You are on page 1of 5

300 Chapter 8 Liquid-Liquid Extraction with Ternary Systems

Light employed for liquid-liquid extraction. The choice of pack-


liquid ing material, however, is somewhat more critical. A
material preferentially wetted by the continuous phase is
6 preferred. Figure 8.6 shows performance data, in terms of
HTU, for Intalox saddles in an extraction service as a
function of con-tinuous, Uc, and discontinuous, UD,phase
superficial velo-cities. Because of backrnixing, the HETP is
generally larger than for staged devices. For that reason,
. . packed columns are used only where few stages are needed.
. ..
Plate Columns
liquid Sieve plates in a column also reduce axial mixing and achieve
a more stagewise type of contact. The dispersed phase may be
the light or the heavy phase. In the former case, the dispersed
Heavy phase, analogous to vapor bubbles in dis-tillation, flows
liquid
vertically up the column, with redispersion at each tray. The
(a) heavy phase is the continuous phase, flowing at each stage
Figure 8.5 Spray columns: (a) light liquid dispersed, heavy liquid through a downcomer and then across the tray the way a liquid
continuous; (b) heavy liquid dispersed, light liquid continuous. does in a trayed distillation tower. If the heavy phase is
dispersed, upcomers are used for the light phase. Columns
dispersed phase are generated only at the inlet, usually by have been built and successfully operated for diameters larger
spray nozzles. Because of lack of column internals, than 4.5 m. Holes from 0.64 to 0.32 cm in diameter and 1.25 to
throughputs are large, depending upon phase-density differ- 1.91 cm apart are commonly used. Tray spacings are much
ence and phase viscosities. As in gas absorption, axial dis- closer than in distillation-10 to 15 cm in most applications
persion (backmixing) in the continuous phase limits these involving low-interfacial-tension liquids. Plates are usually
devices to applications where only one or two stages are built without outlet weirs on the downspouts. A variation of
required. Axial dispersion is so serious for columns with the simple sieve column is the Koch Kascade Tower, where
large diameter-to-length ratio that the continuous phase perforated plates are set in ver-tical arrays of complex designs.
may be completely mixed. Therefore, spray columns are
rarely used, despite their very low cost. If operated in the proper hydrodynamic flow regime,
extraction rates in sieve-plate columns are high because the
dispersed-phase droplets coalesce and re-form on each
Packed Columns stage. This helps destroy concentration gradients, which
Axial mixing in a spray column can be substantially re-duced, develop if a droplet passes through the entire column
but not eliminated, by pachng the column. The pack-ing also without distur-bance. Sieve-plate columns in extraction
improves mass transfer by breaking up large drops to increase service are subject to the same limitations as distillation
interfacial area and promotes mixing in drops by distorting columns: flooding, entrainment, and, to a lesser extent,
droplet shape. With the exception of Raschig rings [5], the weeping. Additional problems, such as scum formation at
same packings used in distillation and absorption are interfaces due to small amounts of impurities, are
frequently encountered in all types of extraction devices.

. I

IJ,
I I I I
UD=dispersed phase velocity
I Columns with Mechanically
Assisted Agitation
~-E-~-~-R U = 56.5ft/h If the surface tension is high, and/or the density difference
between the two liquid phases is low, and/or liquid viscosi-
- ties are high, gravitational forces are inadequate for proper
"
UD = 24.6 ftlh
phase dispersal and the creation of turbulence. In that case,
some type of mechanical agitation is necessary to increase
I I I I I I
interfacial area per unit volume and/or decrease mass-
0 50 100 150 200 250 300 transfer resistance. For packed and plate columns, agitation
y, continuous phase velocity, ftlh is provided by an oscillating pulse to the liquid, either by
Figure 8.6 Efficiency of 1-in. Intalox saddles in a column 60 in. mechanical or pneumatic means. Pulsed, perforated-plate
high with MEK-water-kerosene. columns found considerable application in the nuclear
[From R.R. Neumatis, J.S. Eckert, E.H. Foote, and L.R. Rollinson, Chem.
Eng. Progc, 67(1), 60 (1971) with permission.]
8.1 Equipment 301

Wire-mes
packing

Heavy
-+ liquid

Motor
I
+ Light

Heavy liquid out


liquid in

Feed if
operated for
fractional
extraction
nRHH

Figure 8.7 Commercial extractors


with mechanically assisted agitation:
(a) Scheibel column-first design;
(b) Scheibel column-second
design;
(c) Scheibel column-third design;
(d) Oldshue-Rushton (Mixco)
(d) column; (continued)
equilibrium stages in a compact device without large numbers
of pumps and

industry in the 1950s, but their popularity declined because of


mechanical problems and the difficulty of propagating a pulse
through a large volume [6] . The most important mechanically
agitated columns are those that employ rotat-ing agitators,
driven by a shaft that extends axially through the column. The
agitators create shear mixing zones, which alternate with
settling zones in the column. Differences among the various
agitated columns lie primarily in the mix-ers and settling
chambers used. Nine of the more popular arrangements are
shown in Figure 8.7. Agitation can also be induced in a column
by moving the plates back and forth in a reciprocating motion
(Figure 8.7j) or in a novel horizontal contactor (Figure 8.7k).
Such devices are also included in Figure 8.7. These devices
answer the plea of Fenske, Carlson, and Quiggle [7] in 1947
for equipment that can efficiently provide large numbers of
motors, and extensive piping. They stated, "Despite. . .
advantages of liquid-liquid separational processes, the
prob-lems of accumulating twenty or more theoretical
stages in a small compact and relatively simple
countercurrent opera-tion have not yet been fully solved."
Indeed, in 1946 it was considered impractical to design for
more than seven theo-retical stages, which represented the
number of mixer-settler units in the only large-scale,
commercial, liquid-extraction process in use at that time.
Perhaps the first mechanically agitated column of impor-
tance was the Scheibel column [8] (Figure 8.7a), in which
countercurrent liquid phases are contacted at fixed intervals by
unbaffled, flat-bladed, turbine-type agitators (Figure 8.3)
mounted on a vertical shaft. In the unbaffled separation or
calming zones, located between the mixing zones, knitted
wire-mesh packing is installed to prevent backmixing between
mixing zones and to induce coalescence and
302 Chapter 8 Liquid-Liquid Extraction with Ternary Systems

Variable-speed drive

Light liquid
outlet
Settling zone
Heavy
-

liquid
inlet Interface Contact zone

Transport zone

qE $stator ring Shell

Stator
Rotor disk
Light
liquid Agitator
inlet
Settling zone
- Heavy
liquid

oPhase
Variable-speed drive

Light
Heavy

phase in out

phaseLight i n <-7
Figure 8.7 (Continued) (e) rota
(i)
J/ disk-contactor (RDC); (f) asymm
Heavy rotating-disk contactor (ARD); (g)
phase out
tion of ARD contactor; (h) Kuhni
umn; (i) flow pattern in Kuhni colu

designs did not pennit removal of the agitator shaft for inspection and
settling of drops. The mesh material must be wetted by the maintenance. Instead, the entire internal assembly (called the
dispersed phase. For more economical designs for larger- cartridge) had to be removed. To permit removal of just the agitator
diameter installations (>1 m), Scheibel [9] (Figure 8.7b) added assembly shaft, especially for large-diameter columns (e.g., > 1.5 m),
outer and inner horizontal annular baffles to divert the vertical and
flow of the phases in the mixing zone and to ensure complete
mixing. For systems with high interfacial surface tension and
viscosities, the wire mesh is removed. The first two Scheibel
The Oldshue-Rushton extractor [ll] (Figure 8.7d) con-
inspection, cleaning, and repair, Scheibel [lo] sists of a column with a series of compartments
offered a third design, shown in Figure 8 . 7 ~ separated by annular outer stator-ring baffles, each with
Here. the agitator assembly shaft can be removed four vertical baffles attached to the wall. The centrally
because it has a smaller diameter than the opening mounted verti-cal shaft drives a flat-bladed turbine
in the inner baffle. impeller in each compartment.
A third type of column with rotating agitators that ap-
peared about the same time as the Scheibel and Oldshue-
allow an access way through the column for any necessary Rushton columns is the rotating-disk contactor
(RDC)

You might also like