G3P + O2 GPO DAP + H2O2 PROCEDURE

1. Assay conditions:
H2O2 + 4-AP + p-Chlorophenol POD Quinone + H 2O
Wavelength:……………….505 nm (490-550)
Cuvette ………………………………... 1 cm light path
Triglycerides The intensity of the color formed is proportional to the Temperature…………………………….37°C / 15-25°C
triglycerides concentration in the sample. 2. Adjust the instrument to zero with distilled water.
GPO-POD. Liquid 3. Pipette into a cuvett e:
MATERIALS PROVIDED Blank Standard Sample
For in -vitro diagnostic use only REAGENTS R (mL) 1.0 1.0 1.0
GOOD pH 7.5 50 mmol/L Standard (µL) -- 10 --
Store at 2-8°C p-Chlorophenol 2 mmol/L Sample (µL) -- -- 10
R Lipopro tein lipase (LPL) 150000 U/L
Glycerol kinase (GK) 500 U/L
INTENDED USE Glycerol-3-oxidase (GPO) 3500 U/L 4. Mix and incubate for 5 min at 37°C or 10 min at 15-
For the determination of Triglycerides concentration in 4 – Aminophenazone (4-AP) 0.1 mmol/L 25°C.
human serum or plasma. ATP 0.1 mmol/L 5. Read the absorbance (A) of the samples and
calibrator, against the Blank. The color is stable for
INTRODUCTION at least 30 minutes.
Triglycerides are fats that provide energy for the cell. Like
cholesterol, they are delivered to the body’s cells by TRIGLYCERIDES STD: Triglycerides primary calibrator
200 mg/dL CALCULATIONS
lipoproteins in the blood. A diet with a lot of saturated fats (A)Sample x 200 (Standard conc.)= mg/dL
or carbohydrates will raise the triglyceride levels. The EQUIPMENTSNEEDED BUT NOT PROVIDED
(A) Standard triglyceride in the
increases in serum triglycerides are relatively non-specific.  Spectrophotometer or colorimeter measuring at Sample
For example liver dysfunction resulting from hepatitis, 505 nm.
Conversion factor: mg/dL x 0.0113 mmol/L.
extra hepatic biliary obstruction or cirrhosis, diabetes  Matched cuvettes 1.0 cm light path.
mellitus is associated with the increase.  General laboratory equipment.
Clinical diagnosis should not be made on a single test QUALITY CONTROL
result; it should integrate clinical and other laboratory REPARATION  If control values are found outside the defined
data.  Reagent and standard provided are ready to use. range, check the instrument, reagent and
calibration for problems
PRINCIPLE STORAGE AND STABILITY  Each laboratory should establish its own Quality
Sample triglycerides incubated with lipoproteinlipase (LPL),  All the components of the kit are stable until the Control scheme and corrective actions if
liberate glycerol and free fatty acids. Glycerol is expiration date on the label when stored tightly controls do not meet the acceptable
converted to glycerol-3-phosphate (G3P) and adenosine- closed at 2-8°C, protected from light and tolerances.
5-diphosphate (ADP) by glycerol kinase and ATP. Glycerol- contaminations prevented during their use.
3-phosphate (G3P) is then convert ed by glycerol  Do not use reagents over the expiration date. REFERENCE VALUES
phosphate dehydrogenase (GPO) to dihydroxyacetone  Signs of reagent deterioration: Men 40 – 160 mg/dL
phosphate (DAP) and hydrogen peroxide (H2O2).  Presence of particles and turbidity . Women 35 – 135 mg/dL
In the last reaction, hydrogen peroxide (H2O2) reacts with NOTE
 Blank absorbance (A) at 505 nm > 0.40. These values are for orientation purpose; each
4-aminophenazone (4-AP) and p-chlorophenol in
presence of peroxidase (POD) to give a red colored dye: laboratory should establish its own reference range.
SAMPLES
Serum or plasma. PERFORMANCE CHARACTERISTICS
Triglycerides + H 2 O LPL Glycerol + free fatty acids Stability of the sample: Triglycerides are stable for 5 days Measuring range:
at 2-8°C. From detection limit of 5.85to linearity limit of 1000
Glycerol + ATP Glycerol kinase G3P+ ADP
mg/dL.
If the concentration is greater than linearity limit, 4. Young DS. Effects of disease on Clinical Lab.
dilute the sample 1/2 with NaCl 9 g/L and multiply the Tests, 4th ed AACC
result by 2. 5. Burtis A et al. Tietz Textbook of Clinical
Precision: Chemistry, 3rd ed AACC
Intra-assay Inter-assay 6. Tietz N W et al. Clinical Guide to Laboratory
(n=20) (n=20) Tests, 3rd ed AACC
Mean (mg/L) 123 205 124 204
SD 3.08 1.63 4.89 4.28
CV (%) 2.49 0.79 3.92 2.09 ATLAS MEDICAL
Unit 4, William James House
Cowley Rd, Cambridge, CB4 0WX
Sensitivity: Tel: ++44 (0) 1223 858 910
1 mg/dL = 0.0011 (A). Fax: ++44 (0) 1223 858 524
Accuracy: Results obtained using ATLAS reagents (y) did
not show systematic differences when compared with PPI406A01
other commercial reagents (x). The results of the Rev C (04.06.2015)
performance characteristics depend on the analyzer Product For in-vitro
used Reference No. diagnostic use.
Store at 2 -
Caution.
8°C.
INTERFERENCES
Read product Number of
No interferences were observed with bilirubin up to 170 insert before tests in the
mmol/L and hemoglobin up to 10 g/L. use. pack.
A list of drugs and other interfering substances with Lot (batch)
Manufacturer.
cholesterol determination has been reported by Young et. Al number.
Manufacturer
NOTES Expiry date. telephone
 Calibration with the aqueous standard may cause number.
a systematic error in automatic procedures. In Manufacturer
fax number.
these cases, it is recommended to use a serum
Calibrator.
 Use clean disposable pipette tips for its
dispensation.

REFERENCES
1. Buccolo G et al. Quantitative determination of
serum triglycerides by use of enzimes. Clin
Chem 1973; 19 (5): 476-482.
2. Fossati P et al. Clin. Chem 1982; 28(10): 2077-
Kaplan A et al. Tryglycerides. Clin Chem The C.V.
Mosby Co. St Louis. Toronto. Princeton 1984;
437 and Lipids 1194-1206.