Professional Documents
Culture Documents
net/publication/334388794
CITATIONS READS
0 115
6 authors, including:
Some of the authors of this publication are also working on these related projects:
All content following this page was uploaded by Bart HJ Van den Berg on 11 July 2019.
%B
While nano-flow LC-MS proteomics applications yield 50
unsurpassed depth of coverage and sensitivity, they do
not lend themselves well to high-throughput workflows 0
1 7
and are commonly associated with high levels of MS 13 19 25 31
idle time, especially when long columns are employed. Minutes 37 43
Column 1 Column 2
In contrast, capillary-flow LC-MS methods offer a
unique blend of sensitivity and throughput.1 We recently
published a capillary-flow LC-MS method that exploits Figure 1. Flow-gradient diagram depicting high-throughput low-
both of these attributes while also facilitating previously flow tandem LC operation affording >200 samples per day
2
Table 1. Fluidics, columns, and consumable accessories required to run the application. The letter and number assignments are given in
Figure 2A.
Separation pump
+
Column compartment
1
+
Micro-flow loading pump
4
1
6 Equilibration pump
b
4
Autosampler
5
2
3
a to ESI MS
Figure 2. Fluidic (A) and UltiMate 3000 RSLCnano (B) configurations for the high-throughput low-flow tandem pre-concentration
LC setup. Note: The number and letter descriptions for each of the fluidic components (in black) are given in Table 1.
3
2.2 Samples Table 2A. LC solvents and conditions for high-throughput low-flow
analysis
• Thermo Scientific™ Pierce™ HeLa protein digest
(P/N 88328, 20 µg/vial) reconstituted to a final
Property Setting
concentration of 200 ng/µL in loading buffer (see
Table 2A for details). Mobile phase A: 100% Water + 0.1% FA
Mobile phase B: 20%/80%
• Thermo Scientific™ Dionex™ Cytochrome C (CytC)
Water/ACN + 0.1% FA
Digest (P/N 161089, 1.6 nmol/vial) reconstituted to a
final concentration of 1 pmol/µL. Loading solvent
Loading Pump A channel: 100% Water + 0.05% TFA
2.3 LC-MS configuration and separation Loading Pump B channel: ACN + 0.1% FA
conditions
Sample: Cytochrome C digest
Measurements were carried out using an UltiMate 3000
(1 pmol/μL) and HeLa digest
RSLCnano system,3 comprising an UltiMate
(200 ng/µL)
NCS-3500RS pump, Thermo Scientific™ UltiMate™
NCP-3200RS pump, and Thermo Scientific™ UltiMate™ Sampler wash solvent: 100% ACN + 0.1% FA
WPS-3000TPL RS pulled loop autosampler Injection volume: 1 μL
(Figure 2B). Both NCS-3500RS and NCP-3200RS (air-flanked microliter pickup)
pumps were configured with a ProFlow™ flow meter. The Loading time: 0.1 min
device names and signals of the NCP-3200RS pump
Gradient flow rate: 1.5 μL/min
were modified through the addition of the number “2”
at the end of the standard instrument property fields in Gradient: Low-flow separation pump
the Thermo Scientific™ Chromeleon™ Chromatography Time, min %B
Data System (CDS) instrument configuration manager. 0 8
For example, the main device name was changed from
5.3 35
“PumpModule” to “PumpModule2” to avoid a conflict in
the signal names. The system fluidics were configured 6.7 70
as shown in Figure 2A and Table 1. In order to interface 6.7 8
the respective linear Acclaim PepMap column outlet with 7 8
the 10-port valve on the right-hand side of the column
Gradient: Equilibration pump
oven, the fused silica outlet supplied with the columns
was removed and replaced with the “nano column to MS Time, min %B
tubing” (#5 in Table 1). The column outlet is connected to 0 99
the “nano column to MS tubing” using a nano connector
3.0 99
supplied with the column. A description of how to make
3.3 8
the connection is given in the UltiMate 3000 RSLCnano
Standard Applications Guide4 section 2.1.2. Before the 7 8
capillary is installed it must first be cut to a length of Oven temperature: 60 °C
30 cm using the cleavage stone. The LC system was
Sample temperature: 5 °C
interfaced with the MS using the Thermo Scientific™
Nanospray Flex™ ion source (P/N ES071) using stainless Loading and washing Micro-flow loading pump
steel nanobore emitters (P/N ES542). gradient: Time, min Flow, µL/min %B
0 150 95
Solvents and analysis conditions were used as described 0.5 150 95
in Tables 2 and 3.
0.6 150 0
5.6 150 0
6.7 150 95
7 150 95
4
Table 2B. Switching program and commands for high-throughput low-flow analysis
Property Setting
Intelligent switching program for Time, min Command Valve positions
10-port valves in the column 0
compartment. These commands If
must be inserted manually into ColumnOven.ValveLeft 1_2
the script editor: ColumnOven.ValveRight 10-1
Else
ColumnOven.ValveLeft 10-1
ColumnOven.ValveRight 1-2
End If
6
If ColumnOven.ValveLeft 1_2
ColumnOven.ValveLeft 10_1
Else
ColumnOven.ValveLeft 1_2
End If
Commands manually inserted into Time, min Command
the method script editor: 1.5 Sampler.InjectValveToLoad
1.6 Sampler.Wash
3.4 Sampler.InjectValveToInject
3.6 Sampler.PrepareNextInjection
5.8 Sampler.InjectValveToInject
Property Setting
Draw speed: 1 µL/s
Draw delay: 2s
Dispense speed: 8 µL/s
Dispense delay: 2s
Dispense to waste speed: 8 µL/s
Sample height: 2 mm
Puncture depth: 8 mm
Wash volume: 25 µL
Wash speed: 8 µL/s
5
2.4 MS conditions Table 6. MS settings for DDA experiments
6
3. Results and discussion • High retention time reproducibility via use of one
3.1 Tandem high-throughput low-flow LC dedicated low-flow (NCS-3500RS) pump for gradient
method explained elution while the second (NCP-3200RS) pump is used
An optimized LC method is used for tandem LC for column washing and equilibration.
operation that enables ~100% MS sample data
acquisition while maintaining a short analysis cycle time An example data file containing complete LC-MS
and high sensitivity (Figure 3). This method contains the method parameters is available for download at https://
following attributes: appslab.thermofisher.com/App/4176/tandem-lowflow-
lcms.
• Simultaneous washing of separation column 2 and
trap column 2 parallel to gradient elution on separation 3.2 Continuous MS utilization with tandem
column 1.
LC-MS
• Parallel washing of injection fluidics and sample loop Optimized high-throughput tandem LC-MS methodology
during sample elution. using one single pump for gradient elution on both
analytical columns enabled reproducible chromatography
• Look ahead injections: sample 2 is picked up, loaded with peptide data elution covering more than 6.8 of
onto the trap column 2, concentrated, and transferred 7 minutes of each sample run (Figure 4), equivalent to
to the separation column 2 while sample 1 is over 97% MS utility with a throughput of >200 samples
undergoing gradient elution in parallel (Table 2B). per day.
• The delay in sample peptide elution is removed using
post-column flow diversion. Furthermore, the even distribution of PSMs over the
entire elution window (Figure 5A) ensures a constant
• Intelligent valve switching commands enable LC control sample data feed at a complexity level that can be
for both columns using a single LC method (Table 2B).
handled by the Q Exactive HF-X mass spectrometer.
Sampler wash Sample loop Trap 2 strong Prepare for Sample transfer
between strong solvent solvent wash and next injection to 2nd column
reinjections wash equilibration routine Done in parallel
with loading pump,
Sample load syringe, and 2nd nano pump
onto trap 2
Trap 2 washing
7
7 min 7 min
100 3.63 100 3.60 3.62
90 3.65 90
Column 1 Column 2 3.68
80 3.71 80 3.70 4.54
4.50
Relative Abundance
70 70 4.52
4.56
60 60 3.06
3.13
50 1.80 50 3.05
3.42 5.93
5.96
1.82 2.40 5.74 5.77
3.43
40 1.78
2.82 40 2.24 4.16
3.78 4.17 5.80 1.59 1.60
1.62 2.71
2.42 2.26 4.21
30 1.84
2.55 30 2.98
0.80 0.82 1.04 1.13 0.15 1.65 2.28 3.28 5.57
0.16 0.66 5.02 5.03
20 0.34 0.49 1.47 1.97
3.95 4.94
4.97 5.38 20 0.34
0.81
0.99
1.41 3.95 4.99
5.59
6.00
0.30 4.60 5.15 5.85 1.02 2.02 4.74 5.23 6.02
10 5.99 6.346.87 6.85 6.09 6.48
6.46
10 1.99 6.50
6.60
0 0
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0
Minutes Minutes
Figure 4. Typical BPC profile of HeLa protein digests for data sequentially collected on column 1 and 2 using the tandem low-flow
LC-MS method
MS/MS
800 A 14000 B 12948
12000
PSMs 10399
600 10000 Peptide
groups 6689
8000 5587
Counts
400 6129
6000 5374
Protein
4000 groups
200 1175
2000 972
0 0
0.5 1 1.5 2 2.5 3 3.5 4 4.5 5 5.5 6 6.5 7
Single analytical column setup Tandem setup
PSMs - RT [min]
Figure 5. Distribution of PSMs from 200 ng HeLa cells protein digest eluted using the 7-minute tandem LC-MS method (A) and a
comparison of the DDA results from an 8-minute single analytical column LC high-throughput pre-concentration method with the
7-minute tandem pre-concentration method
The proteomics data collected using the tandem LC-MS post-column flow guiding using the switching valve. This
method were compared with data collected using the effect is virtually non-existent using a Thermo Scientific™
recently described high-throughput pre-concentration EASY-Spray™ LC column used in the single analytical
low-flow method employing a single analytical column.2 column setup.2 Nevertheless, our results clearly show
The wider elution window afforded by the tandem low- that the benefits of the increased peptide elution window
flow LC resulted in more MS/MS events and subsequently afforded by the tandem LC method (which covers almost
more protein group identifications when compared to the entire 7 minutes of runtime) vs. 6 out of 7 minutes
the single analytical column-based method (Figure 5B). for the single analytical column method outweighs any
This was despite the fact that the peak width was up negative impact caused by the lower chromatographic
to twice the width for the tandem method. The wider resolution achievable with the tandem LC setup for very
peak width observed for the tandem LC method can be short gradients.
attributed to post-column dispersion that happens during
8
4. Conclusions 5. References
1. Zhang, X.; Gajadhar, A.; Sun, X.; Baynham M. A highly sensitive and robust 150 µm
The versatility of the UltiMate 3000 RSLCnano system column to enable high-throughput proteomics. Thermo Scientific Application Note
provides the flexibility necessary to meet the individual 21744; January 2018.
demands of modern proteomics research and beyond. 2. Boychenko, A.; Pynn, C.; van den Berg, B.; Arrey, T, N., Baynham, M.; Decrop, R.;
Ruehl, M. High-throughput capillary-flow LC-MS proteomics with maximum MS
The system shows superior performance both in terms utilization. Thermo Scientific Technical Note 72777; August 2018.
of retention time precision when employed for long 3. Thermo Scientific UltiMate 3000 RSLCnano System – Versatility and Performance –
shallow gradient elution experiments typical of deep- The Ultimate Solution for All Separation Workflows, BR-71898; 2016.
4. Thermo Scientific UltiMate 3000 RSLCnano Standard Applications Guide v3.0; 2018
dive proteomic analysis6 as well as short gradient
5. The Complete and Easy Guide to Configuring your Thermo Scientific Nano LC for Mass
high-throughput methodology, making maximum use Spectrometric Analysis.
of MS time with the added benefit that both types of 6. Thermo Scientific Poster Note 64618. https://assets.thermofisher.com/TFS-Assets/
experiments can be performed without any changes CMD/posters/PN-64618-LC-MS-Human-Proteomes-Label-Free-Quantitative-Profiling-
HUPO2015-PN64618-EN.pdf
required in either LC hardware or capillary fluidics.
©2018 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its
subsidiaries unless otherwise specified. ProFlow is a trademark of Restek Corp. SEQUEST is a registered trademark of the
University of Washington. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It
is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others.
Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local
sales representatives for details. TN72827-EN 0918S