PLX101D Portable digital X-ray machine installation condition confirmation form

To ensure good performance and satisfactory clinical work, PLX101D Portable digital X-ray machine must be installed in accordance with below
requirements:

ITEM CONTENT Yes/No Name / Signature


Power supply Power supply capacity within 220v ±10%
1
requirements Power supply frequency 50 Hz
Power Supply Internal Resistance: ≤ l Ω
Room X-ray room area ≥ 2m x 2m
2
condition Room height ≥2.5m,room length ≥2m
The machine base occupies an area of about 200 Kg

Door requirements The width of the road to delivery the machine r≧1.5m
3
The height of the door≧2m
The width of the door≧1m
Equipment Temperature: -5~55℃
4
storage relative humidity: 10 - 90% (no condensation)
environment Atmospheric pressure: 50-106 kPa;
requirements indoor environment without corrosive gases;
packing box should not be stacked with more than 30 kg
weight
Working Temperature: +10℃~+40℃
5
environment relative humidity: 30%~+75% (no condensation)
requirements Atmospheric pressure: 70~106kPa;
No corrosive and explosive gas. No dust.
Optional A laptop with battery need to be prepared, memory more
6
requirements than 4GB, our engineer will help to do a back up for the
workstation.

NOTE:
 Recommend UPS in 2000VA power to support the instrument .

 The X ray building must have proper grounding.

User Requirements

Requirements Yes/No Name / Signature

The hospital’s operator on this instrument must be


on-site.
The operator details:
Full name:
Location: ( Hospital name)
Contact number:
If the installation is failed because the above installation requirement can not met, the manufacturer and distributor are exempt from liability.