You are on page 1of 8

Sr.

Inst Calibration
Instrument / Equipment Due date
No. No. range

1. Thermohygrometer TH-12 12.07.2020 Temperature:


2.0˚C to 8.0˚C

Relative
Humidity:
30% to 70% rh

2.
HPHV INST-
126

a) Compound Gauge
Forbes Marshall 16.07.2020
(1802/MAC/03)
-1-1.5 kg/cm2
(1802/MAC/04) -1-4 kg/cm2
(1802/MAC/05) -1-4 kg/cm2
(1802/MAC/06) -1-4 kg/cm2
1-5 kg/cm2
b) Pressure Gauge 1-5 bar
(1802/QCSS/01) 1-5 bar
(1802/QCSS/02) 1-5 bar
(1802/QCSS/03)
(1802/QCSS/04)
-1-1.5 kg/cm2
c) Compound Gauge
(1802/MAC/01)
-1-1.5 kg/cm2
(1802/MAC/02)

3. Thermohygrometer
TH-01 17.07.2020 Relative
Humidity, rh:

30 % rh to 70
% rh
Temperature,
°C:

20.0°C to
30.0°C

4. Thermohygrometer
TH-03 17.07.2020 Relative
Humidity, rh:

30 % rh to 70
% rh

Temperature,
°C:

20.0°C to
30.0°C

5.

Incubator Newtronic INST- 30.08.2020 20˚C - 25˚C


005

6.

Thermometer TM-11 06.09.2020 -10˚C to 250˚C

7.

Thermohygrometer TH-25 10.09.2020 Relative


Humidity, rh:

25% rh to 75%
rh
Temperature,
°C (IN):

2.0°C to 8.0°C

Temperature,°C
(OUT):

20.0°C to
40.0°C
8.

Precision Balance INST- 26.09.2020 1 gram to


149
310 gram

9.

Analytical Balance INST- 26.09.2020 1 mg to


148
200 gram

10.

Standard weight SW-02 14.102020 1g -100g

11.
Airflow
Laminar Air Flow INST- 21.10.2020 velocity test
123 Minimum:
0.36m/s
HEPA filter
installation leak
test

Maximum:
0.01%

Airborne
particle count
test
<3520 count/m3
at particle size
≥ 0.5µm

12.
Biohazard Safety Cabinet INST- Downflow
168 21.10.2020 velocity
0.45 ± 20%

Faceflow
velocity
> 0.5m/s

HEPA filter
installation leak
test

<0.01%

Airborne
particle count
test

< 3520
count/m3

Light intensity

>800 Lux

UV light
intensity
>40µw/cm

1. High Pressure High INST-


Vacuum Sterilizer 126

a) Compoud Gauge
Forbes Marshall
(1802/MAC/03) 20.03.2020 -1-1.5 kg/cm2
(1802/MAC/04) -1-4 kg/cm2
(1802/MAC/05) -1-4 kg/cm2
(1802/MAC/06) -1-4 kg/cm2
1-5 kg/cm2
b) Pressure Gauge 1-5 bar
(1802/QCSS/01) 1-5 bar
(1802/QCSS/02) 1-5 bar
(1802/QCSS/03)
(1802/QCSS/04)
-1-1.5 kg/cm2
c) Compound Gauge -1-1.5 kg/cm2
(1802/MAC/01)
(1802/MAC/02)

2.

Thermometer TM- 01.04.2020 0˚C to 250 ˚C


05

3.

Sartorius Balance INST- 15.04.2020 0 gram to


113 420 gram
4. pH & conductivity INST- 15.04.2020 pH 4 - pH 10
140

5.

Laminar Air Flow INST- 21.04.2020 Airflow


(Sterility room) 045 velocity test
Minimum
0.36m/s

HEPA filter
installation
leak test
Maximum:
0.01%

Airborne
particle count
test
<3520
count/m3 at
particle size ≥
0.5µm
<29 count/m3
at particle size
≥ 5.0µm

6.

Standard weight SW- 13.06.2020 200 gram


03
7.

Thermohygrometer TH- 17.06.2020 Temperature:


13 15.0˚C to
35.0˚C

Relative
Humidity:
30% to 90%

8.

Dry Heat Sterilizer INST- 19.06.2020 Max: 400˚C


012

9.

BOD Incubator INST- 19.06.2020 30˚C to 35˚C


108

10.

Incubator Carbolite INST- 19.06.2020 30˚C - 35˚C


049

11.

Incubator Pathak INST- 19.06.2020 40 ˚C to 45˚C


016

12.
Digital caliper NA 19.06.2020 0 mm to 150
mm

You might also like