EP0186 - Thermo-desinfecting machines – lack of efficacy for the removal of C.

difficile spores
Rafael Perdiz¹, RN; Jorge Sampaio², MD, Phd; Mirian Dalben Corradi¹, MD, Msc; Andrea Alfaya Acuna¹, RN; Renata Desordi Lobo¹, RN, Phd; Maria Beatriz Dias1, MD, Phd.
1 - Hospital Sírio Libanês; 2 - Fleury Diagnostic Medicine; E-mail: jorge.sampaio@grupofleury.com.br; biagsd@gmail.com

Background: C. difficile infection (CDI) is a leading Before the cycles a positive control was obtained from the spore suspension. In each load a Media with and without thaurocholate agreed in
healthcare associated infection, with about half a sterile urinal was included to serve as a negative control and to check for potential contamination 71/81 (87.6%) culture pairs.
million cases in USA and a high prevalence in North during the disinfecting cycle. Cultures were collected with Rodac® plates with C. difficile
hemisphere hospitals. Sírio-Libanês Hospital is a moxalactam norfloxacin agar (Thermo) with and without sodium thaurocholate. Thaurocholate was Conclusions:
private 400 bed tertiary hospital, loicated in São added in order to improve spore germination. Swabs were utilized to sample hard to reach areas. - Contrary to previously published data(1,2),
Paulo, Brazil. The CDI incidence is 8.8 CDI/10,000 After each cycle in both machines, a point close to the draining area was sampled with a Rodac our results show that neither thermo-
patient days, high by Brazillian benchmarks. We plate. Plates were incubated in anaerobiosis for three days at 36˚C. Species identification was disinfecting machine nor the different
take many measures to avoid C. difficile transmission confirmed by MALDI-ToF (Bruker). detergent solutions showed efficacy in
between patients, including sterilization of bedpans removing the C. difficile spores from
and male urinals, which constitute a high workload contaminated bedpans and urinals.
for the central sterilization service. Therefore we - Sterile urinals placed in the same cycle
evaluated the efficacy of two automatic thermo- came out contaminated with C. difficile.
disinfecting machines, which would be placed in the - After a disinfecting cycle C. difficile spores
wards, in order to diminish this central service were recovered from the thermo-desinfecting
demand. machines except when Liquezime® was
used in the Clinox 3A Auto Arcania®.
Material/methods: The two machines had similar Figure – Wokflow used in this study. - According to our data the use of these
technical characteristics: Arcania® Clinox 3A Auto (5 automated thermo-disinfecting machines
Results: Results are detailed in Table.
minute cycle at 90oC max temp) and Getinge® FD was not effective in removing C. difficile
Table – Performance of thermo-disinfecting machines on C. difficile spores removal. spores and may contribute to foster its
1600 (8 minute cycle at 91oC max temp). A fecal
suspension was sterilized in autoclave and then Detergent solution Liquezime® Sekumatic® Prolistica® ArjoHuntleigh® dissemination in hospitals.
contaminated with spores from a clinical strain of C.
difficile in other to obtain a final concentration of 107 Thermo-disinfecting machine Clinox 3A Clinox 3A Clinox 3A
FD 1600
/ml. The fecal suspension with spores was utilized to Auto Auto Auto References
Getinge®
contaminate bedpans and urinals. After Variables Arcania® Arcania® Arcania® 1.Alfa MJ, Olson N, Buelow-Smith L. Simulated-use testing
of bedpan and urinal washer disinfectors: Evaluation of
contamination, devices were let at ambient Positive controls (spore suspension) 1/1(100%) 1/1(100%) 1/1(100%) 3/3 (100%) Clostrium difficile spore survival and cleaning efficacy. Am J
temperature for 24 hours and then submitted to Infect Control, 2008;36:5-11.
disinfection in the machines (Figure). Contaminated devices tested (N) 15 15 15 9 2.Alfa MJ, Olson N, Buelow-Smith L, Murray B. Alkaline
The detergents utilized were: Devices with C. difficile growth after detergent combined with a routine ward bedpan washer
100% 93.3% 93.3% 88.9% disinfector cycle eradicates Clostridium difficile spores from
- Sekumatic® (acid pH 3), Liquezime® (multi- disinfection cycle (%)
the surface of plastic bedpans. Am J Infect Control, 2013:
enzymatic pH 9.5) and Prolistica® (alkaline pH > 13) Negative controls (sterile urinals) 41:381-383.
3 cycles each detergent, with the Arcania® machine; which grew C. difficile after 3/3 (100%) 1/ 2 (50%) 3/3 (100%) 3/3 (100%)
- ArjoHuntleigh® (alkaline pH 14) - 3 cycles, with disinfection cycle (%) Financing sources: Hospital Sírio Libanês and
Machine contamination Fleury Diagnostic Medicine. Machines and
the Getinge® machine.
(culture of drainage area with C. 0/3 (0%) 3/3 (100%) 2/3(66.6%) 3/3 (100%) detergents were provided by the thermo-desinfecting
difficile after the disinfecting cycle) machines manufacturers