Guide: CFG 107

For more information login:


www.averyweigh-tronix.com

Product: Z Series
Tech Guide: TCP/IP Communication
Introduction
This technical guide will provide instructions on how to configure a Z-Series indicator for TCP/IP
communication using the Ethernet port. This example will show how to setup a Server
connection and create a protocol for sending displayed weight over the Ethernet connection
when the PRINT button is pressed.

Supplements
Tech Guide: GEN 102 PC LAN Setup
E-NET Menu
Setup
Calib Ports
Serial E-net
IP Addr Subnet Gateway DHCP Conn.NO
E-Net 1
OFF
Port 1 Type 1 Host 1

23 Server
Procedure – Ethernet Server Connection
In this example we will use E-Net Conn No 1 and Port 23 (Telnet port)
1. Press and hold F1 until PASS is displayed…release F1 > 0 is displayed
2. Enter 3 0 8 8 and press ZERO > BUSY is displayed and then SETUP
3. Press SELECT > CALIB is displayed
4. Press UNITS until PORTS is displayed
5. Press SELECT > SERIAL is displayed
6. Press UNITS > E-NET is displayed
7. Press SELECT > IP ADDR Is displayed. No changes are needed to use the default IP
8. Press UNITS until DHCP is displayed
9. Press SELECT and set to OFF
10. Press ZERO > DHCP is displayed
11. Press UNITS > CONN.NO is displayed
12. Press SELECT > E-NET 1 is displayed
13. Press SELECT > PORT 1 is displayed
14. Press SELECT and then enter 2 3 and press ZERO > PORT 1 is displayed
15. Press UNITS > TYPE 1 is displayed
16. Press SELECT and then UNITS until SERVER is displayed

CFG 107 TCPIP Communication.docx Page 1 of 2


Tech Guide: TCP/IP Communication continued Guide CFG 107

17. Press ZERO > TYPE 1 is displayed


18. Press TARE until E-NET is displayed
Next configure the protocol setup
PROTCL Menu
E-net Protcl

Prot 2

Type 2 Bind 2 Attr 2

Prnft

Port 1 Port 2 Enet 1

None Print ENQ SMA B-Cast

Protocol Setup
For this example we will use Protocol 2 and Print Format 5 for displayed weight output
1. From E-NET press UNITS > PROTCL is displayed
2. Press SELECT > PROT 1 is displayed
3. Press UNITS > PROT 2 is displayed
4. Press SELECT > TYPE 2 is displayed
5. Press SELECT and then UNITS until PRINT is displayed
6. Press ZERO > TYPE 2 is displayed
7. Press UNITS > BIND 2 is displayed
8. Press SELECT and then UNITS until ENET 1 is displayed
9. Press ZERO > BIND 2 is displayed
10. Press UNITS > ATTR 2 is displayed
11. Press SELECT > PRNFT is displayed
12. Press SELECT and then enter 5 and press ZERO > PRNFT is displayed
13. Press TARE until SAVE NO is displayed
14. Press UNITS > SAVEYES is displayed
15. Press ZERO > BUSY is displayed. The indicator will reboot and return to normal operation
Use HyperTerminal or Hercules (or any TCP Client terminal program) and test the
configuration

CFG 107 TCPIP Communication.docx Page 2 of 2