You are on page 1of 1

22 ERROR DISPLAYS

22 ERROR DISPLAYS
Error code Description Countermeasures
When area calculating, there are less Confirm the file data and calculate
3 points than 3 points coordinate data in again.
required
selected file.

CALC Calculation is impossible from the Confirm the input data.


ERROR data input.

DELETE When deleting coordinate data, it can not Confirm the data and delete again.
ERROR be done successful.
Displayed when you measure a prism of short Measure the distance by using a target
SIGNAL distance with non-prism mode, or when the without using a prism, or change the mode
OVER
telescope received too much light. from non-prism mode to prism mode.
Displayed when REM measurement Operate in the range out of ± 6° from the
E35 carried out to the range from zenith or zenith or nadir.
nadir ±6°.
Any abnormality occurs with EDM (distance Repair is required.
E60’s
measuring system).
Displayed when vertical angle 0 position is Confirm the procedure and readjust.
E71
set with incorrect procedure.
Displayed when Vertical angle -position is Repair is required.
E72
adjusted in wrong position.
The instrument was not leveled when Vertical Level the instrument then carry the
E73
angle 0-position is adjusted. adjustment work.
Mainly at the time data transmission Confirm operation procedure is correct or the
E80’s between GPT-3000 series and external connection cables are correct.
instrument.
Abnormality in internal memory Repair is required.
E90’s
system.

FILE The same file name exists. Use another file name.
EXISTS

FULL When making a file, 30 files already exist. If necessary, send or delete files.
FILES

FAILED Initializing can not be done successful. Confirm initializing data and try to
INITIALIZE initialize again.

LIMIT Limit of input data exceeded. Input again.


OVER

MEMORY Any abnormality occurs with internal Initialize the internal memory.
ERROR memory.

MEMORY Shortage of capacity of the internal Download data from internal memory
POOR memory. to PC.
The data is not found in the search Confirm the data and search again.
NO DATA
mode
NO FILE There is no file in internal memory. If necessary, make files.

FILE NOT When using a file, no file is selected. Confirm the file and select a file.
SELECTED

P1-P2 When in point to line measurement, the The horizontal distance between first
distance horizontal distance between first point and point and second point must be more
too short second point is within 1m. than 1m.

22-1