You are on page 1of 1

Periodic Test for GammaTRACER XL2 Probes

Location: Sublocation:

Shortname: Site-ID. Coordinate (Longitude):

Contact person: Telephone: Coordinate (Latitude):

Probe type: Serial number: Test source used:


GammaTRACER XL2 GSM/SMS

Performing the test:


1. Mount the GammaCHECK at the GammaTRACER and adjust the position.
2. Connect the notebook to the GammaTRACER using RS-232 cable (or use IR adapter).
3. Start the notebook, use DataEXPERT to start the communication with the
GammaTRACER (communication module for GammaTRACER).
4. Switch Standard measurement cycle to 1 min and activate Maintenance Mode.
5. Mount the test source at the GammaCHECK.
6. Wait for the number of measurements sufficient to get the accuracy achieved (about 5
minutes will be sufficient for 1 sigma).
7. Dismount the test source from the GammaCHECK.
8. Reconfigure Standard measurement cycle to standard value and deactivate
Maintenance Mode.
9. Check date and time of the probe.
10.Remove notebook and GammaCHECK. Finished.

Test protocol:
• Date of check:
• Optical check of the GammaTRACER: ok not ok
• GammaCHECK mounted and adjusted correctly: yes
• Measurement cycle = 1 min and Maintenance Mode active: yes
• Test source mounted: yes
• Start time of irradiation: ____________
• End time of irradiation ____________
• Test source removed yes
• Measurement cycle restored / Maintenance Mode deactivated yes
• Date and time of GammaTRACER are correct ? yes no
Remarks: ...........................................................................................................................................
................................................................................................................................................................

Sign of testing person: ...................................

You might also like