HumaSed 100
| Quick Guide
Front View
Cover
• To protect tubes during
operation and to have
results indicated by
opening and closing (not
applicable to random
operation)
LCD Display
• Channel numbers
LED
• To display results
• Green = Machine ready and
• To indicate status of
available
channels
• Yellow = Mixing or reading
• AVL = Channel is available
operation
• XXX = Channel is occupied
• Blue = Machine busy
• LVC = Level checking
initializing
• ERL = Error level
• ERR = Error
• MAX = > 140 mm/hr
Rear View
Power on/off
Memory
|
Cat.No. 15010/05 01/2011-05
Turning the System ON
Connect the power supply to the input jack situated at the rear of the instrument, marked DC IN. Plug the
power adapter into a grounded 110-240 volt power line. Press the ON/OFF rocker switch at the rear of the
mix
instrument to turn on the HumaSed 100 and follow the sequence of information which will appear on the LCD
display.
Step 2 Selection of temperature compensation option (refer to correction table according to Manley).
Description SW3
Temperature compensation disabled OFF
Temperature compensation enabled ON
Step 5 Selection of results presentation on the LCD and printout. Analysis Time
Description SW6 SW7 Fast Mode Normal Mode
Result presentation: 1 hr OFF OFF 12 min. 20 min.
Results presentation: 30 mins, 1 hr ON OFF 12 min. 20 min.
Results presentation: 1 hr, 2 hrs OFF ON 20 min. 40 min.
Results presentation: 30 mins, 1 hr, 2 hrs ON ON 20 min. 40 min.
(*) Only applicable when mm/2Hrs ESR RESULT is selected. “INVALID” ESR RESULT is displayed if such pre-requisite is not met.
mix
Step 7 Power off and power on the HumaSed 100 to allow read in the new configuration settings.