Professional Documents
Culture Documents
Performa Technique Guide PDF
Performa Technique Guide PDF
Performa
Technique Guide
Film: Kodak Min R 2000, Screen: Kodak Min R 2000 and Min R 2190
Film: Fuji AD-M, Screen: Fuji AD Mammo Fine
Film: Agfa HDR and HDR-C, Screen: Agfa HD and HD-S
Compressed breast thickness Technique
kVp Filter Density
< 3 cm 23 Mo 0
4 - 5 cm 24 - 26 Mo 0
6 - 7 cm 27 - 29 Mo or Rh* 0
> 8 cm 30 - 32 Mo or Rh* 0
NOTE
Implant displaced views and specimens may be phototimed if sufficient breast tissue covers the
photocell closest to the chest wall.
Comments:
1 Maintain exposure times between 0.5 and 2.5 seconds. Decrease kVp if times are less than
0.5 seconds and increase if they are more than 2.5 seconds.
2 If exposure times are within indicated times, use + or - density settings to achieve optimal
film density.
3 Your technique may vary from those indicated above depending on film density preferences,
processor or film irregularity, or different film/screen combinations.
4 *Rhodium - use for very dense breasts, 6 cm or greater. No need to increase kV.
5 Magnification views - increase technique by 2 - 3 kVp, if exposure time exceeds 5 seconds.
37147-IMG rev 3
Performa Check code guide
Term “Performa” in this manual refers to Performa MGF-110 unit. If any other error code appears,
consult the error message section in the Operation Instructions (37094-IMG).
For more detailed information on the subject, please refer to World Headquarters
operation instructions of the unit, code 37094-IMG. GE Healthcare
P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A.
General Electric Company reserves the right to make changes
in specification and features shown herein, or discontinue the European Headquarters
product described at any time without notice or obligation. GE Healthcare - Europe:
Contact your GE representative for the most current Paris, France - Fax +33-1-30-70-94-35
information.
Asia Pacific
Copyright © 2000 by General Electric Company Inc. All rights GE Healthcare -Asia:
reserved. Tokyo, Japan - Fax +81-425-85-5490
GE Healthcare Finland Oy, a General Electric company, going Hong Kong - Fax: +852-2559-3588
to market as GE Healthcare.
www.gehealthcare.com
37147-IMG rev 3 10/2007