You are on page 1of 1

GE HEALTHCARE

DIRECTION 5180263-100, REV 5 LOGIQ E9 S ERVICE M ANUAL

5-2-5 Signal flow overview (cont’d)


Figure 5-4 LOGIQ E9 Signal Flow - MRX

(GTX 64 Ch future)
Front Plane
256 GTX 64 Ch
GTX 64 Ch
David GTX
Pulser 0 64 Ch

MRX
CW
Option David
1

...
A/D
Converters David
I/ Q 2

David
Filter Pulser 3
(MRX 64 Ch future)
GRLY MRX 128 Ch GDIF
128

X I Analog
X Q Beamformer
TGC A/D Nathan
TGC A/D 32 Ch
R
192

TGC A/D Nathan


e 256 32 Ch Front End
TGC A/D
l T/R Nathan Interface
a 32 Ch
Nathan
192

y 32 Ch
s Nathan
DSP
32 Ch
Nathan
192

32 Ch
TGC A/D SED

TGC Test
GFE
Gen Gen

The MRX Board combines the functionality of the Receiver Boards (DRX and GRX) and the GFI.Weak
ultrasound echoes from body structures and blood cells are received by the probes and routed via the
Relay board and the XD bus to the MRX board. The MRX board amplifies the ultrasound signal and
connects it with an A/D converter to the digital domain. The digital signals are then further processed
on the MRX boards.

5-8 Section 5-2 - LOGIQ E9 description