86.0367.00 Rev B 1
EUROPEAN TO NORTH AMERICAN PLUG AND SOCKET COMPARISON 2
Earth Ground
(Green Insulation with Yellow Stripe)
SAFETY CHECKS ON A-DEC 200, 1040, 1221, 8000 SYSTEMS
Test 1 (limit > 1 MΩ)
L
24 V~
0 V~
N
PE
L
24 V~
0 V~
N
PE
L
L
24 V~
24 V~
0 V~
0 V~
N
N
PE
PE
Ω Ω
L
24 V~
L
V~ 24 V~
0 V~ V~
N
0 V~
N
PE
PE
Safety Checks on A-dec 200, 1040, 1221, 8000 Systems
86.0367.00 Rev B 5
SAFETY CHECKS ON A-DEC 500 SYSTEMS 6
L
24 V~
0 V~
N
PE
L
24 V~
0 V~
N
PE
Ω
Safety Checks on A-dec 500 Systems
L
24 V~ L
24 V~
0 V~
N 0 V~
PE N
PE
Ω
Ω
L
L 24 V~
24 V~ V~
V~ 0 V~
N
0 V~
N PE
PE
86.0367.00 Rev B 7
Safety Checks on A-dec 500 Systems 8
L
24 V~
0 V~
N
PE
L
24 V~
0 V~
N
PE
86.0367.00 Rev B 9
Safety Checks on A-dec 300 and 400 Systems 10
L
L
24 V~
24 V~
0 V~
N 0 V~
PE N
PE
Ω Ω
L
L 24 V~
24 V~ V~
V~ 0 V~
N
0 V~
N PE
PE
Safety Checks on A-dec 300 and 400 Systems
86.0367.00 Rev B 11
System Safety Checklist
Model: 12
Serial Number:
Date Test 1 Test 2 Test 3 Test 4 Test 5 Test 6 Test 7 Test 8
Limit > 1 MΩ Limit > 1 MΩ Limit < 5Ω Limit > 1 MΩ Limit ≈ 26 V~ with no load Limit = Mains ± 10% Limit < 0.5 mA) Limit < 0.1 mA AC,
< 0.01 mA DC
If a system safety check indicates that there may be a problem with the equipment, contact A-dec Customer Support.
13