You are on page 1of 6

776028

research-article2018
DSTXXX10.1177/1932296818776028Journal of Diabetes Science and TechnologyPayne et al

Symposium/Special Issue

Journal of Diabetes Science and Technology

Capabilities of Next-Generation
1­–6
© 2018 Diabetes Technology Society
Reprints and permissions:
Patch Pump: Improved Precision, sagepub.com/journalsPermissions.nav
DOI: 10.1177/1932296818776028
https://doi.org/10.1177/1932296818776028

Instant Occlusion Detection, and journals.sagepub.com/home/dst

Dual-Hormone Therapy

Forrest W. Payne, PhD1, Bradley Ledden, PhD1,


and Greg Lamps, MS, MBA1

Abstract
Insulin pumps allow patients to attain better blood glucose control with more lifestyle flexibility. Their size and cost,
however, limit their usefulness. Current CSII pumps are bulky, intrusive, and expensive. SFC Fluidics is addressing
these problems by developing a new type of wearable patch pump based on the patented electro-chemiosmotic (ECO)
microfluidic pumping technology. This nonmechanical pumping technology allows accurate and precise delivery of very
small amounts of insulin and/or other drugs, including concentrated insulin. The pump engine is small and can be made
inexpensively from injection molded parts, allowing its use in a disposable or semidisposable pod format. In addition,
a single ECO pump engine can be used to deliver two drugs through independent pathways. Other features of SFC
Fluidics’ pod include latching safety valves that prevent accidental overdosing of insulin due to pressure changes and an
instantaneous occlusion sensor that can immediately detect delivery failure at the first missed dose. These features allow
for the development of a series of patch pumps that will offer users the benefit of CSII therapy in a more discreet and
reliable patch pump form.

Keywords
insulin patch pump, occlusion detection, dual-hormone therapy, automated insulin delivery, artificial pancreas, concentrated
insulin

There is significant evidence that people with diabetes who


use continuous subcutaneous insulin infusion (CSII) have
better clinical outcomes compared to those who use multiple
daily injections (MDI),1,2 however, only about 30% of people
suffering from type 1 diabetes use this therapy. SFC Fluidics
has developed new technology that will enable a new genera-
tion of much smaller insulin pumps. SFC is using electrical,
rather than mechanical pumping, to create a range of innova-
tive products to serve the needs of all insulin-using people
with diabetes. The core electro-chemiosmotic (ECO) pump-
ing technology delivers excellent precision, which coupled
with a valve system and flow sensor, deliver a unique safety
factor. Together these three technologies form a “pump
engine” (Figure 1) that will be utilized for a simple patch Figure 1.  ePump and valves.
pump, an insulin-only automated insulin delivery (AID) sys-
tem, and a dual-hormone artificial pancreas (AP) system.
1
SFC Fluidics, Inc, Fayetteville, AR, USA

Technical Details of ECO Pumping Corresponding Author:


Greg Lamps, MS, MBA, SFC Fluidics, Inc, 534 W Research Center Blvd,
Currently marketed insulin delivery pumps generally use a Ste 260, Fayetteville, AR 72701, USA.
mechanical motion to drive a piston or a shuttle through a Email: glamps@sfcfluidics.com
2 Journal of Diabetes Science and Technology 00(0)

Figure 2.  Schematic of ECO pump operation.

syringe to push the insulin into the patient. While this is a take place using established electrochemical techniques and
well understood and straightforward method, it does have simple electronics. The direction of the insulin flow is con-
drawbacks that limit its usefulness for delivering the small trolled using SFC’s latching safety valves, described later in
doses required by people with type 1 diabetes. The limita- this paper. A single stroke of this reciprocating displacement
tions of the current syringe pump technology include:3 pump can be infinitely varied to deliver any amount between 0
and 1 U of U100. Larger doses are delivered using multiple
•• precision limitations due to discrete motor movement strokes at a rate of 1U/min. This system is able to deliver nor-
•• inaccurate dispenses caused by stiction of the plunger mal, square-wave, dual-wave, or any other shape of bolus
•• strict design limitations due to the straight-line form dose. The 1U/stroke limit ensures that the pump isn’t damaged
requirements during operation. A schematic diagram showing the operation
•• engineering constraints that prevent further miniatur- of the ECO pump is shown in Figure 2.
ization of mechanical components This type of system has several advantages over a mechan-
ical system. First, there is no discrete stepping mechanism
The heart of SFC Fluidics patch pump system, the patented that limits the precision of doses that can be delivered. While
ePump® (an ECO pump), solves these problems. The ECO existing mechanical insulin patch pumps are very good at
pump is a nonmechanical pump that is highly accurate and delivering the small volumes required when using U100, they
readily scales with delivery requirements, allowing for a typically are limited to 0.05 U dose step sizes. For U500, this
more discrete product. same minimum step size becomes a dose of 0.25U for exist-
Fundamentally, the ECO pump is two electrochemical half- ing systems. Increments of 0.05 U are fine for adults, but
cells separated by a semipermeable membrane. Each half-cell smaller controlled dosing increments are needed for children
contains an electrode, is bound on one side by a flexible, and are essential for migration to the use of more concen-
impermeable diaphragm and is filled with a nontoxic proprie- trated insulins. Other advantages stem from the construction
tary solution that supports fully reversible aqueous electro- of the ECO pump itself. For example, the liquid working
chemical reactions. Initially, each half-cell contains the same material virtually eliminates engineering constraints on the
solution, and there is no movement in the pump. The applica- shape and size of the pump. Drugs that require smaller doses,
tion of a small voltage (≤1.5 VDC) between the two electrodes such as concentrated insulin, can be delivered using a smaller
drives chemical reactions which change the osmotic pressure pump. The design freedom in the shape of the pump allows
between the two chambers. When this happens, ions and asso- the pump to be tailored to fit into an organically shaped pod
ciated solvent move from one half-cell to the other through the with consideration of the other components. A third benefit of
semipermeable membrane to equilibrate the pressure. As the this construction is that the pump can be made almost entirely
fluid moves between chambers, the flexible diaphragms move of injection molded parts and is amenable to high-speed,
to accommodate the volume change in each cell. This dia- automated manufacture and assembly, permitting its use in a
phragm movement is used to pump insulin or other drug into discreet, disposable pod/patch pump type application.
and out of a metering chamber that is adjacent to the flexible Another not-so-obvious advantage of this type of recipro-
diaphragm. After the correct fluid volume has been pumped, cating pump with potentially huge significance for diabetic
the applied potential is reversed to drive the half-cells back to therapy is that the diaphragms of both chambers move in con-
equilibrium. The flow rate and total volume pumped are easily cert as fluid is moved across the semipermeable membrane. By
controlled by controlling the rate and number of reactions that adding a second set of the latching safety valves, two different
Payne et al 3

hormones can be independently dosed through separate fluidic


pathways using a single pump engine. Each side of a dual-hor-
mone ECO pump will have the same accuracy and precision as
a single hormone ECO pump. This means that two hormones,
for example insulin and glucagon, can be delivered as neces-
sary in an AP system to either lower or raise blood sugar in
response to a continuous glucose monitor (CGM) reading.
Dual-hormone therapy has been shown to significantly reduce
the occurrence of hypoglycemia in studies using two CSII
pumps—one filled with insulin and the other with glucagon.4-13
Although these studies have shown benefits to dual hormone
therapy, the mental, physical, and financial burden of maintain- Figure 3.  Rendition of insulin delivery pod. The pod as designed
ing two pump systems and a CGM is too much for many peo- is 51 mm × 55 mm × 13 mm.
ple to manage, even with the prospect of reduced hypoglycemic
events and better-long term outcomes. SFC Fluidics will take (nanoamperes) electrical current. This characteristic current
advantage of the dual-hormone delivery capacity of the ECO depends on fluid composition such as ionic strength and vis-
pump in meeting patient needs for comfort and discretion. The cosity. During pumping, the sensor confirms fluid flow by
dual-hormone pod will be larger than the insulin-only AID measuring an increase in current due to convection augmenting
device by only the size of the glucagon reservoir; dramatically movement of charge carriers over diffusion alone. In addition
reducing patient burden. SFC’s single-pod AP system will to being fast and accurate, SFC’s occlusion sensor is also quite
make dual-hormone therapy a practical treatment option for small with a sensing length of 1.5 mm and a diameter of 0.5
both children and adults with type 1 diabetes. mm. The sensors are fabricated by mass manufacturable meth-
The latching safety valves are designed, as the name sug- ods common in the electronics industry and can be made at a
gests, with safety of the patient in mind. These mechanical size and cost suitable for disposable use. The power require-
inlet and outlet valves are created as a set and interact with ments are very low as well, requiring a few microamperes dur-
each other during operation so that they can never both be ing fluid flow but with the ability to be switched off completely
open at the same time. During the switching of the valves, between dispenses. Finally, the processing power needed to
both valves must close momentarily before the other can confirm flow is low as well, and easily within the capabilities
open. Once switched, the open valve physically holds the of a pod control system. Rapid detection of dispense failure is
other closed so that there is never an open path from the drug key to diagnosing occlusions, whether from kinked cannulas or
reservoir to the patient. A failure of the pump or external pres- other possible faults. Accurate occlusion detection holds the
sure change such as during an airplane flight will not cause an promise of reducing time in hyperglycemia.
overdosing of the insulin to the patient. The outlet valve to the Figure 3 shows one potential design layout of the compo-
patient is always closed when the pump is not actively dis- nents that would be used for the Bluetooth®-enabled, 3-day
pensing insulin to the patient. When pumping is initiated, use, fully disposable insulin-only CSII pod.
both valves are closed and then the outlet valve is open. The
pump delivers the prescribed dose and then the valves close
again before the valve to the drug reservoir is open. The pump Performance Details
then returns to equilibrium by drawing the same amount of
ECO pump and valves
drug from the reservoir. In this way, failure of the pump can-
not cause an overdelivery of insulin to the patient. During the development of the ECO pump technology, sev-
The final piece of technology SFC has incorporated into the eral key aspects of the technology were proven before it
pod, adding yet another layer of safety and reliability to the could be considered as a technology capable of delivering
device, is the instantaneous occlusion sensor. A main failure life-saving drugs in a miniaturized embodiment. One of the
point in continuous insulin infusion is silent occlusions.14-16 early questions was how this technology would be affected
While this is primarily an infusion set/insertion issue, existing by any backpressure in the system. Many nonmechanical
occlusion sensors are unable to reliably detect the occlusion in pumps are not capable of pumping reliably against backpres-
a timely manner. Too often, occlusions are “silent” and detected sure. Electroosmotic pumps that use surface charge of a
when correction boluses are required to remedy a hyperglyce- porous frit to generate fluid movement are significantly
mic condition. A key benefit of SFC’s pod is the ability to con- affected by small changes in backpressure. Conversely, the
firm drug flow in real-time. Instantaneous confirmation of ECO pump has been shown to operate well against backpres-
individual basal dispenses may allow for detection of infusion sure and was able to generate enormous amounts of pressure
set faults or other occlusions within minutes. The flow confir- when pumping into a closed chamber. A well-fortified ECO
mation sensor sits in line with the drug delivery path between pump was used to generate a pressure of 23 atm (300 psi) in
the valve and cannula tip and consists of a series of electrodes. a closed chamber (see Figure 4).17 This capability assures
When a potential is applied, the fluid supports a very small reliable operation during subcutaneous delivery of drugs.
4 Journal of Diabetes Science and Technology 00(0)

Figure 6.  Trumpet curve for 48-hour basal testing of ECO


pump and valves. 48-hour total error is 0%.

Figure 4.  An ECO pump is used to generate 23 atm of pressure


in a closed chamber. The pump was operated at 20 mA reaction
current to generate the pressure. After a brief rest period, the
reactions were reversed at −20 mA to release the pressure.

Figure 7.  Shows the trumpet curves from 48-hour dual


hormone AP dosing schedule onto dual IEC 60601-2-24
compliant test setups. The trumpet curves for insulin (black dots)
and glucagon (green dashes) were calculated separately.

Once these technical characteristics were understood,


the ECO pump technology was optimized for use in a wear-
Figure 5.  Shows the efficiency of an ECO pump remains able drug delivery device for delivery of insulin to people
constant at temperatures ranging from 5-55°C while the flow rate who suffer from type 1 and type 2 diabetes. The flow rates
increases linearly with temperature.
and energy requirements could be met with a pump with
chambers that were only 12.5 mm in diameter and 3 mm
In addition, the accuracy of the pump had to be character- high. Once this pump was designed and built, IEC testing
ized over a wide range of potential operating environments. was begun.
Unlike a motor driven pump, there is no mechanical dis- The International Electrotechnical Commission (IEC) has
placement that can be measured to determine how far a shut- defined the standard for measuring and presenting the error
tle or plunger has moved. Instead of counting the number of of dispenses for ambulatory drug infusion pumps in their
steps of a motor, ECO pump calibration is based on the num- IEC 60601-2-24 document.20 In this procedure, a pump is
ber of reactions that take place inside the pump. While this is used to pump a dose of water onto a balance repeatedly, and
an easy measurement to make, there are many factors that the measured change in mass is used to calculate the volume
can affect the result. It was shown that different solutions and of liquid that was dispensed. Pump accuracy and precision
solvents and even different ionic concentrations affect the are clearly presented in a “Trumpet Curve.” This chart con-
amount of fluid that moves across the semipermeable mem- veniently displays the maximum amount that the pump over
brane per unit charge.18,19 It was also shown that for a given or under dispensed over any specified time interval during
solution, the pumping efficiency, measured as volume of the testing. As the time interval is increased, the values
fluid delivered per number of reactions, was constant over approach the overall average value, giving a trumpet shape.
the broad temperature range applicable to drug delivery The ECO pump and valves have been rigorously tested using
(5-55°C), though the pump works faster at higher tempera- an IEC 60601-2-24 compliant test setup. The results of these
tures (see Figure 5). The ECO pump can be tailored to work tests are shown in Figure 6. The figure shows the trumpet
at a certain efficiency and nominal flow rate while maintain- curve for a basal delivery routine of 0.5 U/h dispensed as
ing dosing accuracy over a broad temperature range. 0.05 U every 6 minutes over a total of 48 hours.
Payne et al 5

Figure 8.  Carrier solution for three rapid acting insulin analogs show similar response during pumping of 1U equivalent volume.

Figure 9.  Graph of sensor response at different flow rates relevant for SFC pod. Signal amplitude is indicative of pumping speed.

In a separate experiment, an ECO pump was coupled with mixed based on formulations available from the package insert
two sets of latching safety valves, and two IEC 60601-2-24 for the three-predominant rapid acting insulin analogs on the
compliant test apparatus were used to measure the ability of market. The graph for each solution shows four dispenses of 1
the ECO pump to independently deliver doses using both U (10 µL) given over a 5-minute time span. Apidra® carrier
sides of the pump. After initial verification that the system solution is shown in green while Humalog® and Novolog®
works, the dual-hormone pump prototype was used to repli- are depicted with red and blue respectively. Each increase in
cate the doses given to a patient in a published dual-hormone signal corresponds to a dispense of 1 U through the occlusion
AP trial study (Patient 303).21 One side of the pump dosed sensor at a flow rate of 20 µL/min for 30 seconds. Data were
the insulin protocol and the other delivered the required glu- taken for each formulation separately and overlaid to create
cagon. In the published dual-hormone study, an algorithm the figure. For each dispense there is a rapid increase in the
was used to calculate a dose every 5 minutes using readings current from the sensor when flow is initiated and a corre-
from a CGM (sometimes the dose was 0 mg). The dispenses sponding quick decrease when flow is stopped.
ranged from 0-20 µL (0-2 U) of insulin and 0-6 µL (0-30 µg) The current measured by the sensor during pumping is
for glucagon. The total error for insulin was −2.8% over the significantly higher than when the solution is at rest, with
48-hour period. For glucagon the dosing error was +1.8%. signal amplitude indicative of flow rate (Figure 9). The data
Each side of the two-sided ECO pump prototype demon- in Figure 9 show current response at various rates from
strated precision comparable to that of the single-sided ECO 5-20 µL/min, indicating the sensor can be used to identify
pump and existing insulin only pumps, as shown in the trum- partial occlusions as well. In this figure, dispenses were
pet curves in Figure 7. fixed to 30 seconds, so 10 µL/min and 5 µL/min dispenses
are equivalent to 0.5U and 0.25U, respectively. In case of a
partial occlusion the pump would dispense the correct
Occlusion Sensor Test Results amount, but the low flow rate would occur over a long-time
Commercially available rapid acting insulins give similar period. By measuring both the flow rate as well as duration
response from the occlusion sensor due to their similar fluid of flow, the sensor can confirm dosage dispensed. The
composition (Figure 8). The graphs in Figure 8 are of solutions occlusion sensor responds quickly to initiation of flow and
6 Journal of Diabetes Science and Technology 00(0)

can sense occlusion of volumes as small as 0.05 U of U100 6. Jacobs PG, El Youssef J, Castle JR, et al. Development of a
rapid acting insulin, allowing confirmation of individual fully automated closed loop artificial pancreas control system
basal dispenses. with dual pump delivery of insulin and glucagon. In: 33rd
Annual International Conference of the IEEE, EMBS. New
York, NY: IEEE;2011:397-400.
Conclusion 7. Jacobs PG, El Youssef J, Castle J, et al. Automated control
of an adaptive bi-hormonal, dual-sensor artificial pancreas and
SFC Fluidics’ microfluidic technologies represent a signifi- evaluation during inpatient studies. IEEE Trans Biomed Eng.
cant step forward in the development of CSII treatment of 2014;61(10):2569-2581.
diabetes, allowing for safe, accurate continuous subcutane- 8. Jacobs PG, El Youssef J, Reddy R, et al. Randomized trial of a
ous insulin injection in a reduced footprint for type 1 and dual-hormone artificial pancreas with dosing adjustment during
type 2 diabetes and greatly increasing the speed at which exercise compared with no adjustment and sensor-augmented
occlusions are detected. pump therapy. Diabetes Obes Metab. 2016;18(11):1110-1119.
9. van Bon AC, Luijf YM, Koebrugge R, Koops R, Hoekstra
Abbreviations JB, DeVries JH. Feasibility of a portable bihormonal closed-
loop system to control glucose excursions at home under
AID, automated insulin delivery; AP, artificial pancreas; CGM, con- free-living conditions for 48 hours. Diabetes Technol Ther.
tinuous glucose monitor; CSII, continuous subcutaneous insulin infu- 2014;16(3):131-136.
sion; ECO, electro-chemiosmotic; IEC, International Electrotechnical 10. Blauw H, van Bon AC, Koops R, DeVries JH. Performance
Commission; MDI, multiple daily injections; U100, 100 units/ml insu- and safety of an integrated bihormonal artificial pancreas
lin; U, insulin unit; VDC, direct current voltage. for fully automated glucose control at home. Diabetes Obes
Metab. 2016;18(7):671-677.
Declaration of Conflicting Interests 11. El-Khatib F, Jiang J, Damiano E. Adaptive closed-loop con-
The author(s) declared the following potential conflicts of interest trol provides blood-glucose regulation using dual subcutaneous
with respect to the research, authorship, and/or publication of this insulin and glucagon infusion in diabetic Swine. J Diabetes Sci
article: Forrest Payne, Bradley Ledden, and Greg Lamps are all full- Technol. 2007;1(2):181-192.
time employees of SFC Fluidics, Inc. 12. Russell SJ, El-Khatib FH, Sinha M, et al. Outpatient glycemic
control with a bionic pancreas in type 1 diabetes. N Engl J Med.
Funding 2014;371(4):313-325.
13. Russell SJ, Hillard MA, Balliro C, et al. Day and night glycae-
The author(s) disclosed receipt of the following financial support mic control with a bionic pancreas versus conventional insulin
for the research, authorship, and/or publication of this article: Some pump therapy in preadolescent children with type 1 diabetes: a
of this research was supported by the US Army Research Office and randomized crossover trial. Lancet Diabetes Endocrinol. 2016;
the US Army Research Laboratory (Contract DAAD19-03-1-0053). 4: 233-243.
This material is based on work supported by the National Science 14. Heinemann L, Krinelke L. Insulin infusion set: the Achilles
Foundation under Grant 0848253. Research reported in this publi- heel of continuous subcutaneous insulin infusion. J Diabetes
cation was supported by the NIDDK of the National Institutes of Sci Technol. 2012;6(4):954-964.
Health under Award R43DK110972. 15. Gibney M, Xue Z, Swinney M, Bialonczyk D, Hirsch L.

Reduced silent occlusions with a novel catheter infusion set
References (BD FlowSmart): results from two open-label comparative
1. Majedah M. A., Mousa M, Al-Mahdi M, Al-Sanaa H, Dalia studies. Diabetes Technol Ther. 2016;18(3):136-143.
A-A, Al-Kandari H. A comparison of continuous subcutane- 16. Klonoff DC, Freckmann G, Heinemann L. Insulin pump occlu-
ous insulin infusion vs. multiple daily insulin injection in chil- sions: for patients who have been around the (infusion) block.
dren with type I diabetes in Kuwait: glycemic control, insulin J Diab Sci Tach. 2017;11(3):451-454.
requirement, and BMI. Oman Med J. 2015;30(5):336-343. 17. Evans C, Noble R, Koval C. A nonmechanical, membrane-
2. Misso ML, Egberts KJ, Page M, O’Connor D, Shaw J. based liquid pressurization system. Ind Eng Chem Res.
Continuous subcutaneous insulin infusion (CSII) versus mul- 2006;45(1):472-475.
tiple insulin injections for type 1 diabetes mellitus. Cochrane 18. Norman M, Evans C, Fuoco A, Noble R, Koval C.

Database Syst Rev. 2010(1):CD005103. Characterization of a membrane-based, electrochemically
3. Ferrari R, Breech DR. Infusion pumps: guidelines and pitfalls. driven pumping system using aqueous electrolyte solutions.
Aust Prescr. 1995;18:49-51. Anal Chem. 2005;77(19):6374-6380.
4. Haidar A, Lagault L, Dallaire M, et al. Glucose-responsive 19. Jeerage K, Novble R, Koval C. Investigation of an aqueous
insulin and glucagon delivery (dual-hormone artificial pan- lithium iodide/triiodide electrolyte for dual-chamber electro-
creas) in adults with type 1 diabetes: a randomized crossover chemical actuators. Sens Act B. 2007;125(1):180-188.
controlled trial. CMAJ. 2013;185(4):297-305. 20. IEC 60601-2-24:1998. Medical electrical equipment—Part
5. Haidar A, Legault L, Messier V, Mitre TM, Leroux C, Rabasa- 2-24: Particular requirements for the safety of infusion pumps
Lhoret R. Comparison of dual-hormone artificial pancreas, and controllers. International Electrotechnical Commission;
single-hormone artificial pancreas, and conventional insulin 1998.
pump therapy for glycaemic control in patients with type 1 21. Russell S, El-Khatib F, Nathan D, Magyar K, Jiang J, Damiano
diabetes: an open-label randomised controlled crossover trial. E. Blood glucose control in type 1 diabetes with a bihormonal
Lancet Diabetes Endocrinol. 2015;3(1):17-26. bionic endocrine pancreas. Diabetes Care. 2012;35:2148-2155.

You might also like