You are on page 1of 4

Internal

Laboratory
Scope
Laboratory Scope TP F-LS 01

Issue: 02 Issue date: 4 August 2016 Rev. 00 Rev. date :

Parameter/ Measured Quantity/ Remarks/ Standard Equipment &


S. No. Range
Device Under Calibration * Method Used

Caliper
(Vernier / Digital / Dial) Using Slip Gauge & Slip Gauge
1 L.C. 0.01 mm 0-300 mm Accessories,
L.C. 0.02 mm

Depth Caliper (Vernier/Dial/Digital) Using Slip Gauge Accessories,


2 0-300 mm
L.C 0.01mm , Surface plate

External Micrometer
3 L.C. 0.001 mm 0-100 mm By Using Gauge Blocks
L.C. 0.01 mm

Plunger type Dial Indicator


( Analogue / Digital)
4 0-10 mm By Using Dial Calibration Tester
L.C. 0.001 mm
L.C. 0.01 mm

Lever type Dial Indicator


( Analogue / Digital)
5 0-2 mm By Using Dial Calibration Tester
L.C. 0.001 mm
L.C. 0.01 mm

Dial Bore Gauge


6 0-1 mm By Using Dial Calibration Tester
L.C. 0.001 mm

Comparator Stand
7 300*300 mm By Using Flatness Gauge with 1µm
Granite/Cast Iron

Scope

MECHANICAL CALIBRATION:
Laboratory Scope TP F-LS 01

Issue: 02 Issue date: 4 August 2016 Rev. 00 Rev. date :

S.No. Reference Master Used For Calibration

1 Slip Gauge Set

3 Flatness Gauge

4 1µm Dial Indicator

2 Electronic Dial Calibration Tester

5 Comparator Stand

LIST OF INSTRUMENTS USED IN LABORATORY:

LIST OF APPLICABLE STANDARDS:

S. Standard
Reference Standard Used For Calibration
No. Number Rev No.
General requirements for the competence of
1 IS.17025:2005 NOVEMBER 2005
testing and calibration laboratories

2 IS.2092.1983 Plunger Dial Gauge JANUARY 2006

IS.11498.1985 Lever Dial Gauge NOVEMBER 2005


3 IS.2967.1983 External Micrometre MARCH 1998

4 IS.3651.1.1982 Vernier Calliper L.C 0.1 & 0.05 MAY 2007

5 IS.3651.2.1985 Vernier Calliper L.C 0.02 NOVEMBER 1985

6 IS.4213.1991 Vernier Depth Gauge 300mm AUGUST 2004

7 IS.2285.2003 Cast Iron Surface Plate AUGUST 2003

LIST OF WORK INSTRUCTION USED IN LABORATORY:


S. no Work Instruction Reference WI Used For Calibration Rev no
Number.
Laboratory Scope TP F-LS 01

Issue: 02 Issue date: 4 August 2016 Rev. 00 Rev. date :

1 WI-18 Dial Gauge Calibration 01

2 WI-19 Comparator Gauge Calibration 01

3 WI-20 Vernier Caliper Calibration 01

4 WI-21 Micrometer Calibration 01

5 WI-24 Depth Vernier Gauge Calibration 01

6 WI-25 Length Rod Calibration 01

7 WI-27 Height Gauge Calibration 01

Note:
1. Calibration is done inhouse with the help of Reference Masters calibrated
through external NABL certified Laboratory.
2. For Inhouse Calibration TP has appointed external agency Unique
Instruments from which trained person do the calibration activity in
house.