1. Output power : 5.1kW

2. Line :
(1) Voltage : AC 220 ~ 240
(2) Phase : Single
(3) Power Capacity : 7.5 kVA

3. kV regulation:(INPUT : 220V 50 Hz) 40 – 110kV


KV mAs mA KV mAs mA
40KV~50KV 1.0~10mAs 100 71KV~80KV 1.0~10mAs 60
12.5~20mAs 85 12.5~20mAs 50
25~50mAs 70 25~50mAs 40
63~100mAs 60 63~100mAs 30
125~200mAs 40 81KV~90KV 1.0~10mAs 50
51KV~60KV 1.0~10mAs 85 12.5~20mAs 40
12.5~20mAs 70 25~50mAs 30
25~50mAs 60 63~100mAs 20
63~100mAs 50 91KV~100KV 1.0~10mAs 50
61KV~70KV 1.0~10mAs 70 12.5~20mAs 40
12.5~20mAs 60 25~50mAs 30
25~50mAs 50 63~100mAs 20
63~100mAs 40 101KV~110KV 1.0~10mAs 40
12.5~20mAs 30
25~50mAs 20
63~80mAs 15
4. mAs regulation : 1.0, 1.25, 1.6, 2.0, 2.5, 3.2, 4.0, 5.0, 6.3, 8.0, 10, 12.5, 16, 20,
25, 32, 40, 50, 63, 80, 100, 125, 160, 200mAs

5. Maximum Out Power : 5.1kW @60kV/85mA Input: single-phase AC 220V

6. Maximum deviation :
(1) kV : +/- 5%
(2) mAs : +/- 10%

7. Compensation of the supply mains : Automatic

8. X-ray tube :
(1) Name : OX/110-15
(2) Focal spot : 1.8mm
(3) Target angle : 15 degrees
(4) Heat dissipation : 350HU / sec

9. Filtration :
(1) X-ray tube : At least 0.5 mm Al equivalent at 50 kV
(2) Total filtration : 2.5 mm Al – eq. @80kV
(1.0 mm Aluminum Plates)

10. Reference axis : is paralleled to SID scale(page23,13) and (center of the X-ray beam)
centerofthecrosslineofthecollimatorlightfield
DK Medical System
ELMO-T3
PRODUCT DATA
11. Exposure switch : Two stage dead man type

12. Collimator :
(1) Name : BC-
(2) Total filtration : 2.5mm Al – eq. @80kV
(3) Min. X-ray field size : ≦ 5cm x 5cm @ 1m SID
(4) Max. X-ray field size : 40cm x 40cm @ 75cm SID
(5) Selectable X-ray field size :
SID Selectable X-ray field size
Minimum Maximum
75cm 3.75cm x 3.75cm 40cm x 40cm
100cm 5cm x 5cm 53.8cm x 53.8cm
(6) X-ray field adjustment : Proportional
(7) Illumination : ≧ 160lux @ 1m SID
(8) Lamp : 24V / 150W Halogen
(9) Timer : 30 sec.
(10) SID measure : 2.0m Accuracy ±5%
(11) Field size indicator : Dial Accuracy ±2%

13. Dimensions : 497 (W) x 803 (D) x 1460 (H) mm

14. Weight : 150 kg

15. Tube stand movement:


(1) Vertical : 358mm-2086mm

16. Environmental conditions:

17. Operation :
(1) Temperature : 10°C - 40°C
(2) Relative humidity : 30% - 75%
(3) Atmospheric pressure : 700 hPa – 1060 hPa

18. Transport and storage condition :


(1) Temperature : -40°C ~ 70°C
(2) Relative humidity : 10 ~ 90%, including condensation
(3) Atmospheric pressure : 500 hPa – 1060 hPa