Energy Dispersive X-Ray Fluorescence Spectrometer EDX-2: Leader in Lighting & Electrical Test Instruments
Energy Dispersive X-Ray Fluorescence Spectrometer EDX-2: Leader in Lighting & Electrical Test Instruments
Fluorescence Spectrometer
(EDX-2)
Brochure
For the purpose of restricting the pollution caused by the dispose of electronic and
electrical products, the RoHS or Lead Free directive was announced by EU on Feb
13th of 2003 and became effective on Jul 1st of 2007. Lisun Group developed an
high accuracy EDXRF spectrometer. The EDX-2 can fully meet the RoHS test
requirements
-Page 1 -
Lisun Electronics Inc. Sales@Lisungroup.com www.Lisungroup.com
Part 1 Specification:
Nondestructive measurement, no chemical and physical damage on measured
object
High accuracy which can be similar as chemical test
Detect and analyze more than 75 chemical elements from Sulfur to Uranium,
test Range: 2ppm~99.99%
Measure the details for the 5 restricted elements in RoHS Directive:
Cd/Pb/Hg/Cr/Br)
esting time: 60~300s
Adopt world-famous electrical cooled detector, reliable and stable, energy
resolving power is less than 160eV
X-ray tube voltage: 5~50kV; Tube current: 50~1000uA
High definition camera built-in; USA made high pressure module built-in; Light
enhancement system
SNR design enhanced, quick test circuit and precise numbering system
Multi-function software and multi-substrate effect calibration module built-in
The system includes a PC and printer which have already installed the English
software
Part 2 Installation
2.1 Structure
2.1.1 Front view:
①: Cover
②: Strut Bar
③: Control Column
④: Sample Hole
⑤: Body
-Page 2 -
Lisun Electronics Inc. Sales@Lisungroup.com www.Lisungroup.com
⑥: Power
⑦: Indicator
⑧: Detector
⑨: Collimator
2.1.2 Rear view
①: Fan
②: RS-232C port
③: USB port
④: Outlet
2.2 Installation
2.2.1 Preparation for operation. The RoHS test system needs enough space as
follows:
-Page 3 -
Lisun Electronics Inc. Sales@Lisungroup.com www.Lisungroup.com
2.2.2 Connection
-Page 4 -
Lisun Electronics Inc. Sales@Lisungroup.com www.Lisungroup.com
Part 3 Operation
3.1 RoHS testing process
START
CHECK
CONNECTION
YES
YES
YES
CHOOSE CURVE
CHOOSE COLILIMATOR
TEST
START TO
YES PRINT
PRINT
NO
END
-Page 5 -
Lisun Electronics Inc. Sales@Lisungroup.com www.Lisungroup.com
3.2 Preparation
Before switching on EDX, check all the connections. Then power on PC, printer,
EDX-2, and preheat 30min.Initiallize EDX-2, put in silver slice, choose 1# collimator,
Notice:
EDX-2 need to calibrate after switching on every time
Wrong mode and collimator and too short RoHS test time lead to additional
error.
Too small unknown sample or not in the center of he cross of camera lead error.
-Page 6 -
Lisun Electronics Inc.
www.Lisungroup.com
Tel:+86(21)51083342
Y-计计
4500
4360
4000
3500
3000
2500
2000
1500
1000
500
2048
Hg
As BrPb Cd
0
0 200 400 600 800 1000 1200 1400 1600 1800 2000 X-通通
Element Intensity Content(ppm) Limitation Results
As 0.0020 28.02 1000 Pass
Hg 0.0014 25.26 1000 Pass
Pb 0.0054 94.04 1000 Pass
Br 0.1454 538.68 1000 Pass
Cd 0.0024 121.97 100 False