Professional Documents
Culture Documents
AE600N Trouble Display
AE600N Trouble Display
1. Check whether all indications are defective. If only parts related to parameters are defective, there
may be a problem with RAM. For example, measurement item and result in test result are parameter
items.
Parameters
2. If all indications are defective, check whether the radiating fan works. If the radiating fan
works, there may be a problem with fuse or Power Supply 2, which is smaller than Power
Supply 1.
3. If the radiating fan does not work, check the function of ASP and Keyboard. Press ASP button to
check its function. If it works, it makes a beep sound.
If ASP works but Keyboard does not work, check Power Supply 1.
If both ASP and Keyboard do not work, check CUP Board.
If both ASP and Keyboard works, check Power & Drive Board.
All indications
are defective
Check
Check whether the Radiating fan woks
Fuse
radiating fan works
Power Supply 2
Radiating fan
Check
ASP Working