This document is an installation confirmation form for a PLD7300A medical equipment model. It summarizes the power and room requirements that must be met for proper installation and use of the equipment. These include requirements for the distribution box height, circuit breaker capacity, power supply voltage and frequency stability, grounding resistance, room size and dimensions, floor bearing capacity, temperature and humidity ranges, and observation window specifications. The user must check off that all listed project and field measurement results are qualified according to the requirements.
This document is an installation confirmation form for a PLD7300A medical equipment model. It summarizes the power and room requirements that must be met for proper installation and use of the equipment. These include requirements for the distribution box height, circuit breaker capacity, power supply voltage and frequency stability, grounding resistance, room size and dimensions, floor bearing capacity, temperature and humidity ranges, and observation window specifications. The user must check off that all listed project and field measurement results are qualified according to the requirements.
This document is an installation confirmation form for a PLD7300A medical equipment model. It summarizes the power and room requirements that must be met for proper installation and use of the equipment. These include requirements for the distribution box height, circuit breaker capacity, power supply voltage and frequency stability, grounding resistance, room size and dimensions, floor bearing capacity, temperature and humidity ranges, and observation window specifications. The user must check off that all listed project and field measurement results are qualified according to the requirements.
ZHPLD-JL-SH05/A Country Contacts Tel Equipment model PLD7300A Project requirements are Field project content qualified or not Measurement Qualified Unqualified results Dsitribution box installation height l.5m to 1.6m □ □ Power Install a 125A circuit breaker in the distribution box □ □ condition Power supply capacity≥65KVA □ □ Power supply voltage stability, three-phase 380V±5% □ □ power frequency 50Hz/60Hz±1Hz □ □ Power supply internal resistance≤0.17Ω, □ □ Independent grounding, grounding resistance < 4Ω, □ □ neutral and ground voltage< 5VAC Independent power cord ≥25mm²,Copper core cable □ □ X-ray room area ≥20m²;Room length ≥ 5m, width ≥ 4m, □ □ height≥2.9m(Note:net size) X-ray protection(3mm thick lead in X-ray direction, 2mm □ □ thick lead in other directions) Equipped with dehumidifier and air conditioner □ □ Room The size of the observation window is 120cm×80cm, the condition height from the ground is 900mm, and lead glass is □ □ installed. Ground bearing capacity≥600kg, foundation is concrete □ □ structure, thickness ≥ 200mm X-ray room door (width≥1.5m, height≥2.1m) □ □ Open a 100mm hole between the machine room □ □ and the control room The room floor has been installed and the floor has □ □ been cleaned Ambient temperature: +10 ~35'C □ □ Relative humidity: 30 %~75 % □ □ There is equipment operator or no. Yes □ No□ Whether to install by Perlove Install Need to arrange for engineers to yourself□ install□ Note: in order to be able to serve you faster and better, please put√ in □and mail to our company. User units(stamp): User signature: date: