H30/H50

Regular and Preventive Maintenance

Revision History:

Version Date Created By Details


V.1.0 Initial Version

1
1. Automatic Bleach Cleaning Maintenance
Go to “SERVICE””MAINTENANCE”, on the screen, the users can determine the
frequency of performing bleach cleaning cycle by setting the following items:
 Number of Cycles (ranging from 50-300 and the default value is 300)
 Bleach Frequency (Days) (ranging from 1-30 and the default value is 7)
No matter which one is reached, the system will pop out Screen to prompt users to
perform bleach cleaning cycle when you press Shutdown to turn off analyzer.

Note:
1) Shutdown: we need to perform Shutdown every day and to leave at least 3
hours the instrument under Shutdown status every 24 hours.
2) When the machine is giving the message about performing bleach cycle, user
must perform this maintenance cycle.

2. Manual Bleach Cleaning Maintenance


Go to “SERVICE””TROUBLESHOOTING””BLEACH CLEANING”

2
3. How to Perform Bleach Cleaning
Following the pop out messages to perform bleach cleaning procedure step by step.

Adding 4mL of bleach (Sodium Hypo chloride solution) into RBC bath and WBC bath
respectively, then click on
Note:
1) When the user open the right chamber door to perform the bleach procedure,
kindly check whether there is any leakage from the valves, syringe, counting
chamber. If there is leakage, contact the service engineer as soon as possible.
2) If there is any crystal or blood stain on the chamber protection shield, try to
clean it.

3
4. Regular Maintenance
The following maintenance table is recommended to user when:
 50 samples are tested per day.
 6x52 working days per year
If the number of tested samples increases, the frequency of maintenance needs to
be increased proportionally.

5. Cleaning and Disinfecting the Analyzer


Clean and disinfect the exterior surfaces to remove dust, splatters, etc.
Policies regarding the cleaning and disinfecting intervals are at the discretion of your
individual institution.
 Turn off the analyzer.
 Disconnect the power cord and the power adaptor.
 Disconnect the connecting cables if the system is connected to other pieces of
equipment.
 Dampen a lint-free cloth with detergents or disinfectants.
NOTE:
The cloth should be wet but not dripping.
The validated detergents for cleaning the system are:
 Ethanol (75%)
 Isopropanol (70%)
The validated disinfectant for cleaning the system is:
 HB300 Bleach (produced by Edan)
 Clean the exterior surfaces with the wet cloth.
 Clean the surfaces with the wet cloth to disinfect them after the surfaces are
completely dry.
NOTE:
Clean the surfaces prior to disinfecting them. Make sure the surfaces are
thoroughly dry before disinfecting them.
 When the surfaces are thoroughly dry, reconnect the power cord, power
adaptor and other connecting cables.

4
Contact Information
Edan Instruments, Inc. EDAN Diagnostics, Inc.
Monday to Friday 09:00-17:00 (UTC Monday to Friday 09:00-17:00 (UTC -08:00)
+08:00) Service: +1 858 750 3066
Service: +86-755-26898321 E-mail: Services@edandiagnostics.com
E-mail: info@edan.com.cn
EDAN GmbH
Monday to Friday 09:00-17:00 (UTC
+01:00)
Tel: +49 (0) 6103 202 0781
Email: service.eu@edan.com.cn