Professional Documents
Culture Documents
Service
Manual
D
ii
i
5
:
5
P
!
si
E
50
5
a
i
e
5
f RADIOMETERm
g ORd
a lometer A/S, Denmark, 1989, All Rights Reserved. 0
COPENHAGENi
Preface
GENERAL
The purpose of this Service Manual is to give a
basic understanding of the instrument and to give
INFORMATION
\
the instructions necessary for servicing it.
m
The manual has been divided into 10 main sections CHECKOUT and
which are:
GENERAL INFORMATION, which gives a short ADJUSTMENT
ail-round description of the instrument.
DISMANTLING and
pretation of error messages and gives propo-
sals for quick and systematic location of
faults.
Service Policy
Test Equipment and Tools
Service Notes
3. Troubleshooting
Introduction 3
Operator Record 3.1
Sample Record 3.2
Transducer Record 3.3
System Records, General 3.4
System Records, Last Cali bration 3.5
System Records, Last Measurement 3.6
System Records, Last Manual Calibration 3.7
Other Record 3.8
Electronics and Software Error Messages 3.9
5. Maintenance
Introduction 5
User 5.1
Service (yearly) 5.2
a
Section Page
6. Programs
Introduction 6
User programs 6.1
Wet Section Measurement Programs 6.2.1
Wet Section Calibration Programs 6.2.2
Wet Section Maintenance Programs 6.2.3
Selftest Programs 6.3
Service Programs 6.4
7. Electronics Description
Introduction 7
Main Computer (901-782) 7.1.1
Main Computer (902-146) 7.1.2
Main Memory 7.2
Numerical Keyboard 7.3
RSISO, Serial Output 7.4
Printer and Printer Control 7.5
Wet Secti n Control, and 7.6
Wet SI ction Distribution 7.6
pH, pO2 and pCO2 Preamplifiers 7.6
Gas Mixer Interface 7.7
CRT-Monitor 7.8
CRT-Deflection 7.8
Video Amplifier 7.8
Power Supply 7.9
RSI 100, Serial Interface 7.10
floppy Disk Control 7.11
Monochromator Interface 7.12
8. Mechanics Description
Introduction 8
Wet Section 8.1
Optical System 8.1.1
Gas Mixer 8.2
10. Drawings
Survey 10
Mechanic Drawings 10.1
Electronic Drawings 10.2
Service Policy
Test Equipment and Tools
Service Notes
3. Troubleshooting
Introduction 3
Operator Record 3.1
Sample Record 3.2
Transducer Record 3.3
System Records, General 3.4
System Records, Last Cali bration 3.5
System Records, Last Measurement 3.6
System Records, Last Manual Calibration 3.7
Other Record 3.8
Electronics and Software Error Messages 3.9
5. Maintenance
Introduction 5
User 5.1
Service (yearly) 5.2
Introduction
The service policy for the ABLTM500 has been outlined to obtain a maximum degree of reliability and
a minimum of repair time.
This is obtained by demanding the advanced and time-consuming troubleshooting, repair and test-
ing to be carried out at the factory.
Further, a combination of comprehensive troubleshooting procedures in the service manual, a variety
of selfcheck and service programs, and a high degree of modulization enables the service engineer to
quickly localize a fault in theABL500.
Continuous reliability is ensured by using the test equipment, procedures and spare parts listed in this
service manual.
Electronics
All PCBs are defined as modules which must be returned to the factory for replacement under the
trade-in arrangement for modules. Repairs are not to be carried out in the field.
I _
1, The reason for this is, that most of the electronics are based on the surface mounted technology
i m
(SMT) which involves a high component density and thereby a high complexity. Repair of such PCBs
requires advanced troubleshooting, repair and test facilities, which are only available at the factory.
For the few PCBs based on conventional technology (leaded components), such as the Power Supply
PCB, we have decided to follow the same policy to ensure a continuous reliability of the ABLSOO.
Since repair of the electronics is not allowed in the field, detailed circuit diagrams have not been in-
cluded in the Service Manual.
Modules which must be returned to the factory for repair are marked with an M in the spare parts
list (M for Module replacement).
One exception is the complete CRT Monitor which may to some extend be repaired to component
level. Therefore, the complete CRT Module is defined as a module which may be returned to the fac-
tory for replacement under the trade-in arrangement.
The CRT Monitor is therefore marked with an C in the spare parts list (C for Component replace-
ment).
Service Policy -1
SERVICE POLICY
Gas Mixer
The gas mixer may to some extent be repaired to component level. The membrane pump, stepping
motor, etc. may be replaced in the field. However, parts such as the mixing coils cannot be repaired in
the field. Therefore, the complete gas mixer is defined as a module which may be returned to the fac-
tory for replacement under the trade-in arrangement.
Wet Section
Most of the electromechanical and pure mechanical parts of the wet section may be repaired to com-
ponent level; tubings, valves and pump motors and the like are regarded as components. The various
programs enables the service engineer to check and adjust the various parts and also to verify their
function after repair.
Some mechanical reassemblings and adjustments are critical for their function. Such parts are not to
be dismantled further than indicated in the spare parts list. A valve, for instance, is not to be disman-
tled and this is indicated by not listing any subparts in the spare parts list.
None of the wet section parts are included in the trade-in arrangement for modules; they can either
be repaired in the field or replaced by a new part.
Service Policy -2
TEST EQUIPMENT AND TOOLS
Introduction
Below is given a list of the various test equipment and tools required for servicing ABL500/510/520.
Description Code
In the event of important changes in procedures or components taking place, the Service Manual will be
upgraded by means of Service Notes, which will be forwarded to all implicated Radiometer Distributors.
These future Service Notes for the ABL500 should be inserted here in chronological order.
Code Notes
917217 Update of the Service Manual. Check-out, troubleshooting, dismantling and mainte-
nance procedures changed. Programs, spare parts list and drawings updated.
917-225 Inclusion of ABL510 information. Electronics, optical pans, procedures, programs, and
spare parts.
917- 232 Inclusion of ABL520 information. Revised hardware due to e.g. FEPRBMs and
reduced power consumption.
917-237 Update of the Service Manual. Check-out, troubleshooting, dismantling, spare parts
list. Addition of software versions.
917-241 Update of the Service Manual. Spare Parts List section 9.4 Software Versions.
917-242 Update of the Service Manual. Check-out, troubleshooting, Wet Section programs,
mechanical description, spare parts and mechanical drawings.
917-246 Introduce RSl50/100 version 2 software, increasing the buffer capacity for each
channel.
917-249 Inclusion of ABL505 information. Electronic, procedures, program, spare parts and
drawings.
917-258 New Printer Control PCB, 902-241 with reduced power consumption and better opto-
couplers.
917-272 pO2 drift after short power breaks. Fitter capacitors causing polarization voltage to
drop.
917-273 Inclusion of ABL50 information. Electronic, procedures, program, spare parts and
drawings.
917-280 Adaptation of CRT Monitor 902-186 to early runs of Top Part 902-l 39.
917-282 Rubber tube for old type of ABL500 with the narrow grove.
BACKGROUND: The service manual has been revised. Your attention is especially drawn to the
following:
Section 3 has been extended to comprise Electronic and software error mess-
ages.
Two new service programs have been added to section 6: Shut down proce-
dure (Wet section service program) and Clear all data in RAM (Other service
programs).
DOCUMENTATION: Please discard the following pages of the existing ABLSOO service manual:
instead, please insert the revised pages enclosed with this Service Note.
RADIOMETER-_
C O P E N H A G E N0
-d
b
Background The ABL500 Service Manual has been updated to include the information about
the ABL510.
and insert instead the revised sheets enclosed with this Service Note.
Section 2 2.13,2.14,2.15,and2.16
Section 7.12 Monochromator Interface
Section 8.1-l Optical System
*) It has been reported that the detailed program descriptions have no field ser-
vice relevance. We have, therefore, replaced them by the program sum-
maries included with this Service Note. If you wish to keep the detailed de-
scriptions, please store them in a separate binder.
MEDICAL SERVICE NOTES
C o d e 917-232 Issue 9104A Instrument ABLSOO
0dN.J~OU~d/
summary In connection with the ABL520 release, some changes have been introduced
from the following production runs:
New power supply board 902-154 includes 2 ring net relays only. For ABL500 run
50 onwards/ABL510/520 run 8 onwards only.
Existingboard907-786for ABL500 run 1 - 49/ABL510/520 run 1 - 7 only.
New Wet Section Control Board 902-157 consumes less power and includes the
waste interface circuits. Fully compatible to all ABL500/510/520 from run 1.
Existing board 901-774. Limited compatibility. For details, please refer to section
7.6.
New Main Computer Board 902-146 includes electrical erasable PROM S (FLASH-
P R O MS ). Makes Main Memory Board superfluous. For ABLSOO run 50 on-
wards/ABL510/520 run 8 onwards on/y.
Existing board 901-782 for ABL500 run 1 - 49/ABL51 O/520 run 1 - 7 only.
New waste detection system is calibration-free and consumes less power. For de-
tails, please refer to section 7.6.
New RSl50 902-148 consumes less power. Includes ring net relay. Fully com-
patible to all ABL500/51 O/520 from run 1.
Existing board 902-021. Limited compatibility. For details, please refer to section
7.4.
New RSllOO 902-147 consumes less power. Includes ring net relay. Fully com-
patible to all ABL500/510/520 from run 1.
Existing board 901-773. Limited compatibility. For details, please refer to section
7.10.
Miscellaneous:
New mains transformer. Common secondary winding for the option boards.
Reduced rinse solution consumption (ABL510/520 only).
Minor cabinet changes, e.g. painting is now white-structure, printer paper
knife is now non-transparent, gas cylinder compartment redesigned.
Documentation Please replace the obsolete service manual pages with the revised and new
sheets enclosed with this service note.
RADIOMETERm
0
C O P E N H A G E N il
MEDICAL SERVICE NOTES
Code 917-237 Issue 9 107A Instrument: ABL5XX
Distribution General
BACKGROUND: The Service Manual has been revised. Your attention is especially drawn to the
following:
Section 5.2 has been updated to reflect the latest changes in the parts replaced
during yearly maintenance.
Section 9 has been divided into three sections. The latest changes in the parts list
have been added. A section has been added to comprise Software Versions.
DOCUMENTATION: Please discard the following pages of the existing ABLSOO Service Manual:
Instead, please insert the revised pages enclosed with this Service Note.
RADIOMETER-
0
COPENHAGEN u
MEDICAL SERVICE NOTES
Code 9 17-240 I s s u e 9108A Instruments ABL5XX
Distribution General
Background There have been several changes in the Spare Parts List as well as additional in-
formation to the Service Manual which we would like to emphasize.
Spare parts 1. Pump Tubing 842-170 has been replaced by 842-194. The new Tubing can be
used in all the ABL5XX instruments.
2. The Pump Housing with Rotor 901-753 has been replaced by 902-188. The
new Pump can be used in all ABL5XX. However, there is a color difference
between the two code numbers. Please note that the new Pump 902-188 can
only use the new Pump Tubing 842-194.
The new Pump Housing has been implemented as follows:
AB L500 Run 54 and onwards.
ABL51 O/520 Run 20 and onwards.
3. The Rotor for the Pumps is now made avaiable as a separate spare part item,
code No. 923-628. The rotor fits both Pump Housing types.
If you have problems with the Rollers in the Pump you may only need to re-
place the Rotor and not the complete Pump.
4. The Tubing for Vl has been given a separate spare part number, code No.
841-757. New softer material.
5. The Tubing 841-644 for V2 in ABLSOO has been replaced by 841-750, this is
the same Tubing as the one used in ABL510/520. This is to eliminate intermit-
tent high ~02 readings on a few ABLSOOs.
6. The front part of Valves V4/5 is now available as a separate spare part. This
means that you do not have to replace the solenoids when you have a
broken Nipple.
Questions 1. Why have we increased the holding voltage for the Valves?
The background for increasing the voltage was an in-depth test of the hold-
ing force versus the applied holding voltage. It was found that by increasing
the voltage the holding force was increased significantly when passing a cer-
tain point in the supply voltage. The double voltage was selected as this volt-
age was above the point found in the test. The low voltage used earlier is not
a problem, the change was made to get an optimal function of the Valves.
RADIOMETERm
0
COPENHAGEN ii
SERVICE NOTES
from Radiometer Medical A/S
917-241 September 24,199l
BACKGROUND: The existing software has been modified in the following respect:
IMPLEMENTATION: ABL500 from run 53 onwards and ABL51 O/520 from run 22 onwards are shipped
with the new software.
UPDATING: The following updating kits are available from stock free of charge:
RADIOMETER-
0
COPENHAGENi
SERVICE NOTES RADIOMETER MEDICAL A/S
917-241 Paae 2
Each kit includes the software (either EPROMsor a Disk), update instructions, and
a Users Handbook insert.
Please note that the kits are bilingual (except the English EPROM
kits) i.e. the German and French have English as their second language. The
English EPROM kiis are English-only in order to save EPROMs.
Some of the kiis may not yet be on stock or be incomplete due to limited
capacity of EPROMs and UH-inserts. We will do our outmost to fulfil your
0
-
We recommend that all instruments are updated. You are kindly requested to
return the replaced EPROMs immediately upon completion of the updating.
Please note that all data in the internal log will be lost when updating ABLs
with main processor software version lower than 6.32. Therefore, make a
log-printout before updating!
DOCUMENTATION: Please discard the following pages of your ABLSOO service manual:
Instead, please insert the revised pages enclosed with this service note.
0
. SERVICE NOTES
MI Radiometer Medical A/S
917-242 October 31,1991
On version 1 of 902-l 54, the power supply board which was introduced in
ABL500 from run 50 and ABL51 O/520 from run 8, the connector pins of J2
are in reverse order with pin No. 12 being closest to J3.
1 0
/
The silk-screening was corrected (pin No. 1 being closest to J3) from board
version 2 onwards. The board version No. is printed on the component side.
Version 1: 971-462Kl Sl KM1 . Version 2: 971-462K2S2KM2.
DOCUMENTATION: Please discard the following pages of your ABL500 service manual:
Contents Page 2
Section 2.1 Page 2.1-2
Section 2.13 Page 2.13-l
Section 2.16 Pages 2.16-l and 2.16-2
Section 3 Page 3-2
Section 3.4 Page 3.44
Section 3.9 Page 3.9-5
Section 6.2 Page 6.2-46
Section 8.1 .l Page 8.1 .l-2
.@ Section 9.2 All pages
Section 9.4 All pages
Section 10.1 Pages lO.l-3,10.1-5, 10.1-15, 10.122 and 10.1-23.
Instead, please insert the revised pages enclosed with this service note.
RADIOMETERS
0
COPENHAGENii
SERVICE NOTES
POIII Radiometer Medical A/S
917-24s November IS,1991
Background To further improve the New Design ABL5xx we are introducing a new CRT Monitor
which provides the following benefits:
The front surface of the CRT tube is frosted which prevents reflections being
caught in the screen. This improves the readability from any angle.
The front of the tube includes an extra layer of glass which functions as a
protection shield (against blows etc.). The tube is UL-recognized as an implosion
protected type.
This enables us to remove the plastic protection shield which was needed for the
original tube. The plastic shield was prone to catch dust (also on the inner side)
which meant that cleaning was necessary during e.g. the yearly maintenance.
The new tube is a 9 kV type (the original was 11 kV), this means that the radiation
for the new tube is less than for the original. The radiation for the original is,
however, below the limit stated in the standards.
Faint vertical shadows could be observed across the original screen - these have
been eliminated on the new monitor.
Introduction The new CRT Monitor and its subparts will be included in:
l RADIOMETERS
0
COPENHAGENi
SERVICE NOTES RADIOMETER MEDICAL A/S
917-245 Page 2
The new parts are all defined as components, that is, they are not included in the
Trade-In Arrangement for modules. They are, however, priced so the net price equals
the trade-in price for the old version parts.
Compatibility The new parts are not compatible with the old neither mechanically nor electrically. Do
not attempt to use new parts in place of old or vice versa.
-
SERVICE NOTES RADIOMETER MEDICAL A/S
917-246 Paae 2
Act ion We recommend that the software is upgraded in all ABL5XXs, however, since it
is very unlikely that the error occurs, we suggest the upgrade to be carried out in
connection with other service on the instruments (e.g. in connection with the
yearly maintenance).
Spare part The new PROM, code 368-863, is available from stock free of charge.
_
SERVICE NOTES
0 POD Radiometer Medical A/S
917-246 November 30.1991
Background Earlier the RSl50/100 would only accept to store one single ASTM message (e.g.
one patient result) at a time. If a second message was sent to the RSI before it
had completed the transmission to an external computer, it would abort the
second message which would then be lost.
The above would only happen in very extreme cases as a result of a combination
of several factors:
the baud rate for the external computer is low, e.g. 300 or 600 baud.
the internal printer is set to off (causing less activity on the internal ring
network).
Till now we have only received one report concerning a lost message.
New software The buffers on the RSl50/100 have been rearranged so each channel now has
more than 10 kbyte available (the buffer assigned for the unused fourth channel
on RSIl 00 has been divided between the other three channels). This enables
l each buffer to store up to three ASTM messages - in addition, the RSI will now
also accept more than one message to be stored.
Introduction The revised software (version 2.00) will be introduced generally in:
The revised software has, however, already been included in PCBs in our stock
and will also be included in the earlier versions of RSl50/100 (902-021/901-773)
returned for repair.
RADIOMETER-
0
COPENHAGENi
SERVICE NOTES
POIII Radiometer Medical A/S
917 - 249 Instrument: ABL5XX February 14,1992
SUBJECT: In connection with the ABL505 release, the service manual has been updated.
DOCUMENTATION: Please note that the entire section 6.2 has been maked up into section 6.2.1
Wet Section Measurement Programs, 6.2.2 Wet Section Calibration Programs,
and 6.2.3 Wet Section Maintenance Programs.
Please discard the following pages of the existing ABL500 service manual:
Instead, please insert the revised pages enclosed with this service note.
RADIOMETERS
0
COPENHAoENi
. SERVICE NOTES
POITI Radiometer Medical A/S
917 - 258 Instrument: ABL5XX July 16, 1992
Background: To further improve the New Design ABL5XX we are introducing a new Printer Control
PCB which provides the following benefits:
a) A new and more reliable type of opto-coupler is included for connection to the
internal ring network.
d) The production time is reduced by having components only on one side of the
PCB.
Introduction: The new Printer Control PCB is introduced in the production of new ABL5XXs by the
end of September 1992, but is released as a spare part now.
Compatibility: The new Printer Control PCB is fully compatible with the original PCB.
However, the cable between the Printer Control and Main Computer PCBs is
soldered onto 902-241, whereas the two parts were separate in the original design.
Important Note: All 901-781 which are recycled through theTrade-in System are upgraded to include
the changes in paragraphs a) and b) above. This means that these upgraded boards
0 will be just as reliable as the new PCB since the reduction of the power consumption
alone does not increase the reliability.
So, therefore we will continue to use the original Printer Control PCB, 901-781, as
a spare part.
RADIOMETER-
0
COPENHAGENi
. SERVICE NOTES
from Radiometer Medical A/S
917 - 272 Instrument: ABL5xx February 11,1993
Background: When the instrument is turned off, the polarisation voltage on the ~02 electrode is
supplied from a battery. The switching between the battery and normal voltage
supply is done by a FET transistor.
In some cases during power down, the switching is delayed due to the fact that the
capacitors for the positive and the negative voltage on the electrode amplifier are
not discharged at the same rate.
This causes in some cases the pO2 electrode to be without polarisation voltage for a
short time. If the instrument is switched on again within short time i.e. not going into
cold start, there might be a Flush Error and ~02 drift on the first Flush and calibra-
tion.
The problem has been rectified by changing the filter capacitors in the power supply
for the ~02 amplifier.
RADIOMETER-
C O P E N H A G E0
N;
, SERVICE NOTES
from Radiometer Medical A/S
917 - 273 Instrument: ABL50 (ABLSXXI Februarv 15.1993
Documentation: In connection with the inclusion of ABL50, a number of sections and pages have
bee revised and a couple of new are added according to the list below:
Replace Sections:
1
2.1, 2.2, 2.3, 2.4, 2.5, 2.6, 2.7, 2.8, 2.9, 2.10, 2.11 and 2.17.
0 3 page l/2, 3.3 page 13114, 3.4 page 9110,
4 page 718, page 1 l/l 2, page 19120,
5.2 page 314, page 516
6, 6.1, 6.2.1 page 314, 6.2.2 page 314, page 718, page 9/l 0, page 1 l/l 2, 6.3.
7, 7.1 and 7.1 .1-l.
8 page l/2, 8.1 page 3/4, ABL50/500/51 O/520 wet section principle diagrams
9,9.1,9.2, and 9.4.
Add Sections:
3.10
4 pages 21/22,23/24, and Up/Downgrade procedure.
6.5 and 6.5.1.
7.1.3
RADIOMETERS
0
COPENHAGENi
SERVICE NOTES
0 from Radiometer Medical A/S
Instrument: ABL5xx May 12,1993
DOCUMENTATION: Please insert the attached section 2.12 into the ABL500 service manual.
0 RADIOMETER-
0
COPENHAGENC,
. SERVICE NOTES
i mm Radiometer Medical A/S
917 - 280 Instrument: ABL5xx May 12,1993
SUBJECT: Adaptation of the CRT Monitor 902-l 86 to top part 902-I 39.
BACKGROUND: The monitor including tube, deflection board and video board is available as a
spare part. It is intended as a replacement unit in case of a defective CRT
tube in the top part 902-139 distributed with the following instruments:
0
1
However, the first 34 runs of the top part were equipped with an aluminum
front panel which does not fit the CRT monitor 902-I 86 and some adaptation
parts, enclosed in the kit listed below, are therefore necessary in case the
monitor needs to be replaced.
Run 1 - 34 of top part 902-139 were distributed with the following instruments:
ABL500 run 50 - 59
ABL505 run 1 - 2
ABL510/520 run 8 - 29
DOCUMENTATION: The spare parts list has been updated according to the above (and a few
other items). Please discard the existing sheet 9.2-l/2 from the service
manual and insert instead the revised one ehclosed with this service note.
SERVICE NOTES
from Radiometer Medical A/S
917 - 281 Instrument: ABL5xx Mav 13.1993
BACKGROUND: The membrane pump circulatesatmosphericair at a flow rate of 180 ml per minute
in order to obtain the correct equilibrium conditions for the creation of the
calibration gasses.
Acontrol system keeps the flow constant by regulating the pump speed, e.g. if the
flow drops due to a pump which is defective or not mounted correctly the control
system tries to compensate by increasing the pump speed. If the air flow is not
brought back into its regulation field within acertain time the system status reports
GAS ERROR B and eventually the pump motor may stall.
PUMP REPLACEMENT 1. Select the gas mixer service program and ensure that the air pump is not
INSTRUCTION: running (gas mixer off) before you proceed with the next step.
2. Remove the left side-panel of the measuring station (behind the gas bottle)
and dismantle the old membrane pump.
3. Mount the new membrane pump by pressing it entirely down into the
chassis slot with the white tube towards the motor. Connect the 2 tubings
to the pump (transparent tubing to white nipple, black tubing to black
nipple).
4. Press the white nylon ring over the bearing of the motor axle. Ensure that
the bearing is pressed completely into the ring.
5. Ensure that the edge of the motor housing does not touch the rubber part
of the membrane pump. If it does, loosen the two fixation screws and turn
the motor away from the pump.
6. Start the air pump (gas mixer on) and check that the gas mixer is running
correctly within half a minute. If it does not, proceed with step 8 overleaf.
0 RADIOMETER-
0
COPENHAGENi
SERVICE NOTES RADIOMETER MEDICAL A/S
If the ABL gasmixer is not running correctly after replacement of the membrane pump:
8. Minimize the start up time by selecting the time saving settings in the
service program (Rinse only at power on) and switch the ABL off and on.
IMPORTANT: The potentiometers on the gas mixer control board need no adiustment in
connection with replacement of the membrane pump (only necessary if the
pressure transducers or the gas mixer control board has been replaced). 0
-
SERVICE NOTES
from Radiometer Medical A/S
917-282 Instrument: ABL500 June 11,1993
Background: In the ABL500 the SBS solution is preheated by having the tube, from the SBS
bottle to LS5, placed in a grove in the thermostatted plate for the measuring
chamber. The same is the case for the tube 841-759 from V4 to the LS3 housing/
preheater 3 (no. 23 on the drawing 22523-A4 in the service manual).
The groves in the plate have been widened and the geometri changed. The
diameter of the rubber tube has been increased at the same time. This change
was introduced in the production in the summer of 1990.
Unfortunatelly this means that the the tubes 840-043 and 841-759 now supplied
as spare parts can not go into the groves in older ABL500s which have the narrow
groves in the thermostatted plate.
A new tube, code number 840-l 04, is introduced for replacement of 840-043 and
841-759 in the above-mentioned places.
The new tube is only for use in old ABLs with the narrow grove.
If replacing the tube for the SBS bottle and the V4/LS3 housing in older ABL500s
with the narrow grove, use the tube 840-l 04.
To distinguish between where to use 840-043 or 840-l 04 please see the sketch
below, which shows the difference in the outlook of the groves in the thermostatted
plates.
Do not use the thinner tube 840-104 in newer ABL5XXs with the wide grove
as the heat transfer from the plate to the tube will not be optimal.
Documentation: Discard the pages SERVICE NOTES-i/2 and lO.l-25/26 in the ABLSOO service
- manual and replace with the revised pages enclosed with this Service Note.
eb
,)
1
Old type with the narrow grove. New type with the wide grove.
Use rubber tube 840-l 04. Use rubber tube 840-043.
RADIOMETER-
0
COPENHAGENi
SERVICE NOTES
POIII Radiometer Medical A/S
917 - 283 Instrument: ABLSO June 17,1993
Background We have found some software errors in the present version of the software,
version 1 .OO. The errors are:
Printed parameters (on PRSl5 or ticket printer) does not correspond with the
parameters selected in setup and displayed.
Manual printout of the measurement result has an incorrect printing date if the
printout is called at a later date than the actual time of the measurement. The
printed date was incorrectly set equal to the measurement date.
Calibration results for pH (low and high) is always displayed and printed as
4.47xxxxxxxxxxxxx.
The error occurs if you interrupt a normal cold start up after having upgraded
to software version 1 .OO by removing the top of the thermostatted box prior to
the startup rinse. The symptom is caused by the fact that the two values are
not properly initialized during startup.
This error may be remedied by letting the ABLSO perform an uninterrupted
cold startup or by using the Clear all RAM facility.
In the Gas mixer check service program the CO* valve was not opened
when activating key 3 (Gas mixer on). This caused the gas mixer to mal-
function in the service program.
Implementation These errors have been corrected in the new software version 1 .OlG and 1 .Ol J
(see overleaf) which is included in ABL50 of Run 16 onwards.
The following instruments of earlier Runs have been modified prior to shipment
from Copenhagen:
RADIOMETER-
I _I .
COPENHAGEN-
SERVICE NOTES RADIOMETER MEDICAL A/S
Action We recommend that the software is upgraded to version 1 .Ol when it is conven-
ient. Please refer to the service manual, Section 2.17, Page 5 for the upgrade
procedure.
Parts The Memory Cards containing the new software are available from stock under
the following part numbers:
On receipt of the new Memory Cards, please return your version 1 .OO to
RADIOMETER DISTRIBUTION A/S for reuse.
Documentation Please:
Discard the existing Service Notes sheet (survey of service notes included in
the manual), and insert instead the sheet enclosed with this service note.
Insert this front page in the Service Notes section in front of the service manual.
Discard the existing sheet, Section 9.4, Pages 5 and 6, and insert instead the
sheet enclosed with this service note.
a 1. GENERAL INFORMATION
Introduction
The ABL500 is a fully automated blood gas analyzer which measures pH, pCOp and p02 on liquid samples
and pC02 and p02 on gas samples.
The ABL50 measures the same parameters as ABL500, but on liquid samples only.
In addition to the above parameters the ABL510 measures tHb and s02., and ABL520 tHb, ~02, MetHb and
COHb.
The ABL505 measures pH, pCO2, pO2 and electrolytes K+, Na+, and Ca++ or Cl-.
Gas Mixer
The Gas Mixer produces three gas mixtures using of pure COP (from a cylinder) and surrounding
atmospheric air (drawn in by means of a membrane pump). The gas mixtures are used for calibration of
the two gas electrodes pCOp and pOz and for flushing the inlet system. The gas mixtures and their use are
outlined below:
Gas 1: 5.61% CO2 and 19.76% 02, 45 ml/min;
used for the first point calibration of both gas electrodes and for flushing the inlet system,
Gas 2: 11.22% COP and 18.58% Oz, 45 ml/min;
used for the second point calibration of the pC02 electrode, and
Gas 3: 100% CO2 from cylinder, 40 mllmin;
used for zero-point calibration of the p0, electrode.
The gas mixer is only running, that is, producing gas mixtures when an activity needing gasses is taking
place in the Wet Section. This is to minimize the CO2 consumption.
A block diagram of the Gas Mixer is given in Figure 8.2.1.
Introduction
The wet section is found in the front part of the base unit. All parts which may contain liquid (tubings,
reference well etc.) are placed on the front side of the wet section mounting plate, whereas all the electronics
( such as pump motors , wires and PCBs) are placed on the rear side. This is to prevent a liquid spillage
from damaging any electronics.
Three principle diagrams for the wet section are given in section (ABL500/50, ABL51 O/520 and ABL505)
are given in section 8.1 .l.
Thermostatted Box
The heart of the thermostatted box is the electrode chamber which include the three electrodes, pH, pC02
and p0, in the left branch and up to three electrolyte electrodes in the right branch (ABL505).
All liquids and gasses on which the ABL is to perform measurements (during measurementsorcalibrations)
are thermostatted to 37YX0.1C to assure correct measurement conditions (human conditions).
Thermostatting of the samples is carried out by a number of preheaters (active and passive). The
thermostatted box itself is thermostatted using a heating foil on rear of the thermostatted box mounting plate
which also functions as a passive preheater for the SBS and rinse solutions.
Outsidethe thermostatted box are two active preheaters, one heating both rinse solution and sample before
it enters the main valve and one between the main valve and the electrode chamber.
Liquid flow
The liquid and gas flow in the wet section is controlled by two pumps which determine flow speed and
direction. The flow path itself is controlled by five valves and a main valve with 9 positions. The flow is
monitored using five liquid sensors.
Electronics
The electronics of the ABL is built-up of a number of intelligent PCBs (including a CPU) which means that
the computer power is distributed. The main function of the various PC& is described briefly below. A block
diagram of the electronics is given in Figure 7.1 or 7.2.
Standard PCBs:
0
l-2 REVISED 93.01 .I5
e GENERAL INFORMATION
Power Supply
This PCB performs two functions. One, to supply voltages - it receives the AC voltages from the secondary
of the mains transformer, passes them through fuses, regulates two voltages (+12.1 V and +5 V for the Main
Computer PCB) and transfer the remaining AC voltages directly to the PCBs. Another, to be an
interconnection PCB for the Internal Ring Network.
CRT-Deflection PCB
This PCB is connected directly to the Main Computer PCB. On the basis of the synchronization signals
received from the Main Computer PCB, the PCB generates the signals for controlling the electron beam
of the CRT screen.
Softkeypad
This keypad is connected directly to the Main Computer PCB and is physically situated below the CRT
screen.
ABLSO PCBs:
Optional PCBs:
0
I-4 REVISED 93.01.15
2. CHECKOUT AND ADJUSTMENT
introduction
This section describes how the ABL is checked and realigned.
The ABL must have been switched on for half an hour before performing any of the described
checkouts and adjustments.
Test Equipment A sheet denoted Test Equipment in front of this Service Manual gives an
overview of the special test equipment needed for servicing the ABL. The test equipment needed for
the individual procedures is mentioned in the various paragraphs.
Test equipment such as the PHA860 must be grounded to the ABL during the checkouts and
alignments.
PHA860 polarity switch On PHA860s with serial numbers from 29R19N01, the polarity of the
output voltage is indicated underneath the POLARITY switch. On PHA860s with serial numbers below
29R19N01, the POLARITY switch should be set to position (ANION) - fl if a positive output voltage is
required and to position M + (CATION) if a negative output voltage is required.
PHA860 output voltage Always use a DVM to fine adjust the output voltage of the PHA860.
Tolerances In the procedures where the electronics may be adjusted, it will be noted that
two tolerances are included, one for checkout and one for adjustment. The checkout tolerance is the
broadest to allow for drift during operation which may occur after adjustment.
CAUTIONS Carry out only the adjustments described in this Section, unless instructed to do
so by the Service Department at Radiometer Medical A/S.
Repairs must not be attempted on any M modules, without specific authori-
zation or instructions from the Service Department at Radiometer Medical A/S.
Mains Supply
1. Check the mains supply according to the Setting [V]
(nominal) Range [VI
setting and the table:
Please note that if the mains supply is offset 240 204 - 264
by e.g. +3 % from its nominal value, all AC 220 187-244
voltages and unregulated DC voltages will 200 170-220
also be offset by +3 %. The procedures for 140 119- 134
these voltages imply that the mains supply is 120 102- 132
equal to its nominal value. 100 85-110
2. Check for 18.5 VAc f 2.0 VAC between J2l4 and J2/5.
3. Check for +5.10 V f 0.15 V between TP2 (+5.1 V, SENSEl) and TP3 (0 V, SENSE2). This is the
voltage actually applied to the Main Computer PCB.
4. Check that the voltage between TPl (+5.1 V) and TP4 (0 V) is not above 5.50 V. This is the voltage
applied to the cable connected to the Main Computer PCB.
(The difference between the voltages obtained in items 3 and 4 expresses the power loss in the cable from
the Power Supply and the Main Computer PCBs).
5. Check for +12.0 V +0.5 V/-O.3 V between TP7 (+12 V) and TP8 (0 V).
6. Check for 29.2 VAc f 3 VAc between J2/2 and J2/3 (AC voltage for the Printer Control PCB).
7. Check for 29.2 VAC f 3 VAc between J2/6 and J2/7 (AC voltage for the Wet Section Control and Gas
Mixer PCBs).
8. ABL510/520 only. Check for 11 .O V ,,c f 1.1 VAC between Jl/5 and Jl/6 (AC voltage for the
Monochromator Interface PCB).
The limits given in item 9 assume that the supplies are not loaded. In case an option is mounted, the limits
are instead 9.6 V,c f 1 .O VA,.
9. Check for 9.9 VAC f 1 .O VAc between J2/9 and J2/10 (AC voltage for the Option 1 PCB).
Check for 9.9 VAC f 1 .O V,,c between J2/11 and J2/12 (AC voltage for the Option 2 PCB).
Check for 9.9 VAc rt 1 .O VAC between Jl/l and Jl/2 (AC voltage for the Option 3 PCB).
Check for 9.9 VAc f 1 .O V,, between Jl/3 and Jl/4 (AC voltage for the Option 4 PCB).
3. Check for +5.10 V f 0.15 V between TP2 (+5.1 V, SENSEl) and TP3 (0 V, SENSE2). This is the
voltage actually applied to the Main Computer PCB.
4. Check that the voltage between TPl (+5.1 V) and TP4 (0 V) is not above 5.50 V. This is the voltage
applied to the cable connected to the Main Computer PCB.
(The difference between the voltages obtained in items 3 and 4 expresses the power loss in the cable from
the Power Supply and the Main Computer PCBs).
5. Check for +12.0 V +0.5 V/-O.3 V between TP7 (+I2 V) and TP8 (0 V).
6. Not ABLSO. Check for 31 .O VAC f 3.0 VAC between J2/2 and J2/3 (AC voltage for the Printer Control
PCB).
7. Check for 28.0 VAc f 3.0 VAC between J2/6 and JZY7 (AC voltage for the Wet Section Control and 0
Gas Mixer PCBs).
8. ABL510/520 only. Check for 11 .O VAc f 1.5 VAc between J2/11 and J2/12 (AC voltage for the
Monochromator Interface PCB).
The limits given in item 9 assume that the supplies are not loaded. In case an option is mounted, the limits
are instead 9.7 V,c f 1 .O VAC.
9. Check for 10.0 V,c f 1 .O VAC between J2/9 and 52/l 0 (AC voltage for the Option PCBs).
0
2.1 - 2 REVISED 92.11 .Ol
POWER SUPPLIES
0
Wet Section Control PCB 901-774
Please note that ground for the voltages mentionen below is found at J24/1.
1. Check for +33 V +9 V/-5 V at J24/2 (supply voltage for pumps and main valve).
3. Check for +15 V f 0.7 V at J24/4 (positive supply voltage for liquid sensor and ND converter circuits).
4. Check for +5000 mV f 20 mV at J24/5 (reference voltage for V/F and D/A converters).
5. Check for +lO.O V f 0.6 V at J24/6 (supply voltage for measuring chamber light).
6. Check for +5.0 V f 0.5 V at J24/7 (supply voltage for digital circuits).
0
7. Check for -15 V f 0.7 V at J2418 (negative supply voltage for liquid sensor and ND converter circuits).
3. Check for +15 V f 0.7 V at J24/4 (positive supply voltage for liquid sensor and ND converter circuits).
4. Check for +5000 mV f 20 mV at J24/5 (reference voltage for V/F and D/A converters).
5. Check for +9.6 V f 0.5 V at J24/6 (supply voltage for measuring chamber light).
6. Check for +5.0 V f 0.5 V at J24/7 (supply voltage for digital circuits).
0
7. Check for -15 V f 0.7 V at J24/8 (negative supply voltage for liquid sensor and ND converter circuits).
p0, channel
A zero point calibration of the @*channel MUST be carried out in case the pD, Preamplifier Unit or the Wet
Section Control PCB has been replaced - or when necessary.
++-D_rl
Electrode plug
I
PHA860 setting
Please read the paragraph concerning setting and adjusting the PHA860 in Section 2 prior to checking the
electrode channels.
l
REVISED 92.11 .Ol 2.2-l
ELECTRODE CHANNELS
Quick check
1 Enter the electrode check(out) service program.
2 Check that the counter values (#) displayed for 0 Vr and 2.5 Vr are within limits:
0 Vr: [21850 ; 274001 (#)
2.5 Vr: [46800 ; 584001 (#)
3 Remove the pH electrode (respective K+,Na+, Ca++/CI- electrodes) and instead mount the dummy
electrode (Code 902-049) for connection to a PHA860. Ground the PHA860 to the ABL.
4 Set the PHA860 to 0.0 mV (source resistance 0 0) and check that the displayed electrode signal is
0.0 mV *I .O mV (zero point).
Then change the source resistance to 1000 MQ and check that the displayed electrode signal does -
not change by more than 1 .O mV.
5 Set the PHA860 to -200.0 mV (source resistance 0 Q) and check that the displayed electrode signal
is -200.0 *3.0 mV (sensitivity).
Then change the source resistance to 1000 MR and check that the displayed electrode signal does
not change by more than 1 .O mV.
6 Exit from the electrode checkout program, remove the dummy electrode, remount the pH electrode
and reassemble the ABL.
Thorough check
1 Withdraw the Wet Section to its service position.
2 Fold down the Wet Section Control PCB (two screws on rear).
4 Locate J22 on the Wet Section Control PCB and check the following voltages:
(Ground is at pin 2)
Pin 1: +5.70 V f 0.5 V (positive supply voltage).
Pin 3: - 5.70 V f 0.5 V (negative supply voltage).
Pin 4: +2500 mV f 10 mV (reference voltage for A/D converter).
6a (pH): Using a PHA860, apply 0.0 mV to J16/1 (Ground to J16/2) and check that the displayed
electrode signal is 0.0 mV f 1 .O mV (zero point).
6b (K+): Using a PHA860, apply 0.0 mV to Jl7/1 (Ground to Jl7/2) and check that the displayed
electrode signal is 0.0 mV f 1 .O mV (zero point).
6c (Na+): Using a PHA860, apply 0.0 mV to Jl8/1 (Ground to J18/2) and check that the displayed
electrode signal is 0.0 mV f 1 .O mV (zero point).
6d (Ca+*/Cl-): Using a PHA860, apply 0.0 mV to Jig/l (Ground to Jl9/2) and check that the displayed
electrode signal is 0.0 mV f 1 .O mV (zero point).
7a (pH): Then apply - 2000 mV to Jl6/1 (Ground is at Jl6/2) and check that the displayed electrode
signal is - 200.0 mV f 1 .O mV (sensitivity).
7b (K+): Then apply - 2000 mV to Jl7/1 (Ground is at Jl7/2) and check that the displayed electrode
7c (Na+): Then apply - 2000 mV to Jl8/1 (Ground is at Jl8/2) and check that the displayed
electrode signal is - 200.0 mV f 1 .O mV (sensitivity).
7d (Ca+*/CI-): Then apply - 2000 mV to Jl9/1 (Ground is at Jl9/2) and check that the displayed
electrode signal is - 200.0 mV f 1 .O mV (sensitivity).
pC0, Channel
0 Quick check
1 Enter the electrode check(out) service program.
2 Check that the counter values (#) displayed for 0 Vr and 2.5 Vr are within limits: I
0 Vr: (21850 ; 274001 (#)
2.5 Vr: [46800 ; 584001 (#).
3 Remove the pC0, electrode and instead mount the dummy electrode (Code 902-049) for connection
to a PHA860. Ground the PHA860 to the ABL.
4 Set the PHA860 to 0.0 mV (source resistance 0 Q) and check that the displayed electrode signal is
0.0 mV f 1 .O mV (zero point).
Then change the source resistance to 1000 MR and check that the displayed electrode signal does
not change by more than 1 .O mV.
5 Set the PHA860 to - 200.0 mV (source resistance 0 Q) and check that the displayed electrode signal
is -200.0 f 3.0 mV (sensitivity).
Then change the source resistance to 1000 MR and check that the displayed electrode signal does
not change by more than 1 .O mV.
6 Exit from the electrode checkout program, remove the dummy electrode, remount the pC0, electrode
and reassemble the ABL.
Thorough check
1 Withdraw the Wet Section to its service position (nine screws marked A on Drawing 22523-A4).
2 Fold down the Wet Section Control PCB (two screws on rear).
4 Locate J15 on the Wet Section Control PCB and check the following voltages:
(Ground is at pin 2)
a) Pin 1: +5.70 V f 0.5 V (positive supply voltage).
b) Pin 3: - 5.70 V f 0.5 V (negative supply voltage).
c) Pin 4: +2500 mV f 10 mV (reference voltage for A/D converter).
6 Using a PHA860, apply 0.0 mV to J14/1 (Ground to J14/2) and check that the displayed electrode
signal is 0.0 mV f 1.0 mV (zero point).
7 Then apply -2000 mV to J14/1 (Ground is at J14/2) and check that the displayed electrode signal is
-200.0 mV f 1 .O mV (sensitivity).
Quick check
1 Enter the electrode check(out) service program.
2 Check that the values displayed for the following parameters are within the limits given:
0 Vr: [425 ; 9001 (#)
- 2.5 Vr: [31800 ; 39650](#)
Pol-r: I-1 200.00 ; -1250.00) (mV)
Pol-e: [-625.00 ; -635.00] (mV)
Batt: [-2700.00 ; -3400.00] (mV).
0
2.2-4 REVISED 92.11 .Ol
ELECTRODE CHANNELS
0
3 Remove the p0, electrode and check that the displayed electrode signal is 0 pA f 15 pA (zero point).
4 Mount the dummy electrode (Code 902-050, 100 MQ resistor) in place of the p0, electrode, and
check that the displayed electrode signal is +6300 pA f 100 pA (sensitivity). Items 3 and 4 verify that
the p0, Preamplifier is working correctly.
5 Exit from the electrode checkout program, remove the dummy electrode, remount the p0, electrode
and reassemble the ABL.
Thorough check
1 Withdraw the Wet Section to its service position (nine screws marked A on Drawing 22523-A4).
6 Check that the displayed value for the -1.2 V reference voltage (pol-r) is -1225 mV f 25 mV.
7 Check that the displayed battery voltage (Batt) is 3000 mV +400/- 300 mV.
r
0 8 Set the PHA860 to -630.0 mV (source resistance 0 SI), and apply the voltage to Jl2/1 (Ground to Jl2/
2) and check that the displayed electrode signal is 0 pA f 50 pA (zero point).
9 Then set the PHA860 to -1260 mV (source resistance 0 a) and and check that the displayed electrode
signal is 6300 pA f 100 pA (sensitivity).
12 Remove the top part of the thermostatted box and disconnect the p0, Preamplifier from the p0,
electrode.
13 Press Calibr(ate). After approximately four seconds a new offset current is displayed - check that
this value is within: [6000.00 ; 6350.00) (PA).
0
REVISED 92.11 .Ol 2.2-5
ELECTRODE CHANNELS
.
14 Press Save if the value is within limits. (Please note that if Save is pressed when the new offset
current is outside [SOOO.OO ; 7050.00], a ? will be displayed and the new offset current will be
deleted.
Quick check
1 Enter the temperature check(out) service program.
2 Check that the values displayed for the following parameters are within the limits given:
0 Vr: [5950 ; 92001 (W
5 Vr: [36000 ; 551751 (#)
Ref-37: [26900 ; 445001 (#I
Ref-37 is the reference resistor for the temperature regulation circuit, 6.1 kL.
This check verifies that the reference resistor and the A/D converter are working.
4 Check that the temperatures displayed for Block and Sense are both within 37.00 C *O.lO C.
Check also that the difference between the two temperatures is less than 0.05 C.
During a warm up period or in case of an error, the temperature difference could be larger.
5 Check that the temperatures displayed for Heatl and Heat2 are both 37.00 C *l .OO C.
0
REVISED 92.11 .Ol 2.3-1
TEMPERATURE (WET SECTION)
0
Thorough check
Withdraw the Wet Section to its service position (nine screws markedA on drawing 22523-A4).
Fold down the Wet Section Control PCB (two screws on rear).
4 Enter the temperature checkout program (wet section service program no. 2).
5 Locate Jl on the Wet Section Distribution PCB (AUX), connect a 6k100 0.05 % resistor (Code 152-
332) between J1/2 and J1/3 and check that the displayed value for AUX is 37.00 C *0.07 C (check
of temperature circuit).
6 Disconnect Preheater 1 and Preheater 2 from the Wet Section Distribution PCB and check that the
impedance between pin 3 and 6 of the respective preheater is 180 f 54 L.
0
2.3 - 2 REVISED 92.11 .Ol
2.4 WASTE
0
Introduction
Two detection systems exist. One is used in ABL5CKf runs l-49, ABL5101520 runs 1-7 and is based on
a thermal conductivity measurement of the waste container and compare the thermal condition of two
sensors. The necessary electronic hardware is mounted on a separate interface board. The system needs
calibration if the Waste Interface Board or the Wet Section Control Board are replaced.
The second system, which is used in ABLSOO run 50 onwards, ABL510/520 run 8 onwards, ABL505
ABL58, is also based on measurement of thermal conductivity, but uses a different algorithm and one
sensor only. The necesary hardware is mounted on the Wet Section Control Board. The system is
calibration- free.
0 Please note that it may take up to 10 minutes from the time the status of the waste container has changed
until the ABL gives the correct indication.
In the procedures below, the Waste Empty Constant is referred to as WEC and corresponds to VWAsrs
(in mV) divided by 10.
Checkout
1 Enter the waste calibration program (wet section service program no. 5).
i
2 Wiih an empty waste container, check (after 10 minutes) that the displayed waste sensor value is
within the range:
500 c V < lOvWEC+300 [mV].
3 With a full waste container, check (after 10 minutes) that the displayed waste sensor value is within
the range:
1 OvWEC+300 c V e 4500 [mV]. ,I-
e
4 With removed waste container, check (after 10 minutes) that the displayed waste sensor value is
within the range (same as for an empty waste container):
500 c V < lOvWEC+300 [mV].
5 Withdraw the Wet Section to its service position (nine screws marked A on drawing 22523-A4).
6 Disconnect the waste detector from the Wet Section Distribution PCB at the waste detector end and
check that the value after a short time decreases to below 500 mV (in fact the value will be close to
0 mV). When reconnecting the plug again, make sure that it is correctly oriented.
The limits used in the procedures above are illustrated in the figure overleaf where the WEC as an example
equals 240. The messages empty, full, and bad shown in the figure may be obtained if the procedure
is carried out using the check of various modules (wet section service program no.9) instead.
,..,I
~
0 500
I I I
I I
bad empty full bad
or removed
Calibration
1 Mount an empty waste container.
2 Enter the waste calibration program (wet section service program no. 5).
3 Wait 10 minutes; then press Calibrate to store the new WEC value.
Please note that only VW,,.sTE values in the range from 2200 mV through 3200 mV should be saved.
The acceptable range for VWAsTE is illustrated in the figure on the previous page (thick line).
If the value is outside the limits, the system is probably defective.
0
2.4 - 2 REVISED 92.11 .Ol
WASTE
2 The three waste sensor voltages each of which reflects the sensor output at 3 different moments
in the measuring sequence, VT(ambj, VTthighj and VT(~~,,,). The voltages are updated as the heating/
cooling cycle isruns.
All ABLa except ABL50: By pressing the Calibr. softkey the actual waste sensor voltages will
be erased and those obtained during the next measuring sequence will be displayed. (Calbr. is
actually a dummy name as no calibration takes place).
3 Wait a couple of minutes until all 3 waste sensor voltages are redisplayed. It may take between 1
l/2 and 5 l/2 minutes depending where on the actual sensor temperature curve you press the
softkey. Check that the difference between the waste sensor voltage for VT(lo,,,j and VTthighj is in
accordance with the table below, e.g. with an empty waste container and VTtamb.) between 1980 and
2170 mV (ambient temperature between 21 and 27 C) the waste sensor voltage difference should
be between 90 and 185 mV.
If the W sensor values are not as indicated, the sensor orthe Wet Section Control Board are defective
and should be replaced.
7 This limit is not used by the software as the uncertainty of the thermal conductivity without waste
container, is too high, i. e. for the ABL, only two waste container conditions exist, empty or full.
Supply Voltages
Please note that no adjustments are required for the voltage supplies which have all been preadjusted at
the factory.
1. On the Wet Section Control PCB, check for between +28.0 and +42.0 V at J24/2 (supply voltage for
the air pump and valves). Ground is at J24/1.
On the Wet Section Control PCB, check for +15.0 V f 0.7 V at J24/4 (positive supply voltage for the
monitor circuits). Ground is at J24/1.
On the Wet Section Control PCB, check for -15.0 V f 0.7 V at J24/8 (negative supply voltage for the
monitor circuits). Ground is at J24/1.
4. On the Gas Mixer Interface PCB, check for + 5000 mV f 20 mV at TPl/i 4 (reference voltage for the
monitor circuits). Ground is at TPl/l.
1. On the Wet Section Control PCB, check for + 5000 mV f 20 mV at J24/5 (reference voltage for the
ND converter). Ground is at J24/1.
0 3. Locate J26 on the Wet Section Control PCB and apply +2300 mV using a PI&60 with the source
resistance set to 0 Sz. Connect +2300 mV to J26/2 and ground to J26/1.
On 901-774: +2300 mV f 10 mV
0
REVISED 92.11 .Ol 2.5-1
2.5 GAS MIXER (ELECTRONICS AND MECHANICS)
0
The voltage supplies and the ND converter for the Gas Mixer signals should be checked before carrying
out any of the checkouts and adjustments of the Gas Mixer which involve the use of the electronics.
Supply Voltages
Please note that no adjustments are required for the voltage supplies which have all been preadjusted at
the factory.
1. On the Wet Section Control PCB, check for between +28.0 and +42.0 Vat J24/2 (supply voltage for
the air pump and valves). Ground is at J24/1.
2. On the Wet Section Control PCB, check for +I 5.0 V f 0.7 V at J24/4 (positive supply voltage for the
monitor circuits). Ground is at J24/1.
0
3. On the Wet Section Control PCB, check for -15.0 V f 0.7 V at J24/8 (negative supply voltage for the
monitor circuits). Ground is at J24/1.
4. On the Gas Mixer Interface PCB, check for + 5000 mV f 20 mV at TP1/14 (reference voltage for the
monitor circuits). Ground is at TPl/l.
1. On the Wet Section Control PCB, check for + 5000 mV f 20 mV at J24/5 (reference voltage for the
A/D converter). Ground is at J24/1.
0 3. Locate J26 on the Wet Section Control PCB and apply +2300 mV using a PHA860 with the source
resistance set to 0 Q. Connect +2300 mV to J26/2 and ground to J26/1.
On 901-774: +2300 mV f 10 mV
0
REVISED 92.11 .Ol 2.5-l
GAS MIXER
0
CO, Flow Regulator
This checkout/adjustment should always be carried out in case the CO, flow regulator has been replaced.
Checkout
1. Connect the flow meter (5-l 00 mUmin) as shown in Figure 2.5.1 (left part) below.
p+i-1
Exhaust
0
Towe; (Gas 1 + Gas 2)
To wet section
i (Gas 3)
Figure 2.5.1.
2. Enter the Electrode check(out) service program, and press Gas1 (fill); the gas mixer will deliver
Gas1 for 60 seconds.
3. When the gas mixer has stabilized (after 15 seconds), check that the total gas flow (Gas1 + Gas2)
is 90 f 10 mUmin. If outside, skip items 4 and 5 below and jump to the adjustment procedures.
4. Press Stop to stop the gas mixer, if not already stopped automatically. Connect the flow meter (5
100 mUmin) as shown in Figure 2.5.1 (right part) above. Then press Gas3 (fill) to restart the gas
mixer which will now deliver gas3 for 60 seconds.
5. When the gas mixer has stabilized (after 15 seconds), check the flow for 40 f 10 mUmin (please note
that if the gas mixer tester A7303 is used, it will indicate a flow of only 32 mUmin when the flow is
correct -this is due to the fact that the A7303 is calibrated using atmospheric air which has a higher
viscosity than pure COJ.
0
Adjustment
To carry out the adjustment a U-shaped yoke, see Figure 2.5.2 below, should be used.
Figure 2.5.2.
1. Connect the flow meter (5-100 mUmin) as shown in Figure 2.5.1 (left part) above.
2. Enter the Electrode check(out) service program, and press Gas1 (fill); the gas mixer will deliver
Gas1 for 60 seconds.
0
REVISED 92.11 .Oi 2.5 - 3
GAS MIXER
0
3. When the gas mixer has stabilized (after 15 seconds), adjust the CO, flow regulator until the total flow
is 90 f 2 mUmin using the above mentioned yoke. Then checkthe flow of each individual gas mixture
(Gas1 and Gas2) for 45 f IO mUmin.
4. Press Stop to stop the gas mixer, if not already stopped automatically. Connect the flow meter (5
100 mUmin) as shown in Figure 2.51 (right part) above. Then press Gas3 (fill) to restart the gas
mixer which will now deliver gas3 for 60 seconds.
5. When the gas mixer has stabilized (after 15 seconds), check the flow for40 f 10 mUmin (please note
that if the gas mixer tester A7303 is used, it will indicate a flow of only 32 mUmin when the flow is
correct - this is due to the fact that the A7303 is calibrated using atmospheric air which has a higher
viscisity than pure C02).
The monitor must be checked/adjusted if the CO2 pressure transducer or the CO2 flow regulator has been
replaced.
3. Press Gas mixer on (the gas mixer is turned on) and check that the mean voltage after stabilisation
(15 seconds) is 1900 f 50 mV.
- Check the total flow is 90 f 10 mUmin, see Page 3 of this section (paragraph concerning CO2 0
flow regulator).
Adjust, if outside limits.
- Check the differential pressure for < 0.7 mm water column, see page 6 of this section
(paragraph concerning differential pressure regulator).
Adjust, if outside limits. L
- Adjust RI 13 on the Gas Mixer lnteface PCB until the mean voltage displayed is 1900 f 10 mV.
4. If RI 13 has been adjusted in item 3, press Gas mixer off and Gas mixer on (to turn the gas mixer
off and then on again), and then check after stabilisation (15 seconds) that the voltage is 1900 f 20
mV.
0
2.5 - 4 REVISED 92.11 .Ol
GAS MIXER
0
Air Flow
The air flowthrough the pump which should be 180 mUmin is measured indirectly by measuring the amount
of air taken in from the surroundings (only 82 of the 180 mUmin is applied to the mixing coils, whereas the
remaining 98 mUmin are recirculated; this means that 82 mUmin is taken in from the surroundings, see
Figure 2.5.3 below).
82 mllmin
from surroundings
II I-R- Differential
pressure regulator
Figure 2.5.3.
Checkout
This checkout should always be carried out in case any part of the atmospheric air side has been replaced.
1. Remove the screws securing the gas mixer, and then turn the gas mixer inside the ABL (this is to
obtain access to the air exhaust hole of the differential pressure regulator which is found on the rear
side of the gas mixer when looking at it through the left side of the ABL).
2. Loosen the screw securing the air exhaust cover plate and then turn the plate away from the exhaust
hole.
3. Connect the test equipment as shown in Figure 2.5.4 below (use as short tubings as possible).
82 f 10 ml/min Differential
b pressure regulator
Figure 2.5.4.
4. Enter the gas mixer service program and press Gas mixer on to start the gas mixer (delivers Gasl).
5. When the gas mixer has stabilized (after 15 seconds), check the flow for 82 f 10 mUmin.
If within limits, exit from the service program, remove the test equipment and reassemble the ABL.
0
REVISED 92.11 .Ol 2.5 - 5
GAS MIXER
0
If outside limits, check the mean air flow voltage for 1900 f 50 mV.
- If the air flow voltage is within limits, adjust the air flow monitor as described below.
- If the air flow voltage is outside limits, check the air pump speed for a value below 1200 rpm - if
above, replace the membrane pump.
- Check the gas mixer for disconnected, blocked or leaking tubings; check filters, stepping motor,
air intake on the differential pressure regulator, membrane inside the differential pressure
regulator, pressure transducer, Gas Mixer Interface and Wet Section Control PC&.
1. Connect the test equipment as shown in Figure 2.5.4 above (use as short tubings as possible). 0
3. Check that the air flow voltage displayed equals 100 f 30 mV (with the gas mixer off).
If outside limits, adjust R210 on the Gas Mixer Interface PCB until the voltage is 100 f 10 mV.
4. Press Gas mixer on (the gas mixer is turned on) and check that the flow after stabilization (15
seconds) is 82 f 10 mUmin. If within limits, exit from the service program, remove the test equipment
and reassemble the ABL.
If the flow is outside limits, use the Decrease (air flow) and Increase (air flow) keys to adjust the
pump speed until the flow is 82 f 5 mUmin.
5. Then adjust R213 on the Gas Mixer Interface PC6 until the displayed voltage varies symmetrically
around 1900 mV.
6. If R213 has been adjusted in item 5, press Gas mixer off and then Gas mixer on (to turn the gas
mixer off and then on again) and then check after stabilizatm (15 seconds) that the air flow is 0
regulated to be 82 f 5 mUmin.
Please note, that if the air pump speed after adjustment of the air flow monitor is above 1200 rpm, the
membrane pump should be replaced.
This checkout/adjustment should always be carried out in case the membrane of the differential pressure
regulator has been replaced.
Before carrying out thischeckout/adjustment, boththe C&pressure monitor and the airflow monitor should
be checked and adjusted, if necessary.
0
2.5 - 6 REVISED 92.11 .Ol
GAS MIXER
0
Checkout
1. Remove the stoppers on the two test nipples situated on the differential pressure regulator and
connect instead the test equipment as shown in Figure 2.5.5 below (use as short tubings as possible).
CO2 side
Differential
Adjustment screw
pressure regulator
f
left input
i
I Air side
Figure 25.5.
2. Enter the electrode check(out) service program and press Gas1 (fill).
3. Zero the differential manometer (inject Manometer Liquid, type S5224 or similar, until the front-end
reaches the long mark on the right slant gauge).
Adjustment
1. If the manometer liquid moves to the right (COP pressure > air pressure), tighten the adjustment screw
very little (turn clockwise) until the manometer liquid does not move when shifting between the
ZERO and MEASURE positions on the differential manometer.
-
If the manometer liquid moves to the left (CO* pressure < air pressure), loosen the adjustment screw
very little until the manometer liquid does not move when shifting between the ZERO and
MEASURE positions on the differential manometer.
2. Disconnect the test equipment and carefully remount the stoppers at the test nipples.
3. Check that the total output flow (Gas1 + Gas2) is still within limits (refer to the paragraph concerning
the COP flow regulator).
If the flow has changed, it could be caused by a leak in one of the stoppers.
0
REVISED 92.11 .Ol 2.5 - 7
2.6 VALVES
0
Please note that no adjustments are required for the valves; only checkouts.
Quick check
1 Enter the valve test service program.
3 Activate Pull and release and check (audibly and visually) that the valve is pulled momen-tarily.
4 Activate Pull and hold and check that the valve is pulled and held (visually).
The valve is released again by activating Release all.
Please note that it is possible to pull and hold all the valves one after the other.
5 Exit from the valve test program - all valves are released upon exit.
Thorough check
Wet Section: 1 Check the +24 V supply on the Wet Section Control PCB between J24/1 (0 V) and
J24/3 (+24 V) for +23.4 V 1tl.2 V.
2 On the Wet Section Distribution PCB, disconnect the valves in succession and
check that the +24 V above is pin 1 of each of the valve plugs (J25 through J30).
3 For each valve, check that the impedance of the coil is 123 L 210 L.
2 On the Gas Mixer Interface PCB, disconnect the plug J5 and check that the +33
V above is present on pins 3, 5, 7, and 9 respectively (on the PCB) - ground is
present at TP1/20.
3 For each valve, check that the impedance of the coil is 288 L +25 L. The pin
numbers refer to the plug.
VI : pins 1 and 3
V2: pins 4 and 5
V3: pins 6 and 7
V4: pins 8 and 9
1 Enter the pump and main valve (MV) test service program.
3 Exit from the pump and flow selector test program after completion of the test.
Pumps
1 Set the speed to 60 uL/s using Increase and Decrease, then press Start/stop (to stop the pump
without deleting the setup).
2 Disconnect the tubings connected to the pump tubing, and connect instead two test tubings, one
going to a liquid supply beaker and one to a measuring beaker.
3 Press StarVstopto start the pump -when the liquid starts flowing into the measuring beaker, press
Start/stop to stop the pump.
4 Press Start/stop to start the pump and at the same time start a stop watch. When 60 seconds has
elapsed, press Start/stop again to stop the pump.
5 Check that the measuring beaker contains 3.6 ml +0.9 /_0.7 ml.
Main Valve
1 Use Increase and Decrease to move the Main Valve between its various positions.
Please refer to Section 4, Page 7 for a leakage test procedure for the Main Valve.
0
REVISED 92.11 .Ol 2.7 - 1
2.8 LIQUID SENSORS
Please note that no adjustments are required for the liquid sensor channels. All voltages, etc, are
preadjusted at the factory.
3 Check that the calibration values for each sensor is above 30 counts.
If below, troubleshoot the failing liquid sensor as described below and then repeat procedure
from item 1.
4 Press Fill (liquid sensors), and check after completion theat the present value is below the latest
calibration value for each liquid sensor.
5 Press Empty, and check after completion that the present value is above the latest calibration value
for each liquid sensor.
Full Empty
I I I I
I I
60 % 100%
(0 ZZts) (min. 30 counts) (Empty value
(calibration value) during last
calibration
min. 34 counts)
3 Disconnect the optical fibres at the transmitter/receiver end (Wet Section Distribution PCB).
4 Connect instead another optical fibre directly from the transmitter to the receiver.
-
REVISED 92.11 .Ol 2.8-l
2.9 INLET MODULE
0
This checkout may be used to find out whether the Inlet Module is defective in case the service
program (Check of various modules or Modules check) indicate an error.
Checkout
1 Remove the Inlet Module from the ABL.
2 Place the module on the work bench with the rear side upwards as shown.
VII,, 00 01 Flap
0
0 0 0
0
GND 0 00 Inlet
3 Check that the resistance between Flap and GND is in accordance with:
Flap closed: 0 Q.
Flap open: infinite a.
4 Check that the resistance between Inlet and GND is in accordance with:
Aspirate button deactivated: Approx. 237 k&Z.
Aspirate button activated: Approx. 100 kQ.
2 Momentarily short circuit the two pins of TP401 (calls a therm0 head CAL).
The same function may be obtained by momentarily shorting TPi/l (0 V) to TP1/5 (TEST).
3 Momentarily short circuit the two pins of SW1 01 (calls a test printout).
After completion of the test printout, check the parameters listed in item 3 of the quick check above.
24 VOLT SUPPLY :
OPAHP. S U P P L Y :. E
OPRMP. O U T P U T
PRINTER CURRENT ; E
P R I N T E R HERD T E M P . : OK
C U R R E N T FRCTOR : + 09x C-25%;+35%1
2 4 V O L T FRCTOR : + 01x
IMPEDANCE FACTOR : - 10X
TEHPERRTURE FRCTOR :+ 24x [-56X;+l21Xl
PRINTER HERD TEMP. : 28OC C+5~C;+5SCl
f#$%&()*+,-./012345678S:;<=>?@R8CDEFGHlJK
LllNOPORSTUVWXYZC~l- abcdef~hl.)klmnopqrstuvwx
uz<: ~-aCUCela3Sc~eBTTI# It 8o&bl3uusu f2lL I*BlbirP(k
L-J n
e 8%3O@tiu f&I*
n ’ am”++
2.11 SERIAL COMMUNICATION
0
1 Enter the loop-back test service program.
3 Connect the test plug required for the channel in question according to the table below.
4 Start the test and check that the status message Test OK or OK appears. In case of a fault,
the message Test failed or Failed will appear.
The test runs continuously so that it is possible to check for bad connections by working with the
wires, connectors etc. during the test.
0
r- ABL500/505/510/520
Channel
RS150/100 1
ABL50
Channel
WI50
Test Plug
Numerical
and 902-029
Alphanumerical
Keybords
1) In case the red LED on the rear of the %150/l 00 is permanently on or is blinking, the test
cannot be initiated since the PCB is disconnected from the ring network.
2) The slide switch on the test plug should be set to position loop back.
0
Introduction
This section includes checkout and adjustment procedures for the ABL5XX CRT monitor.
Two procedures exist, one for early instruments with CRT deflection board 901-770 and video amplifier
board 901-771 (top part 901-818 and 902-l 39 run l-34) and one for later instruments with CRT deflection
board 902-l 84 and video amplifier board 902-l 85 (top part 902-139 run 35 onwards).
The two procedures are quite similar and includes checkout and adjustment of the horizontal and vertical
parameters, light, contrast and focus. The adjustments are performed on the CRT deflection board and the
Video amplifier board.
Figure 2.12.1 indicates the frame size and the relative vertical and horizontal picture position.
,o Adjustment of the picture geometry (horizontal and vertical picture centering) is performed by help of
magnets on the CRT tube. Such adjustment requires special instrumentation and s&i//and must not
beattemptedcarried through in the field. In case of considerable unlinearity (trapezium and/or pillow
distortion) the entire CRT monitor must be replaced.
WARNING. For safety reasons the CRT tube must not be replaced in the field. In case of a defective
CRT tube the entire CRT monitor must be replaced.
WARNING. High Voltage! Do not disconnect the CRT anode lead from the tube while the ABL is
on power!!
If the anode lead is to be disconnected (replacement of the CRT deflection board):
Switch off the mains power, disconnect the top part cover and discharge the anode to chassis by
a screwdriver before disassembling the anode cap.
2.12.1 CRT Deflection Board 901-770 and Video Amplifier Board 901-771
Horizontal position. Check that the picture is positioned in the middle of the frame. If necessary, adjust
R33 (L. PHASE.) on the CRT deflection board. Seal the potentiometer after the adjustment.
Horizontal linearity. Select the vertical-line test picture in other service program no. 6 (test picture 3).
Check the linearity, i.e. that the distance between the lines is constant across the picture. If necessary,
adjust Ll (lower coil) on the CRT deflection board for best linearity. Use a non-metallic trimming tool,
quadratic 3 x 3 mm.
Vertical synchronization/position. Check that the picture is not scrolling or has a tendency to do so in
connection with picture changes. If necessary, adjust R52 (F. FREQ.) on the CRT deflection board until the
picture synchronizes. Fine-adjust for correct position with reference to figure 2.12.1. Seal the potentiometer
after the adjustment.
Vertical linearity. Select the horizontal-line test picture in other service program no. 6 (test picture 2).
Check the linearity, i.e. that the distance between the lines is constant. If necessary, adjust R59 (F. LIN.)
on the CRT deflection board for best linearity. Seal the potentiometer after the adjustment.
5mm 5mm
7
J -
I
8 mr
7-
-JIT
08 mm
k
2mm
T 145 mm ..
0
2.12 - 2 WRITTEN 93.04.20
/
CRT MONITOR
; 0
2.12.1.3 Contrast
The video amplitude has been factory adjusted for a normal contrast, i.e. a contrast which is considered
to be convenient for most users at normal light intensity. However, the contrast may be adjusted by R2 on
the video amplifier board. Seal the potentiometer after the adjustment. Please note that a very high contrast
may distort the picture.
2.12.1.4 Brightness
Check that the users brightness control (rear of cabinet) has a good covering range. If necessary, adjust
R6 (internal potentiometer LIGHT) on the CRT deflection board. Seal the potentiometer after the
adjustment.
t 0 2.12.1.5 Focus
Select the Ready-Picture. Adjust for normal light intensity. Check that the focus is good in the center as
well as in the corners. If necessary, adjust for best compromise by potentiometer R4 (FOCUS) on the CRT
deflection board. Do not sea/ the potentiometer after the adjustment (high voltage on the potentiometer).
2.12.2 CRT Deflection Board 902-l 84 and Video Amplifier Board 902-l 85
Horizontal synchronization. Check that the picture is not tilting or has a tendency to do so in connection
with picture changes. If necessary, select the vertical-line test picture in other service program no. 6 (test
picture 3). Short circuit P4 (near ICI) and adjust PR4 (H. FREQ.) on the CRT deflection board until the
picture is as vertical as possible. Disconnect the short of P4.
Horizontal position. Check that the picture is positioned in the middle of the frame. If necessary, adjust
PR5 (PHASE) on the CRT deflection board.
Horizontal linearity. Select the vertical-line test picture in other service program no. 6 (test picture 3).
Check the linearity, i.e. that the distance between the lines is constant across the picture. If necessary,
adjust Ll (lower coil) on the CRT deflection board for best linearity. Use a non-metallic trimming tool,
quadratic 3 x 3 mm.
Picture-width. Checkthe picture-width wii h reference to figure 2.12.1. If necessary, adjust L2 (upper coil)
on the CRT deflection board. Use a non-metallic trimming screwdriver (3 mm).
0
WRITTEN 93.04.20 2.12-3
CRT MONITOR
Vertical synchronization/position. Check that the picture is not scrolling or has a tendency to do so in
connection with picture changes. If necessary, adjust PR2 (V. SHIFT) on the CRT deflection board until
the picture synchronizes. Fine-adjust for correct position with reference to figure 2.12.1.
Vertical linearity. Select the horizontal-line test picture in other service program no. 6 (test picture 2).
Check the linearity, i.e. that the distance between the lines is constant. If necessary, adjust PR3 (V. LIN.)
on the CRT deflection board for best linearity.
Picture-height. Check the picture-height with reference to figure 2.12.1. If necessary, adjust PRl
(HEIGHT) on the CRT deflection board.
2.12.2.3 Contrast 0
The video amplitude has been factory adjusted for a normal contrast, i.e. a contrast which is considered
to be convenient for most users at normal light intensity. However, the contrast may be adjusted by PRl
(CONTR.) on the video board. Please note that a very high contrast may distort the picture.
2.12.2.4 Brightness
Check that the users brightness control (rear of cabinet) has a good covering range. If necessary, adjust
PR7 (internal potentiometer PRE. BRIGHT.) on the CRT deflection board.
2.12.2.5 Focus
Select the Ready-Picture. Adjust for normal light intensity. Check that the focus is good in the center as
well as in the corners. If necessary, adjust for best compromise by potentiometer PR8 (FOCUS) on the
CRT deflection board.
1 Place the Wet Section in the service position and fold down the Wet Section Control PCB.
3 Slightly loosen the screw securing the Lamp Unit to the Hemolyter Unit (marked 4 in Dra-
wing 23741-A4).
0
4 Call Wet Section Service Program 13, Photodiode checkout.
7 Swing the Lamp Unit forth and back until the maximum mean diode current [nAl is obtained
(bottom parameter in the middle column).
Then tighten the screw firmly.
-1 Loosen the lock nut on the lamp socket hex-head adjustment screw (marked 5 in Drawing
23741-A4).
2 Adjust the socket hex-head screw until the maximum mean diode current [nA] is obtained.
Then tighten the lock nut.
Please note that the mean photodiode current must be within 30 - 400 nA (if outside, check the
various parts of the optical system).
3 Tighten the lock nut and exit from the service program.
4 Call a CAL1 calibration and check after completion that the tHb zero point status is within 30 -
400 nA.
This procedure is to be carried out in case the optical system has been dismantled or when suspecting
an error.
1 Place the Wet Section in the service position and fold down the Wet Section Control PCB.
2 Check the + 1 S V, and - 1 S V supplies on the Wet Section Control PCB and the 11 VAC on the
Power Supply PCB.
4 Disconnect the photodiode lead for the channel to be checked and the ground lead from the
Monochromator interface PCB. see the drawing below.
@3 Q4
-5 Connect instead the code 617-849 Test Cable between the PHA860 and the analog channel in
question (black plug on Test Cable to ground socket on Monochromator Interface PCB).
0 6 Set the voltage of the PHA860 to + 100.0 mV 20.2 mV and the out-put resistance to 10 MQ
(this applies 10 nA to the channel).
a Set the out-put voltage of the PHA860 to + 1000 mV + 2 mV (this applies 100 nA to the chan-
nel).
13 When the reading has stabilized, check that each of the diode currents are within the limits gi-
ven in the table below.
If outside, check the Lamp Unit, the Hemolyzer Unit and the Monochromator Unit.
Hemolyzer Driver
This procedure is to be carried out in case the Hemolyzer Unit or the Monochromator Interface PCB
has been replaced.
The procedure is to adjust the kernel of Tl on the Monochromator Interface PCB until the lowest
possible voltage is measured across the hemolyzer crystals -this is also the point where the noise from
the hemolyzer is loudest. (The cuvette should be full to amplify the noise).
The oscillator has the following data: Frequency: approx. 30.5 kHz
Amplitude: approx. 25-100 VP, or
approx. 1 O-35 VRM~
1 Place the Wet Section in the service position and fold down the Wet Section Control PCB.
0 2 On the PCB on the Lamp Unit, lift the hemolyzer plug a little to get access to the two pins lea-
ding to the hemolyzer crystals; then connect a DVM between the two pins.
When using a DVM or a moving coil meter which cannot measure up to 40 kHz, connect a dio-
de in series with the meter (half wave rectification) and measure the mean voltage.
The diode must be orientated with its anode towards TP2/1 and its cathode towards the meter.
Cuvette Compensation
This procedure is to be carried out in case the Hemolyzer Unit or the Wet Section Control PCB has be-
en replaced.
What the procedure does is, that it measures and stores the length of the light path through the cu-
vette and at the same time performs a tHb calibration.
NOTE: Please note that the photometer lamp and the hemolyzer driver must be chec-
ked/adjusted before carrying out this procedure.
Ensure that the instrument is READY and that the latest tHb zero point calibration has been
completed without questionmarks.
Open an ampoule with S2190 and aspirate the contents into a syringe.
Open the inlet flap and press Blood if not selected. Then inject at least 200 PL 52190 and clo-
se the inlet flap.
When the measurement has completed, call Other Service Program 11, Cuvette compensation.
Check that the parameters shown are within the limits given.
Press Save and check that Current calibration factor is set to 1.000 and that Current tHb
cuvette factor is set to Calculated tHb cuvette factor.
Perform a measurement on 52190 and check that the tHb is within limits.
Introduction
This adjustment is normally only to be carried out in some cases if the measured ~02 on a blood
sample with ~02 = 100 % is outside the range 100 % + 0.4 % /- 0.6 % after replacing the Monochro-
mator Adaptor or the Monochromator Unit (or when the two parts have been taken apart).The
procedure requires the use of blood saturated 100 % (for preparation of the blood sample, see over-
leaf), a Code 920-700 Adjustment Dial and a torque wrench (4 kpcm).
2 Carry out a Protein Removal program to clean the cuvette and then call a CAL1 calibration.
5 Open the inlet flap, inject the sample, and close the inlet flap.
When the acoustic alert is heard, the ~02 value is displayed (as sample #l).
7 If ~02 is within 100.0 % +0.4 % /-0.6 %, proceed with item 14 below. If it is not, proceed with
item 8.
8 Loosen the lock nut on the free side of the Monochromator Unit.
0 9 Mount the Code 920-700 Adjustment Dial on the adjustment screw of the Monochromator
Unit.
10 so2 >lOO.l %: Turn the Adjustment Dial a little clockwise and press Repeat meas. to up-
date the screen according to the new adjustment (the new result is dis-
played as sample #l, the previous results are pushed one step down).
so2 < 99.9 %: Turn the Adjustment Dial counter-clockwise then a little clockwise. Then
press Repeat meas. to update the screen according to the new adjust-
ment (the new result is displayed as sample #l, the previous results are
pushed one step down).
11 Repeat item 10 until SOI is within 100.0 % fi 0.1 %. Note that the final movement of the Ad-
justment Dial must be carried out clockwise.
12 Remove the Adjustment Dial and tighten the lock nut (torque: 4 kpcm). If this torque is ex-
ceeded, it may change the adjustment.
13 Press Repeat meas. to update the screen and check that the ~02 is still within the limits. If
not, repeat the adjustment.
16 Check that the new zero point status of the photometer is above the lower limit (30 nA). If the
new status value is close to the lower limit, use Wet Section Service Program 13, Diode
checkout to check that all the individual diode currents are above their lower limits.
1 Collect an 0.5 ml blood sample in a 10 ml test tube. Make sure that a suitable anticoagulant
(i.e. sodium heparinate 1000 I.U./ml) is added to the blood and that blood and anticoagulant
are well mixed.
2 Oxygenate the blood by allowing a gas mixture of 5.7 % CO2 balanced with 02 to bubble
through the blood for at least three minutes (the CO2 must be present to adjust the pH value
of the blood since it influences ~02).
The oxygenation can alternately be carried out as follows: Tilt the tube to give the blood the
largest possible surface area (forming a thin liquid film) and roll it between your hands for at
least three minutes.
3 Fill the sample into a syringe or capillary tube. Note that the sample should be used as soon as
possible after preparation.
1. Select Rinse only at power on in the Service Setup (No. 5) of Other Service Programs (Menu - Util.
- Other Util. - Service - Other).
3. Loosen carefully the disk drive blind plate below the printer. The blind plate can easily be removed
0 by first lifting out the right side.
4. Check that the strap settings of your service disk drive are in accordance with the instructions in
section 4, page 4-9 and connect the two blind plate cables to the disk drive connectors.
.
5. Turn the ABL main power on.
6. Select Software Upgrade in Other Service Programs (Menu - Util. - Other Util. - Service - Other)
7. Insert the upgrade disk in the disk drive and press the Prog. softkey.
The interpreter program FLASH.HEX will be loaded into the ABL ram. The loading process can be
followed on the screen.
8. By using a small screwdriver or similar, set the SW switch (next to the serial number plate on the rear
panel) ON (left position as seen from the ABL front) when the loading is successfully finished.
9. Set the SW switch OFF (right position as seen from the ABL front) when instructed to do so by the
screen text.P
0
REVISED 92.09.11 2.17 - 1
SOFTWARE UPGRADE
11. Remove the upgrade disk from the service disk drive, disconnect the two cables and reconnect the
blind plate into the ABL front panel.
12. Restart the ABL by turning the main power on. The restart time is 15 minutes.
In case of an error during the upgrade procedure, an error code will be displayed on the screen. For
decoding, please refer to section 3.9, page 8.
1. Select Software Upgrade in Other Service Programs (Menu - Ml. - Other Util. - Service - Other)
2. Replace the normal data disk in the disk drive by the software upgrade disk and press the Prog.
softkey. -
The interpreter program FLASH.HEX will be loaded into the ABL ram. The loading process can be
followed on the screen.
3. By using a small screwdriver or similar, set the SW switch (next to the serial number plate on the rear
panel) ON (left position as seen from the ABL front) when the loading is successfully finished.
In case of &I error during the upgrade procedure, an error code will be displayed on the screen. For
decoding, please refer to section 3.9, page 8.
If a fatal error occurs during the software upgrade, e.g. a main power drop when the FEPROMs have been
erased, the microcomputer does not know how to get started and the boot program must be loaded into the
ABL by means of the Boot Board and the following procedure:
1,
2. Carefully connect the Boot Board into TPI on the Main Computer Board accessible through a cut in
the CRT compartment bottom. The connector key should be directed towards the rear.
Now a Boot Prom menu appears on the screen. Do not select the RAM choice (1) as it is intended
for the ABL production department and requires a special terminal.
If the SW switch on the rear panel is ON (left position as seen from the ABLfront) when the main power
is turned on, a softkey test is called. Carry out the test and continue with step 5.
The interpreter program FLASH.HEX will be loaded into the ABL ram. The loading process can be
followed on the screen.
6. By using a small screwdriver or similar, set the SW switch on the rear panel (next to the serial number
plate on the rear panel) ON (left position as seen from the ABL front) when the loading is successfully
finished.
7. Set the SW switch on the rear panel OFF (right position as seen from the ABL front) when instructed
to do so by the screen text.
8. Turn the main power off and remove the Boot Board.
9. For ABLs without disk drive: Remove the upgrade disk from the service disk drive, disconnect
the two cables and reconnect the blind plate into the ABL front
panel.
For ABLs with disk drive: Replace the software upgrade disk with the normal data disk.
10. Restart the the ABL by turning the main power on again. The restart time is 15 minutes.
In case of an error during the upgrade procedure, an error code will be displayed on the screen. For
decoding, please refer to section 3.9, page 8.
ABLSO
1. Switch off the ABL, mount the Memory Card in the slot on the edge of the Main Computer PCB.
2. Locate the software upgrade switch which is accessible through a hole in the chassis just above the
Memory Card slot. Using a pen or similar press and hold the software upgrade switch while switching
on the ABL and release the switch after two seconds.
The programming is now in progress. During the programming the four LEDs next to the Memory Card
connector indicate the stage of the programming, see the table below (CR717 is the upper LED).
If anything goes wrong during the programming, the programming halts and some of the LEDs start flashing
and the acoustic alarm sounds at 1 second intervals; the table below provides interpretation of the fault and
a corrective action.
NOTE: The message RINGNET COMMUNICATION ERROR may appear during the programming.
This does, however, not influence the programming.
3. When the acoustic signal sounds and all LEDs are off (after up to 120 seconds) the programming is
complete. Switch off the ABL, remove the Memory Card and switch on the ABL again.
4. The message System setup invalid - default selected. Press continue! appears.
Press Continue and then check that the ABL starts up correctly.
CR CR CR CR Corrective Action
717 709 708 805 Interpretation
0 0 0 0 Initialization or termination NA
0 0 I 0 Checksum error in Memory Card Try reprogramming. Replace Memory Card.
0 1 0 0 Incorrect programming voltage. Try reprogramming.
Replace Main Computer PCB.
0 1 I 0 Device code error for FPROM(s) Try reprogramming.
0 I 1 0 Replace Memory Card.
0 II 1 I 1 0 Replace Main Computer PCB.
I 0 0 0 Main Computer RAM test error. Try reprogramming.
Replace Main Computer PCB.
1 0 I 0 FPROM erase error. Try reprogramming.
I 0 1 0 Replace Main Computer PCB.
1 0 I 0
1 I1 0 0 FPROM programming error. Try reprogramming.
I 1 0 0 Replace Memory Card.
I I 0 0 Replace Main Computer PCB.
1 n1I 1 0 FPROM verify error. Try reprogramming.
m1 1 1 0 Replace Memory Card.
n1I 1 1 0 Replace Main Computer PCB.
I
/
0 = LED off 1 =LEDon I = LED flashing
- \
m 3. TROUBLESHOOTING
I%
GENERAL INFORMATION
Monitoring The ABL performs a multitude of checks during operation. If these checks are not
facilities satisfactory, the errors are registered in relevant records in the System status.
All records are also filed in the System status log.
The ABL may, depending on the severity of a detected error:
- interrupt its present activity
- carry on, but question the results
- disregard invalid data entry,
Question marks in front of measured and derived parameters indicate that the measurement results are
unreliable due to errors.
AB L500/505/510/520
Messages The ABL indicates the presence of a record by issuing messages on the screen in the
instrument status field (upper left) on the second and third line.
The following general messages can be displayed, one at a time, on line 2 in the
following order of priority:
Measurement aborted during measurement
Calibration aborted during calibration
Maintenance aborted during one of the maintenance programs
Insufficient sample during measurement
Inhomogeneous sample during measurement
No Cleaning solution (1) during cleaning
No Salt-bridge solution (2) during refill/rinse/Cal l/Cal 2/Total cal
No Cal 1 solution (3) during refill/Cal l/Cal 2/Total cal
No Cal 2 solution (4) during refill/Cal 2/Total cal
No Rinse solution (5) during refill/rinse/Cal l/Cal 2/Total cal
The messages disappear after a rinse is completed.
The following messages can be displayed, one at a time, on line 3 in the following order
of priority:
Check record in System status in case of system record
Messages in System status in case of operator, sample, and transducer
records.
To enter System
Status Press the softkeys: M &nu / System status. (The System status softkey isvisible in this
case).
Record groups All detected errors are divided into 4 groups displayed in a pop-up window. Only the
groups that the record belongs to are visible, together with the corresponding softkey.
The other record groups aredisplayed with low light intensity, and the softkeys forthem
are not displayed. The record groups are:
Transducer
record are detected during monitoring of transducers and electrodes. (section 3.3)
System record refers to mechanical or electronic errors, faulty gas/solution supply, or temperature.
The system record is subdivided into:
- general (section 3.4)
- from last calibration (section 3.5)
- from last measurement (section 3.6)
- from last manual calibration (section 3.7)
- other records (section 3.8)
ABL50
To enter System
Status Press the Menu softkey - 4, System Status
1. RECORDS
2. CALIBRATION STATUS.
General messages
Measurement messages
Calibration messages
The messages are numerical codes which are listed and described in section 3.10.
In the ABL50 standard setup the numerical codes are compressed. In order to
make ABL50 print the detailed numerical error codes, Print extended errors must
be selected in service &rogram no. 13.
The Calibration Status includes 4 display pages with the present calibration status, i.e.
Main Display with information on the last and next calibration. Barometer status included.
Calibration status for the pH electrode
Calibration status for the pCO2 electrode
Calibration status for the pC2 electrode
OPERATOR RECORD
INLET CLOSED
Interpretation: Gas measurement:
Inlet closed before requested.
Liquid measurement:
Inlet closed before LS4 has detected liquid.
ABL action: Rinse.
Opera for action: Repeat measurement on sample.
Service action: 1 Test function of inlet flap.Wet section service program no. 9.
2 Test function of LS4.Wet section service program no. 8
3 Test inlet module, test procedure 2.9.
4 Check mounting of optical fibers for LS4.
5 Replace wet section control PCB.
6 Replace wet section distribution PCB.
7 Replace LS4.
NO SAMPLE
Interpretation: No sample detected during aspiration in last measurement.
ABL action: Acoustic alarm, Rinse.
Opera for action: Repeat measurement on sample.
Service action: 1 Test function of aspirate button.Wet section service program no.9.
2 Test function of LS4.Wet section service program no. 8.
3 Test inlet module, test procedure 2.9.
4 Check mounting of optical fibers for LS4.
5 Replace wet section control PCB.
INHOMOGENEOUS SAMPLE
Interpretation: Air bubbles detected in sample. The Air counts greater than 1.5%
and less than 5.0% of total counts during sample check.
ABL action: All parameters marked with I?.
Operator action: Perform a new measurement.
Service action: 1 Test function of LS2. Wet section service program no. 8.
2 Check wet section tubing.
3 Perform leakage test of main valve Page 4-8.
4 Replace wet section control PCB.
5 Replace wet section distribution PCB.
INSUFFICIENT SAMPLE
Interpretation: Too small sample volume detected. The Air counts greater than 5%
during sample check and LS2 empty.
ABL action: All parameters marked with ?.
Opera for action: Perform a new measurement.
Service action: 1 Test function of LS2. Wet section service program no. 8.
2 Replace wet section control PCB.
3 Replace wet section distribution PCB.
0
REVISED 92.01 .I7 3.20 1
._ TROUBLESHOOTING
0
TRANSDUCERRECORD
Transducer
record is obtained from flush of pC0, and p0, electrodes performed every hour, gas flush of p0,
electrode performed every 10 minutes, calibration, and measurement.
Ca++ DRIFTING
Interpretation: Last calibration
Ca++ drift value exceeds user-selected limits.
Operator action: Check for other errors in SYSTEM RECORD, GENERAL. Remedy if
any are listed.
Error is remedied when Cal 1 or Cal 2 is accepted (If protein deposits
removed from the electrode were the cause of the error, the calibra-
tion show a considerable change in drift value. It can be necessary to
call another calibration).
If limits are too narrow, redefine them via Setup program.
Check measuring chambers for clots - remove electrode(s) and clean
. with cotton stick moistened with distilled water, if necessary.
0 Perform Protein removal program.
Call tubing refill program.
Remembrane Ca electrode.
Replace Ca electrode.
Service action: 1 Check the records in SYSTEM STATUS.
2 Check Ca channel, test procedure 2.2.
3 Check SBS line and dosage.
4 Check wet section tubing.
5 Check continuity between calomel electrode and shield of Ca
Preamplifier cable.
6 Replace Wet Section Control PCB.
7 Replace Ca Preamplifier.
meq/L.
Operator action: Check for other errors in SYSTEM RECORD, GENERAL. Remedy if
any are listed.
Error is remedied when Cal 2 is accepted.
Call Tubing refill program.
Ensure electrode is mounted correctly in measuring chamber.
Ensure there is sufficient Salt-bridge Solution in calomel electrode
well. Ensure calomel electrode plug is connected.
Clean Ca electrode plug if dirty or wet.
Check measuring chambers for clots - remove electrode(s) and clean
with cotton stick moistened with distilled water, if necessary.
Check calibration for status value trend. Remembrane electrode if
status value has gradually reached the limit.
Replace Ca electrode.
Service action: See Ca++ DRIFTING.
Cl- DRIFTING
Interpretation: Last calibration
Cl- drift value exceeds user-selected limits.
Operator action: Check for other errors in SYSTEM RECORD, GENERAL. Remedy if
any are listed.
Error is remedied when Cal 1 or Cal 2 is accepted (If protein deposits
removed from the electrode were the cause of the error, the calibra- i
tion show a considerable change in drift value. It can be necessary to
call another calibration).
If limits are too narrow, redefine them via Setup program.
Check measuring chambers for clots - remove electrode(s) and clean
with cotton stick moistened with distilled water, if necessary.
Perform Protein removal program.
Call tubing refill program.
Remembrane Cl electrode.
Replace Cl electrode.
Service action: 1 Check the records in SYSTEM STATUS.
2 Check Cl channel, test procedure 2.2.
3 Check SBS line and dosage.
4 Check wet section tubing. )
5 Check continuity between calomel electrode and shield of Cl
Preamplifier cable.
6 Replace Wet Section Control PCB.
7 Replace Cl Preamplifier.
l
REVISED 92.01.24 3.3 - 3
TROUBLESHOOTING
Cl- UNSTABLE
Interpretation: Last calibration/measurement
Cl- response fault.
Opera for action: Check for other errors in SYSTEM RECORD, GENERAL and/or
SAMPLE RECORD. Remedy if any are listed.
REVISED 92.01.24
TROUBLESHOOTING
0
Error is remedied when Cal 2 is accepted.
Check measuring chambers for clots - remove electrode(s) and clean
with cotton stick moistened with distilled water, if necessary.
If Cl electrode is contaminated, remembrane electrode.
Ensure analyzer is grounded.
Ensure plug of calomel electrode is connected.
Ensure there is sufficient Salt-bridge Solution in calomel electrode
well.
Cal Tubing refill program.
Service action: See Cl- DRIFTING.
CAL 2 ERROR
Interpretation: Last e-point pH, K+, Na+, Ca++/CI-calibration.
Operator action: Check specific error.
Call Cal 2 program (Calibration).
Service action: See actions for individual electrodes.
Hb EFFECTIVE LOW
Interpretation: Last measurement:
Effective hemoglobin too low fwreliable determination of SO,.
Operator action: Effective hemoglobin too low for reliable determination of SO,.
Service action: 1 Check photodiode channels.
2 Replace Monochromator Interface PCB.
K+ DRIFTING
Interpretation: Last calibration
K+ drift value exceeds user-selected limits.
Operator action: Check for other errors in SYSTEM RECOYD, GENERAL. Remedy if
any are listed.
Error is remedied when Cal 1 or Cal 2 is accepted (if protein deposits
removed from the electrode were the cause of the error, the calibra-
tion show a considerable change in drift value. It can be necessary to
call another calibration).
If limits are too narrow, redefine them via Setup program.
Check measuring chambers for clots - remove electrode(s) and clean
with cotton stick moistened with distilled water, if necessary.
Perform Protein removal program.
Call Tubing refill program.
Remembrane K electrode.
Replace K electrode.
Service action: 1 Check the records in SYSTEM STATUS.
2 Check K channel, test procedure 2.2.
3 Check SBS line and dosage.
4 Check wet section tubing.
5 Check continuity between reference electrode and shiels of K
Preamplifier cable.
6 Replace Wet Section Control PCB.
7 Replace K Preamplifier.
K+ OUT OF LIMITS
Interpretation: Last calibration
K+ status value out of limits: 1.2-13.3 mmol/L, or 1.2-13.3 meq/L.
Operator action: Check for other errors in-SYSTEM RECORD, GENERAL. Remedy if
any are listed.
Error is remedied when Cal 2 is accepted.
Call Tubing refill program.
Ensure electrode is correctly mounted in electrode chamber.
Ensure that there is sufficient Salt-bridge Solution in calomel elec-
trode well.
Ensure calomel electrode plug is connected.
Clean K electrode plug if dirty or wet.
Check measuring chambers for clots - remove electrode(s) and clean
with cotton stick moistened with distilled water, if necessary.
Check calibration log for status value trend.
Remembrane electrode if status has reached the limit.
Replace the K electrode.
Replace calomel electrode.
Service action: See K+ DRIFTING
K+ OUT OF RANGE
Interpretation: Last measurement:
K+ out of coverage range: 0.5-25 mmol/L or 0.5-25 meq/L.
Operator action: Check for other errors in SYSTEM RECORD, GENERAL and/or
SAMPLE RECORD. Remedy if any are listed.
Error is remedied when Cal 2 is accepte . Perform quality control.
P
Check measuring chambers for clots - remove electrode(s) and clean
with cotton stick moistened with distilled water, if necessary.
Call refill program.
Ensure there is sufficient Salt-bridge Solution in calomel electrode
well.
0 Service action:
Check plug of calomel electrode.
See K+ DRIFTING.
K+ SENS. ERROR
Interpretation: Last calibration:
K+ sensitivity out of range: 92-l 05 %
Operator action: Check for other errors in SYSTEM RECORD, GENERAL. Remedy if
any are listed.
Ckeck K drift value during Cal 1 and Cal 2 (if protein deposits re-
moved from the electrode were the cause of the error, the calibration
show a considerable change in drift value. It can be necessary to call
another calibration).
Error is remedied when Cal 2 is accepted.
Clean K electrode plug if dirty or wet.
Check calibration log for status value trend. Remembrane electrode if
status has reached the limit.
Replace K electrode.
Service action: See K+ DRIFTING
K+ UNSTABLE
Interpretation: Last calibration/measurement
K+ response fault
Operator action: Check for other errors in SYSTEM RECORD, GENERAL and/or
SAMPLE RECORD. Remedy if any are listed.
Error is remedied when Cal 1 or Cal 2 is accepted. Perform quality
control.
Check measuring chambers for clots - remove electrode(s) and clean
with cotton stick moistened with distilled water if necessary.
If K electrode is contaminated, remembrane electrode.
Ensure analyzer is grounded.
Ensure plug of calomel electrode is connected. 0
-
Ensure there is sufficient Salt-bridge Solution in calomel electrode
well.
Call Tubing refill program.
Service action: See K+ DRIFTING
Na+ DRIFTING
L
Interpretation: Last calibration
Na+ drift value exceeds user-selected limits.
Opera for action: Check for other errors in SYSTEM RECORD, GENERAL. Remedy if
any are listed.
Error is remedied when Cal 1 or Cal 2 is acypted (If protein deposits
removed from the electrode were the cause of the error, the calibra-
tion show a considerable change in drift value. It can be necessary to
call another calibration).
If limits are too narrow, redefine them via Setup program.
Check measuring chambers for clots - remove electrode(s) and clean
with cotton stick moistened with distilled water, if necessary.
Perform Protein removal program.
Call tubing refill program.
Replace Na electrode.
Service action: 1 Check the records in SYSTEM STATUS.
2 Check Na channel, test procedure 2.2.
3 Check SBS line and dosage.
4 Check wet section tubing.
5 Check continuity between calomel electrode and shield of Na
Preamplifier cable.
6 Replace Wet Section Control PCB.
7 Replace Na Preamplifier.
3.3 - 8
l
TROUBLESHOOTING
3.3 - 9
l TROUBLESHOOTING
pC0, UNSTABLE
Interpretation: Last calibration:
pC0, response fault during calibration.
Last measurement: >
pC0, response fault during electrode monitoring
Operator action: See pCO,out of range.
Service action: See pC0, out of range.
pH DRIFTING
Interpretation: Last calibration:
pH drift value exceeds user-selected limits.
Operator action: See pH out of range.
Service action: See pH out of range.
pH OUT OF LIMITS
Interpretation: Last calibration:
pH status value out of limits: 6.820 - 7.946
Operator action: See pH out of range.
Service action: See pH out of range.
pH OUT OF RANGE
Interpretation: Last measurement:
pH out of coverage range: (6.3 - 8.0).
Operator action: Check for other errors in SYSTEM RECORDS GENERAL. Remedy
if any are listed.
Check for sample errors in SAMPLE RECORD. Remedy if any are
l
3.3 - 11
TROUBLESHOOTING
pH SENS. ERROR
Interpretation: Last calibration:
pH sensitivity out of range: 92 - 103% or 0.92 - 1.03.
Operator action: See pH out of range.
Service action: See pH out of range.
pH UNSTABLE
Interpretation: Last calibration:
pH response fautt during calibration.
Last measurement:
ph response fault during electrode monitoring.
Operator action: See pH out of range.
Service action: See pH out of range.
3.3 - 12
t- -
TROUBLESHOOTING
0
2 Adjust lamp unit and lamp socket, refer to Section 2.13.
3 Check Photodiode channels, refer to Section 2.14.
4 Replace Monochromator if one or more Photodiode channels are
defective.
5 Replace Monochromator Interface PCB.
6 Check Hemolyzer Unit, replace if defective.
p0, DRIFTING
Interpretation: Last flush:
p0, drift value exceed 75% of user- selected limits during 1 -hour
check.
Last calibration:
p0, drift value exceeds user-selected limits.
Operator action: See p0, out of range.
Service action: See p0, out of range.
p0, UNSTABLE
Interpretation: Last calibration:
p0, response fault during calibration.
Last measurement:
p0, response fault during electrode monitoring.
Operator action: See PO, out of range.
Service action: See p0, out of range.
SPECTRUM ERROR
Interpretation: Last measurement:
Spectrum mismatching the combination of measured derivatives.
One or more parameters > 5% outside measuring range.
Operator action: Call Cal 1 and repeat measurement.
Service action: 1 Check photodiode channels Section 2.14.
2 Check the photolamp Section 2.13.
3 Check voltages for the Monochromator Interface PCB (5V, +15
and - 15V on the wet section control board) and 11 VAC on the
power supply board.
4 Replace Monochromator Interface PCB.
5 Replace Monochromator Unit.
defective.
5 Replace Monochromator Interface PCB.
6 Check Hemolyzer Unit, replace if defective.
TURBIDITY HIGH
Interpretation: Last measurement:
Sample turbidity to high for reliable measurement.
Opera for action: Hyperlipemic sample: decrease lipemic content by, e. g. centrifugation
a /
3.3 - 15
TROUBLESHOOTING
or extraction.
Perform measurement on blood sample from healthy donor.
Service action: 1 Check Hemolyzer Unit section 2.15, replace if necessary.
2 Replace Monochromator interface PCB.
3.3 - 16
TROUBLESHOOTING
0
SYSTEM RECORDS, GENERAL
Operator
actions are listed in order of priority.
CAL 1 NEEDED
Interpretation: Gas Cal 1 required after maintenance procedure, or last Gas
Cal 1 not completed due to errors.
ABL action: All data with ? after measurement.
Operator action: Check for specific error in SYSTEM RECORD IN LAST
CALIBRATION. Remedy the error and recalibrate analyzer.
Service action: Same as operator action.
CAL 2 NEEDED
Interpretation: Gas Cal 2 required after maintenance procedure, or last Gas
Cal 2 not completed due to errors.
ABL action: All data with ? after measurement.
Operator action: Check for specific error in SYSTEM RECORD IN LAST
CALIBRATION. Remedy the error and recalibrate analyzer.
Service action: Same as operator action.
l CLEANING ERROR A
Interpretation: No cleaning solution (1) detected by LS2
Operator action: Replace Cleaning Solution bottle (ABL5xO: S5345, ABL505:
S5375).
Call Refill program (Maintenance).
Service action: 1 Test LS2, wet section service program no. 8.
2 Check optical fibres, replace if faulty.
3 Check wet section tubing.
4 Test valves, wet section service program no. 6.
5 Test Pi, wet section service program no. 7.
6 Replace wet section control PCB.
7 Replace wet section distribution PCB.
6 Replace Electrode Chambers or LS4 if faulty.
CLEANING ERROR B
Interpretation: Malfunction in liquid system transport, LSl detecting liquid
when expected empty.
Operator action: Check System Status for other errors, e.g. RINSE ERROR
A.
Perform Protein Removal Program (Maintenance).
Check the Wet Section for leakage or salt deposits at:
Electrodes: Remove electrodes and clean upper surface
of the electrode chambers.
Pump 1: Replace pump tubing if necessary.
Call liquid sensor adj. (Maintenance).
Call Cal 2 (Calibration).
Service action: 1 Test LSl, wet section service program no. 8.
2 Check optical fibres, replace if faulty.
3 Check wet section tubing. 0
4 Test valves, wet section service program no. 6.
5 Test Pi, wet section service program no. 7.
Replace wet section control PCB. L
7 Replace wet section distribution PCB.
8 Replace Electrode Chambers if faulty.
CONTAMINATED CUVETTE
Interpretation: Contaminated cuvette detected during last rinse check. tHb
0.2 mmol/L, or 0.32 g/dL, or 3.3 g/L.
ABL action: Oxi parameters in next measurement have ?.
Opera for action: Perform protein removal procedure.
Call Cal 1.
Service action: 1 Check photo lamp, replace if necessary.
2 Adjust lamp unit and lampsocket refer to Section 2.13.
3 Check photodiode channels refer to Section 2.14.
4 Replace Monochromator if one or more diode channels
are defective.
5 Replace Monochromator Interface PCB.
6 Check Hemolyzer Unit section 2.12, replace if defective.
FLUSH ERROR
Interpretation: LSl, LS2, or LS4 detected liquid during Flush program.
Operator action: Check System Status for other errors, e. g. RINSE ERROR
A.
Perform Protein Removal program (Maintenance).
Check the Wet Section for leakage or satt deposits at:
Electrodes: Remove electrodes and clean upper surface
of the electrode chambers.
Pump 1: Replace pump tubing if necessary.
Call liquid sensor adj. (Maintenance).
Call Cal 2 (Calibration).
Service action: 1 Test function of LSl , LS2, LS4.
0
3.4 - 2 REVISED 92.01.17
TROUBLESHOOTING
1)
Check optical fibre, replace it if faulty
Wet section service program no. 8.
Test function of Vi and V2
Wet section service program no. 6.
Test Pi. Wet section service program no. 7.
Check wet section tubing.
Replace wet section control PCB.
Replace wet section distribution PCB.
GAS ERROR A
Interpretation: CO2 pressure too low.
A BL action: All data with ? after measurement. No automatic calibrations
and calibrations not accepted.
Operator action: Replace CO2 cylinder.
Check tubing from CO2 cylinder.
Service action: 1 Check CO2 pressure.
2 Wet section service program no. 3.
3 Test CO2 pressure monitor. Test procedure 2.5.
4 Replace gas mixer interface PCB.
5 Replace wet section control PCB.
6 Replace CO2 pressure transducer.
7 Replace gas mixer.
GAS ERROR B
Interpretation: Gas mixer not operating.
A BL action: All data with ? after measurement. No automatic calibrations
and calibrations not accepted.
Operator action: Check for loose tubing.
Service action: 1 Check gas mixer. Test procedure 2.5
2 Check air filter and replace if dirty.
3 Check membrane pump and replace if necessary.
LSl ERROR
Interpretation: LSl error during Liquid Sensor adjustment. Emty counts <
34.
Opera for action: Check System Status for other errors, e.g. RINSE ERROR
A.
Perform Protein Removal Program (Maintenance).
Check the Wet Section for leakage or salt deposits at:
Electrodes: Remove electrodes and clean upper surface
of the electrode chambers.
Pump 1: Replace pump tubing if necessary.
Call liquid sensor adj. (Maintenance).
Call Cal 2 (Calibration).
Service action: 1 Test LSl , wet section service program no. 8.
2 Check optical fibres, replace if faulty.
3 Check wet section tubing.
4 Test valves, wet section service program no. 6.
5 Test Pl , wet section service program no. 7.
6 Replace wet section control PCB.
7 Replace wet section distribution PCB.
8 Replace Electrode Chambers if LSl is faulty.
LS2 ERROR
Interpretation: LS2 error during Liquid Sensor adjustment. Emty counts c
34.
Operator action: Check System Status for other errors, e.g. RINSE ERROR
A.
Perform Protein Removal Program (Maintenance).
Check the Wet Section for leakage or salt deposits at:
Electrodes: Remove electrodes and clean upper surface
of the electrode chambers.
Pump 1: Replace pump tubing if necessary.
Call liquid sensor adj. (Maintenance).
Call Cal 2 (Calibration).
Service action: 1 Test LS2, wet section service program no. 8.
2 Check optical fibres, replace if faulty.
3 Check wet section tubing.
4 Test valves, wet section service program no. 6.
5 Test PI, wet section service program no. 7.
6 Replace wet section control PCB.
LS3 ERROR
Interpretation: LS3 error during Liquid Sensor adjustment. Emty counts <
34.
Operator action: Check System Status for other errors, e.g. RINSE ERROR
A.
Perform Protein Removal Program (Maintenance).
Check the Wet Section for leakage or salt deposits at:
Pump 2: Replace pump tubing if necessary.
Call liquid sensor adj. (Maintenance).
Call Cal 2 (Calibration).
Service action: 1 Test LS3, wet section service program no. 8.
2 Check optical fibres, replace if faulty.
3 Check wet section tubing.
4 Test valves, wet section service program no. 6.
5 Test P2 wet section service program no. 7.
6 Replace wet section control PCB.
7 Replace wet section distribution PCB.
8 Replace LS3 if faulty.
LS4 ERROR
Interpretation: LS4 error during Liquid Sensor adjustment.
Operator action: Check System Status for other errors, e.g. RINSE ERROR
A.
Check the washer for inlet. Replace if leaky or damaged.
Perform Protein Removal Program (Maintenance).
Check the Wet Section for leakage or salt deposits at:
Pump 2: Replace pump tubing if necessary.
Call liquid sensor adj. (Maintenance).
Call Cal 2 (Calibration).
Service action: 1 Test LS4, wet section service program no. 8.
2 Check optical fibres, replace if faulty.
3 Check wet section tubing.
4 Test valves, wet section service program no. 6.
5 Test P2 wet section service program no. 7.
Replace wet section control PCB.
7 Replace wet section distribution PCB.
8 Replace inlet tube if LS4 is faulty.
LS5 ERROR
Interpretation: LS5 error during Liquid Sensor adjustment. Emty counts <
34.
Operator action: Check System Status for other errors, e.g. RINSE ERROR
A.
Checkthe Salt-bridge solution bottle and nipple for clogging.
RINSE ERROR A
Interpretation: No rinse solution detected by LS4.
ABL action: Calibrations are not accepted. Analyzer not READY.
Operator action: Replace Rinse Solution bottle (ABL5xO: S4901, ABL505:
S4921).
Call a rinse or Refill program (Maintenance).
Service action: 1 Check Rinse Solution tubing.
RINSE ERROR B
Interpretation: Malfunction in liquid system transport. LSI seeing liquid
when expected empty.
ABL action: Calibrations are not accepted. Analyzer not READY.
Operator action: Rinse solution bottle mounted properly? Check and correct
if necessary.
Check System Status for other errors, e.g. RINSE ERROR
A.
Perform Protein Removal Program (Maintenance).
Check the Wet Section for leakage or saft deposits at:
Electrodes: Remove electrodes and clean upper sur-
face of the electrode chambers.
Pump 1: Replace pump tubing if necessary.
Call liquid sensor adj. (Maintenance).
Call Cal 2 (Calibration).
Service action: 1 Test LSl, wet section service program no. 8.
2 Check optical fibre, replace it if faulty.
3 Check wet section tubing.
4 Test valves, wet section service program no. 6.
5 Test Pl, wet section service program no. 7.
6 Replace wet sectioncontrol PCB.
7 Replace wet section distribution PCB.
8 Replace Electrode Chambers.
0 RINSE ERROR C
Interpretation: Malfunction in liquid system transport. LSl, LS2 or LS4
detecting liquid when expected empty.
Opera for action: Check System Status for other errors, e.g. RINSE ERROR
A.
Perform Protein Removal Program (Maintenance).
Check the Wet Section for leakage or salt deposits at:
Electrodes: Remove electrodes and clean upper surface
of the electrode chambers.
Pump 1: Replace pump tubing if necessary.
Call liquid sensor adj. (Maintenance).
Call Cal 2 (Calibration).
Service action: 1 Test LSl , LS2 and LS4, wet section service program no.
8.
2 Check optical fibers, replace if faulty.
0
REVISED 92.01 .I7 3.4 - 7
TROUBLESHOOTING
SBS ERROR A
Interpretation: No Salt-bridge solution (2) detected. LS5 did not sense liquid
when expected.
ABL action: All pH calibration and measurement data have?, and . . . . . for
pH reading
Operator action: Replace Salt-bridge Solution (ABL5xO: S4915, ABL505:
S4935). Call Tubing refill program (Maintenance).
Service action: 1 Test function of LS5.
2 Check optical fibres, replace if faulty.
3 Wet section service program no. 8.
4 Check wet section tubing.
5 Test function of Pl .
Wet section service program no. 7.
7 Check function of V2 and V6.
8 Wet section service program no. 6.
9 Replace wet section control PCB.
10 Replace wet section distribution PCB.
SBS ERROR B
Interpretation: Malfunction in Salt Bridge Solution transport, LSI detected
liquid during replacement of SBS front in Rinse program.
ABL action: All pH calibration and measurement data have ?, and . . . . . for
pH reading
Operator action: Check Salt-bridge solution level.
Check System Status for other errors, e.g. RINSE ERROR
A.
Perform Protein Removal Program (Maintenance).
Check the Wet Section for leakage or salt deposits at:
Electrodes: Remove electrodes and clean upper surface
of the electrode chambers.
Pump 1: Replace pump tubing if necessary.
Call liquid sensor adj. (Maintenance).
Call Cal 2 (Calibration).
Service action: 1 Test LSl, wet section service program no. 8.
2 Check optical fibres, replace if faulty.
3 Check wet section tubing.
4 Test valves, wet section service program no. 6.
5 Test Pi, wet section service program no. 7.
6 Replace wet section control PCB.
7 Replace wet section distribution PCB.
TEMP. ERROR
Interpretation: Temperature in thermostatted unit outside 37.0 +/- 0.2 C.
A BL action: All measurement data have ?. Automatic calibration not
performed. Calibrations are not accepted.
Operator action: Check ambient temperature are within 15 C to 34 C.
ABL 500 has performed a Startup from cold condition, wait
for TEMP. error to disappear.
Check fan filter. Replace if dirty.
Service action: 1 Check wet section temperature, test procedure 2.3.
2 Replace wet section control PCB
3 Replace heating foil for heated panel.
4 Replace thermistors for electrode chambers.
WASTE FULL
Interpretation: Waste Detector has detected waste container full.
A BL action: Message: WARNING! Waste container is full! Replace con-
tainer.
Operator action: Replace waste container and press the Continue softkey.
Service action: 1 Perform Waste check sec. 2.4.
2 Replace Waste Detector.
3 Replace Wet Section Control PCB.
4 Replace Wet Section Distribution PCB.
CAL ERROR A
Interpretation: Liquid detected in sensor LSI .
ABL action: Rinse. CAL NEEDED.
Operator action: Check System Status for other errors, e.g. RINSE ERROR A.
Perform Protein Removal Program (Maintenance).
Check the Wet Section for leakage or salt deposits at:
Electrodes: Remove electrodes and clean upper surface of the
electrode chambers.
Pump 1: Replace pump tubing if necessary.
Call liquid sensor adj. (Maintenance).
Call Cal 2 (Calibration).
Service action: 1 Test LSI, wet section service program no. 8.
2 Check optical fibre, replace it if faulty.
3 Check wet section tubing.
4 Test valves, wet section service program no. 6.
5 Test Pi, wet section service program no. 7.
6 Replace wet section control PCB.
7 Replace wet section distribution PCB.
8 Replace Electrode Chambers.
CAL ERROR B
Interpretation: Liquid detected in sensor LS2.
ABL action: Rinse. CAL NEEDED.
Operator action: Check System Status for other errors, e.g. RINSE ERROR A.
Perform Protein Removal Program (Maintenance).
Check the Wet Section for leakage or salt deposits at:
Electrodes: Remove electrodes and clean upper surface of the
electrode chambers.
Pump 1: Replace pump tubing if necessary.
Call liquid sensor adj. (Maintenance).
Call Cal 2 (Calibration).
Service action: Test LS2, wet section service program no. 8.
Check optical fibre, replace it if faulty.
Check wet section tubing.
Test valves, wet section service program no. 6.
Test Pl , wet section service program no. 7.
Replace wet section control PCB.
Replace wet section distribution PCB.
Replace Electrode Chambers.
CAL ERROR C
Interpretation: Calibrating solution (3) not detected by liquid sensor LSl .
ABL action: Rinse. CAL NEEDED.
Operator action: Check System Status for other errors, e.g. RINSE ERROR A.
Perform Protein Removal Program (Maintenance).
0
REVISED 92.01.24 3.5 - 1
TRO
CAL ERROR D
Interpretation: Calibrating solution (3) not detected by LS2.
ABL action: Rinse. CAL NEEDED.
Operator action: Replace Calibrating Solution container.
Service action: 1 Check Rinse Solution tubing.
2 Test LS2 wet section service program no. 8.
3 Check optical fibre, replace it if faulty
4 Replace measuring chambers if LS2 is defective.
5 Replace wet section distribution PCB.
6 Replace wet section control PCB.
7 Check Pl wet section service program no. 7.
8 Check Valves wet section service program no. 6.
CAL ERROR E
Interpretation: S1565 Calibrating Solution (3) not detected by liquid sensor LS3.
ABL action: Rinse. CAL NEEDED.
Operator action: Check System Status for other errors, e.g. RINSE ERROR A. -
Perform Protein Removal Program (Maintenance).
Check the Wet Section for leakage or salt deposits at:
Pump 2: Replace pump tubing if necessary.
Call liquid sensor adj. (Maintenance).
Call Cal 2 (Calibration).
Service action: 1 Test LS3, wet section service program no. 8.
2 Check optical fibres, replace if faulty.
Check wet section tubing.
4 Test valves, wet section service program no. 6.
5 Test P2 wet section service program no. 7.
6 Replace wet section control PCB.
7 Replace wet section distribution PCB.
8 Replace LS3 if faulty.
CAL ERROR G
Interpretation: Calibrating solution (4) not detected by LS2.
ABL action: Rinse. CAL NEEDED.
Operator action: Replace Calibrating Solution container.
Service action: 1 Check Rinse Solution tubing.
2 Test LS2 wet section service program no. 8.
3 Check optical fibre, replace it if faulty
4 Replace electrode chambers if LS2 is defective.
5 Replace wet section distribution PCB.
6 Replace wet section control PCB.
7 Check Pi wet section service program no. 7.
8 Check Valves wet section service program no. 6.
CAL ERROR H
Interpretation: Calibrating Solution (4) not detected by liquid sensor LS3.
A BL action: Rinse. CAL NEEDED.
Operator action: Check System Status for other errors, e.g. RINSE ERROR A.
Perform Protein Removal Program (Maintenance).
Check the Wet Section for leakage or salt deposits at:
Pump 2: Replace pump tubing if necessary.
Call liquid sensor adj. (Maintenance).
Call Cal 2 (Calibration).
Service action: 1 Test LS3, wet section service program no. 8.
2 Check optical fibres, replace if faulty.
3 Check wet section tubing.
4 Test valves, wet section service program no. 6.
5 Test P2 wet section service program no. 7.
CAL ERROR I
Interpretation: Unhomogenous Calibrating Solution (3).
ABL action: Rinse. CAL NEEDED.
Operator action: Check Calibration Solution properly mounted. Remedy if necessary.
Check System Status for other errors, e.g. RINSE ERROR A.
Perform Protein Removal Program (Maintenance).
Check the Wet Section for leakage or salt deposits at:
Electrodes: Remove electrodes and clean upper surface of the
electrode chambers.
Pump 1: Replace pump tubing if necessary.
Call liquid sensor adj. (Maintenance).
Call Cal 2 (Calibration).
Service action: 1 Test LS2, wet section service program no. 8.
2 Check optical fibre, replace it if faulty.
3 Check wet section tubing.
4 Test valves, wet section service program no. 6.
5 Test Pl , wet section service program no. 7.
6 Replace wet section control PCB.
7 Replace wet section distribution PCB.
8 Replace Electrode Chambers.
CAL ERROR J
Interpretation: lnhomogenous Calibrating Solution (4).
ABL action: Rinse. CAL NEEDED.
Operator action: Check Calibration Solution properly mounted. Remedy if necessary.
Check System Status for other errors, e.g. RINSE ERROR A.
Perform Protein Removal Program (Maintenance).
Check the Wet Section for leakage or salt deposits at:
Electrodes: Remove electrodes and clean upper surface of the
electrode chambers.
Pump 1: Replace pump tubing if necessary.
Call liquid sensor adj. (Maintenance).
Call Cal 2 (Calibration).
Service action: 1 Test LS2, wet section service program no. 8.
2 Check optical fibre, replace it if faulty.
3 Check wet section tubing.
4 Test valves, wet section service program no. 6.
5 Test Pl , wet section service program no. 7.
6 Replace wet section control PCB.
7 Replace wet section distribution PCB.
8 Replace Electrode Chambers.
CAL ERROR L
Interpretation: Leakage in oxi-system.
A BL action: Oximetry data with ?.
Operator action: Check System Status for other errors, e.g. MEAS ERROR 0 and MAN.
CAL ERROR 0. J
Check the Wet Section for leakage or salt deposits at Hemolyzer Unit
Service action: 1 Check wet section tubings at Hemolyzer Unit.
2 Test valves V2 and V3, wet section service program no. 6.
3 Replace Hemolyzer Unit.
MEAS ERROR A
Interpretation: Liquid sensor LSl detected liquid with open inlet flap.
ABL action: Acoustic alarm. Rinse.
Operator action: Check System Status for other errors, e.g. RINSE ERROR A.
Perform Protein Removal Program (Maintenance).
Check the Wet Section for leakage or salt deposits at:
Electrodes: Remove electrodes and clean upper surface of the
electrode chambers.
Pump 1: Replace pump tubing if necessary.
Call liquid sensor adj. (Maintenance).
Call Cal 2 (Calibration).
Service action: 1 Test LSl, wet section service program no. 8.
2 Check optical fibre, replace it if faulty.
3 Check wet section tubing.
4 Test valves, wet section service program no. 6.
5 Test Pl , wet section service program no. 7.
6 Replace wet section control PCB.
7 Replace wet section distribution PCB.
8 Replace Electrode Chambers.
MEAS ERROR B
Interpretation: Liquid sensor LS2 detected liquid with open inlet flap.
ABL action: Acoustic alarm. Rinse.
Opera for action: Check System Status for other errors, e.g. RINSE ERROR A.
Perform Protein Removal Program (Maintenance).
Check the Wet Section for leakage or salt deposits at:
Electrodes: Remove electrodes and clean upper surface of the
electrode chambers.
Pump 1: Replace pump tubing if necessary.
Call liquid sensor adj. (Maintenance).
Call Cal 2 (Calibration).
Service action: 1 Test LS2, wet section service program no. 8.
2 Check optical fibre, replace it if faulty.
3 Check wet section tubing.
4 Test valves, wet section service program no. 6.
5 Test Pl , wet section service program no. 7.
6 Replace wet section control PCB.
7 Replace wet section distribution PCB.
8 Replace Electrode Chambers.
MEAS ERROR C
Interpretation: Liquid sensor LS4 detected liquid with open inlet flap.
ABL action: Acoustic alarm. Rinse.
Operator action: Check System Status for other errors, e.g. RINSE ERROR A.
Perform Protein Removal Program (Maintenance).
l
REVISED 92.01.24 3.6 - 1
TROUBLESHOOTING
MEAS ERROR D
Interpretation: Liquid sensor LS2 detected liquid during blood sample injection.
ABL action: Acoustic alarm. Rinse.
Operator action: Check System Status for other errors, e.g. RINSE ERROR A.
Perform Protein Removal Program (Maintenance).
Check the Wet Section for leakage or salt deposits at:
Electrodes: Remove electrodes and clean upper surface of the
electrode chambers.
Pump 1: Replace pump tubing if necessary.
Call liquid sensor adj. (Maintenance).
Call Cal 2 (Calibration).
Service action: 1 Test LS2, wet section service program no. 8.
2 Check optical fibre, replace it if faulty.
3 Check wet section tubing.
4 Test valves, wet section service program no. 6.
5 Test Pl , wet section service program no. 7.
6 Replace wet section control PCB.
7 Replace wet section distribution PCB.
8 Replace Electrode Chambers.
MEAS ERROR E
Interpretation: Liquid sensor LSl detected liquid during gas sample injection.
A BL action: Acoustic alarm. Rinse.
Operator action: Check for condensation in gas sample.
Check System Status for other errors, e.g. RINSE ERROR A.
Perform Protein Removal Program (Maintenance).
Check the Wet Section for leakage or salt deposits at:
Electrodes: Remove electrodes and clean upper surface of the
electrode chambers.
Pump 1: Replace pump tubing if necessary.
Call liquid sensor adj. (Maintenance).
Call Cal 2 (Calibration).
Service action: 1 Test LSl, wet section service program no. 8.
2 Check optical fibre, replace it if faulty.
MEAS ERROR F
Interpretation: Liquid sensor LS2 detected liquid during gas sample injection.
ABL action: Acoustic alarm. Rinse.
Operator action: Check for condensation in gas sample.
Check System Status for other errors, e.g. RINSE ERROR A.
Perform Protein Removal Program (Maintenance).
Check the Wet Section for leakage or salt deposits at:
Electrodes: Remove electrodes and clean upper surface of the
electrode chambers.
Pump 1: Replace pump tubing if necessary.
Call liquid sensor adj. (Maintenance).
Call Cal 2 (Calibration).
Service action: 1 Test LS2, wet section service program no. 8.
2 Check optical fibre, replace it if faulty.
3 Check wet section tubing.
4 Test valves, wet section service program no. 6.
5 Test Pl , wet section service program no. 7.
6 Replace wet section control PCB.
7 Replace wet section distribution PCB.
8 Replace Electrode Chambers.
MEAS ERROR G
Interpretation: Liquid sensor LS4 detected liquid gas sample injection.
ABL action: Acoustic alarm. Rinse.
Operator action: Check for condensation in gas sample.
Check System Status for other errors, e.g. RINSE ERROR A.
Perform Protein Removal Program (Maintenance).
Check the Wet Section for leakage or salt deposits at:
Pump 2: Replace pump tubing if necessary.
Call liquid sensor adj. (Maintenance).
Call Cal 2 (Calibration).
Service action: 1 Test LS4, wet section service program no. 8.
2 Check optical fibres, replace if faulty.
3 Check wet section tubing.
4 Test valves, wet section service program no. 6.
5 Test P2 wet section service program no. 7.
6 Replace wet section control PCB.
7 Replace wet section distribution PCB.
8 Replace LS4 if faulty.
0 3.6 - 3
REVISED 92.01.24
TROUBLESHOOTING
MEAS ERROR K
Interpretation: Liquid sensor LSl detected liquid during sample processing.
ABL action: Rinse.
Opera for action: Check System Status for other errors, e.g. RINSE ERROR A.
Perform Protein Removal Program (Maintenance).
Check the Wet Section for leakage or salt deposits at:
Electrodes: Remove electrodes and clean upper surface of the
electrode chambers.
Pump 1: Replace pump tubing if necessary.
Call liquid sensor adj. (Maintenance).
Call Cal 2 (Calibration).
Service action: 1 Test LSl , wet section service program no. 8.
2 Check optical fibre, replace it if faulty.
3 Check wet section tubing.
4 Test valves, wet section service program no. 6.
5 Test Pl , wet section service program no. 7.
6 Replace wet section control PCB.
7 Replace wet section distribution PCB.
8 Replace Electrode Chambers.
MEAS ERROR L
Interpretation: Liquid sensor LS2 detected liquid during sample processing.
A BL action: Rinse.
Operator action: Check for condensation in gas sample.
Check System Status for other errors, e.g. RINSE ERROR A.
Perform Protein Removal Program (Maintenance).
Check the Wet Section for leakage or salt deposits at:
Electrodes: Remove electrodes and clean upper surface of the
electrode chambers.
Pump 1: Replace pump tubing if necessary.
Call liquid sensor adj. (Maintenance).
Call Cal 2 (Calibration).
Service action: 1 Test LS2, wet section service program no. 8.
2 Check optical fibre, replace it if faulty.
3 Check wet section tubing.
4 Test valves, wet section service program no. 6.
5 Test Pl , wet section service program no. 7.
6 Replace wet section control PCB.
7 Replace wet section distribution PCB.
8 Replace Electrode Chambers.
MEAS ERROR M
Interpretation: LS3 detected liquid during sample processing.
ABL action: Rinse.
Operator action: Check System Status for other errors, e.g. RINSE ERROR A.
Perform Protein Removal Program (Maintenance).
Check the Wet Section for leakage or salt deposits at:
MEAS ERROR N
Interpretation: No sample detected during sample processing.
ABL action: Rinse.
Operator action: Clots in sample? Too small sample volume?
Check System Status for other errors, e.g. RINSE ERROR A.
Perform Protein Removal Program (Maintenance).
Check the Wet Section for leakage or salt deposits at:
Electrodes: Remove electrodes and clean upper surface of the
electrode chambers.
Pump 1: Replace pump tubing if necessary.
Call liquid sensor adj. (Maintenance).
Call Cal 2 (Calibration).
Service action: 1 Test LSl, LS2 and LS4 wet section service program no. 8.
2 Check optical fibre, replace it if faulty.
3 Check wet section tubing.
4 Test valves, wet section service program no. 6.
5 Test Pl , wet section service program no. 7.
6 Replace wet section control PCB.
7 Replace wet section distribution PCB.
8 %$?!iZ&e E.&%sssc&ec~~~bses~Q~ cs4.
MEAS ERROR 0
Interpretation: Leakage in oxi-system.
ABL action: Oximetry data with ?.
Operator action: CheckSystem Statusforothererrors, e.g. CAL ERROR Land MAN. CAL
ERROR 0.
Check the Wet Section for leakage or salt deposits at Hemolyzer Unit
Service action: 1 Check wet section tubings at Hemolyzer Unit.
2 Test valves V2 and V3, wet section service program no. 6.
3 Replace Hemolyzer Unit.
OTHER RECORD
Wet section
error Codes TheStart up/Restart screen will be displayed with the following message:
Wet section control fault. Error code x
where x = 0 : power on reset
1 : unknown fault
2 : prom error
3 : ram error.
The message will be printed out as well.
The start up program is performed automatically after 10 seconds.
QUESTION MARKS
Purpose To view and print out the list of records registered by the ABL during its operation. The
following information is available:
Record fype describes the record group the error belongs to.
State describes the instrument status during which the record has been
registered. Calibration, measurement, and QC measurement are
identified with number (e.g. QC meas. # . ..).
Date/Time gives the date and time for the registered record.The latest detected
record is placed last in the list.
To enter System
status log Press the softkeys: Menu / Data Management I System status log.
The cursor is positioned on the latest record.
All internal errors will display error messages on the screen. The screen will be black with white text,
in the format shown below:
Report format:
_
mm## FaOO for fatal errors. For other internal error mm is error type and
##a code number.
vvvv-vv Version identification of system software.
....
5555-1111 Place in program for the error and identification of FaOO errors.
Interpretation Internal errors are divided into the following four types:
of internal Calc: Calculation program module.
error messages: Fatal error: Exception errors.
OS: Operating System program module.
Unpack: Display program module.
The type code refersto the part of the software where the error is located.
Calc:
Ca03
t- 2 Tabel overflow Same as above
Fatal error:
FaOO 1,2 Attemp to write to prom address Replace Main CPU PCB or Main
Memory PCB.
FaOO 1, ? Attention key pressed If Attn. key has not been pressed
replace Main CPU PCB.
Fatal error:
QCplot failure
QCrestore failure
L
FaOO Total clearing complete This message will appear after
activating Clear all soft key in
other service program No. 10.
After this message switch the
ABL500 off and on again
Fatal error:
FaOO 2 Warmstart
FaOO 2 DM returns n
2 Logcomb error
OS:
I OS04 1 2 r Overflow
~~~~~ ~~in rcb table I I
OS10 2 Illegal pGroup
l Illegal
I I I
I
I I 2 date I I
F
OS16
2 Illegal context
2 Illegal timeout
2 Illegal bank
2 Illegal int id
I OS30 I 2 I Timeout I I
1
I
OS31 12
I
I I
Part order not observed I
I
II
I I
I
Unpack:
Unoack:
Cant scroll
Picture mismatch
Illegal testpicture
0 Return code nn
0 2
3
Replace Wet Section Control PCB.
Replace the Power Supply PCB in Base Unit.
4 Replace the Main Computer PCB.
Error code 1
ABL action: Performs a Power On Reset of the ABL.
Operator action: None
Interpretation: Unknown cause of reset.
Service action: 1 Check the wet section voltages.
2 Replace Wet section Control PCB.
3 Replace the Power Supply PCB in Base Unit.
4 Replace the Main Computer PCB.
Error code 2
ABL action: Performs a Power On Reset of the ABL.
Operator action: None
Interpretation: PROM checksum error detected on the Wet Section Control PCB.
Service action: 1 Check the wet section voltages.
2 Replace Wet Section Control PCB.
3 Replace the Power Supply PCB in Base Unit.
4 Replace the Main Computer PCB.
Error code 3
ABL action: Performs a Power On Reset of the ABL.
Operator action: None
Interpretation: RAM read/write error detected on the Wet Section Control PCB.
Service action: 1 Check the wet section voltages.
2 Replace Wet Section Control PCB.
3 Replace the Power Supply PCB in Base Unit.
4 Replace the Main Computer PCB.
Note: After remedy of the error the ABL must be turned off and then back on and the program
restated.
Error code 2
Interpretation: Erase of FEPROMs failed.
Service action: 1 Restart the program. If Bank 8 and 15 have erased use Boot PCB.
2 If the error is still there, replace Main Computer PCB.
Error code 3
Interpretation: Programming of FEPROMs failed.
Service action: 1 Restart the program. If bank 8 and 15 have been erased use Boot
PCB.
2 If the error is still there, replace Main Computer PCB.
Error code 4
Interpretation: Vpp error.
Service action: 1 Check SW- switch on the rear of Main Computer Unit.
Error code 10 1
Interpretation: No disk drive detected.
Service action: 1 Chek mounting of disk drive.
2 Replace disk drive.
3 Replace Main Computer PCB.
Revised 91.04.08
a TROUBLESHOOTING
This section includes the compressed as well as the extended system status record codes.
In order to make the ABL50 display extended error codes, select Print extended errors in Service Program
No. 13.
17. SENDING ERROR 1150 Transmission error during last data transfer to connected
computer.
0 Check that the computer is properly connected.
Theck that Computer has been selected for RSl50 in
Interface setup (Setup program 8).
Check that the computer setup (RS150) is correct.
Call send report/status in the Print/send menu.
Correct setup if necessary.
Internal errors will cause a stop in the normal main computer program execution and an error code will be
displayed on the LCD:
AA-BBBB-V.W-DDDD
The errors are divided in 4 main groups depending upon the source of the problem.
We-OOOO-v.w-0001 Wet section is not Check the ringnet (flat cable connections). Restart the
responding ABL. Replace wet section control board, power supply
board, main computer board or Keypad/display if error
reoccurs.
Error in WS trans- Check the ringnet (flat cable connections). Restart the
mission protocol. ABL. Replace main computer board if error
reoccurs.
We-OOOO-v.w-0011 WS timer overflow Error in wet section timer. Restart the ABL. Replace wet
section control board if error reoccurs.
We-OOOO-v.w-0012 Main valve error Continuously incorrect main valve position. Check
photosensor connection to wet section control board.
Check fixation of photosensor and strobo-disc. Check
that holes in strobo-disc are free from dust ect.
We-0000~v.w-0013 RAM error RAM read/write error on wet section control board.
Restart the ABL. Replace wet section control board if
error reoccurs.
EPROM error EPROM checksum error on wet section control board.
Restart the ABL. Replace wet section control board if
error reoccurs.
We-0000~v.w-0050 Calculation error Error in program execution. Restart the ABL. Replace
and main computer board if error reoccurs.
We-0000~v.w-0051
Counts out of range Wet section signal out of range. Check electrode sig-
nals. If OK, restart the ABL. Replace wet section control
board if error reoccurs.
We-OOOO-v.w-0059 Unknown count Wet section signal unknown. Check electrode signals.
If OK, restart the ABL. Replace wet section control
board if error reoccurs.
We-0000~v.w-0053 Calculation error Error in program execution. Restart the ABL. Replace
through main computer board if error reoccurs.
We-OOOO-v.w-0058
and
We-0000~v.w-0060 Calculation error Error in program execution. Restart the ABL. Replace
through main computer board if error reoccurs.
We-OOOO-v.w-0064
Clear all RAM Displayed when the RAM is cleared (service program
no. 14). Restart the ABL by switching off/on.
Rt-0000~v.w-0003 Errors in program Restart the ABL. Replace main computer board if error
through execution. reoccurs.
Rt-oooo-v.w-0005
Sd-0000~v.w RAM error RAM read/write erroron main computer board. Restart
the ABL. Replacemaincomputerboard ifertorreoccurs.
Sd-0001 -v.w FEPROM error FEPROM checksum error on main computer board.
Restart the ABL. Replace main computer if error
reoccurs.
0 3.10 - 5
WRll-l-EN 93.01 .I5
TROUBLESHOOTING
0
EPROM/RAM Error
During start-up of ABL50 and when entering the service program the Keypad & Display CPU performs an
EPROM checksum test and an RAM write/read test. If one of these tests fails the instrument displays:
FATAL ERROR, RAM ERROR or FATAL ERROR, EPROM ERROR, respectively.
Switch the instrument off and then on again. Replace the Keypad & Display Board if the error reoccurs.
The Keypad 8, Display CPU continuously checks the communication hardware including the ringnet by a
loop-back test. If this test fails the instrument displays: RINGNET COMMUNICATION ERROR.
Perform the ringnet test procedure (section 6.5.1-27) in order to diagnose whether the Ringnet Commu-
nication Error is caused by an error in the Keypad & Display hardware (cable or CPU board) or a problem
elsewhere in the ringnet.
IMPORTANT! During software updating the communication check is disabled and hence the
RINGNET COMMUNICATION ERROR will a/ways be displayed. Under this circumstance the error
message should be ignored.
Introduction
This section describes the dismantling and replacement procedures for those of the modules and
parts, for which it is not obvious how to proceed, or where special precautions should be taken.
Note positions when dismantling, attention should be paid to the positioning and connection of the
tubing and cables, use of screws, etc, which will ease reassembly. You should make a final inspection
after reassembly before switching on.
Special Tools
For further information on the special tools required in the procedures, please refer to the listing in
front of this Service Manual.
The Spare Parts List in Section 9 lists the parts available for servicing the ABLSOO. In addition, the
Spare Parts List states to which extent a unit should be serviced in the field.
Removal of Pumps
1 Select Other Service programs and select program 5 Service Setup. Set Rinse only at power
on to YES, Disregard gas & temp errors * to YES and Inlet and Cover off removal ig-
nored to YES, and then turn off the ABL.
2 Remove the inlet module.
3 Remove the pump tubing from the pump which is to be removed
4 Remove the two screws holding the pump housing, marked E for the left pump and F for
the right pump on drawing 22523-A4
5 Pull the pump straight forward.
0 15 Place the new o-ring, coming with inlet tube and LS4 on the front of the Main Valve.
16 Mount the inlet preheater with inlet tube and LS4 on front of the Main Valve.
17 Pull the optical fibres through the wet section panel. Mount the clips on the fibres and connect
the fibres to the Wet Section Distribution PCB.
18 Pull the cable for the inlet preheater through the wet section panel and mount it on the Wet
Section Distribution PCB.
19 Reconnect tubing to the inlet preheater.
20 Remount the clear tubing, coming from Pl, to the waste stopper.
21 Remount the inlet module and the inlet washer.
22 Before mounting of the Wet Section on the Base Unit perform check of the Liquid sensor, Wet
Section Service Program No. 8.
CAUTION: When taking the Main Valve apart it must be done in a clean area.
Removal of Floppy Disk Unit (ABLSOO Run 1 - 49, ABL51 O/520 Run l-7)
1 Remove the printer unit
2 Disconnect the cable between Floppy Disk Control PCB and Main Computer PCB at the Floppy
Disk Control PCB.
3 Remove the four 4 mm screws, two on each side holding the Floppy Disk Unit.
4 Slide the Floppy Disk Unit back out of the front panel and lift out.
Removal of Floppy Disk Unit (ABLSOO Run 50 onwards, ABL510/520 Run 8 onwards,
ABL505)
1 Disconnect the cable between floppy disk drive and main computer at the Floppy Disk Drive.
2 Remove the two 4 mm screws, one on each side holding the Floppy Disk Drive.
3 Slide the Floppy Disk Drive back out of the front panel and lift out.
Epson SMD340-100
FIGURE 1
Straps:
Revised 92.02.14
4-9
DISMANTLING AND REPLACEMENTS
Note: This alignment should only be performed when the CRT monitor has been replaced.
Inspect the mounting of the right side rollers, if the screws are mounted from the top remove
the screws and interchange the top and the bottom metal plates. Then mount the two screws
from the bottom.
Loosen the two screws holding the rollers in each side. Loosen the screws holding the vertical
mounting plates marked with A, see next page.
Mount the printer in the CRT compartment.
Mount spring (920-713) between the hole in the middle of the holder for right side rollers and
the hook in the middle of the holder for left side rollers.
Adjust the printer unit up, down and sideways to center the unit in the cutout in the front
panel. Tighten the screws for the elevation plates A. Tighten the screws for the right rollers
from the bottom and then the screws for the left rollers from the top.
Remove the printer unit from the CRT compartment and install the CRT compartment on the
main computer unit in reverse order of removal of the compartment.
l
3 After installation turn on the ABL and select Wet Section Service program 13, Photodiode
Checkout.
4 Press Lamp on/off to turn on the photometer lamp.
5 Loosen the lock nut on the mirror adjustment screw (item No. 1 on drawing 23742-A4). Adjust
the mirror for maximum mean diode current.
6 If the Optical Fibre has been replaced the tip of the Optical Fibre at the Hemolyzer Unit must
also be adjusted. Loosen the screw holding the optical fibre to the hemolyzer. Adjust the fibre
to maximum mean diode current and make sure that the fibre is not touching the surface of
the hole in the Hemolyzer. There must be light all the way around the tip of the Optical Fibre.
7 Perform Monochromator check page 2.16.
2. Install the new main computer PCB in ABLSXX. If the ABL does not have a floppy disk in-
stalled, connect the service disk drive. If the Main Computer PCB already has a software
loaded then install the Boot PCB 902-198 as well.
First-time loading of software in a spare part PCB:
3. Insert the program disk in the disk drive.
4. Turn the main power on. The screen below will appear.
Enter Command :
1) Diskboot
1:
Screen picture
5. Select Diskboot by pressing softkey 1.
The interpreter program FLASH.HEX will be loaded into the ABL ram. The loading process
can be followed on the screen.
6. By using a small screwdriver or similar, set the SW switch on the rear panel (next to the
serial number plate on the rear panel) ON (right position as seen from the back of the
ABL) when requested on the screen.
The SW switch is an on/off switch for the FEPROM programming voltage.
CAUTION: Do NOT set SW-switch OFF and
Do NOT turn power OFF and
Do NOT remove floppy disk
during software upgrade.
After a while the FEPROM erasing/programming/verifying procedure starts. It takes ap-
proximately 4 minutes to complete the procedure.
7. Set the SW switch on the rear panel OFF (left position as seen from the back of the ABL)
when instructed to do so by the screen text.
8. Turn the main power off and remove the Boot Board.
9. For ABLs without disk drive: Remove the upgrade disk from the service disk drive, dis-
connect the two cables and reconnect the blind plate into
the ABL front panel.
For ABLs with disk drive: Replace the software upgrade disk with the normal data
disk.
10. Restart the the ABL by turning the main power on again. The restart time is 15 minutes.
General:
11. On the removed Main Computer PCB, move strap SW705 to position 2 and 3 thereby dis-
connecting the battery.
In case of an error during the upgrade procedure, an error code will be displayed on the
screen. For decoding, please refer to the Service Manual section 3.9, page 8.
Setting up Please refer to the Users Handbook for detailed information about how to set-up the
ABL, Book 1, set-up programs.
AEL60 (one RS150 only) and first option in ABL~OO run 50 onwards,
ABL510/520 run 8 onwards and ABL505.
1 Switch off the ABLSXX and disconnect it from the mains supply.
2 Remove the blind plate for the second option slot on the rear, counting
from the terminal plate.
3 Withdraw the flat cable marked 1 from within the Base Unit and con-
nect it to the matching connector on RSISO/lOO. It may be necessary to
remove the blind plate of the neighbour option slot in order to get hold
of the cable.
4 Insert the RSl!SO/lOO into the slot and secure it. Remount also the blind
plate of the neighbour option slot, if removed.
5 Connect the ABLSXX to the mains supply and switch on.
6 Verify that the channel(s) are operating correctly by using the loop-back
test service program (other service programs No. 1).
Procedure:
Warning: parts inside connected to mains. Switch main supply off. Remove the power cable.
Remove terminal panel (978-683) to be able to see the internal power plugs on the Power Supply PCB.
Remove Main Computer Board (902-224) from the rear of the Measuring Station.
Remove internal power cable (636-301) from Main Computer Board to Power Supply.
Mount the internal cable from the Cannon connector Jl mounted on the rear of the Base Unit to the
Power Supply PCB connectors J3 & J4.
Mount the cable #2 in the Panel with Software Gate, if any. (Software gate is necessary if floppy disk is
used in Control Station).
Install the Software Gate Panel and the Blind Panel(s) in the rear of the Measuring Station.
Mount the cable (617-877) between the Control Station and the Measuring Station.
Turn power on. Check the software versions in Wet Section and Control Station according to chapter
9.4 in the Service Manual.
Perform normal start-up procedure. Check that the total calibration results are without any question
marks.
Fill in the ABL50/500 Up/Downgrade Report. Sign it with name and date. File it in a central file.
Procedure:
Warning: parts inside connected to mains. Switch main supply off. Remove the power cable.
Remove terminal panel (978-683) to be able to see the internal power plugs on the Power Supply PCB.
Remove the two most left modules from the rear of the Measuring Station. If any of the modules
contain electronic parts, remount it/them in the next vacant slot(s).
Unplug the internal power cable from the Power Supply PCB J3 & J4 (two green plugs). Fasten the
cable with a tyrap as shown in figure 1 to prevent the unconnected end to move around inside the unit.
Do not remove the other end with Cannon plug attached to the rear of the Base Unit.
Mount the internal power cable (636-301) from Power Supply J3 & J4 to Main Computer Board. Fasten
the cable to the cable from Power Supply J2 with a tyrap as shown in figure 2.
Mount the Main Computer board (902-224) in the two most left slots.
Insert Memory Card with Software in the main Computer. While pressing the switch through the hole
above the Memory Card, turn power on. Release the switch after 5 seconds. Wait approximately two
minutes until one beep is heard. Switch power off. Remove Memory Card. If the beeper beeps with an
interval of 1 Hz, the programming has failed, please refer to chapter 2.17 Software Upgrade in the
Service Manual.
Mount the cable between the Keypad with Display and the Main Computer.
Cover the name ABLSOO on the serial number plate with the black lable 980-669.
Turn power on. Display will show: System setup invalid, default selected. Press continue to confirm!
Check the software versions in Wet Section and Control Station according to chapter 9.4 in the Service
Main -.-
Keypad -*-
WS program -.-
WS control -*-
RSl50 -.-
List the software versions shown in Other Service Programs 4 Software Versions:
The undersigned confirms that the Up/Downgrade has been carried out according to the
procedure described in the Service Manual page 4-21 and that the information stated is
correct.
Name (Block letters)
Date: Signature
RADIOMETERS
0
COPENHAC3EN;
0 5. MAINTENANCE
Introduction
Maintenance of the ABL should normally be performed by the user as described in the ABL Users
Handbook and in the ABL Maintenance Schedules. In addition to this maintenance, the service engin-
eer should always make a maintenance check after having serviced the instrument. The yearly main-
tenance programme must also be carried out by the service engineer.
Maintenance Check
Clean the inlet module and the inlet flap. If necessary, replace the inlet washer.
Check that there are no deposits present on the electrode chambers or in the KCI well. Clean
with distilled water.
Check that the rollers of the peristaltic pumps rotate freely. Check that the pump tubing is
properly mounted.
Check the setup with the printout taken before performing maintenance.
Revised 91.04.08
5.2. YEARLY MAINTENANCE
a
Yearly Maintenance
The following instructions are to be used together with the Code 904-882, Yearly Maintenance Service
Kit, when carrying out the recommended ABL yearly overhaul. TheService Manual will also be required
during the procedures. The page references used throughout these instructions refer to those of the
Service Manual. The maintenance procedures and replacements described should be performed only
by service engineers from RADIOMETER or its distributors.
WARNING: Avoid direct contact with blood-contaminated parts. Use approved rubber gloves to
avoid infection!
Survey
1 Decontamination
2 Emptying of the Wet Section
3 Cleaning inside the ABL
4 Replacements in the Wet Section and the Gas Mixer
5 Electrical and mechanical check
6 Cleaning the exterior of the ABL
7 Final check
8 Yearly Maintenance Service Kit
6 Remove the Calomel Reference Electrode from the ABL and place it in the protecting sheath. Empty
the well with a syringe.
0
REVISED 92.01.27 5.2 - 1
YEARLY MAINTENANCE
7 Remove all the electrodes. Place the pC0, and p0, electrodes in their protecting jackets. Lay aside
the electrodes.
8 Carefully remove the waste container. Due to the infection risk, disposal of the waste container must
be in accordance with your local regulations. Replace with a new item from the kii when asked to do
so.
9 Close the Main Valve on the CO, cylinder, and remove the cylinder from the ABL.
10 Select Other Service Programs No. 5 and set Rinse only at power on to Yes.
11 Turn off the main power switch on the ABL and disconnect the main power cable from the wall outlet.
2 Remove the left-hand panel from the base unit, see page 4-2.
3 Remove the right-tiand panel from the base unit, see page 4-2.
5 Clean the rear side of the wet section with a soft bmsh. Be careful not to damage the optical fibres.
7 Remove the inlet module and discard the inlet washer (836-l 47). Replace with new item from the kit
when asked to do so.
8 Remove the tubing between Pump 1 and electrode chambers (LSl) and discard the tubing.
9 Remove tubing between Pump 1 and waste, and discard the tubing.
12 Clean the electrode chambers using a cotton stick wetted with protein remover. Rinse afterwards using
a cotton stick and distilled water.
15 Pull up the clamp located at the top front of the CRT (Marked with Danger - High Voltage) and slide
up the the protecting screen in front of the CRT.
16 Clean the front of the CRT using a cloth moistened with mild soapy water.
17 Clean the protecting screen on both sides using a cloth moistened with mild soapy water.
2 Replace the gasket with tracks (834-612), the support disc (834-613), and the gasket for the front part
(834-620) in the Main Valve, see page 4-6. Do not perform the leakage test of the Main Valve until asked
to do so.
6 Cut 240 mm of tubing (840-227) and replace the tube between Pump 1 and the waste stopper.
9 Install a new tubing between Pump 1 and electrode chambers (LSl) (841-757 Vl Tubing).
10 Replace V3 Tubing (ABL505,841-761) (ABL51 O/520,841 -760) or tubing between Rinse Preheater
and V4 (ABL50/500,841-759).
12 Fill up the reference electrode with Salt-Bridge Solution to the red mark. Mount the reference electrode.
14 Mount pH electrode.
2 Clean the front screen using a cloth moistened with mild soapy water.
3 Clean the painted surface, using a cloth moistened with mild soapy water.
4 Clean the outside of the power cable using a cloth moistened with mild soapy water.
Final Check
1 Remount the power cable and turn on the ABL.
4 Check the set-up report with the print-out taken in step 1.2. Correct any discrepancies.
5 Record date and time for the annual overhaul in the maintenance log.
0
Qualicheck is a trademark of Radiometer A/S, Denmark and is registered in USA and in some other countries.
0 5.2 - 5
REVISED 92.01.27
YEARLY MAINTENANCE
The software of the ABL includes a number of programs available for the user and/or the service engineer.
The various programs are accessed using the softkeys below the CRT screen or display.
Survey
Section 6.1: provides an overview of the various calibration programs (which electrodes are
calibrated in which program).
Section 6.2: provides a brief description of the various wet section control programs.
This Section provides an overview of the calibration programs included in the software. It indicates the
extend of the calibration of each electrode.
NOTE: For information about the various other user programs (description and use), please refer to
the appropriate User Handbook.
The calibration programs are called automatically according to the setup program. They may also be called
manually via the softkeys. The purpose of the various calibration programs is indicated below:
Please Note that the two gas calibrations, Gas Cal 1 and Gas Cal 2, are not available in ABL50.
Introduction.
This section includes flow charts, tables and descriptions of the wet section activities during
measurements for ABL500/505/5 1 O/520.
The flow charts include information about the important wet section activities such as liquid sensors,
reference data collection, electrode data collection etc. Timing information for the main events are
included as well.
Main computer activities (softkeys, screen texts etc.) are included if important for the overall
understanding of the wet section function.
The tables include information about duration and liquid consumption for the individual wet section
programs.
The program descriptions at the end of the section include a summary, the duration, the liquid
consumption (if any) and a list of the possible errors which can turn up during the program execution.
1. The inlet flap is opened, the sample is injected or aspirated and the inlet flap is closed.
2. The sample is in the electrode chambers 9.3 seconds after the inlet flap has been closed.
3. The measurement result will be displayed 40 - 54 seconds after the inlet flap has been closed
depending on the pO2IpCO2 levels.
Sample injection:
Sample aspiration:
I MEASUREMENT
RINSE I
70.7 73.1 SBS
RINSE
54
1160
54
1177
3. The measurement result will be displayed 30 seconds after the inlet flap has been closed.
1. The inlet flap is opened, the capillary tube is inserted in the inlet and the aspirate button is
activated.
3. The measurement result will be displayed 45 (ABL500)/46 (ABL510/520) seconds after the inlet
flap has been closed.
1. The inlet flap is opened, the capillary tube is inserted in the inlet and the aspirate button is
activated.
3. The measurement result will be displayed 39 seconds after the inlet flap has been closed.
Summary:
1. The inlet flap is opened, the sample is injected or aspirated and the inlet flap is closed.
2. The sample is placed in the electrode chambers 20.8 seconds after the inlet flap has been
closed.
3. The measurement result will be displayed 53 - 65 seconds (depending on the pCdpC0, level)
after the inlet flap has been closed.
Sample injection:
MEASUREMENT 79 SBS 55
RINSE I RINSE 1438
l ) The time to open the inlet flap, inject the sample and close the inlet flap again is
estimated to 3 seconds (INITIALIZATION + INJECTION).
Sample aspiration:
I TOTAL
I
112.1 SBS
RINSE
76
1438
Summary:
1. The inlet flap is opened and the aspiration key is activated within 0.9 second.
3. The measurement result will be displayed 30 seconds after the inlet flap is closed.
Summary:
1. The inlet flap is opened and aspiration is started within 0.9 second.
2. The sample is aspirated and the inlet flap is closed within 10.3 seconds.
3. The sample is placed in the electrode chambers 14.7 seconds after the inlet flap has been
closed.
4. The measurement result will be displayed 47 seconds after the inlet flap has been closed.
Summary:
1. The inlet flap is opened and aspiration is started within 0.9 second.
2. The sample is aspirated and the inlet flap is closed within 10.8 seconds.
3. The sample is placed in the electrode chambers 15 seconds after the inlet flap has been
closed.
4. The measurement result will be displayed 48 - 59 seconds (depending on the @#CO, level)
after the inlet flap has been closed.
BG INJECTION, ABL500/510/520
Prior to LS4 signal or aspirate key activation, the user can select (by softkeys) gas measurement, pH-
only measurement, or oxi-only measurement.
SUMMARY OF EVENTS: After sample injection (LS4 signal) the busy LED is switched on and
the following message appears on the screen: CLOSE INLET FLAP
NOMINAL LIQUID
CONSUMPTION: None.
SYSTEM ERRORS: If the inlet flap is closed before LS4 is full, the program is aborted
and the following error messages appear:
Screen status field:
Measurement aborted
System status:
INLET CLOSED
INLET CLOSED WITHOUT INTRODUCING ENOUGH SAMPLE IN
LAST MEASUREMENT
If the inlet flap is not closed within 20 seconds (acoustic alert after 10
seconds) from sample injection, the program is aborted and the
following error messages appear:
Screen status field:
Measurement aborted
System status:
INLET NOT CLOSED
INLET FLAP NOT CLOSED WHEN REQUESTED IN LAST
MEASUREMENT
SUMMARY OF EVENTS: The sample is aspirated by Pl. After aspiration (LS4 signal) the busy
LED is switched on and the following message appears on the
screen: CLOSE INLET FLAP
When the inlet flap is closed, the program jumps to BG MEASURE.
NOMINAL LIQUID
CONSUMPTION: None.
SYSTEM ERRORS: If the inlet flap is closed before activation of the aspiration key, the
program is aborted and the following error messages appear:
Screen status field:
Measurement aborted
System status:
INLET CLOSED
INLET CLOSED WITHOUT INTRODUCING ENOUGH SAMPLE IN
LAST MEASUREMENT
If the inlet flap is closed before LS4 is full, the program is aborted
and the following error messages appear:
Screen status field:
Measurement aborted
System status:
INLET CLOSED
INLET CLOSED WITHOUT INTRODUCING ENOUGH SAMPLE IN
LAST MEASUREMENT
If the inlet flap is not closed within 20 seconds (acoustic alert after 10
seconds) from CLOSE INLET FLAP appears on the screen, the
program is aborted and the following error messages appear:
Screen status field:
Measurement aborted
System status:
INLET NOT CLOSED
INLET FLAP NOT CLOSED WHEN REQUESTED IN L A S T
MEASUREMENT
SUMMARY OF EVENTS: The sample is aspirated by Pl. After aspiration (LS4 signal) the busy
LED is switched on and the following message appears on the
screen: CLOSE INLET FLAP
When the inlet flap is closed, the program jumps to BG MEASURE.
NOMINAL LIQUID
CONSUMPTION: None.
a
-
ENTRY FROM: BG INJECTION.
SYSTEM ERRORS: If the inlet flap is closed before activation of the aspiration key, the
program is aborted and the following error messages appear:
Screen status field:
Measurement aborted
System status:
INLET CLOSED
INLET CLOSED WITHOUT INTRODUCING ENOUGH SAMPLE IN
LAST MEASUREMENT
If the inlet flap is closed before LS4 is full, the program is aborted
and the following error messages appear:
Screen status field:
Measurement aborted
System status:
INLET CLOSED
INLET CLOSED WITHOUT INTRODUCING ENOUGH SAMPLE IN
LAST MEASUREMENT
if the inlet flap is not closed within 20 seconds (acoustic alert after 10
seconds) from CLOSE INLET FLAP appears on the screen, the
program is aborted and the following error messages appear:
Screen status field:
Measurement aborted
System status:
INLET NOT CLOSED
INLET FLAP NOT CLOSED WHEN REQUESTED IN LAST
MEASUREMENT
BG MEASURE, ABLSOO
The sample is sucked from inlet to the electrode chambers.
NOMINAL LIQUID
CONSUMPTION: None.
SYSTEM ERRORS: If LS2 senses liquid before the sample has arrived, the program is
aborted and the following error messages appear:
Screen status field:
“Measurement aborted”
System status:
“MEASURE ERROR L”
“UNEXPECTED LIQUID DURING SAMPLE PROCESSING (LS2)”
If LS2 does not sense liquid when supposed to, during sample
transportation, the program is aborted and the following error
messages appear:
Screen status field:
“Measurement aborted”
System status:
“MEASURE ERROR N”
“NO SAMPLE DETECTED DURING SAMPLE PROCESSING”
If LSl senses liquid before the sample has arrived, the program is
aborted and the following error messages appear:
Screen status field:
“Measurement aborted”
System status:
“MEASURE ERROR K”
“UNEXPECTED LIQUID DURING SAMPLE PROCESSING (LSI)”
If LSl does not sense liquid when supposed to, during sample
transportation, the program is aborted and the following error
messages appear:
Screen status field:
Measurement aborted
System status:
MEASURE ERROR N
NO SAMPLE DETECTED DURING SAMPLE PROCESSING
If LS2 does not sense liquid when LSl does, during electrode
chamber filling, the following error messages appear:
Screen status field:
Insufficient sample
System status:
INSUFFICIENT SAMPLE
Sample volume too small
BG MEASURE, ABL510/520
SUMMARY OF EVENTS: LS2 is tested for liquid. Must be empty.
Blood sample is sucked from inlet to LS2 by P2. Aircount is started
(LS2) and the sample is sucked/pushed by PI/P2 through the
electrode chambers and the cuvette. LSI, LS2 and LS3 are tested for
liquid (all must be full). The sample is parked in the electrode
chambers and the cuvette, aircounting is stopped and electrode data
collection starts. P2 establishes pressure in the cuvette and V3 closes
The program jumps to MEASUREMENT RINSE where the OXI sample
is hemolyzed and data collection is completed.
NOMINAL LIQUID -
CONSUMPTION: 24 ~1 rinse solution (used to maintain the hemolyzer pressure).
SYSTEM ERRORS: If LS2 senses liquid before the sample has arrived, the program is
aborted and the following error messages appear:
Screen status field:
Measurement aborted
System status:
MEASURE ERROR L
UNEXPECTED LIQUID DURING SAMPLE PROCESSING (LS2)
If LSl senses liquid before the sample has arrived, the program is
aborted and the following error messages appear:
Screen status field:
Measurement aborted
System status:
MEASURE ERROR K
UNEXPECTED LIQUID DURING SAMPLE PROCESSING (LSl)
If LS3 senses liquid before the sample has arrived, the program is
aborted and the following error messages appear:
Screen status field:
Measurement aborted
System status:
MEASURE ERROR M
UNEXPECTED LIQUID DURING SAMPLE PROCESSING (LS3)
If LSl, LS2 or LS3 do not sense liquid when supposed to, during
sample transportation, the program is aborted and the following
error messages appear:
Screen status field:
Measurement aborted
System status:
MEASURE ERROR N
NO SAMPLE DETECTED DURING SAMPLE PROCESSING
If LS2 does not sense liquid when LSl and LS3 do, during electrode
chamber filling, the following error messages appear:
Screen status field:
Insufficient sample
System status:
INSUFFICIENT SAMPLE
Sample volume too small
SUMMARY OF EVENTS: When the ASP key has been activated, INJECT GAS SAMPLE SLOW-
LY appears on the screen.
The inlet tubings are emptied by P2. LSl, LS2 and LS4 are tested for
liquid.The gas sample is introduced and sucked through the elec-
trode chambers by Pl for 25 seconds.
The tubings are tested for liquid once again and CLOSE INLET
FLAP appears on the screen.
Electrode data collection is started and the program jumps to
MEASUREMENT RINSE during which the data collection is com-
-
pleted.
NOMINAL LIQUID
CONSUMPTION: None
SYSTEM ERRORS: If the inlet flap is closed before activation of the aspiration key, the
program is aborted and the following error messages appear:
Screen status field:
Measurement aborted l
System status:
INLET CLOSED
INLET FLAP CLOSED DURING INJECTION IN LAST GAS
MEASUREMENT
If the inlet flap is not closed within 20 seconds (acoustic alert after 10
seconds) from CLOSE INLET FLAP appears on the screen, the
program is aborted and the following error messages appear:
Screen status field:
Measurement aborted
System status:
INLET NOT CLOSED
INLET FLAP NOT CLOSED WHEN REQUESTED IN LAST
MEASUREMENT
SUMMARY OF EVENTS: After sample aspiration (LS4 signal) the busy LED is switched on and
the following message appears on the screen: CLOSE INLET FLAP
When the inlet flap is closed, the program jumps to pH MEASURE.
NOMINAL LIQUID
CONSUMPTION: None.
NOMINAL DURATION: ABLSOO: 5.6 seconds from activation of the aspiration key.
ABL510/520: 6.7 seconds from activation of the aspiration key.
SYSTEM ERRORS: If the inlet flap is closed before activation of the aspiration key, the
program is aborted and the following error messages appear:
Screen status field:
Measurement aborted
System status:
INLET CLOSED
INLET CLOSED WITHOUT INTRODUCING ENOUGH SAMPLE IN
LAST MEASUREMENT
If the inlet flap is closed before LS4 is full, the program is aborted
and the following error messages appear:
Screen status field:
Measurement aborted
System status:
INLET CLOSED
INLET CLOSED WITHOUT INTRODUCING ENOUGH SAMPLE IN
LAST MEASUREMENT
If the inlet flap is not closed within 20 seconds (acoustic alert after 10
seconds) from CLOSE INLET FLAP appears on the screen, the
program is aborted and the following error messages appear:
Screen status field :
Measurement aborted
System status:
INLET NOT CLOSED
INLET FLAP NOT CLOSED WHEN REQUESTED IN LAST
MEASUREMENT
SUMMARY OF EVENTS: After sample aspiration (LS4 signal) the busy LED is switched on and
the following message appears on the screen: CLOSE INLET FLAP
When the inlet flap is closed, the program jumps to pH MEASURE.
NOMINAL LIQUID
CONSUMPTION: None.
NOMINAL DURATION ABL500: 5.6 seconds from activation of the aspiration key.
ABL510/520: 6.7 seconds from activation of the aspiration key
SYSTEM ERRORS: If the inlet flap is closed before activation of the aspiration key, the
program is aborted and the following error messages appear:
Screen status field:
Measurement aborted
System status:
INLET CLOSED
INLET CLOSED WITHOUT INTRODUCING ENOUGH SAMPLE IN
LAST MEASUREMENT
-
WET SECTION MEASUREMENT PROGRAMS
If the inlet flap is closed before LS4 is full, the program is aborted
and the following error messages appear:
Screen status field:
Measurement aborted
System status:
INLET CLOSED
INLET CLOSED WITHOUT INTRODUCING ENOUGH SAMPLE IN
LAST MEASUREMENT
If the inlet flap is not closed within 20 seconds (acoustic alert after 10
seconds) from CLOSE INLET FLAP appears on the screen, the
program is aborted and the following error messages appear:
Screen status field:
Measurement aborted
System status:
INLET NOT CLOSED
INLET FLAP NOT CLOSED WHEN REQUESTED IN LAST
MEASUREMENT
pH MEASURE, ABL500/510/520
The sample is sucked from inlet to the electrode chambers.
NOMINAL LIQUID
CONSUMPTION: None.
SYSTEM ERRORS: If LS2 senses liquid before the sample has arrived, the program is
aborted and the following error messages appear:
Screen status field:
Measurement aborted
System status:
MEASURE ERROR L
UNEXPECTED LIQUID DURING SAMPLE PROCESSING (LS2)
If LSl senses liquid before the sample has arrived, the program is
-
aborted and the following error messages appear:
Screen status field:
Measurement aborted
System status:
MEASURE ERROR K
UNEXPECTED LIQUID DURING SAMPLE PROCESSING (LSl)
If LSl or LS2 do not sense liquid when supposed to, during sample
transportation, the program is aborted and the following error
messages appear:
Screen status field:
Measurement aborted
System status:
MEASURE ERROR N
NO SAMPLE DETECTED DURING SAMPLE PROCESSING
If LS2 does not sense liquid when LSl does, during electrode
chamber filling, the the following error messages appear:
Screen status field:
Insufficient sample
System status:
INSUFFICIENT SAMPLE
Sample volume too small
If 4.2 % < aircounts < 14.3 % of the total counts during aircount-
ing, the following error messages appear:
Screen status field:
Inhomogeneous sample
System status:
INHOMOGENEOUS SAMPLE
Air bubbles detected in sample
If aircounts > 14.3 % of the total counts during aircounting and LS2
is empty after the aircounting, the following error messages appear:
Screen status field:
Insufficient sample
System status:
INSUFFICIENT SAMPLE
Too small sample volume detected
If aircounts > 14.3 % of the total counts during aircounting and LS2
is full after the aircounting, the following error messages appear:
Screen status field:
Inhomogeneous sample
System status:
INHOMOGENEOUS SAMPLE
Air bubbles detected in sample
SUMMARY OF EVENTS: The sample is aspirated by Pl. After aspiration (LS4 signal) the busy
LED is switched on and the following message appears on the
screen: “CLOSE INLET FLAP”
When the inlet flap is closed, the program jumps to OXI MEASURE.
NOMINAL LIQUID
CONSUMPTION: None.
SYSTEM ERRORS: If the inlet flap is closed before activation of the aspiration key, the
program is aborted and the following error messages appear:
Screen status field:
Measurement aborted
System status:
INLET CLOSED
INLET CLOSED WITHOUT INTRODUCING ENOUGH SAMPLE IN
LAST MEASUREMENT”
If the inlet flap is closed before LS4 is full, the program is aborted
and the following error messages appear:
Screen status field:
Measurement aborted
System status:
INLET CLOSED
INLET CLOSED WITHOUT INTRODUCING ENOUGH SAMPLE IN
LAST MEASUREMENT
If the inlet flap is not closed within 20 seconds (acoustic alert after 10
seconds) from CLOSE INLET FLAP appears on the screen, the
program is aborted and the following error messages appear:
Screen status field:
Measurement aborted
System status:
INLET NOT CLOSED
INLET FLAP NOT CLOSED WHEN REQUESTED IN LAST
MEASUREMENT
SUMMARY OF EVENTS: The sample is aspirated by Pl. After aspiration (LS4 signal) the busy
LED is switched on and the following message appears on the
screen: CLOSE INLET FLAP
When the inlet flap is closed, the program jumps to OXI MEASURE.
NOMINAL LIQUID
CONSUMPTION: None.
-
ENTRY FROM: BG INITIALIZATION.
SYSTEM ERRORS: If the inlet flap is closed before activation of the aspiration key, the
program is aborted and the following error messages appear:
Screen status field:
Measurement aborted
System status:
INLET CLOSED
INLET CLOSED WITHOUT INTRODUCING ENOUGH SAMPLE IN
LAST MEASUREMENT
If the inlet flap is closed before LS4 is full, the program is aborted
and the following error messages appear:
Screen status field:
Measurement aborted
System status:
INLET CLOSED
INLET CLOSED WITHOUT INTRODUCING ENOUGH SAMPLE IN
LAST MEASUREMENT
If the inlet flap is not closed within 20 seconds (acoustic alert after 10
seconds) from CLOSE INLET FLAP appears on the screen, the
program is aborted and the following error messages appear:
Screen status field:
Measurement aborted
System status:
INLET NOT CLOSED
INLET FLAP NOT CLOSED WHEN REQUESTED IN LAST
MEASUREMENT
NOMINAL LIQUID
CONSUMPTION: 10 bl rinse solution (used to maintain the hemolyzer pressure).
SYSTEM ERRORS: If LS2 senses liquid before the sample has arrived, the program is
aborted and the following error messages appear:
Screen status field:
“Measurement aborted”
System status:
“MEASURE ERROR 1”
“UNEXPECTED LIQUID DURING SAMPLE PROCESSING (LSZ)”
If LS3 senses liquid before the sample has arrived, the program is
aborted and the following error messages appear:
Screen status field:
Measurement aborted
System status:
MEASURE ERROR M
UNEXPECTED LIQUID DURING SAMPLE PROCESSING (LS3)
If LS2 does not sense liquid when LS3 does, during hemolyzer filling,
the following error messages appear:
Screen status field:
Insufficient sample
System status:
INSUFFICIENT SAMPLE
Sample volume too small
If 4.2 % < aircounts < 14.3 % of the total counts during aircount-
ing, the following error messages appear:
Screen status field:
Inhomogeneous sample
System status:
INHOMOGENEOUS SAMPLE
Air bubbles detected in sample
If aircounts > 14.3 % of the total counts during aircounting and LS2
is empty after the aircounting, the following error messages appear:
Screen status field:
Insufficient sample
System status:
INSUFFICIENT SAMPLE
Too small sample volume detected
If aircounts > 14.3 % of the total counts during aircounting and LS2
is full after the aircounting, the following error messages appear:
Screen status field:
Inhomogeneous sample
System status:
INHOMOGENEOUS SAMPLE
Air bubbles detected in sample
SUMMARY OF EVENTS: Busy lamp on, measure light on, ref. data collection.
Gas mixer pump start and GAS1 selection.
INLET RINSE:
2x4 rinse/air segments created by P2, V4, V5 through inlet and out
to waste through W7.
P2 is dosing rinse sol. into W7 tubing. LS4 is tested for liquid.
P2 sucks rinse sol. back from W7 tubing. MV alternates between
position 0 and 1, and 4 rinse sol./GASl segments are hence
created. P2 changes direction and the tubings are emptied.
P2 sucks rinse solution again. P2 changes direction and the rinse -
sol. is pumped back.
Electrode datacollection stop.
ELECTRODE CHAMBER RINSE:
Pl expels the sample/sucks rinse solution parked in the WO
tubing. P2, V4, V5 make 3 rinse/air segments.
Pl/P2 suck/push rinse/air segments through electrode chambers.
Pl sucks rinse sol./GASl via MV. LS2 is tested for liquid.
Pl and MV suck 3 rinse/GAS1 segments through preheater 2 and
electrode chambers. P2 sucks rinse sol. from V4.
P2 changes direction and doses the rinse sol. into the electrode
chambes. Pl emties the electrode chambers.
SBS front is changed. LS5 checks the SBS homogeneity.
Pl and MV create B rinse sol./GASl segments which are
sucked/pushed forward and backward through the electrode
chambers. Finally Pl empties the electrode chambers. LS2 and LSl
are tested for liquid.
NOMINAL DURATION: 70.7 seconds (incl. 1.2 second pause after inlet rinse)
NOMINAL LIQUID
CONSUMPTION: 1160 ~1 rinse solution.
54 II SBS.
SYSTEM ERRORS: Errors occurring during rinse programs do not cause aborts in order
not to create infinite loops (an aborted wet section program will
always cause a rinse program to be called).
If the inlet flap is opened during the inlet rinse (first 30 seconds of
the Measurement Rinse), another Measurement Rinse will be called
in succession of the first.
If the inlet flap is open when the Measurement Rinse has been
finished (when the program jumps to READY), the ABL goes into
“hold” and must be restarted.
If LSl, LS2 or LS4 do not sense liquid when expected, the following
error messages appear:
Screen status field: “No rinse solution”
System status:
“RINSE ERROR A”
“No rinse solution detected”
If LS2 does not sense liquid while LSl does, the following error
messages appear:
Screen status field: “No rinse solution”
System status:
“RINSE ERROR B”
“Malfunction in liquid system while fetching rinse solution”
If LSl, LS2 or LS4 sense liquid when not expected, the following error
messages appear:
System status:
“RINSE ERROR C”
“Unexpected liquid during rinse”
If LS5 does not sense liquid while L51 does, the following error
messages appear:
Screen status field: “No SBS solution”
System status:
“SBS ERROR B”
“Malfunction in liquid system while fetching SBS”
If neither L55 nor L51 sense liquid when expected, the following
error messages apper:
Screen status field: “No SBS solution”
System status:
“SBS ERROR A”
“No Salt Bridge Solution Detected”
SUMMARY OF EVENTS: Busy lamp on, measure light on, ref. data collection.
Gas mixer pump start and GAS1 selection.
Hemolyzation.
INLET RINSE:
2x4 rinse/air segments created by P2, V4, V5 through inlet and out
to waste through W7.
P2 isdosing rinse sol. into W7 tubing. LS4 is tested for liquid.
P2 sucks rinse sol. back from W7 tubing. MV alternates between
position 0 and 1, and 4 rinse sol./GASl segments are hence
created. P2 changes direction and the tubings are emptied.
OXI data collection while gasmixer and other pumps are stopped in
a
-
NOMINAL LIQUID
CONSUMPTION: 1177 ~1 rinse solution.
54 ~1 SBS.
SYSTEM ERRORS: Errors occurring during rinse programs do not cause aborts in order
not to create infinite loops (an aborted wet section program will
always cause a rinse program to be called).
If the inlet flap is opened during the inlet rinse (first 30 seconds of
the Measurement Rinse), another Measurement Rinse will be called
in succession of the first.
If the inlet flap is open when the Measurement Rinse has been
finished (when the program jumps to READY), the ABL goes into
“hold” and must be restarted.
If LSl, LS2 or LS4 do not sense liquid when expected, the following
error messages appear:
Screen status field:
No rinse solution”
System status:
RINSE ERROR A No rinse solution detected
If LS2 does not sense liquid while LSl does, the following error
messages appear:
Screen status field:
“No rinse solution”
System status:
RINSE ERROR B Malfunction in liquid system while fetching
rinse solution”
If LSl, LS2 or LS4 sense liquid when not expected, the following error
messages appear:
System status:
RINSE ERROR C Unexpected liquid during rinse
If LS5 does not sense liquid while LSl does, the following error
messages appear:
Screen status field:
No SBS solution”
System status:
“SBS ERROR B Malfunction in liquid system while fetching SBS
If neither LS5 nor LSl sense liquid when expected, the following
error messages apper:
Screen status field: No SBS solution”
System status:
“SBS ERROR A No Salt Bridge Solution Detected
LS3 is tested for liquid after the OXI datacollection. If it is not full the
following error messages appear:
System status:
“CAL ERROR L”
Leakage in OXI system
Introduction
This section includes flow charts, tables and descriptions of the wet section activities during
calibrations for ABL50/500/505/51 O/520
Two flow charts are included. One showing the activities which are automatically executed when the
ABL is in the READY state (calibrations, flushes, decompression etc.). The second flow chart shows the
activities which are executed in connection with power on and restart.
The tables include information about duration and liquid consumption for the individual wet section
programs.
The program descriptions at the end of the section include a summary, the duration, the liquid
consumption (if any) and a list of the possible errors which can turn up during the program execution.
GASCALl, summary
Auto GASCALl intervals are selectable in the SET UP menu (l/2, 1 or 2 hours).
The ~02 and the pCO2 electrodes are l-point calibrated by GAS1 (5.6 % CO2/19.7 % 02)
GM UPSTART 28 28 28 RINSE 51
GASCALZ, summary
Auto GASCALZ irftervals are selectable in the SET UP menu (1,2 or 4 hours).
WET SECTION
CALIBRATION
RINSE CAL1 2 2 CAL1 2 2 CAL1 22
CAL2 2 2 CAL2 2 2 CAL2 22
RINSE 592 RINSE 7 7 2 RINSE 592
CALI, summary
Auto CAL1 intervals are selectable in the SET UP menu (l/2, 1,2 or 4 hours).
, WET SECTION
I NOMINALDURATION IN SECONDS NOMINAL LIQUID CONSUMPTION IN (rl
~ PROGRAM
ABL50/500 ABLSOS ABL5 101520
GM UPSTART 28 28 28 RINSE 51
The calibration results are displayed 191 (ABL50/500) / 197 (ABL505) / 195 (ABL510/520) seconds
after start.
CALZ, summary
Auto CAL2 intervals are selectable in the SET UP menu (2,4, or 8 hours)
GA52 END
QH~ START
CALIBRATION
50.9
107.2
I 51.5
116.1
I 52.4
106.8
SBS
CAL2
5B5
6 6 5B5
3 8 2 CAL2
54 SBS
6 6 SBS
3 3 8 CAL2
55 SBS
66
396
54
RINSE CAL1 22 CAL1 22 CAL1 22
CAL2 22 CAL2 22 CAL2 22
RINSE 592 RINSE 772 RINSE 592
I
TOTAL 399.5 423.5 406.5 5B5 225 SBS 226 SBS 225
CAL1 349 CAL1 361 CAL1 356
CAL2 404 CAL2 360 CAL2 418
RINSE 808 RINSE 1120 RINSE 808
The calibration results are displayed 313 (ABL50/500) / 328 (ABL505) / 320 (ABL510/520) seconds
after start.
Auto TOTAL CAL intervals are selectable in the SET UP menu (8, 12, or 24 hours).
The ~02 electrode is 2-point calibrated by GAS1 (5.6 % CO2/19.7 % 02) and GAS3 (100 % CO$O %
02).
The pCO2 electrode is 2-point calibrated by GAS1 (5.6 % CO2) and GAS2 (11.2 % CO2).
The pH and electrolyte electrodes are 2-point calibrated by CAL1 solution (7.4 pH) and CAL2 solution
(6.8 pli).
GM UPSTART
GAS1 START -
GAS1 END
GAS2 START
GAS2 END
pH1 START +
GAS3 CAL 131.2 133.7 131.2 CAL1 210 CAL1 528 CAL1 210
CAL2 78 CAL2 156 CAL2 78
RINSE 477 RINSE 378 RINSE 477
The calibration results are displayed 427 (ABL50/500) 1435 (ABL505) 1434 (ABL510/520) seconds
after start.
AUTO CLEANING intervals are selectable in the SET UP menu (8, 12 or 24 hours).
GASFLUSH
The CO2 partial pressure in the electrode chambers and the inlet tubings is maintained. The ABL
remains READY during the entire program execution.
NOMINAL LIQUID
CONSUMPTION: None.
SYSTEM ERRORS: If LSl, LS2 or LS4 are not empty during the electrode chamber flush,
the following error messages appear:
System status:
FLUSH ERROR
Unexpected liquid during flush
REVISED 92.02.14
0 WET SECTION CALIBRATION PROGRAMS
FLUSH UPSTART
Gasmixer start and GAS1 (5.6% CO*) selection before GASFLUSH.
NOMINAL LIQUID
CONSUMPTION: None.
GAS1 CHECK
The program checks the 1 point calibration of the pCO2 and pO2 electrodes. The program is automati-
cally executed once every hour and is followed by FLUSH. The ABL remains READY during the entire
program execution.
NOMINAL LIQUID
CONSUMPTION: None.
SYSTEM ERRORS: If LSl is not empty during the electrode chamber flush, the program
is aborted and the following error messages appear:
Screen status field:
Calibration aborted
System status:
CAL ERROR A
Unexpected liquid in LSI
If LS2 is not empty during the electrode chamber flush, the program
is aborted and the following error messages appear:
Screen status field :
Calibration aborted
System status:
CAL ERROR B
Unexpected liquid in LSZ
FLUSH
Tubings and electrode chambersare flushed by fresh rinse solution in order to avoid measurement er-
rors caused by concentrated rinse solution. The program is automatically called once every hour and
the ABL is busy during the program execution.
SUMMARY OF EVENTS: Pl and P2 suck rinse solution via V4. Pl sucks rinse sol. via the main
valve. Tubings and electrode chambers are emptied. LSl test.
NOMINAL LIQUID
CONSUMPTION: 132 11 rinse solution.
l
-
ENTRY FROM: GAS 1 CHECK.
SYSTEM ERRORS: If LSl is full when not expected, the following error messages ap-
pear:
System status:
FLUSH ERROR
Unexpected liquid during flush
GAS1 START
1 point calibration (5.6 % CO2) of pCO2 and ~02 electrodes is started in this program.
SUMMARY OF EVENTS: Pl flushes the electrode chambers with GASI. Electrode data
collection is started. LSl and LS2 are tested for liquid (both must be
empty).
NOMINAL LIQUID
CONSUMPTION: None.
SYSTEM ERRORS: If LSl is not empty during the electrode chamber flush, the program
is aborted and the following error messages appear:
Screen status field:
Calibration aborted
System status:
CAL ERROR A
Unexpected liquid in LSI
If LS2 is not empty during the electrode chamber flush, the program
is aborted and the following error messages appear:
Screen status field:
Calibration aborted
System status:
CAL ERROR B
Unexpected liquid in LSZ
GM UPSTART
Gasmixer start and GAS1 (5.6% CO2) selection before GASISTART
NOMINAL LIQUID
CONSUMPTION: None.
GAS1 END
If the ABL is performing a GASCALZ, CALl, CAL2 or TOTAL CAL. the pO2 and pCO2 GAS1 electrode
data collection is completed in this program.
SUMMARY OF EVENTS: The gasmixer change to GASZ, 11.2 % CO2 (preparation of the 2nd
point gas calibration)
Electrode data collection is completed.
P2 flushes the inlet tubings with GASZ.
NOMINAL LIQUID
CONSUMPTION: None.
GAS2 START
The 2nd point calibration (11.2 % COz) of the pCO2 electrodes is started in this program.
SUMMARY OF EVENTS: Pl flushes the electrode chambers with GASZ. Electrode data
collection is started. LSl and LS2 are tested for liquid (both must be
empty).
NOMINAL LIQUID
CONSUMPTION: None.
SYSTEM ERRORS: If LSl is not empty during the electrode chamber flush, the program
is aborted and the following error messages appear:
Screen status field:
Calibration aborted
System status:
CAL ERROR A
Unexpected liquid in LSl
If LS2 is not empty during the electrode chamber flush, the program
is aborted and the following error messages appear:
Screen status field:
Calibration aborted
System status:
CAL ERROR B
Unexpected liquid in LSZ
GAS2 END
If the ABL is performing a CAL1 , CAL2 or TOTAL CAL. the GASZelectrode data collection is completed
in this program.
NOMINAL LIQUID
a CONSUMPTION: None
NOMINAL LIQUID
CONSUMPTION: 66 @I SBS.
327 ~1 CAL1 solution.
SYSTEM ERRORS: If LS2 does not sense liquid within 24 seconds during the electrode
chamber CAL1 solution flush, the calibration is aborted and the
following error messages appear:
Screen status field:
Calibration aborted
System status:
CAL ERROR D
No high CAL solution detected
If LS5 does not sense liquid within 20 seconds during the SBS front
replacement, the following error messages appear:
Screen status field:
No SBS solution
System status:
CAL SBS ERROR
No Salt-Bridge Solution during high pH calibration
If LSl does not sense liquid within 10 seconds during the electrode
chamber CAL1 filling, the calibration is aborted and the following
error messages appear:
Screen status field:
Calibration aborted
System status:
“CAL ERROR C”
“LSI did not sense high CAL solution
If LS2 does not sense liquid within 4 seconds during the electrode
chamber CAL1 filling or the liquid is not homogeneous, the ABL will
try another pHISTART by jumping to the first program step. Max. 3
attemps are performed. If still not successful1 the calibration is
aborted and the following error messages appear:
Screen status field:
Calibration aborted
System status:
“CAL ERROR I”
High CAL solution inhomogenous
This program is used to perform a pH and an OXI calibration using CAL1 (high pH) solution.
NOMINAL LIQUID
CONSUMPTION: 66 &I SBS.
334 11 CAL1 solution.
SYSTEM ERRORS: If LS2 does not sense liquid within 24 seconds during the electrode
chamber CAL1 flush, the calibration is aborted and the following
error messages appear:
Screen status field: Calibration aborted
System status:
CAL ERROR D
No high CAL solution detected
If LS5 does not sense liquid within 20 seconds during the SBS front
replacement, the following error messages appear:
Screen status field: No SBS solution
System status:
CAL SBS ERROR
No Salt-Bridge Solution during high pH calibration
If LSl does not sense liquid within 10 seconds during the electrode
chamber CAL1 filling, the calibration is aborted and the following
error messages appear:
Screen status field: Calibration aborted
System status:
CAL ERROR C
LSI did not sense high CAL solution
If LS3 does not sense liquid within 1 second during the cuvette
filling, the calibration is aborted and the following error messages
appear:
Screen status field: Calibration aborted
System status:
CAL ERROR E
LS3 did not sense high CAL solution
If LS2 does not sense liquid within 6 seconds during the electrode
chamber CAL1 filling or the liquid is not homogeneous, the ABL will
try another pHISTART by jumping to the first program step. Max. 3
attemps are performed. If still not successful1 the calibration is
aborted and the following error messages appear:
Screen status field: Calibration aborted
System status:
CAL ERROR I
High CAL solution inhomogenous
-
NOMINAL DURATION: 7 1.7 seconds.
NOMINAL LIQUID
CONSUMPTION: 39 @I SBS.
216 ~1 rinse solution.
SYSTEM ERRORS: If neither LS5 nor LSl sense liquid within 20 seconds during the SBS
front replacement, the following error messages appear:
Screen status field: No SBS solution
System status:
SBS ERROR A
No Salt-Bridge Solution detected
If LS5 does not and LSl does sense liquid within 20 seconds during 0
-
the SBS front replacement, the following error messages appear:
Screen status field: No SBS solution
System status:
SBS ERROR B
Malfunction in liquid system while fetching SBS
If neither LS2 nor LSl sense liquid within 24 seconds during the
electrode chamber rinse, the following error messages appear:
Screen status field: No rinse solution
System status:
RINSE ERROR A
No rinse solution detected
If LS2 does not and LSl does sense liquid within 24 seconds during
the electrode chamber rinse, the following error messages appear:
Screen status field: No rinse solution
System status:
RINSE ERROR B
Malfunction in liquid system while fetching rinse solution
If LS2 senses liquid where it should not while the electrode chamber
is emptied, the calibration is aborted and the following error
messages appear:
Screen status field: Calibration aborted
System status:
CAL ERROR B
Unexpected liquid in LSZ
If LSl senses liquid where it should not while the electrode chamber
is emptied, the calibration is aborted and the following error
messages appear:
Screen status field: Calibration aborted
System status:
CAL ERROR A
Unexpected liquid in LSI
If the ABL is performing a CAL2 or TOTAL CAL,.the electrode chambers and the OXI part are rinsed
before pH2START.
-
NOMINAL DURATION: 52.8 seconds.
NOMINAL LIQUID
CONSUMPTION: 39 11 SBS
216 ~1 rinse solution
SYSTEM ERRORS: LS3 is tested for liquid after the OSM datacollection. If it is not full
the following error messages appear:
System status:
“CAL ERROR L”
“Leakage in OXI system”
If neither LS5 nor LSl sense liquid within 20 seconds during the SBS
front replacement, the following error messages appear:
Screen status field: “No SBS solution”
System status:
“SBS ERROR A”
“No Salt-Bridge Solution detected”
If LS5 does not and LSl does sense liquid within 20 seconds during
the SBS front replacement, the following error messages appear:
Screen status field: “No SBS solution”
System status:
“SBS ERROR B”
“Malfunction in liquid system while fetching SBS”
If neither LS2 nor LSl sense liquid within 24 seconds during the
electrode chamber rinse, the following error messages appear:
Screen status field: No rinse solution
System status:
RINSE ERROR A
No rinse solution detected
If LS2 does not and LSl does sense liquid within 24 seconds during
the electrode chamber rinse, the following error messages appear:
Screen status field: No rinse solution
System status:
RINSE ERROR B
Malfunction in liquid system while fetching rinse solution
If LS2 senses liqtiid where it should not while the electrode chamber
is emptied, the calibration is aborted and the following error
messages appear:
Screen status field: Calibration aborted
System status:
CAL ERROR B
Unexpected liquid in LSZ
If LSl senses liquid where it should not while the electrode chamber
is emptied, the calibration is aborted and the following error
messages appear:
Screen status field: Calibration aborted
System status:
CAL ERROR A
Unexpected liquid in LSl
NOMINAL LIQUID
CONSUMPTION: 66 ~1 SBS
382 ~1 CAL2 solution
SYSTEM ERRORS: If LS2 does not sense liquid within 24 seconds during the electrode
chamber CAL2 solution flush, the calibration is aborted and the
following error messages appear:
Screen status field:
Calibration aborted
System status:
CAL ERROR G
No low CAL solution detected
If LS5 does not sense liquid within 20 seconds during the SBS front
replacement, the following error messages appear:
Screen status field:
No SBS solution
System status:
CAL SBS ERROR
No Salt-Bridge Solution during low pH calibration
If LSl does not sense liquid within 10 seconds during the electrode
chamber CAL2 filling, the calibration is aborted and the following
error messages appear:
Screen status field:
Calibration aborted
System status:
CAL ERROR F
LSl did not sense low CAL solution
If LS2 does not sense liquid within 4 seconds during the electrode
chamber CAL2 filling or the liquid is not homogeneous, the ABL will
try another pH2START by jumping to the first program step. Max. 3
attemps are performed. If still not successful1 the calibration is
aborted and the following error messages appear:
Screen status field:
Calibration aborted
System status:
CAL ERROR I
Low CAL solution inhomogenous
This program is used to initialize a pH and an OXI calibration using CAL2 (low pH) solution.
NOMINAL LIQUID
CONSUMPTION: 66 ~1 SBS
396 ~1 CAL2 solution
SYSTEM ERRORS: If LS2 does not sense liquid within 24 seconds during the electrode
chamber CAL2 solution flush, the calibration is aborted and the
following error messages appear:
Screen status field: “Calibration aborted”
System status:
“CAL ERROR G”
“No low CAL solution detected”
If LS5 does not sense liquid within 20 seconds during the SBS front
replacement, the following error messages appear:
Screen status field: No SBS solution”
System status:
“CAL SBS ERROR”
“No Salt-Bridge Solution during low pH calibration”
If LSl does not sense liquid within 10 seconds during the electrode
chamber CAL2 solution filling, the calibration is aborted and the
following error messages appear:
Screen status field: Calibration aborted”
System status:
“CAL ERROR F”
“LSl did not sense low CAL solution”
If LS3 does not sense liquid within 1 second during the cuvette
filling, the calibration is aborted and the following error messages
appear:
Screen status field: Calibration aborted
System status:
CAL ERROR H
LS3 did not sense low CAL solution
If LS2 does not sense liquid within 6 seconds during the electrode
chamber CAL2 filling or the liquid is not homogeneous, the ABL will
try another pH2START by jumping to the first program step. Max. 3
attemps are performed. If still not successful1 the calibration is
aborted and the following error messages appear:
Screen status field: Calibration aborted
System status:
CAL ERROR J
Low CAL solution inhomogenous
pH2 END
If the ABL is performing a TOTAL CAL, the pH 2nd. point calibration initiated in pH2START is
completed in this program.
NOMINAL LIQUID
CONSUMPTION: 39 ill SBS.
ABLSOO: 241 @I rinse solution.
ABL510/520: 216 ~1 rinse solution.
SYSTEM ERRORS: If LS2 senses liquid where it should not, while the electrode
chambers are emptied, the calibration is aborted and the following
error messages appear:
Screen status field: Calibration aborted
System status:
CAL ERROR B
Unexpected liquid in LSZ
GAS3 CALIBRATION
The pO2 electrode is calibrated with GAS3 (sensitivity calibration).
NOMINAL LIQUID
CONSUMPTION: 477 ~1 rinse solution.
210 11 CAL1 solution.
78 ~1 CAL2 solution.
SYSTEM ERRORS: If LSl senses liquid where it should not, during the electrode data
collection, the calibration is aborted and the following error
messages appear:
Screen status field: Calibration aborted
System status:
CAL ERROR A
Unexpected liquid in LSI
If LS2 senses liquid where it should not, during the electrode data
collection, the calibration is aborted and the following error
messages appear:
Screen status field: Calibration aborted
System status:
CAL ERROR B
Unexpected liquid in LS2
NOMINAL DURATION: 87.2 seconds (After GASCALl and GASCALZ: + 10 seconds updating)
(After CAL1 and CALZ: + 20 seconds updating)
NOMINAL LIQUID
CONSUMPTION: 592 111 rinse solution.
54 11 SBS.
22 PI CAL1 + 22 &I CALZ.
SYSTEM ERRORS: Errors occurring during rinse programs do not cause aborts in order
not to create infinite loops (an aborted wet section program will
always cause a rinse program to be called).
If the inlet flap is opened during the inlet rinse (first 30 seconds of
the program), another Calibration Rinse will be called in succession
of the first.
If the inlet flap is open when the Calibration Rinse has been finished
(when the program jumps to READY), the ABL goes into hold and
must be restarted.
If LSl, LS2 or LS4 do not sense liquid when expected, the following
error messages appear:
Screen status field: No rinse solution
System status:
RINSE ERROR A
No rinse solution detected
If LS2 does not sense liquid while LSl does, the following error
messages appear:
Screen status field: No rinse solution
System status:
“RINSE ERROR B”
“Malfunction in liquid system while fetching rinse solution
If LSl, LS2 or LS4 sense liquid when not expected, the following error
messages appear:
System status:
“RINSE ERROR C”
“Unexpected liquid during rinse”
If LS5 does not sense liquid while LSl does, the following error
messages appear:
Screen status field: No SBS solution
System status:
SBS ERROR B”
“Malfunction in liquid system while fetching SBS”
If neither L55 nor LSl sense liquid when expected, the following
error messages apper:
Screen status field: No SBS solution
System status:
SBS ERROR A
No Salt Bridge Solution Detected
NOMINAL DURATION: 86.8 sec. (After GASCALl and GASCALZ: + 10 seconds updating)
(After CALZ: + 20 seconds updating)
NOMINAL LIQUID
CONSUMPTION: 592 @I rinse solution.
54 ~1 SBS.
22 PI CAL1 + 22 @I CALZ.
SYSTEM ERRORS: Errors occurring during rinse programs do not cause aborts in order
natto create infinite IOOQS (an aborted wet section program will
always cause a rinse program to be called).
If the inlet flap is open when the Calibration Rinse has been finished
(when the program jumps to READY), the ABL goes into hold and
must be restarted.
If LSl, LS2 or LS4 do not sense liquid when expected, the following
error messages appear:
Screen status field: No rinse solution
System status:
RINSE ERROR A No rinse solution detected
If LS2 does not sense liquid while LSl does, the following error
messages appear:
Screen status field: No rinse solution
System status:
RINSE ERROR B
Malfunction in liquid system while fetching rinse solution
If LSI, LS2 or LS4 sense liquid when not expected, the following error
messages appear:
System status:
RINSE ERROR C Unexpected liquid during rinse
If LS5 does not sense liquid while LSl does, the following error
messages appear:
Screen status field: No SBS solution
System status:
“SBS ERROR B Malfunction in liquid system while fetching SBS
If neither LS5 nor LSl sense liquid when expected, the following
error messages apper:
Screen status field: No SBS solution
System status:
SBS ERROR A No Salt Bridge Solution Detected
LS3 is tested for liquid when the electrode chamber rinse is started. If
it is not full the following error messages appear:
System status:
“CAL ERROR L”
Leakage in OXI system
Introduction
This section includes descriptions of the wet section maintenance programs.
Each program description includes a summary, the duration, the liquid consumption (if any) and a list
of the possible errors which can turn up during the program execution.
NOMINAL LIQUID
CONSUMPTION: 844 ~1 rinse solution.
52 &I CAL1 solution.
104 @I CAL2 solution.
SYSTEM ERRORS: If the inlet flap is closed before LS4 signal, the program is aborted
and the following error messages appear:
Screen status field:
Maintenance aborted
System status:
PROGRAM NOT STARTED
No solution injected in last protein removal
If the inlet flap is not closed within 20 seconds (acoustic alert after 10
seconds) from the LS4 signal, the program is aborted and the
following error messages appear:
Screen status field:
Maintenance aborted
System status:
INLET NOT CLOSED
Inlet flap not closed when requested in last protein removal
If LS2, LSl or LS3 are full when not expected the main valve is rinsed
with rinse solution and the program is aborted. The following error
messages appear:
System status:
MAINT. PROG. ERROR
*Maintenance program did not finish due to error(s)
If LS2 or LSl are not full within 8 seconds from pump start when the
protein removing solution issucked through the electrode
chambers, the main valve is rinsed with rinse solution and the
program is aborted. The following error messages appear:
System status:
MAINT. PROG. ERROR
Maintenance program did not finish due to error(s)
NOMINAL LIQUID
CONSUMPTION: 628 ~1 rinse solution.
52 @I CAL1 solution.
104 ~1 CAL2 solution.
SYSTEM ERRORS: If the inlet flap is closed before LS4 signal, the program is aborted
and the following error messages appear:
Screen status field:
Maintenance aborted a
-
System status:
PROGRAM NOT STARTED
No solution injected in last protein removal
If the inlet flap is not closed within 20 seconds (acoustic alert after 10
seconds) from the LS4 signal, the program is aborted and the
following error messages appear:
Screen status field:
Maintenance aborted
System status:
INLET NOT CLOSED
Inlet flap not closed when requested in last protein removal
If LS2, LSl or LS3 are full when not expected the main valve is rinsed
with rinse solution and the program is aborted. The following error
messages appear:
System status:
MAINT. PROG. ERROR
Maintenance program did not finish due to error(s)
If LS2 or LSl are not full within 8 seconds from pump start when the
protein removing solution is sucked through the electrode chambers
the main valve is rinsed with rinse solution and the program is
aborted. The following error messages appear:
System status:
MAINT. PROG. ERROR
Maintenance program did not finish due to error(s)
If the inlet flap is not closed within 20 seconds (acoustic alert after 10
seconds) from the LS4 signal, the program is aborted and the
following error messages appear:
Screen status field:
Maintenance aborted
System status:
INLET NOT CLOSED
Inlet flap not closed when requested in last protein removal
If LS2, LSl or LS3 are full when not expected the main valve is rinsed
with rinse solution and the program is aborted. The following error
messages appear:
System status:
MAINT. PROG. ERROR
Maintenance program did not finish due to error(s)
If LS2 or LSl are not full within 8 seconds from pump start when the
protein removing solution is sucked through the electrode chambers
the main valve is rinsed with rinse solution and the program is
aborted. The following error messages appear:
System status:
MAINT. PROG. ERROR
Maintenance program did not finish due to error(s)
If the inlet flap is not closed within 20 seconds (acoustic alert after 10
seconds) from the LS4 signal, the program is aborted and the
following error messages appear:
Screen status field:
Maintenance aborted
System status:
INLET NOT CLOSED
Inlet flap not closed when requested in last decontamination
NOMINAL LIQUID
CONSUMPTION: 505 (~1 cleaning solution.
64 ~1 SBS.
91 ~1 CAL1 solution.
65 ~1 CAL2 solution.
2229 @I rinse solution.
SYSTEM ERRORS: If both LS2 and LSl are empty during the electrode chamber filling,
the following error messages appear:
Screen status field:
No cleaning solution
System status:
CLEANING ERROR A
No cleaning solution detected
If LS2 is empty and LSl is full during the electrode chamber filling,
the following error messages appear:
Screen status field:
No cleaning solution
System status:
CLEANING ERROR B
Malfunction in liquid system while fetching cleaning solution
REFILL
This program refills the tubings with rinse solution, CAL1 solution and CAL2 solution when liquid
containers are replaced and in the start-up procedure after power on.
0 NOMINAL LIQUID
CONSUMPTION: 900 @I rinse solution.
648 ~1 CAL1 solution.
648 @I CAL2 solution.
SYSTEM ERRORS: If CAL2 solution is not homogeneous during refill, the following
error message appears:
Screen status field:
No calibration solution (4)
SBS REFILL
Possible air bubbles are removed from the SBS system.
SUMMARY OF EVENTS: Measuring light on. Gasmixer pump on. GAS1 selection.
Pl sucks SBS. Out to waste.
LS5 is emptied. Pl sucks SBS again. LS5 test (must be full).
Dynamic SBS cut off with GASl.
NOMINAL LIQUID
CONSUMPTION: 182 LLI SBS
SYSTEM ERRORS: If the salt bridge solution is not homogeneous during the SBS refill,
the following error messages appear:
Screen status field:
“No SBS solution”
System status:
“SBS ERROR A”
“No salt bridge solution detected”
The membrane valves V4. VS and V6 are activated 5 times in order to prevent sticking. Main valve
position is adjusted.
NOMINAL LIQUID
CONSUMPTION: None.
LS ADJUST
SUMMARY OF EVENTS: Gasmixer start and GAS1 selection. Measure light on.
LSl, LS2, LS3 (ABLSlO only) and LS4 are emptied and calibrated.
LS5 is emptied, calibrated and refilled.
NOMINAL LIQUID
CONSUMPTION: 131 PI SBS.
ENTRY FROM: Maintenance soft key. The program is automatically called during
the autocleaning procedure.
The membrane valves V4, V5 and V6 are activated 5 times in order to prevent sticking. Main valve
position is adjusted.
NOMINAL LIQUID
CONSUMPTION: None.
NOMINAL LIQUID
CONSUMPTION: None.
GM FLUSH
The gas mixer is started and GAS3 selected for flushing after replacement of the gas cylinder.
NOMINAL LIQUID
CONSUMPTION: None.
This rinse program is used after maintenance programs. Compared to MEASUREMENT RINSE, a main
valve rinse is performed during the electrode chamber rinse.
SUMMARY OF EVENTS: Busy lamp on, measure light on, ref. data collection.
Gas mixer pump start and GAS1 selection.
INLET RINSE:
2x4 rinse/air segments created by P2, V4, V5 through inlet and out
to waste through W7.
P2 is dosing rinse sol. into W7 tubing. LS4 is tested for liquid.
P2 sucks rinse sol. back from W7 tubing. MV alternates between
position 0 and 1, and 4 rinse sol./GASl segments are hence
created. P2 changes direction and the tubings are emptied.
P2 sucks rinse solution again. P2 changes direction and the rinse
sol. is pumped back.
Electrode datacollection stop.
ELECTRODE CHAMBER RINSE:
Pl expels the sample/sucks rinse solution parked in the WO
tubing. P2, V4, VS make 3 rinse sol./air segments.
Pl/P2 sucks/pushes sample and rinse/air segments through
electrode chambers.
MAIN VALVE RINSE:
Pl sucks fresh rinse sol. into the Main Valve which rotates twice.
Pl refills tubings with liquids (CAL1 , CALZ, CLEANING and RINSE).
ELECTRODE CHAMBER RINSE (continued):
Pl sucks rinse sol./GASl via MV. LS2 is tested for liquid.
Pl and MV suck 3 rinse sol./GASl segments through preheater 2
and electrode chambers. P2 sucks rinse sol. from V4.
P2 changes direction and doses the rinse sol. into the electrode
chambes. Pl emties the electrode chambers.
SBS front is changed. LSS checks the SBS homogeneity.
Pl and MV create 8 rinse/ GAS1 s e g m e n t s w h i c h a r e
sucked/pushed forward and backward through the electrode
chambers. Finally Pl empties the electrode chambers. LS2 and LSl
are tested for liquid.
SYSTEM ERRORS: Errors occurring during rinse programs do not cause aborts in order
not to create infinite loops (an aborted wet section program will
always cause a rinse program to be called).
If the inlet flap is opened during the inlet rinse (first 30 seconds of
the program), another Maintenance Rinse will be called in succession
of the first.
If the inlet flap is open when the Maintenance Rinse has been
finished (when the program jumps to READY), the ABL goes into
hold and must be restarted.
If LSl, LS2 or LS4 do not sense liquid when expected, the following
error messages appear:
Screen status field: No rinse solution
System status:
“RINSE ERROR A”
“No rinse solution detected
If LS2 does not sense liquid while LSl does, the following error
messages appear:
Screen status field: No rinse solution
System status:
“RINSE ERROR B”
“Malfunction in liquid system while fetching rinse solution
If LSl, LS2 or LS4 sense liquid when not expected, the following error
messages appear:
System status:
RINSE ERROR C
Unexpected liquid during rinse
If LSS does not sense liquid while L51 does, the following error
messages appear:
Screen status field: No SBS solution”
System status:
“SBS ERROR 8”
“Malfunction in liquid system while fetching 585”
If neither LS5 nor LSl sense liquid when expected, the following
error messages apper:
Screen status field: No SBS solution
System status:
SBS ERROR A
No Salt Bridge Solution Detected
PROM Test
The PROMS are checked by using a checksum test. A predetermined checksum is stored in each PROM.
This checksum is compared with a checksum which is calculated during the checksum test.
0 RAM Test
The RAMS are checked via a write/read test in which the following is carried out for all cells in succession:
Two patterns are written into and read from the cell in question, first 01010101 (55,-J and then 10101010
(AA,,), to check for stuck at faults.
Watchdog Test
The Watchdog test circuit is a hardware test circuit which will reset the satellite CPU when it fails in the
execution of the corresponding program in the software.
0
REVISED 92.11 .Ol 6.3 - 1
SELF-TEST PROGRAMS
, nnnnnRnnn
1
I_ -----__________--____-----__----_-- -I
Figure 1.
ABL500 Run 50 onwards, ABL5101520 Run 8 onwards and ABL505 Run 1 onwards 0
-
The bars representing the required PROMS and the RAM have a normal intensity. PROMS not required by
the system have a lower intensity. During the test of a PROM or the RAM the corresponding bar is
highlighted. If the PROM is OK the intensity is returned to normal, but if the PROM is rejected the intensity
is reduced to a little higher than normal and flashes with 3 Hz.
The test continues until finished even if there is a defect PROM. If there was an error during the test the
picture will remain on the screen indicating the error, otherwise it will continue the start-up program by
testing the remaining PROMS and initiating the ABL.
During the normal operation the CPU performs a PROM checksum test and a write/read check of the RAM.
If an error occurs an error message is shown on the screen.
r-_--------------------------,
Where hhhh is the segment address and 1111 is the offset address of the faulty RAM cell.
The last line in the error message is for Radiometer Medical A/S R&D department. This line contains
information about software version and where in the software the error is located.
EPROM Error
When the power is switched on the CPU will perform a checksum check of the EPROM. During this test
the display will show the picture seen in figure 4. If an error isdetected the CPU will stop all activity and show
the picture seen in figure 5.
During normal operation the CPU is continuously checking the EPROM. If a checksum error is detected the
CPU will stop all activity and show the picture seen in figure 5.
RADIOMETER
Figure 4.
KEYPAD -UNIT
VERSION: V.VV
EPROM-TEST: ERR. SS RR
Figure 5.
Where SS is the sum written in the EPROM and RR is the calculated sum.
The only way to restart the Printer Unit when one of the above errors has been detected is to turn
off the ABL and then turn it on again.
Watchdog Error
The Watchdog error circuit (WD) is a hardware circuit monitoring the function of the CPU. If the CPU does
not send a Watchdog trigger to the WD circuit within every 71 ms, the WD will reset the CPU. This will
happen twice. If the CPU has not cleared the WD circuit when the third reset occurs, the WD circuit will shut
down the CPU and remove the RSl50/100 from the ring network. The LED will be turned on with a steady
light.
Each time the CPU is reset by the watchdog circuit the CPU sends a message to the Main Computer PCB.
The Main Computer PCB will then retransmit the setup of the RSl50/100.
RAM Error
When power is switched on the CPU will carry out a complete RAM test. If a fault is detected the LED will
start flashing with a frequency of 1 Hz and the PCB will remain inactive.
During normal operation the CPU is continuously performing tests of the RAM, one cell per 1 ms. If a fault
is detected the CPU will remove the PCB from the ring network and the RED LED will start flashing with
1 Hz.
0 6.3 - 7
REVISED 92.11 .Ol
SELF-TEST PROGRAMS
EPROM Error
When the power is switched on the CPU will perform a checksum check of the EPROM. If an error is
detected the RED LED will start flashing with a frequency of 2 Hz and the PCB will remain inactive.
During normal operation the CPU will continuously check the EPROMs by adding one cell every 1 ms. If
a checksum error is detected the CPU will remove the PC6 from the ring network and the RED LED starts
flashing with 2 Hz.
The only way to restart the RS150/100 when one of the above errors has been detected is to turn off
the ABL and then turn it on again.
Watchdog Error.
, The Watchdog error circuit (WD) is a hardware circuit monitoring the function of the CPU.
If the CPU does not send a Watchdog triggerto the WD circuit within every 71 ms the WD will reset the CPU.
This will happen twice and if the CPU has not cleared the WD circuit when the third reset occurs the WD
circuit will shut down the CPU and remove the Wet Section Control PCB from the ring network.
The LED will be turned on with a steady light.
Every time the Wet Section CPU is reset it will send a message to the Main Computer PCB. The Main
Computer PCB will display the following message on the screen: Wet section control fault, Error Code
0. Then the Main Computer PCB will perform a Power On Reset of the ABL.
RAM Error
When power is switched on the CPU will carry out a complete RAM test. If a fault is detected the Busy
and Ready light will start flashing with a frequency of 1 Hz for approximately 5 seconds. Then the CPU
sends a message to the Main CPU. The Main CPU will display on the screen Wet section control fault Error
code S, and then it will perform a Power On Reset of the ABL.
During normal operation the CPU is continuously performing tests of the RAM, one cell per 1 ms. If a fault
is detected the Busy and Ready light will start flashing with a frequency of 1 Hz for approximately 5
seconds. Then the CPU sends a message to the Main Computer PCB. The Main Computer PCB will display
the following on the screen: Wet section control fault Error code S, and then perform a Power On Reset
of the ABL.
6.3 - 8
SELF-TEST PROGRAMS
EPROM Error
When the power is switched on the CPU will perform a checksum checkof the EPROM. If a checksum error
is detected the Busy and Ready light will start flashing with a frequency of 2 Hz for approximately 5
seconds. Then the CPU sends a message to the Main Computer PCB. The Main CPU will display the
following on the screen: Wet section control fault Error code 2 and then perform a Power On Reset of
the ABL.
During normal operation the CPU will continuously check the EPROM by adding one cell every 1 ms. If a
checksum error is detected the Busy and Ready light will start flashing with a frequency of 2 Hz for
approximately 5 seconds. Then the CPU sends a message to the Main Computer PCB. The Main Computer
PCB will display this on the screen: Wet section control fault Error code 2 and then perform a Power On
Reset of the ABL.
a 6.3 - 9
a 6.4 SERVICE PROGRAMS
Introduction
The instrument provides a number of service programs to improve the servicability of the instrument.
This section provides a description of the individual service programs. For the detailed checkout, ad-
justment and troubleshooting procedures, please refer to the appropriate sections.
1) ABL5101520 only.
2) ABL500/510/520 from Run 50/8/8 onwards only. In ABL500 this program is listed as Other
service program No. 11 (since Cuvette compensation does not exist).
Entry
The service programs are available whenever the Menu softkey is visible. Press the sequence of
softkeys as shown below.
Password: After having pressed “Service”, it is necessary to key in a password and press ENTER to
confirm in order to obtain access to the service menu. From the factory, the service password is set to
“89:452”, however, this may be changed or deleted using the other service program no. 8 (Password).
Wet Sect. Programs: In case an activity is taking place in the Wet Section when entering the ser-
vice menu, “Wet section service programs” is displayed with low intensity and the corresponding
softkey is extinguished to indicate that these programs are presently not available.
Program selection: When one of the above two service program menus is displayed, a specific
program is selected either by:
keying in the program number followed by ENTER, or
moving the curser to the program using the finger wheel and then pressing ENTER.
moving the curser to the program using the finger wheel and then pressing “Enter program”.
Auto Exit: Please note that if the instrument is left idle with one of the submenus displayed for
three minutes, it will automatically exit from the submenu and return to the highest level (where the
“Menu” softkey is visible).
Screen
1. ABL500/5l O/520:
2. ABL505:
a
Soft keys
Fill menu: Provides a menu for filling the electrode chambers (and cuvette, ABL510/520) with the
following liquids:
Cal 1: Cal 1 solution after which a liquid junction is established.
Cal 2: Cal 2 solution after which a liquid junction is established.
Sample: Enables introduction of a liquid sample followed by establishment of liquid
junction.
Exit: Returns to the main menu for the electerode checkout program.
When one of the softkeys in the fill menu is activated, the main menu for this
program is shown. TheFill menu key is, however, not available, but an Empty
is available instead.
Gas menu: Provides a menu for flushing gasses through the measuring chamber as follows:
Gas 1: Flushes Gas 1 through the electrode chambers for 60 seconds.
Gas 2:
Gas 3:
Flushes Gas 2 through the electrode chambers for 60 seconds.
Flushes Gas 3 through the electrode chambers for 60 seconds. l
-
stop: Terminates the gas flush.
Exit: Returns to the main menu for the electrode checkout program.
When one of the softkeys in the Gas menu is activated, the main menu for this
program is shown.
Only elect. channels: Enables quick updating (every 0.57 second) of electrode signals only (the
remaining parameters are frozen).
Replaced by All channels which enables updating of all parameters again
(approx. every 6 seconds).
Print: Provides printout of the data displayed.
Exit: Returns to the Wet Section Service Program Menu.
A rinse flag will be set when entering this program. The rinse will, however, not be called
until exit from the wet section service program menu is carried out.
Please note that if one of the softkeys in the Fill menu" or Gas menu has been activated,
the sample stays in the electrode chambers as long as you stay in the wet section service
programs (unless, of course, Empty is activated). This means that it is possible, for
l
example, to inject a cold sample in the temperature checkout program to see how the
temperatures vary and then afterwards go into the electrode checkout program to check
the elecrode signals.
Screen :
Note 1: When stabilized. The monitor is linear for the range [33 C ; 41 Cl.
Note 2: Max. difference between Block and Sense is 0.05 C when stabilized.
Note 3: When the test resistor is connected (6.1 kQ, 0.05 %, Code 152-332)
Softke ys
Fill elect. chamber: Fills electrode chambers (and cuvette, ABL510/520) with CAL1 solution and
establishes the liquid junction. When activated, it is replaced by Empty.
Sample: Enables introduction of a liquid sample followed by establishment of liquid junction.
Print: Provides printout of the data on the screen.
Exit: Returns to the Wet Section Service Program Menu.
A rinse flag will be set when entering this program. The Rinse will, however, not be
called until exit from the wet section service program menu is carried out.
Please note, that if Fill elect. chambers has been activated, the sample stays in the
electrode chambers as long as you stay in the wet section service programs (unless, of
course, Empty is activated). This means that it is possible, for example, to inject a cold
sample in the temperature checkout program to see how the temperatures vary and
then afterwards to go into the electrode checkout program to check the elecrode sig-
nals.
Purpose: Checkout and alignment of the Gas Mixer mechanics and electronics.
Screen:
Note 1: Dynamic range given. Limits for GM on are: 1550 + voffset ; 2050 + Voffset.
Note 2: Limits given for GM on. When off, the value is 0.
Note 3: With GM off, the value is 0 rpm. With GM on the upper limit before replacing the
membrane pump is 1200 rpm.
Note 4: Adjusted to 50 mV at factory. The maximum drift is only limited by the dynamic range of
the monitor.
Note 5: Dynamic range given. Limits for GM on are: 1440 + Voffset ; 2 160 + Joffset,
Softkeys
a Decrease air flow: Decreases the air flow by decreasing the pump speed in steps of 10 rpm per ac-
tivation. Used when adjusting the air flow monitor.
Backspace may be used as an alternative for this softkey.
Increase air flow: Increases the air flow by increasing the pump speed in steps of 10 rpm per acti-
vation. Used when adjusting the air flow monitor.
Enter may be used as an alternative for this softkey.
Gas mixer on: Turns the gas mixer on (the gas mixer supplies Gas 1). The ENTER key may be
used as an alternative to the Gas mixer on softkey.
Gas mixer off: Turns the gas mixer off.
Print: Provides printout of the data on the screen.
Exit: Returns to the Wet Section Service Program Menu. The gas mixer is turned off
upon exit.
A Rinse flag will be set when entering this program. The Rinse will, however,
not be called until exit from the wet section service program menu is carried
out.
Purpose: Offset Calibration of the ~02 channel. The program is to be used in case the pO2 Pre-
amplifier Unit or the Wet Section Control PC3 has been replaced.
Screen:
Softkeys
Calibr. : Provides a new calibration value.
Please note that the pO2 electrode must be disconnected before pressing Calibrate.
Save: Used for saving the new calibration value. It is possible to save values in the range
[SO00 ; 7000 PA].
However, in a service situation only values in the range [6000 ; 6350 pAI should be
saved.
kin t: Provides a printout of the data on the screen.
Exit: Returns to the Wet Section Service Program Menu.
A Rinse flag will be set when entering this program. The Rinse will, however, not be
called until exit from the wet section service program menu is carried out.
Purpose: Calibration of the waste sensor 902-053 (mounted from the factory in ABL500/510/
520 below Run 50/8/8). To be used when the waste detector has been replaced.
Check of the Waste detector 902-155 (mounted from the factory in ABL500/510/520
from Run 50/8/8 onwards).
Please refer to Section 2.4 for calibration/checkout procedures.
Screen: The two upper lines aaply only for the earlier type waste detector and are always
shown. The three lower lines apply only for the new type waste detector and are only
shown in case the new type is actually mounted.
Softkeys
Calibr. : Earlier waste detector Stores the output of the waste detector in the E*PROM.
In a service situation only values in the range [2200 ; 3200 mV] should be saved
(only the three most significant digits are stored, e.g. 220).
Please note that an empty waste container must have been mounted for at least
10 minutes before pressing Calibrate.
New waste detector Initiates a measuring cycle to update the displayed values.
Purpose: Checkout of the valves in the Wet Section and the Gas Mixer.
Screen:
off
off
of f
off
off
off
off
of f
off
off
Softkeys
Pull and release: Pulls and releases the selected valve. Used for checking the pull circuit.
Pull and hold: Pulls and holds the selected valve. Used for checking the pull and hold function.
It is possible to setup a combination of held valves by pulling and holding the
valves one after the other.
Release all: Releases all valves which are held activated.
Exit: Returns to the Wet Section Service Program Menu. All valves are released upon
exit.
A Rinse flag will be set when entering this program. The Rinse will, however, not
be called until exit from the wet section service program menu is carried out.
Enter key: May be used as a toggle key for a selected valve. May alternately be used as pull
and hold and release for the selected valve.
NOTE Please note that the valve, V3, is not included in the ABLSOO.
Purpose: Checkout of the pumps and the main valve in the Wet Section.
Screen :
Softkeys
Decrease: Is used for decreasing the pump speed or the position of the main valve.
The Backspace key may be used as an alternative to this softkey.
Increase: Is used for increasing the pumpspeed or the position of the main valve.
The Enter key may be used as an alternative to this softkey.
The function of increase and decrease are visualized overleaf. 4
Both pumps may be set up to run simultaneously one after the other. When any of the pumps are
turning, the main valve is at position 0. If any of the pumps are turning when the main valve is moved
away from position 0, the pumps are immediately stopped.
Start/stop: When the pumps have been setup, they may be stopped simultaneously and
restarted using this softkey.
When the main valve has been set to a certain position, say pos. 4, this softkey
may alternately rotate the main valve to position 0 and 4 respectively.
Exit: Returns to the Wet Section Service Program Menu. The pumps are turned off
and the main valve is returned to position 0 upon exit.
A Rinse flag will be set when entering this program. The Rinse will, however, not
be called until exit from the wet section service program menu is carried out.
Softkeys
Fill liquid sensors: Fills all the liquid sensors. When activated, it is replaced by Empty. All liquid
sensors are empty at entry to the program.
(Print: Provides printout of the data on the screen).
Exit: Returns to the Wet Section Service Program Menu.
A rinse flag will be set when entering this program. The rinse will, however, not
be called until exit from the wet section service program menu is carried out.
Screen :
Inlet nodule on
tnlet flap closed
&pirate button releaseci
covet on
Uaste container mPtY
Fa? regulate
neas light off
Ready/Busy LED busy
Softkeys
Ready Busy LED: A toggle switch for alternately switching on the Ready and Busy LEDs.
Meas light: Used for turning on/off the measuring chamber light.
Fan: Used for controlling the fan. Three settings are available; Regulate (default)
where the ambient temperature controls the fan, Off, On.
Exit: Returns to the Wet Section Service Program Menu.
A Rinse flag will be set when entering this program. The Rinse will, however, not
be called until exit from the wet section service program menu is carried out.
a
6.4.1-12 Revised 91.04.05
SERVICE PROGRAMS
0
Wet Section Control Step, Wet Section Service Program No. 10
Purpose: For checking the wet section. It is possible to design an individual wet section control
program with up to five steps.
Screen (main):
54 0 Cl0 0 0(1 0 II 0 0 0 0 0 :: :
Status o f L i q u i d Smsnrs 0 0 0 0 1 Current Step 0
Purpose: For manual entry of gas mixer percentages, e.g. in case the EEPROM on the Gas Mixer
Interface PCB is defective. In this case the percentages may be read on the Gas Mixer
itself and then be manually entered in this program. In order to avoid accidental
change of the gas mixer percentages, a check value must be calculated and entered
together with the percentages before the ABL will accept the change.
Screen :
Softkeys
Save and restart: Saves the keyed in values after which the ABL is restarted. That is the wet section
control programs are reloaded and the ABL goes into STARTUP.
Exit: Returns to the Wet Section Service Program Menu.
A Rinse flag will be set when entering this program. The Rinse will, however, not
be called until exit from the wet section service program menu is carried out.
Purpose: Used for cleaning and drying the Wet Section in a controlled way when shutting
down the ABL for a longer period. The program is used in the production department
after having completed the final test and before transferring the instrument to the
warehouse
Screen: The screen provides the shut down procedure step by step (only the initial screen is
shown).
2:
Softkeys
Start: Initiates the next part of the cleaning/drying procedure. Please note that once
started, the program cannot be interrupted.
Purpose: Checkout of the six photo diode channels of the optical system. Each channel may be
checked as a whole (optics and electronics) or electronics alone.
Screen : The screen is devided into four columns providing channel number, present counter
values, present current values (individual and mean), and individual current values
during the latest zero point calibration (the fourth column also provides drift and
status values).
Softkeys
Fill elect. chamber: Fills electrode chambers and cuvette with CAL1 solution and establishes the
liquid junction. When activated, it is replaced by Empty.
Sample: Enables introduction of a liquid sample; liquid junction is establish.
Lamp on/off: Used for turning on/off the bulb in the lamp unit.
Hemol. on/off: Used for turning on/off the hemolyzation circuit
Exit: Returns to the Wet Section Service Program Menu.
Print: Provides printout of the data on the screen.
Purpose: Checkout and adjustment of the Monochromator Unit. To ensure that the wet section
is empty at entry, it is important that the program is called directly from the ready
mode where ti Menu s is visible.
Screen : The screen may display up to five measurements showing ~02, tHb and cturb. Sample
#l is the latest measurement (earlier results are pushed downwards).
AEL520S02100% ADJUSTMENT
Softkeys
Sample: Enables introduction of the sample with an ~02 of 100 %. The key is extinguished
after activation.
After introduction, the sample is transported into the cuvette, hemolyzed and
measured. When the results are displayed, the acoustic alarm sounds.
In case any of the three parameters is outside the specified range, the parameter will be questioned.
Screen:
Softkeys
Select channel: Used for selecting the channel to be tested.
Start test: Used for starting the test. When activated, the test will be repeated infinitely
until Stop test is activated. This makes it possible to work with cables, plugs,
etc. to reveal errors caused by bad connections, etc.
Stop test: Used for ending a test.
Exit: Returns to Other Service Program Menu. An ongoing test is stopped upon exit.
Purpose: Used for checking the Printer and Printer Control PCB.
Screen:
Purpose: Used for setting up the ABL temporarily in a service situation without changing the
users own setup.
Note: Please note, that if a temporary setup is used, the contents of the status field (e.g.
Ready) will be displayed in black letters. This is to inform you that the instrument is
not operating according to the users setup.
Screen :
3. I
Softkeys
Change setup: Used for selecting which part of the RAM memory the subsequent changes in
the setup is to be stored in. The possibilitiesare:
Permanent which is battery backed-up (default setting). If Permanent is
selected, the ABL will be in the normal mode where all changes to the setup
will overwrite the original setup.
Screen
ABL500/51 O/520:
AB L.505:
Softkeys
Yes/no: Used for setting up.
If Disregard gas & temp errors, Cover off removal ignored, Disable gas checkand/
or Disable waste check are set to Yes, it is possible to run the ABL in the normal
mode as if no errors were present.
If Rinse only at power on is selected, the ABL will only perform a Rinse when
switching on the power (that is, no refill, no LS adjustment, no warming period, etc.).
It is possible to work with the ABL as soon as the upstart rinse has completed.
Exit: Returns to the Other Service Program Menu. The service setup remains unchanged
after having activated Exit.
Screen:
Softkeys
Print: Provides a printout of the data on the screen.
Exit: Returns to the Other Service Program Menu. It is possible to exit, even though a
printout of the data on the screen has not yet been completed.
Purpose: Used for setting up the control of the wet section during service. It is possible to change
or cancel the ABLs normal action caused by the status of the various parts of the wet
section.
Note: If a service setup is used, the contents of the status field (e.g. Ready) will be displayed
in black letters. This is to inform you that the instrument is not operating according to the
Users Handbook.
Above note does not apply to *‘Additional decimal digit for bytes’?’
Purpose: ABLSOO: Provides the possibility of obtaining a printout of the electrode updat-
ings. It is possible to select the printout for measurements, calibrations, and/or gas
flushes. For measurements and gas flushes both electrode signals (mV or PA) and the
calculated values (e.g. mmHg) are printed out. However, for calibrations it is possible
to choose the updatings to be printed out as electrode signals and/or as calculated
values.
ABLSlOl520: Provides in addition to the above the possibility of obtaining a prin-
tout of the photo diode currents (dark and light) during an OXI measure-
ment/calibration.
Examples of printouts of the electrode updatings are shown at the end of this section.
Note: Please note, that if printout of updatings has been selected, the contents of the status
field (e.g. Ready) will be displayed in black letters. This is to inform you that the ABL
is not operating according to the Users Handbook.
Screen :
ACTIVITY OXI
Measurements No
Calibratcons No
Softkeys
Yeslno: Used for setting up the updating printout, that is, for which activities the updat-
ings printout should be generated and in which unit the parameters should be
printed out.
Automatic: The updatings may be printed out automatically just after the
normal printout for the activity in question.
Purpose: Used for aligning the picture on the CRT monitor. The program provides three
different test pictures, each serving their own function for performing an align-
ment of the picture.
Screen :
Softkeys
Test picture 1: White screen; used for adjusting the height and width of the picture.
Test picture 2: Horizontal lines; used for adjusting the vertical linearity.
Test picture 3: Vertical lines; used for adjusting the horizontal linearity.
When any of the above softkeys has been activated, the test picture, which uses the entire screen, is
displayed. This leaves no space for displaying any softkeys, however, all softkeys have the same func-
tion, Exit, which causes a return to the screen shown above.
Exit: Returns to Other Service Program Menu.
Softkeys
Clear user password: Deletes the user password protecting access to the setup programs.
Exit: Returns to the Other Service Program Menu. The setup remains unchanged
at exit..
Purpose: Used for entering the factory code (run number of the ABL) if desired, or reentering it
in case the Main Computer PCB has been replaced.
_
Screen:
Softkeys
Exit: Returns to the Other Service Program Menu. The setup remains unchanged at
exit.
Format: The factory code has the following format Rxx Nxx (e.g. R04 N14). When using
the numerical keyboard, the ID is keyed in as follows:
“.04..14”.
When pressing Enter to confirm, the decimal points are replaced by R and m
N * respectively.
Purpose: Used for keying in a password for protecting the service programs from unauthorized
entry.
Screen:
Purpose: Used for compensating for variations in the lenght of the light path through different
cuvettes. The program is used after having replaced the Hemolyzer Unit or the Wet
Section Control PCB. Refer to Section 2.15 for the actual procedure.
IMPORTANT: The program MUST be called directly from READY and the latest activ-
ity which MUST be carried out prior to entry is either a tHb calibration
or a measurent on S2190.
Screen:
*I
ABL520 CUVETTE COMPENSATION
I I I-
I Save
Softkeys
-
Save: Used for saving the tHb factor calculated on basis of the last measurement or tHb cali-
bration on 52190; the calculated tHb factor will then replace the current tHb factor.
When pressed, Current tHb cuvette factor is set equal to m Calculated tHb cuvette
factor and Current tHb calibration factor is set to 1 .OOO. In fact, a tHb calibration is
carried out.
The ABL510 will not save the calculated tHb cuvette factor in case:
- calculated tHb cuvette factor is outside the range displayed. The factor is then
questioned and the acoustic alarm will sound if Save is pressed.
- tHb from last S2190 meas/cal is outside the range displayed. Both tHb from
last 52190 meas/cal and Calculated tHb cuvette factor are then questioned and
the acoustic alarm will sound if Save is pressed.
Exit: Returns to the Other Service Program Menu. The setup remains unchanged at exit.
Print: Provides a printout of the data on the screen
Purpose: Used for resetting the sample and calibration counters to zero and for deleting all
data stored in the data management system (except the contents of the optional
floppy disk). That is, the patient report, calibration, quality control and system status
logs are deleted.
In addition, the contents of the setup and service programs is reset to factory stan-
dard (default).
Screen:
Softkeys
Clear all: Erases the parity bit of the data stored in the RAM memory.
a
Note A few seconds after having pressed Clear all, the ABL will show a fatal
error (due to the erased parity bits). When this appears, switch the ABL off
and then on again. During the power upstart the contents of the RAM
memory is reset to default.
Exit: Returns to the Other Service Program Menu. The setup remains unchanged at
exit.
Screen : Displayed after activation of Program on the initial screen. This basic screen will
during the software upgrade include various messages indicating the status of the
upgrade together with prompts, cautions and error messages. Please refer to Section
2.17 for the actual software upgrade procedure.
Purpose: Used for loading new software into the flash PROM S on the Main Computer PCB.
NOTE: The program is only available in ABL500/510/520 of Run 50/8/8 onwards (that
is, instruments including flash PROM S on the Main Computer PCB).
NOTE: The complete procedure for loading new software into the flash PROMS is
found in Section 2.17.
LADY ~ ~ 15.30
16.00
15'00
Softkeys
Program : Used for initiating the software upgrade. The software upgrade may take up to 20
minutes. On activation the screen changes to the one shown overleaf.
Exit: Returns to the Other Service Program Menu.
8 7 5 6 nnHg
a
53 -78.98 7.400 -30.74 79.0 503. 25.9
54 -30.74 79.0 574. 25.4
55 -30.74 79.0 573. 25.4
56 -30.73 79.0 569. 25.4
57 -30.73 79.0 569. 25.4
58 -30.71 79.1 565. 25.2
59 -30.71 79.1 564. 25.3
60 -30.71 79.1 565. 25.4
61 -30.71 79.1 562. 25.3
62 -30.70 79.1 566. 25.6
63 -30.71 79.1 558. 25.3
64 -30.71 79.1 559. 25.3
65 -30.71 79.1 562. 25.6
66 555. 25.3
67 553. 25.2
66 550. 25.0
69 550. 25.1
70 549. 25.1
71 550. 25.2
72 554. 25.5
73 553. 25.5
signals.
Na 1 4 5 nnol/L Or i ft 0 St at us 135 nnol/L
50 nnol/L D r i f t 0 Sens . 96.9 %
pH updating no. 1, 8, 15, 22 and
-0 .Ol St at us 1 .26 nnol/L
29 refer to K+. No. 2,9, 16, 23 and Ca l 1 . 2 5 nnol/L Drift
5 . 0 0 nnol/L Orlft 0 .oo Sens . 101.4 %
30 refer to Na+, and no. 3,10,17,24
and 31 are for Ca++/CI-. B 7 6 2 nntig
R A D I O M E T E R AELSIO
B 7 6 4 nnHg
P r i n t e d N o v e n b e r 1 3 , 1 9 9 0 lo:57
OXYGEN STRTUS
sD2. 97.9 %
tDzs . . . . . Uol %
# PH pCD2 P=
CnUl C 1 InUl CnnHg 1 CpAl CnnHg 1
# K’ Na’ Ca"
CnUl CnnoIA_l [Ml Cnnol/Ll CnUl CnnoIA_l
Program Stack: The programs are organised as a stack consisting of 1 through 5 displays. In some of
the programs are included help text pages showing the function of the various keys in
the service program in question.
ENTER key: Used for stepping between the various underlying choices in a service program.
Exit Softkey: Softkey for stepping back up in the menu hieraki; in a service program Exit will return
to the part of the service main menu from which the program was called; in the service
main menu Exit will return to the Utilities menu.
Printout: All the relevant displays may be printed out; for this purpose the Paper feed key is used
(functions as screen dump).
None of the help pages can be printed; this is also the case for some of the displays
containing figures. If the paper feed key is activated in a non printable display the
acoustic signal will sound.
The software versions for the various modules is to be found in the user Setup menu. Do as follows:
1, Press MENU,
2. Select 3 Utilities,
3. Select 2 Setup,
4. Press Page,
5. Select 10 Analyzer ID,
6. Press Page.
The software versions may be printed out as part of the user setup.
Entry
Please note that the entry into the service programs is hidden (not shown in the menu).
Do as follows:
1. Press MENU,
2. Select 3 Utilities,
3. Key in 89.452 (pass word)
The leftmost service main menu display (see below) is shown upon entry.
In case a wet section activity is in progress at the time of entry, the service programs enabling control of
the wet section are unavailable; this is indicated by displaying a bar rather than the program number in the
service main menu.
In case an incorrect pass word is entered it is ignored; if this is the case exit from the Utilities menu and
reenter.
Auto Exit: Please note that if the ABL50 is left idle in the service main menu for two minutes,, it will
automatically exit and return to the highest level (where the Menu softkey is visible).
Keys:
Liquid fill: Menu for filling the electrode chambers with the liquids listed and establishment of liquid
junction:
1 Call fill: Cal 1 solution.
2 Cal2 fill: Cal 2 solution.
3 Sample: Enables introduction of a liquid sample.
Procedure:
- Press 3 Sample,
- Open the inlet flap (BUSY will be lit for a few seconds during renewal of
the SBS solution),
- Wait for READY to be lit, and then introduce the sample,
- Close the inlet flap.
4 Empty: Empties the electrode chambers.
Gas fill: Menu for flushing gasses through the measuring chamber for 60 seconds as follows:
5 Gas7 fill: Flushes Gas 1.
6 Gas2 fill: Flushes Gas 2.
7 Gas3 fill: Flushes Gas 3.
8 stop: Terminates the gas flush.
Printout: of displayed data; press paper feed (help displays cannot be printed out).
Electrode Displays:
Help Displays:
Page 1 Exit
Displays:
Note 1: When stabilized. The monitor is linear for the range [33 C ; 41 Cl.
Note 2: Max. difference between Block and Sense is 0.05 C when stabilized.
Note 3: When the test resistor is connected (6.1 kohm, 0.05 %, Code 152-332).
0
WRITTEN 93.02.08 6.5.1-3
SERVICE PROGRAMS (ABL50)
Printout: of displayed data, press paperfeed (help displays cannot be printed out).
Purpose: Checkout and alignment of the Gas Mixer mechanics and electronics.
Displays:
Note: A ?I in the upper right corner of the screen indicates that the gas mixer is not running correctly.
Keys: 1 Airflow down: Decreases the air flow by decreasing the pump speed in steps of 10
9 rpm per activation. Used when adjusting the air flow monitor.
2 Air flow up: Increases the air flow by increasing the pump speed in steps of 10
rpm per activation. Used when adjusting the air flow monitor.
3 Gas mixer on: Turns the gas mixer on (the gas mixer supplies Gas 1).
4 Gas mixer off: Turns the gas mixer off.
Printout: of displayed data; press paper feed (help displays cannot be printed out).
Note 1 Dynamic range given. Range for running GM is: 1550 mV + Offset ; 2050 mV + Offset
Note 2: Limits given for running GM. When off, the value is 0.
Note 3: With GM off, the value is 0 rpm. When on, the upper limit before replacing the membrane
pump is 1200 rpm.
Note 4: Adjusted to 100 mV at the factory. The maximum drift is limited by the dynamic range of
the monitor.
Note 5: Dynamic range given. Range for running GM is: 1440 mV + Offset ; 2160 mV + Offset.
Purpose: Offset Calibration of the p0, channel. The program is to be used in case the p0,
Preamplifier Unit or the Wet Section Control PCB has been replaced.
Displays:
Calibrate 1 Exit I
Save I Exit I Exit
Printout: of displayed data, press paper feed (rightmost display cannot be printed out).
Display:
SERUICE 12:34)
Waste detect check
I Exit
I I
Purpose: Checkout of the valves in the Wet Section and the Gas Mixer.
Display:
Menu keys: 1 Pull/release: Pulls and releases the selected valve. Used for checking the pull circuit.
2 Pull/hold: Pulls and holds the selected valve. Used for checking the pull and hold
function. It is possible to setup a combination of held valves by pulling and -
holding the valves one after the other.
3 Release all: Releases all valves which are held activated.
4 Toggle: Used as a toggle key for a selected valve. May alternately be used aspull/
hold and release for the selected valve.
ENTER: Is used for selecting the valve to be tested (moving the cursor).
Printout: of displayed data, press paper feed (help display cannot be printed out).
NOTE Please note that the valve, V3, is not included in the ABL50.
Purpose: Checkout of the pumps and the main valve in the Wet Section.
Displays:
SERVICE 12:34)
Pump/MU test
1 Increase
Pl 2 Decrease
P2 3 Start&top
I MU pas. 0
Exit
I
L Page I I
Menu keys: 1 Increase: Is used for increasing the pump speed or the position of the main valve.
2 Decrease: Is used for decreasing the pump speed or the position of the main
valve.Both pumps may be set up to run simultaneously one after the other.
When any of the pumps are turning, the main valve is at position 0. If any
of the pumps are turning when the main valve is moved away from position
0, the pumps are immediately stopped.
3 Start/stop: When the pumps have been setup, they may be stopped simultaneously
and restarted using this softkey.
When the main valve has been set to a certain position, say pos. 4, this
softkey may alternately rotate the main valve to position 0 and 4 respec-
tively.
ENTER: Is used for selecting which of the three electromechanical parts is to be manipulated.
Displays: First display: First column contains liquid sensor #; second and third columns contain
the latest calibration values (up to two, first and second) for each liquid sensor (calibration
value=60 % of the empty value during the calibration).
The fourht column indicate the present value of each liquid sensor together with the status
(E=empty / F=full).
Second display: Contains time and date for the up to two liquid sersor calibrations
stored.
1 321 311
12:341
Liquid sensor test
2E
SERUICE
_iquid s e n s o r t e s t
1
_iquid s e n s o r t e s t
1 Fill
a
-
Values: The calibration and present values for the liquid sensors are shown in counts. The dynamic
range for present value is from 0 through 1023 counts.
The absolute maximum range for the calibration value is from 16 to 612 counts.
However, to allow for drift after a service visit, the calibration value should be in the range
from 30 to 550 counts.
Please note that the ABL50 will give General Error #15 if the empty value for a liquid sensor
is below 34 during a liquid sensor calibration (adjustment). In addition it will not update the
calibration value, but simply repeat the latest accepted value.
Since LS3 is not used in ABL50, the numbers in the corresponding line on the display are
not shown.
Printout: of displayed data, press paper feed (help display cannot be printed out).
Display:
SERVICE 12:341
Status: Fan r=regulate / O=off / 1 =on
Modules c h e c k
(Ready/Busy) LED r=ready / b=busy
Fan r Inlet 0
(Meas) Light O=off / l=on
Led Flap 0I
Inlet (module) O=on / l=off
Light L RSP 0I
Cww 0I (Inlet) Flap O=closed / 1 =open
Waste e Asp(irate button) O=released / 1 =pressed
Cover (thermostat. box) O-on I 1 =off
Waste (container) e=empty / f=full / b=bad
Change 1 Exit
t J
Sofikeys Change: Changes status of the selected module (ENTER is used for selecting the
module (Fan, LED or Light).
Exit: Returns to the Service Main Menu.
A Rinse flag will be set when entering this program. The Rinse will be
called upon exit from the service main menu.
Important: If it is desirable to enter into another service program which
exercises the wet section after having worked with this program, you
should exit from the service main menu and then reenter (tocall the rinse
cycle). This ensures a welldefined condition in the wet section before
entering the second service program.
ENTER: Used for selecting the module (Fan, LED or Light) whoose status is to be changed
using Change.
Purpose: For checking the wet section. It is possible to design an individual wet section control
program with up to four steps.
Displays:
Explanation of display
Each line on the display describes a programmed wet section control step; each column of the line
(denoted A, B, C etc.) describes the state of the electromechanical parts of the wet section as follows:
Printout: of displayed data, press paper feed (help displays cannot be printed out).
Purpose: For manual entry of gas mixer percentages, e.g. in case the EEPROM on the Gas Mixer
Interface PCB is defective. In this case the percentages may be read on the Gas Mixer itself
and then be manually entered in this program. In order to avoid accidental change of the
gas mixer percentages, a check value must be calculated and entered together with the
percentages before the ABL50 will accept the change.
Display:
1 SERUICE 12:34(
Gas mixer EEPROM
Present
5.6400 11.2915
New
0.0000 0.0000
Check 0.0000
kcept 1 Exit I
-
WRITTEN 93.02.08 6.5.1-13
SERVICE PROGRAMS (ABL50)
Softkeys Accept: Saves the keyed in values (see note in Exit paragraph).
Exit: Returns to the Service Main Menu.
A Rinse flag will be set when entering this program. The Rinse will be called
upon exit from the service main menu.
NOTE: In case new values have been accepteed above, the ABL50 will
enter the startup upon exit from the service main menu. This is to
ensure that measurements are not performed until a total calibration
based on the new values has been accepted.
ENTER: is used for moving the cursor between the input fields.
Procedure: a) Key in the new CO, percentages for Gas 1 (left column) and Gas 2 (right).
b) Add the above percentages.
c) Add 12.3456 to the result of the above.
d) Key in the result obtained in step c) as Check value.
Purpose: Used for testing the serial communication channels. Please note that it is not possible to
test the printer port (Port3).
Display: Pott2 is always displayed. RSl50 isonlydisplayed in case the board is actually installed.
I stop I Exit I
Softkeys Start: Used for starting the test. When activated, Start is replaced by Stop.
?, The test will be repeated infinitely until Stop, Exit or ENTER is activated.
This makes it possible to work with cables, plugs, etc. to reveal errors caused
by bad connections, etc.
During an ongoing test, the status field will indicate either OK or Failed.
Please refer to Section 2.11 for the actual test procedure.
stop: Used for ending a test.
Exit: Returns to Service Main Menu. An ongoing test is stopped upon exit.
ENTER: is used for selecting the channel to be tested (moving the cursor). At entry the cursor is
placed at Port2.
Purpose: Used for setting up the control of e.g. the wet section during service. It is possible to change
or cancel the ABL50s normal action caused by the status of the various sensors.
Note: If a service setup is used, the contents of the status field (e.g. Ready) will be inversed.
This is to inform you that the instrument is not operating according to the Users Handbook.
Rinse only must be selected before power is switched off.
Inlet cover off must be selected before removing the partzotherwise the ABL50 will enter
into the Hold mode.
l . 6.5.1-15
WRITTEN 93.02.08
r
Display / Printout:
11-5
SERUICE 12:341
hruice setup
Rinse only Y6ZS 5-401
Print updatings No
-Gas/temp err-w5 Yes
Inlet/cover off No
-Waste check Yes
Print ext. errors No mf% Meatuvemen t l-701
Change 1 Exit c
Softkeys Change: Used for setting up. Default selection is No for all setups.
Rinse only: forces the ABL50 to only perform a Rinse when switch-
ing on the power (that is, no refill, no LS adjustment, no
warming period, etc.). It is possible to work with the
ABLSO as soon as the upstart rinse has completed.
Print upda tings: Provides the possibility of obtaining a printout of the
electrode updatings (measurements and calibrations).
Both electrode signals (mV or PA) and the calculated
values (e.g. mmHg) are printed out for measurements,
for calibrations only the electrode signals are printed..
Examples of printouts of the electrode updatings are
shown at the end of this section. The printout takes
longer than a complete measuring cycle and the ABL50
is locked during the printing (not possible to carry out
measurements etc - this is indicated by Printing in the
status field). The locked time may be up to 30 seconds
after the measurement rinse has completed.
-Gas/temp errors: forces the ABL50 to disregard gas mixer errors,
/n/et cover of: forces the ABLSO to disregard removal of the inlet
module or top part of thermostatted box (cover),
-Waste check: forces the ABL50 to stop monitoring the waste level.
Print ext.effofs: enables printout of extended error codes. All error
messages will be printed out as they appear (however,
not in the middle of e.g. a measurement result.
The various extended error codes are described in the
Troubleshooting Section.
Purpose: Used for resetting the sample and calibration counters to zero. In addition, the contents of
the user setup and service programs is reset to factory standard (default).
a Display:
SERUICE 12:34)
Clear all RQM
Char I Exit
l 6.5.1-17
WRITTEN 93.02.08
SERVICE PROGRAMS (ABL50)
Purpose: Used for printing out the error log which contains the up to IO lat?st different errors which
have occured on the ABL50. If the program is selected Printing error log,,is displayed for
approximately 3 seconds while the printer starts printing out the error log.
Note: Please note, that this program is unavailable in case the pririter is already busy
printing or in case the printer has not been selected in the setup program.
R A D I O M E T E R ABL5O
ERROR LOG
l
Cal lbvat ion 3-509 22:03 11.01.93
Cal ibvat ion 2-505 12:05 11.01.93 -
11:32 1 3 . 0 1 . 9 3 ANALYZER tt 1234567890
Purpose: The program enables suppression of frequent calibrations after startup and some mainte-
nance for e.g. demonstration purposes.
Note: If Yes is selcted, the contents of the status field (e.g. Ready) will be displayed in black
letters. This is to inform you that the instrument is not operating according to the Users
Handbook.
Display:
SERUICE 12:341
Ruxiliary setup
Supp. freq. cals No
Change Exit
? PH do2 PO2
UllI 1 WI uk 1 Ml imlb 1
a
28 -5u 7.143 -ulm 54.3 1035 70.0
29 -5a.47 7.143 -a% 54.3 103U 69.7
30 -50.47 7.143 -4oJ6 54.3 1027 69.5
31 a45 7.143 -10.84 54.3 1022 69.2
32 58.46 9.143 -40.81 54.4 1019 68.9
33 58.45 7.143 -40.64 54.4 1018 68.9
34 58.46 7.143 -40.63 54.4 1016 68.7
35 -50.44 7.143 -a@ 54.4 1011 6a.4
36 58.44 7.143 -40.82 54.4 10% 9.3
37 58.44 7.143 -4tl.81 54.4 Ial7 a.1
38 -5&u 7.143 -ul.Bl 54.4 1003 67.9
39 58.43 7.143 -10.81 54.4 loal 67.8
40 58.40 7.142 -ul.E2 54.4 999 67.6
41 a42 7.143 -4lm s.4 999 67.6
42 -58.40 7.142 -40.81 54.4 996 67.4
43 58.41 1.142 -40.7V 54.5 996 61.3
U -5e.39 7.142 -4oo.w 54.4 953 67.1
45 58.39 7.142 -am w.4 992 61.1
46 -5l.39 7.142 -40.80 54.4 990 66.9
47 -58.39 1.142 -4o.m 54.4 989 66.9
48 *.3V 7.142 -40.81 54.4 9m Q.8
49 58.39 1.142 -4a.m 54.4 9% 66.7
To enter the Service Test Program in the Keypad with Display unit press key number 3, 6 and 9
simultaneously and then turn power on. Alternatively, unplug and re-insert one of the plugs on the cable to
the unit while pressing the above-mentioned keys simultaneously.
During the test the Keypad with Display unit is off the Ring Net. This causes the LED CR805 on the Main
Computer PCB to light. At the same time the Ring Net connection is looped back in the Main Computer PCB
to be able to test the connection.
The test consists of four sequences. To step to the next sequence press any of two keys simultaneously.
When all sequences are passed, the unit performs a normal start-up procedure as described in chapter 6.3
Keypad with Display Self-test Program.
EPROM test.
When the service test program is entered the CPU performs a checksum check of the EPROM. During this
test the display will show the picture seen in figure 1. If an error is detected the CPU will stop allactivity.
KEYPAD - UNIT
VERSION: V.VV
EPROM-TEST: XXXX SS RR
Figure 1.
RAM test.
When the EPROM test is passed without error, the CPU performs a RAM test. This test is a write/read test
in which the following is carried out for all cells in succession: two patterns are written into and read from
the cell in question, first 01010101 (55) and then 10101010 (AA) to check for stuck at faults.
During this test the display will show the picture seen in figure 2. If an error is detected the CPU will stop
all activity.
KEYPAD - UNIT
VERSION: V.VV
EPROM-TEST: OK
RAM-TEST: XXXX aaaa
Figure 2.
When both the EPROM and RAM tests are passed without error, the display will look like figure 3.
~KEYPAD - UNIT
VERSION: V.VV
EPROM-TEST: OK
RAM-TEST: OK FCOO
Figure 3.
To continue the service test program press any of two keys simultaneously.
Display test.
When entering this test the display shows alternately black and white vertical lines, see figure 4. Each time
one key is pressed, the display is inverted (black lines get white and reverse). To continue the Display test
press any of two keys simultaneously. Now horizontal lines are displayed, as shown in figure 5. Again, each
time one key is pressed the display is inverted.
Figure 4.
Figure 5.
To continue the service test program press any of two keys simultaneously.
Keypad test.
In this Keypad test the display has a text for each key, as shown in figure 6. Pressing a key inverts the
corresponding text in the display.
KEY-TEST
SOFT1 SOFT2
7 8 9 FEED
4 5 6
1 2 3
0 . BACK ENTER
Figure 6.
To continue the service test program press any of two keys simultaneously.
KEYPAD - UNIT
VERSION: V.VV
COM-TEST: OK nnnn
ERR. nnnn
Figure 7.
To continue press any of two keys simultaneously. This causes the unit to perform a normal start-up
procedure, as described in chapter 6.3 Keypad with Display Self-test Program.
CRT Compartment
CRT-Deflection PCB
Video Amplifier PCB
Printer Control PCB
Therm0 Printer
Softkey Pad
Flomw Disk Control PCB ‘)
Floip) Disk Drive
Numerical Keyboard
Numerical Keyboard PCB /
Numerical Keypad
Computer Unit
Main Computer PCB
Main Memory PCB *)
Wet section
Wet section Control PCB
Wet section Distribution PCB
pH Preamplifier PCB Base unit
~02 Preamplifier PCB Gas Mixer Interface PCB
pCO2 Preamplifier PCB Power Supply PCB
Monochromator Interface PCB ) Software Gate
Waste Detector PCB *) Up to three (four *) ) optional
PCBs
Wet section
Wet section Con
Base unit
Gas Mixer Interface PCB
pH Preamplifier PCB
Power Supply PCB
~0, Preamplifier PCB
Main Computer PCB
pC0, Preamplifier PCB
One optional PCB
Memory Map
The total address domain for the 8OC186 Main processor is 1 Mbyte. However, since the memory
needed in the ABL is larger, up to 3 Mbyte, bank switching is included for the EPROM and the
EzPROM situated on the Main Memory PCB, and the NVRAM (Non Volatile RAM) situated on the
Main Computer PCB itself. The memory map is outlined as shown below:
Ban& switching The bank switching is obtained by the four bank select signals, BSO through BS3,
which are provided via the output port on the DUART. The banks are shown below.
- Bank 0 is special since word access is not possible (only byte access). Selection of which EZPROM is
selected when is determined by strap connectors mounted on the PCBs at the factory.
Bank
0000 AOOOOn-AFFFF, even addr.: NVRAM
odd addr.: E2PROM QD19
80000,-BFFFF, e v e n addr.: EZPROM QD20
odd addr.: EZPROM QD21
0001 EPROM 0 (QDl)
0010 EPROM 1 (QD.3
0011 EPROM 2 (QW
0100 EPROM 3 (QW
0101 EPROM 4 (Qbs)
0110 EPROM 5 (Qbs)
0111 EPROM 6 (Qo7)
DRAM interface
The address lines connected to the DRAM Controller itself (AXl-AX18) are internally converted into
row and column addresses, each consisting of nine bits. In this way, each DRAM is organized as a ma-
trix consisting of 512 rows and 512 columns. The DRAM Controller provides a time multiplexed ad-
dress output for the DRAM S; the CASx and RASx signals indicate to the DRAMS that Column or Row
address, respectively, is now available. Each bank has its own set of CASx and RASx signals.
The combined write enable and bank select signals (WE-BANK-LOW and -HIGH) for the DRAMS
shown in Figure 7.1.1.8. AX0 (active low) are used for selecting the lower byte, and BHeX (active low)
is used for selecting the high byte. The AACK signal is used for synchronizing the bank select (and
therewith the write enable) to the DRAM refresh cycle. The synchronized bank select is gated
together with the actual write signal WE, providing combined bank select and write signals for the
DRAMS. The AACK signal is also used for deciding which bank (or both banks) is to be accessed at a
given time (high or low byte, or word access).
DRAM refresh
The storage elements of each DRAM cell is a capacitor which eventually discharges due to leak cur-
rents. The DRAMS are continuously refreshed to prevent the stored data from becoming lost. During
a refresh cycle, the charge of each capacitor is read, amplified and stored on the capacitor; this all
happens within the DRAM chip itself. The refresh cycle consists of 512 subcycles in which one row is
refreshed. The row address output of the DRAM Controller is incremented continuously, and when
the RASx signal is low, the addressed row is refreshed. The entire DRAM domain is refreshed every 7
ms. The DRAM Controller sets up the SRDY signal for the main CPU and can thereby hold the CPU in
case it tries to access the DRAM during a refresh cycle.
Parity check
The parity check circuit consists of a parity generator and some gates. The purpose of the circuit,
which is shown in Figure 7.1.1.8 overleaf, is to check if the data, stored in the DRAM, do not change
during storage. When data are written into the DRAM, a parity bit (even ) is calculated and stored in a
separate DRAM. When the same data later on is read again, the parity bit is re-calculated and com-
pared with the one calculated previously. If the two parity bits differ, the Parity Error signal goes high
causing a NMI signal to the main CPU.
First three control words are transferred, a Request indicating to the other PCBs that the Main
Computer wishes to use the Ring Network, then a Destination indicating to which PCB it wishes to
transfer data (destination address), and finally a Length indicating the length of the message
(number of bytes to be transferred) to the destination PCB. Then the data are transferred byte by
byte, and then finally the control word Stop is sent to indicate that the transmission has been com-
pleted and that the Ring Network is available for the other PCBs.
If two PCBs output a Request at the same time, the one with the highest priority will be granted
permission to use the Ring Network.
The destination PCB will acknowledge the Destination and Length words to tell the source that
the words have been received and that the destination is ready to receive data. The destination PCB
will echo all received data bytes together with an acknowledgement to tell the source that the data
have been received and to enable the source to check for transmission errors.
PCBs which do not communicate on the Ring Network at a given time, function as repeaters for the
PCBs which do.
110 ports
The eight bit wide output port provides the following signals:
- BSO through BS3 providing a four bit bank select for the various types of memories, please see the
paragraph concerning memory map. The bank select signals are applied to the decoders which
generate the actual chip select signals.
- Two outputs for controlling the LEDs, CR 501 and CR502. The LEDs are used for indicating RAM
and PROM write errors in the I/O processor circuit.
The seven bit wide input port is used for the following purposes:
- One bit is connected directly to the Attention button (ATTN).
- Five bits are connected to the test points TPl/l through TP1/5 which are used for test purposes at
the factory.
Processor/Memory
The Processor/Memory circuit is built-up in an conventional way with a latch separating the time
multiplexed data and low order address bus and an address decoder generating the various chip se-
lect and latch enable signals.
The memory consists of a 32 kbyte PROM (software) and a 8 kbyte RAM (for temporary storage of
data-data buffer for external communication)
Housing: Shield
Beeper
P1.5 of the I/O processor is used for controlling the beeper, which is used for several purposes, e.g. to
indicate a correct reception of bar code and to indicate error conditions.
QUALICHECKTM is a trademark of Radiometer A/T, Denmark and IS registered in the USA and in some other countries.
Memory Map
The total address domain for the 8OC186 Main processor is 1 Mbyte. However, since the memory
needed in the ABL is larger, up to 1.5 Mbyte, bank switching is included for the FEPROM and the
EZPROM, and the NVRAM (Non Volatile RAM). The memory map is outlined as shown below:
10 banks each of 128 kbytes (FEPROM, EZPROM, NVRAM) AOOOO, through BFFFF,
Bank switching The bank switching is obtained by the four bank select signals, BSO through 853,
which are provided via the output port on the DUART. The banks are shown below.
- Bank 0 is special since word access is not possible (only byte access)
Bank
0000 AOOOO,-AFFFF, even addr.: NVRAM
80000,-BFFFF, odd addr.: E*PROM QD2 1
0001 FEPROM (QD202 + QD2 12)
0010 FEPROM (QD202 + QD212)
0011 FEPROM (QD203 + QD213)
0100 FEPROM (QD203 + QD213)
0101 FEPROM (QD204 + QD2 14)
0110 FEPROM (QD204 + QD214)
0111 FEPROM (QD205 + QD215)
1000 FEPROM (QD205 + QD215)
1001 empty
1010 empty
1011 empty
1100 empty
1101 empty
1110 FEPROM (QD201 +QDZll) l )
1111 FEPROM (QD201 +QDZll) l )
l) These two FEPROMs are replaced by the Boot PCB in the memory domain when connected during upgradmg of the
software.
rectly the status signals SO, Sl and 52 from the main processor; the decoding is used for controlling
the WE (write enable) signal.
DRAM interface
The address lines connected to the DRAM Controller itself (AXI-AX18) are internally converted into
row and column addresses, each consisting of nine bits. In this way, each DRAM is organized as a ma-
trix consisting of 512 rows and 512 columns. The DRAM Controller provides a time multiplexed ad-
dress output for the DRAMS; the CASx and RASx signals indicate to the DRAM S that Column or Row
address, respectively, is now available. Each bank has its own set of CASx and RASx signals.
The combined write enable and bank select signals (WE-BANK-LOW and -HIGH) for the DRAMS
shown in Figure 7.1.2.10. AX0 (active low) are used for selecting the lower byte, and BHEX (active
low) is used for selecting the high byte. The AACK signal is used for synchronizing the bank select
(and therewith the write enable) to the DRAM refresh cycle. The synchronized bank select is gated
-
together with the actual write signal WE, providing combined bank select and write signals for the
DRAMS. The AACK signal is also used for deciding which bank (or both banks) is to be accessed at a
given time (high or low byte, or word access).
DRAM refresh
The storage elements of each DRAM cell is a capacitor which eventually discharges due to leak cur-
rents. The DRAMS are continuously refreshed to prevent the stored data from becoming lost. During
a refresh cycle, the charge of each capacitor is read, amplified and stored on the capacitor; this all
happens within the DRAM chip itself. The refresh cycle consists of 512 subcycles in which one row is
refreshed. The row address output of the DRAM Controller is incremented continuously, and when
the RASx signal is low, the addressed row is refreshed. The entire DRAM domain is refreshed every 7
ms. The DRAM Controller sets up the SRDY signal for the main CPU and can thereby hold the CPU in
case it tries to access the DRAM during a refresh cycle.
Parity check
-
The parity check circuit consists of a parity generator and some gates. The purpose of the circuit,
which is shown in Figure 7.1.2.10 overleaf, is to check if the data, stored in the DRAM, do not change
during storage. When data are written into the DRAM, a parity bit (even ) is calculated and stored in a
separate DRAM. When the same data later on is read again, the parity bit is re-calculated and com-
pared with the one calculated previously. If the two parity bits differ, the Parity Error signal goes
high.
The real time clock puts 19 or 20 in front of the year and is capable to keep track of leap years.
The real time clock includes battery backup, see the description below.
Battery backup
The Real Time Clock and the NVRAM include battery backup to prevent the data in the NVRAM to be
erased in case the ABL is switched off or in case of a power failure. The circuit consists of a programm-
able voltage detector, a 3.0 V battery, and a number of transistors.
The voltage detector prevents the processor from accessing the real time clock and the NVRAM, in
case the + 5 V supply decreases to below + 4.6 V. This is done by disabling the RD, WR, and chip select
signals (via transistors).
The supply voltage switches from the + 5 V supply to the battery when the + 5 V has decreased to
below 2.4 V (the battery voltage).
First three control words are transferred, a Request indicating to the other PCBs that the Main
Computer wishes to use the Ring Network, then a Destination indicating to which PCB it wishes to
transfer data (destination address), and finally a Length indicating the length of the message
(number of bytes to be transferred) to the destination PCB. Then the data are transferred byte .~y
byte, and then finally the control word Stop is sent to indicate that the transmission has been com-
pleted and that the Ring Network is available for the other PCBs.
If two PCBs output a Request at the same time, the one with the highest priority will be granted
permission to use the Ring Network.
The destination PCB will acknowledge the Destination and Length words to tell the source that
the words have been received and that the destination is ready to receive data. The destination PCB
will echo all received data bytes together with an acknowledgement to tell the source that the data
have been received and to enable the source to check for transmission errors.
PCBs which do not communicate on the Ring Network at a given time, function as repeaters for the
PCBs which do.
110 ports
The eight bit wide output port provides the following signals:
- BSO through BS3 providing a four bit bank select for the various types of memories, please see the
paragraph concerning memory map. The bank select signals are applied to the decoders which
generate the actual chip select signals.
- NMI-EN is used for enabling the output of the NMI circuit.
- FLASH-PROG is used for enabling the 12 V supply for programming the FEPROMs.
The seven bit wide input port is used for the following purposes:
- One bit is connected directly to the Attention button (ATTN).
- Three bits are connected to the test points TPl/l through TP1/3 which are used for test purposes
at the factory.
- 12V-FL-ERR (LO-HI) indicates if the 12 V supply is OK.
_ 12V-FL-ON indicates if the SW switch on the rear of the Main Computer PCB is in the correct posi-
tion (for programming or for normal operation).
- FDC-INST detects if a floppy disk drive is installed.
Processor/Memory
The Processor/Memory circuit is built-up in an conventional way with a latch separating the time
multiplexed data and low order address bus and an address decoder generating the various chip se-
lect and latch enable signals.
The memory consists of a 32 kbyte PROM (software) and a 8 kbyte RAM (for temporary storage of
data - data buffer for external communication)
bit 11 brt 1
sv ._r___T___T___T___T___T___T--:___T---I_
I I I I I I I I I
I I I I I I I I I I
I I I I
nv
_
:
I
I I
___I___I___ ___L_--J.-__1__-L___ : ___J
I
I II 1 I
I 8 data bits (ASCII) I
Stop bit Start bit
If the Main Computer has commands to the keyboard, it first checks if the keyboard is transmitting
data. If not, it will set the clock line low (request to send) for 60 ms. Then, it outputs the start bit and
frees the clock line, enabling the keyboard to time the data flow from the Main Computer to the key-
board. Since the Main Computer is the master, it may overrule the transmission from the keyboard by
forcing the clock line low (this will cause the keyboard to stop transmission).
Pin designation: Pin 1: Clock out / in
Pin 2: Data in/out
Pin 3: NC
Pin 4: Ground
Pin 5: +5V
Housing: Shield
When connecting both an alphanumerical keyboard and an external CRT monitor, it is possible to set-
up a remote working place, for example in connection with editing one of the data bases (patient in-
formation, etc.)
Housing: Shield
Beeper
P1.5 of the I/O processor is used for controlling the beeper, which is used for several purposes, e.g. to
indicate a correct reception of bar code and to indicate error conditions.
QUALICHECKTM is a trademark of Radiometer AK, Denmark and is registered in the USA and In some other countries.
The total address domain for the 8OC188EB Main processor is 1 Mbyte. The Main Computer PCB
includes 512 Kb FPROM, 128 Kb SRAM (Static RAM) and a Memory Card connector with up to 512 Kb
addressable memory area. To be able to boot the Main Computer when there is no functional program
in the FPROM, a Memory Card Boot Switch is mounted. If the Memory Card Boot Switch is pressed
during power up, an alternative memory map is selected. The two memory maps are outlined as shown
below:
FFFFFh
QD651
FPROM-0
256Kb
COOOOh Memory Card
512Kb
QD652
FPROM-1
256Kb
80000h
QD651
60000h FPROM-0
256Kb
Memory Card
256Kb 40000h
QD652
20000h FPROM-1
Static RAM 256Kb
128Kb
OOOOOh
Normal Boot Memory Card
Boot
(SW update)
The Main Processor 8OC188EB may receive interrupts from a number of external and internal sources,
as shown below.
Battery Backup
The Main Computer PCB includes a 3V Lithium cell to keep the information in the RAM and the clock
running when the ABL is switched off or in case of power failure. A special switching circuit separates
the ground of the Real Time Clock and the RAM from the digital ground when the mains power is off.
This is done to extend the battery life by avoid current to leak backwards through the control pins to the
unpowered circuit. When measuring the battery voltage be sure to use the negative pole on the battery
and not a normal ground pin.
Programming
The FPROMs are programmed via a special program located on the Memory Card. Before applying the
programming voltage to the FPROMs, a window comparator checks that the voltage is within the limit
11.4V to 12.6V. If the voltage is outside the limits, the beeper beeps and an error code is displayed on
the four LEDs located under the Memory Card plug. For further information please refer to chapter
2.17: Software Upgrade.
Interface
The Main Computer PCB contains a USART with one parallel interface and one serial interface.
The parallel interface is connected to a 25 pin female plug for use with a printer using standard
Centronic interface as the PRS15. The serial interface is intended for connecting a ticket printer and
NOTa computer, due to missing galvanically isolation and missing protocol support.
Pin 1: DCD In
Pin 2: RXD In
Pin 3: TXD out
Pin 4: DTR out
Pin 5: GND ___
Pin 6: DSR In
Pin 7: RTS out
Pin 8: CTS In
Pin 9: RI In
0
7.1.3 - 6 WRITTEN 92.11 .Ol
MAIN COMPUTER
a
Jl Internal connector to Power Supply
Memory Map
Bank
0000 AOOO+,-AFFFF, odd addr.: EZPROM QD19
BOOOOH-BFFFF, even addr.: EZPROM QD20
odd addr.: EzPROM QD2 1
0001 EPROM 0 (QD1)
0010 EPROM 1 (QD2)
0011 EPROM 2 (QD3)
0100 EPROM 3 (QD4)
0101 EPROM 4 (QW
0110 EPROM 5 (QW
0111 EPROM 6 (407)
Revised 91.04.11
RSl50, SERIAL OUTPUT
0
Each UART includes the two data lines TxD and RxD. One output and two input lines are used for the
hardware handshake signals RTS, CTS, and DSR. The signals from/to the UART are passed through
buffers and optocouplers. The RS-232 driver, which converts output signals from TTL levels to + 12 V
(RS-232-C levels) and input signals from RS-232-C levels to TTL levels. The output connector is a nine
pin D-connector -the pin configuration is shown below.
Communication speed (50 through 19200 baud), protocol, frame format, etc. may be set-up, please
refer to the setup programs described in the Users Handbook.
Each time one of the UARTs receives a data byte, an INTl to the CPU is generated -this causes the CPU
to read the data.
Each UART includes two status signals, RxRDY and TxRDY. RxRDY indicates that the UART has
received a data byte from an external device, and TxRDY indicates that the UART is ready to receive
another data byte from the CPU for transmission to an external device. These status signals are
connected to port 1 of the CPU.
1 NC Not connected
2 RxD Data received by the Serial Output PCB
3 TxD Data transmitted from the Serial Output PCB
4 DTR Data Terminal Ready (ON when the ABL is turned on)
5 GND Signal ground
6 DSR Data Set Ready (ON when the external equipment is on)
Request To Send (ON when the output buffer of the Serial
7 RTS
Output PCB is not empty)
Clear To Send (ON when the external equipment is ready
8 CTS
to receive data)
9 NC Not connected
Voltage Supplies
The Serial Output PCB receives a 7 VAC from the Power Supply. This AC voltage is full-wave rectified
at the input of the RS-232-C Output PCB providing an unregulated 7 VDC. The voltage supplies
needed for the various circuits are all generated locally on basis of the full-wave rectified voltage. All
voltages are fixed and cannot be adjusted. The voltages are:
+SV, This voltage is regulated to + 5 V, using an LM2925 regulator. The + 5 V regulator
generates a reset signal for the CPU directly (see description below).
The voltage is used for supplying the digital ICs which are not connected to the ABL and
the external equipment.
The following supplies are derived using a DC/DC converter built-up around a tranformer. The
individual supplies are stabilized using zener diodes.
-+ 12 v For RS-232-C driver.
+5V For RS-232-C driver and optocouplers.
Reset The + 5 V regulator provides a reset signal for the CPU. The reset is removed (goes from
low to high) approximately 80 ms after the + 5 V supply has reached its nominal value.
The reset from the + 5 V regulator is gated together with the reset from the CPU
watchdog before it is applied to the reset terminal of the CPU and the DUART. The CPU
can then determine the source of a reset by checking the status of the WI-STATUS
signal.
Computer/Memory
The PCB is built-up around an 8OC535 CPU. The computer/memory part is built-up in a conventional
way with a latch separating the time multiplexed data and low order address bus and a decoder for
generating the various chip select signals.
The memory consists of a 64 kbytes EPROM (software and character table) and a 32 kbytes RAM (stor-
age of the data to be printed out).
The clock circuit consists of a 7.3728 MHz crystal which supplies the frequency used for the CPU and
the presettable counter.
Printer Interface
The printer interface consists of a number of subcircuits which are described below.
Transfer of data This circuit consists of a five bit presettable counter (an EEPLD, Electrical
Erasable Programmable Logical Device) and an eight bit shift register. The transfer is carried out in 64
subcycles for each dot line, that is, one for each character, as follows:
Subcycle: The dot pattern for the left-most character (No. 1) is loaded into the shift regis-
ter. At the same time the presettable counter loaded is with eight (8). Then the counter starts count-
ing down to zero, transferring the data to the shift register in the thermal head at a rate of 3.6864
MHz (half the crystal frequency). When the counter reaches zero, it stops. Then the counter is loaded
l
with two (2), after which it is restarted (In case font B is used, the counter is loaded with six (6) the sec-
ond time instead of two). When the counter reaches zero, it stops again. Then the subcycle for char- _
acter 2 commences, and so forth.
Four port bits on the CPU are used for loading the four upper bits of the counter (bit 0 is always
loaded with 0, internally in the EEPLD).
Clk out and Prn elk are inverted in relation to each other, so that the one clocks the data out of
the shift register on the Printer Control PCB, and the other clocks the data into the shift register of
the therm0 head.
By setting the Width signal low, the frequency of Prn elk is doubled - this enables printout of the
character fonts with twice normal width (used for printout headings, e.g. RADIOMETER ABL500).
Printout of dot line When the 640 bit wide shift register in the therm0 head is filled, the CPU lat-
ches the dot pattern into the output latch (using a port bit).
Then it sets Enable 1 - 5 high simultaneously (five port bits) - these signals are applied to a buffer
which is enabled by the Heat signal. Then the Heat signal is set high for a period of time to activate
the heating elements.
l
7.5-4
PRINTER and PRINTER CONTROL
the CPU. This is done during adjustment of the printing intensity and when enabled via
a service program.
The + 24 V supply is held disabled until the selftest program has completed using the
24 V enable signal. The 24 V enable signal controls a transistor which causes the
+ 24 V to be decreased to approximately + 7.5 V during the selftest. This means that the
+ 5 V regulator which is powered from the + 24 V supply is unaffected by the disabling.
+ 15v This voltage is used for supplying the operational amplifier. It is derived from the + 33 V
supply using a 15 V zener diode.
The level of the + 15 V supply may be monitored via one of the A/D converter inputs of
the CPU. This is done during the final test in the production and when enabled via a ser-
vice program.
+SV, This regulated voltage is used for supplying the digital KS. The voltage is regulated
using an LM2925 regulator. The + 5 V regulator generates a reset signal for the CPU di-
rectly (see description below).
+sVAD This regulated voltage is used as a reference voltage for the ND converter in the CPU. It
is taken from the + 5 Vo via a low pass filter.
Reset The + 5 V regulator provides a reset signal for the CPU. The reset is removed (goes from
low to high) approximately 80 ms after the + 5 V supply has reached its nominal level.
7.57
0 7.6 WET SECTION CONTROL
Overview
Figure 7.617 on Page 19 gives an overview of the Wet Section Control PCB.
The main functions on the Wet Section Control PCB are:
- A/D conversion Of electrode, temperature and gas mixer signals, and additional signals from the
Monochromator Interface PCB (ABL510/520 only).
- Control of pump motors, tubing/membrane valves and the main valve.
- Control of temperatures.
- Control of the gas mixer.
- Control and monitoring of the liquid sensors and inlet.
- Communication via the Ring Network.
- Control of the optical system, ABL51 O/520 only.
The PCB includes its own CPU, and its function is in principle independent of the Main Computer PCB.
The Main Computer PCB is the master which may instruct the Wet Section Control PCB to Perform a
GM Upstart and a Gas Start. After having received the instruction, the CPU on the Wet Section Con-
trol PCB takes over the control, setting up the wet section and gas mixer to move the gas, obtain
results from the electrodes, perform the A/D conversion etc. During the calibration, the Wet Section
Control PCB presents the data to the Main Computer PCB. When the calibration has ended, the Main
Computer PCB then instructs the Wet Section Control PCB to perform either a GAS1 End or a Rinse de-
pending on which calibration is performed.
Two versions
Please note that the Wet Section Control PCB is found in two versions as follows:
901-774 Included in both ABLSOO and ABL510 from Run 1.
902- 157 Introduced in ABLSOO from Run 50 and ABL51 O/520 from Run 8
The basic differences between the two are shown below, a compatibility table is given on Page 18.
Power supply Cascade coupled linear regulators Switching circuit with transformer
Waste detector circuit On separate PCB, two NTCs On Wet Section Control, one NTC
*) 901-774 is modified to have these specifications when returned for replacement under the trade-
in arrangement.
Computer/Memory
The Computer/Memory part is shown in Figure 7.6.2 overleaf.
The circuit is built-up around a 8032 CPU. The computer/memory is built-up in a conventional way
with a latch separating the time multiplexed data and low order address bus and an address decoder
generating the various chip select signals.
The memory map is as follows:
FF7F,
8000,
7FFF,
32 kbytes of PROM
OOOOH
The memory consists of 64 kbytes of PROM (only the lower 32 kbytes are used for startup, communi-
cation programs, etc.). During opstart, that is, when power is switched on, approximately 24 kbytes of
program (wet section control) from the Main Memory PCB is loaded into the 32 kbyte RAM memory
from which it is then executed. A 128 bytes EZPROM is used for storing pO2 zero and waste calibra-
tion values (the latter only for the original waste detector, 902-053). The remaining part of the RAM
(32 kbytes - 24 kbytes - 128 bytes) is used for temporary storage of data. The latter 128 bytes are used
for the various chip select signals. Communication with the PROM and RAM is carried out in a parallel
manner, whereas with the EJPROM it is serial.
A buffer is able to disconnect the PROM from the data bus during test of the PCB at the factory (when
running external programs (EP-CE, external program chip enable) or when setting the CPU to the free
run mode (F&EN, free run enable).
The data signals (TxD and RxD) are transferred through opto-couplers to obtain galvanical isolation.
The clock circuit consists of a 7.3728 MHz crystal and a counter which generates the various clock sig-
nals used on the Wet Section Control PCB as follows:
7.3728 MHz is used as a basic clock for the CPU. In addition, it is used as a reference clock for the
A/D converter for the various analog signals (except electrode signals), for generating
strobe signals for the heat transistors, and for generating the clock signal for the
pump stepping motor in the gas mixer.
460.8kHz is used by the programmable interval timers for generating clock signals for the pump
stepping motors, and is used by the CPU Watchdog.
115.2 kHz is used for generating the AC voltages for the electrode channels, as reference frequen-
cies for the respective A/D converters, and in the valve holding circuit for generating the
strobe signal.
57.6 kHz is used in the valve holding circuit for generating the strobe signal.
rents in the plugs). The preamplifiers, which are built-up around ADSlSs, function both as impedance
converters and preamplifiers.
The preamplifier for the pH and pCO2 channels amplify the input voltage by a factor of 10.
The preamplifier for the ~02 channel functions as a current to voltage converter (100 pV / PA).
Multiplexer: Each electrode channel include at the input end a multiplexer via which the CPU may
select which signal is to be converted. The signals are: the electrode signal, a factory adjusted ref-
erence voltage, which is +2,500 mV for pH and pCOz and -2,S.OO mV for the ~02 channel, and
ground (the tolerence for the reference voltages is + 0.5 mV). The reference voltage and GND are
used for determining the transfer function of the V/F converter (automatic calibration of the zero
point and the sensitivity) -this eliminates the need for manual adjustments. For the ~02 channel also
the polarization voltage and the battery voltage are connected to the multiplexer. The selected sig-
nal is buffered before it is applied to the V/F-converter.
Voltage to Frequency Converter: The V/F converter is in principle a voltage to duty cycle converter
where the output is gated together with the reference frequency (115.2 kHz). The V/F converter
generates a pulse train where the average frequency is proportional to the input voltage - the maxi-
a
-
l-l l-l n
225 Hz, 24 V, 12.5% DC
;,.:g:?.;.i:;.y: .: ]
\0\ :..y
-b , ..::,: ..:::.
::$T.:;-i
Channel
1
Background
OptIcal / light
fibres
L
The LEDs in the liquid sensors are driven by a 225 Hz, 24 V 12.5% duty cycle signal. The output of the
phototransistor will also be a 225 Hz signal, however, it will include a DC component caused by the
background light (the liquid sensor housings are made of clear plastic). The phototransistor signal is
therefore applied to a differential amplifier, where the DC component is stored on a capacitor, SO
that it may be subtracted from the total signal, see overleaf.
The output of the amplifier is then applied to a comparator, where the liquid sensor signal is com-
pared to a Trigger voltage derived from a D/A converter, During auto cleaning (every 24/12/8 hours) a
liquid sensor adjustment program is automatically called to calibrate the sensors. The sensors are
emptied, and then the trigger voltage is increased until the output of the comparator no longer tog-
gles (= maximum output of amplifier). This voltage is set to 100%; during normal operation the trig-
ger voltage is set to 60% of maximum (calibration value). If the output is above 60%, the sensor is
hereafter regarded as being empty; if the output is below 60%, the sensor is regarded as being full.
Differential
amplifier
Photo
transistor
to pot-t
Digital value
Resolution
In the service program, the present value is obtained by increasing the trigger voltage until the
comparator output no longer toggles. During normal operation, the circuit only has two states,
empty and full since the trigger voltage is fixed. -
The trigger voltage for the latest six calibrations is stored so the drift may be obtained in the liquid
sensor service program.The maximum output of the differential amplifier (trigger voltage) may
change over time due to contamination of the channel through the sensor, deteriation of the LED
etc.
The trigger voltage may be set in steps of 20 mV if the Resolution Voltage is above 1.25 V (low resolu-
tion) and in steps of 5 mV if below 1.25 V (high resolution).
The CPU reads the status of the port syncronuously with the 225 Hz clock, that is, syncronuously with
the LEDs being lit.
As for the various other parts of the wet section, the WD-ERROR signal disables the liquid sensor sup-
ply in case the CPU is faulty.
Dosage
pllsec ( 0 / 6 ( 8 1 13 ( 22 1 36 ( 60 1 100
Pl-DIR Decides the pump direction by changing the phase of the current applied to the
motor windings.
PQX-X Outputs for the motor windings.
Valve driver
The six valves, Vl through V6 (V3 is not mounted in ABLSOO), are supplied from the + 24 V DC via the
driver circuit shown overleaf.
To ensure a quick activation and to minimize the current consumption, the valves are driven as fol-
lows:
Pulling: The valve pull circuit is built-up around the counter and Decoder No. 1. When a valve is to
be pulled, the CPU first enables the valve itself via the programmable peripheral interface setting the
appropriate VALVE-X line low (X may be 1 through 6). Then it enables the pull circuit by setting up
the lines VALVE-SEL-X (1 through 3) via the programmable peripheral interface. Then it resets the
counter applying + 24 V DC to the selected valve. The counter counts at a rate of 28.8 kHz and stops
itself after 71 ms (when Ql 1 goes high) disabling the pull circuit.
Holding: When the valve has been pulled, it is still held activated by the hold circuit built-up
around Decoder No. 2. This decoder is coupled as a Johnson counter which strobes all enabled valves
in succession as follows:
901-774: One valve at the time is energized with 4.3 ps 24 V pulses at a rate of 28.8 kHt. The re-
sulting hold signal for each valve is a 28.8 kHz 24 V square wave signal with a duty cycle
of 12.5% (the power applied to the valve corresponds to a 3 V DC voltage).
902- 157: Two valves at the time are energized with 4.3 ps 24 V pulses at a rate of 57.6 kHz. The
resulting hold signal for each valve is a 57.6 kHz 24 V square wave signal with a duty
cycle of 25% (the power applied to the valve corresponds to a 6 V DC voltage).
See the note concerning compatibility on Page 18.
Revised 91.04.05
WET SECTION CONTROL
two different speeds as follows: When the mean duty cycle of the Block heat signal decreases to
below 2 % (ambient temperature approx. 28 C), the fan is turned on with the low speed (supply volt-
age is 9.0 V). If the duty cycle of the Block heat signal decreases to below 2 % with the fan at the slow
speed (ambient temperature approx. 34 C), the fan is set to turn at the high speed (supply voltage is
13 V). The speed of the fan is controlled by means of the D/A converter (which is also used in the
liquid sensor circuit). The maximum current supplied to the fan is limited to 250 mA.
Cover off A reed relay in the base plate for the thermostatted box senses whether the lid of the
thermostatted box is mounted or not - if removed, COVER-OFF is set high and a message is sent to the
Main Computer PCB.
l Heat1 and Heat2 The temperature of Heat1 and Heat2 is regulated to 37C * 1C. The regulation
circuit consists of a switching FET transistor, a ceramic heating substrate and an NTC resistor which is
calibrated to be 6.1 kQ at 37.OC using a parallel resistor.
l AUX In a test situation, it is possible to connect e.g. a variable resistor to the AUX input of the
temperature circuit.
To
Gas Mixer
Interface
PCB
On 902-157 the above three AC voltages are obtained on three separate windings on
the transformer in the fly back converter.
The AC voltages are regulated in the electrode channels themselves.
All voltages except the + 5 Vref are fixed and cannot be adjusted.
Reset The + 5 V regulator provides a reset signal for the CPU. The reset is removed (goes from
low to high) approximately 80 ms after the + 5 V supply has reached its nominal value.
The reset from the + 5 V regulator is gated together with the reset from the CPU watch-
dog before it is applied to the reset terminal of the CPU. The CPU can then determine
the source of a reset by checking the status of the WD-STATUSsignal.
Compatibility Survey
ABL510
ABLSOO ABLSOO ABL510 ABL520 ABL505
AB L520
Rl-R49 R50 + Rl-R7 Rl-R7
R8 +
90 l-774 YE5 NO 2) NO NO NO 2) NO 2)
w. 368-673
1) Only PCBs delivered from RADIOMETER after week 42/1990 can be used due to hardware re-
quirements (improvement of A/D converter and inclusion of connector for the Software Gate)
Voltage Supplies
The Wet Section Control PCB receives an unregulated + 33 V DC supply from the Power Supply. The
voltage supplies needed for the various circuits are all generated locally on basis of the + 33 V. How-
ever the way in which the voltages are generated on 901-774 and 902-l 57 respectively is different.
On 901-774 the voltages are generated using a forward converter with cascade coupled regulators.
On 902-157 the voltages are generated using a flyback converter built up around a transformer and
an LTl172CT regulator (reduces power consumption). The flyback converter is selfstarting (oscillates
at approx. 100 kHz), but as soon as the clock circuit has started up it is syncronized to the 115.2 kHz
clock.
The voltages are:
+33VDC This unregulated supply is received directly from the Power Supply. It is used for sup-
0 plying the stepping motors for the two main pumps Pl and P2 and the main valve MV. It
is also used for the three heating elements. In addition, the voltage is supplied to the
Gas Mixer Interface PCB.
+24VDC This regulated voltage is derived from the + 33 V supply using an LM317 regulator and is
used for supplying the liquid sensors LSl through L55 (the voltage is also supplied to the
auxillary connector). It is also used for supplying the valves Vl through V6.
+18VDC This unregulated supply is used for the fan (only found on 902-l 57).
+lSVDC This supply is regulated using an LM340T15 regulator and is used for supplying the inlet
section, the V/F converter (except for electrode signals), and the liquid sensor detection
circuit. On 901-774 it is derived from the + 24 V supply - on 902-l 57 it is derived via the
flyback converter from the + 18 V supply.
- 15VDC This supply is regulated using a LM320T15 regulator and is used for supplying the analog
circuits. On 901-774 it is derived via a transformer (DC/DC converter) - on 902-157 it is
derived via the flyback converter.
+lOVDC This supply is used for the measuring chamber light. On 901-774 it is derived from the
+ 33 V supply via a switch mode regulator (4193), the supply includes a 300 mA current
limitation - on 902-157 it is derived from the flyback converter. On 902-157 the 10 V is
fed back to the flyback regulator to adjust the power transferred.
+SVDC This supply is used for the various digital circuits. On 901-774 the supply is derived from
the + 10 V supply using a LM2925 regulator - on 902-157 is derived from the fly back
converter using a L387A regulator.
The + 5 V regulator generates a reset signal for the CPU directly (see description below).
The voltage is used for supplying the digital KS (CPU, memory I/O ports, gates, etc.). A
manual reset may be obtained by shorting together the two pins of J2B momentarily.
+SV,fDC This regulated voltage is derived from the + 15 V supply using a reference zener diode.
The voltage is used as a reference voltage for the temperature monitoring circuit, the
V/F converter (except for electrode signals), and the D/A converter in the liquid sensor
detection circuit.
AC supplies On 901-774 the transformer used in the DC/DC converter includes three additional wind-
ings providing three separate AC voltages, one for each electrode channel.
Vertical Deflection
The vertical deflection circuit is built-up around a TDA2653 chip. The input to the circuit is the VSYNC
signal (50 or 60 Hz). The TDA2653 includes an oscillator which is triggered by the VSYNC signal to run
at a frequency of 50 or 60 Hz (picture frequency). The oscillator will continue to oscillate (at a fre-
quency of approximately 50 Hz), in case the VSYNC signal from the Main Computer PCB becomes lost;
this ensures that the beam will not be concentrated in a single horizontal line. The free running fre-
quency may be adjusted.
A constant current generator charges a capacitor, providing a ramp output. The voltage across this
capacitor determines the current through the deflection coil and thus the deflection of the electron
beam. When the electron beam reaches the bottom of the screen, the capacitor is shortened via a
diode. This causes a reversed field in the deflection coil, causing the electron beam to start its flyback.
The maximum magnetic field in the deflection coil, that is, the picture height, as well as the linearity,
S-correction of the deflection curve, may be adjusted.
7.84
POWER SUPPLY
SENSE The + 5.1 V supply is regulated so that the voltage on the Main Computer is cor-
rect. The SENSE line is a current less line for monitoring the level on the Main Computer PCB. In this
way the Power Supply compensates power loss in the interconnecting cable.
12VDC
This voltage is supplied to the Main Computer PCB where it is used for generating a 1.59 reference
voltage for the R, G, and B signals provided by the VSDD (Video Storage and Display Device). In addi-
tion, the voltage is supplied to the CRT Deflection PCB and the Floppy Disk Control PCB (the latter
only in ABL500/510/520 below Run 50/B/B).
The switching frequency of the regulator is approximately 50 kHz.
Soft start The regulator employs a soft start-up which slowly increases the maximum cur-
rent that may be drawn from the supply. This ensures that the regulator is not overloaded when
switching the ABL on.
Current limitation The maximum current to be drawn from this supply is 3.2 A (set by the fuse F3,
3.2 A).
Power Valid The regulator provides a power valid signal which is applied to the Main Com-
puter PCB. The power valid signal will go low in case the 20 V DC at the input of the regulator de-
creases to below 16 V, that is, when the mains supply decreases to below 80% of its nominal value.
Protection The regulator will automatically shut off in case it becomes overheated.
11 VAC
The transformer also provides 11 V AC which is connected to the Monochromator Interface PCB in
ABL510/520. The supply is used for the hemolyzer and lamp drivers.
33VDC
Another secondary of the transformer Tl provides a 33 V DC voltage for the Wet Section Control PCB.
The voltage is also supplied to the Gas Mixer Interface PCB. The voltage is used for supplying the
Introduction
The power supply is found in two versions (where both the transformer and the Power Supply PCB)
are different. The Power Supply consists of the following parts:
- Mains filter including the main fuse.
- Mains voltage selector with six positions: 100 V, 120 V, 140 V, 200 V, 220 V and 240 V.
- Mains transformer, Tl, including four separate primaries] a thermal switch, and eight (earlier)
or five (new) separate secondaries.
- Power Supply PCB including secondary fuses, two regulation circuits, and relays for
(dis)connecting the various PCBs (from) to the internal Ring Network.
The reason for the change is reduction of power loss (= heat emission).
Introduction-compatibility: The new parts which were introduced in ABL500/510/520 of Run
50/8/8 onwards are NOT compatible with the earlier.
Voltage Supplies
The Power Supply PCB provides two regulated voltages for the Main Computer PCB, 5.1 V and 12 V.
The two regulated voltages are derived from an unregulated 20 V supply obtained from a separate
secondary of the transformer. The voltages are generated using two integrated switching regulators
of type L296.
The Power Supply provides, in addition, a number of AC voltages which are applied to the various
PCBs where they are rectified and regulated locally.
5 . 1 VDC
This voltage is supplied to the Main Computer PCB where it is used for the digital circuits. It is, fur-
thermore, used for supplying the opto-couplers connected to the internal Ring Network on the Wet
Section Control PCB and the four optional PCBs.
The switching frequency of the regulator is approximately 50 kHz.
Soft start The regulator employs a soft start-up which slowly increases the maximum cur-
rent that may be drawn from the supply. This ensures that the regulator is not overloaded when
switching on the ABL.
Current /imitation If the current drawn from the 5.1 V supply exceeds 4 through 7 A, the regulator
will be shut off (the wide range is caused by the tolerances of the regulator). The slow blow fuse F2
through which both the load on the 5.1 V and the 12 V supplies flow is a 3.2 A fuse.
Reset The regulator provides a reset signal used on the Main Computer PCB. When
switching power on, the reset signal will go high for at least 250 ms after the output has increased to
5.0 V. The duration of the reset pulse is not critical as long as it is above 250 ms (it may be up to 750
ms). When switching power off, Reset will go low when the output has decreased to 4.75 V.
Protection If the output voltage increases to above 6.1 V, e.g. in case of a short circuit to
the 33 V supply, the SCR CR9 will be triggered shortening the 20 V DC to ground and thus blowing the
fuse.
The regulatorwill automatically shut off in case it becomes overheated.
valves, pumps, heating elements, etc. The fuse F4, which is a slow blow 4 A type, protects against
overloads. In case the + 33 V DC is too low, the message MAINS TOO LOW appears.
33 VAC
The last secondary of the transformer Tl provides a 33 V AC voltage for the printer (supplied to the
Printer Control PCB via the Main Computer PCB). On the Printer Control PCB itself, the AC voltage is
rectified and regulated. The fuse Fl, which is a slow blow 3200 mA type, protects against overloads.
If no PCBs are connected (other than the Main Computer PCB itself), all relays are deactivated, that is,
switched to the upper positions thus looping TxD back to RxD (both of the Main Computer PCB). The
various relays are controlled by the connected PCBs themselves as follows:
When the power is switched on, the connected PCB, say the Wet Section Control PCB, will perform a
selftest and initiation program. When this has been carried out successfully, that is, when the PCB has
started up and is running, it activates the appropriate relay (switches to the lower position), thus con-
necting itself to the Internal Ring Network. A CPU-Watchdog circujt found on each of the connected
PCBs, monitors if the CPU is working. If so, the PCB remains in the ring. However, if the CPU fails, the
CPU-Watchdog will automatically deactivate the relay, thus disconnecting the defective PCB from the
ring.
The above means that only well-functioning PCBs will be in the ring; this system prevents a fault on
one PCB from disabling all communication via the Internal Ring Network.
Compatibility of RSl50/100
The RSl50/100 have been changed for two reasons - to reduce power consumption and to change the
Interconnection in ring network.
Earlier PCBs are: 902-021 (RSISO) and 901-773 (RSllOO)
New PCBs are: 902-148 (RSl50) and 902-147 (RSllOO)
The new type PCBs are fully compatible with the earlier type whereas only one of the earlier type
PCBs can be connected in new instruments (because of missing relay). In case both an RS150 and an
RSllOO is to be connected in a new ABL and one of the PCBs is of the new type and the other is of the
earlier type they must be connected as follows: The new type PCB must be connected as Option 1 and
the earlier type as Option 2 (refer to figure 7.9.2).
The three external communication channels all origin from two DUARTs controlled by the CPU. Each
DUART includes four data lines (two TxD and two RxD lines) together with eight output and six input
lines. The output part of each channel is galvanically isolated from the rest of the RSilOO PCB by pas-
sing all signals through optocouplers and using separate power supplies.
The voltage is used for supplying the digital ICs which are not connected to the ABL and
the external equipment.
The following supplies are all derived using a DC/DC converter built-up around a tranformer with
three separate secondary windings - one for each of the RS-232-C channels. The individual supplies
are stabilized using zener diodes.
,+ 12 VI For RS-232-C driver for channel 1.
+ 5 Vl For RS-232-C driver and optocouplers for channel 1.
2 12 v2 For RS-232-C driver for channel 2.
+ 5 v2 For RS-232-C driver and optocouplers for channel 2.
f 1 2 v, For RS-232-C driver for channel 3.
+ 5 v3 For RS-232-C driver and optocouplers for channel 3.
Reset The + 5 V regulator provides a reset signal for the CPU. The reset is removed (goes from
low to high) approximately 80 ms after the + 5 V supply has reached its nominal value.
The reset from the + 5 V regulator is gated together with the reset from the CPU watch-
dog before it is applied to the reset terminal of the CPU and the DUARTs. The CPU can
then determine the source of a reset by checking the status of the WI-STATUS signal.
General Information
This section describes the wet section and gas mixer mechanics.
The wet section part includes a flow volume table, principle diagrams and the ABL51 O/520 optical system.
LS3 Liquid sensor, testing that liquid samples are filling the hemolyzer (ABL510/520 only)
or the right electrode chamber branch (ABL505 only) completely.
General information
1. Front part with liquid containers, inlet module, thermostatted electrode chamber,
valves, liquid sensors, tubings, and pumps.
2. Rear part with PC-board, stepping motors for the main valve and pumps, wiring,
and optical fibres for the liquid sensors.
The two parts are separated by a common mounting plate of painted aluminium.
The front part appears with an easy-to-clean, smooth surface. The inlet module is designed as a snap-
on, and completely water-proof unit.
The liquid containers are suspended electrically isolated from the rest of the wet section.
The front part of the wet section is covered by a transparent cover which is magnetically suspended
assuring an easy mounting/dismantling.
The entire wet section is suspended and fastened to the base unit by 4 guide pins and 9 screws ac-
cessible from the front.
When servicing the ABL, the entire wet section is easily dismountable by unscrewing the 9 front
screws.
Below is a short general description of the wet section components. For an overview please refer to
the wet section principle diagrams included in this section:
PI and P2 The wet section includes 2 pumps in order to reduce the measuring and calibration
time by allowing liquid and gas transport in the right and left tubing circuit simulta-
neously, e.g. P2 is rinsing the inlet while the ABL is measuring on the blood sample in
the electrode chamber.
VI ,V2 and V3 Tubing valves which control the liquid and gas flow in the electrode chambers and
the hemolyzer (V3 in ABLS05/510/520 only).
v4 Membrane valve for rinse solution intake to the right tubing circuit.
LSl Liquid sensor, testing that liquid samples are filling the left electrode chamber branch
completely.
LS2 Liquid sensor, testing for sufficient and homogenous liquid samples.
t I I I
I Nominal Flow,
&sec
100
I
Minimum Flow, Maximum Flow,
Fl/sec
90
I
&ec
120
I
60 54 72
36 32.3 43.2
22 19.4 25.9
1 13 11.6 15.5
I
l
I 6
I
5.63
I
7.5
I
8.1-4
WET SECTION MECHANICS
The following table shows the wet section flow volumes for ABL50/500/505/510/520. The figures in
(brackets) are for ABL510/520 only, and in [brackets] are for ABL505 only
I T-junction - V2
I
4.6
I
3.9
I
5.3
Monochromator
The Monochromator Unit converts the light transferred through the sample into a continuous spec-
trum as follows:
Due to the behavior of light, the beam transferred through the entrance slit of the Monochromator
Unit is cone-shaped. The beam is directed to the mirror unit which contains a concave mirror conver-
ting the cone-shaped beam into a parallel light beam.
The parallel light beam is reflected to the grating unit. The grating unit includes a plane grating that
converts the beam into a continous spectrum and again returns it to the mirror unit.
The spectrum is then reflected to the slit plate which, apart from the entrance slit, also includes six
output slits. These output slits are located so that they allow six monochromatic light beams, corre-
sponding to the six wavelengths in the continuous spectrum (535, 560,577,622,636, and 670 nm), to
pass through it. The monochromatic light beams are picked up by six photo diodes placed behind
each slit.
Due to low light intensity it is necessary to measure the photodiode current immediately before the
photo lamp is turned on (dark current) during each measurement and calibration and then when the
photo lamp is turned on. The dark current is substracted from the current measured with the photo
lamp on. The resulting current is then converted into absorbances for each of the six wavelengths.
CO2 Flow
Pure CO2 is obtained from the cylinder and is passed through a reduction valve before being applied
to the gas mixer itself. The output pressure of the reduction valve is 2.5 + 0.5 atmospheres. At the
input, the CO2 passes the COz input valve, which is only open when CO2 is needed. Then the CO2
passes the flow regulator which has two outputs, one (Gas 3 - pure CO2) in which the flow is
approximately 40 ml/min, and another in which the flow is approximately 8 ml/min the latter being
applied to the mixing coils. On the CO2 flow regulator a pressure transducer is mounted which
monitors the CO2 pressure; if the pressure is outside the nominal value ?r 20 %, an error message will
appear.
output
There is only one tubing going from the gas mixer to the flow selector in the wet section. Therefore,
one pass-through (on/off) valve for pure CO2 and two valves, each with two outputs for the gas
mixtures, are provided at the output (for selection of gasses). When the gas mixer is running, the
unused gas mixtures are exhausted via a separate outlet, whereas pure CO2 is only flowing when
actually needed.
Poiseuilles Law
The principle of operation of the gas mixer is based on a patented method using Poiseuilles law
which applies for gas flowing in a capillary tube. In this form Poiseuilles law states that:
Ap x d4
Q= X constant
PXl
8.2-2
0 8.2. GAS MIXER
Introduction
The Gas Mixer produces three gas mixtures on basis of pure CO2 (from a cylinder) and atmospheric air
from the surroundings (drawn in by means of a membrane pump). The gas mixtures are used for
calibration of the two gas electrodes, pC02 and ~02, and for flushing the inlet system. The gas
mixtures and their use are outlined below:
Gas 1: 5.61% CO2 and 19.76% 0~~45 ml/min;
used for the first point calibration of both gas electrodes and for flushing the inlet system,
Gas 2: 11.22% CO2 and 18.58% 02,45 ml/min;
used for the second point calibration of the pCOz electrode, and
Gas 3: 100% CO2 from cylinder, 40 ml/min;
used for zero-point calibration of the ~102 electrode.
The gas mixture percentages are as indicated above when the content of CO2 in atmospheric air is
0.04% and the relative humidity is 50% (variations in the mixtures caused by a change in the two
parameters are insignificant for the function of the ABL).
The gas mixer is only running, that is, producing gas mixtures, when there is a need for it. This is to
minimize the CO2 consumption. On Page 4 of this section, a flow diagram of the gas mixer is shown.
Air Flow
When running, the membrane pump will circulate atmospheric air at a flow rate of 180 ml/min. The
atmospheric air is flowing through a capillary tube (resistance) to create a pressure drop (which is
proportional to the flow). Across the capillary tube is a pressure transducer, which controls the speed
of the membrane pump to ensure a flow rate of 180 ml/min. If the air flow is outside 180 mUmin f 25
ml/min, an error message will appear.
The atmospheric air taken from the surroundings is passed through two air filters, which also
function as vibration dampers. Of the 180 ml/min, only approximately 82 ml/min are passed on to the
mixing coils; the remaining 98 ml/min (already filtered air) are recirculated. This means that only
approximately 82 mI/min is drawn from the surroundings when the gas mixer is running. Please see
below.
Differential
. pressure regulator
I I
I 4
98 ml/min
180 ml/min
pump flow
8.2-l
GAS MIXER
2 Resistance caused by curvature of the capillary must be negligible compared to the Poiseuille
resistance, and
3 The dynamic pressure loss at the capillarys intake and exhaust must be negligible compared to
the Poiseuille pressure drop.
Q
co2
x 100
02% = Qair + Q,,
2
That is:
1 Qair
1 0 0 x - = -+1
CO,% Q,,
2
and since the pressure drops (Ap) across the two mixing coils are the same, Poiseuilles law states:
1 pair x 5
100 x - = x constarrtl + 1
CO,% AP x 4
a That is:
Pco2 d; x I2
1
100 x - = - - x cXM.cq + 1
CO,% P,, x 4x1,
At a particular temperature, PCns/pair is constant, and within a certain temperature range, pCoalPair
only changes slightly, since the correction factor is 0.033 %/C. Therefore:
4X12
100 xc;z%
- = -
4 x 1, x constant2 + l
That is, l/COz% and therewith COz% is only dependent on the diameter and the length of the mixing
tubes.
8.2-3
9. SPARE PARTS LIST
0
Introduction
All spare parts available for the ABL50/500/505/51 O/520 are included in this list. Accessories available for
the end-user are listed in the Order Form, and have therefore not been included in the Spare Parts List.
Modules
Page 9.1-I gives a survey of the mechanical and electrical modules.
Defective modules marked M must be returned to the Radiometer Distribution A/S for replacement
under the Trade-In Arrangement with the exception of modules which have been damaged due to
attempted repairs. Repairs can only be made where special permission has been given by the Service
Department at Radiometer Medical A/S.
Defective modules marked C may be repaired by the Distributors personnel with the precautions
given for the modules in the Check-out and Adjustment procedures (Section 2), and using the
components given in this Spare Parts List only. C - modules may also be returned to Radiometer
Distribution A/S under the trade-in arrangement.
Component marking
A: Components only for the following series of instruments
ABL500 Run l-49
ABLSI O/520 Run l-7
Components
The components available for service are listed in section 9.2. All parts are manufactured by RADIOM-
ETER, and can only be purchased from RADIOMETER.
Ordering
When ordering spare parts, please include the following information:
e
REVISED 92.11 .Ol 9-l
9.1 MODULES
0
Computer Unit (ABL500/505/510/520)
A 901-782 Main Computer PCB M
B 902-l 46 Main Computer PCB M
A 901-783 Main Memory PCB, Without EPROMS
(See section 9.2 for EPROM kits) M
Wet Section
901-577 Monochromator Unit (ABL51 O/520 only) M
A 901-774 Wet Section Control PCB M
901-775 Monochromator Interface PCB (ABL51 O/520 only) M
902-I 57 Wet Section Control PCB M
Base Unit
A 90 l-773 Serial Interface PCB, RSllOO, Optional M
901-785 Gas Mixer Control PCB M
A 901-786 Power Supply PCB M
901-822 Gas Mixer, Complete C
A 902-021 Serial Output PCB, RSl50, Optional M
B 902-l 47 Data Interface PCB, RSllOO, Optional (not ABL50) M
B 902-148 Serial Output PCB, RSl50, Optional M
B 902-l 54 Power Supply PCB M
902-224 Main Computer PCB (ABL50) M
The following assemblies are not considered Spare Parts/ Modules, and cannot be returned under the
Trade-in Arrangement. The listed items must be repaired by the distributors service center.
Computer Unit
CRT Compartment
Printer Unit
Floppy Disk Unit
Numerical Keyboard
Base Unit
Wet Section Unit
Keypad & Display Unit (ABL50)
0
REVISED 93.02.08 9.2 - 1
COMPONENTS/MODULES
The tubing for the Gas Mixer must be cut in accordance with drawing 22520-A4, Section
10.1 page 3. The length shown in parenthesis is the total requirement for one Gas Mixer.
Base Unit
460-033 Line Voltage Selector
617-902 Cable for Software Gate (ABL500/505/51 O/520)
636-050 Cable with 3 plugs W9, Power Supply PCB to Rear Panel,
(ABL500/505/51 O/520)
a
A 636-051 Cable Flat with 6 plugs W6, Power Supply PCB to Ring Net
636-052 Cable Flat with 2 plugs W8, Gas Mixer to Wet Section Control PCB
B 636-l 57 Cable Flat, Power Supply to Wet Section Control
B 636-l 59 Cable Flat, Power Supply to Option Modules in Ring Net
636-301 Main Computer to Power Supply (ABL50)
A 770-728 Power Transformer
B 770-743 Power Transformer
803-208 Jack, Phone, Green
802-218 Line Cord Receptacle with Filter and Switch
855-l 30 Foot, plastic for Measuring Station
885-837 Panel, Left Side
885-842 Cabinet, Base Unit
885-919 Panel, Right Side
885-932 Panel, Blind Cover for Option Slots
888-270 Panel, Blind Cover for Option Slots
888-275 Panel Right side
888-297 Cabinet, Base Unit
0
REVISED 93.02.08 9.2 - 5
COMPONENTS/MODULES
Screws
009-325 Screw, Countersunk M3x25, For Wet Section D
009-904 Screw, Countersunk M4x16, for 901-818
009-906 Screw, Lens head M4x20, for Wet Section
009-910 Screw, Countersunk M4x16, for 902-l 39
011-410 Screw, Hex M4/10
020-327 Screw, Finger for 904-999
021-312 Screw, Selftapping, 3x8
021-313 Screw, Selftapping, 3x12
021-314 Screw, Selftapping, 3x20
021-315 Screw, Selftapping, 4x10
021-407 Screw, Self tapping, 2.9x9.5,CS
021-410 Screw, Self tapping, 3x8
021-411 Screw, for 886-352
025-035 Screw, for Fan
032-905 Nut, Ml 0x1, for 923-320
805-343 Nut, Locking for D25 Connector
Miscellaneous
450-030 Main Fuse, 1.6 A slow for 220 Volt Line Voltage (EU)
450-l 25 Main Fuse, 3.2 A slow for 120 Volt Line Voltage (US)
617-877 Cable WI, between Computer Unit and Base Unit
617-888 Cable for OSM3
617-889 Cable for ICA, KNA, RISl
617-890 Cable, Video for Ext. Monitor NEC.
617-891 Cable for Ticket Printer
836-i 21 Gasket for CO, Cylinder
Components
655-001 Optical Fibres with One Clip for Liquid Sensors
816-283 Clip for Optical fibre Transceiver end (10 PCS.)
834-575 Drop Catcher, for KCI Well
834-612 Main valve Gasket with Tracks
834-613 Support disc for Main Valve Gasket
834-620 Gasket for Front part of Main Valve (2 PCS.)
835-351 O-ring for LS3 and LS4 (5 PCS.)
836-l 47 Inlet Washer (2 PCS.)
837-035 Rubber Washer for Pump Housing
837-472 Soft Rubber Plate for Dual Membrane Valve, V4 and V5
840-027 Rubber Tube, EPDM 2X1 5
840-043 Rubber Tube, Black 0.9X1.3
840-044 Rubber Tube, Santoprene l.Oi4.0
840-204 Silicone Tube, 2.5X1.5
840-227 Silicone Tube, pump No.1 to Waste
841-727 Rubber tube, Main valve to Preheater 2 (2 pcs.)
841-738 Rubber tube, Preheater 2 to Electrode Chambers (2 pcs.)
841-750 Tubing for V2 (2~~s.)
841-757 Tubing for Vl
841-759 Tubing, Saline preheater - V4, (ABL500)
841-760 Tubing for V3 (ABL51 O/520)
841-761 Tubing for V3 (ABL505)
842-l 94 Tubing for Pump (5 PCS.)
846-l 46 Holder for Tubing
846-l 49 Holder for Tubing for Vl
848-017 Membrane Pump
901-754 Valve, dual, V4 and V5
901-788 Reference Well Assembly (V6, LS5)
901-789 Valve (Vl ,V2)
901-790 Liquid Sensor (LS3) with Rubber Washer and Ring
901-792 Electrode Chambers with LSl and LS2 (ABL500/51 O/520)
901-799 Filter, Air (2 PCS.)
902-023 Inlet Tube with LS4 (ABL500)
902-l 05 Inlet Tube with LS4 (ABLSI O/520)
902-l 65 Inlet Tube with LS4 (ABL505)
902-l 71 Electrode Chambers (ABL505)
923-582 Stand-Off with Connecting Tip for Bottles
923-604 Stand-Off, Holders for Bottles
Modules
901-775 Monochromator Interface PCB (ABL51 O/520) M
901-781 Printer Control PCB M
901-822 Gas Mixer, complete C
B 902-l 46 Main Computer PCB M
B 902-l 54 Power Supply PCB M
B 902-l 57 Wet Section Control PCB M
Components
039-800 Locking Tie Rod for Inlet Module
552-030 Numerical Keypad
552-027 Keypad for Softkeys
580-042 Motor, Stepping, for Pl and P2
580-161 Fan
617-877 Cable, Computer Unit - Base Unit
901-752 Main valve
901-758 Lamp Unit (ABL51 O/520)
901-759 Optical Fiber (ABL51 O/520)
901-760 Monochromator Adaptor (ABL51 O/520)
901-761 Hemolyzer Unit with LS3 (ABL51 O/520)
901-762 Inlet Module without Flap and Washer
901-763 Tie Rod for Inlet Module with Wire and Plug
901-768 Numerical Keyboard PCB
901-791 Preheater 2 ,(Main Valve - Electrode Chamber)
901-794 Thermistors for Electrode Chambers
901-795 Inlet Flap
901-811 Wet Section Distribution PCB
901-920 Preheater 1 without Inlet Tube (ABL500/510/520)
901-988 Pre-Amplifier Unit p0,
902-018 Gear Box 27:l (Pl and P2)
902-155 Waste Detector
902-181 Waste Stopper with Nipples
902-184 CRT Deflection PCB
902-185 Video Amplifier PCB
902-188 Pump Housing with Rotor
902-204 Pre-Amplifier Unit pH/pCO,
902-212 Preheater 1 without inlet Tube (ABL505)
910-199 Therm0 Printer
912-091 Heating Foil for Heated Panel, with Cable and Plug
REVISED92.01.15 9.3 - 1
SOFTWARE VERSIONS
Wet Section Control software version 5.45 (5.39, 5.40, 5.43 and 5.44 fully compatible)
Wet Section Control PCB EPROM (QD3) part No. 368-858
INTRODUCTION
This section lists the latest released software. For ABL500 below run 50 and ABL51 O/ABL520 below run
8 the main processor software, the operating system and the wet section program software are stored in
EPROMs on the main memory board.
For ABL500 run 50 onwards, ABL510/ABL520 run 8 onwards and ABL505 run 01 onwards, the main
processor software, the operating system and the wet section program software are stored in electrical
erasable FLASHPROMs on the main computer board.
The software is bilingual, i.e. all ABL5xx delivered from factory have English as their second language
(German/English, French/English etc.). In EPROM versions, languages other than English are defined in
main memory EPROMs QD5 and QD6 without which the ABL speaks English only.
QD3 368-970
QD4 368-994
QD5 Not mounted English only
368-995 German
368-951 French
368-953 Japanese
Spanish (not yet released)
368-893 Italian
QD6 Not mounted English
368-996 German
368-952 French
368-954 Japanese
Spanish (not yet released)
368-894 Italian
QD7 368-888
QD8 368-997
QD9 368-890
Wet Section Control software version 5.45 (5.40, 5.43 and 5.44 fully compatible)
Wet Section Control PCB EPROM (QD3) part No. 368-858
Memory Cards
Part No.: 914-001 English/Japanese (Vl .Ol J)
914-006 English/German (Vl .Ol G)
Keypad software version 1.09 (also compatible with Main software Vl .OO)
Keypad & Display PCB EPROM (QD) part No. 367-010
WS control software version 5.45 (also compatible with Main software Vl .OO)
Wet Section Control PCB EPROM (QD3) part No. 368-858
0
-
913-173 English/Italian
English/Spanish (not yet released)
Page
Mechanics 10.1
Electronics 10.2
Drawing Page
Gas Mixer Assembly 22520-A4 10.1-2
Wet Section Front View (ABL500) 22523-A4 10.1-4
Top of Waste Stopper and Front of Main Valve 22577-A4 10.1-6
Electrode Chamber and Main Valve with Inlet tube and
Preheaters, Right View 22589-A4 10.1-8
Wet Section Rear View (ABL500) 22595-A4 10.1-10
Main valve 22599-A4 10.1-12
Wet Section Front View (ABL51 O/520) 23674-A4 10.1-14
_.. ._-. _._ .___. --_-- _
Wet Section Rear View (ABL51 O/520) 23675-A4 io.i-16
Hemolyzer and Lamp Unit (ABL5101520) 2374 1-A4 10.1-18
Monochromator Adapter (ABL51 O/520) 23742-A41 10.1-20
Wet Section Front View (ABL505) 24395-A4 10.1-22
Wet Section Rear View (ABL505) 24396-A4 10.1-24
Tabel for Tubing in Wet Section 10.1-26
0 22
23
Tubing Santoprene
Tubing Santoprene (2 PCS)
L=
L=
24 mm
32 mm
840-101
840-101
Y..k_I__ ~__*-___-_ ,rn
24 I wing Banroprene Lz IP _-
11,111 0nn ,A,
0-+-1I
l 22
23
Valve Vl, V2
Drawing 22589-A4 Electrode chamber and Main valve with Inlet Tube and Prehea-
ters Right View
4 Valve V2 90 l-789
L
J Tb.~r...:..+~r
III~:II,II>LI 0..w+
TQI LIA-FTherm:r+rrrr 4-c
lllF111113LV13
I1
Electrode Chambers 90 l-794
7 Valve VI 901-789
20 l”lF f-r
ILlnln ,“I nnt;rd
“yrmL.cll fihvn
ll”l F ILad
c2
0 22
23
Hole for Optical fibre LS4
Hole for Optical fibre LS4
Revised 91.03.18
0 MECHANIC DRAWINGS
4 Valve V2 90 l-789
7 Valve VI 90 l-789
24 Valve V3 90 I-789
Drawing No. 23741-A4. Hemolyzer and Lamp Unit. right View (ABL51 O/520).
Screw holding Optical fibre
Clamp holding Optical fibre
Hemolyzer mounting clamp
Screw holding Lamp Unit
Adjusment screw flamp.
Adjustment screw for diaphragm (Factory set, do not adjust)
Cable Connectror
Photodiode
Retaining sprin for lamp
Halogen Lamp 400-007
Kuvette
12 Optical Fibre 901-759
3 Optical Fi bre
0
WRITTEN 92.02.14 10.1 - 23
MECHANIC DRAWINGS
0
0
WRITTEN 92.02.14 10.1 - 25
MECHANIC DRAWINGS
Specie1 tubings: Tubing Type 840-l 04 00.8/2.0 for older ABL500s with narrow grove in
thermostted plate ( see Service Note 917-282)
Lenght in millimeters:
28: Bottle No. 2 Nipple (6) LS5 280
29: Rinse preheater v4 205
0
WRITTEN 92.02.14 10.1 - 25
ELECTRONIC DRAWINGS
e
Page
Block Diagrams 10.2-l