You are on page 1of 8

Energy Dispersive X-ray

Fluorescence Spectrometer
(EDX-2)

Brochure

Global Office of Lisun Electronics Inc.


http://www.Lisungroup.com

Lisun Group (Hong Kong) Limited


Add: Room 803, Chevalier House, 45-51 Chatham Road South, Tsim Sha Tsui, KL, HK
Tel: 00852-68852050 Fax: 00852-30785638
Email: SalesHK@Lisungroup.com

Lisun Electronics (Shanghai) Co., Ltd


Add: 113-114, No. 1 Building, Nanxiang Zhidi Industry Park, No. 1101, Huyi Road, Jiading
District, Shanghai, 201802, China
Tel: +86(21)5108 3341 Fax: +86(21)5108 3342
Email: SalesSH@Lisungroup.com

Lisun Electronics Inc. (USA)


Add: 445 S. Figueroa Street, Los Angeless, CA 90071, U.S.A.
Email: Sales@Lisungroup.com

Lisun China Factory


Add: NO. 37, Xiangyuan Road, Hangzhou City, Zhejiang Province, China
Tel: +86-189-1799-6096
Email: Engineering@Lisungroup.com

Leader in Lighting & Electrical Test Instruments


Rev. 10/18/2019
Lisun Electronics Inc. Sales@Lisungroup.com www.Lisungroup.com

The quotation includes the following items:

For the purpose of restricting the pollution caused by the dispose of electronic and
electrical products, the RoHS or Lead Free directive was announced by EU on Feb
13th of 2003 and became effective on Jul 1st of 2007. Lisun Group developed an
high accuracy EDXRF spectrometer. The EDX-2 can fully meet the RoHS test
requirements

-Page 1 -
Lisun Electronics Inc. Sales@Lisungroup.com www.Lisungroup.com

Part 1 Specification:
 Nondestructive measurement, no chemical and physical damage on measured
object
 High accuracy which can be similar as chemical test
 Detect and analyze more than 75 chemical elements from Sulfur to Uranium,
test Range: 2ppm~99.99%
 Measure the details for the 5 restricted elements in RoHS Directive:
Cd/Pb/Hg/Cr/Br)
 esting time: 60~300s
 Adopt world-famous electrical cooled detector, reliable and stable, energy
resolving power is less than 160eV
 X-ray tube voltage: 5~50kV; Tube current: 50~1000uA
 High definition camera built-in; USA made high pressure module built-in; Light
enhancement system
 SNR design enhanced, quick test circuit and precise numbering system
 Multi-function software and multi-substrate effect calibration module built-in
 The system includes a PC and printer which have already installed the English
software

Part 2 Installation
2.1 Structure
2.1.1 Front view:

①: Cover
②: Strut Bar
③: Control Column
④: Sample Hole
⑤: Body

-Page 2 -
Lisun Electronics Inc. Sales@Lisungroup.com www.Lisungroup.com

⑥: Power
⑦: Indicator
⑧: Detector
⑨: Collimator
2.1.2 Rear view

①: Fan
②: RS-232C port
③: USB port
④: Outlet

2.2 Installation
2.2.1 Preparation for operation. The RoHS test system needs enough space as
follows:

-Page 3 -
Lisun Electronics Inc. Sales@Lisungroup.com www.Lisungroup.com

2.2.2 Connection

-Page 4 -
Lisun Electronics Inc. Sales@Lisungroup.com www.Lisungroup.com

Part 3 Operation
3.1 RoHS testing process

START

CHECK
CONNECTION

FIRST TEST NO PREHEAT 30min

YES

SET 200s NO TESTING


TIME 200s

YES

NO PUT IN SILVER TO CALIBRATE


INITIALIZATION?

YES

CHOOSE CURVE

CHOOSE COLILIMATOR

TEST

START TO
YES PRINT
PRINT

NO

END

-Page 5 -
Lisun Electronics Inc. Sales@Lisungroup.com www.Lisungroup.com

3.2 Preparation
Before switching on EDX, check all the connections. Then power on PC, printer,
EDX-2, and preheat 30min.Initiallize EDX-2, put in silver slice, choose 1# collimator,

click” ”, system calibrate automatically.

3.3 Sample test


1. According element needed of sample, choose the right rohs testing mode
(choose the wrong mode lead to a bad accuracy)
2. Put in matching collimator (choose the wrong collimator lead to a bad
accuracy)
3. Choose testing time, 200s is recommended(too long or too short lead a
bad accuracy)
4. Put the unknown sample into the testing bed, observe the cross of
camera is in the center of unknown sample, or you can adjust it by
moving the sample and control column.
5. Close the cover, the protective light on
6. Click “Start”
7. After testing, you can save the data and name
8. If need to print, click ”Total analysis report”, then go to ”report info
setting”

Notice:
 EDX-2 need to calibrate after switching on every time
 Wrong mode and collimator and too short RoHS test time lead to additional
error.
 Too small unknown sample or not in the center of he cross of camera lead error.

 Next page is the test report

-Page 6 -
Lisun Electronics Inc.
www.Lisungroup.com
Tel:+86(21)51083342

ANALYSIS REPORT FOR RoHS REGULATIONS


Sample Name:CFL Tube Sample Batch:No.1250
Sample Source:TUV Operator:Jacky Jiang
Analysis institution:Lisun EDX-2 Analysis Date:2014-03-14
Analysis Duration:200s Tube Voltage(KV): 45kV
Tube Current(uA):400uA Work Curve Used:PE

Y-计计
4500
4360

4000

3500

3000

2500

2000

1500

1000

500
2048

Hg
As BrPb Cd
0
0 200 400 600 800 1000 1200 1400 1600 1800 2000 X-通通
Element Intensity Content(ppm) Limitation Results
As 0.0020 28.02 1000 Pass
Hg 0.0014 25.26 1000 Pass
Pb 0.0054 94.04 1000 Pass
Br 0.1454 538.68 1000 Pass
Cd 0.0024 121.97 100 False

Note:1)ND stands for contents not larger than 2ppm.


2)All Data are based on sample surface analysis with this X-ray fluorescence Instrument.
3)The Data for Cr&Br are the summation contents respective, their above-limitation
doesn't mean the same thing with VI-valence Cr,PBB and PBDE.

You might also like