This enclosure provides the installation instructions for software ARC_2012 and
higher versions.
► Referenced documents
# Document Reference
None
CONFIDENTIALITY NOTE:
Use, dissemination, distribution or reproduction of this document by unauthorized personnel is not permitted and may be unlawful.
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
01-2014 Printed in Germany Document ID: 34181094
Enclosure_e_template_v07_modified
Agfa Company Confidential Copyright © 2014 Agfa HealthCare N.V.
Enclosure
► Manufacturer
Agfa HealthCare N.V.
Published by
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstraße 161
D - 81539 München
Germany
Copyright 2014 Agfa HealthCare N.V.
All rights reserved.
Technical modifications reserved.
Agfa and the Agfa rhombus are trademarks of Agfa-Gevaert N.V., Belgium, or its
affiliates.
All other trademarks mentioned in this document are held by Agfa HealthCare N.V. or
the respective owners and are used in an editorial fashion with no intention of
infringement.
Nothing contained in this legal notice nor in any text in this document shall be
construed as granting by implication, estoppel or otherwise, any license or right to use
any of the trademarks, service marks, trade names or logos appearing in this document
without the express prior written consent of their respective owner.
WARNING:
Improper operation or service activities may cause damage or injuries.
INSTRUCTION:
(1) Read the "Generic Safety Directions" prior to attempting any operation, repair or
maintenance task on the equipment.
Refer to Document ID 11849633, Agfa Intranet / Agfa Portal via Internet.
(2) Strictly observe all safety directions within the "Generic Safety Directions" and on
the product.
IMPORTANT:
The installation and service of the product(s) described herein is to be performed
by qualified personnel who are employed by Agfa HealthCare N.V or one of its affiliates
or who are otherwise authorized by Agfa HealthCare N.V. or one of its affiliates to
provide such services.
LIST OF CONTENTS
1 INTRODUCTION ......................................................................................................................4
2 SOFTWARE INSTALLATION VIA USB FLASH DRIVE ...........................................................5
2.1 Checking current Software Version ..........................................................................................5
Subject Software installation instructions for Software ARC_2012 and higher versions
SCOPE OF DELIVERY:
ZIP File ARC_2012*
* or higher version
REQUIRED TOOLS:
Service PC with USB port
REQUIRED TIME:
Approximately 15 minutes for the installation via USB flash drive
NOTE:
Register a software upgrade in the local service management system as modification,
e.g. in SMS as “MOD”.
1 Introduction
NOTE:
The software installation instructions provided here can be used for software upgrade
or software downgrade.
The software upgrade via USB flash drive is performed in following steps:
# Step Section
1 Check current software version. 2.1
2 Uninstall the old Error Viewer. 2.2
3 Prepare the USB flash drive. 2.3
4 Perform the software upgrade. 2.4
5 Install the new Error Viewer. 2.5
6 Create a backup. 2.6
7 Verify the software upgrade. 2.7
Figure 1
Figure 2
Figure 4
5151_ENC_001d.cdr
Figure 5
(3) Extract the downloaded zip file into a folder on the Service PC.
Contents of the ZIP file:
Software package ARC_2012.zip (or higher version)
This document
Depending on the software version a readme.txt
Figure 6
IMPORTANT:
Do not switch off the Digitizer during software installation as the PMI and Power Board
are re-programmed during the software installation procedure.
Figure 7
NOTE:
The Digitizer LED activity sequence during software installation is as follows:
Slow Red blinking: Approx. 30 seconds
Off: Approx. 10 seconds
Slow red blinking: Approx. 40 seconds
Fast red blinking or green*: After approx. 80 seconds
* Digitizer LED only gets green if the right Error Viewer is installed. If the Digitizer software does not detect
the Error Viewer or detects a version that does not fit, the LED is fast red blinking.
User Name:
Figure 8
NOTE:
By keeping the default settings of the Error Viewer installation wizard it is
guaranteed, that the Error Viewer access rights for the different Windows users
(administrators / users) are set properly. See Figure 10 and Figure 11.
Between 1 to 15 seconds after installation of the Error Viewer the Digitizer status
LED turns green.
Figure 10 Figure 11
Figure 12
NOTE:
Software upgrade via browser can be performed from any PC in the network.
It is tested with Firefox and Internet Explorer.
Figure 13
(9) Browse for the software ZIP file ARC_2012.zip (or higher version).
(10) Select: Upload and Install Software
(11) Wait until the Digitizer reboots automatically and the Digitizer LED turns fast red
blinking or green*. This takes approximately 1,5 minutes.
(12) Install the Error Viewer. See section 2.5.
(13) Create a backup. See section 2.6.
(14) Verify the software upgrade. See section 2.7.
* for details see Note on page 8.
NOTE:
Software installation incl. formatting of the flash memory on the PMI board is required
only, if a Digitizer error description instructs to do so.
It can be used to perform a normal software update or upgrade, too.
(1) On the NX workstation: Uninstall the Error Viewer. See section 2.2.
Note, that on other workstations (e.g. SE 2.0 or higher) it is not required to
uninstall the Error Viewer.
(2) Extract the downloaded zip file into a folder on the Service PC.
Contents of the ZIP file:
Software package ARC_2012.zip (or higher version)
This document
Depending on the software version a readme.txt
(3) Extract all files from ARC_2012.zip*.
Folder ARC_2012* is created automatically.
* or higher version
Figure 14
(7) Release the USB flash drive from the Service PC.
(8) Perform the software installation. See section 2.4.
(9) Install the Error Viewer. See section 2.5.
Note that this step may not be performed if the Digitizer is connected to SE.
See NOTE below.
NOTE:
If the Digitizer is connected to an SE suite workstation, download following file from the
Agfa HealthCare Library: SE Image Management > SE Image Management >
SE Suite 2.0 > Software > SE_2.0_SW_Digitizer_CR10_CR12_Installation.ZIP.
This ZIP file contains a file Prepare_Configuration_for_SE.cmd.
Instead of installing the Error Viewer proceed as follows:
Copy file Prepare_Configuration_for_SE.cmd to directory ARC_2012* on the
Digitizer USB flash drive.
Double-click file Prepare_Configuration_for_SE.cmd
Boot the Digitizer with the USB flash drive: Now the Digitizer is prepared for the
Error Protocol which is implemented in SE 2.0.
* or higher version
NOTE:
After software installation including formatting the Digitizer LED will usually not get
green. Reason: The destination information is reset to default. The up-to-date
destination information is only available after restoring the latest backup.
Figure 15
IMPORTANT:
BAC_CR10-5151_<Subtype>_<DigitizerSN>_<DDMMMYYYY>_<HH-MM-SS>.zip
Or
BAC_CR12-5151_<Subtype>_<DigitizerSN>_<DDMMMYYYY>_<HH-MM-SS>.zip
5 Appendix
Figure 16