You are on page 1of 1

Solenoid Data/+5V/5V Ret.

SDM/CDM
SOLENOID CONTROL BLOCK DIAGRAM
J51 3-1 WC4 Vacuum Supply (12" Hg)
J52 3-2 WC4 Pressure (12 psi)
J53 3-3 RBC Diluent Syringe Supply
J54 3-4 CS Probe Diluent Flush (NC) CDM Connectors
J56 3-6 CS Vent Trap Drain (WC3) J3 J1
J57 3-7 CS Probe Drain (WC4) +24V UR
J9
J58 3-8 CS Vent Trap Vent 5 1
J61 4-1 NOC Sample Staging From 4 +12V UR 2
J3 P. S. 3 UR RET 3 To
CLK MDM
J1 J6 SDM5 1 5 SDMA
Y 6
J68 J2 DC J1 5-1 HGB Flow Cell Vent J6
ON/OFF J6 Valve 7
FCM J67 J1 Motor J9 J2 5-2 Staging Pump Input On CLK A
CLK CW/CCW J3 5-3 WBC Cup Vent 1
CLK J1 Hi CLK A
J2 J4 5-4 RBC/PLT Sample Staging 2 On CLK B J2
CLK CLK J5 5-5 WBC Sample Staging 3 Hi CLK B
J4 J2 CDM2 4
J19 J6 5-6 Optical Flow Cell Output From 1
FCM 5 On CLK C
J1 J10 J7 5-7 Optical Flow Cell Drain Hi CLK C 2
J2 J8 5-8 Spare 6 3 To
Peripheral Bus J1 1-1 Shear Valve Diluent Flush On CLK D
7 Hi CLK D 5 SDMB
J9 J2 1-2 Sample Aspiration Vacuum 8 Reset 6
J3 1-3 Shear Valve Drain (WC4)
9 7
J4 1-4 Open Sample Probe Drain (WC4) SDM6
J16 J5 1-5 Open Sample Probe Diluent Flush (NC) J4
1-6 Spare J6 J1 6-1 Sheath Reservoir Vacuum (12" Hg)
CPU/DCM J6 J9 J2 6-2 Diluent Reservoir Vacuum (12" Hg)
J10 J7 1-7 WC2 Vacuum Supply (4" Hg) 1
J17 J1 J3 6-3 Diluent Reservoir Pressure (5 psi) 2
J8 1-8 WC2 Pressure Supply (12 psi)
J4 6-4 Sheath Reservoir Pressure (9 psi) To
3
SDM1 J5 6-5 Sheath Reservoir Output (NC) 5 SDMC
SDM2 J6 6-6 Diluent Reservoir Output (NC)
6
J10 J7 6-7 WC1 Pressure (12 psi)
7
J1 2-1 RBC Mixing Chamber Drain J8 6-8 WC1 Vent
J9 J2 2-2 Sample Injection Syringe Sheath Supply J5
J3 2-3 Main WBC Lyse Supply (NC)
J4 2-4 HGB Lyse Syringe Output 1
SDM9 2
J5 2-5 WBC Lyse Syringe Output
J6 2-6 Diluent Syringe HGB Dilution J1 9-1 Optical Flow Cell Sheath Flow 3 To
J10 J7 2-7 Diluent Syringe RBC Dilution J9 J2 9-2 WBC Mixing Chamber Drain (WC3) 5 SDMD
J8 2-8 Main HGB Lyse Supply (NC) J3 9-3 HGB Flow Cell Drain (WC3) 6
J4 9-4 HGB Flow Cell Diluent Flush 7
J5 9-5 RBC Mixing Chamber Diluent Flush
J6 9-6 WBC Mixing Chamber Diluent Flush
J10 J7 9-7 WC3 Vacuum (12" Hg)
J8 9-8 WC3 Pressure (12 psi)

VPM
FCM Connectors SDM Connectors & Driver Outputs
1 12 psi Accumulator Vent J9
J17 J1 J1 J3
PD0 +24VUR +12VUR
J2 2 9 psi Pressure Supply On CLK 1 PD0 1
3 5 psi Pressure Supply 1 1 PD1
J3 Hi CLK 1 PD1 2
2 2 PD2
J4 4 12" Hg Accumulator Vent On CLK 2 PD2 3
5 Vacuum #2 Vacuum Supply 3 3 PD3
J5 Hi CLK 2 PD3 4
4 4 PD4
On CLK 3 To PD4 5 To 5
5 PD5 Hi/Lo
Hi CLK 3
SHM1 Aspiration Tower SHM Solenoid Connector 6 CDM1 PD5 6 SDMs 6
PD6
Driver
+24V On CLK 4 7 PD6 7 7
RS-485 Bus GS2 Stop PD7 TP1
J16 J1 Hi CLK 4
8 PD7 8 8
+5V
J2 Door Lock Reset +5V 9
J1-J8 9 9
J3 Spare 10 10
J4 Spare 15
1
J5 Spare
J6 Spare 2 J1-J8
AGND DGND DGND
J7 Spare 1
J8 Spare J2 J4 J10 Solenoid Drive
On/Off On/Off 2
Driver On CLK 5 +24VUR +24VUR TP2 Driver
1 1 1
SHM2 Sample Loader Hi CLK 5 +12VUR
2 +12VUR 2 2
On CLK 6 UR RET UR RET
RS-485 Bus V1 Rack Advance 3
J16 J1 Hi CLK 6
3 To On CLK 9
3 To
5 On CLK
J2 V2 Load Side Cross Transfer 4 CDM2 5 SDM9 Hi CLK
J3 V3 Mixer Lift Reset Hi CLK 9 6
9 6 Reset
J4 V4 Mixer Bladders 15 Reset 7
7
J5 V5 Unload Side Cross Transfer
J6 Spare
J7 Spare AGND
J8 Spare
H3_9052b