★ Before using this system, be sure to thoroughly read this manual and familiarize
yourself with the system.
★ After reading this manual, keep it near the system, in a handy location where you
can refer to it easily.
Tokyo, Japan
Copyright © Hitachi Medical Corporation. 2009-2014. All rights reserved.
Q2E-ES1155-15
About the ultrasound scanner display
Filing
Media Handling 24
Formatting 25
Recording and Loading Images 26
Image Search and Play 27
Copy to media and Edit 28
※
option
(i) Q2E-ES1155
A table of contents
Contrast Echo ※
Startup and Exit 29
Adjustment 30
MTI・TIC 31
Advanced Measurement ※
IMT 44
%WT 45
Color Tissue Tracking 47
LA Tracking 49
※
option
( ii ) Q2E-ES1155
A table of contents
4D (Real-time 3D) ※
Preparation 51
Scanning, Volume operation 52
Function 53
Volume Data Save and Load 54
HI DEF 3D (Scanning) 55
HI DEF 3D (Function) 56
HI DEF 3D (Multi Slice mode) 57
HI DEF 3D (CFM 3D mode) 58
STIC function ※ 59
3D (Freehand 3D) ※
Preparation 60
Scanning 61
※
option
( ⅲ ) Q2E-ES1155
Before starting examination
Turning the Power ON or OFF
Breaker switch
Start-up Examination
screen ready screen
③ Cancel the FREEZE key to start examination.
1 Q2E-ES1155
Before starting examination
Panel Key Description
『PROBE』 Key
Displays the probe switching dialog 『Dual Doppler Mode』 Key
box. Activates Dual Doppler.
『Patient』 Key
『HI Zoom』 Key
Displays dialog boxes for entering or
Turns high-definition zoom on or off.
modifying patient information.
『PSS』 encoder
Switches between the image 『PAN Zoom』 Key
adjustment parameter settings (data Turns PAN zoom on or off.
registered from the PSS setup menu).
『Function1』・『Function2』 Key
Activates the preset option functions or
displays a list of options in the form of
screen buttons.
2 Q2E-ES1155
Before starting examination
Examination screen (Screen buttons)
- Use the Function keys on the operation panel. Scrolls through functions when the
button is pressed.
Rotation
Executes the selected function
when pressed.
Push
The actual buttons displayed vary depending on the examination mode.
Function keys
Multi encoder
Performs operations related to the
Encoder Func. button selected
from the screen button area.
3 Q2E-ES1155
Before starting examination
Selecting the Examination Conditions
1 2
The default application functionality allows you to assign frequently used application settings for each probe.
For details on how to assign application settings for each probe, see the instruction manual for the device.
You can add a new setting to the applications registered for each probe of the equipment, or edit the
registered application settings.
For further details, see the instruction manual.
4 Q2E-ES1155
Before starting examination
ID Input
Use the trackball to move the pointer on the window, and then press the
ENTER key to select the next item to be entered.
Enter patient information, and then click the Save button to save
the data. The next time you don’t have to enter ID, you can
select from the list of IDs displayed in the lower part of the
window.
OB/GYN
5 Q2E-ES1155
Examination of each mode
B Mode ①
Brighter Higher
frequency
Deeper
Focus settings
To display the focus-related screen buttons, press the Focus Set. button.
Focus Set.
To go back to the general menu, press the Focus Set. button again.
6 Q2E-ES1155
Examination of each mode
B Mode ②
Display settings
To display the display setting menu, press the Display Set. screen button.
Display Set. To go back to the general menu, press the Display Set. button again.
To invert the displayed image horizontally, To invert the displayed image vertically,
press the Invert L/R button. press the Invert U/D button.
Shallower
To change the depth, activate the Up/Down Shift To steer the image, activate the Image Steering
button, and then use the trackball. button, and then use the trackball.
# #
Image Size Angle
7 Q2E-ES1155
Examination of each mode
B Mode ③
Steering HI Com
dTHI-W-R: High resolution
(Standard) mode HI Com HI Com is on.
To change the compound direction, use the
Down dTHI-W-P: High resolution
(Deep scan) mode
OBLIQUE toggle. Changes can be made at a
maximum of 3 levels in each direction.
Left Steering Right
dTHI-R: Standard mode OFF
dTHI-P: Deep scan mode Level 3 Level 2 Level 1 Level 1 Level 2 Level 3
*Different by probes.
※ Level 3 can be selected only during a linear probe, with the
emphasized effect of a puncturing needle.
Oblique B Mode image
(for linear probe only)
HI Com HI Com is off. Trapezoid display (for linear probe only)
Trapezoid
B Mode
Image
GUIDE
8 Q2E-ES1155
Examination of each mode
B Mode ④
Enlarge
If you press the ENTER key again, or do not touch Use the trackball to change the display position of
the trackball for over five seconds, the mark the image.
disappears, and the ROI can be moved again. Press the PAN Zoom key again to cancel the zoom
mode.
Press the UPDATE (NEXT) key
to the zoom mode.
.
Press the HI Zoom key again to
cancel the zoom mode.
Real-time
Cine memory images can be played back by frame
Real-time Frozen
with a trackball. Frozen
Frozen
Frozen Real-time
Real-time
9 Q2E-ES1155
Examination of each mode
M Mode
Darker
To play the M Mode cine memory images frame by
frame, freeze the display, and then turn the M ODM ODM
Mode encoder.
(If this causes the gain to change, press the center
of the encoder and then turn it.)
Moving the M Mode cursor Rotating and moving the ODM cursor line
Use the trackball to move the cursor right or left. Turn the Angle encoder so that the cursor line rotates
around the { mark.
Press the center of the encoder to turn the cursor line
45 degrees at a time.
Use the trackball to move the cursor line position.
To switch between Single Directional ODM and Dual Directional ODM, press the Dual ODM button.
The ODM image indicated by
the ODM Symbols mark .
Dual ODM
During Dual ODM you can use screen buttons to switch between the two cursor line control options.
Independent Both cursors have their own rotation center, and therefore move separately.
Independent Both cursors share the same rotation center, and therefore move together.
10 Q2E-ES1155
Examination of each mode
Color Mode
Color Doppler (CFI Mode) display Tissue Doppler image (TDI Mode) display
Press the center of the CFI Mode encoder. Press the Elasto/TDI Mode encoder to switch TDI
To exit CFI Mode, press the CFI Mode encoder again. Mode ON or OFF.
Brighter Turn the encoder to adjust the
gain.
* Depending on the probe or
Turn the encoder to adjust the gain. application being executed,
this might cause the Elasto to
be activated. For details, see
Darker the Instruction Manual.
Switching between the Power Doppler Changing the position and size of the
(CFA Mode) and Fine Flow modes color box
Press the Fine Flow/CFA Mode key.
To exit Fine Flow mode, press the Fine Flow/CFA
Mode key again.
Fine Flow mode is selected. CFA is selected. To change the size of the ROI, press the ENTER
(High-definition blood flow key. A dot (z) appears at the lower right corner of
power display) the ROI. Use the trackball to change the size of the
ROI.
11 Q2E-ES1155
Examination of each mode
Doppler Mode ①
Less Narrower
PRF
Lower volume
Remove lower frequency Select the Sound Volume button, and then use the
Multi encoder to adjust the volume.
Simult.
blood flow.
Simult. Counterclockwise
ON Real-time
Every time the encoder is pushed, the display in
Real-time registered angle, reversing, and erasing of the angle
bar are repeated.
12 Q2E-ES1155
Examination of each mode
Doppler Mode ②
If only the Doppler cursor is displayed with no waveform shown, press the UPDATE (NEXT) key.
Doppler image
corresponding to
Doppler cursor 2
Once frozen, you can also change the Doppler cursor by
pressing the screen button marked Sel.Dop..
PW PW TDI
PW TDI TDI
13 Q2E-ES1155
Real-time Tissue Elastography
Display option
ROI adjustment
① Make sure that the mark is selected and illuminated in yellow in the trackball priority section in the
lower right of the examination window.
If not selected, press the TRACKBALL key to select the trackball.
② Use the trackball to move the ROI or change its size. To switch between the two functions, press the
ENTER key.
The ROI size is fixed. Use the trackball to move If the z mark appears in the lower right corner of
the ROI to the desired location. the ROI, you can change the size of the ROI. Use
the trackball to adjust the size of the ROI.
The ROI can be adjusted not only in real time image but also in frozen image.
Press the ROI ReConfig. button, and
then set the ROI as described in steps 1
ROI ReConfig. and 2 above. Pressing the ROI ReSet
button returns the ROI to its state
immediately after the freeze. Canceling
Frozen the freeze state causes the ROI to
Use the trackball to
select the frame. return to the location before the freeze.
To make other adjustments, press the Image Menu key, and then use the displayed screen buttons.
Prev.Page Next
E-Dyn. MapL MapR Frame Rata Frame Rej. Noize Rej. Persistence Density
14 Q2E-ES1155
Real-time Tissue Elastography
Strain Graph option
The strain graph displays the variation over time of tissue strain caused by compression in real time.
Pressing the tissue which you want to image so that the periods and amplitudes of the Strain Graph are
constant makes it possible to create a stable Elasto Mode image.
Pressing probe (applying pressure to tissue) → Positive value
Withdrawing probe (releasing pressure from tissue) → Negative value
* Stronger pressure increases the amplitude. Gentler pressure reduces the amplitude.
Sector mode
Select Strain Graph Sector
Press the Sector button and use the encoder
to set the number of sectors.
If the frame is changed with the
# trackball, the sector box operates
Sector
synchronizing with the time phase bar.
15 Q2E-ES1155
Real-time Tissue Elastography
Strain Ratio, FLR, Strain Histogram Measurement option
Strain Ratio
① Select a Str. Ratio from Measurement Menu.
② Move the cross cursor displayed in the center of the screen to
the desired starting point of ROI-A, and then press the ENTER
key.
③ Move the next displayed cross cursor to the desired end point
of the ROI-A, and then press the ENTER key to draw the ROI.
The measurement results are displayed when ROI-A is set.
④ A cross cursor appears outside ROI-A. Repeat steps 2 and 3
above to set ROI-B.
FLR * In Dual Elasto mode, ROI-A and ROI-B can be
specified in either the left or right (up or down)
① Select a FLR from Measurement Menu.
window.
② Move the cursor to the center of a lesion using the trackball, and
then press the ENTER key.
Lesion and Fat are detected automatically.
16 Q2E-ES1155
ECG and Physiologic Module ① option
ECG
Down Up
* You can
To adjust the amplitude of the waveform, press change the
the ECG Gain button, and then turn the Multi position even
encoder. after a freeze.
Fast
Slow
ECG Invert
Screen buttons
ECG display Adjust ECG Adjust ECG ECG Invert ECG Filter
ON/OFF wave amplitude Position
R-R Detection R-R unit Display R-R time ECG Auto gain
ON/OFF detect filter
* R-R Detection is a function of * S,M,L means display
Adjustment of RRp/RRpp ratio. Tachycardia/Brad
automatically analysis interval ycardia cut filter size of ECG wave.
time of R to R.
17 Q2E-ES1155
ECG and Physiologic Module ② option
Details
In B mode, when the time bar reaches the detected interval, the details of the preceding R-R interval and
second preceding R-R interval are displayed.
(In M Mode or Doppler Mode, the details are displayed when the detected interval is shown on the screen.)
Time phase bar Number of the detected sector of Total detected sectors in the
the time phase bar position Ratio
cine memory
Broader search
conditions Display R-R time
R-R Interval(ms) button is display R-R time under the
ECG.
Narrower search
conditions
* The conditions cannot be changed during measurement or
during real-time Doppler measurement.
18 Q2E-ES1155
BodyMark
When using dual-window mode, you can use the L or R key to select the window in
which the body mark (probe mark) is to be set.
Move
① Make sure the mark is selected in the trackball priority section in the lower part of the examination
window. If not selected, press the PROBE MARK encoder to select this mark.
② Use the trackball to move the probe mark to the desired location.
Push the screen button of BodyMark In the ball of a finger. It is necessary to enlarge the contact area.
You can also select a body mark by pressing the PROBE MARK encoder.
You can scroll through the pre-registered body marks by pressing the
encoder.
(Example)
Clear
Screen buttons
Mark Move : Allows you to move the display position of the body mark by using the trackball.
19 Q2E-ES1155
Comments and Annotations
Comments
Activation and input
Press the Char. key from the keyboard to display a cursor. Use
the trackball to move the cursor to the target location, and then cursor
enter the desired text. ABCD
Press the Char. key again to finish entering the comment.
Annotations
Press the either ANNOTATION key on the operation panel or Annot. key on the keyboard.
From the displayed list, select the target text, use the trackball to move it to the desired location, and then
press the ENTER key.
Press the either ANNOTATION key or Annot. key again to finish entering the annotation.
Liver I Liver
or or
Liver
or
* This will also cause the measurements to be deleted.
20 Q2E-ES1155
Basic Measurement ①
The basic measurement methods are caliper measurement and trace measurement.
Other special measurements are available by either pressing the MEAS. MENU key
or by selecting one of the screen buttons.
Caliper measurement
The measurement to be executed
B Mode : Distance
M Mode : Depth difference, time, and slope
D Mode : Speed (pressure gradient) and time
Measurement result
Measurement procedure
① Press the CALIP key to display the caliper mark .
② Press the ENTER key to set the start point, and then move the next caliper mark to the end point.
Press the UNDO key to set the end point and go back to moving the start point again.
③ Press the ENTER key to set the caliper. After a caliper mark has been set, a new caliper mark appears so
that the next measurement can be specified.
In Doppler Mode, press the CALIP key to switch between speed
measurement and time measurement. (Measurement mode can Caliper;Time Caliper;Velocity
also be changed by using the screen button.)
Trace measurement
Measurement that is executed
B Mode : Area
M Mode : Distance, circumference, and area
D Mode : Average flow velocity
Measurement result
Deleting measurements
Press the CLEAR key to delete the measurements displayed in the examination window.
*This will also cause the comments (annotations) to be deleted.
21 Q2E-ES1155
Basic Measurement ②
Measurement result
Press the TRACE key while In the case of Abdomen Press the CLEAR key
observing the Doppler waveform, application, measurement to end the processing.
trace processing starts up, the items are Vp, Vd, Vm, PI,
measurement results is updating RI ,VTI.
in real-time.
Threshold : Adjusts the threshold so that the Dir. : Available trace direction options include up,
trace line fits the waveform. down, and up/down.
Smooth Line : Performs smoothing processing Average. : Display measurement result by averaging
in the direction of temporal axis. heartbeats.
FlowVolume measurements
① Press MEAS. MENU key and start FlowVolume from the list that appears.
② Tracing starts automatically. If the trace line is correct, freeze the image. The
most recent single heartbeat is used as the section to measure. Move the
section to measure as required.
③ Use the UPDATE (NEXT) key to finalize the detection waveform.
④ Measure blood vessel diameter using the CSA measurement caliper that
appears next. FlowVolume is displayed.
Moving section to measure and cine memory for PW waveform
Turn the PW Mode encoder. The cine memory moves with the section to measure.
22 Q2E-ES1155
Basic Measurement ③
Continues FlowVolume measurement.
Angle measurement
Angle 3 point measurement
23 Q2E-ES1155
Filing
Media Handling
Supported Media
Supported media: DVD-RAM, DVD-R*, USB memory devices, and USB hard disk (USB-HDD).
The use of reliable DVD-RAM media is recommended.
(For details on the model names of the recommended media, see the Instruction Manual HI VISION Preirus
~Preparation~.)
* Note that no additional data can be written to a DVD-R disk.
Disk ejection
Press the EJECT key.
Manual
Edit Patient
Mode Adjust
① From the tab menu on the left side of the screen,
Media select Tool and then Remove Hardware.
Remove Har・・・
Tool
Filing
Cine Memory * The Stop a Hardware device window is displayed.
Display
Account Menu ③ Click the OK button.
Annotation
Physiology
Video/ORIG
* Hardware can be removed.
PinP ④ Click the Close button.
24 Q2E-ES1155
Filing
Formatting
Formatting media
① Select an image in the filing window, and then click Copy.
The menu shown in the figure to the right is displayed.
② Click d under Destination, select the media, and then
select Format. A message is displayed.
③ Click the Yes button if you agree with the message. Select the media to
be formatted.
④ Follow the following message to start formatting.
【Format window】
【Format window】 Click the Start button.
Click the Start button.
Click OK butt0n if you agree with the message, and then execute formatting.
25 Q2E-ES1155
Filing
Recording and Loading Images
Prolonged multi image recording * In prolonged multi image recording mode, you
Press the assigned REC# key. can record multi images even during the freeze.
* The image can be recorded either directly to the
Stopping prolonged multi image recording DVD-RAM or to the device's HDD.
* The assigned key might vary depending on the
Press the assigned REC# key again to stop settings.
recording.
26 Q2E-ES1155
Filing
Image Search and Play
Searching images
① Press the READ key during the freeze to display Example: Searching by Study Term
the Filing window. ① Click ▼ in the Search field and select
Study Term.
Playing images
Selecting and playing images
① Using the trackball and the ENTER key, select the
thumbnail representing the image you want to play.
* Press the ENTER key to select the
② Click the Retrieve button to play the image. image. A check mark is displayed
on the selected image.
Filing
Filing
Playing and stopping multi images Switching between the Retrieve window and
(AVI/MPEG2) Examination window
Use the screen button to play or stop the Press either the READ or FREEZE key to go back to
multi image. the examination window.
■
Play Stop
27 Q2E-ES1155
Filing
Copy to media and Edit
28 Q2E-ES1155
Contrast Echo
Startup and Exit option
Selecting an application
For SonazoidⓇ, Hitachi recommends selecting an application whose name starts with Contrast-Mid.
For SonoVueⓇ, Hitachi recommends selecting an application whose name starts with Contrast-Low.
Measurement Menu;Contrast
If the above menu is not displayed, select the Contrast button to display the menu.
Contrast Contrast
* Make the screen button OFF to turn OFF contrast harmonic mode.
Contrast harmonic OFF Contrast harmonic ON
Body Mark Contrast T.Start Alternate I.Flash MTI To FI User Define * Display of screen buttons would change.
M
M
Contents of buttons differ according to the
Measurement UCI Flash Frame Rate Menu;Contrast
situation.
29 Q2E-ES1155
Contrast Echo
Adjustment option
Alternate Alternate
You can independently adjust MI and gain for the left (Contrast image) and the right (Monitor image).
Contrast Fundamental
Contrast image Monitor image image image
(Monitor image)
Reduce MI
dCHI-W-P: Priority given to depth sensitivity
T.Clear T.Start
Press the T.Clear(Timer Clear) button to
reset the counter to 00:00.
30 Q2E-ES1155
Contrast Echo
MTI・TIC option
MTI improves the visibility of the effects of a contrast agent and the visibility of where blood vessels
architecture by displaying, per frame and through superposition, the shadowed parts of a contrast image.
TIC is a function for drawing, based on recorded an AVI file, a time brightness level curve that shows the
relationship between contrast brightness level and time.
It is helpful in observing the changes in brightness level over time.
TIC
③ Select an ROI shape from the following and place an ROI where you want to watch the changes in
brightness level .
ダミー
31 Q2E-ES1155
Real-time Virtual Sonography
Preparation ① option
Real-time Virtual Sonography (RVS) allows the real-time creation of cross-section (virtual) images that match the
ultrasonic image reflecting the position and inclination of the probe. Cross-section images are retrieved out of the
volume data (CT/MR/US) that has been captured beforehand.
Virtual images are intended only for reference purposes. No puncture procedure or diagnosis must be determined
based solely on the observation of the virtual image.
Raised line
Orientation marks
Raised line
Attachment L73S
(For C715) Orientation mark of C715 L74M
the magnetic sensor C514 C532 L65 L75
When attaching the magnetic sensor, make sure that the orientation mark of the sensor
is aligned with the raised line of the attachment.
Connect the attachment to the probe, with its raised line aligned with the orientation
mark of the probe.
In EUP-L65, hang hook of the attachment on raised line of the probe and fix.
Attachment installation of L65
EUP-B512, EUP-B514
B512 B514
Attach the orientation mark of the magnetic sensor facing towards the outer side of the probe.
* The color and shape of the sensor is different depending on the specification.
32 Q2E-ES1155
Real-time Virtual Sonography
Preparation ② option
Attachment
B712 B715 C511
Orientation mark of
the magnetic sensor
U533
Orientation mark of
the magnetic sensor
Slide the
attachment.
Attachment
O732T
Attach the orientation mark of the magnetic sensor facing towards the inner side of the probe.
* The color and shape of the sensor is different depending on the specification.
Positioning
Transmitter
RVS Startup
① Make sure that the RVS sensor unit and that the lamp is
illuminated in green.
② Press the key ( F1 or F2 ) to which the RVS is assigned.
Press the F2 key to which list is allocated, and then use the screen button.
④ ⑤ ⑥
⑥ Assign volume data from the RVS Setting window, set a probe point and press the Start button.
The RVS mode window opens and the RVS top menu screen buttons are displayed.
BodyMark Create US Volume US Capture Volume;# Icon Menu Virtual Freeze US Freeze User Define
Measurement To Contrast RVS Setting Registration Reset Settings 1st Point Connection Menu;RVS
RVS top menu
* You can also use the Close button in the Image Library window and the RVS Setting window to change
the RVS mode with the RVS top menu screen buttons.
RVS Exit
RVS can be terminated by pressing the same key ( F1 or F2 ) that has been used to start up RVS.
Just as with the startup process, if this does not directly cause the RVS to shut down, press the screen button
marked RVS to terminate RVS.
34 Q2E-ES1155
Real-time Virtual Sonography
Loading volume data option
Saving volume data to and deleting volume data from the internal hard disk
【Image Library window】 【Import window】
RVS Setting
① Press the Import button of the Image Library window to open the Import window.
② Select the folder in which the volume data is stored from the Import window.
③ Click the Import button, the volume data is stored the hard disk.
④ When the required folders have been imported, press the Close button to return to the Image Library
window. The imported data is added to the list.
Saving volume data from DICOM server (Q/R Viewer option required)
① Press the Q/R button in 【Image Library window】 【Q/R Viewer window】
the Image Library
window to open the Q/R ①
Viewer window. ②
② Select the target DICOM
data from the Q/R Viewer
③ ④
window.
③ Press the Save button to store the data on the internal hard disk.
④ When data has been saved, press the X button in the top right corner
to close the Q/R Viewer window and return to the Image Library
window. The saved data is added to the list.
Deleting volume data
Delete no longer needed volume data from the internal hard disk as necessary.
Perform deleting and backup operations by pressing the Image Library Manager button at the bottom of the
Image Library window.
* For details refer to the separate instruction manual entitled "Real-time Virtual Sonography Software EZU-RV8 Position Sensor Unit
(Magnetic Sensor Unit for Real-time Virtual Sonography) EZU-RV3S."
35 Q2E-ES1155
Real-time Virtual Sonography
Volume data adjustment option
① Set the misalignment volume to “Async”. ③ Adjust the vertical and horizontal directions of the
② Adjust the depth direction of the volume. volume.
Place the pointer over the image, press the Set the adjustment point +A at the same locations
ENTER key, and then roll the trackball up and of each volume. The point appears whenever
down to switch to different cross-sections in the clicking the Adjustment point button.
* If the volumes are “Sync”, set the adjustment points at only
depth direction. Press the ENTER key again to one of the volumes.
fix the position. ④ Click the Adjust button.
+A +A
+A
+A
Image adjustment
Click a target icon from the trackball priority column. Move the pointer over the image and push the ENTER
key. Then, the adjustment with the trackball becomes possible. Press the ENTER key again to fix.
* To make adjustment by volume, sever the links between volumes.
Slice by slice advance Movement of image
Outward Arbitrary direction
Inward
Adjusting the image display size Adjusting the Window width and level
Reduce Window level ↓
Window Window
width ↑ width ↓
36 Q2E-ES1155
Real-time Virtual Sonography
Corresponding to the SYNAPSE 3D data option
The SYNAPSE 3D analysis image can be fused with the CT base image etc.
Fusion Reading
① Like normal volume data, import the SYNAPSE 3D analysis image (DICOM image) to the system hard disk
as well, and select the image from the list in the Image Library window.
② Double click the thumbnail display of the RVS Setting window to read the normal volume data.
③ Drag and drop the thumbnail of the Thumbnail
SYNAPSE 3D analysis result onto the
data that you wish to fuse with in the
volume data already read in ②.
④ The Parts color setting window is displayed. Make
adjustments where necessary and then press the OK button.
⑤ The Fusion registration window is displayed. Make
adjustments where necessary and then close the screen
with the Registration button and Close button.
⇒Refer to the various screen explanations below for the
adjustments in ④ and ⑤.
Parts color setting window Fusion registration window
Adjust the color of each part Adjust the relationship between the position of the
and the transparency level to base image and the position of the fusion color image
be used in the MPR image from the Fusion registration window.
display from the Parts color
setting window.
Base image Color image Fusion image
Select the button to open the display area display area display area
color setting dialog box.
The Parts color ① Align the display positions in the depth direction of
setting window can the base image and color image.
also be opened from Change the slicing by moving the trackball
the icon in the Fusion up and down in the respective image display
registration window.
areas.
Press the OK button to transit ② Set the adjustment point at the same
to the Fusion registration position of the base image and color image.
window. Or move the image positions so that both
images are aligned.
③ Press the Registration button and then the Close
button.
Parts Selection
【Parts select window】
Select the parts for the fusion from
the Parts select window.
Select the fusion items from the list
with the ENTER key and tick off the
items.
Up to 8 types of parts can be
selected.
37 Q2E-ES1155
Real-time Virtual Sonography
3D body mark option
3D body mark
From the RVS Settings window, click the 3D Object Setting button to bring up the 3D Object Setting
window.
Overview of how to create a 3D body mark
Select the volume to make the body mark from the
pull-down menu.
To create a body mark, several parameters must be
set from both Body Surface and Internal Body tabs.
Internal Body Superimposed
Ex.) Body Surface
Image
+ =
Select the icon of your choice from the dialog (left fig.)
Move the pointer over the image and press the ENTER
key.
It is necessary to display the pop-up menu to return the area cut by the cut icon.
Place the pointer over the image and press the UNDO key to display it.
38 Q2E-ES1155
Real-time Virtual Sonography
Marking function option
Function to set a marker to the region where users would like to mark (e.g. tumor on the volume data.)
3 types of markers (Sphere, Line, and Target). Sphere marker and Line marker disappear when going
out of the marking region. On the other hand, Target marker is displayed bigger in same situation, and
users can tell how much the marker is away from the target from its extent.
Marker settings
Marker setting from RVS settings window
Pick out one center section of the area (ex. Mass) that tries to be marked.
Move the pointer over the image and push the ENTER key to pick out the cross section. Then, it becomes
possible with trackball.
Icon menu Select the type of marker from icon menu and follow the same procedure as the above
mentioned one.
39 Q2E-ES1155
Real-time Virtual Sonography
RVS mode start option
Initial alignment
According to the probe point setting, the volume data taken beforehand is correlated with the body
position of the patient during real-time ultrasound examination.
There are Plane Start, Point Start, and 3 Point Start in the method.
For example, ideal reference points are the xiphoid process for the abdomen, nipple for the breast,
or internal urethral orifice for the prostate.
Plane Start and Point Start
① Pick out the cross section of the Xiphoid process.
Confirm the Paging button has been selected, and
move the trackball to adjust the depth position.
Xiphoid process
Xiphoid process
* If multiple series have been loaded, align their display
positions with one another and keep them linked.
RVS menu Measurement To Contrast RVS Setting Registration Reset Settings 1st Point Plane Start Menu;RVS
Menu;Contrast
Volume;# Click the Volume button to scroll through images starting from 1.
40 Q2E-ES1155
Real-time Virtual Sonography
Adjust option
Function to adjust the section when a gap appears between the virtual image and US image in scanning.
To make adjustments, you can use either Plane Adjust or, for positioning using internal or external
marks, use Point Adjust.
Plane Adjust
Adjustment by Virtual Freeze Adjustments of virtual image cross-sectional
and display positions can be performed
① Press the Virtual Freeze button
Virtual Freeze through probe scanning and also by using the
to freeze the virtual image.
trackball.
② Take US image which matches the frozen virtual
image. Select the type of movement from icon menu,
and make adjustment by trackball.
③ Press the Plane Adjust button to
release frozen image, and Plane Adjust
Inward and outward
synchronize the virtual image and
US image.
Adjustment by US freeze Vertically and horizontally
① Press the US Freeze button to
US Freeze
freeze US image. When either of above icons is selected,
② Move the probe to display the virtual image which rotation of virtual image is possible.
matches the frozen US image.
Virtual image rotates around
③ Press the Plane Adjust button to Z axis by M Mode encoder
release frozen image, and Plane Adjust
synchronize the virtual image and * This operation can be performed not only in virtual
US image. freeze image, but also in real-time image.
Point Adjust
Simple Adjust
Place a point mark on the same section of a virtual
image and US image and press the ENTER key
respectively to lock it in place (in any order).
Press the UNDO key before companion to the point mark
is locked, the adjust point can be canceled.
* The point mark is displayed as a pointer outside of the examination
window. When a pointer moved inside the examination window does
not turn into a point mark, select the 1st Point screen button.
To return to the initial location state, press the Registration Reset screen button. Registration Reset
To reset the probe points, press the RVS Setting. RVS Setting
41 Q2E-ES1155
Real-time Virtual Sonography
Performing when RVS mode is executed option
Registration Function
This function allows you to save and load misalignment data.
Store: Press this icon to save the current misalignment data (sync
data) of the virtual image and US image.
US Capture
① Press the US Capture button. The probe moves and scans at constant speed.
Capturing operation can be canceled by pressing the Cancel button, which is
displayed during the capture.
② The data captured up until the US Capture button is pressed the second time
will be loaded.
③ After reconstruction, the virtual image of the US volume data is displayed
US Volume automatically.
US image
Virtual image
In case of replacing the current volume
data by the US volume data to be captured
newly, delete the previous data from pop
up menu and perform the US Capture.
① Select Save Volume icon from icon menu, and display the dialogue.
*The Save Volume icon has the same function as the Save button in the RVS Setting window.
42 Q2E-ES1155
Real-time Virtual Sonography
Combination of RVS and Contrast mode option
To use RVS in combination with Contrast, execute the probe application for Contrast (for example,
Contrast-Mid-Abd) before starting RVS.
Screen buttons
The following screen buttons are available for RVS and Contrast, respectively:
RVS menu BodyMark Create US Volume US Capture Volume;# Icon Menu Virtual Freeze US Freeze User Define
Measurement To Contrast RVS Setting Registration Reset Settings 1st Point Plane Start Menu;RVS
Contrast menu BodyMark Contrast T.Start Alternate I.Flash MTI To FI User Define
Switching between the RVS mode window and Contrast mode window
During alternate mode (monitor mode)
To switch between the RVS mode and Contrast mode windows, use the To Virtual/To Contrast buttons.
The image selected during Contrast mode is displayed as the US image (right side) in the RVS mode window.
During non-alternate mode (non-monitor mode)
The To Virtual/To Contrast buttons are disabled.
Disabling alternate mode (monitor mode) from the Contrast mode window causes the Contrast image to be
displayed in single-window state.
To return to the RVS mode window, press the DUAL/SINGLE key.
Alternate
43 Q2E-ES1155
Advanced Measurement
IMT option
IMT measurement is the measurement to calculate the thickness of the internal media complex (IMC)
of a carotid artery through automatic trace processing.
“Trace method” and “three-point method”, which calculate only the mean value of three points are
available.
44 Q2E-ES1155
Advanced Measurement
%WT ① option
This performs measurement by using B Mode video (moving images). Pixels of two points (left ventricular
wall) specified by the user are automatically traced, and based on the results of that tracing, changes in
the distance between the two points are calculated. From the distance between the two specified points
and from the distance of the maximum wall thickness, the rate of wall thickness increase (%WT) will be
calculated.
Start
Re-measurement
Move the cursor over the position of a
Press the READ key to open the Image Filing
measurement point, and the cursor's shape
window.
changes to the following. If you press the UNDO
Select a AVI file used for measurement from the
key, the measurement pair is deleted and new
image list and press the Retrieve button or
positions can be specified. (It is possible to correct
directly press double click on the image.
multiple pairs)
45 Q2E-ES1155
Advanced Measurement
%WT ② option
#
Speed
Displaying graphs
Saving data
Saving the locus image
RevCapture
You can output the following measurement data in CSV file format
• Distance per pixel [mm/pixel]
• %WT [%] of each measurement pair
• Maximum thickness [mm] of each measurement pair
• Initial thickness [mm] of each measurement pair
• Thickness [mm] of each frame for each measurement pair
• Coordinates on the window of each frame for each measurement value
46 Q2E-ES1155
Advanced Measurement
Color Tissue Tracking ① option
Color Tissue Tracking (hereafter CTT) is a function for calculating the radial strain of a cardiac short axis
image. CTT automatically determines a tracing point from the ROI specified on an AVI file. Calculated
strain and strain values will be displayed by using color images and time sequence graphs.
47 Q2E-ES1155
Advanced Measurement
Color Tissue Tracking ② option
① Align the measurement cursor with ② Move the trackball to adjust ③ Press the Start button.
the intima, click the ENTER key, the intima elliptically and
align the measurement cursor also click the ENTER key.
with the intima on the opposite
side, and click the ENTER key.
Command functions
Changing the image and graph positions Changing the %WT value and color
You can change the display positions of video To change the %WT value, retype the value in the
images and graphs. dialog box. To change the color, click the color box
*Measurement on graphs is enabled only when to display the color palette. Choose a color you
graphs are displayed in large size on the right side. prefer.
Analysis of trimming
For a AVI file of two heartbeats or more, you can perform an analysis between the specific frames of each
heartbeat.
Use the frame bar (frame 1 to frame 45 in the figure) to specify a start point and an end point. Use the pointer
to move the frame bar.
Press the Save button to save the measurement You can correct the ROI.
result data (used for Bull's Eye) in CSV file format. The ROI returns to the pre-analysis status.
Saving a AVI file on basic area Saving AVI file on full area
A CTT video (moving images) is created from For a AVI file, you can save all frames synchronized
the analyzed video. You can check the created with a graph's time bar.
video from the Image Filing window.
48 Q2E-ES1155
Advanced Measurement
LA Tracking ① option
LA Tracking (hereafter LAT) is tracking system of traced line on an B-mode AVI file. LAT calculates the
area and volume from the analyzed result, and displays the change graph of time.
LAT measurement
Press the MEAS. MENU key, select LA Tracking from the
items displayed, and start LAT.
49 Q2E-ES1155
Advanced Measurement
LA Tracking ② option
Command functions
Start analyze
Press the Start button to perform LAT analysis.
For Bi plane measurements, switch between
A4C and A2C for AVI file.
Analysis of trimming
For a AVI file of two heartbeats or more, you can perform an analysis between the specific frames
of each heartbeat.
Use the frame bar (frame 1 to frame 45 in the figure) to specify a start point and an end point. Use
the pointer to move the frame bar.
Changing the image and graph positions Changing the graph type
Saving AVI file on full area Saving a AVI file on basic area
50 Q2E-ES1155
4D (Real-time 3D)
Preparation option
51 Q2E-ES1155
4D (Real-time 3D)
Scanning, Volume operation option
Top;Right Top;Left
* Select which side of the window, right or left, is to display the head side of the fetus.
【 3D ROI Setting window 】 * A 3D image is displayed with the selected side as its top.
Top;Right Top;Front
Top;Back Top;Left
* The orientation of the 3D image changes with each press of the screen button.
Starting 4D scanning
Press the UPDATE (NEXT) key, 4D scanning is started.
Rotating a 3D image
Use the following encoder to rotate a 4D (3D) image:
52 Q2E-ES1155
4D (Real-time 3D)
Function option
② Move the cursor, press the ENTER key to start tracing, specify the range of the
unnecessary part by using the trackball, and then press the ENTER key to exit.
The volume data within the specified range will be cut out.
You can also remain the region
Cut;Inside Cut;Outside
surrounding the specified range by
using the screen button.
Cut Menu Press the Cut Menu screen button again to finish cutting and close the menu related to cutting.
53 Q2E-ES1155
4D (Real-time 3D)
Volume Data Save and Load option
Save Volume
Filing menu
【Filing window】
③ Press the Edit Volume screen button to edit or process the volume data.
Edit Volume
Press the Edit Volume screen button and press the Filing button.
54 Q2E-ES1155
4D (Real-time 3D)
HI DEF 3D (Scanning) option
HI DEF 3D is a 3D mode in which observation of still 3D images and MPR images with a higher definition
than the real-time 3D mode is made possible through a slow single-swing.
45
Swing Angle
Rendering Menu
55 Q2E-ES1155
4D (Real-time 3D)
HI DEF 3D (Function) option
* You can rotate the image and save the volume data in the same way as 4D.
MPR_Y MPR_Y
2
MPR Bright
56 Q2E-ES1155
4D (Real-time 3D)
HI DEF 3D (Multi Slice mode) option
57 Q2E-ES1155
4D (Real-time 3D)
HI DEF 3D (CFM 3D mode) option
Switch of B/W & blood flow 3D image and Blood flow 3D image
Press B mode encoder.
Plane Cut function Blend adjustment of B/W & Blood flow 3D image
Select the object to cut by screen button, and move Press Image Menu key, select CFM 3D Blend
trackball pressing ENTER key to cut the 3D image to make adjustment.
from a plane.
Increase
50
CFM 3D Blend
58 Q2E-ES1155
4D (Real-time 3D)
STIC function option
3D image display
Select Display Setting and then Display 3D
from screen buttons.
* You can save the currently * The load method of STIC data
Save STIC
displayed STIC data. is the same as the 4D data.
* For review and edit of STIC data, see the HI VISION Preirus Three-dimensional Display Software (Real-Time) INSTRUCTION
MANUAL.
59 Q2E-ES1155
3D (Freehand 3D)
Preparation option
:direction of view.
60 Q2E-ES1155
3D (Freehand 3D)
Scanning option
* A progress bar appeared with remaining time. Please scan complete 【MPR window】 【3D Single image window】
in time.
* You can also use the FREEZE key to end capturing.
※NOTICE
If you move the probe in a direction that is not the specified direction, a mirror 3D image will be created.
30mm
Body Mark Move Dir;# Scan Distance Top;# ViewLine;# 3D Scan Cancel User Define
61 Q2E-ES1155