Professional Documents
Culture Documents
REV: M
TITLE: PB560 FINAL TEST PROCEDURE
INSTRUCTIONS Page 1 of 12
PREPARED APPROVED
REV ECO DATE SUMMARY
BY BY
N/A N/A N/A Refer to 10039030 rev G for
N/A N/A
earlier revision history
Added instruction to
complete final test
equipment verification log
P/N 10039304. Move
instruction on alarm test daily
check, sound level meter
L EC075186 Refer to Agile Refer to Agile
(PN 10039030, pg.9) to front
of procedure. Move O2 valve
leak test (pg.3) and O2 valve
functional test box (pg.8) to
front (pg.3 & 7 of P/N
10036007).
Updated following G-
M EC102789 Refer to Agile Refer to Agile
QAP056 FY18 review
P/N: 10039030
REV: M
TITLE: PB560 FINAL TEST PROCEDURE
INSTRUCTIONS Page 2 of 12
TOOLING REQUIRED
Pentium Based PC with the OpenVent software P/N 10060473 loaded
Dual-limb Adult Patient Circuit with Exhalation Valve, DAR P/N 5094000 or approved equivalent
Dual-limb Pediatric Patient Circuit with Exhalation Valve, DAR P/N 5093900 or approved equivalent
Patient Hose, 22mm, 40cm length, P/N G-061440-00
, Pneuflo Resistor, Rp50, P/N 13394-03
, Pneuflo Resistor, Rp5, P/N 13394-05
Test Lung, 50 Litre bottle, See Doc Ref [8]
Test Lung, 3 Litre bottle, See Doc Ref [8]
PTS 2000 Performance Test Instrument with RS232 serial comm. cable
Sprint LC Multi Air Tester with 8mm tubing and PB540/PB560 O2 Inlet connector
FiO2 Sensor Kit, Airox P/N 3814199
Sound Level Calibrator Type 4231
TEST NOTES
1. If the ventilator fails any test step, describe the parameter out of spec in the Discrepancy Log of the DHR
and notify line technician or Manufacturing Engineer before switching off. Refer to procedure TBD.
2. The Alarm Silence Reset button may be used to suppress alarms as needed.
3. If at any time you notice an abnormality of any kind then notify the line Engineer/Technician for
clarification before continuing.
4. For each test section on the DHR, indicate Pass or Fail by signing initials in the appropriate Pass or Fail
column on the DHR.
5. Leave the PTS2000 switched ON. If the PTS2000 power switch is turned off and on, the pressure
transducers are automatically zeroed. Do not apply pressure to low or high pressure ports during power-
up. Wait approximately 10 minutes for the PTS2000 to warm-up.
6. Daily Check; Daily log checks to be completed at the start of each shift.
PRIOR TO COMMENCING TESTING ENSURE PB500 SERIES SYSTEM CALIBRATION / FINAL TEST
VERIFICATION LOG P/N 10039304 IS COMPLETED AS FOLLOWS;
REV: M
TITLE: PB560 FINAL TEST PROCEDURE
INSTRUCTIONS Page 3 of 12
O2 VALVE LEAK TEST [REF [1] PRD38, REF [10] SEC 4.6]
Daily Check: Air Pressure Gauge Check
This check is to be performed once before testing each day. Verify that the air supply pressure gauge is set to
100-120psi and record the setting on the PB500 Series System Final Test Equipment Verification Log, P/N
10039304. If this check fails, stop testing and notify the Line Supervisor or Manufacturing Engineer.
REV: M
TITLE: PB560 FINAL TEST PROCEDURE
INSTRUCTIONS Page 4 of 12
4. Check that the vent Date is current and that the vent Time is equivalent to the plant clock ± 1 minute.
5. Switch the AC Power Switch at the rear of the Ventilator OFF.
.
O2 VALVE LEAK TEST [REF [1] PRD38, REF [10] SEC 4.6]
Test Procedure:
1. Connect the tubing from the Sprint LC Multi Air Tester #1 test port to the O2 Inlet at the rear of the
ventilator.
2. Turn the ventilator power ON and ensure the ventilator is not ventilating.
3. On the Sprint LC Multi Air Tester press the square white button to start the test.
4. When the test has finished, record the Sprint LC Multi Air Tester leak measurement value (Leak Tester
Reading) on the DHR.
NOTE:
This measurement is indicated to the right of ‘Ch 1’ on the Sprint leak tester screen. Limit: <0.073 psig.
5. .Disconnect the tubing from the O2 Inlet at the rear of the ventilator.
P/N: 10039030
REV: M
TITLE: PB560 FINAL TEST PROCEDURE
INSTRUCTIONS Page 5 of 12
NOTE: The FiO2 sensor needs to be calibrated for each vent used for testing the FiO2.
2) Turn ON the ventilator if not already on, and press the Menu key to enter the Alarm menu.
3) Move the cursor to the FiO2 % row, and then press the Accept key twice to move to the ‘Current’ column.
4) Press the up arrow key to select YES and then the Accept key to perform the calibration operation.
5) If no calibration errors are displayed, then the calibration is successful.
6) Indicate the Pass or Fail test result by signing initials in the appropriate Pass or Fail column on the DHR.
7) Exit the Alarm menu to return to the ventilation menu.
P/N: 10039030
REV: M
TITLE: PB560 FINAL TEST PROCEDURE
INSTRUCTIONS Page 6 of 12
FiO2 Sensor Detection Test
1) Connect the Inspiratory Cap Test Assembly to the FiO 2 Sensor as shown in the picture below.
Fig 2 Fig 3
7) Press the Alarm Silence button to silence the Intentional Vent Stop alarm.
8) Disconnect the Inspiratory assembly and FiO2 Sensor from the vent.
P/N: 10039030
REV: M
TITLE: PB560 FINAL TEST PROCEDURE
INSTRUCTIONS Page 7 of 12
Fig.1 Connection of the patient circuit Fig.2 Connection of Patient hose to 3L Test Lung
for the volume and pressure test
Test Note: During the Paediatric and Adult Volume and Pressure Testing, the PB560 will
display the Intentional Vent Stop Alarm when ventilation is switched OFF and the alarm
will cease when ventilation is resumed. This is normal behaviour for the PB560.
REV: M
TITLE: PB560 FINAL TEST PROCEDURE
INSTRUCTIONS Page 8 of 12
16. From the Ventilation menu, select and enter the Preferences menu.
17. Set the ‘Paediatric Circuit’ option to OFF, and then return to the Ventilation menu.
18. Print the Test Report, Sign and Date the Test Report Title Page at the bottom and attach to the DHR.
19. Disconnect the paediatric patient circuit tubing from the vent and the 3 L test lung from the PTS2000.
REV: M
TITLE: PB560 FINAL TEST PROCEDURE
INSTRUCTIONS Page 9 of 12
REV: M
TITLE: PB560 FINAL TEST PROCEDURE
INSTRUCTIONS Page 10 of 12
Fig 3
Fig 2
17. Press the Alarm Silence button to silence the Intentional Vent Stop alarm.
18. Disconnect the adult patient circuit tubing from the vent and the 50 L test lung from the PTS2000.
REV: M
TITLE: PB560 FINAL TEST PROCEDURE
INSTRUCTIONS Page 11 of 12
REV: M
TITLE: PB560 FINAL TEST PROCEDURE
INSTRUCTIONS Page 12 of 12
N°1
Inspiratory flow measurement fault Calibrate the flow sensor
N°6
Abnormal Turbine speed Change the Turbine or CPU card
N°9
Low Proximal Pressure Change the CPU card
N°10
Low Internal Pressure Change the CPU card
N°11
Low Valve Pressure Change the CPU card
N°12
Event Log Error Change the CPU card