Specifications
2. Line :
(1) Voltage : AC 220 ~ 240
(2) Phase : Single
(3) Power Capacity : 7.5 kVA
6. Maximum deviation :
(1) kV : +/- 5%
(2) mAs : +/- 10%
8. X-ray tube :
(1) Name : OX/110-15
(2) Focal spot : 1.8mm
(3) Target angle : 15 degrees
(4) Heat dissipation : 350HU / sec
9. Filtration :
(1) X-ray tube : At least 0.5 mm Al equivalent at 50 kV
(2) Total filtration : 2.5 mm Al – eq. @80kV
(1.0 mm Aluminum Plates)
10. Reference axis : is paralleled to SID scale(page23,13) and (center of the X-ray beam)
centerofthecrosslineofthecollimatorlightfield
DK Medical System
ELMO-T3
PRODUCT DATA
11. Exposure switch : Two stage dead man type
12. Collimator :
(1) Name : BC-
(2) Total filtration : 2.5mm Al – eq. @80kV
(3) Min. X-ray field size : ≦ 5cm x 5cm @ 1m SID
(4) Max. X-ray field size : 40cm x 40cm @ 75cm SID
(5) Selectable X-ray field size :
SID Selectable X-ray field size
Minimum Maximum
75cm 3.75cm x 3.75cm 40cm x 40cm
100cm 5cm x 5cm 53.8cm x 53.8cm
(6) X-ray field adjustment : Proportional
(7) Illumination : ≧ 160lux @ 1m SID
(8) Lamp : 24V / 150W Halogen
(9) Timer : 30 sec.
(10) SID measure : 2.0m Accuracy ±5%
(11) Field size indicator : Dial Accuracy ±2%
17. Operation :
(1) Temperature : 10°C - 40°C
(2) Relative humidity : 30% - 75%
(3) Atmospheric pressure : 700 hPa – 1060 hPa