Professional Documents
Culture Documents
OPTIMAS™ manual
Document no. 0060.170-11
Alias™ Autosampler
Integrity™ Autosampler
Mistral™ Column Thermostat
Reliance™ Autosampler
Symbiosis™ system
Symbiosis™ Pro system
Symbiosis™ Pico system
SPH1240™ Gradient pump
Hysphere™ SPE cartridges
P. de Keyserstraat 8
7825 VE EMMEN
The Netherlands
www.sparkholland.com
OPTIMAS™ is a registered trademark of Spark Holland B.V. All other brands, company and product names are or may be trademarks of their
respective holders and are used to identify their products or services.
Copyright © 2014 Spark Holland B.V. All rights reserved. No part of this publication may be reproduced, stored in a retrieval system or
transmitted in any form by any means, without written permission from Spark Holland B.V.
Spark Holland B.V. reserves the right to make changes to the design and specifications of the system and this manual without prior notice.
0060.170-11 Optimas manual
Contents
2 Introduction 9
3 Identification of parts 11
3.1 Keyboard and display 14
4 Injection principles 17
4.1 Full loop injections 17
4.2 Partial loopfill injections 19
4.3 µL pickup injections 22
6 Programming reference 31
6.1 Programming overview 31
6.2 Ready screen options 33
6.2.1 <WASH> 33
6.2.2 <MAINTENANCE> 34
6.2.3 <TEMP> 34
6.2.4 <UTILS> 35
6.2.5 <SERVICE> 35
6.3 System settings programming 36
6.3.1 <GENERAL> 36
6.3.2 <METHOD> 38
6.3.3 <TRAY> 38
6.3.4 <IO> 39
6.3.5 <PREP> 39
6.4 Method programming mode 40
6.4.1 <INJECTION> 41
6.4.2 <WASH> 42
6.4.3 <MIX> 43
6.4.4 <TIMED EVENTS> 45
1
Optimas manual 0060.170-11
7 Maintenance 53
7.1 Injection valve and rotor seal 53
7.2 Needle and tubing 54
7.3 Sample needle 57
7.3.1 Sample needle replacement 58
7.3.2 Air needle replacement 58
7.3.3 Sample needle height setting 58
7.4 Syringe 59
8 Specifications 61
9 Trouble shooting 65
9.1 Start up problems 66
9.2 Analytical problems 67
9.2.1 No injection 68
9.2.2 Bad reproducibility 69
10 Installation 71
10.1 Fluid connections 74
10.1.1 HPLC connections 75
10.1.2 Waste tubing 75
10.1.3 Wash solvent and system rinse 76
10.1.4 Syringe 77
10.2 Preparation of vials 77
10.2.1 Types of vials 77
10.2.2 Filling and sealing the vials 81
10.2.3 Loading the sample tray 81
10.3 Control I/O connections 81
10.3.1 Contact closures output (P1) 81
10.3.2 TTL input (P2) 82
2
0060.170-11 Optimas manual
Index 103
3
Optimas manual 0060.170-11
4
0060.170-11 Optimas manual
This manual is intended for laboratory technicians who use the OPTIMAS
autosampler for execution of analytical runs.
Please read this manual and the safety instructions carefully before you start to use
the autosampler.
This manual provides the following information on the autosampler:
section 1 offers a short introduction to the autosampler
section 2 offers identification of parts and a description of the injection
principles of the autosampler
section 3 offers procedures for use of the autosampler
section 4 offers maintenance procedures
section 5 lists specifications
section 6 offers trouble shooting information
section 7 describes the installation procedure for the autosampler
appendix A lists available spares and accessories
appendix B lists error codes and how to resolve them
appendix C offers a test procedures
appendix D describes the Prep option
appendix E contains sample log book pages for the OPTIMAS
appendix F describes procedures for storage, packaging and end-of-life.
An index is provided for easy reference.
1.1 Pictorials
5
Optimas manual 0060.170-11
6
0060.170-11 Optimas manual
System Operation
To keep up the high performance of the system we
recommend that the system is checked regularly and
maintenance procedures are carried out.
Do not use the device if there is visible damage.
Note that the OPTIMAS is a high-pressure device. Always
make sure pressure in the device is built up slowly. Reduce
the pressure gradually to 0 before disconnecting any tubing
or before opening the purge valve. Sudden loss of pressure
may damage the pump and the analytical column connected
to it.
Electrical safety
Removal of panels may expose users to dangerous voltages.
For that reason this should only be done by authorized
service engineers. Disconnect the system from all power
sources before removing protective panels.
Replace or repair faulty insulation on power cords
immediately after discovery of the fault.
Check that the actual power voltage is the same as the
voltage for which the instruments are wired. Make sure power
cords are connected to correct voltage sources.
The system must only be used with appliances and power
sources with proper protective grounding to prevent damage
through buildup of static electricity.
Do not place the system in the vicinity of equipment that emit
EMC radiation.
Solvents
The solvents used may be flammable and toxic. The room in
which the system is installed should be well ventilated to
prevent that solvent vapors cause poisoning or ignite and
cause a fire.
Use of open fire in the vicinity of the system must be strictly
prohibited. Do not install the system in the same room with
any other equipment that emits or could potentially emit
sparks.
Provide protective equipment near the instrument. If solvent
gets into the eyes or on the skin, it must be flushed away
immediately. Provide equipment, such as eye wash stations
and safety showers, as close to the system as possible.
Solvents bottles should be sealed to minimize any risks
related to solvent vapors.
Do not allow flammable and/or toxic solvents to accumulate.
Follow a regulated, approved waste disposal program. Never
dispose of flammable and/or toxic solvents through the
municipal sewage system.
Perform periodic leak checks on supply lines.
7
Optimas manual 0060.170-11
Biohazard
When you analyze biological fluids you need to take all
possible precautions and treat all specimens as potentially
infectious. Always wear protective goggles and gloves when
handling toxic or biologically infectious samples to prevent
bio hazards or toxic hazards while working with the
OPTIMAS.
The following typographical conventions are used to refer to the various controls
and display text of the OPTIMAS autosampler:
control keys and Start/Stop, in square brackets.
Hold/Cont., keys of numerical Examples: [Start/Stop], [Escape]
keypad
8
0060.170-11 Optimas manual
2 Introduction
OPTIMAS has been designed to meet the challenge of the modern analytical
laboratory: it is robust, cost-effective and easy to implement. Sample cooling is
available for consistent results. High-resolution syringe control guarantees superior
precision for injection and reagent addition. The use of these and many more
features is described in this manual to help you exploit the full potential of
OPTIMAS.
9
Optimas manual 0060.170-11
10
0060.170-11 Optimas manual
3 Identification of parts
1. Hold the cover with both hands on either side of the autosampler.
11
Optimas manual 0060.170-11
1. Needle arm
2. Injection valve
3. Wash position
8. Sample tray
9. Syringe dispense
12
0060.170-11 Optimas manual
1. CE label
5. Power connector
6. On/off switch
13
Optimas manual 0060.170-11
1. Display
14
0060.170-11 Optimas manual
15
Optimas manual 0060.170-11
16
0060.170-11 Optimas manual
4 Injection principles
The OPTIMAS uses a syringe to aspirate the sample from a vial into the sample
loop. To prevent contamination of the syringe the OPTIMAS is equipped with a
buffer tubing between the syringe and the injection valve.
The wash solvent is used to remove the sample from the buffer tubing and sample
needle and rinse the buffer tubing and sample needle.
The switching sequence for a full loop injection is schematically shown below:
1. The Initial situation: the injection valve is in the INJECT position. The sample
needle with air needle has entered the vial. The headspace pressure, applied
through the outer air needle, ensures that no air or vapor bubbles are formed
during sample aspiration.
17
Optimas manual 0060.170-11
2. The syringe dispenser aspirates the "flush volume" from the sample vial to fill
the sample line with sample and remove wash solvent.
3. The injection valve switches to the LOAD position, placing a distinct sample
plug at the inlet of the sample loop.
5. The injection valve switches into the INJECT position. The sample loop is now
part of the HPLC mobile phase flow path: sample is transported to the column.
The analysis starts.
18
0060.170-11 Optimas manual
The next injection sequence will start with a flush of 50% of the programmed flush
volume, in case an injection from the same vial and no wash routine is
programmed. Otherwise it will start with a flush of the programmed flush volume. If
the withdrawal of sample for the next injection will exceed the total volume of the
sample buffer tubing, the buffer tubing is rinsed before the next injection. The next
injection will start with the programmed flush.
Air segment
An air segment can be used to reduce the amount of flush volume. This air
segment is at the front of the flush volume and will not be injected and therefore will
not influence the injection. The air segment may be selected in the system settings
(ON/OFF, see section 6.3.1).
With a standard needle, the flush volume must be at least 30 µL for injections with
air segment and 35 µL for injections without air segment. If the samples are highly
viscous it may be necessary to program larger flush volumes and reduce the
syringe speed for better performance.
Figure 6: Situation after filling the loop (A=with air segment; B=with air segment)
The switching sequence for a partial loopfill injection is schematically shown in the
following figures:
19
Optimas manual 0060.170-11
1. The Initial situation: the injection valve is in the INJECT position. The sample
needle with air needle has entered the vial. The headspace pressure, applied
through the outer air needle, ensures that no air or vapor bubbles are formed
during sample aspiration.
2. The syringe dispenser aspirates the "flush volume" from the sample vial to fill
the sample line with sample and remove wash solvent.
3. The injection valve switches to the LOAD position, placing a distinct sample
plug at the inlet of the sample loop.
4. The programmed injection volume is now aspirated into the sample loop.
20
0060.170-11 Optimas manual
5. The injection valve switches to the INJECT position. The sample loop is now
part of the HPLC mobile phase flow path: sample is transported to the column.
The analysis starts.
The next injection sequence will start with a flush of 50% of the programmed flush
volume, in case an injection from the same vial and no wash routine is
programmed. Otherwise it will start with a flush of the programmed flush volume. If
the withdrawal of sample for the next injection will exceed the total volume of the
sample buffer tubing, the buffer tubing is rinsed before the next injection. The next
injection will start with the programmed flush.
Air segment
An air segment can be used to reduce the amount of flush volume. This air
segment is at the front of the flush volume and will not be injected. The air segment
can be selected in system settings (ON/OFF, see section 6.3.1).
With a standard needle the flush volumes must be minimal 30 µL for injections with
air segment and 35 µL for injections without air segment. If the samples are highly
viscous it may be necessary to program larger flush volumes and reduce the
syringe speed for better performance.
Figure 7: Situation after filling the loop (A=with air segment; B=without air segment)
21
Optimas manual 0060.170-11
1. The Initial situation: the injection valve is in the INJECT position. The sample
needle has entered the vial of transport liquid (mobile phase, this to avoid
disturbance of the chromatogram with an additional peak of the transport
solvent).
2. For the first injection, after a wash or after emptying of the buffer tubing. The
syringe dispenser aspirates transport liquids from the transport vial to fill the
sample line with transport liquid and remove wash solvents.
3. The needle moves from the transport vial to the sample vial. The injection
valve switches to the LOAD position.
22
0060.170-11 Optimas manual
5. The sample needle moves back to the transport vial. The sample is
quantitatively transported into the loop, with transport liquid (mobile phase)
from the transport vial.
6. The injection valve switches into the INJECT position. The sample loop is now
part of the HPLC mobile phase flow path: sample is transported to the column.
The analysis starts.
For the next sequence the first withdrawal of transport solvent is omitted, unless a
wash routine is performed or the OPTIMAS has emptied the buffer tubing to the
waste position. In these cases the sequence is completely repeated.
Air segment
If an air segment has been programmed, it appears at the front of the first plug of
transport liquid and at the front of every sample plug. An air segment can be
programmed in the system settings (see section 6.3.1).
23
Optimas manual 0060.170-11
Figure 8: Situation after filling the loop (A=with air segment; B=without air segment)
24
0060.170-11 Optimas manual
Use this section to get started with the OPTIMAS, after the installation instructions
(see section 10) have been executed.
Three different examples are described below; make sure you adapt values for the
type of application you wish to use.
1. Allow the OPTIMAS to adopt ambient temperature for at least one hour
2. Check OPTIMAS for visual damage resulting from transport; if this is the case,
please contact your distributor immediately.
4. Remove the safety screw on the right-hand side of the front cover.
Always keep the keyboard front cover closed during
operation.
1. Check fuses and voltage settings on the rear side of the instrument.
3. After turning "ON" the power, the Ready screen appears displaying the
firmware revision number.
4. Connect the drain tubing to the wash outlet on the lower front side of the
instrument.
5. Fill the wash solvent bottle with distilled water/ 2-Propanol (80/20 v/v%) or
mobile phase. Only water/organic solvents should be used. Do not use any
crystalline or buffer solutions; these may block the system and cause severe
damage. To avoid air bubbles in the syringe, degas the wash solvent.
6. Fill the wash solvent tubing, syringe and buffer tubing by washing the system 2
or 3 times (press the <WASH> function key from the Ready screen). Check if
there are air bubbles trapped in the syringe; they can be removed by gently
tapping the syringe.
9. Check for leakage and let the system equilibrate for at least 5 minutes.
25
Optimas manual 0060.170-11
1. First you define a method such as the injection, mix or wash method. These
methods are stored in battery back up.
Wash Program:
26
0060.170-11 Optimas manual
If you press <START> OPTIMAS will start searching for vial 1 and perform a 10 µL
injection.
This example describes execution of three 10 µL injections from one vial without
sample loss using the µL Pick-up injection routine, followed by a wash routine
between each injection.
If the Air segment is set ON in System Settings, a small air
segment will also be injected into your HPLC system (for
more detailed information, see section 4.3 of this manual).
27
Optimas manual 0060.170-11
Wash Program:
Press keys: Description:
<WASH> Selecting wash program
<INJECTION> [ENTER} Perform a wash after each injection
[1] [Enter] Enter the number of syringe volumes
to be used for washing, e.g. 1 syringe
volume.
Press [Escape] three times to go back to the Ready screen.
If you press <START> OPTIMAS starts searching for the transport vial and perform
3 injections of 10 µL. After every injection the needle system will be washed with 1
syringe volume of wash solvent.
28
0060.170-11 Optimas manual
Wash Program:
Press keys: Description:
<WASH> Selecting Wash program
<NONE> Disable the wash
Press [Escape] once to go up one level.
29
Optimas manual 0060.170-11
If you press <START>, OPTIMAS starts searching for vial 86 and transport 180 µL
to the destination vial twice, subsequently 40 µL of sample will be added and after
mixing 3 times a 10 µL injection will be performed.
30
0060.170-11 Optimas manual
6 Programming reference
This section describes all programming options for the OPTIMAS autosampler.
This overview gives a short description of all programming options. Please note
that all options are described in more detail in following sections.
[SYSTEM] options:
<GENERAL> Loop volume
Needle volume
Syringe volume
Syringe speed
Skip missing vials
Air segment
Headspace pressure
<METHOD> Use calibration vials
<TRAY> Tray type
<IO> Inject marker pulse length
Input edge next injection
Freeze input active
Reset output after last Series
<PREP> Select Prep mode; a number of system settings will be
set to a fixed value.
31
Optimas manual 0060.170-11
[METHODS] options:
<INJECTION> <FULL> Flush volume
No. Injections/vial
Analysis time
<PARTIAL> Flush volume
No. injections/vial
Injection volume
Analysis time
<PICK-UP> No. injections/vial
Injection volume
Analysis time
<WASH> Wash volume
Wash between: injections/vials/series
<TIMED EV.> timed events program:
4 x AUX-1
4 x AUX-2
END TIME
<MIX> Mix program: Add, Mix, Wait
32
0060.170-11 Optimas manual
After power up, the OPTIMAS will run an initialization routine in which it performs a
self-test and every item in the autosampler is set to default positions.
[MENU]
READY (X.XX)
TRAY:--°C
<WASH> <MAINTENANCE> <TEMP>
In the bottom line of the display, control keys are available that can be used. If
[Menu] is displayed in the top right-hand corner, the [Menu] function key can be
used to navigate to other available control keys.
[MENU]
READY (X.XX)
TRAY:--°C
<UTILS> <SERVICE>
At the right-hand side of the screen the tray temperature is displayed (only when
this option is installed).
This section describes all options available from the Ready screen.
6.2.1 <WASH>
The <WASH> option activates the manual wash of the OPTIMAS. All tubing, the
needle and the syringe will be rinsed.
33
Optimas manual 0060.170-11
6.2.2 <MAINTENANCE>
<MAINTENANCE> offers access to three options for controlling the needle arm,
the syringe and the injection valve:
<VLV LOAD> With this function the injection valve can be switched.
Press <VLV LOAD> to switch the valve from INJECT to
LOAD.
When the valve is in the LOAD position it can be switched
back to INJECT by pressing <VLV INJ>.
It is not possible to leave this function when the valve is in
the LOAD position, first switch back by pressing <VLV
INJ> and then press [Escape].
6.2.3 <TEMP>
Press <TEMP> to enter the temperature control screen for sample tray cooling
(optional).
<TRAY-COOLING>
34
0060.170-11 Optimas manual
<ON> <OFF>
Switch the tray cooling on by pressing <ON> and program the new setpoint. After
pressing [Enter] the new tray cooling setpoint will be activated.
6.2.4 <UTILS>
<Utilities> offers the following options:
6.2.5 <SERVICE>
The service mode can only be entered if you have a service entry code. This code
is only known to authorized service engineers.
1. Press [SERVICE].
2. Enter the code with the help of the keys of the numeric keypad.
3. Press [E] on the numeric keypad. The OPTIMAS is now in Service mode.
35
Optimas manual 0060.170-11
Press [System] to enter the System Settings. In this menu all system variables of
the OPTIMAS can be set.
<PREP>
The System Settings are divided in five sub-sections: to enter a sub-section, press
the corresponding function key. To step through a sub-section press the [Enter]
key.
After all values have been defined correctly within a sub-section, return the menu
screen of the System Settings by pressing [Escape]. Return to the Ready screen
by pressing [Escape] one more time. The values are stored in battery backup
memory.
The following sections describe the programmable system settings, with their
ranges.
6.3.1 <GENERAL>
Press <GENERAL> to enter general system settings.
36
0060.170-11 Optimas manual
Range Comment
Skip missing vials: <YES> Define if the OPTIMAS has to stop if a vial
<NO> is not found or continue with the next
programmed vial.
Program <NO> to stop the OPTIMAS in
case a vial is not found.
Air segment: <YES> Inject a sample with or without an air
<NO> segment between the sample and the
wash solvent. If the air segment is ON
(YES) the OPTIMAS will draw an air
segment of 5 µL before the flush volume
prior to an injection.
The air segment will not be injected in the
analytical system.
In case of µL pick-up the sample will be
preceded by an air segment, which will be
injected in the analytical system (see also
section 4.3).
Headspace pressure: <YES> The OPTIMAS normally uses headspace
<NO> pressure in combination with the syringe
to transport sample in to the loop. If, for
some reason, no headspace pressure can
be used, switch if off by selecting <NO>.
The compressor will always be used in the
wash position of the OPTIMAS. The
accuracy and reproducibility of the
OPTIMAS may decrease if <NO>
headspace pressure is used.
37
Optimas manual 0060.170-11
6.3.2 <METHOD>
Press <METHOD> to define the use of calibration vials
Range Comment
Use calibration vials: <ENABLED> Determine if calibration vials can
<DISABLED> be used in Series.
6.3.3 <TRAY>
Press <TRAY> to indicate the type of tray to be used:
Range Comment
Tray type: <84+3> Standard tray with 84 vials, 1.5 mL
and 3 vials of 10 mL for special use.
38
0060.170-11 Optimas manual
6.3.4 <IO>
Press <IO> to configure Inputs and outputs:
6.3.5 <PREP>
Press <PREP> to enter settings for the Prep Mode (optional; see also Appendix D):
Range Comment
Use Prep Mode: <YES> Define if the OPTIMAS is a standard
<NO> autosampler, or a Prep mode
autosampler.
In the Prep Mode the mix, the full loop and µL pick-up
injection mode, will automatically be disabled.
39
Optimas manual 0060.170-11
In the Prep Mode not all of the System Settings are programmable. see table below
for an overview.
For information how to install the Prep option to a standard OPTIMAS refer
appendix D.
A method is any combination of the above and is saved as combination under its
method number.
40
0060.170-11 Optimas manual
Enter the method number; OPTIMAS will display the main screen of method
programming:
From this screen the various sections can be programmed after pressing the
corresponding function key.
Existing programs in this method are displayed between brackets on the first line:
I for the injection program
W for the wash program
M for the mix program
T for the timed events program.
When you have programmed the method, press [Escape] to leave the method
programming mode.
The methods are stored in battery backup memory.
6.4.1 <INJECTION>
Press <INJECTION> to enter the injection programming mode.
Depending on the injection mode you selected, the following parameters can be
entered:
Parameter: Range: Comment:
Flush volume 0-9999 µL The amount of sample used to fill
the needle plus tubing connection
to the injection valve (not the loop-
fill volume). Only for full loop and
partial loopfill injections.
41
Optimas manual 0060.170-11
When all parameters have been programmed, return to the main method
programming screen by pressing [Escape].
6.4.2 <WASH>
Press <WASH> to enter the wash programming mode:
42
0060.170-11 Optimas manual
When all parameters have been programmed, return to the main method
programming screen by pressing [Escape].
6.4.3 <MIX>
Press <MIX> to enter the mix programming mode. A question will appear, to ask if
the mix program must be used in this method. Press <YES> to define the mix
program.
The mix method may contain a maximum of 15 programmable lines (including the
line for END OF MIX).
To scroll through the program steps use the <DOWN> and <UP> function keys.
Press <DELETE> to delete the displayed step.
Press [Enter] to edit the displayed step.
43
Optimas manual 0060.170-11
44
0060.170-11 Optimas manual
Example: MIX 3 times with 250 µL. This will result in the
following actions:
Aspirate an air segment of 50 µL, to separate the wash
solvent in the buffer tubing from solvent to be mixed.
Unload the syringe to the syringe-waste position.
Aspirate 250 µL the solvent and dispense it back into
the vial.
Repeat the last action two times.
Rinse buffer tubing and needle with wash solvent.
<WAIT> x:xx:xx
Wait a programmed period of time before continuing
with the next step (reaction time).
When all parameters have been programmed, return to the main method
programming screen by pressing [Escape].
45
Optimas manual 0060.170-11
When all parameters have been programmed, return to the main method
programming screen by pressing [Escape].
46
0060.170-11 Optimas manual
PROGRAMMING SERIES
SERIES NUMBER: 1
Each time a value has to be entered, key in a number [1-9] or press [CL] or 0 to
select NONE and press [Enter] to go to the next screen.
47
Optimas manual 0060.170-11
Figure 9: Run sequence with 3 calibration vials between every 5 vials (First calibration vial: 1, last
calibration vial: 3 and number of samples between calibrations: 5).
When all parameters have been programmed, return to the Series programming
screen by pressing [Escape]. Another Series can now be programmed. If all Series
have been programmed, return to the Ready screen by pressing [Escape].
This section describes the options that can be used during a run.
6.6.1 [Start/Stop]
Use [Start/Stop] when the Ready screen is displayed to start a Run.
The OPTIMAS will ask for the Series number to start with and the last Series
number.
Enter each number and press [Enter] to go to the next screen.
48
0060.170-11 Optimas manual
EXECUTE SERIES 1
<START> <REMOTE>
To return to the Ready screen without starting the OPTIMAS, press [Escape].
Press [Start/Stop] again to stop the Run. The OPTIMAS will execute a shutdown
sequence by removing all the sample from the buffer tubing and performing a
standard wash routine.
49
Optimas manual 0060.170-11
6.6.2 [Priority]
OPTIMAS offers the possibility to analyze an emergency sample without stopping
the current Run.
PROGRAMMING PRIORITY
PRIORITY SAMPLE: 12
Return to the Run screen by pressing [Escape], the priority sample will be
processed after the current sample has finished with the method that is running at
the moment of entering the vial number.
To indicate the priority sample has been activated, a blinking "p" is displayed in the
Run screen.
As long as the "p" is blinking it is possible to change or
remove the priority sample.
50
0060.170-11 Optimas manual
With the Next Injection Input OPTIMAS will start the next injection in the Series.
When the Series is completed, it starts the next injection from the next Series, until
all Series have been completed.
A blinking "r" in the lower left corner of the Run screen indicates that the
autosampler is running in Remote Control mode.
After completion of all Series the OPTIMAS will display the following screen:
To return to the Ready screen press [Escape]. From the Ready screen the
OPTIMAS can be re-started by pressing the [Start/Stop] key.
The programming possibilities are the same as described in sections 6.4 and 6.5.
51
Optimas manual 0060.170-11
It is not possible to add Series to the Run, but you can program empty Series
between the first and last Series in a Run. As long as the empty Series have not
been executed, they can be (re-) programmed. In this way Series with vials can be
added later to the Run.
52
0060.170-11 Optimas manual
7 Maintenance
1. Disconnect all tubing from the valve. Only the sample loop can stay in place.
3. Remove the valve: push the pushers A backward to release the valve.
4. Pull the valve out of the instrument while turning the valve counter-clockwise.
2. Gently open the valve and take out the rotor seal.
53
Optimas manual 0060.170-11
4. Place the stator back on the rotor and fasten the screws.
1. Hold the valve for the mounting with the port 1 position facing to the top.
2. Push the valve into the mounting and turn the valve clockwise till it clicks in the
fitting.
4. Insure that the valve is in the INJECT position. Switching the OPTIMAS OFF
and ON can do this, all Series will be reset.
Two sizes of syringes are available for the dispenser: 250 and 1000 µL.
The 250 µL syringe is the standard syringe; combined with the standard 500 µL
buffer and the standard 100 µL sample loop, the following injection volume range is
available for the various injection modes:
Full loop : 100 µL
Partial loopfill: 1 - 50 µL
µL pick-up : 1 - 27 µL
The maximum injection volumes are calculated with the following formulas:
Full loop : injection volume =loop volume
Partial loopfill: max. inj. volume =½ x of loop volume
µL pick- up : max. inj. volume =½ x loop volume - 1½ × needle
volume
54
0060.170-11 Optimas manual
55
Optimas manual 0060.170-11
Buffer tubing:
Syringe: Buffer tubing:
250 µL 500 µL
1000 µL 2000 µL
56
0060.170-11 Optimas manual
1. Ferrule
2. Nut
3. Tefzel tubing
5. Ferrule (PTFE)
6. Conical ring
7. Sample needle
9. Needle holder
57
Optimas manual 0060.170-11
1. Place the sample needle in the front position, press <MAINTENANCE>, <NDL-
HOR>.
2. Loosen needle connection nut (4) that fixates the sample needle (7).
3. Loosen the nut (1) that connects the PTFE tubing (3) to the injection valve.
4. Remove sample needle by pulling it out of its fitting with the PTFE tubing that
usually sticks to the needle.
5. Put in a new needle assembly, watch the air seal (6) is around the needle.
8. Place the sample needle back in the home position and leave the maintenance
mode, press <NDL HOME> and [Escape].
9. Check the needle penetration depth, see section and adjust if necessary.
10. Perform a wash routine to clean the new needle by pressing <WASH>.
58
0060.170-11 Optimas manual
b. Program a Series with the programmed method and vial 2 as first and last
vial.
c. Start the Series.
d. As soon as the sample needle has entered the vial, switch off the power
supply.
4. Use the needle connection nut (4) as a grip, adjust needle penetration depth.
Needle should stop not less than 1 mm above the bottom of a standard vial
and not less than 2 mm above the bottom of a conical vial.
6. Switch on the power supply; the OPTIMAS will initialize and is ready for use.
7.4 Syringe
The autosampler is factory equipped with a 250 µL syringe, but can also be
equipped with a 1000 µL syringe.
59
Optimas manual 0060.170-11
5. Fill the new syringe with wash solvent, be sure most air bubbles are removed
from the syringe.
6. Connect the plunger of the filled syringe to the syringe drive and connect the
syringe at the syringe valve.
8. Remove the air from the syringe by pressing <SYR HOME>. The syringe will
move to its HOME position and dispense its contents to the syringe waste.
9. If some air remains in the syringe press <SYR END> again. The syringe is
filled with wash solvent.
10. Press <SYR HOME> to dispense the wash solvent to waste. If there is still air
in the syringe, repeat the previous action and tap the syringe gently as the
wash solvent is dispensed to the syringe waste.
11. Leave the maintenance screen by pressing [Escape] and press <WASH> to
perform a standard wash routine. All tubing connected to the syringe valve will
be refilled and flushed.
60
0060.170-11 Optimas manual
8 Specifications
General
Sound pressure level LAeq <70 dB
Working temperature 10 - 40°C, indoor use only
Humidity 20 - 80% RH, indoor use only
Safety and EMC compatibility According to EC-directives
Installation class II
Pollution degree 2
Altitude up to 2000m
Dimensions (W x D x H) 300 mm x 500 mm x 340 mm
11.8". x 19.7" x 13.4"
Weight 19 kg (42 lb.)
21 kg (46 lb.) with cool option
Power requirements 115/230 VAC; ±15%; 50 Hz / 60 Hz; 250 VA
Fuses For 115 VAC: two 5.0 AT-fuses
(slow, ¼" x 1¼", UL/CSA)
For 230 VAC: two 2.5 AT-fuses
(slow, 5 x 20 mm, IEC 127)
Sampling
Sample capacity: Standard tray:
84 vials of 1.5 mL (Std) and
3 vials of 10 mL (LSV)
Vial dimensions (cap Maximum vial height: 47 mm
included): Minimum vial height: 32 mm
Loop volume 1 - 5000 µL
61
Optimas manual 0060.170-11
Analytical performance
Reproducibility RSD <= 0.3% for full loop injections
RSD <= 0.5% for partial loopfill injections,
injection volumes >10 µL
RSD <= 1.0% for µL pick-up injections,
injection volumes >10 µL
Memory effect < 0.05% with programmable needle wash
Programming
Methods 9 programmable methods
Injection method Injection mode:
Full loop injection
Partial loopfill injection
µL pick-up injection
Injection volume 1 µL - 5000 µL, with 1 µL increment
Full loop injection: max. 5000 µL
Partial loopfill injection: max. 2500 µL
µL pick-up injection: max. 2477 µL
Injections per vial max. 9 injections
Analysis time max. 9 hr 59 min 59 sec
Wash Programmable:
Wash between injections
Wash between vials
Wash between Series
Mix max. 15 programmable steps
Timed events Programmable:
4 x AUX-1 ON/OFF
4 x AUX-2 ON/OFF
Series 9 programmable Series
Priority sample Programmable
Communication
Outputs Inject marker (Relay & TTL)
2 Auxiliary outputs (Relay)
Alarm output (Relay)
Inputs Next injection input (TTL)
Freeze input (TTL)
Stop input (TTL)
Options (user-installable)
Prep option Including software, large volume tray, 2.5 mL
syringe and large bore Valco injection valve
with 10 mL loop.
Bio-compatible valve Special PEEK injection valve, quick connect
mounting.
62
0060.170-11 Optimas manual
Option (factory-installed)
Sample tray cooling Built-in Peltier cooling
Range: 4°C - 15°C
Accuracy: ± 2°C (at temperature sensor)
(Temperature at relative humidity of 80% and
ambient temperature of 30°C)
Cooling 20°C below ambient.
capacity:
63
Optimas manual 0060.170-11
64
0060.170-11 Optimas manual
9 Trouble shooting
This section offers basic troubleshooting information. Contact your service provider
if you come across a problem that you cannot solve.
65
Optimas manual 0060.170-11
66
0060.170-11 Optimas manual
In case of analytical problems, determine whether the cause of the problem is the
autosampler, or the rest of the system.
Quick check!
Replace the valve by a manual injection valve to
discriminate between valve problems and other problems.
Perform some manual Full loop injections. If the results are
fine the fault has to be found in the autosampler, if not the
HPLC system should be checked.
67
Optimas manual 0060.170-11
9.2.1 No injection
68
0060.170-11 Optimas manual
69
Optimas manual 0060.170-11
70
0060.170-11 Optimas manual
10 Installation
Unpacking
Inspect the OPTIMAS for signs of damage. Damage that occurs to the OPTIMAS
as in transit, indicated by damaged containers, is the responsibility of the carrier
and should be reported to the carrier immediately. Shipping containers should be
inspected by the carrier if a claim is filed.
For contents of shipping containers refer to the packing list in the container.
Lift the OPTIMAS with both hands under the instrument or with one hand under the
front while you hold the top rear of OPTIMAS with your other hand.
71
Optimas manual 0060.170-11
Power connections
Before plugging in the power cable, check voltage setting of the OPTIMAS at the
input socket on the rear-panel. Make sure that the voltage setting is identical with
the voltage of your local power supply.
If the indicated voltage is not correct select the proper voltage by removing,
inverting, and then re-entering the voltage selector-cartridge.
Check if the right fuses are installed, if not replace them with the right fuses.
RISK OF FIRE,
REPLACE FUSES AS MARKED.
Location
The OPTIMAS needs approximately 30 cm (12 inches) of bench space and one
power supply (230 VAC or 115 VAC, 50/60 Hz).
The best place to put your OPTIMAS is at the left-hand side of your HPLC system,
next to your HPLC pump and detector. From the right-hand side you have the
shortest connection to the injection valve of the OPTIMAS.
72
0060.170-11 Optimas manual
Installation steps
Execute the following steps to install the OPTIMAS (if fluid connections have been
created already):
1. Place the OPTIMAS in the desired location and leave it there to adopt ambient
temperature for at least one hour.
2. Check OPTIMAS for visual damage; if this is the case, please contact your
distributor immediately.
4. Remove the safety screw on the right-hand side of the front cover.
5. Check fuses and voltage settings on the rear side of the instrument.
7. Switch on the OPTIMAS. The display shows the firmware revision number.
8. Connect the drain tubing to the wash outlet (figure 3, number 5).
9. Fill the wash solvent bottle with distilled water/ 2-Propanol (80/20 v/v%) or
mobile phase. Only use water/organic solvents; do not use any crystalline or
buffer solutions, as these may cause damage to the system. Degas wash
solvents to prevent air bubbles in the syringe.
10. Fill the wash solvent tubing, syringe and buffer tubing by washing the system 2
or 3 times: press <WASH> in the Ready screen.
11. Check if there are any air bubbles trapped in the syringe. Remove them by
gently tapping the syringe.
13. Connect the column (or capillary) to port 6 of the injection valve.
14. Check for leakage and let the system equilibrate for at least 5 minutes.
73
Optimas manual 0060.170-11
The OPTIMAS is factory installed with a 250 µL syringe, a 100 µL loop, a 500 µL
buffer tubing and a stainless steel sample needle.
A schematic presentation of all parts that have contact with any kind of fluid is
shown below. This sticker can be found on the inside of the front cover.
74
0060.170-11 Optimas manual
75
Optimas manual 0060.170-11
Syringe waste
Place the syringe waste tubing into a small bottle next to the OPTIMAS. If no
injection volumes are programmed that are larger than the buffer tubing can
contain, the syringe waste will only be wash solvent.
1. Put the wash solvent tubing in the filled wash solvent bottle.
3. Fill the tubing using the OPTIMAS soft-function keys <SYR END> and <SYR
HOME>.
With <SYR END> a syringe volume of wash solvent is aspirated from the wash
solvent bottle and the wash solvent tubing is filled.
With <SYR HOME> the syringe contents will be dispensed to the syringe
waste.
4. Repeat this action until the wash solvent tubing and the syringe are completely
filled.
76
0060.170-11 Optimas manual
After the wash solvent tubing and the syringe are filled, press <WASH> to perform
a standard wash routine. All tubing connected to the syringe valve will be rinsed
with wash solvent.
10.1.4 Syringe
The OPTIMAS is factory equipped with a 250 µL syringe, but can also be equipped
with a 1000 µL syringe. See section 7.4 for more information on how to replace the
syringe.
The OPTIMAS will give the best results if all air is removed
from the syringe.
Use <WASH> to remove air from the syringe, if this is not
satisfactory, see section 7.4 for more tips on removing air.
The OPTIMAS is standard equipped with a tray which can contain 84 standard
sample and three large volume vials for reagents and transport liquid.
The following section contains examples of Chromacol vials witch can be used in
the OPTIMAS. See appendix A, List of accessories, for the corresponding article
numbers for the vials, seals and caps.
77
Optimas manual 0060.170-11
78
0060.170-11 Optimas manual
To reduce the volume of standard vials, use a Chromacol insert (02–MTV, 02–
MTVWG or 03–MTV) in combination with the appropriate vial and support sleeve or
spring.
79
Optimas manual 0060.170-11
80
0060.170-11 Optimas manual
Do not fill vials to the edge to prevent that sample is forced into the air needle,
risking extra cross-contamination of samples and fouling of the needle pair.
It is important that the seal is airtight, to maintain a pressure on the vial for air
bubble prevention and to prevent evaporation of volatile samples.
Check the seal after crimping: if the cap can be turned easily, the seal is not airtight
(re-adjust the hand–crimper).
81
Optimas manual 0060.170-11
82
0060.170-11 Optimas manual
From the Ready screen a Next Injection Input will start the latest programmed
Series. In this case the OPTIMAS will not wait for the Next Injection Input before
continuing with the next injection. The OPTIMAS will execute the complete Run as
if it was started with the [Start/Stop] key.
Freeze input:
The OPTIMAS will freeze the analysis time for the time this input is active. If the
Freeze Input is activated while the analysis time is not running, the OPTIMAS will
perform all programmed pre-injection sample handling (mix method and sample
loop). But the OPTIMAS will wait with injecting the sample until the Freeze Input is
no longer active.
Stop Input:
With this input the run of the OPTIMAS is immediately aborted and return to the
Ready screen. In case the OPTIMAS is in Remote Control, the Run of the
OPTIMAS is immediately aborted but the OPTIMAS remains in Remote Control
and can not be re-started with a Next Injection Input.
83
Optimas manual 0060.170-11
84
0060.170-11 Optimas manual
85
Optimas manual 0060.170-11
86
0060.170-11 Optimas manual
Injection valve
ERROR 211 Injection valve is not in a valid position.
ERROR 212 The injection valve did not switch within 1.5 seconds.
ERROR 213 The switching time of the injection valve exceeds 500
msec
Tray unit
ERROR 251 Incorrect tray rotation
ERROR 252 No sample tray positioned in the OPTIMAS.
ERROR 253 The sample needle arm is not in the home position
while rotating the tray.
Vials
ERROR 260 Missing vial. Only available when Skip
Missing Vial is set to NO in the
System Settings (see section )
and during the execution of the
87
Optimas manual 0060.170-11
Mix of a method.
ERROR 262 Missing transport vial. Only available when Skip
Missing Vial is set to NO in the
System Settings (see section ).
ERROR 264 Missing vial for reagent A
ERROR 265 Missing vial for reagent B
ERROR 268 Missing destination vial.
ERROR 269 Not enough transport liquid available due to missing
transport vials.
ERROR 270 Not enough reagent A available.
ERROR 271 Not enough reagent B available.
Electronics
ERROR 275 Error occurred during initialization, the OPTIMAS
cannot start.
88
0060.170-11 Optimas manual
The OPTIMAS is factory tested for reproducibility, carry-over and mixing according
the following test procedure.
Analytical parameters:
Pump: Flow: mL/min
Eluent: distilled water
Detector: Wave length: 254 nm
Sample: Uracil in distilled water (50 ppm)
89
Optimas manual 0060.170-11
Peak Area
number
1 206342
2 204773
3 205425
4 -
5 -
6 -
7 205203
8 205566
9 205568
10 206156
11 206174
12 205956
13 206165
14 205765
14 205960
90
0060.170-11 Optimas manual
16 206296
17 205617
18 205619
19 205648
20 206141
21 207070
From the integration results the Relative Standard Deviation (RSD%) and the
carry-over can be calculated. For this example the results are:
RSD% = 0.25%
Carry over < 0.5%
Mix test
For the mix test one sample vial (50 ppm Uracil), one empty vial and the full
Reagent-A vial, filled with eluent, are placed in the OPTIMAS tray. A 1:10 dilution
of the sample will be made in the empty destination vial. After the dilution from both
vials (sample and destination) 10 µL will be injected. The wash solvent bottle is
filled with 80% H2O / 20 % Isopropanol.
Method 2:
Injection
Injection mode: Partial loopfill
Flush volume: 30 µL
No of inj./vial: 3
Injection volume: 10 µL
91
Optimas manual 0060.170-11
From the integration results the Relative Standard Deviation (RSD%) for the two
vials can be calculated.
Peak Area
number
1 38802
2 38968
3 39088
4 395258
5 394090
6 394471
For this example the results are:
RSD% (vial 1, 1:10 dilution) = 0.37%
RSD% (vial 2, sample) = 0.15%
1:10 dilution = 1 : 10.13
92
0060.170-11 Optimas manual
By just choosing PREP in your system settings, you can use the OPTIMAS to inject
all of your sample into a Preparative LC system or in other areas where large
injection volumes are required. Also in combination with the PROSPEKT On-line
Solid Phase Extraction System, the OPTIMAS can be your ideal autosampler.
The combination of large sample vials (10 mL) a large sample volume needle and
a 2.5 mL syringe enable you to inject large volumes very reproducible with high
speeds and only 45 µL of sample loss. The installed large bore valve (0.75 mm)
with 10 mL sample loop enables you to inject from microliters to milliliters with the
same autosampler. Flow rates up to 200 mL per minute are no problem for the
OPTIMAS autosampler, used in Prep mode.
Specifications
The Specifications of the Prep option differ in some ways from the standard
autosampler. The different specifications are listed below.
Sampling
Sample capacity LSV tray: 24 vials of 10 mL (LSV)
Loop volume Not programmable, injection volume
determines the aspirated sample volume
Dispenser syringe 2500 µL syringe
Analytical performance
Reproducibility RSD <= 1.0% for partial loopfill injections,
injection volumes >10 µL up to 50% of the
installed sample loop.
Memory effect < 0.1% with programmable needle wash
Programming
Methods 9 programmable methods
Injection method Partial loopfill only
Injection volume 1 µL – 19,999 µL, with 1 µL increment
Mix Not available
93
Optimas manual 0060.170-11
Option
Bio-compatible valve Not available
Installation
To install the Prep option on the standard autosampler perform the following
installation steps:
1. Replace the standard injection valve with the special Prep valve (VALCO #C2-
3006 SPHMI).
2. Replace standard sample needle, air needle and buffer tubing with the ones
supplied in the Prep kit. See section 7.3 for detailed information.
4. Replace the standard syringe with the 2500 µL syringe; see section 7.4 for
detailed information.
Injections
The injection routine differs from the standard autosampler. The Prep version has
only the Partial loopfill injection routine, which differs from the partial loopfill
injection routine of the standard autosampler.
Therefore note the following for the best injection performance:
94
0060.170-11 Optimas manual
The software does not check the injection volume with the
installed loop volume. It is therefore possible to program
injection volumes larger than 50% of the loop volume.
95
Optimas manual 0060.170-11
96
0060.170-11 Optimas manual
Customer information:
Name of user
Company
Department
Address
Telephone
Fax
OPTIMAS information:
Serial number
Firmware revision
Purchase date
Installed options
Local Dealer
Service Engineer
Address
Telephone
Fax
E-mail
97
Optimas manual 0060.170-11
General
Loop volume: µL
Needle volume: µL
Syringe volume: 250-µL syringe
1000-µL syringe
Syringe speed: Low
Normal
High
Skip missing vials: Yes
No
Air segment: Yes
No
Headspace pressure: Yes
No
Method
Use calibration vials: Enabled
Disabled
Tray
Tray type: <84+3>, standard tray.
<96>, tray with 96 standard vials
<24>, large volume tray.
Vial type: <STANDARD>, standard 1.8 mL vials.
(only when <84+3> or <96> is selected) <2.5 mL>, 2.5 mL vials.
I/O configuration
Inject marker pulse length: sec.
Input edge next injection: Falling
Rising
Freeze input active: Low
High
Reset outputs after last Series: Yes
No
Prep Mode
Use Prep Mode: Enabled
Disabled
98
0060.170-11 Optimas manual
No:
INJECTION WASH
Injection Type: Full loop Wash: between injections
Partial loopfill between vials
µL Pick-up between Series
None None
Flush volume: µL Wash volume: X syringe volume
Injection / vial:
Injection µL
volume:
Analysis time:
TIMED EVENTS
1. AUX-1 ON AT TIME:
1. AUX-1 OFF AT TIME:
2. AUX-1 ON AT TIME:
2. AUX-1 OFF AT TIME:
3. AUX-1 ON AT TIME:
3. AUX-1 OFF AT TIME:
4. AUX-1 ON AT TIME:
4. AUX-1 OFF AT TIME:
1. AUX-2 ON AT TIME:
1. AUX-2 OFF AT TIME:
2. AUX-2 ON AT TIME:
2. AUX-2 OFF AT TIME:
3. AUX-2 ON AT TIME:
3. AUX-2 OFF AT TIME:
4. AUX-2 ON AT TIME:
4. AUX-2 OFF AT TIME:
END OF TIMED EVENTS AT:
MIX
1.
2.
3.
4.
5.
99
Optimas manual 0060.170-11
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
100
0060.170-11 Optimas manual
2. Disconnect and remove all tubing, except for the sample loop.
4. In case of biohazard materials: remove and throw away the needle and other
flow path materials, in accordance with an approved waste disposal program.
6. Fill in a health & safety form for the instrument and include the form in the
package.
7. Dispatch the package; make sure any tax/import/export requirements are met.
Contact your local distributor if you need more information.
Decontamination
The instrument shall be decontaminated before decommissioning and all local
regulations shall be followed with regard to scrapping of the equipment.
Hazardous substances
The instrument contains hazardous substances. Contact the manufacturer for more
detailed information.
101
Optimas manual 0060.170-11
102
0060.170-11 Optimas manual
E
Emergency sample • 50
Index
Example - 10 uL Partial loopfill injection from
vial number 1 • 26
Example - 3 x 10 µL pickup injections • 27
[
Example - a 1
[Hold Cont.] • 50
10 dilution + 10 µL partial loopfill injection •
[Priority] • 50
29
[Start/Stop] • 48
F
<
Filling and sealing the vials • 81
<GENERAL> • 36
Fluid connections • 74
<INJECTION> • 41
Full loop injections • 17
<IO> • 39
H
<MAINTENANCE> • 34
Hold analysis time • 50
<METHOD> • 38
HPLC connections • 75
<MIX> • 43
I
<PREP> • 39
Identification of parts • 11
<SERVICE> • 35
Injection principles • 17
<TEMP> • 34
Injection valve and rotor seal • 53
<TIMED EVENTS> • 45
Injections • 94
<TRAY> • 38
Installation • 71, 94
<UTILS> • 35
Introduction • 9
<WASH> • 33, 42
K
µ
Keyboard and display • 14
µL pickup injections • 22
L
A
Loading the sample tray • 81
About this manual • 5
M
Air needle replacement • 58
Maintenance • 53
Analytical problems • 67
Method log sample • 99
Appendix A. Spares and accessories • 85
Method programming mode • 40
Appendix B. Error codes • 87
N
Appendix C. Test procedure • 89
Needle and tubing • 54
Appendix D. Prep option • 93
No injection • 68
Appendix E. Log book • 97
P
Appendix F. Storage, packaging, recycling
Partial loopfill injections • 19
and disposal • 101
Pictorials • 5
B
Preparation of vials • 77
Bad reproducibility • 69
Priority sample • 50
C
Programming during the run • 51
Contact closures output (P1) • 81
Programming overview • 31
Continue analysis time • 50
Programming reference • 31
Control I/O connections • 81
R
Ready screen options • 33
103
Optimas manual 0060.170-11
Remote control • 51
Rotor seal • 53
Run options • 48
S
Safety instructions • 6
Sample needle • 57
Sample needle height setting • 58
Sample needle replacement • 58
Series programming mode • 47
Specifications • 61, 93
Start up problems • 66
Syringe • 59, 77
System settings programming • 36
System settings sample • 98
T
Trouble shooting • 65
TTL input (P2) • 82
Types of vials • 77
Typographical conventions • 8
U
Using the OPTIMAS • 25
W
Wash solvent and system rinse • 76
Waste tubing • 75
104
Spark Holland B.V.
P.O. Box 388
7800 AJ Emmen
The Netherlands
Spark Holland -
Main office and factory:
P. de Keyserstraat 8
7825 VE Emmen
The Netherlands
Phone: +31 591 631700
Fax: +31 591 630035
E-mail: sales@sparkholland.com