You are on page 1of 10

5E-HGT2321

Direct Total Mercury Analyzer


5E-HGT2321 Direct Total Mercury Analyzer

5E-HGT2321 Direct Total Mercury Analyzer is able to directly measure the total
mercury in solid, liquid and gas samples without the need of complex sample
pretreatment, effectively avoiding the loss of mercury in the pretreatment
process, cross contamination and experimental environmental pollution, and
ensuring the accuracy of the analysis data.

Compared with the traditional chemical vapor generation- atomic fluorescence


spectrometry method, the pyrolysis method is more efficient, accurate, stable
and reliable in measuring mercury, which can greatly reduce labor intensity, and
save 70% of analysis time and cost.

Working Principle

Sample is dried by heating and burnt (pyrolysis) in oxygen or air. Its decomposed
product is oxidized by the high temperature catalyst carried via oxygen.
Meanwhile, distractors like halogen, nitric oxides, and sulfur oxides are absorbed
by catalyst. Moisture is exhausted by pneumatic truncation, and the rest of
decomposed product is conveyed to sub-nanometer Hg-trap further. Mercury
atoms are selectively captured, and finally experience secondary atomization by
heating Hg-trap. The flowing oxygen carries mercury atoms into the cold atomic
absorption detector (253.7nm) for determination, and the total mercury content
in the sample is directly calculated by the software.

Integral prepared catalyst


and gold tube
5E-HGT2321 Direct Total Mercury Analyzer

Low Effort for Ideal Efficiency


 Wide Applicability:Suitable for  Low Cost:No need for dehydration
solid, liquid and gas samples, with consumables, consumable cost is 70%
detection capability of 6 orders of lower than the traditional method,
magnitude, 0.005ng ~ 1000ng 40% lower than similar products
 Easy Operation:Weigh sample  Eco-friendly:No harmful chemicals
and place it on auto feeder without are used in the whole analysis process.
digestion. Complete analysis with The exhaust gas produced by the
just one click on Start key analysis is effectively absorbed.
 High Efficiency:Single test only Equipped with the human eye
takes 2-5mins, one batch includes recognition color, which can be an
45 samples, and emergency early warning of the activity of
samples can be inserted at any mercury adsorbents
time  No Memory Effect:Unique cutting-
 Splendid Repeatability & off dehydration and whole pipelines
Accuracy:Heating intergrated heating design are able to effectively
detection cell and intergrated overcome problems of mercury
catalytic-amalgamation tube memory effect
ensure measurement repeatability  Double Screen Display:With
effectively; composite catalytic PC/PAD dual screen display and
agent of high efficiency guarantees control function, users can monitor
the accuracy of test results the test data without entering the lab
5E-HGT2321 Direct Total Mercury Analyzer

Critical Design
Replaceable Sample Feeder
• Carousel sample feeder, available for 45 samples per batch.
Cyclic addition ensures unattened operation.
• Sample can be weighed and placed on the spare sample tray,
which can be directly replaced after the former one has been
tested.

Reduce Weighing Error


• The balance can be placed next to the instrument and
computer to directly weigh the sample, effectively reducing
the transfer time.
• Press balance Print key to input sample weight to the
software, greatly avoiding errors in sample weighing.

Intergrated Catalytic-Amalgamation Tube


• Unique intergrated tube can effectively improve the reliability
of connection in high temperature of catalytic-amalgamation
tube and avoid mercury residue at the connection.
• Greatly eliminate the moisture condensation generated by
combustion and improve the analysis speed.

Double Screen Display & Control


• Equipped with a removable micro-display screen and can be
controlled by PAD in addition to computer.
• Adopt wireless connection for micro screen to facilitate
monitoring the testing process and results.
5E-HGT2321 Direct Total Mercury Analyzer

Specification

Principle Pyrolysis-Mercury Amalgamation-Cold Atomic Absorption

Control Method PC / PAD

Sample Type Solid, liquid, and gas

Max Sample QTY 1000mg (solid),1000μL (liquid)

Optical Source Low-pressure mercury lamp

Wave Length 253.7nm

Detector Double UV photodetector

Repeatability RSD<1% @ 5ng

Analysis Time 2~5min

Detection Limit 0.001ng

Working Range 0.005~1000ng

Analysis Method Adjustable according to samples

Heating Temp. Room temperature-950℃ adjustable

Carrier Gas Oxygen(>99%) or compressed air

Energy Consumption Peak<3000W,standby<100W

Calibration Standard Solution or CRM

Weight 40kg

Dimension(L×W×H) 590×460×330mm

Power Supply 110/220VAC,50/60HZ


5E-HGT2321 Direct Total Mercury Analyzer

Simple Analysis Process

Sample

少图

Take Samples Weigh Samples Auto Feed Samples Get Results in 3min

Analysis Accuracy

Sample Certified Material Reference Value 5E-HGT2321 Test Value


No. (mg/kg) (mg/kg)

Henan weat GBW10046 0.0022 0.0021±0.0002

Rice GBW(E)100362 (0.0017) 0.0016±0.0002

Scallions GBW10048 0.0146±0.0024 0.0153±0.0006

Citrus leaf GBW10020 0.150±0.020 0.149±0.003

Celery GBW10048 0.0146±0.0024 0.0145±0.0005

Chicken GBW10018 0.0036±0.0015 0.0033+0.001

Soil GSS-2 0.015±0.03 0.014±0.005

Soil GSS-5 0.29±0.03 0.30±0.01

Coal GBW10060 0.096±0.06 0.092±0.003

Coal of South SARM20 0.25±0.03 0.25±0.005


Africa
5E-HGT2321 Direct Total Mercury Analyzer

Typical Results

Type Food Water Soil

Sample Rice Wheat Waste Irrigation Chernozem Red Brown Soil

QTY(mg) 100 100 100 1000 100 100

Drying Temp.(℃) 300 280 280 280 330 330

Drying Time(s) 60 60 60 60 30 30

Decomposition Temp.(℃) 750 750 750 750 750 750

Decomposition Time(s) 70 70 120 120 120 120

Total Time (min) 2.5 2.5 3.5 3.5 3 3

Results of 5 Times (μg/kg) 4.4±0.1 1.2±0.1 6.5±0.2 2.1±0.2 65.7±0.8 53.8±0.6

02
5E-HGT2321 Direct Total Mercury Analyzer

Comparative Analysis of the Efficiency between Pyrolysis Mercury Method and Chemical
Vapor Generation Atomic Fluorescence Method

Take completing 24 samples as example

Traditional method (microwave


Pyrolysis-mercury amalgamation-
Method digestion-atomic fluorescence
cold atomic absorption
spectrometry)

Nitric acid, hydrochloric acid, perchloric


Reagent acid, potassium borohydride, sodium None
hydroxide/potassium

High-pressure argon gas, balance,


Required microwave digestion apparatus, atomic Oxygen or air, balance, and
Apparatus fluorescence photometer, and common mercury analyzer
containers in laboratories

Weigh-digestion-filtration-constant
Process Weigh-auto determination
volume-determination

Laboratory environment, residual


Influence memory effect, baseline drift, cross
None
Factors contamination, pretreatment loss, and
mood of operator

Hazardous Acid mist, unused sample solution, and


None
Wastes acid base solution, etc

Time 8h 1h

1. Low working intensity, short


1. High working intensity and reagent
time consumption, and no
consumption.
reagent required.
Conclusion 2. Time-consuming.
2. The results are accurate with
3. Hard to measure the sample of low
unattended operation and
concentration accurately.
high automation.
5E-HGT2321 Direct Total Mercury Analyzer

Conforms to Method
Standard Title

GB/T 5009.17-2014 Determination of total mercury and organic-mercury in foods

GB/T 31947-2015 Iron ores - Determination of mercury content - Solid sampling and direct
mercury analysis method

HJ 923-2017 Soil and sediment—determination of total mercury—catalytic pyrolysis-


cold atomic absorption spectrophotometry

HJ 910-2017 Ambient air — determination of gaseous mercury — gold amalgamation


collection and analysis by cold vapor atomic absorption
spectrophotometry

HJ 917-2017 Stationary source emission—determination of total gaseous mercury—


carbon sorbent traps / thermal cracking atomic absorption
spectrophotometric method

SN/T 3010-2011 Determination of mercury in paint - solid and liquid direct injection
method

SN/T 3511-2013 Determination of mercury in minerals. General rules of direct mercury


determining with solid injection method

SN/T 3912.3-2014 Determination of mercury content in imported condensate - direct


injection method

SN/T 4697-2016 Determination of mercury in paper packaging- solid injection - direct


mercury analyzer method

SN/T 4763-2017 Determination of mercury in coal - solid injection - direct mercury analyzer
method

DB45/T 939-2013 Soil, fertilizer, feed, hair - determination of mercury content - direct
mercury analyzer method

DB22/T 1979-2013 Determination of mercury in cosmetics-automatic mercury analyzer

EPA 7473 Mercury in Solids and Solutions by Thermal Decomposition, Amalgamation,


and Atomic Absorption Spectrophotometry

ASTM D6722 Standard Test Method forTotal Mercury in Coal and Coal Combustion
Residues byDirect Combustion Analysis

ASTM D7623 Standard Test Method for Total Mercury in Crude Oil Using Combustion-
Gold Amalgamation and Cold Vapor Atomic Absorption Method
Changsha Kaiyuan Instruments Co., Ltd.
No.1259 Liangtang East Road
Changsha Economic & Technological
Development Zone Changsha, P. R. China.
Website: www.ckic.net

Sales Contact
Tel: 86-731-84021617
Email: info@ckic.net

Service Contact
Tel: 86-731-84879027
Email: service@ckic.net

You might also like