B.11 Set Reagent Rack and Reaction Tube Rack ........................................................................................... B-13
Based on the following check list, check the packings before installation.
To make maintenance and service easy, give consideration to heat radiation by the instrument, provide at least
50 cm clearance between the wall and the instrument’s side panels, rear and top panels.
Table B-2:
Table B-3:
487
540
470
CAUTION: • Be sure to place the rinse bottle and waste bottle on the base which the instrument
is set.
• Do not put them on the instrument, or there is a possibility the instrument may
break down or fail to produce correct results.
CAUTION: • Unless the fixing metals are removed, the instrument cannot operate.
(4) Raise the sample probe by hand to a place where it is easy to operate.
Remove the catcher retainer.
Catcher Retainer
WARNING! • When draining the trap chamber, always wear rubber gloves.
• After completing the operation, be sure to wash hands in anti-septic
solution.
• If hands should be contaminated with blood or the like, there is a hazard of
being infected by pathogenic bacteria.
CAUTION: • Even at a facility equipped with the waste channel (drain system),
the waste bottle should be connected.
• Also, put the rinse bottle and the waste bottle at the same level as
the instrument. Be sure not to use any other tube than the
furnished one; otherwise, the instrument's hydraulic system may fail
to operate properly.
CAUTION: • Remove the rubber tube that locks the float switch in the rinse bottle
and waste bottle. This rubber tube served to prevent vibration in
transit.
Rubber tube
7 Power Connector
AC outlet
CAUTION: Confirm the power switch is OFF, at "O," before routing the power cord.
Make sure to ground the AC outlet; otherwise, there is a hazard of electrical
shock.
CAUTION: Confirm the power switch is OFF, at "O," before routing the connection cord;
otherwise, there is a hazard of electrical shock.
NOTE: • For setting host computer interface parameters, refer to Operator’s Manual
Chapter 10: Section 5.1: Host Computer.
• The connection cord for the host computer is not included in the
accessories.
Printer Cover
Lock Lever
Print Paper
Lock Lever
Ready HC IP
Replace Rack? YES!
Sysmex
Error List
Temperature
P. FEED
Main
Menu
Light
Ready HC IP
Sysmex Replace Rack? YES!
Special Operation
System Tests
Cycle Counter
Main
Menu
(3) Press [Rinse & Prepare] key on the Special Operation Menu screen.
The Supply Liquid Confirmation screen will display the message "Water supply?"
Ready HC IP
Sysmex R e p la c e R a c k ? YE S !
S up p l y L i q u i d
Wa t e r s u p p l y ?
Set Cancel
(4) Press [Set] key or [Cancel] key on the Supply Liquid Confirmation screen.
[Set] key: Executes water supply.
[Cancel]: Cancels water supply and returns the screen to the Root Menu.
NOTE: Filling the hydraulic line with rinse solution takes 35 sec or so.
Ready HC IP
Sysmex R e p la c e R a c k ? Y E S !
S u p p l y L i q u id
W a t e r s u p p ly r u n n in g
Cancel
Reagent Rack
Figure B-19: Setting the Reagent Rack and Reaction Tube Rack
(1) Remove the left side panel of the CA-500 main unit.
Loosen the screws as shown to remove the panel.
Panel
Duct Cover
3) Pull out the connector from the duct, and disconnect the connector.
Connector
Lever
(3) Attach the slide rail to the sampler with the ID bar code scanner.
1) Remove the slide rail from the removed sampler.
Loosen the screws as shown and remove the slide rail from the sampler.
Take care not to lose the slide rail, screws, or washers that were removed, as they
must be attached to the sampler with the ID bar code scanner.
Screws Screws
2) Attach the slide rail to the sampler with the ID bar code scanner.
Mount the duct-attached slide rail to the sampler onto the side where the cable
protrudes. All four screws should be temporarily tightened.
3) Insert the cable from the sampler into the slide rail duct.
Inner Rail
CAUTION: Until you are sure the sampler with ID bar code scanner will not
disconnect, do not let go hands.
CAUTION: The table lock is only 1 mm apart from the lock guide
arms. In pushing, take care not to allow the lock guide
arm to contact the under panel.
Table Lock
Lock Guide
Arm
Connector
Duct Cover
5) Repeat pushing in and pulling out the sampler with ID bar code scanner several times.
6) With the sampler pulled out in parallel, fully tighten the screws that were temporarily
tightened before.
7) Make sure that the sampler with ID bar code scanner will slide in and out smoothly.
(6) Attach the left side panel of the CA-500 main unit.