DINAMAP™ Compact Vital Signs Monitor

6.5 Procedural & Error Alarm Code Table

Alarm LED LCD Display Audio Tone & Effect of Effect of Probable Cause
Code Display Volume Alarm Clear
Silence via
switch Rotor
N99 No N99-NIBP Unable to make an NIBP determination
change FAILED due to insufficient signal.
N55
N55- High priority 2 minutes Clear One cuff pressure for > 1 minute. Motion
N44 TIMEOUT: alarm. Volume silence artifact.
PRESS adjustable.
N33 Determination time exceeds 2 minutes.
N44- Motion artifact.
N00 TIMEOUT:
TOTAL Inflation time > 40 seconds or air leak
detected.
N33-
TIMEOUT: Overpressure detected.
INFLATE

N00- OVER
PRESSURE

N12 Systolic N0 Display Steady tone, No effect No effect System pressure whilst idle 5 or
Display maximum 15mmHg above ambient for 30 seconds
‘N12’ volume for a neonate or an adult cuff
respectively
N13 Systolic No Display Steady tone, No effect No effect Pump on/off status determined by
Display maximum measurement of pump current does not
‘N13’ volume concur with Main board software
expected status. Unit will power down
within 30 seconds of detected failure..
P55 P55-SpO2: NO High priority No or very low SpO2 signal. Check or
No SIGNAL alarm. Volume 2 minutes Clear reposition sensor.
P00 change adjustable. silence
P00-SpO2: NO SpO2 sensor not connected. No sensor
SpO2 code detected. Sensor failure.
SENSOR

E33 E33-Temp: Clear Temperature probe not connected or


FAIL High priority 2 minutes inoperable.
E11 No alarm. Volume silence Clear Predictive temperature loss of tissue
change E11-Temp: adjustable. contact > 30 sec.
E00 FAIL Clear Removal of probe before temperature
determination complete.
E00-Temp:
FAIL Predictive temperature determination
exceeds 60 sec.

75
DINAMAP™ Compact Vital Signs Monitor
Procedural & Error Alarm Code Table continued

Alarm LED LCD Display Audio Tone & Effect of Effect of Probable Cause
Code Display Volume Alarm Clear via
Silence Rotor
switch
No Low Battery 3 beeps every 2 minutes No effect Replace or recharge battery. From
change Flashing 10 sec, silence onset of alarm, 5 NIBP
battery icon adjustable measurements available. Beep
Blank volume. No effect No effect linearly increases rate as battery
No code Low Battery - discharges.
System Steady tone,
No Disabled maximum 2 minute Clear Replace or recharge battery. NIBP
change volume silence measurement disabled.
Printer - No
Blank Paper High priority 2 minute Clear
alarm silence Paper expired or printer door open.
Volume
NIBP Range adjustable
error NIBP algorithm returned value
High priority outside specified accuracy rage.
alarm
Volume
adjustable
Other: Blank Error code, Steady tone, No effect No effect Internal system fault
N,P,E,I,S. description maximum
volume