Fresenius Medical Care

Technical Training
on
4008 Dialysis Machine

Shanghai, April 2002

Hotin

May 2001
Dialyzer Membrane
Fresenius Medical Care

May 2001
Fresenius Polysulfone - Structure
Fresenius Medical Care

May 2001
Fresenius Polysulfone - Structure
Fresenius Medical Care

HOLLOW FIBRE DIALYSER

PORE
PORESIZE
SIZEisisdependant
dependanton
onthe
thetype
type
ofofmembrane;
membrane;

HF Dialyser:
Pore size ~ 0.3nm

LF Dialyser:
Pore size ~ 0.1nm

May 2001
Venous Pressure
Measurement

P
Fresenius Medical Care

Blood out

Venous
Bubble Catcher
Dialysate in

Dialyzer

Dialysate out

Venous Clamp
Blood in

Blood Pump

Arterial Pressure
Measurement
HEMODIALYSIS HD

The transport of uremic wastes by diffusion, i.e. the process in which


of particles move across a semipermeable membrane from the area
of higher concentration (blood) to the area of lower concentration
(dialysate).

May 2001

j:\sd\ch\folien\tpf_nma\hd_arm_e.cdr
SD / TRAINING CENTER SCHWEINFURT