Sample Conditioning System (SCS)

The sample transport line (installed by others) must be vented to


the atmospheric vent header through the bypass flowmeter to
avoid pressure surges.

All valves, regulators, switches, etc. should be operated in


accordance with site lock-out/tag-out procedures.

1. Close the sample supply shut-off valve. Refer to Figure 3–4.

SAMPLE SYSTEM
PANEL
SAMPLE TRANSPORT
(BY OTHERS)

MEMBRANE CELL FLOW


SEPERATOR CONTROL
0.8M SAMPLE CELL

ORIFICE
PI
BYPASS
ANALYZER VENT
ELECTRONICS BYPASS FLOW SAMPLE EXTRACTION
N
ADJUST REGULATION AND
SAMPLE ISOLATION VALVE
VENT (BY OTHERS)
CELL
FI FLOW

Figure 3–4 Shutting of sample supply shut-


off valve

2. Allow the sample to flow until all residual gas has dissipated from the
lines as indicated by no flow on the sample and sample bypass
flowmeters.
3. Turn off power to the analyzer or begin maintenance procedure.

If the system will not be out of service for an extended period, it


is advised that power remain applied to the sample transport line
electric tracer, if applicable.

Hardware Installation and Maintenance Manual 3–11