You are on page 1of 1

da Vinci SP Reprocessing

SP CAMERA

1 Remove Accessories 9 Rinse


Remove sheath. Rinse under running cold water. 60 sec

2 Inspect 10 Ultrasonic Cleaning


Inspect entire camera for damage. Prime and clean in ultrasonic bath. 15 min

3 Prepare Solution 11 Flush


Prepare enzymatic solution (pH 7-11). Flush all ports with pressurized cold water. 2 x 20 sec

4 Prime 12 Final Rinse


Prime primary port with enzymatic solution. Rinse with critical water. 60 sec

5 Soak 13 Inspect
Soak camera and cable. 30 min Inspect entire camera for soil.

6 Flush 14 Leak Test


Flush all ports with pressurized cold water. 2 x 20 sec Submerge tip and use pressurized air. 30 sec

7 Spray 15 Thermal Disinfection (optional)


Submerge and spray areas (holes and wrist). 2 x 30 sec Disinfect per hospital policy.

8 Brush
Brush submerged or under running water. 60 sec

da Vinci Universal Reprocessing

INSTRUMENTS AND STAPLER

1 Remove Accessories 8 Rinse


Remove all accessories before cleaning. Rinse under running cold water. 60 sec

2 Prepare Solution 9 Ultrasonic Cleaning


Prepare enzymatic solution (pH 7-11). Prime and clean in ultrasonic bath. 15 min

3 Prime 10 Flush
Prime primary port with enzymatic solution. Flush ports with pressurized cold water. 20 sec

4 Soak 11 Final Rinse


Soak the device. 30 min Rinse with critical water. 60 sec

5 Flush 12 Inspect
Flush all ports with pressurized cold water. 20 sec Inspect entire device for soil.

6 Spray 13 Thermal Disinfection (optional)


Submerge and spray all areas. 30 sec Disinfect per hospital policy.

7 Brush
Brush submerged or under running cold water. 60 sec

ACCESSORIES GROUP 1
Cannulae, Obturators, Release Kits, Hasson Cone, Instrument Introducers, Gage Pins, Suction Irrigator Kit, etc.

1 Remove Accessories 6 Ultrasonic Cleaning (optional)


Remove all accessories before cleaning. Clean in ultrasonic bath. 15 min

2 Prepare Solution 7 Final Rinse


Prepare enzymatic solution (pH 7-11). Rinse with critical water. 10 sec

3 Soak 8 Inspect
Soak and actuate moving parts. 10 min Inspect entire device for soil.

4 Rinse 9 Thermal Disinfection (optional)


Rinse under running cold water. 20 sec Disinfect per hospital policy.

5 Brush
Brush submerged or under running cold water. 60 sec

ACCESSORIES GROUP 2
Cautery Cord, Stapler Cable, PK Cords, Light Guide Cable, etc.

1 Disconnect Cords or Cables 5 Final Rinse


Disconnect all cords and cables from device. Rinse with critical water. 60 sec

2 Prepare Solution 6 Inspect


Prepare enzymatic solution (pH 7-11). Inspect entire accessory for soil.

3 Wipe 7 Thermal Disinfection (optional)


Wipe thoroughly using enzymatic solution. Disinfect per hospital policy.

4 Brush
Brush submerged or under running cold water. 60 sec

Intuitive Surgical, Inc.


1266 Kifer Road
Sunnyvale, CA 94086 USA
Customer Service 800.876.1310

Copyright © 2018 Intuitive Surgical, Inc. All rights reserved. For Intuitive Surgical da Vinci® Surgical System Instruments, Accessories, Stapler,
Product names are trademarks or registered trademarks of their respective holders. and SP Camera, refer to Reprocessing Instructions. This chart is not intended to
replace an in-service or a manual.

PN 554061-01 Rev. A