You are on page 1of 1

PROTEUS XR/a

GE MEDICAL SYSTEMS Operator Manual


REV 22 DIRECTION 2259724-100
mode) or kV, mA, s (in 3 point mode) and
exposure parameter: include: kV, mAs,
mA, s , filmsize and SID
4. Press display button The console will redisplay the last
exposure parameter: kV, mAs, mA, s for
15s, and press any button will return to
previous interface. If no exposure has
been done: it will display no exposure.

5-3-5 Console Indicator


1. System error Led If there is any system error, this Led will
be on.
2. Thermal error Led If tube temperature is over limitation, this
Led will be on.
3. Error message When any error occurs, console will
display some error message and error
code: technique error, parameter error,
AEC error, rotor thermal error, inverter
thermal error, tank thermal error, Door
Open in exposure.
4. PREP Led Press and hold PREP button, the PREP
lamp will be on.
5. Exposure Led During X-ray emission, this lamp will be
on.
6. Buzzer During X-ray emission, it will be on.
7. Turn auto/manual switch Manual lamp on or off
8. Table in PBL: when table in Exposure ready is on, or exposure holder
center, SID in detent, Cassette is on
in place (if have tray selected)
9. WALL in PBL: when wall SID Exposure ready is on, or exposure holder
in detent, Cassette in place is on
10. TOMO in PBL: when table in Exposure ready is on, or exposure holder
center, SID in detent Cassette is on
in place (if have tray selected)
11. IN manual Exposure ready is on, exposure holder is
off
5-3-6 APR
1.Press any APR button on main Enter the sub-APR interface and the
screen parameter display will not change
2.Press home in any sub-APR Go back to the upper interface
interface
3.Press sub-APR button The button will be selected and the
parameter displayed will change
accordingly.
4.Insert the APR disk, then press Enter main screen edit interface, in this
procedure edit button in main screen, the parameter or technique can’t
screen. be changed or selected.
5.Press name cat button and Enter the name edit interface and edit the
then press any APR button APR name (only in English)
6.Press Save/Retrieve button Enter save & retrieve backup interface
7.In save & retrieve interface: Save the APR name to the floppy disk or
press save button the APR&AEC USB disk

8.In save & retrieve interface: If the floppy disk or the APR&AEC USB
press retrieve button disk has the relative file, it restores the file

5-17

You might also like