to "MSK" and press "SET";
3. Move the cursor to "Hip Angle" and press "SET";
4. Move the cursor, combined with "SET" key, do the hip angle measurement.
5. The measurement results are displayed at the bottom right of the image.
- 116 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 117 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 118 -
Color Ultrasonic Diagnostic Apparatus User Manual
3. Move the cursor to "Rt Thyroid" or "Lt Thyroid" and press "SET";
4. Move the cursor to "Volume" and press "SET";
5. Move the cursor and combine the "SET" key to do the measurement.
6. The measurement results are displayed at the bottom of the image.
- 119 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 120 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 121 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 122 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 123 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 124 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 125 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 126 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 127 -
Color Ultrasonic Diagnostic Apparatus User Manual
Distance
Distance
measurement
Area measurement Area measurement
Stenosis diameter
Stenosis D
measurement
Stenosis area
Stenosis A
measurement
Diameter Diameter
measurement measurement
Venous distance
Venous distance
measurement
- 128 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 129 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 130 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 131 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 132 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 133 -
Color Ultrasonic Diagnostic Apparatus User Manual
with the atria, the wall thickness of some of the heart, because the heart of the blood flow only with its
ventricle; left ventricular wall thicker than the right ventricle. The left ventricle blood to the body, so
the need for greater contractility, the wall will be thicker. The right ventricle to the blood delivery to
its very close to the lungs, the need for some smaller shrinkage, the wall is thinner. Normal, the right
ventricular wall thickness of only 1/3 of the left ventricular wall.
Operation
Under B, B/B, PW, CF mode, adjust the image, freeze, press "DIS" key to enter the measurement
menu:
1. Move the cursor to "Application" and press "SET";
2. Move the cursor to "Cardiac" and press "SET" to enter the cardiac measurement menu;
3. Move the cursor to ”RVWd” and press "SET";
4. Move the cursor to the measurement start point and press "SET";
5. Move the cursor to the end point of the measurement and press "SET";
6. Repeat steps 3-5 to do several measurements;
7. The measurement results are displayed at the bottom right of the image.
- 134 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 135 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 136 -
Color Ultrasonic Diagnostic Apparatus User Manual
ventricular wall inward retraction, diastolic outward expansion. Dilated cardiomyopathy patients,
ventricular wall thinning, if part of the relatively thin, there will be reverse movement, that is, this part
of the heart wall contraction, outward protruding, while the diastolic recovery. "Ventricular septum
and left ventricular posterior wall was reverse movement" that the ventricular septum and left
ventricular posterior wall is relatively thin, its function has been weakened.
Operation
Under B, B/B, M, PW, CF mode, adjust the image, freeze, press "DIS" key to enter the
measurement menu:
1. Move the cursor to "Application" and press "SET";
2. Move the cursor to "Cardiac" and press "SET" to enter the cardiac measurement menu;
3. Move the cursor to ”LVDs” and press "SET";
4. Move the cursor to the measurement start point and press "SET";
5. Move the cursor to the end point of the measurement and press "SET";
6. Repeat steps 3-5 to do several measurements;
7. The measurement results are displayed at the bottom right of the image.
- 137 -
Color Ultrasonic Diagnostic Apparatus User Manual
preload.
- 138 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 139 -
Color Ultrasonic Diagnostic Apparatus User Manual
5. Move the cursor to the end point of the measurement and press "SET";
6. Repeat steps 3-5 to do several measurements;
7. The measurement results are displayed at the bottom right of the image.
- 140 -
Color Ultrasonic Diagnostic Apparatus User Manual
4. Move the cursor to the measurement start point and press "SET";
5. Move the cursor to the end point of the measurement and press "SET";
6. Repeat steps 3-5 to do several measurements;
7. The measurement results are displayed at the bottom right of the image.
10.12.23 LA Measurement
Operation
Under B, B/B, M, PW, CF mode, adjust the image, freeze, press "DIS" key to enter the
measurement menu:
1. Move the cursor to "Application" and press "SET";
2. Move the cursor to "Cardiac" and press "SET" to enter the cardiac measurement menu;
3. Move the cursor to ”LA” and press "SET";
4. Move the cursor to the measurement start point and press "SET";
5. Move the cursor to the end point of the measurement and press "SET";
6. Repeat steps 3-5 to do several measurements;
7. The measurement results are displayed at the bottom right of the image.
- 141 -
Color Ultrasonic Diagnostic Apparatus User Manual
4. Move the cursor to the measurement start point and press "SET";
5. Move the cursor to the end point of the measurement and press "SET";
6. Repeat steps 3-5 to do several measurements;
7. The measurement results are displayed at the bottom right of the image.
- 142 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 143 -
Color Ultrasonic Diagnostic Apparatus User Manual
2. Move the cursor to "Cardiac" and press "SET" to enter the cardiac measurement menu;
3. Move the cursor to ”RVOT Diam” and press "SET";
4. Move the cursor to the measurement start point and press "SET";
5. Move the cursor to the end point of the measurement and press "SET";
6. Repeat steps 3-5 to do several measurements;
7. The measurement results are displayed at the bottom right of the image.
- 144 -
Color Ultrasonic Diagnostic Apparatus User Manual
Press the " " key on the keyboard, there will be ROI sampling box, to enter into the 3D/4D
mode of operation.
It is recommended to adjust the depth to 170mm or less, the sampling frame angle transferred to
moderate, to achieve the best 2D image results.
Use the trackball with " " key to adjust the size of the sampling frame, move the trackball
to adjust the position.
- 145 -
Color Ultrasonic Diagnostic Apparatus User Manual
After selecting the area of interest, press the " " freeze button on the keyboard to enter the
3D / 4D image display interface.
Note: Select the area (regional size should be moderate, too large or too small will prompt to re-select).
If the region of interest is not selected, the region of interest is the default selection. Click the "Set"
key to adjust the size and position of the frame.
Press the " " key twice to enter the 2D mode of operation.
- 146 -
Color Ultrasonic Diagnostic Apparatus User Manual
corresponding to the left and right arrows, adjust the 4D images of different profiles.
Y axis
X axis B profile
A profile
Z axis
C profile 4D display
Move the trackball to the A profile, move the cursor to the desired position, hold down the
"SET" key, the cursor changes from white to yellow, move the trackball, and the A profile to the
rotation on the Z axis.
Move the trackball to the B profile, move the cursor to the desired position, hold down the
"SET" key, the cursor changes from white to yellow, move the trackball, and the B profile to the
rotation on the X axis.
Move the trackball to the C profile, move the cursor to the desired position, hold down the
"SET" key, the cursor changes from white to yellow, move the trackball, and the C profile to the
- 147 -
Color Ultrasonic Diagnostic Apparatus User Manual
Rotate around the X axis to move the Rotate around the Y axis to move the
trackball to the B profile and rotate trackball to the C profile and rotate
Rotate around the Z axis to move the trackball to A profile and rotate
- 148 -
Color Ultrasonic Diagnostic Apparatus User Manual
Function Menu
Press the "3D / 4D" key to enter the 3D/4D preparation mode. After adjusting the sampling frame,
press " " to start the 3D / 4D data acquisition and enter the 3D / 4D operation interface. The left
side of the interface for the 3D/4D function menu, use the trackball to move the cursor and with
"SET" button to select the operation.
4 diaplay mode
Default value
2D mode color
3D mode color
Profile movement
selection
Profile rotate
selection
Image optimization
Auto cine
Image storage
Video storage
Exit measure
- 149 -
Color Ultrasonic Diagnostic Apparatus User Manual
Use the trackball to move the cursor to the cursor of the A profile, hold down the "SET" key
while moving the trackball, the green, blue dot of the other three profiles also follow the trackball
movement, any points on the other profiles and the 3D/4D render image display the image effect
along the movement trajectory.
- 150 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 151 -
Color Ultrasonic Diagnostic Apparatus User Manual
Real-time 3D mode allows cine loop. Press the " " key to freeze the image, move the cursor
to the lower left of the interface "Auto Cine", 4 profiles at the same time into the cine loop. You can
also use the trackball with the SET button to stop the cine loop operation.
Exit measure
- 152 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 153 -
Color Ultrasonic Diagnostic Apparatus User Manual
Save as path
- 154 -
Color Ultrasonic Diagnostic Apparatus User Manual
the diagnosis, place it in the probe box and press "FREEZE" key to make the probe in the "freeze"
state.
12.3.2 The probe will be broken even was slightly impacted. Be especially careful not to be hit or hit
by hard objects.
12.3.3 Probe head (sound window), should be careful protected and avoid scratch.
12.3.4 Non-conductive "medical ultrasound gel" should be used in the inspection, and no conductive
or corrosive gel shall be used.
- 155 -
Color Ultrasonic Diagnostic Apparatus User Manual
12.3.5Cleaning, sterilizing and disinfecting the probe: After each diagnosis, it should be disinfected
after each use. 5% sodium hypochlorite can be diluted with water 100 times to produce effective
chlorine content of 500mg / L dilution, with the dilution of the probe or spray the probe surface for 10
minutes.
12.3.6The probe is an ordinary anti-soaking part, prohibit dipping any conductive liquid, so as not to
corrosion probe; probe immersed in water position shall not exceed 5mm of the sound window.
12.3.7 Once the probe and the host are connected, do not disassemble it, to avoid probe plug and
socket poor contact.
Note:
1. A long-term repeated gel contact with the probe may damage the probe.
2. Clean the probe after each use.
3. Do not use a surgical brush to clean the probe, even if you use a soft brush may damage
the probe, use soft cloth only.
4. Probe and the patient's contact time should not be too long, to avoid patient discomfort.
To extend the life of the probe and get the best performance, do the following:
1. Periodically check the probe cable, socket and sound window.
2. Swicth off the machine before connecting or removing the probe.
3. Do not drop the probe to the floor or hard objects, it is strictly forbidden to hit the probe sound
window, or the probe is easily to damage.
4. Do not heat the probe.
5. Do not bend or pull the probe cable, otherwise it may cause the internal cable fracture.
6. Use only the gel on the head of the probe and clean the probe after each use.
7. After cleaning the probe, you must carefully check the probe's sound window, shell and cable, do
not use the probe if you find any cracks or damage on the probe.
- 156 -
Color Ultrasonic Diagnostic Apparatus User Manual
connect the battery to the charger (AC adapter) when not charging. The battery is connected to the
charger (AC adapter) for no more than 10 hours, otherwise it may shorten battery life. If the battery is
not used for a long time, the battery will discharge itself over time.
5. To prevent battery failure, the battery should be charged one time for each 3 months.
6. Extreme temperature (subcooling or overheating) The environment will affect the battery charging
effect. Do not charge the battery in fire or extremely hot conditions! Do not use or store batteries near
heat sources (such as fire or heaters)! If the battery leaks or smells, it should be removed immediately
from the open flame.
7. Do not continue using damaged batteries and chargers (AC adapter).
8. Do not try to disassemble the battery.
9. Do not short the battery.
10. Do not throw the battery in the fire or heat the battery, otherwise it will cause an explosion.
11. Do not put the battery into the water or wet it.
12. Do not connect the battery directly to a wall outlet or on-board cigarette lighter outlet.
13. Do not use the wire or other metal objects to shorten the battery positive and negative, prohibit the
battery with the necklace, hairpin or other metal objects with the transport or storage.
14. Do not use a nail or other sharp objects to pierce the battery shell, prohibit hammering or pedaling
the battery.
15. Do not hit, throw or use the battery to be mechanically shocked.
16. It is forbidden to disassemble the battery in any way.
17. Prohibited with a battery (such as dry batteries) or different capacity, model, variety of batteries
used in combination.
18. Stop using it if the battery is exposed to odor, heat, deformation, discoloration, or any other
abnormality. If it is being used or charged, please remove it from the appliance or charger
immediately.
19. Dispose of used batteries in accordance with local regulations on proper disposal of waste.
- 157 -
Color Ultrasonic Diagnostic Apparatus User Manual
13.2 Storage
Packaging should be stored in the room, the ambient temperature -22 ℃ ~ +55 ℃, relative humidity
less than 80%, indoor should avoid strong sunshine, no cause corrosion of the instrument gas, indoor
ventilation should be good. If the storage period is more than 6 months, you should take out of the
apparatus from the package, in the normal working environment conditions, power on 4h and then
re-inspection and pack the apparatus into package, store it in the required room.
Fragile
Temperature limit
Upper
Stacking limit
Afraid of rain
- 158 -
Color Ultrasonic Diagnostic Apparatus User Manual
In the process of using the apparatus, if you encounter difficulties, and have no way to solve,
please consult the company's professional technical personnel or the company after-sales service
department. The following is a list of common troubleshooting methods, for reference only.
Fault Detection
a) Check whether the external power supply is normal, the host power cord is connected, and has
been plugged into the power outlet; host adapter indicator light is on, the adapter output plug is
normal, the adapter plug and the host is connected.
b) Check the probe and the host connection is correct;
c) Check the connection of the monitor is correct and secure.
Troubleshooting
a) Turn on the apparatus power switch, the panel light is off:
Check the power supply is normal; check the power cord and plug connection is reliable; power
switch is bad.
b) No signal display:
Check whether the monitor power switch is on, adjust the apparatus brightness and contrast knob;
check the monitor signal input line and the host connection is reliable.
c) The image on the display is not clear:
Adjust the gain, near field, far field button; adjust the brightness, contrast keys;
d) The monitor appears intermittent stripes interference, snow-like interference:
Check whether the power supply is normal, whether the interference by other equipment; equipment
around the space whether there is electricity, magnetic interference; instrument power and probe plug,
socket contact is good, ground column is full and earth connection.
- 159 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 160 -
Color Ultrasonic Diagnostic Apparatus User Manual
15.15 The apparatus is based on PC platform, please do not arbitrarily insert flash disk, and to ensure
that the device flash disk without external program.
15.16 If the apparatus fails, please immediately shut down the power and contact the supplier or its
authorized agents.
15.17 If you need to connect other companies' electronic or mechanical devices, please contact the
supplier or its authorized agents before connecting.
- 161 -
Color Ultrasonic Diagnostic Apparatus User Manual
- 162 -