Professional Documents
Culture Documents
H t
Fig. 2. Addition of a time varying homogeneous offset field ( ) to a gradient
field 0Gx x t
causes a shifting of the field free point ( ). For a gradient 0G ,
x t G H t
the location of the FFP can be solved for as ( ) = ( ).
(1)
Fig. 3. Simulated MPI 1-D image of a complex phantom (solid line) and the
Fundamentally, MPI relies on the rapid movement of a field free source distribution (dotted line).
point (FFP) across the sample to elicit a signal. We can describe
the location of the field free point by solving the above equation
for . This yields where the , , and axes correspond to the physical , , and
axes. This tells us the total magnetization in the system is a
convolution of the magnetic particle density with a Langevin
function kernel. This assumes that the magnetic nanoparticles
Substituting FFP position into (1), the magnetic field at
do not change the H field significantly. This is generally true
position can be rewritten
since the average of magnetite at physiologically well toler-
ated concentrations [1] is small [6].
However, the point spread function (PSF) of a 1-D MPI
The magnetization of magnetic nanoparticles in response to system is not the Langevin function. We typically receive the
an applied magnetic field in one dimension can be described by MPI signal using an inductive detector that sees changing mag-
the Langevin equation (see Section III) netic fields with sensitivity [7]. Visually described
in Fig. 1, we obtain
MPI 1-D signal equation in volts
Fig. 4. (Top) The Langevin function, [kH ]. (Bottom) The derivative of the
Langevin function _ [kH ]. The derivative of the Langevin function is analyti- Fig. 5. Intrinsic MPI resolution for various SPIO nanoparticle diameters when
cally FWHM 4:16. imaged using different gradient field strengths.
TABLE I
VALUES OF k FOR VARIOUS DIAMETERS are separated by the full-width at half-maximum (FWHM) [10].
OF SPHERICAL SPIO NANOPARTICLES We use this classical criterion in this paper. We also note that
super-resolution methods exist for increasing image resolution
beyond the classical limit at the expense of SNR [11].
From the image (3), we see that the point spread function
of the MPI process is the derivative of the Langevin function.
Solving for the FWHM of this well behaved function, we can
estimate the intrinsic resolution of the MPI process analytically
as . This gives [4]
of applied field is
where
which shows that system resolution can be improved by in-
creasing the diameter of the magnetic nanoparticles , or by in-
creasing gradient strength . We can then calculate the intrinsic
and is Boltzmann’s constant, is the temperature, and is resolution of the 1-D MPI process as a function of particle size
the vacuum permeability. We note that gives and gradient strength (Fig. 5 and Table II). Because spatial res-
the magnetic moment of a spherical, single domain magnetic olution improves with the cube of magnetic nanoparticle diam-
nanoparticle [4]. For magnetite SPIO nanoparticles, eter, it may be possible to use a large particle with resistive mag-
. nets. This would even enable adjusting the intrinsic resolution
The derivative of the Langevin function is a well behaved of the system in real time by simply changing the strength of
even function the gradient. Note that operating at reduced temperatures, such
as 77 K or 5 K, could dramatically increase and the resulting
resolution for preserved histological samples, provided we do
not go below the blocking temperature which would causes the
The operand of the Langevin function, , is a dimensionless SPIO particles to behave ferromagnetically.
value. Scalar has units of [m/A] and example values are shown
in Table I. V. BANDWIDTH
To design a MPI system, we must know the bandwidth re-
IV. SPATIAL RESOLUTION quirements to achieve the desired resolution. We can approxi-
Spatial resolution of the MPI system can be thought of as the mate the bandwidth required to represent the Langevin function
ability to accurately depict two distinct points of equal intensity derivative through Fourier analysis.
in space [9]. There are many classical criteria for resolution in- We found earlier that the derivative of the Langevin function
cluding Rayleigh, Schuster, Houston, and Buxton. Houston pro- is our PSF and defines the intrinsic resolution of the 1-D MPI
posed a criterion where two points are just resolved if two points process. The derivative of the Langevin function does not have
1854 IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 29, NO. 11, NOVEMBER 2010
TABLE II
INTRINSIC RESOLUTION FOR GIVEN SPIO NANOPARTICLE SIZES AND
GRADIENT STRENGTHS. LARGER CORE DIAMETERS ENABLE SIGNIFICANTLY
IMPROVED RESOLUTION. A 5:1 T=m= MAGNETIC FIELD GRADIENT IS
REASONABLE TO BUILD FOR SMALL ANIMALS
This lets us introduce what we term the magnetic field slew rate,
[A/m/s], which is the product of the scanning rate and the
gradient size . We will also find that governs a number of
important parameters including bandwidth requirement, SAR, , ) results in a . We
and magnetostimulation. can see the relationship between the particle size, the intrinsic
For particles located at the origin, resolution, and the bandwidth in Table III. The particle
, the 1-D signal (2) and substituting a Lorentzian size strongly controls the resolution and required bandwidth.
for the derivative of the Langevin curve gives
VI. RESOLUTION-BANDWIDTH TRADEOFFS
The finite receiver bandwidth of a real system reduces the
bandwidth from the intrinsic resolution to what we term the
Taking the Fourier transform of yields achievable resolution. As a result, we can trade off receive band-
width, achievable resolution, and SNR. To see this, let us assume
a receive bandwidth, , with a brick-wall filter. Then, the re-
ceived signal in Fourier and real space is
Fig. 7. MPI F bandwidth and maximum resolution as a function of multiples of the F bandwidth (a) Linear relationship between F bandwidth for
various particle sizes and magnetic field slew rate. Note that the magnetic field slew rates are much faster than those seen in MRI and are given in T=ms= (b)
Relationship between F bandwidth and resolution assuming a brick-wall receive filter. The achievable resolution of the MPI process is 150% of the intrinsic
4
resolution when the receiver bandwidth is f 2:2F . We do not reach 110% of the intrinsic resolution until f 4 3:8F .
VII. SIGNAL-TO-NOISE RATIO SNR, which we define as the peak signal divided by the stan-
dard deviation of the noise, is
We can calculate the signal and noise from first principles.
We note that our primary assumption is that the system acts
quasi-statically since the SNR may be significantly less if we ac-
count for particle relaxation. From the MPI Signal (2), the peak
signal for a linearly ramping magnetic field for a point source
of particles with total magnetization
is
which assumes the nanoparticles are at the coil temperature,
. This describes the SNR for a single pass across the sample,
and does not take into account time averaging or acquisition
We can consider as the maximum magnetization the SPIO time.
nanoparticles at the point source can produce. This is useful for
understanding the SNR as a function of the available magneti- A. 3-D SNR With Averaging
zation, which is constant for a given weight of magnetic tracer. The signal to noise ratio should be normalized to scan time
The received noise occurs from three sources, the noise figure and take averaging into account. If we assume a regular sam-
of the preamplifier, NF, the noise from the receive coil, , pling of the field of view with readout in the direction and
and body noise, . Since the MPI received signal is typ- acquisition of and lines in the and directions, re-
ically at a higher frequency than noise present in Si-BJT spectively, during a total sampling time of , we can estimate
and Si-JFET preamplifiers [12], we can model the body and coil the number of averages as
noise as spread across a noise bandwidth, . Then, assuming
Boltzmann noise, the standard deviation of the resistive voltage
noise after the preamplifier is
TABLE IV
THEORETICAL F , SAR AND SNR, AND DETECTION LIMIT FOR REAL-TIME (RT) AND HIGH RESOLUTION (HR) SCANS FOR d = 30 nm
PARTICLES AT THE
SAR LIMIT OF 4 W/KG WITH AVERAGING TIME IN SECONDS. NOTE THAT WE REQUIRE BW =22
: F TO ACHIEVE 150% OF THE INTRINSIC RESOLUTION.
WE HAVE CHOSEN THE AVERAGING TIMES TO REFLECT TYPICAL TIMESCALES OF INTEREST TO THE REAL TIME OR HIGH-RESOLUTION SCANS
where arises from the increased SNR due to the con- task to magnetically isolate the excitation field. Since a trian-
tribution of receiver coils in multiple axes. This is perhaps the gular excitation field would be composed of a sum of odd har-
only imaging modality where our signal increases linearly with monics, we instead excite with a simple sinusoid that has limited
the scanning rate as faster scanning not only increases the raw frequency content. We then isolate geometrically and with pas-
signal, but also increases the number of averages possible. As sive electrical filters.
expected, the SNR increases with greater time averaging, , Geometric isolation [1], [15] reduces coupling between the
and decreases with increasing field of view size. This does not transmit and receive coils by putting the receive coil in a gra-
take into account the increased resistance of the body as we diometer configuration. This also reduces coupled noise from
move to high frequencies [13] since , however at outside the system bore. Unfortunately, even perfect geometric
the low frequencies employed in MPI we typically operate in a isolation between the transmit and receive coils is difficult—if
coil-noise dominated regime. not impossible—to achieve because of eddy current coupling. If
Optimizing a tracer to the MPI process has many benefits. we place a conductive sample in the bore, eddy currents in the
Increasing the particle diameter increases the signal dramati- sample induce voltage in the receive coil.
cally as . This also reduces the gradient magni- Because we cannot perfectly geometrically isolate the two
tude required for the same resolution. Increasing the saturation coils, existing systems isolate with passive notch and high-pass
magnetization, , of the magnetic nanoparticles will also en- filters [1], [3], [15]. The combination can achieve million-fold
hance SNR. However, the saturation magnetization is difficult to reduction in the excitation frequency from the received signal.
change since this is an intrinsic property to the magnetic mate- Unfortunately, this comes with drawbacks. With a sinusoidal
rials used. excitation, the SNR drops near the edges of the scan volume
System modifications can also increase the signal to noise because of the reduced magnetic field slew rate. As a result, we
ratio. SNR increases with gradient strength, the FFP speed, re- do not obtain optimal SNR throughout. In this paper, we do not
ducing coil noise, and improvement of the preamplifier noise discuss filtering, but we do excite with a sinusoidal signal.
figure. MPI’s SNR increases as we increase the scanning speed,
even though we will continue to increase our bandwidth require- IX. SPECIFIC ABSORPTION RATE
ments. Reducing the gradient strength, , also increases SNR For a sinusoidally oscillating magnetic field, SAR can be an-
for a fixed sample size because the scanning rate can be in- alytically estimated for a cylinder as [16]
creased while maintaining the same magnetic field slew rate.
From (4) we estimate the SNR for various scenarios in
Table IV. The calculations are made for specific applications
with well defined fields of view and resolutions at the SAR where is the excitation frequency, is the excitation mag-
limit assuming a single receive coil and sensitivity and noise nitude, is the diameter of the sample, is the tissue resistivity
values from Gleich et al. [1]. For a small animal, we assume at the excitation frequency, and is the specific gravity of the
a coil sensitivity value of and a noise of sample. If we assume sinusoidal excitation, FOV, gradient size
. For a human scanner and , and scanning rate , we can estimate the excitation magni-
. It is clear that, theoretically, MPI should detect tude and excitation frequency as
a single nanogram of magnetic material with reasonable SNR.
These sensitivity numbers are similar to those calculated in
an illuminating simulation study on the optimal SPIO core and
diameter for MPI imaging [14].
Fig. 9. MPI spectrometer for testing point spread function and bandwidth of
the MPI process. The excitation magnet generates a 160 mT peak-to-peak os-
cillating magnetic field at 6.23 kHz. The bias coil supplies a dc magnetic field
6
of up to 80 mT. The signal received from the gradiometric receive coil is dig-
itized at 1.25 MSPS without filtering.
XII. RESULTS
In Fig. 10(a), we see the received signal plotted as a function
of virtual FFP position. This is equivalent to the explanatory
figure shown in Fig. 1. The measured data fits remarkably well
to theoretical predictions. Normalizing to the relative speed of
the FFP position shows that the point spread function does not
change across the FOV.
The magnitude of the magnetic fields used here are signif-
icantly beyond the magnetostimulation limit. Large excitation
fields are necessitated by the small core diameters of the SPIO
nanoparticles used in the study.
We fit the measured point spread function to what would be
Fig. 10. Experimental data from zero-dimensional MPI imager (a) [Top]
measured if we considered a distribution of nanoparticle diam- Parametric plot of FFP position and received signal with multiple offset field
eters. Assuming a log-normal distribution, we estimate that the strengths. The peaks correspond to the offset field generated by the bias coil.
mean magnetic core diameters is 12.3 nm with a standard devia- [Bottom] Received signal divided by instantaneous FFP velocity. Only half the
data is shown due to alignment issues which we believe are due to hysteresis.
tion of 3.2 nm. This corresponds well to particles from the same This is equivalent to the 1-D image [see (II.3)] (b) [Top] Measured PSF
manufacturer measured with a TEM [22] as well as Resovist compared with simulated PSF of a particle distribution, also shown with the
particles measured using a different type of MPI-based instru- point spread function of a hypothetical 30 nm particle. [Bottom] Log-normal
particle size distribution used to generate simulated data.
ment [23].
XIII. CONCLUSION
XIV. ACKNOWLEDGEMENTS
Magnetic Particle Imaging is a new imaging modality with
great promise to detect nanogram quantities of iron oxide tracers The authors would like to thank A. Tamrazian and R. Pida-
deep in the body of an animal or a human. To develop the tech- parthi for their excellent assistance in acquiring data.
nology into human capable systems, we need to understand the
basis for the imaging method. In this paper we have derived the REFERENCES
1-D x-space formulation of the MPI signal. We use this founda- [1] J. Weizenecker, B. Gleich, J. Rahmer, H. Dahnke, and J. Borgert,
tion to discuss resolution, bandwidth requirements, SNR, spe- “Three-dimensional real-time magnetic particle imaging,” Phys. Med.
Biol., vol. 54, pp. L1–L10, Feb. 2009.
cific absorption rate, and slew rate limitations. We concluded [2] T. F. Sattel, T. Knopp, S. Biederer, B. Gleich, J. Weizenecker, J.
by briefly discussing the construction of a system to test our Borgert, and T. M. Buzug, “Single-sided device for magnetic particle
theoretical predictions, which we tested by measuring the point imaging,” J. Phys. D: Appl. Phys., vol. 42, p. 022001, Dec. 2008.
[3] P. Goodwill, G. Scott, P. Stang, and S. Conolly, “Narrowband mag-
spread function and estimating the core diameter of a commer- netic particle imaging,” IEEE Trans. Med. Imag., vol. 28, no. 8, pp.
cially available SPIO nanoparticle. 1231–1237, Aug. 2009.
GOODWILL AND CONOLLY: HE X-SPACE FORMULATION OF THE MAGNETIC PARTICLE IMAGING PROCESS 1859
[4] J. Rahmer, J. Weizenecker, B. Gleich, and J. Borgert, “Signal encoding [15] B. Gleich and J. Weizenecker, “Tomographic imaging using the
in magnetic particle imaging: properties of the system function,” BMC nonlinear response of magnetic particles,” Nature, vol. 435, pp.
Med. Imag., vol. 9, no. 1, p. 4, 2009. 1214–1217, Jun. 2005.
[5] T. Knopp, S. Biederer, T. Sattel, and T. Buzug, “Singular value analysis [16] P. A. Bottomley and E. R. Andrew, “RF magnetic field penetration,
for magnetic particle imaging,” in IEEE Nucl. Sci. Symp. Conf. Rec. phase shift and power dissipation in biological tissue: Implications for
(NSS’08), Jan. 2008, pp. 4525–4529. NMR imaging,” Phys. Med. Biol., vol. 23, pp. 630–643, Jan. 1978.
[6] J. F. Schenck, “The role of magnetic susceptibility in magnetic res- [17] W. Irnich and F. Schmitt, “Magnetostimulation in MRI,” Magn. Reson.
onance imaging: MRI magnetic compatibility of the first and second Med., vol. 33, pp. 619–623, May 1995.
kinds,” Med. Phys., vol. 23, pp. 815–850, Jun. 1996. [18] B. A. Chronik and B. K. Rutt, “Simple linear formulation for magne-
[7] D. Hoult and R. Richards, “The signal-to-noise ratio of the nuclear tostimulation specific to MRI gradient coils,” Magn. Reson. Med., vol.
magnetic resonance experiment,” J. Magn. Reson., vol. 24, pp. 71–85, 45, pp. 916–919, May 2001.
Jan. 1976. [19] B. J. Recoskie, T. J. Scholl, and B. A. Chronik, “The discrepancy be-
[8] T. Yoshida and K. Enpuku, “Simulation and quantitative clarification tween human peripheral nerve chronaxie times as measured using mag-
of ac susceptibility of magnetic fluid in nonlinear Brownian relaxation netic and electric field stimuli: The relevance to MRI gradient coil
region,” Jpn. J. Appl. Phys., vol. 48, no. 12, 2009. safety,” Phys. Med. Biol., vol. 54, pp. 5965–5979, Oct. 2009.
[9] A. J. den Dekker and A. van den Bos, “Resolution: A survey,” J. Opt. [20] D. Schaefer, J. Bourland, and J. Nyenhuis, “Review of patient safety
Soc. Am. A, vol. 14, no. 3, pp. 547–557, 1997. in time-varying gradient fields,” J. Magn. Reson. Imag., vol. 12, pp.
[10] W. Houston, “A compound interferometer for fine structure work,” 20–29, Jan. 2000.
Phys. Rev., vol. 29, pp. 478–485, Jan. 1927. [21] J. P. Reilly, “Maximum pulsed electromagnetic field limits based on
[11] M. Shahram and P. Milanfar, “Imaging below the diffraction limit: A peripheral nerve stimulation: Application to IEEE/ANSI C95.1 elec-
statistical analysis,” IEEE Trans. Image Process., vol. 13, no. 5, pp. tromagnetic field standards,” IEEE Trans. Biomed. Eng., vol. 45, no. 1,
677–689, May 2004. pp. 137–141, Jan. 1998.
[12] V. Radika, “Low-noise techniques in detectors,” Ann. Rev. Parr. Sci., [22] V. Schaller, G. Wahnström, A. Sanz-Velasco, P. Enoksson, and C. Jo-
vol. 38, no. 1, pp. 217–277, 1988. hansson, “Monte carlo simulation of magnetic multi-core nanoparti-
[13] P. Röschmann, “Radiofrequency penetration and absorption in the cles,” J. Magn. Magn. Mater., vol. 321, no. 10, pp. 1400–1403, 2009.
human body: Limitations to high-field whole-body nuclear magnetic [23] S. Biederer, T. Knopp, T. Sattel, K. Lüdtke-Buzug, B. Gleich, J.
resonance imaging,” Med. Phys., vol. 14, no. Jan., pp. 922–31, 1987. Weizenecker, J. Borgert, and T. Buzug, “Magnetization response
[14] R. M. Ferguson, K. R. Minard, and K. M. Krishnan, “Optimization of spectroscopy of superparamagnetic nanoparticles for magnetic
nanoparticle core size for magnetic particle imaging,” J. Magn. Magn. particle imaging,” J. Phys. D: Appl. Phys., vol. 42, p. 205007,
Mater., vol. 321, pp. 1548–1551, Jan. 2009. 2009.