To ensure good performance and satisfactory clinical work, PLX101D Portable digital X-ray machine must be installed in accordance with below
requirements:
Door requirements The width of the road to delivery the machine r≧1.5m
3
The height of the door≧2m
The width of the door≧1m
Equipment Temperature: -5~55℃
4
storage relative humidity: 10 - 90% (no condensation)
environment Atmospheric pressure: 50-106 kPa;
requirements indoor environment without corrosive gases;
packing box should not be stacked with more than 30 kg
weight
Working Temperature: +10℃~+40℃
5
environment relative humidity: 30%~+75% (no condensation)
requirements Atmospheric pressure: 70~106kPa;
No corrosive and explosive gas. No dust.
Optional A laptop with battery need to be prepared, memory more
6
requirements than 4GB, our engineer will help to do a back up for the
workstation.
NOTE:
Recommend UPS in 2000VA power to support the instrument .
User Requirements