Laboratory G.B.
Central Laboratory, Trichy
Accreditation Standard ISO/IEC 17025:2005
Field Mechanical Calibration Issue Date 11.12.2010
Certificate Number C-0520 Valid Until 10.12.2012
Last Amended on - Page 1 of 3
Measured Qty / Range *Calibration Remarks
Instrument Measurement
Capability ()
AT LABORATORY
1. VERNIER CALIPER
L.C : 0.01 mm 0 to 300mm 15 μm Using Caliper Checker
L.C : 0.02 mm 0 to 300mm 17 μm
0 to 600mm 20 μm
2. VERNIER HEIGHT GAUGE
L.C : 0.02 mm 0 to 300mm 18 μm Using Caliper Checker
0 to 600mm 21 μm
3. VERNIER DEPTH GAUGE
L.C : 0.02 mm 0 to 300mm 18 μm Using Caliper Checker
4. EXTERNAL MICROMETER 0 to 25mm 2.9 μm Using Slip Gauge Set
L.C : 0.001 mm ‘O’ Grade
L.C : 0.01mm 0 to 25mm 6.4 μm Using Slip Gauge Set
‘O’ Grade
25 mm to 300mm 9.0 μm Using Long Slip Gauge
Set ‘O’ Grade
5. INTERNAL MICROMETER 50 mm to 300 mm 11.0 μm Using Long Slip Gauge
L.C : 0.01mm Set & Accessories Set
‘O’ Grade
Laboratory G.B. Central Laboratory, Trichy
Accreditation Standard ISO/IEC 17025:2005
Field Mechanical Calibration Issue Date 11.12.2010
Certificate Number C-0520 Valid Until 10.12.2012
Last Amended on - Page 2 of 3
Measured Qty / Range *Calibration Remarks
Instrument Measurement
Capability ()
6. STEEL TAPE 1 meter 582 μm x √L Using Digital Steel Tape /
L.C : 1mm 3 meter L=Length of tape in meter Scale Calibrator
15 meter
30 meter
50 meter
7. FEELER GAUGE Upto 1 mm 3.08 μm Using Digital Micrometer
8. STEEL RULE
L.C : 0.5 mm 1000 mm 3.0 μm Using Digital Steel Tape /
L.C : 1 mm 1000 mm 581 μm Scale Calibrator
9. BORE DIAL GAUGE
L.C : 0.01 mm 0 to 1.2 mm 7.0 μm Using Dial Calibration
(Transmission Only) Tester
10. PLUNGER DIAL GAUGE
L.C : 0.001mm 0 to 12.5 mm 3.5 μm Using Dial Calibration
Tester
L.C : 0.01mm 0 to 10 mm 7.0 μm Using Dial Calibration
Tester
11. LEVER DIAL GAUGE
L.C : 0.01mm 0 to 1.0 mm 7.0 μm Using Dial Calibration
Tester
12. COATING THICKNESS FOILS 9 μm to 1000 μm 2.9 μm Using Digital Micrometer
Laboratory G.B. Central Laboratory, Trichy
Accreditation Standard ISO/IEC 17025:2005
Field Mechanical Calibration Issue Date 11.12.2010
Certificate Number C-0520 Valid Until 10.12.2012
Last Amended on - Page 3 of 3
Measured Qty / Range *Calibration Remarks
Instrument Measurement
Capability ()
13. REFERENCE BLOCK FOR 1 mm to 10 mm 2.9 μm Using Digital Micrometer
ULTRASONIC THICKNESS
GAUGE
14. PRESSURE CALIBRATION
DIAL / DIGITAL PRESSURE 0 to 40 kg/cm² 0.41% rdg. Using Digital Pressure
GAUGE 0 to 300 kg/cm² 0.15% rdg. Calibrator
L.C : 0.1 kg/cm² 0 to 700 kg/cm² 0.09% rdg.
15. PRESSURE TRANSMITTER
WITH INDICATOR 0 to 600 kg/cm² 0.17% rdg. Using Digital Pressure
L.C : 1 kg/cm² Calibrator
-X-X-X-X-X-X-X-X-X-X-X-X-X-X-X-X-