TROUBLESHOOTING
PAGE 1 OF 7
BLANKETROL II
TROUBLESHOOTING
PAGE 2 OF 7
BLANKETROL II
TROUBLESHOOTING
PAGE 3 OF 7
BLANKETROL II
TROUBLESHOOTING
P. The unit is operating in Low line voltage or Check line voltage. Reset
Manual or Automatic power source. Setpoint display and resume
mode. When cycling to operations.
Cool mode, all displays
flicker on and off, the unit
goes to CK SET PT in the
Status display or
REMOVE FROM
SERVICE SENSOR or HI
LIMIT.
Q. Unit is operating in The back-up High Limit
Automatic or Manual Safety device is triggered
mode, the Status display which shuts down the unit
flashes HI LIMIT and when the circulating water
REMOVE FROM and reservoir reach 46°C
SERVICE. +2°C (115°F + 4°F)
because:
PAGE 4 OF 7
BLANKETROL II
TROUBLESHOOTING
PAGE 5 OF 7
BLANKETROL II
TROUBLESHOOTING
W. Unit is set to operate in Operator error; switches not Correctly set switches.
Automatic or Manual set correctly.
mode; the water flow
indicator does not move,
i.e.: water not circulating
Disengaged quick Check all couplings for proper
disconnect couplings fit.
Kink in connecting hose Straighten out the hose.
and/or blanket
PAGE 6 OF 7
BLANKETROL II
TROUBLESHOOTING
or
PAGE 7 OF 7