Measurement Uncertainty
(VIT / EMU / 17)
Uncertainty Evaluation
Accuracy in Accuracy MU from Cal.
Equipment Name Parameter Range Resolution
value in % Certificate
Uncertainty Evaluation for Voltage Measurement
Reading Rn Parameter Mean [R] Diff [Rn-R] Diff2
R1 n= 10
Type A R2 n-1 = 9
Evaluation R3 value in P
R4
R5
Mean Value
( R1+R2….+Rn)/n
sum of Diff2 =
where n = number of
readings
Standard Deviation = Sqrt [[ 1/n-1]* sum of Diff 2 ]=
Standard Uncertainty [UA] = Std. Deviation / Sqrt [n] = P
Degree of freedom = VA = n-1 =
Standard Uncertainty [ % UA] = UA *100/ Mean value = %
Type B Evaluation
1. Uncertainty based on the Uncertainty declared in the calibration
certificate of the instrument [ U1 ]
Uncertainty [ A1 ] declared in the calibration certificate
at Normal Distribution and coverage factor [k =2] at 95%
%
confidence level =
Combined Uncertainty [ Uc ]
Uc = sqrt (UA2 + U12 +U22 + U32 ) = P
Standard Uncertainty [ % Uc] = Uc *100/ Mean value = % or P
Expanded Uncertainty [ U ]
Expanded Uncertainty [ U ] at approximately 95% of confidence level,
P k= 2
with a coverage factor k = 2 = k*Uc =
%
% of Expanded Uncertainty [ % U ] = [U]*100/Mean Value P
or
Probability Degrees
Source of Uncertainty
Distribution Standard Uncertainty of
Uncert- Estimate Limits Contribution
Type A or Freedom
ainty
Type B
P % P %
Repeat-
Normal
ability ---
Type A
[UA]
Normal
U1 -- Type B
k=2
Rectangular
U2 -- Type B
Sqrt(3)
Rectangular
U3 -- Type B
Sqrt(3)
Uc -- -- --
Expande
d Uncert- -- -- k=2 -- --
ainty [U]