You are on page 1of 11

E.

CAM
Floor Leveling Procedure
NM

43 87 929, Rev. 05
Siemens Medical Systems, Inc. has prepared this document for use by its personnel and
licensees. Siemens reserves the right to change any products described in this document
as well as information included herein without prior notice. Although the information
presented in this document has been thoroughly tested and reviewed and is considered
reliable, this document does not convey any license or warranty beyond the terms and
conditions set forth in the written contracts between Siemens and its customer.

Any software described herein is licensed from and is a trade secret of Siemens Medical
Systems, Inc.

Siemens Medical Systems, Inc. reserves the right to make any content changes to this
document at any time.

The contents of this document might be affected by one or more Service Document
Updates (SDUs). Contact Service Technical Writing for a listing of the SDUs available
for this document.

E.CAM is a trademark of Siemens Medical Systems, Inc.

© 2000 Siemens Medical Systems, Inc. - Nuclear Medicine Group


2501 North Barrington Road,
Hoffman Estates, IL 60195-5203

Telephone: (847)304-7700 Telex: 687-2026 SIEGAM UW


Telefax: (847)304-7701

All rights reserved. No part of this document may be reproduced, in any form or by any
means, without direct permission in writing from Siemens Medical Systems, Inc.
SMS-NMG E.CAM Floor Leveling Procedure

Revision History

Rev Date By ECO Action

01 28 Oct 96 Gary Kershner 541904 Initial Release

02 06 Jun 97 Gary Kershner 542839 Rounded Foam Frame Corners.


Added Template.

03 20 Jan 98 T. E. Moi 5434476 Removed dimensions from Ffigure


1. Corrected minor errors.

04 13 Oct. 99 T. Brellinthin 544944 Changed Epoxy & Hardener

05 15 Sep 00 J. Healey 546158 Added Duet Floor Plate picture


and dimensions.

43 87 929, Rev. 05 i
E.CAM Floor Leveling Procedure SMS-NMG

ii 43 87 929, Rev. 05
SMS-NMG E.CAM Floor Leveling Procedure

1.1 Use Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1


1.2 Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.3 Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.4 Preparing Floor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.5 Determine Amount of Epoxy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.6 Mixing Epoxy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.7 Completing Floor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

43 87 929, Rev. 05 iii


E.CAM Floor Leveling Procedure SMS-NMG

iv 43 87 929, Rev. 05
SMS-NMG E.CAM Floor Leveling Procedure

1 E.CAM Floor Leveling Kit


This procedure takes approximately 2 hours to complete.

1.1 Use Conditions


Epoxy and hardener must be above 65°F before use. May have to
store in room for 24-48 hrs. if temperature is lower.
Floor must be above 65°F. Epoxy will not harden below 65°F
(18.3.°C)
Entire contents of hardener must be added and thoroughly mixed
before pouring.

1.2 Precautions
– Epoxy resins may sensitize certain personnel.
– Wear protective clothing, plastic gloves and chemical splash
goggles.
– Avoid inhaling fumes.
– Wash skin and clothes with soap and water only.
– If eye contact occurs, flush immediately with clean water
and consult a physician.
– Seller makes no warranty and user assumes all risks with
respect to the use of this product.
– Allow 8 hours for curing.
– Suggested to pour epoxy floor late in afternoon when most
department personnel have left. This also allows over night
for curing.

1.3 Material
43 87 911 E.CAM Floor Leveling Kit contents:

Quantity Part Number Description

1 43 87 929 E.CAM Floor Leveling Procedure

1 35 60 799 Floor Plate Template

1 59 70 897 Epoxy and Hardener

1 43 87 903 5 Pieces Foam Frame

1 N/A Automobile Paste Wax


1 52 22 950 Pkg Tube

43 87 929, Rev. 05 1
E.CAM Floor Leveling Procedure SMS-NMG

Quantity Part Number Description

1 29 71 765 3/4” Glass Reinforced Tape

1pr N/A Rubber Gloves

Tools needed:

Quantity Part Number Description

1 N/A Drill (at least 1/2” drill, not battery


operated)

1 N/A Rasp or Rough Flat File


1 N/A Pencil

1.4 Preparing Floor


1) Use planning guide to determine exact gantry location.
2) Seal any holes or large cracks in floor with the tape sup-
plied.
3) Cut out floor plate template (35 60 799). Be sure to cut out
rear patient bed template and tape to main floor plate tem-
plate (Figure 1)

Note
Note
Floor plate template is designed to be used with standard and
extended base plates. See fig. 2

Tape

Cut
Cut Rear Patient Bed template Foam
Foam

Figure 1 Foam Layout

2 43 87 929, Rev. 05
SMS-NMG E.CAM Floor Leveling Procedure

167(65.75")
167(65.75”)

36.8(14.5") 93.4(36.75") 36.8(14.5")


Extended Base
79.4(31.25")

71.8(28.25")

Standard Base
95.6(37.65")
Floor Plate under Gantry

23.9(9.4")

57.8(22.75") 85.1(33.4")
24.1(9.5")

Figure 2 Floor Plate


4) Tape floor plate template to the floor at gantry location.
5) Layout foam frame on outline. Use serrated or razor knife to
cut foam. Cut foam so that all pieces end at corners. For
rounded corners, do not cut foam all the way through. In
this way the foam partially cut will more easily form to the
curve (Figure 1).
6) Peel white tape from foam and affix to floor on outline with
brown paper tape facing in towards where the floor plate
will be situated.
7) Remove template from inside foam.
8) Seal all corners (foam junctions) with tape on inside of
frame (Figure 1).

43 87 929, Rev. 05 3
E.CAM Floor Leveling Procedure SMS-NMG

9) Apply wax with finger to floor/foam joint to seal (Figure 3)


.
Wax Seal (apply
with finger) Foam

Floor
Figure 3 Wax Seal
10) Apply generous wax coating to entire floor area inside foam
frame.

1.5 Determine Amount of Epoxy


To prevent the E.CAM from being too high above floor in relation
to the Patient Bed and Collimator Servers, only enough epoxy to
level the floor beneath the floor plate is desired. On a mostly level
floor pour only slightly more than one-half (1/2) the contents of
the epoxy mixture. For more uneven floors, pour more. In most
cases two-thirds (2/3) of the epoxy is enough.

1.6 Mixing Epoxy


1) Open 50lb. container of epoxy.
2) Using drill-stirrer combination (conventional drill with mix-
ing paddle), pre-mix contents of 50lb. can for approxi-
mately 2 minutes.
a) Stir tool must reach bottom of bucket.
b) Move stir tool around perimeter of bucket and up and
down.
c) Do not move stir tool up so high as to cause air to be
mixed with liquid grout.
3) Open hardener and carefully add to epoxy. POUR ENTIRE
CONTENTS OF HARDENER INTO GROUT. Avoid
splashing.
4) With drill-stirrer, mix epoxy and hardener for approxi-
mately 3 minutes or until thoroughly mixed.
a) Stir tool must reach bottom of bucket.
b) Move stir tool around perimeter of bucket and up and
down.
c) Do not move stir tool up so high as to cause air to be
mixed with liquid grout.
5) Pour contents into frame area where floor seems to be the
highest. Pour in one spot only. Do not move bucket
around while pouring.

4 43 87 929, Rev. 05
SMS-NMG E.CAM Floor Leveling Procedure

1.7 Completing Floor


1) After curing (approximately 8 hours) remove foam from
perimeter of poured floor.
2) Discard foam.
3) Use a rasp to put a 1/8” chamfer on edges of floor. Smooth
epoxy of any rough spots caused by bubbles with rasp. End.

43 87 929, Rev. 05 5