Professional Documents
Culture Documents
Kemppi Oy
CONTENTS
Kemppi Oy
Validation instructions Version 1.0 19.7.2019
Before removing any covers or commencing any testing or measurement disconnect the power source
from the mains voltage
Dangerous DC voltage may still exist after the removal of the input voltage. Machine discharges the
voltage while it is turned off, but it is always better to ensure this by measuring the voltage
Digital multimeters (later DMM) may give different values depending the specifications they have. For
example diode measuring values may vary between the DMM models. In this manual Fluke 179 DMM is
used.
The device may be repaired only by a person legally authorized to perform electric work.
2. GENERAL
This manual holds the instructions to do a successful validation. Validation must be done at least once in
a year for standard grade and twice for precision grade. Customer may demand more frequent validation
process.
Validation must be done after the repair work and new label must be put in to the machine.
2.1. Terms
Calibration: Measuring and recording the errors that measuring instruments have.
Validation: Accepting or discarding the machine according to measurements and determined errors.
3. MEASURING INSTRUMENTS
To execute calibration several equipment is needed. Measuring instruments Following values must be
measured: voltage, current, wire feed speed gas flow and water flow.
For calibration measurement load bank is needed as well. There is no big difference if the load is manual
or automatic. Only requirements are that it must be adjustable and accept welding voltage and current
range up to 500 amps.
Calibration must be done with annually calibrated equipment that are at least 2 times more accurate than
equipment under validation.
1
Kemppi Oy
Validation instructions Version 1.0 19.7.2019
Measuring machines output values and comparing them to display values must be done in certain way.
Current to voltage (MMA and TIG) or voltage to current (MIG/MAG) has relation that is called nominal
curve. Idea is that curve is standardized and results can always be compared to earlier ones during the
machine life cycle.
Notice that nominal curve minimum point is limited higher that many modern welding machines can
reach. This is the reason why examples are giving suggested minimum point for each welding process.
∗ 0,04 10
So if using 300A TIG machine current and voltage values should be following in maximum.
Minimum point is determined to 15A in practice, because going below that point voltage stays almost
constant but current goes to zero according to formula. In practice machine can go lower values. Good
compromise it to select 15A minimum current and voltage will be 10,6V.
Maximum setting can be done according to machine maximum output current value.
∗ 0,04 20
Also in MMA mode 15A is good current value for minimum. Voltage will be 20,6V:
2
Kemppi Oy
Validation instructions Version 1.0 19.7.2019
MIG/MAG mode is CV type and instead of adjusting current voltage is adjusted. But still power sources
are ranged by current it is better to determine output values in function of current:
∗ 0,05 14
Maximum setting can be determined normally according to machine output voltage and current. For
example 300A machine output voltage will be 29V:
Minimum setting is better to select again above machine’s real minimum value. Nominal curve formula
cannot give values below 14V but modern inverters can actually weld below that point. 20A is good value
to start with an voltage will be 15V:
The error limits are determined for standard grade (appendix 1). If precision grade is required see
appendix 2.
Following examples describes standard grade power source and wire feeder validation.
- Set value: The value that is set by welding machine user interface, i.e. machine is asking some
voltage (CV) or current (CC) output
- Display value: The display reading that is shown
- Measured value (called sometimes true value or real value): The value that measuring instrument
measures in the welding machine output. This is the value where set value and display value are
compared.
Both of these have different error limits and they can be determined by certain rules.
According to EN 50504 only one is required to be recorded but it is suggested to record both.
3
Kemppi Oy
Validation instructions Version 1.0 19.7.2019
When machine is set to certain current or voltage set value (only one can be selected and another is
determined by set load) machine’s real output value is compared to this value.
• In the range of 0-25% of voltage and current set value error can be ±2,5% of maximum setting
• In the range of 25-100% of voltage and current set value error can be ±10% of true value
(measured value)
Example 1:
Example 2:
Displays have different error limits depending if they are voltage or current meters.
- Current display may have maximum error of ±2,5% of maximum welding current
- Voltage display may have maximum error of ±2,5% of no load voltage (OCV)
Example 3:
For both set value and display value error range is same. According to EN 50504 only one is required to
be recorded but it is suggested to record both.
4
Kemppi Oy
Validation instructions Version 1.0 19.7.2019
Example 4:
6. VALIDATION PROCEDURE
6.1. Preparations
5
Kemppi Oy
Validation instructions Version 1.0 19.7.2019
1. Check that water is flowing through the system and record PASS or FAILED
2. Check that gas is flowing through the system and record PASS or FAILED
6.5. Certificate
6.6. Labels
Failed label
- Failed
- Date
- Inspectors name or ID
- Brand, model and serial number of the welding machine
6
Kemppi Oy
Validation instructions Version 1.0 19.7.2019
APPENDIX 1
7
Kemppi Oy
Validation instructions Version 1.0 19.7.2019
APPENDIX 2
8
Kemppi Oy